Quantification of Polysorbate Using the ACQUITY UPLC H-Class Bio System with ELS Detection

Similar documents
Method Development for Size-Exclusion Chromatography of Monoclonal Antibodies and Higher Order Aggregates

Analysis of Free Bromate Ions in Tap Water using an ACQUITY UPLC BEH Amide Column

Robert Birdsall, Eoin Cosgrave, Henry Shion, and Weibin Chen Waters Corporation, Milford, MA, USA APPLICATION BENEFITS INTRODUCTION WATERS SOLUTIONS

Analysis of Vegetable Oils by High Performance Liquid Chromatography

Using Natural Products Application Solution with UNIFI for the Identification of Chemical Ingredients of Green Tea Extract

Analysis of Polyphenols in Fruit Juices Using ACQUITY UPLC H-Class with UV and MS Detection

High-Throughput 3-D Chromatography Through Ion Exchange SPE

A Generic Kit-Based Approach for Quantifying Monoclonal Antibody Drugs Through Direct Digestion of Discovery Study Samples

Henry Shion, Robert Birdsall, Steve Cubbedge, and Weibin Chen Waters Corporation, Milford, MA, USA APPLICATION BENEFITS INTRODUCTION WATERS SOLUTIONS

Oasis HLB Cartridges and 96-Well Plates

How To Test For Contamination In Large Volume Water

How To Use An Acquity Qda Detector

Jennifer L. Simeone and Paul D. Rainville Waters Corporation, Milford, MA, USA A P P L I C AT ION B E N E F I T S INT RO DU C T ION

Fast, Reproducible LC-MS/MS Analysis of Dextromethorphan and Dextrorphan

Analysis of Fat-Soluble Vitamin Capsules using UltraPerformance Convergence Chromatography (UPC 2 )

The Use of Micro Flow LC Coupled to MS/MS in Veterinary Drug Residue Analysis

LC-MS/MS Method for the Determination of Docetaxel in Human Serum for Clinical Research

Overview. Purpose. Methods. Results

Extraction of Epinephrine, Norepinephrine and Dopamine from Human Plasma Using EVOLUTE EXPRESS WCX Prior to LC-MS/MS Analysis

Protein Separation Technology

Analysis of the Vitamin B Complex in Infant Formula Samples by LC-MS/MS

CONFIRMATION OF ZOLPIDEM BY LIQUID CHROMATOGRAPHY MASS SPECTROMETRY

UPLC-MS/MS Analysis of Aldosterone in Plasma for Clinical Research

1.1 This test method covers the qualitative and quantitative determination of the content of benzene and toluene in hydrocarbon wax.

Conversion of a USP Gas Chromatography Method to Convergence Chromatography: Analysis of Atropa Belladonna

In-Depth Qualitative Analysis of Complex Proteomic Samples Using High Quality MS/MS at Fast Acquisition Rates

The Automated SPE Assay of Fipronil and Its Metabolites from Bee Pollen.

Waters Corporation. Waters 2690/5 USER & TROUBLESHOOTING GUIDE

Guide to Reverse Phase SpinColumns Chromatography for Sample Prep

Introduction to the ACQUITY UPLC H-Class Amino Acid Analysis System

Thermo Scientific SOLAµ SPE Plates Technical Guide. Consistent excellence. for bioanalysis

Quantitative analysis of anabolic steroids in control samples from food-producing animals using a column-switching LC-HESI-MS/MS assay

Step-by-Step Analytical Methods Validation and Protocol in the Quality System Compliance Industry

Project 5: Scoville Heat Value of Foods HPLC Analysis of Capsaicinoids

High sensitivity assays using online SPE-LC-MS/MS -How low can you go? Mohammed Abrar Unilabs York Bioanalytical solutions, York, UK

Discovery of Pesticide Protomers Using Routine Ion Mobility Screening

Reversed Phase High Presssure Liquid Chromatograhphic Technique for Determination of Sodium Alginate from Oral Suspension

23 The Thermal Conductivity Detector

ORGANIC SAMPLE PREPARATION

Fast and Automatic Mapping of Disulfide Bonds in a Monoclonal Antibody using SYNAPT G2 HDMS and BiopharmaLynx 1.3

