ACR CT Accreditation: The New Phantom Paradigm Douglas Pfeiffer, MS, DABR Boulder Community Hospital

Similar documents
2/28/2011. MIPPA overview and CMS requirements. CT accreditation. Today s agenda. About MIPPA. Computed Tomography

American College of Radiology CT Accreditation Program. Testing Instructions

Moving Forward What does this mean for the Medical Physicist and the Imaging Community?

Cynthia H. McCollough b) and Michael R. Bruesewitz Department of Radiology, Mayo Clinic College of Medicine, Rochester, Minnesota 55905

Assessing Radiation Dose: How to Do It Right

CT Protocol Optimization over the Range of CT Scanner Types: Recommendations & Misconceptions

CT RADIATION DOSE REPORT FROM DICOM. Frank Dong, PhD, DABR Diagnostic Physicist Imaging Institute Cleveland Clinic Foundation Cleveland, OH

How To Improve Your Ct Image Quality

MDCT Technology. Kalpana M. Kanal, Ph.D., DABR Assistant Professor Department of Radiology University of Washington Seattle, Washington

CT QC Under the ACR QC Manual. What? There s a Manual?? Learning Objectives 8/6/12! ! YES!!! Almost. ! Doesn t have a pretty cover yet

Prepublication Requirements

CT: Size Specific Dose Estimate (SSDE): Why We Need Another CT Dose Index. Acknowledgements

The disclaimer on page 1 is an integral part of this document. Copyright February 23, 2016 by AAPM. All rights reserved.

WHERE IN THE WORLD JILL LIPOTI?

The disclaimer on page 1 is an integral part of this document. Copyright March 1, 2016 by AAPM. All rights reserved.

The Challenge of CT Dose Records

REGULATION: QUALITY ASSURANCE PROGRAMS FOR MEDICAL DIAGNOSTIC X-RAY INSTALLATIONS N.J.A.C. 7:28-22

Tracking Radiation Exposure From Medical Diagnostic Procedures: Siemens Perspectives

Purchasing a cardiac CT scanner: What the radiologist needs to know

Protocol Management and Review Strategies The MD Anderson Experience

SUBCHAPTER 22 QUALITY ASSURANCE PROGRAMS FOR MEDICAL DIAGNOSTIC X-RAY INSTALLATIONS

Breast MRI Quality Control

Practical exercise: Effective dose estimate in CT

Role of the Medical Physicist in Clinical Implementation of Breast Tomosynthesis

Implementation of Cone-beam CT imaging for Radiotherapy treatment localisation.

Physics testing of image detectors

Multi-slice Helical CT Scanning of the Chest

Quality control of CT systems by automated monitoring of key performance indicators: a two-year study

Chest CT protocols. Mannudeep K. Kalra, MD, DNB. Dianna D. Cody, PhD. Massachusetts General Hospital Harvard Medical School

Acknowledgement. Diagnostic X-Ray Shielding. Nomenclature for Radiation Design Criteria. Shielding Design Goal (Air Kerma):

An Overview of Digital Imaging Systems for Radiography and Fluoroscopy

Copyright March 1, 2016 by AAPM. All rights reserved.

Update on ACR Digital Mammography QC Manual

PET/CT QC/QA. Quality Control in PET. Magnus Dahlbom, Ph.D. Verify the operational integrity of the system. PET Detectors

Please follow all instructions carefully.

Gestión global de la dosis en TC. Sistema de registro y gestión

Performance testing for Precision 500D Classical R/F System

Table of Contents. Scan acquisition and user interface basics. Dose modulation and reduction tools. Multi-Slice Detector Geometry

QUANTITATIVE IMAGING IN MULTICENTER CLINICAL TRIALS: PET

Diagnostic Exposure Tracking in the Medical Record

INTRODUCTION. A. Purpose

Patient-centered CT imaging: New methods for patient-specific optimization 1 of image quality and radiation dose

SITE IMAGING MANUAL ACRIN 6698

The disclaimer on page 1 is an integral part of this document. 1. Copyright December 14, 2015 by AAPM. All rights reserved.

OPTIMIZING PATIENT EXPOSURE TO IONIZING RADIATION (OPEIR) MEASURES GROUP OVERVIEW 2015 PQRS OPTIONS FOR MEASURES GROUPS:

Development of SCENARIA New Version Software MEDIX VOL. 62 P.32 P.35

Quality Control of Full Field Digital Mammography Units

The AAPM does not endorse any products, manufacturers, or suppliers. Nothing in this publication should be interpreted as implying such endorsement.

