Pump module pressure measurement system FAQs

Similar documents
Regents of the University of Michigan 5/7/2015. Review pump functionality, sets and accessories Understand key process steps for

A4.7 Management of a totally occluded central catheter and persistent withdrawal occlusion (PWO)

Single-Needle Cross-Over System. Brief description

Multi-Therapy Smart Infusion for Home Care

User Manual. Air-conditioner Controller SB-DN-HVAC (MAC01.331)

SIGMA Spectrum Infusion System Compatible Baxter IV Sets

Easy Machining Center Setup

AUTOMATIC SWINGING DOOR

Calibrating Differential Pressure Sensors

Automatic Fire Sprinkler Submittal Requirements

with FLASH TIPS PLUM A+ INFUSION SYSTEM OPERATING INSTRUCTIONS REFER TO SYSTEM OPERATING MANUAL FOR COMPLETE OPERATING INSTRUCTIONS

Alarm Security Kit - NVR

Element D Services Plumbing

OilonChooser Operating Instructions

CARDIOHELP INTERFACE PRESENTER (CIP) USER GUIDE GENERAL INFORMATION ABOUT THE CIP:

Impact Uni-Vent 754 Portable Ventilator

DUAL PUMPS COMMUNICATIONS CABLE

OPERATOR S MANUAL Vet Pump

Infusovet Veterinary Infusion Pump

Centrifugal Fans and Pumps are sized to meet the maximum

Section Two: Arterial Pressure Monitoring

Using Flow Control with the HEAD Recorder

Smoke Density Monitor application documentation

The Versatile Differential Pressure Transmitter. By David Gunn Honeywell Process Solutions

Example. Fluid Power. Circuits

Product Catalog. WalkMed Infusion, LLC 6555 S. Kenton Street Ste. 304, Centennial, CO walkmed.net. Order Toll Free:

Hydraulic Pipeline Application Modules PSI s Tools to Support Pipeline Operation

Course outline. Know Your Machine From A Programmer s Viewpoint 11 If you ve had experience with conventional (non-cnc) machine tools 11

SD Memory Card Recorder Instructions. ver

Quick Reference Guide VT55 & OBDII. Customer Service TPMS Version: QSG /12

Basic AutoSketch Manual

Date effective: Jan 2005 Date revised: June 2015

CHAPTER II. This chapter contains a simplified description of the pipeline safety requirements. The complete text can be found in 49 CFR Part 192.

Experiment 3 Pipe Friction

Users Guide. By Dr. Don J. Wood Dr. Srinivasa Lingireddy. Featuring the Surge Wizard

Helping to protect you from exposure to hazardous drugs

Windows XP Pro: Basics 1

CRRT with Prismaflex LEADS TO More Flexibility, Ease of Use and Safety

FLOW CALCULATION SOFTWARE. Version 4 User s Manual

Table of contents. Software Installation 2. User Software )AMV Convert Tool ) AMV Player 12

Direct Arterial Blood Pressure Monitoring Angel M. Rivera CVT, VTS (ECC) Animal Emergency Center Glendale, WI March 2003

Creating a Poster Presentation using PowerPoint

Name Partners Date. Energy Diagrams I

535T Window Automation System

HOBO U14 Data Logger User Manual

Maya 2014 Basic Animation & The Graph Editor

Covering the Spectrum

Reviewing documents with track changes in Word 2013

PERIPHERALLY INSERTED CENTRAL CATHETERS (PICC) Fong So Kwan APN, Haematology unit Medical Department, QMH

StructureScan 3D. User Guide ENGLISH.

ÄKTA system Training Guide

How to use the OMEGALOG software with the OM-SQ2010/SQ2020/SQ2040 Data Loggers.

