Thermopile Support: FAQs, Troubleshooting, Care & Maintenance

Similar documents
THERMAL TRANSFER PRINTING PROBLEM SOLVER

Stone Developments. Caring for Irish Blue Limestone Products. Irish Blue Limestone Maintenance Guide

Care and Maintenance for Wood Flooring

Photographic Printing on Non-Traditional Surfaces

10 Ways to Prolong the Life of your Carpet

Infrared Viewers. Manual

DIRECT THERMAL PRINTING PROBLEM SOLVER

In the previous presentation, we discussed how x-rays were discovered and how they are generated at the atomic level. Today we will begin the

LASER DIAGRAM TABLE OF CONTENTS. SKY Technologies Inc. Laser Diagram. Technical Specifications. Laser Operation.

Applicable Models ME511L, ME551i2, MS31i2, MS33i2, MS51i2, MS53i2

Features: Mounting the Optic

GB Tubulars RECOMMENDED PRACTICE For CASING on LOCATION

Section 6 Brake Drum Failure Analysis. Brake Drum Wear Conditions. What is Normal Wear? What is Deep or Excessive Wear? WARNING. Deep, Uniform Wear

PREVENTIVE MAINTENANCE OF FIBER OPTIC CABLES AND OPTICS

MAKING A BETTER MARK WITH INK JET As substrates, industry regulations and customer demands change, fluids are also evolving

Operating manual ICC Camera. Version Menu

Owner s Manual

Optical Power Meter. Specification & User Manual

WET CLEANING MANUAL. Wet Cleaning instructions Wet Cleaning best practices Liquid Solution Maintenance

VC 104+ Rigid Grade / Rigid Grade Imagine VC 104 Rigid Grade Commercial Customised

Application Note. Soldering Methods and Procedures for 1st and 2nd Generation Power Modules. Overview. Analysis of a Good Solder Joint

CELL PHONE REPAIR SPECIALISTS REPAIR GUIDE. A resource for troubleshooting common handset problems

DRY CLEANING MANUAL. Operating an Arctic Butterfly Dry Cleaning best practices Arctic Butterfly Maintenace Limited Warranty Exclusions & Limitations

National Optical & Scientific Instruments Inc Landmark 35 Drive San Antonio, Texas Phone (210) Fax (210)

INFRARED MONITORING OF 110 GHz GYROTRON WINDOWS AT DIII D

Why Expanded Beam types are essential in military and other harsh environment applications

Spandrel Glass Types and Recommendations

Pictorial Surface Preparation Standards For Painting Steel Surfaces

Troubleshooting Guide for Digital Printing on Creative Papers

LASER PRINTING PROBLEM SOLVER

FT-100 Fire Training Fog Machine User Manual English

AKRAPOVIC SLIP-ON EXHAUST SYSTEM for the HONDA CB1000R

ALUMINUM ELECTRICAL WIRING

1) My Dolphin power supply doesn t turn on when I push the power button: 2) My Dolphin power supply turns on, but the Dolphin does not move:

LBS-300 Beam Sampler for C-mount Cameras. YAG Focal Spot Analysis Adapter. User Notes

SCREAMIN' EAGLE "HEAVY BREATHER" PERFORMANCE AIR CLEANER KIT

Screen Melts. Introduction. Selecting Screen Material. Prefire Before Using. Spalling. Kiln Shelf

Scanners and How to Use Them

WEATHERING SCALE AIRCRAFT

Metal Rescue Rust Remover Technical Data

Survival Laser SL-001BB Laser Parts Bundle Assembly & Operation Instructions

SINGLE RANGE DS METERS Operation Instructions

Stainless Steel Marking Guide

Fixture One Power Cord One User Manual One Omega Bracket One DMX Cable - One

Adjustment functions for both span and shift have been incorporated

How to Build a Printed Circuit Board. Advanced Circuits Inc 2004

Tube Control Measurement, Sorting Modular System for Glass Tube

P R E A M B L E. Facilitated workshop problems for class discussion (1.5 hours)