EKSIGENT EKSPERT NANOLC 400. Unmatched flexibility for low flow LC/MS

Daniel M. Mueller, Katharina M. Rentsch Institut für Klinische Chemie, Universitätsspital Zürich, CH-8091 Zürich, Schweiz

Chapter 28: High-Performance Liquid Chromatography (HPLC)

A Novel Approach to Low Volume Sample Preparation

SIMULTANEOUS DETERMINATION OF NALTREXONE AND 6- -NALTREXOL IN SERUM BY HPLC

The Characterization of Perfume Fragrances Using GC/MS, Headspace Trap and Olfactory Port

Automated Method Development Utilizing Software-Based Optimization and Direct Instrument Control

# LCMS-35 esquire series. Application of LC/APCI Ion Trap Tandem Mass Spectrometry for the Multiresidue Analysis of Pesticides in Water

MultiQuant Software 2.0 for Targeted Protein / Peptide Quantification

An Improved SPE-LC-MS/MS Method for the Quantification of Bradykinin in Human Plasma Using the ionkey/ms System

for SolEx TM HRP Cartridges SolEx TM HRP RSLC Columns

Determination of Anabolic Steroids in Horse Urine by SPE and LC-MS/MS

The Scheduled MRM Algorithm Enables Intelligent Use of Retention Time During Multiple Reaction Monitoring

Intelligent use of Relative Response Factors in Gas Chromatography-Flame Ionisation Detection

CORESTA RECOMMENDED METHOD N 72

Analysis of Casein and Whey Protein in Whole, 2%, and Skim Milk by Capillary Gel Electrophoresis

SPE, LC-MS/MS Method for the Determination of Ethinyl Estradiol from Human Plasma

ACETALDEHYDE and ISOVALERALDEHYDE (Gas Chromatography)

Hydrophilic-Interaction Chromatography (HILIC) for LC-MS/MS Analysis of Monoamine Neurotransmitters using XBridge BEH Amide XP Columns

α-cyclodextrin SYNONYMS α-schardinger dextrin, α-dextrin, cyclohexaamylose, cyclomaltohexaose, α- cycloamylase

HPLC Analysis of Acetaminophen Tablets with Waters Alliance and Agilent Supplies

Anion chromatography using on-line recycled eluents

LC-MS/MS for Chromatographers

Determination of Pesticide Residues in Drinking Water Using Automated Solid-Phase Extraction and Gas Chromatography with Nitrogen Phosphorus Detection

Method Development of LC-MS/MS Analysis of Aminoglycoside Drugs: Challenges and Solutions

Purification of reaction mixtures using flash chromatography.

Therapeutic Drug Monitoring of Antiretroviral Drugs with HPLC-MS

Automation of Solid Phase Extraction and Column Chromatographic Cleanup

of Biologics Using LC/MS in Compliant Laboratories

MEDICINAL CHEMISTRY APPLICATIONS BOOK

ß-CYCLODEXTRIN SYNONYMS

Eleanor Riches Waters Corporation, Manchester, UK INTRODUCTION

PrepTip. Reverse Phase PrepTip User Guide

Determination of Food Dye Concentrations in an Unknown Aqueous Sample Using HPLC

Simultaneous Metabolite Identification and Quantitation with UV Data Integration Using LightSight Software Version 2.2

RAPID MARKER IDENTIFICATION AND CHARACTERISATION OF ESSENTIAL OILS USING A CHEMOMETRIC APROACH

Fast and Accurate Analysis of Vitamin D Metabolites Using Novel Chromatographic Selectivity and Sample Preparation

Salman Azimi, 1 Nayan S. Mistry, 2 and Michelle Wood 2. Drug Quality Control Laboratory, Supreme Council of Health, Doha, Qatar 2

Pesticide Analysis by Mass Spectrometry

Technical Report. Automatic Identification and Semi-quantitative Analysis of Psychotropic Drugs in Serum Using GC/MS Forensic Toxicological Database

PosterREPRINT AN LC/MS ORTHOGONAL TOF (TIME OF FLIGHT) MASS SPECTROMETER WITH INCREASED TRANSMISSION, RESOLUTION, AND DYNAMIC RANGE OVERVIEW