Intravenous Infusion Lines Enhancing Patient Safety During IV Therapy

First Steps with CoDeSys. Last update:

A small, safe, reliable and user-friendly device suitable for inpatient and outpatient use in a variety of applications

JORVET INFUSION PUMP J1060M OPERATOR MANUAL

CADD-legacy PCA pump Model ml Cassette

Programming Logic controllers

The Secret of Hydraulic Schematics. BTPHydraulics

RS Stock No Instruction Manual RS-1340 Hot Wire Anemometer

AKRON EDUCTORS TROUBLESHOOTING GUIDE OPERATION & THEORY OF EDUCTORS GENERAL OPERATING AND MAINTENANCE INSTRUCTIONS

How Reed Switches are used with a Permanent Magnet

Impact Uni-Vent 750 Portable Ventilator

Module 3 Training for the HomeChoice SmartCare (version 10.4) Software. Participant s Workbook

PATIENT GUIDE. Understand and care for your peripherally inserted central venous catheter (PICC). MEDICAL

FRACTION COLLECTOR. Microcomputer Controlled CHF122SC

Stop Alert Flasher with G-Force sensor

Experiment 5: Magnetic Fields of a Bar Magnet and of the Earth

Factors Affecting Blood Pressure. Vessel Elasticity Blood Volume Cardiac Output

Excel -- Creating Charts

Maxi Pressurisation Units (Maxi, Maxi Plus & HP)

Operational Overview and Controls Guide

Nicholas Mezei CSCI 6448 OOA&D Homework #3: Use Cases. Use Case Diagram: Home Security System. Alarm System

Window Glass Design 5 According to ASTM E 1300

Traffic Engineering Management Concepts

MD5-26 Stacking Blocks Pages

How To Control A Car Alarm On A Car With A Remote Control System

Session 7 Bivariate Data and Analysis

Alaris VP Plus Guardrails Volumetric Pump. Directions For Use en

Using Microsoft Word. Working With Objects

CREATE A 3D MOVIE IN DIRECTOR

PLC Control Unit for a CSM-C Steam Compact Clean Steam Generator

INSTRUMENTATION AND CONTROL TUTORIAL 3 SIGNAL PROCESSORS AND RECEIVERS

LS-1 Series Tungsten Halogen Light Sources Installation and Operation Instructions

OREGON FIRE CODE Interpretations and Technical Advisories

Tutorial for Tracker and Supporting Software By David Chandler

1 DESCRIPTION OF THE APPLIANCE

Clark County Fire Prevention

p atmospheric Statics : Pressure Hydrostatic Pressure: linear change in pressure with depth Measure depth, h, from free surface Pressure Head p gh

Understanding Anesthetic Delivery Systems Dean Knoll, CVT, VTS (Anes.) Anesthesia Technician Supervisor University of Wisconsin Madison, WI May 2003

1. The Kinetic Theory of Matter states that all matter is composed of atoms and molecules that are in a constant state of constant random motion

Features. Emerson Solutions for Abnormal Situations

Introduction to Solid Modeling Using SolidWorks 2012 SolidWorks Simulation Tutorial Page 1

OPL BASIC. Dosing System for Professional Laundry machines. Contents

HILLS Series LED Code Pad User Manual

Tube Control Measurement, Sorting Modular System for Glass Tube

Release Notes. Asset Control and Contract Management Solution 6.1. March 30, 2005

Technical Manual. FAN COIL CONTROLLER COOLING or HEATING ANALOG or PWM Art A

TEACHING TOMORROW S INDUSTRY LEADERS TODAY. Manufacturing Small Parts on Machining Centers

Transcription:

Alaris Pump module pressure measurement system FAQs Purpose: The purpose of this document is to provide general information and answers for clinicians to the most frequently asked questions regarding the operation of the Alaris Pump module Pressure Measurement System. 1. Exactly what pressure is being measured by the pump? The Alaris Pump module Pressure Measurement System senses fluid pressure in the IV set at the outlet or patient side of the pump. It estimates this pressure by measuring force on the exterior of the IV tube wall much like we take a patient s pulse by pressing over an artery. If the pump is lower than the patient, the pressure measured will be higher than that at the patient end of the delivery set. 2. How does pressure change when a hard occlusion occurs? If a hard occlusion occurs in the delivery path, pressure will build up in proportion to the flow rate. When multiple sets are connected together or air eliminating filters are present in the patient side of the delivery set, the rate of pressure build up is slowed. 3. How is the hard occlusion pressure limit established for the Alaris Pump module? For each of the profiles or care areas, the hospital specifies one of two possible occlusion detection modes. If the mode specified is Selectable Pressure, then a numeric initial and maximum limit is entered. When Selectable Pressure mode is specified, the maximum limit and default values are set in mmhg from 50 to 525 mmhg. The default must be less than or equal to the maximum. When Pump mode is chosen, for flow rates at/above 30 ml/h, the limit is 525 mmhg; between 10.1 and 29.9 ml/h, the limit reduces in proportion to flow rate, and remains constant at 150 mmhg for flow rates of 10 ml/h and below. In Selectable Pressure for limits below 200 mmhg and in Pump mode for flow rates less than 30 ml/h, the pressure limit is set relative to the pressure measured at the time Start is pressed. 2011 CareFusion All rights reserved. Page 1

In all modes, when the measured pressure reaches the limit an alarm will alert the clinician. This should not be confused with the pressure with which the device pumps (see the following question). 4. What determines how much infusion pressure is created when there is no hard occlusion in the IV path? With no hard occlusion, the pressure measured at the pump is determined by the flow rate, the size and length of the delivery path, including catheter and the viscosity of the fluid. Also, the patient s venous or arterial pressure and the elevation of the patient above the pump add to the total pressure measured by the pump. Movement of the catheter within the vessel may create brief increases in pressure if the outlet of the catheter is pressed against the vessel wall or valves. 5. What should the hard-occlusion pressure alarm limit setting be? While the range of pressure limits depends greatly on the application, we can say this in general: the limit must be set higher than the greatest pressure expected to be produced under normal (non-occluded) conditions. If it is not, there is a likelihood of false alarm. Keep this especially in mind when infusing through small bore catheters such as PICC s (Peripherally Inserted Central Catheters). Higher gauge numbers have smaller internal diameters, these catheters may require an increased downstream occlusion alarm limit to potentially prevent false alarms). 6. What pressure units are used to specify the hard-occlusion alarm limit? The Alaris Pump module Pressure Measurement System uses millimeters of mercury (mmhg) to specify pressure limits. 7. What is the relationship between the pressure units millimeters of mercury (mmhg), inches of water (inch-h 2 O), and pounds per square inch (psi)? 51.7 mmhg = 1 psi = 27.6 inchh 2 O. 8. What are the Pressure Mode selections with the Alaris Pump module? There are two pressure modes available to determine the patient-side occlusion limit with the Alaris Pump module: (Note that occlusion pressure limits are established in the data set {configured for each individual care area or Profile} and are adjustable only in Selectable Pressure mode.) Pump mode, where downstream occlusion alarm threshold is 525 mmhg at flow rates of 30 ml/h or greater. For rates less than 30 ml/h, the occlusion pressure is ratedependent, and the limit is relative to the pressure measured when Start is pressed. This is to provide improved time response to hard occlusions. Selectable Pressure mode, where the downstream occlusion alarm threshold can be adjusted by the user in 25 mmhg increments from 50 mmhg up to the value set as the profile s Maximum Occlusion Pressure up to a maximum of 525 mmhg. Pressure must reach hospital established pressure limit before occlusion warning will alarm. For limits from 50 to 175 mmhg, the pressure limit is set relative to the pressure 2011 CareFusion All rights reserved. Page 2