X-RAY TUBE SELECTION CRITERIA FOR BGA / CSP X-RAY INSPECTION

THE IMPOSSIBLE DOSE HOW CAN SOMETHING SIMPLE BE SO COMPLEX? Lars Hode

VISM CQB Scope Series

II. Monochrome Image Quality Problems

1. INTRODUCTION ABSTRACT

MQSA Quality Control Manual for Monochrome Displays for Mammography

SCREAMIN' EAGLE "HEAVY BREATHER" PERFORMANCE AIR CLEANER KIT

Best Practices for Ensuring Fiber Optic System Performance. David Zambrano

BLACK JACK ALL WEATHER ROOF COATING

K E E P E TH T I H S I MA M N A U N A U L A FO F R

National Optical & Scientific Instrument Inc Landmark 35 Drive San Antonio, Texas Phone (210) Fax (210)

RAIL ADAPTER SYSTEM DRAFT TC-23-AIMSS CHAPTER DESCRIPTION

81000 Series Aluminum Gate Valve Installation and Repair Manual

National Optical & Scientific Instruments Inc Landmark 35 Drive San Antonio, Texas Phone (210) Fax (210)

Colored Pencil art 101. An excerpt from The Ultimate Guide To Colored Pencil by Gary Greene

MIST DECORATIVE LAMP WITH ESSENTIAL OIL DIFFUSER

Thermo Scientific Rotor Care Guide. Protect your rotor investment

Piston Diagnosis A rough Guide

Infrared Thermometry. Introduction, History, and Applications. Optical Pyrometry. Jason Mershon, Advanced Energy Industries, Inc

GNOME PELLET E.I. Pellet Heater Owner's Manual Installation and Operating Instructions. Please read this entire manual before installation.

Care and use guide. Contents. a handy solution made easy. page

Tube Liquid Sensor OPB350 / OCB350 Series

Photovoltaic System Technology

SC, FC and ST * Hot Melt Fiber Optic Connectors

Air Hockey Training Camp Care and repair of your Air Hockey Playfield

OPERATION MANUAL INFRARED THERMOMETER

Active and Passive Sash Replacement

Sewer Jetter GuideSM. How to Unclog Drains with Your Pressure Washer. Use this Guide To:

RAUFOSS EXPLOSIVE COMPRESSION FITTINGS

Ignition Coil Sealing Instructions Revision 1.0

FINGERPRINTING IS A SCIENCE - TAKE A SCIENTIFIC APPROACH

Module Connectors and Cable Specifications

SPECIFICATION Aluminum Module Frames. Allowed anodization before fabrication for clear frames.

FOR THE FOLLOWING MODELS: EE-8075W EE-8075O EE-8075R EE-8075BK

Guide Specification STAYFLEX CORROSION CONTROL AND THERMAL INSULATION SYSTEM

By the way, do not get ticked off at me if your results are not fabulous. What you do with this information is up to you.

The Peak of Protection

U LED modules, converters and systems GENERAL ILLUMINATION. Umodule SPOT P330-2 umodule SPOT PHASED OUT NTC

How to Save a Wet Cell Phone

TM Advanced Tracking Technologies, Inc ATTI All rights reserved

U LED modules, converters and systems GENERAL ILLUMINATION. Umodule SPOT P340-2 umodule SPOT PHASED OUT NTC

Experimental Study on Non Destructive Testing Techniques (NDTT)

HVAC Super Pro FAQ. Can HVAC SUPER PRO be used on both hard and soft copper and aluminum tubing?

The soot and scale problems

USER MANUAL PLEASE KEEP THIS FOR FUTURE USE

Reasons for reissue are provided in Section 6, REVISION SUMMARY.

WARNING: FAILURE TO FOLLOW THESE RULES MAY RESULT IN SERIOUS PERSONAL INJURY CAUTION: INSTALLATION LOCATION:

Transcription:

Thermopile Support: FQs, Troubleshooting, Care & Maintenance Table of Contents 1. Introduction 2. How long does a thermopile sensor last and how accurate does it remain over time? 3. What are typical causes of damage on thermopile sensors? 4. What does damage look like on thermopile sensors and how can I tell if my sensor is working properly? 5. Examples of thermopile disks in varying conditions a. Figure 1: Healthy Thermopile Sensor with No Damage b. Figure 2: Discoloration and Scratches Caused by Moving the Sensor during Measurement c. Figure 3: Thermopile near Its Damage Threshold Caused by Moving Sensor during Measurement d. Figure 4: Thermopile Damage as the Result of Too Low of a Power Density e. Figure 5: Severe Thermopile Damage Caused by Too High of a Power Density f. Figure 6: Damaged Caused by Too High Total Power Level and Density g. Figure 7: Streaks on Thermopile Caused by Contamination h. Figure 8: Discoloration Due to Contamination i. Figure 9: kw Level Power Sensor near Its Damage Threshold j. Figure 10: Significant Damage to kw Level Power Sensors Due to Too High of a Power Density or Insufficient Water-Cooling 6. How do I avoid damage with thermopile sensors? 7. How can I clean a thermopile sensor if it gets dirty? 8. How are sensors repaired when they have damage on them? Introduction Thermopile power sensors have been commonly used for laser power measurements for decades. They are a fairly robust measurement technology and can be used with lasers from relatively low power levels in the mw range up to very high power levels in the kw range. question that comes up relatively often, is how long thermopiles last and what to look for to make sure a thermopile is in good working condition. long with this, come questions about maintenance and cleaning of thermopiles. This document is intended to help answer some of the common questions related to those topics. www.coherent.com I lmc.sales@coherent.com I (800) 527-3786 I (408) 764-4983 1