Method development for analysis of formaldehyde in foodsimulant. melamine-ware by GC-MS and LC-MS/MS. Internal Technical Report

SUPPLEMENTARY FIGURES

Simultaneous qualitative and quantitative analysis using the Agilent 6540 Accurate-Mass Q-TOF

INSTRUCTIONS Edition AC

Protocol: HPLC (amino acids)

RESOURCE Q, 1 ml and 6 ml RESOURCE S, 1 ml and 6 ml

INTERNATIONAL JOURNAL OF PHARMACEUTICAL RESEARCH AND BIO-SCIENCE

IC-MS and LC-MS Determination of Ionic Liquids, Counterions, and Impurities

Lean Process Improvements for the GMP Laboratory

Increasing the Multiplexing of High Resolution Targeted Peptide Quantification Assays

Introduction. Methods. Sample Processing

Simultaneous determination of L-ascorbic acid and D-iso-ascorbic acid (erythorbic acid) in wine by HPLC and UV-detection (Resolution Oeno 11/2008)

Transcription:

Quantification of Polysorbate Using the ACQUITY UPLC H-Class Bio System with ELS Detection Sean M. McCarthy Waters Corporation, Milford, MA, USA APPLICATION BENEFITS The ACQUITY UPLC H-Class Bio System, configured with an ACQUITY UPLC Column Manager and ACQUITY UPLC ELS Detector, is an ideal system for the quantification of detergents commonly used in biopharmaceutical formulations. INTRODUCTION Biotherapeutics formulations are complex and their composition is tightly controlled to ensure product stability, and efficacy. Formulations of biotherapeutics contain detergents that are used to prevent denaturation and aggregation of the active pharmaceutical ingredient. Of the detergents used, polysorbates 20 and 80 are very common due to their low toxicity and biocompatibility. By their nature, polysorbates are structurally and chemically heterogeneous mixtures of species, making their quantification challenging. There are recent reports that demonstrate the use of a trap-and-elute strategy for detergent analysis in formulated biopharmaceuticals. 1 Often, analysis of detergent in formulated samples is necessary to ensure stability of the detergents. These strategies provide a rapid and robust process for separating the therapeutic and other excipients from the detergent in the formulation. WATERS SOLUTIONS ACQUITY UPLC H-Class Bio System with Column Manager ACQUITY UPLC ELS Detector In this work, we present the use of the ACQUITY UPLC H-Class Bio System with a Column Manager and an evaporative light scattering detector (ELSD) for determining the concentration of polysorbate in formulated samples. Using the mixed-mode Oasis MAX online cartridge, we effectively trap polysorbate at low ph while the therapeutic and other excipients are unretained. Using the ACQUITY UPLC Column Manager equipped with a switching valve, unretained components are diverted to waste, then the valve position is changed so flow is directed to the ELSD prior to elution of polysorbate from the column. The proposed workflow represents a straightforward and efficient method for analyzing polysorbate content in biotherapeutic formulations. Empower 3 CDS Software Oasis MAX Solid-Phase Extraction KEY WORDS Detergents, polysorbate, evaporative light scattering detection, biopharmaceutical 1

E X P E R IM E N TA L LC conditions QSM Mobile phase A: 2% formic acid in water Mobile phase B: 2% formic acid in isopropanol Seal wash: 10% acetonitrile in water Seal wash period: 5 min. Gradient: Time Flow %A %B %C %D Curve (ml/min) Initial 1.0 90 10 0 0 Initial 1.0 1.0 80 20 0 0 6 3.4 1.0 80 20 0 0 6 3.5 1.0 0 100 0 0 6 4.5 1.0 0 100 0 0 6 4.6 1.0 90 10 0 0 6 6.6 1.0 90 10 0 0 6 Polysorbate 20, Polysorbate 80, and bovine serum albumin were purchased from Sigma Aldrich. Stock solutions of each were prepared in Milli-Q water at 1 mg/ml. Stock solutions were prepared for each polysorbate as follows: 0.4, 0.3, 0.2, 0.1, and 0.05 mg/ml for polysorbate 20 and 0.3, 0.2, 0.1, 0.05, and 0.025 mg/ml for polysorbate 80. Each sample was spiked with bovine serum albumin to a concentration to 30 mg/ml. Sample manager Wash solvent: Purge solvent: 50/50 water/isopropanol 50/50 water/isopropanol Column manager Advanced mode: For plumbing diagram, see Figure 1 Column: Oasis MAX, 30 µm, 2.1 x 20 mm (p/n 186002052) Column temp.: 30 C Events: Time Event Action (min) 2.8 Right valve Position 2 6.5 Right valve Position 1 ELSD Gain: 100 Data rate: 2 pps Time constant: Slow Mode: Heating Power level: 75% Temp.: 80 C Gas pressure: 20 psi (1.64 slpm) 2