measured when Start is pressed. For limits of 200 mmhg and above, the limit is absolute. 9. What is Pressure Status Locked? This is a profile system configuration setting that may be established for each individual care area. When it is locked, the Pressure mode cannot be changed. When it is unlocked the user may chose between Pump mode or Selectable mode. For example, if the Pressure Status is unlocked, the clinician may switch between Selectable and Pump modes. If it is locked, the profile designated Pressure Mode is the only one that can be used. 10. What is the default pressure for the Alaris Pump module? With Selectable Pressure Mode in use, applicable to Guardrails Suite MX version 9 software, this is the occlusion limit applied following initial power up or selection of New Patient. The hospital may establish a default pressure setting for each profile as a starting limit with the ability to increase the pressure limit up to the profile s maximum occlusion pressure limit. 11. How can the clinician access the Pressure modes and adjust the pressure limit on the Alaris Pump module? 1) Press Channel Select on the channel you wish to adjust the pressure limit or Pressure mode. 2) Press Options. Note: Current pressure limit listed in mmhg. 3) Under Pressure Limit Selection, press Selectable. 4) Press up or down. Note: With each key press, the pressure limit will increase or decrease in 25 mmhg increments. 5) Once you have reached the desired new pressure limit, press Confirm. Note: Change tick mark on pressure bar graph on main display for the channel. 12. How does the Alaris Pump Module Dynamic Pressure display work and what exactly is being displayed? The Dynamic Pressure display provides a graphic display of the patient side (downstream) pressure and the occlusion alarm threshold. It may be used to view how close the line pressure is to reaching the occlusion alarm threshold and/or whether the pressure is increasing or decreasing. The bar updates approximately once each 10 seconds except when the pump is paused or in alarm. 2011 CareFusion All rights reserved. Page 3

Pressure is displayed using an open rectangular bar that extends from the left to the right side of the screen, representing a range of 0 525 mmhg. The bar is filled in up to a point, representing the downstream pressure reading. In Selectable Pressure mode, for limits below 200 mmhg and in Pump mode at flow rates less than 30 ml/h, the Occlusion Limit marker is placed on the bar relative to the patient baseline pressure. Patient baseline pressure is the pressure measured when Start was pressed. For example, as shown in the figure below, if the pump measures 50 mmhg in the line when Start is pressed, and the limit is 100 mmhg, the Alarm Limit is set at mmhg. At higher limits in Selectable Pressure and in Pump mode at flow rates of 30 ml/h or more, the limit marker is placed directly in relation to the bar graph 0-525 mmhg range. For example, as shown in the figure below, in Pump mode at 50 ml/h, the limit bar will be set to 250 mmhg (about at the horizontal center); this value is relative to the left end of the pressure bar scale. Note: On the device that the numeric VTBI (Volume To Be Infused) is displayed just above the Dynamic Pressure display bar for each channel. There is no relationship between the VTBI and the Dynamic Pressure display. 13. When more than one pump module is running on the Alaris System, why is the initial baseline pressure bar position and limit marker on the Pressure Display sometimes different for each module? The horizontal position of the filled bar and the limit marker are affected by four variables, each of which affects the total fluid pressure sensed in each line. 2011 CareFusion All rights reserved. Page 4

Previous infusion history of the line Fluid container height of each line relative to the pump Disposable pumping segment variables for each line Sensor variables on each channel When any of these variables is different from one channel to the next, then the initial baseline pressure displayed will appear different. See more details in question 12 above. 14. What should the clinician be looking at on the Pressure display to monitor pressure changes in the line? To monitor changes in the patient side pressure, the clinician should observe the range (or space) between the current pressure reading (filled bar) and the occlusion alarm limit marker (shown as the current available pressure range to occlusion alarm in the diagrams in question 12). If this space is getting smaller, the pressure is increasing and vice-versa. When the pressure bar reaches the occlusion alarm limit marker, an occlusion alarm will be triggered. 15. Can the current pressure reading be displayed as an actual numeric value? While the Alaris Pump module software is internally calculating a numeric pressure value at any point in time and displaying it visually as a filled bar on the Pressure display, the actual numeric value is not currently presented on the Alaris PC unit screen. Since the only purpose of the Alaris Pump module Pressure System is detection of hard occlusions, the precision numeric values are not needed. Further, at any time the absolute pressure is subject to fluctuation variables including: Fluid container height of each line relative to the pump Disposable pumping segment variables for each line Sensor variables on each channel Infusion into different IV sites Flow rate of the channel or other channels connected in tandem Catheter dimensions Fluid viscosity Patient vessel pressure Fluid container elevation A numeric pressure value may be misinterpreted given these variables. Instead, as discussed in question 14 above, the intention is to visually track relative pressure changes in the line, particularly whether it is increasing or decreasing. The user may 2011 CareFusion All rights reserved. Page 5