How long does a thermopile sensor last and how accurate does it remain over time? Thermopile sensors don t have a specific life expectancy or degradation rate. s long as they are used correctly and maintained properly, thermopiles can last for years or even decades and remain in good working condition. The key is typically to just avoid damaging or contaminating the sensor surface. Thermopiles are often built out of aluminum or copper absorbing disks with a fairly strong ceramic-type coating on the surface. These sensors can typically withstand a decent amount of abuse without damaging. But there are definitely some things to watch for. If a sensor can be maintained and used without discoloration or damage to the sensor coating, these will often remain well within spec for a year or longer. We normally recommend getting thermopiles recalibrated on an annual basis to help monitor their performance and keep the sensor within spec. nd as long as nothing severe happens between calibration intervals, thermopiles can have a very long lifetime with very little measurement drift. voiding contamination on the sensor surface is an important part of maintaining thermopiles. ny type of dirt, dust, oil, or grease that settles on this surface can cause a layer of contamination that can lead to problems. Debris or liquids that collect on the sensor surface can actually burn up when a laser beam hits the thermopile disk, which can cause increased heat on the sensor surface and can lead to damage or discoloration. So keeping thermopile sensors in a relatively clean environment can help the lifetime and accuracy of the product. ny significant changes in the color of the absorbing disk due to contaminants on it will most likely lead to changes in the absorption properties of the sensor and can lead to accuracy problems. Even having water seep into the sensor coating can cause problems with the strength of that coating. So it s best to try and avoid contact with most liquids on the sensor surface of these products. What are the typical causes of damage on thermopile sensors? The most common cause of damage to these sensors comes from being used with too high of a power density from the laser beam. Power density has components of both the total power level and the laser beam size at the sensor surface. If the laser beam is being focused or is coming out of the end of a fiber, the beam size is very small at those locations. With small beams, the power density can be very high even if the total power from the laser is not significant. The goal with thermopile power sensors is to try and avoid positioning the sensor at a point that is near the focus of the beam. Thermopile sensors measure the total power in the laser beam and do not directly measure power density. Whether a laser beam is 5 mm in diameter or 10 mm in diameter is irrelevant to the total power measurement the thermopile is capturing. However, differences in beam size can make a large difference in the power density of that beam as it is hitting the sensor surface. It is important to understand the maximum power density specifications for the thermopile being used and to try and avoid measuring very small beam diameters. Typically the larger the beam diameter is at the sensor surface, the less chance there is of damaging the coating on the thermopile. nother source of damage could be total power. If a thermopile is rated for a maximum power level of 10 W and it is used to measure the output of a 50 W laser, the thermopile can overheat, causing internal damage to the sensor which may or may not be visible on the absorbing surface. nother possible source of damage is from objects coming in direct contact with the coating on the thermopile disk. For example, if the thermopile is being stored in a toolkit or case, it is important to ensure other tools (including the connector on the end of the sensor s cable) do not come in contact with the absorbing disk on the thermopile. Objects www.coherent.com I lmc.sales@coherent.com I (800) 527-3786 I (408) 764-4983 2