RESULTS AND DISCUSSION Instrument configuration and plumbing To prevent fouling of the ELS detector, we configured our system to efficiently divert unretained protein and other excipient material to waste following injection. This was accomplished using 6-port, 2-position valves in the column manager while operating in advanced mode. As shown in Figure 1, the right valve is configured so that the LC flow can be directed as needed during the separation. In position 1 (top) LC flow is directed to waste. In position 2 (bottom) LC flow is directed to the ELS detector. At the end of the separation, the valve is returned to position 1 for the following injection. Programming the valve switch is accomplished within the inlet method editor without the need for additional components or connections. Evaluation of chromatographic reproducibility For routine use, methods used for quantitation need to be reproducible in retention time and area. Since our configuration utilized both a trap-and-elute method as well as a diversion valve, we evaluated the reproducibility of the separation at each of the mass load levels for polysorbate 20 and 80. Shown in Figure 2 is an overlay of triplicate injections for each polysorbate 80 standard prepared as described above, containing 30 mg/ml protein with varying levels of polysorbate. As shown, the separations are reproducible both within and between polysorbate concentrations. From this data we determined that both the instrument and the plumbing configuration used were acceptable for routine polysorbate analysis. Figure 1. Fluidic diagram of column manager. Figure 2. Overlay of triplicate injections of polysorbate 80 at five concentration levels. 3

Evaluation of assay linearity After determining that our system configuration was suitable for polysorbate analysis, we generated calibration curves for polysorbate 20 and 80. The calibration curves were constructed in Empower 3 Software using Quickset and defining concentration levels. The data was then processed using a basic processing method and the calibration curve was generated within Empower. The data was plotted using a log vs. log scale, yielding a linear relationship between concentration and response. As shown in Figures 3 and 4, excellent linearity was achieved for polysorbate 20 and 80, respectively. Figure 3. Polysorbate 20 calibration curve. Data points were collected in triplicate. Figure 4. Polysorbate 80 calibration curve. Data points were collected in triplicate. 4

CONCLUSION The analysis of detergents in biopharmaceutical formulations is a required but often challenging task in the laboratory. The ACQUITY UPLC H-Class System with ELSD and column manager in a trap-and-elute configuration provides a comprehensive system for performing these analyses. As shown, the integrated valves on the ACQUITY UPLC Column Manager provides reproducible results without the need for additional external switching valves. In addition, the ACQUITY UPLC ELSD yields a linear response for quantification. References 1. Hewitt D, Zhang T, Kao YH. Quantitation of polysorbate 20 in protein solutions using mixed-mode chromatography and evaporative light scattering detection. J Chromatogr A. 2008 Dec 26;1215(1 2):156-60. doi: 10.1016/j.chroma.2008.11.017. 2. Megoulas NC, Koupparis MA. Twenty years of evaporative light scattering detection. Crit. Rev. Anal. Chem. 2005;35:301 16. Waters, The Science of What s Possible, ACQUITY UPLC, Oasis, and Empower are registered trademarks of Waters Corporation. All other trademarks are the property of their respective owners. 2015 Waters Corporation. Produced in the U.S.A. March 2015 720005323EN AG-PDF Waters Corporation 34 Maple Street Milford, MA 01757 U.S.A. T: 1 508 478 2000 F: 1 508 872 1990 www.waters.com