Surveying and QC of Stereotactic Breast Biopsy Units for ACR Accreditation

M D Anderson Cancer Center Orlando TomoTherapy s Implementation of Image-guided Adaptive Radiation Therapy

R/F. Efforts to Reduce Exposure Dose in Chest Tomosynthesis Targeting Lung Cancer Screening. 3. Utility of Chest Tomosynthesis. 1.

Patient sample criteria for the OPEIR Measures Group are all patients regardless of age, that have a specific CT procedure performed:

GE Healthcare. DoseWatch. Gathering Radiation Dose Data

Chest 1: Pulmonary Nodule Follow-up: Low-Dose Helical CT (Unenhanced) (Non-metastatic) Gantry Rotation Time. mas (Reg-Lg) 40-80

3/12/2014. Disclosures. Understanding IAC CT Accreditation. Outline. Learning Objectives. Who is the IAC?

Interim policy effective January 1, 2016, through December 31, 2017

Spiral CT: Single and Multiple Detector Systems. AAPM Refresher Course Nashville, TN July 28,1999

Q.A. Collectible. Sponsored by CRCPD s Committee on Quality Assurance in Diagnostic X-Ray (H-7)

Introduction. Digital Mammography Detector QC FFDM. What is FFDM QC and why is it. important? What to know before you start. Eric A. Berns, Ph.D.

Clinic. ED Trauma Trauma Stroke. OR Neuro/Spine. Critical Care. Neuro ENT. Diagnostic. Pediatric. Radiology. Plastics Thoracic. Neuro.

Fundamentals of Cone-Beam CT Imaging

Patient Dose Tracking for Imaging Studies. David E. Hintenlang, Ph.D., DABR University of Florida

Siemens Computed Tomography

Digital Breast Tomosynthesis QC Requirements

GE Healthcare. Revolution EVO. More than just high tech. Higher purpose.

Quality Reports With. PlanIQ. *Export only. 510(k) pending. Patent pending. Your Most Valuable QA and Dosimetry Tools

Nuclear Medicine Accreditation Program Requirements

A software tool for Quality Assurance of Computed / Digital Radiography (CR/DR) systems

The IAC Standards and Guidelines for CT Accreditation

Cardiac CT for Calcium Scoring

Thinking ahead. Focused on life. REALIZED: GROUNDBREAKING RESOLUTION OF 80 µm VOXEL

DICOM Correction Item

Computed Tomography Radiation Safety Issues in Ontario

Scan Time Reduction and X-ray Scatter Rejection in Dual Modality Breast Tomosynthesis. Tushita Patel 4/2/13

X-Rays Benefits and Risks. Techniques that use x-rays

IAC Ch 41, p.1. Procedure means a stereotactically guided breast biopsy performed on a patient for diagnostic purposes.

Pediatric Hospitals Bring Low-dose CT to the Middle East

PET Accreditation Program Requirements

ESTIMATING POPULATION DOSES FROM MEDICAL RADIOLOGY

Cirrus 0.2T. MRI for Everyone. North America, Asia, Europe. contact:

SECTION 1: REQUIREMENTS FOR CERTIFICATES OF COMPLIANCE FOR CLASSES OF RADIATION APPARATUS

AAPM Medical Physics Practice Guideline 1.a: CT Protocol Management and Review Practice Guideline

SWABIK Project. Software Tools for DICOM Media Exchange in Clinical Research: the. SWABIK Project

CONTENT SPECIFICATIONS FOR THE FLUOROSCOPY EXAMINATION

TISSUE MIMICKING GEL QUALITY LE PHANTOM SERIES DESIGN. performance the ultrasound labs ofand. icking material has the same attenuation mim-

Brilliance CT 64-channel confi guration

X-ray (Radiography) - Chest

Master s Program in Medical Physics. Physics of Imaging Systems Basic Principles of Computer Tomography (CT) III. Prof. Dr. Lothar Schad.

