Transition from screen-film to digital radiography: Practical Advice

Similar documents
The Exposure Index and Its Standardization

Best Practices in Digital Radiography

Name: Date: Team: Lab Experiment # 3. Focused Grid Positioning Errors. Computed Radiography and Direct Digital Radiography

Standards for accreditation of educational programs and approval of continuing education courses.

ACR AAPM SIIM PRACTICE PARAMETER FOR DIGITAL RADIOGRAPHY

CONTENT SPECIFICATIONS FOR THE FLUOROSCOPY EXAMINATION

MEDICAL IMAGING. 90/30 Remote Controlled RF System

Digital radiography has become the norm in U.S.

SUBCHAPTER 22 QUALITY ASSURANCE PROGRAMS FOR MEDICAL DIAGNOSTIC X-RAY INSTALLATIONS

Study the Quality Assurance of Conventional X-ray Machines Using Non Invasive KV meter

General requirements. 1. Administrative Controls.

Page: 1 of 6 Page: 1 of 6

CHAPTER 5 QC Test For Radiographic Equipment. Prepared by:- Kamarul Amin bin Abu Bakar School of Medical Imaging KLMUC

REGULATION: QUALITY ASSURANCE PROGRAMS FOR MEDICAL DIAGNOSTIC X-RAY INSTALLATIONS N.J.A.C. 7:28-22

Quality Assurance. The selection of the equipment. Equipment Specifications. Medical Exposure Directive 97/43 Euratom. Quality Assurance Programme

QUALITY ASSURANCE PROGRAMS FOR DIAGNOSTIC RADIOLOGY FACILITIES. (a) Applicability. (b) Definitions

The Practice Standards for Medical Imaging and Radiation Therapy. Radiography Practice Standards

Role of the Medical Physicist in Clinical Implementation of Breast Tomosynthesis

PRACTICAL TIPS IN ENSURING RADIATION SAFETY IN THE USE OF MEDICAL DIAGNOSTIC X-RAY EQUIPMENT

COMMUNICATING RADIATION RISKS IN PAEDIATRIC IMAGING

SECTION 1: REQUIREMENTS FOR CERTIFICATES OF COMPLIANCE FOR CLASSES OF RADIATION APPARATUS

An Overview of Digital Imaging Systems for Radiography and Fluoroscopy

SECTION 1: REQUIREMENTS FOR CERTIFICATES OF COMPLIANCE FOR CLASSES OF RADIATION APPARATUS

Performance testing for Precision 500D Classical R/F System

SECTION 1: REQUIREMENTS FOR CERTIFICATES OF COMPLIANCE FOR CLASSES OF RADIATION APPARATUS

SECTION 1: REQUIREMENTS FOR CERTIFICATES OF COMPLIANCE FOR CLASSES OF RADIATION SOURCES

COST EFFECTIVE FLAT PANEL DIGITAL RADIOGRAPHY UPGRADE SOLUTIONS

Analysis of Causes for Reject X-Ray Films as a Quality Assurance Element in Diagnostic Radiology in Jos University Teaching Hospital(JUTH)

Dental Radiography Core Subject. Digital Radiography

Preface to Practice Standards

Physics testing of image detectors

SECTION 1: REQUIREMENTS FOR CERTIFICATES OF COMPLIANCE FOR CLASSES OF RADIATION APPARATUS

Quality Control and Maintenance Programs

Understanding Digital Modalities: System Integration and Use

Optimization of Digital Industrial Radiography (DIR) Techniques for Specific Applications: An IAEA Coordinated Research Project

Radiation safety in dental radiography

Radiation safety in dental radiography

Surveying and QC of Stereotactic Breast Biopsy Units for ACR Accreditation

Q.A. Collectible. Sponsored by CRCPD s Committee on Quality Assurance in Diagnostic X-Ray (H-7)

CT: Size Specific Dose Estimate (SSDE): Why We Need Another CT Dose Index. Acknowledgements

