Organic Carbon, Total

Similar documents
Phosphorus, Reactive (Orthophosphate)

Chlorine, Total. DPD Method 1 Method to 5.00 mg/l Cl 2 Test 'N Tube Vials. Test preparation. Instrument-specific information

USEPA 1 FerroVer Method 2 Method to 3.00 mg/l Fe Powder Pillows or AccuVac Ampuls

Cadmium Reduction Method Method to 30.0 mg/l NO 3 N (HR) Powder Pillows or AccuVac Ampuls

LZV585 (B) The 1-cm path is aligned with the arrow on the adapter

DR/4000 PROCEDURE. CHLORINE, Free

FerroVer Method 1 Method to 3.0, 1.0 to 30.0 and 10.0 to mg/l Fe Powder Pillows

Quaternary Ammonium Compounds DOC

Esterification Method 1 Method to 2800 mg/l (as acetic acid) Reagent Solution

Method 8155 NITROGEN, AMMONIA (0 to 0.50 mg/l NH 3 -N) For water, wastewater, seawater

Color, True and Apparent

Dimethylglyoxime Method Method to 6.0 mg/l Ni TNTplus 856

Color, True and Apparent

Nitrogen, Ammonia. Known Addition ISE Method 1 Method Minimum of 0.8 mg/l NH 3 N. Ammonia ISE. Test preparation. Instrument-specific table

Scope and application: For testing residual corrosion inhibitors (oxygen scavengers) in boiler feed water or condensate

Thermo Scientific Orion AQUAfast Powder and Tablet Reagent Chemistry User Guide

Direct ISE Method Method to 1000 mg/l Na + Sodium ISE

Direct Measurement Method Method to 20.0 mg/l (or 0 to 200% saturation) O 2 Clark-type Amperometric Sensor

Solids, Volatile Dissolved and Fixed Dissolved

Measurement by Ion Selective Electrodes

USEPA Electrode Method Method 8156

Lead Testing and On Site Calibration for Water Testing Detection Range: 2 100ppb

Standard Operating Procedure for Total Kjeldahl Nitrogen (Lachat Method)

Texas Commission on Environmental Quality

Phosphorus, colorimetry, phosphomolybdate, automated-segmented flow

An Introduction to Standards and Quality Control for the Laboratory Barbara Martin

Appendix D lists the Field Services Standard Operating Procedures. Appendix E lists the Biological Monitoring Standard Operating Procedures.

Reagents also compatible in Hach

NANYANG TECHNOLOGICAL UNIVERSITY. School of Civil and Environmental Engineering. CV2701: Laboratory 2A. Laboratory Manual. For

Chapter 16: Tests for ions and gases

Absorbance Spectrophotometry: Analysis of FD&C Red Food Dye #40

Mixtures and Pure Substances

COMMON LABORATORY APPARATUS

Experiment 12- Classification of Matter Experiment

EXPERIMENT 5. Molecular Absorption Spectroscopy: Determination of Iron With 1,10-Phenanthroline

Preparation of frequently used solutions

EXPERIMENT 7. Identifying a Substance by Acid-Base Titration

Specification and Certificate of Analysis

PECTINS. SYNONYMS INS No. 440 DEFINITION DESCRIPTION. FUNCTIONAL USES Gelling agent, thickener, stabilizer, emulsifier CHARACTERISTICS

THE ACTIVITY OF LACTASE

Neutralizing an Acid and a Base

Determining the Quantity of Iron in a Vitamin Tablet. Evaluation copy

FT FIELD MEASUREMENT OF RESIDUAL CHLORINE

Experiment 8 - Double Displacement Reactions

SPIKE RECOVERY AND DETERMINING A METHOD DETECTION LIMIT Pamela Doolittle, University of Wisconsin Madison, pssemrad@wisc.edu 2014

Calcium Analysis by EDTA Titration

OBJECTIVE: To measure Chemical Oxygen Demand. BACKGROUND AND PRINCIPLE: SELECTION OF METHODS. Page 1

5.0 EXPERIMENT ON DETERMINATION OF TOTAL HARDNESS

Acid Base Titrations

Solubility Product Constants

A Volumetric Analysis (Redox Titration) of Hypochlorite in Bleach

OXIDATION-REDUCTION TITRATIONS-Permanganometry

LACHAT METHOD NUMBER D Rev 1, 21 March 2011 SCOPE AND APPLICATION Approximately g CN/L g CN/L. 09e

Determination of the Amount of Acid Neutralized by an Antacid Tablet Using Back Titration

Total Aerobic Bacteria, Yeasts and Molds

HOMEWORK 4A. Definitions. Oxidation-Reduction Reactions. Questions

Standard methods in water analysis

XI. Methods of Analysis DETERMINATION OF POTASSIUM CARBONATE CALCULATIONS REAGENTS PROCEDURE

Chem 131A: Absorbance of Riboflavin

4.0 EXPERIMENT ON DETERMINATION OF CHLORIDES

Ion Selective Electrodes

Experiment 16-Acids, Bases and ph

ph: Measurement and Uses

To determine the equivalence points of two titrations from plots of ph versus ml of titrant added.