coming in contact with the absorbing surface can cause scratches in the coating, which can lead to problems with its operation and accuracy. This is also true for lasers that are being emitted out of a fiber or some type of handpiece. If the beam is exiting the laser through a fiber or a handpiece, it is best not to put the output of the laser directly up to the surface of the thermopile. This can cause physical damage to the thermopile and also to the laser output. It also increases the chances of the power density being too high at that output. When using a laser that has a fiber output, it is best to position that fiber a little ways away from the sensor surface to help protect the coating and to allow the beam to diverge out of the end of the fiber before reaching the absorbing surface. What does damage look like on thermopile sensors and how can I tell if my sensor is working properly? One of the best things that can be done to help determine if a thermopile sensor is working properly is to just do a visual inspection of the coating on the absorbing disk. This can help tell a lot about the condition of the sensor and whether or not it could be measuring accurately. The thermopile coating should be a uniform gray color. In the case of the Excimer-coated sensors, it will be a dull blue color. There should not be any black spots, bleached spots or metal showing through the coating. If any of those signs of damage are showing up on the sensor, there is a decent chance that it may not be measuring correctly. Discolorations or burn spots on the sensor coating can cause the absorption properties of the coating to change, which can cause accuracy and drift issues with thermopiles. If the coating is a uniform gray color, then there s a good chance that at least the absorbing disk on the sensor is in good condition. If there are major burns or discolorations, that s a good sign that the sensor may not be working within spec. Examples of different thermopiles in varying conditions The following photo gallery will illustrate how to properly diagnose various conditions as well as causes and next steps. a. Figure 1: Healthy thermopile sensor with no damage b. Figure 2: Discoloration and Scratches Caused by Moving the Sensor during Measurement c. Figure 3: Thermopile near Its Damage Threshold Caused by Moving Sensor during Measurement d. Figure 4: Thermopile Damage as the Result of Too Low of a Power Density e. Figure 5: Severe Thermopile Damage Caused by Too High of a Power Density f. Figure 6: Very Severe Damage Caused by Too High Total Power Level and Density g. Figure 7: Streaks on Thermopile Caused by Contamination h. Figure 8: Discoloration Due to Contamination i. Figure 9: kw Level Power Sensor near Its Damage Threshold j. Figure 10: Significant Damage to kw Level Power Sensors Due to Too High of a Power Density or Insufficient Water-Cooling www.coherent.com I lmc.sales@coherent.com I (800) 527-3786 I (408) 764-4983 3

Figure 1: Healthy Thermopile Sensor with No Damage The absorbing disk on the sensor is a uniform flat gray color with no discolorations or burn marks on it. sensor in this condition will typically not have any problems with the thermopile disk itself. www.coherent.com I lmc.sales@coherent.com I (800) 527-3786 I (408) 764-4983 4

B Figure 2: Discoloration and Scratches While Moving the Sensor during Measurement. The dark discolorations that cover the entire sensor area are a collection of smaller circular marks caused by using a power density that is slightly higher than recommended while moving the beam around to different areas on the sensor disk. Slight discolorations (either slightly lighter or slightly darker marks) should not cause sensor performance issues. However, the circular marks pictured above are darker than the flat gray color of a healthy sensor surface, which indicate that the coating is starting to burn. In this specific area, the damage has not ablated or melted the coating of the sensor. However, the dark discolorations indicate the coating is starting to burn. We recommend changing the absorbing disk before major damage is caused to the sensor. B. The marks that resemble squiggly scratches are areas where the coating has been ablated. This is the result of using a tightly focused beam and moving the sensor around in the beam path during a measurement. It is evident the coating is missing due to the exposed metallic color within the marks. www.coherent.com I lmc.sales@coherent.com I (800) 527-3786 I (408) 764-4983 5

B D C Figure 3: Thermopile near Its Damage Threshold While Moving Sensor during Measurement. The dark marks indicate the coating is beginning to burn. B. Small metallic spots surrounded by a darker ring indicate the coating has been ablated C. The slightly lighter gray spots are on the boarder of what we consider damage. This is what happens if the sensor is used very near its damage threshold. The coating discoloration typically does not cause performance problems. If you continue to see similar discoloration or if the discoloration darkens, it is necessary to reevaluate the beam size before continued use. In this particular figure we recommend replacing the disk due to the presence of both dark spots and the ablated spots. The linear pattern of marks across this sensor surface appears to indicate that the sensor was moved through the beam path while the power density was fairly high. D. blation on the anodized housing of the sensor does not cause any performance problems with the product, but could be another indication that the power density is too high and that the sensor has most likely been placed too close to the focal point in the beam path. www.coherent.com I lmc.sales@coherent.com I (800) 527-3786 I (408) 764-4983 6