observe whether the space between the filled bar and the occlusion alarm limit marker on the bar graph is increasing or decreasing and take appropriate response such as repositioning line, flushing, etc. Since the full length of the pressure display bar represents 525 mmhg if the bar were half-way filled, the pressure value from that channel would be roughly 250 mmhg. However, this is an estimate, which is subject to the variables above. 16. Shortly after starting an infusion on the Alaris Pump module, why does the Pressure display suddenly shift when there is no apparent change in the infusion conditions? After infusing approximately 0.4 ml after a door closure, or approximately 0.2 ml after a fluid side occlusion alarm, the Alaris Pump module pressure measurement, and thus the location of the filled bar and occlusion alarm limit marker, becomes more accurate. This may show up as a small shift, or series of small shifts in the Pressure display. 17. What should the clinician do if the pressure bar and/or the occlusion alarm limit marker are off the scale to the left and not visible? If the current pressure reading (filled bar) and possibly the occlusion alarm limit marker are off the scale to the left and not visible, depending on the infusion settings, one or more of the following actions will tend to shift them to the right into the visible area: Lowering the fluid container level relative to the pump Lowering the pump relative to the patient level Increasing the occlusion alarm limit if in Selectable Pressure mode 18. Will the occlusion alarm limit ratchet up on the Alaris Pump module? Under some the infusion settings, a limited degree of upward or downward auto adjustment of the pressure limit occurs. If operating in Pump mode with programmed rates less than 30 ml/hr or in Selectable Pressure mode with occlusion alarm settings less than 200 mmhg, the Alaris Pump module sets the baseline pressure each time Start is pressed to begin an infusion, and sets the occlusion alarm limit relative to this baseline. It is therefore possible for the baseline pressure and the occlusion alarm limit to ratchet up over the course of several stops and starts. Using an example of an infusion with a beginning patient side pressure of 20 mmhg and an occlusion alarm limit of 100 mmhg, the baseline pressure would be set at 20 mmhg and the occlusion alarm limit would be set at 120 mmhg. If the pressure rises over several hours to 40 mmhg and the infusion is stopped and then re-started again for some reason, the system would measure the new baseline pressure as 40 mmhg and the alarm limit would be set at 140 mmhg. The opposite is also possible; if the pressure were to drop by 20 mmhg during an infusion and the infusion were stopped and then started again, the baseline pressure and alarm limit would ratchet down. The occlusion alarm limit is only allowed to ratchet up to a maximum setting of 525 mmhg. 2011 CareFusion All rights reserved. Page 6

19. The User Manual states that if the pump restarts more than the allowed time in 10 minutes, it will alarm for a partial occlusion. How does the pump know that a partial occlusion has occurred? In fact, the pump neither knows whether partial or complete or full occlusions have occurred. The pump only knows that the pressure in the line has exceeded the prevailing limit. Since the most likely cause of elevated pressure is occlusion, we conventionally describe the alarms as occlusion alarms. However, other factors such as infusion into an arterial line, an ECMO circuit or other special applications could create pressures above the set limits. In the case of the so called partial occlusion, a likely cause of frequent restarts is a partial occlusion such as would be caused by a catheter intermittently being obstructed due to motion of the patient. However, other sources such as stopcocks, other pumps infusing through a common catheter, etc., may create this alarm. Due to the fact that policies, procedures and current clinical practices of hospitals are extremely specific to the philosophy of the institution, CareFusion does not recommend any values for configuration settings with data set development. We can, however, share what other hospitals anonymously have selected or established for their system configuration settings applicable to specific patient care areas or profiles. If you are interested in reviewing this information, please contact your CareFusion Clinical Operations Project Manager or Clinical Consultant to submit an ad hoc report request. To assist in data set development, these reports can be customized to answer specific questions on system configuration settings. 2011 CareFusion All rights reserved. Page 7