SECTION 1: REQUIREMENTS FOR CERTIFICATES OF COMPLIANCE FOR CLASSES OF RADIATION APPARATUS

Low-dose CT for Pulmonary Embolism

Study the Quality Assurance of Conventional X-ray Machines Using Non Invasive KV meter

X-ray (Radiography) - Bone

PET/CT-MRI First clinical experience

Rb 82 Cardiac PET Scanning Protocols and Dosimetry. Deborah Tout Nuclear Medicine Department Central Manchester University Hospitals

Annual MQSA Inspection Questions NMQAAC April 19, 2004

DENTAL Cone beam 3D X-RAY SYSTEM with

Automated EMR Dose History Extraction and Monitoring

The Exposure Index and Its Standardization

Accreditation Is Coming

Dose Modulation Technique in CT

Transcription:

Introduction ACR CT Accreditation: The New Phantom Paradigm Douglas Pfeiffer, MS, DABR Boulder Community Hospital The CT Accreditation program has been modified into a modular format Most of the application is on-line Submission process, particularly for the phantom, has changed significantly Help? Outline Call the ACR (800-770-0145) Theresa Branham, Dina Hernandez ctacred@acr.org Has individuals who want to help Process is intended to be educational Changes from old system How to scan the phantom What to submit Problems, pitfalls and solutions Wants facilities to become accredited Don t be shy - call! (800-770-0145) Why the Changes? What has Changed? Based on experience with original model Mini annual survey Low yield evaluations Not applicable to some situations Many facilities have no means to print films Subjective evaluations Clinical scans Some evaluations eliminated SMPTE Positioning Water CT # vs. kvp Water CT # vs. image thickness Spatial resolution Low contrast CNR No facility measurements Ped Head is now included, if Ped accreditation 1

Important Document Fundamentally Changed! Following MRI model Scan up to 4 protocols Submit scans on CD in DICOM format has not changed Phantom Scanning Phantom Submission Gammex 464 phantom performed by technologist or physicist If performed by technologist, physicist should check all images and paperwork CTDI portion must be performed by physicist Examination parameters used for phantom submission Adult Head Adult Abdomen Pediatric Head (1 year old) Pediatric Abdomen (average 5 year old [40-50 lb] technique) Use average (typical) techniques; do not use automatic dose reduction options Exams will not necessarily match clinical submission Phantom Submission CT Phantom Site Scanning Data Form CT Phantom Site Scanning Data Form - online Dose spreadsheets - online ACR Phantom Images (Scanned by technologist or physicist) - on CD CTDI Images (Must be performed by physicist) - on CD All parameter information should be consistent on all forms and images Team approach work with the technologists and radiologists Verify that information given to you are the ones actually used for clinical examinations It is not necessary to report ROI values or provide CTDI calculations 2

CT Phantom Site Scanning Data Form page 1 CT Phantom Site Scanning Data Form page 2 Verify that information is complete and correct List all kvp stations and slice thickness options Old Table 1 Complete this form prior to any scanning List average or typical parameters used in clinical practice 70 kg patient Will enter ma, mas, and effective mas (if available) Use definitions from Phantom Testing Instructions Use IEC definition of Pitch I/NT Technologist training may be necessary It is presumed that the images submitted by the site are the best available Phantom Images ACR CT Accreditation Program requires scans for up to four clinical exams Adult Head (if Adult Head/Neck) Adult Abdomen (ALL SUBMISSIONS) Pediatric Head (if Ped and Head/Neck) Pediatric Abdomen (if Ped and Abdomen) Must scan phantom using clinical protocols with only a few exceptions Phantom Images Phantom Images Entire DICOM series submitted on CD Use 21 cm DFOV Do NOT use auto ma feature Match SFOV to phantom Scan entire phantom (0 120) Only primary interpretation reconstruction May provide viewer Ensure that it works on a PC It is NOT required Follows MRAP phantom submission paradigm 3

4/28/11 Phantom Scoring CT# Accuracy Performed by ACR Physicist Reviewers Protocol review Material CT # accuracy in Abdomen protocol Water CT # accuracy in other protocols Contrast-to-noise ratio (in Module 2) - all CT # uniformity in Abdomen protocol CT # Range Water -7 to +7 HU Air -970 to -1005 HU Teflon (bone) 850 to 970 HU Polyethylene -107 to -84 HU Acrylic 110 to 135 HU Artifacts all Best image will be used for CNR But What of the VCT? 20 mm 40 40 mm mm with leading phantom GE VCT CT Numbers 40 mm 4