B C Figure 4: Thermopile Damage as the Result of Too High of a Power Density This is a common example of what happens when too small of a beam size is used. The power density in these small circular spots has zapped the coating right off the sensor.. The green color within the circular spot indicates that the sensor coating has started to melt at that spot. The presence of the color green on the sensor surface represents relatively significant damage due to a melted coating. B. Within this spot, metallic color is surrounded by a darker ring which means the coating has been ablated and there is charring around the ablation. This is caused by too high of a power density hitting the sensor surface. The sensor disk must be replaced in order to recalibrate it. www.coherent.com I lmc.sales@coherent.com I (800) 527-3786 I (408) 764-4983 7

C. Small scratches are caused by something coming in physical contact with the sensor surface. For example, a tool could have hit the sensor surface while being put away, or perhaps the connector on the sensor s cable accidentally rubbed across the sensor surface. These small scratches are usually on the very surface of the sensor and will not cause major problems unless metal can be seen within the scratches of the coating. Typically, no replacement of the thermopile disk is required unless there are a number of scratches and the scratches are deep enough to reveal metal. B C Figure 5: Severe Thermopile Damage Caused by Too High of a Power Density www.coherent.com I lmc.sales@coherent.com I (800) 527-3786 I (408) 764-4983 8

. This is the most serious damaged spot within this figure caused by too high of a power density. The coating has been ablated and there is metal visible. The appearance of this spot would prompt a recommendation of replacing the sensor disk during repair or recalibration. Due to the spot s central location in this particular figure, the sensor cannot be recalibrated until the disk is replaced. B. There are a number of lighter colored marks on the sensor surface that appear to have been caused by a laser beam. These spots are slight discolorations in the coating and may or may not prompt a disk replacement depending on how many there are and their location on the sensor surface. Generally, if there are one or two slightly discolored spots like this, it will not greatly affect the sensor performance and the sensor can be recalibrated to account for any slight changes in absorption due to spots like these. If there are quite a few spots, to the point where the overall absorption of the sensor is compromised, it may prompt recommending a disk replacement. C. There are a few of these small black marks on this sensor which indicate some type of burn mark. It s possible a very small beam caused this type of burn. Or it could be that some piece of debris (maybe a piece of dirt or a drop of grease) landed on the sensor at some point. The next time a laser beam hit that spot, it may have burned the debris onto the sensor coating. Spots like this can burn in the laser beam even if the power density isn t real high. few small spots like this on a sensor may not warrant replacing the disk. But larger spots or large collections of these spots could prompt replacing the disk. www.coherent.com I lmc.sales@coherent.com I (800) 527-3786 I (408) 764-4983 9

B Figure 6: Very Severe Damage Caused by Too High of a Total Power Level and Density. Example of a severely ablated and melted thermopile sensor rendered non-functional. The user most likely used too high of a total power level as well as too high of a power density, puncturing the middle of the sensor disk B. There are also a number of smaller marks around the housing and onto the sensor disk that would have been caused by a very tightly focused beam hitting the sensor. These marks did not cause the significant damage in the center of the sensor, but are a good indication that this sensor was not being used properly. These were most likely caused by a pulsed laser as the sensor was moved in and out of the beam path. www.coherent.com I lmc.sales@coherent.com I (800) 527-3786 I (408) 764-4983 10

Figure 7: Streaks on Thermopile Caused by Contamination This is an example of a sensor that has some marks on it that are most likely not affecting the performance of the product.. The black streaks on this sensor surface were most likely caused by a small amount of grease or dirt that smeared across the sensor surface. Marks like this typically won t cause problems with a sensor, especially since they are not in the very center of the sensor area. There is a chance that marks can burn onto the sensor disk if a laser beam hits them. However, if the marks do not get larger or worse with time, a sensor in this condition will most likely continue to work properly and may not need a disk replacement during recalibration. www.coherent.com I lmc.sales@coherent.com I (800) 527-3786 I (408) 764-4983 11

Figure 8: Discoloration Due to Contamination. This is another case where there are some slight discolorations on the sensor surface most likely caused by small bits of contamination that got on the sensor disk. This type of discoloration is not severe enough to cause problems with the sensor s performance. www.coherent.com I lmc.sales@coherent.com I (800) 527-3786 I (408) 764-4983 12