DRINKING WATER - LAB EXPERIMENTS. Coagulation and flocculation LAB EXPERIMENTS. Jartest

MASSACHUSETTS INSTITUTE OF TECHNOLOGY Department of Chemistry Laboratory Chemistry THE POTENTIOMETRIC TITRATION OF AN ACID MIXTURE 1

Physical and Chemical Properties and Changes

User Guide Dissolved Oxygen Probes

Decomposition. Composition

Canine creatine kinase MB isoenzyme (CK-MB)ELISA Kit

1.4 This method is capable of measuring DO in the range of 0.20 to 20 mg/l.

NITRIC OXIDE and NITROGEN DIOXIDE 6014

III. Chemical Kinetics

MOISTURE (Karl Fischer, Buffered)

Volumetric Analysis. Lecture 5 Experiment 9 in Beran page 109 Prelab = Page 115

Santa Monica College Chemistry 11

STANDARDIZATION OF A SODIUM HYDROXIDE SOLUTION EXPERIMENT 14

METHOD 3510C SEPARATORY FUNNEL LIQUID-LIQUID EXTRACTION

Environmental Water Testing: Surface Water, Groundwater, Hard Water, Wastewater, & Seawater

CONFIRMATION OF ZOLPIDEM BY LIQUID CHROMATOGRAPHY MASS SPECTROMETRY

Customer application report

Dissolving of sodium hydroxide generates heat. Take care in handling the dilution container.

Measuring ph in Water or CaCl2 Using a ph Meter. Compiled by Darren Murray, June 30, 2011 TABLE OF CONTENTS OVERVIEW 3

ACID-BASE TITRATIONS: DETERMINATION OF CARBONATE BY TITRATION WITH HYDROCHLORIC ACID BACKGROUND

Human Free Testosterone(F-TESTO) ELISA Kit

Odor (Threshold Odor, Consistent Series)

CHM 130LL: ph, Buffers, and Indicators

A Visual Method for the Detection of Arsenic 0 to 500 µg/l By Dan Kroll R&D Chemist. Hach Co., Loveland Colorado

Rat Creatine Kinase MB isoenzyme,ck-mb ELISA Kit

Lab #13: Qualitative Analysis of Cations and Anions

BRIEFING Plastic Packaging Systems for Pharmaceutical Use.

To see how this data can be used, follow the titration of hydrofluoric acid against sodium hydroxide below. HF (aq) + NaOH (aq) H2O (l) + NaF (aq)

FIVE-DAY BIOCHEMICAL 7.0 OXYGEN DEMAND

Hot water extractable acidity or alkalinity of paper (Reaffirmation of T 428 om-08) (No changes from Draft 1)

OBJECTIVES: Visitors learn what an antioxidant is and how it behaves. They also learn how to test for the presence of vitamin C..

Recovery of Elemental Copper from Copper (II) Nitrate

Agarose Gel Electrophoresis with Food Color- Teacher Guide

Determination of Citric Acid in Powdered Drink Mixes

Topic 4 National Chemistry Summary Notes. Formulae, Equations, Balancing Equations and The Mole

Bluelab Soil ph Meter Instruction Manual

Transcription:

Method 10129 Direct Method 1 Scope and Application: For water, drinking water, and wastewater 1 U.S. Patent 6,368,870 LR (0.3 to 20.0 mg/l C) Test Preparation Before starting the test: A reagent blank is required for each series of samples. Collect the following items: Quantity Total Organic Carbon Direct Method Low Range Test N Tube Reagent Set 1 Cylinder, graduated, 10-mL 1 DRB200 Reactor 1 ph Paper 1 Flask, Erlenmeyer, 50-mL 1 Magnetic Stirrer 1 Pipet, TenSette, 0.1 to 1.0 ml plus tips 1 Pipet, TenSette, 1.0 to 10.0 ml plus tips 1 Stir Bar, magnetic 1 Test Tube Rack 1 3 Water, organic-free 3.0 ml Wipes, disposable, Kimwipes 1 Note: Reorder information for consumables and replacement items is on page 6. Direct Method Method 10129 HRS MIN SEC 1. Turn on the DRB 200 reactor. Select the TOC program. 2. Use a graduated cylinder to add 10 ml of sample to a 50-mL Erlenmeyer flask that contains a stir bar. 3. Add 0.4 ml of Buffer Solution, ph 2.0. Use ph paper to make sure the sample ph is 2. 4. Place the flask on a stir plate and stir at a moderate speed for 10 minutes. OrganicCarbonTotal_10129_LR.fm Page 1 of 6