Low Contrast Performance Contrast-Noise Ratio Not a scanner test, per se, but an evaluation of protocols Low contrast test was historically too subjective Rod analysis does not apply well to Pediatric protocols CNR provides an objective measure of noise-limited performance Values derived empirically Adult Head and Body Measured CNR with protocols yielding 6 mm rods just visible per the Physics Subcommittee Set CNR value somewhat lower than that to allow for subjective nature of just visible Typically will get CNR ~ 1.2 Contrast-Noise Ratio Contrast-Noise Ratio Pediatric Head and Body Values derived from set of pediatric imaging experts Asked the question: What is the greatest amount of noise I can tolerate and be comfortable making a diagnosis? With that answered, scanned the phantom with that technique CNR limits set a bit lower than resulted from that technique CNR is very dependent upon reconstruction algorithm Required values assume standard reconstruction, as is typical for standard head and abdomen studies Required values assume standard image thickness: 3-5 mm Contrast-Noise Ratio CT# Uniformity Controlling dose is important Lowest dose is not always the best dose Many facilities have driven technique too low, especially in pediatrics Ped Head should pass CNR at a dose of approximately 30-35 mgy This is currently under discussion. Early data may have been misleading (small sample of experts), though the premise remains. All peripheral ROI mean values should be within ± 5 HU of the central value 5

All data will be entered on-line kvp, ma, rotation time must match Data Form Detector configuration should match Data Form if possible Match total beam width Correct table increment in calculation to give proper pitch ACR CT Accreditation Program requires doses for up to four clinical exams Adult Head Adult Abdomen Pediatric Head Pediatric Abdomen Must be measured in axial (sequential) mode Axial Conversion Note actual z-axis (detector) collimation (T) and number of data channels (N) used Do not confuse z-axis collimation with nominal image thickness! May not be able to achieve the same detector configuration in axial as used in helical Axial Conversion (continued) Determine NT of detector configuration underlying the clinical helical protocol Select axial detector configuration most closely matching the helical NT Use this axial detector configuration in all subsequent calculations Adjust table speed in calculation to keep pitch the same Dose Levels Axial Conversion (continued) Example: Siemens Sensation 16 Helical N = 16, T = 1.5 mm NT = 24 mm Pitch = 1.2 Reconstructed scan width = 5 mm Cannot get axial images N = 16 So Use T = 1.5 mm, N = 12 NT = 18 mm Table speed = 21.6 mm in calculation 21.6/(12 x 1.5) = 1.2 All doses are CTDIvol Examination Reference Levels Pass/Fail Criteria Adult Head 75 mgy 80 mgy Adult Abdomen 25 mgy 30 mgy Pediatric Head Pediatric Abdomen No levels set yet 20 mgy 25 mgy 6

Dose P/F Levels Based on CTAP data Head dose limits increased Abdomen dose limits decreased NOTE: Increased head reference dose does not mean that you SHOULD increase dose A number of scanners could not achieve acceptable image quality at original value MANY SCANNERS CAN If images are acceptable at 60 mgy limit, leave it DLP (mgy-cm) = CTDIvol (mgy) total scan length (cm) For ACR, assume total scan length = 17.5 cm for head = 25.0 cm for adult abdomen = 12.0 cm for pediatric head = 15.0 cm for pediatric abdomen Effective Dose (E) = k (msv/mgy-cm) * DLP (mgy-cm) where k = 0.0021 for head = 0.015 for adult abdomen =.0067 for pediatric head = 0.020 for pediatric abdomen At least one (1) dosimetry image must be submitted for each phantom protocol submitted If scanner puts all 6 measurements in one series, submit the whole series If scanner separates each scan into different series, just submit one series Whatever is easiest! Assessment Phantom and dose submission will fail if 7 minor deficiencies 1 major deficiency Phantom Deficiencies Phantom Deficiencies Deficiency Detector configuration inappropriate kvp inappropriate Rotation time inappropriate Table increment/speed inappropriate Reconstruction algorithm inappropriate Minor Major Deficiency Minor Major Phantom kvp differs from Data Form Phantom mas different from Data Form 10% >10% Phantom Pitch different from Data Form 10% >10% CT number not in range CNR too low CT number non-uniformity (N) 5 N 7 >7 7

Phantom Deficiencies Deficiency Minor Major Phantom Deficiencies Deficiency Minor Major Artifacts Sub-clinical Super-clinical CTDI mas does not match Data Form scans not axial phantom incorrect phantom incorrect Detector configuration does not match Data Form* Non-chamber holes not filled Table increment not adjusted to give correct pitch if NT different CTDI kvp does not match Data Form *except for scanner limitations Clinical Image Quality Clinical Image Quality Images are scored for Technique parameters Anatomic coverage / display Filming technique (if hard copy) Artifacts Examination identification Examination protocols Radiologist reviewers Technique factor and anatomic coverage guidelines are given for each exam option Artifacts must not hinder interpretation Exam identification If electronic submission, the information should be on the images or readily accessible in the DICOM header by the reviewer Independent of phantom review Pediatric Studies Pediatric Doses It is very important for you to be aware of the differences between the dose observed for the pediatric abdomen portion of the phantom submission (when a child sized phantom is used) and pediatric doses reported on the scanner (which assume an adult sized phantom). The scanner reported doses (i.e. assuming an adult dose phantom) are approximately 2.4 times lower than the same exact x-ray technique would produce in a pediatric dose phantom. In other words, a dose reported on a pediatric abdomen should ideally be reduced by a factor of 3 or more as compared to adult doses. 8