Figure 9: kw Level Power Sensor near Its Damage Threshold This is a case specific to kw level power sensors such as the PM1K, PM5K and LM-5000 products.. The discoloration in this area was definitely caused by a laser beam, although it is not to the point of melting, charring, or ablation yet. Some slight discoloration with the kw level sensors is somewhat normal just given the high amount of power that these sensors absorb. Therefore, a slightly lighter or darker spot similar to what is seen here isn t necessarily indicating that this has been damaged. This particular sensor technically has not been damaged yet. However, this discolored spot is starting to show a bit of a blue/green color to it, which means that the coating could be getting close to melting. If a spot like this shows up on the sensor, it would be recommended to take steps to reduce the power density that is being used on the sensor or to increase the water flow rate through the sensor to try and help keep it a bit cooler during measurements. www.coherent.com I lmc.sales@coherent.com I (800) 527-3786 I (408) 764-4983 13

It is very important to understand the damage threshold and water-cooling requirements for kw sensor products. water flow requirements note and kw damage threshold note have been created for these products, which should both be reviewed prior to using one of these kw level sensors. To learn more about water-cooling requirements for kw sensor products, read our applications note. C B Figure 10: Significant Damage to kw Level Power Sensor Due to Too High of a Power Density or Insufficient Water-Cooling. In this area a green color is present which indicates that the sensor coating has melted. In addition, there is green smoke in this damaged region. B. Black charring marks are present in this area. C. The copper thermopile disk is exposed near the center of the sensor www.coherent.com I lmc.sales@coherent.com I (800) 527-3786 I (408) 764-4983 14

This sensor is an extreme case, but illustrates key things to watch for on the kw level sensors. ny time that a green melted spot, black charring, or exposed metal show up, the sensor has been significantly damaged and should not continue to be used in that condition. Damage like this can be the result of too high of a power density or from the sensor getting too hot, like what can happen without sufficient watercooling. How do I avoid damage with thermopile sensors? The primary source of damage that we typically see is simply from using too small of a beam size. Users of these products need to understand that laser power measurements do not necessarily need to be taken at the working surface or the focal point in the beam. In fact, most of the time it is ideal to avoid those spots in the beam to help avoid damage. thermopile sensor is capable of measuring the total power in the laser beam, which can be done at a point where the beam is defocused. s long as the beam size fits within the front aperture on the thermopile sensor, the sensor should be able to take an accurate reading on it. nd in most cases, the larger the beam is, the better the chances are of avoiding damage. nother source of damage is just due to overheating the sensor. This can be caused by using a sensor above its maximum rated power level, or by not having sufficient water-cooling available on water-cooled sensors. Understanding the specifications and water-cooling requirements for these products is the easiest way to avoid overheating a thermopile sensor. How can I clean a thermopile sensor if it gets dirty? Cleaning your thermopile sensors can pose a challenge simply because we recommend avoiding any liquids coming in contact with the coating on the sensor. This is due to the coating s porous nature that allows liquids to easily seep in, causing potential corrosion or future damage. If there is dust or debris that has settled on the sensor coating, the safest cleaning method is to use clean pressurized air or compressed nitrogen air to blow the contamination off the sensor surface. The thermopile disks are fairly robust and can take quite a bit of pressure blowing onto them when trying to dislodge something that is on the surface. If needed, a soft brush, like a paint brush or a soft toothbrush can be used to try and brush any material off the sensor surface. This could be used in conjunction with pressurized air. void using a cloth or anything similar that could leave lint behind. If there is oil or grease that has gotten onto the sensor that can t be simply blown off, a small amount of optical cleaning solution (such as alcohol) can be used on the sensor. This would need to be applied using something similar to a soft brush. gain, avoid using material that will leave lint behind. ny type of optical cleaning solution should be used sparingly and dried off immediately using a heat gun or compressed air. ny residual liquid on the sensor surface is at risk of seeping into the coating. www.coherent.com I lmc.sales@coherent.com I (800) 527-3786 I (408) 764-4983 15

How are sensors repaired when they have damage on them? Thermopile sensors like these cannot typically be recoated to repair damage. Once the coating is damaged or ablated, the thermopile disk usually needs to be replaced. This can get a little expensive since the thermopile disk is the only real active component inside most of these sensor products. But this is typically the only way to repair these. Sensor disks cannot be replaced in the field because the response and the calibration of the sensor will change with a disk replacement. So normally to repair these products, the thermopile will need to be sent to the factory to have a disk replacement and a recalibration. www.coherent.com I lmc.sales@coherent.com I (800) 527-3786 I (408) 764-4983 16