5. Label two Low Range Acid Digestion vials sample and reagent blank. 6. Use a funnel to add the contents of one TOC Persulfate Powder Pillow to each Acid Digestion vial (colorless liquid). 7. Use a TenSette Pipet to add 3.0 ml of organic-free water to the reagent blank vial and 3.0 ml of prepared sample to the sample vial. Swirl to mix. 8. Rinse two blue Low Range Indicator Ampules with deionized water and wipe them with a soft, lint-free wipe. Do not touch the ampules sides after wiping. Pick them up by the top. 9. Lower one unopened ampule into each Acid Digestion vial. When the score mark on the ampule is level with the top of the Acid Digestion vial, snap the top off the ampule and allow it to drop into the Acid Digestion vial. Do not invert or tilt the vial after inserting the ampule. 10. Cap the vial assemblies tightly and insert them in the COD reactor for 2 hours at 103 105 C. 11. Carefully remove the vial assemblies from the reactor. Place them in a test tube rack. Allow the vials to cool for one hour for accurate results. The liquid in the reagent blank vial should be dark blue. 12. Select the test. Page 2 of 6 OrganicCarbonTotal_10129_LR.fm

13. Wipe the reagent blank with a damp towel, followed by a dry one, to remove fingerprints or other marks. 14. Insert the reagent blank vial assembly in the 16-mm round cell holder. 15. Press ZERO. The display will show: 0.0 mg/l C. 16. Wipe the sample vial assembly with a damp towel, followed by a dry one, to remove fingerprints or other marks. 17. Insert the sample vial assembly in the 16-mm round cell holder. 18. Press READ. Results are in mg/l C. Interferences The following have been tested for interference and found not to interfere up to the indicated levels: Table 1 Interfering Substances and Levels Substance Aluminum Ammonia Nitrogen ASTM Wastewater Bromide Maximum Level Tested 1000 mg/l as N No effect 500 mg/l Br Bromine 25 mg/l Br 2 Calcium 2000 mg/l as CaCO 3 Chloride 500 mg/l Chlorine Cl 2 Chlorine Dioxide 6 mg/l ClO 2 Copper Cyanide Iodide CN 50 mg/l OrganicCarbonTotal_10129_LR.fm Page 3 of 6

Table 1 Interfering Substances and Levels (continued) Substance Iron (II) Iron (III) Maximum Level Tested Magnesium 2000 mg/l as CaCO 3 Manganese (VII) 1 mg/l Monochloramine 14 mg/l NH 2 Cl as Cl 2 Nitrite 500 mg/l NO 2 Ozone 2 mg/l O 3 Phosphate 3390 mg/l PO 3 4 Silica 100 mg/l SiO 2 Sulfate 5000 mg/l SO 2 4 Sulfide 20 mg/l S 2 Sulfite 50 mg/l SO 3 2 Zinc 5 mg/l Sampling and Storage Accuracy Check If the sample contains greater than 600 mg/l CaCO 3 alkalinity, lower the sample ph to less than 7 before testing by adding Sulfuric Acid Solution. Most sample turbidity is either dissolved during the digestion stage or settled during the cooling period. Sample turbidities up to 50 NTU have been tested without interference. Collect samples in clean glass bottles. Rinse the sample bottle several times with the sample to be collected. Fill the bottle completely full before capping. Test samples as soon as possible. Acid preservation is not recommended. Homogenize samples containing solids to assure representative samples. Standard Additions Method (Sample Spike) 1. After reading test results, leave the sample cell (unspiked sample) in the instrument. 2. Press OPTIONS>MORE. Press STANDARD ADDITIONS. A summary of the standard additions procedure will appear. 3. Press OK to accept the default values for standard concentration, sample volume, and spike volumes. Press EDIT to change these values. After values are accepted, the unspiked sample reading will appear in the top row. See the user manual for more information. 4. Prepare a 150-g/L C standard by transferring 15.00mL of 1000-mg/L C stock solution to a 100-mL Class A volumetric flask. Dilute to volume with organic-free water. Mix. 5. Use the TenSette Pipet to add 0.1, 0.2, and 0.3 ml of the 150-mg/L C standard to each of three Acid Digestion Vials. 6. Add the contents of one TOC Persulfate Powder Pillow to each vial. Page 4 of 6 OrganicCarbonTotal_10129_LR.fm