Pediatric Studies Phantom In other words, a dose reported on a pediatric abdomen should ideally be reduced by a factor of 3 or more as compared to adult doses. Applies only when pediatric dose is report in the 32 cm phantom In the 16 cm phantom, doses should be similar It is not essential for each site to purchase the RMI phantom Technologist and physicist QC may be performed using phantom deemed appropriate by the qualified medical physicist RMI phantom MUST be used for accreditation scans Is This Really True? ACR Practice Guidelines and Technical Standards Yes! The website is slow to be updated Documents can get held up in bureaucracy Not all reviewers come up to speed at the same rate Appeal if necessary The following ACR Practice Guidelines and Technical Standards are pertinent to achieving and maintaining CT Accreditation. These guidelines and standards form the basis of the accreditation program. ACR Practice Guideline for Imaging Pregnant or Potentially Pregnant Adolescents and Women with Ionizing Radiation ACR Practice Guideline for the Use of Intravascular Contrast Media ACR Practice Guideline for Performing and Interpreting Diagnostic Computed Tomography (CT) ACR Practice Guideline for the Performance of Computed Tomography (CT) of the Brain ACR Practice Guideline for the Performance of Computed Tomography (CT) of the Extracranial Head and Neck in Adults and Children ACR Practice Guideline for the Performance of Computed Tomography (CT) of the Spine ACR Practice Guideline for the Performance of Computed Tomography (CT) for the Detection of Pulmonary Embolism in Adults Practice Guideline for the Performance of High-Resolution Computed Tomography (HRCT) of the Lungs in Adults ACR Practice Guideline for the Performance of Computed Tomography (CT) of the Abdomen and Computed Tomography (CT) of the Pelvis ACR Technical Standard for Diagnostic Medical Physics Performance Monitoring of Computed Tomography (CT) Equipment ACR Practice Guideline for Communication of Diagnostic Imaging Findings ACR Practice Guideline for the Performance and Interpretation of Cardiac Computed Tomography (CT) Help? The Future of Accreditation Call the ACR (800-770-0145) Theresa Branham ctaccred@acr.org Has individuals who want to help Process is intended to be educational Wants facilities to become accredited CMS can refine any of its requirements at any time Personnel QC Submission ACR tries to post such information on its website Pay attention to any announcements from ACR or CMS Dont be shy - call! (800-770-0145) 9

Score Your Phantom Use one of the following DICOM viewers PC K-PACS http://www.k-pacs.net/ Clear Canvas http://www.clearcanvas.ca/dnn/ Mac Osirix http://www.osirix-viewer.com/ Common Errors Table scan parameters do not match those seen on the images Incorrect detector configuration used (unless limitation of scanner) Artifacts Using small or large FOV Dose too high Phantom CD Preparation Two series are required for each phantom protocol Gammex 464 Must include at least 1 dosimetry image for each protocol Burn 2 IDENTICAL phantom discs for each unit on your application Separate units must be burned onto separate discs Each disc must contain all required images Phantom and Clinical CD Preparation The discs may include an embedded viewer. The following functions are helpful: Easy access to the complete DICOM header Window/level adjustments Distance measurement Region of interest (area, pixel mean and standard deviation) More importantly, be aware of proprietary formats! (Philips isite) Verify that the images can be imported CD Labeling Adult Head CD1 label Gammex CD1 Adult Abdomen Gammex CTAP #xxxx-01 Adult Head CD2 label Gammex CD2 Adult Abdomen Gammex CTAP #xxxx-01 Place ACR label on case Write CD# and CTAP # on disk DO NOT PUT LABEL DIRECTION ON CD! Anything Weird? Call the ACR Sometimes it is impossible to exactly meet the requirements (e.g. dosimetry) Put a note in with submission Explain the issue Describe actions taken Lets reviewers know what s going on Image quality requirements are not negotiable 10

4/28/11 Questions? 11