Method Performance 7. Add 3.0 ml of sample to each vial. Swirl to mix. Continue the test starting at step 8 on page 2 of this procedure. 8. Analyze each sample spike as described in the procedure above, starting with the 0.1 ml sample spike. Accept each standard additions reading by pressing READ. Each addition should reflect approximately 100% recovery. 9. After completing the sequence, press GRAPH to view the best-fit line through the standard additions data points, accounting for the matrix interferences. Press IDEAL LINE to view the relationship between the sample spikes and the Ideal Line of 100% recovery. Standard Solutions Method 1. Prepare a 1000-mg/L organic carbon stock standard by dissolving 2.1254 g dry primary standard Potassium Acid Phthalate in Organic-free Reagent Water and dilute to 1000 ml. This stock standard is stable for about 1 month at room temperature. Alternatively, open one ampule of TOC Standard Solution, 1000 mg/l C. Perform the TOC procedure as described above. 2. Prepare a 10.0-mg/L C standard by transferring 10.00 ml of the stock standard to a 1000-mL Class A volumetric flask. Dilute to volume using Organic-Free Reagent Water. Stopper and mix thoroughly. Prepare this standard fresh daily. 3. To adjust the calibration curve using the reading obtained with the standard solution, press OPTIONS>MORE on the current program menu. Press STANDARD ADJUST. 4. Press ON. Press ADJUST to accept the displayed concentration. If an alternate concentration is used, press the number in the box to enter the actual concentration, then press OK. Press ADJUST. Precision Standard: 10.0 mg/l C Program 95% Confidence Limits of Distribution 427 9.1 10.9 mg/l C Sensitivity Portion of Curve Abs Concentration Mid-range 0.010 0.2 mg/l C Summary of Method Note: To test for higher ranges of TOC use Method 100173 or Method 10128. The total organic carbon (TOC) is determined by first sparging the sample under slightly acidic conditions to remove the inorganic carbon. In the outside vial, organic carbon in the sample is digested by persulfate and acid to form carbon dioxide. During digestion, the carbon dioxide diffuses into a ph indicator reagent in the inner ampule. The adsorption of carbon dioxide into the indicator forms carbonic acid. Carbonic acid changes the ph of the indicator solution which, in turn, changes the color. The amount of color change is related to the original amount of carbon present in the sample. Test results are measured at 598 and 430 nm. OrganicCarbonTotal_10129_LR.fm Page 5 of 6

Consumables and Replacement Items Required Reagents Description Quantity/Test Unit Cat. No. Reagent Set, Total Organic Carbon Direct Method Low Range Test N Tube, includes: 50 vials 27603-45 Acid Digestion Solution Vials, High Range TOC 1 1 50/pkg Buffer Solution, Sulfate 0.4 ml 25 ml 452-33 Funnel, micro 1 each 25843-35 Indicator Ampule, Low Range TOC 1 1 10/pkg TOC Persulfate Powder Pillows 1 1 50/pkg ph Paper 1 5/pkg 391-33 Water, Organic-free 3.0 ml 500 ml 26415-49 1 Not sold separately. Required Apparatus Description Quantity/Test Unit Cat. No. Cylinder, graduated, 10-mL 1 each 508-38 DRB200 Reactor, 110 V, 15 x 16 mm 1 each LTV082.53.40001 DRB200 Reactor, 220 V, 15 x 16 mm 1 each LTV082.52.40001 Flask, Erlenmeyer, 50-mL 1 each 505-41 Magnetic Stirrer 1 each 28812-00 Pipet, TenSette, 0.1 to 1.0 ml 1 each 19700-01 Pipet Tips, for TenSette Pipet 19700-01 2 50/pkg 21856-96 Pipet, TenSette, 1.0 to 10.0 ml 1 each 19700-10 Pipet Tips, for TenSette Pipet 19700-10 2 50/pkg 21997-96 Stir Bar, magnetic 1 each 45315-00 Test Tube Rack 1 3 each 18641-00 Wipes, Disposable, Kimwipes 1 280/pkg 20970-00 Recommended Standards Description Unit Cat. No. Potassium Acid Phthalate 500 g 315-34 TOC Standard Solution (KHP Standard, 1000-mg/L C) 5/pkg 27915-05 Optional Reagents Description Cat. No. Sulfuric Acid Solution 2449-32 FOR TECHNICAL ASSISTANCE, PRICE INFORMATION AND ORDERING: In the U.S.A. Call toll-free 800-227-4224 Outside the U.S.A. Contact the HACH office or distributor serving you. On the Worldwide Web www.hach.com; E-mail techhelp@hach.com HACH COMPANY WORLD HEADQUARTERS Telephone: (970) 669-3050 FAX: (970) 669-2932 Hach Company, 2005. All rights reserved. Printed in Germany. Updated April 05 Edition 1