Diagnostic X-ray facilities as per quality control performances in Tanzania

Computed Radiography: Acceptance Testing and Quality Control

Training needs for professionals in conventional radiology (radiology technicians, physicists, radiologists) joining digital radiology

CT RADIATION DOSE REPORT FROM DICOM. Frank Dong, PhD, DABR Diagnostic Physicist Imaging Institute Cleveland Clinic Foundation Cleveland, OH

Dental Radiography collimator Ionising radiation image radiolucent area radiopaque area controlled zone scatter radiation intraoral

Radiographic Grid. Principles of Imaging Science II (RAD 120) Image-Forming X-Rays. Radiographic Grids

Cabrillo College Catalog

X-Mind. Instinct for perfection

Recognition. Radiation Survey Objectives. Objectives. Part 1 Documentation Radiation Source Survey Objectives Radiation Detectors Techniques

X-ray (Radiography) - Bone

MDCT Technology. Kalpana M. Kanal, Ph.D., DABR Assistant Professor Department of Radiology University of Washington Seattle, Washington

X-ray reject analysis in Tikur Anbessa and Bethzatha Hospitals

PERFORM-X DIGITAL X-RAY SYSTEM PERFORM-X. with ceiling mounted tube support PERFORM-X

Recommendations for a technical quality control program for diagnostic X-ray equipment

Assessing Radiation Dose: How to Do It Right

Application Note. ipix A Gamma imager to support various applications

Radiation Safety Characteristics of the NOMAD Portable X-ray System

Compliance Guidance for FLUOROSCOPIC QUALITY CONTROL

Requests. Who requests a DXA scan? DXA. DXA Technology. Adequate Clinical Information

EPEX. The Hologic EPEX General. Direct-to-Digital System for General Radiography. Versatility for a complete range of general radiographic exams

INTENSITY MODULATED RADIATION THERAPY (IMRT) FOR PROSTATE CANCER PATIENTS

IAC Ch 41, p.1. Procedure means a stereotactically guided breast biopsy performed on a patient for diagnostic purposes.

Thinking ahead. Focused on life. REALIZED: GROUNDBREAKING RESOLUTION OF 80 µm VOXEL

IAEA-TECDOC-1447 Optimization of the radiological protection of patients: Image quality and dose in mammography (coordinated research in Europe)

European Academy of DentoMaxilloFacial Radiology

Update on ACR Digital Mammography QC Manual

Running head: OBESITY AND MEDICAL IMAGING 1. Obesity and Medical Imaging. November 13, 2012

A Comparison of Screen/Film and Digital Imaging: Image Processing, Image Quality, and Dose

Digital Dental X-ray. Midmark Animal Health Products

X-ray (Radiography) - Abdomen

5.2 ASSESSMENT OF X-RAY TUBE LEAKAGE RADIATION AND X-RAY TUBE OUTPUT TOTAL FILTRATION

Training needs for professionals in conventional radiology (radiology technicians, physicists, radiologists) joining digital radiology

SUMMARY OF CURRENT UK LEGISLATION AND GUIDELINES

The Practice Standards for Medical Imaging and Radiation Therapy. Sonography Practice Standards

CLINICAL PERFORMANCE OBJECTIVES

Image Quality and Radiation Dose for Intraoral Radiography: Hand-Held Held (Nomad), Battery Powered

Scan Time Reduction and X-ray Scatter Rejection in Dual Modality Breast Tomosynthesis. Tushita Patel 4/2/13

MEDICAL IMAGING SCIENCES Information Session

EPA Protocols: Compliance testing of dental X-ray apparatus used for plain radiography

GE Healthcare. DoseWatch. Gathering Radiation Dose Data

Competency Profile. Radiological Technology

HIGH PERFORMANCE MOBILE SURGICAL C-ARM KMC-950

Diagnostic Exposure Tracking in the Medical Record

Parkway College of Nursing and Allied Health School of Nursing and Allied Health. BSc (Hons) Diagnostic Radiography and Imaging Module Synopses

Radiation Protection Series

The Field. Radiologic technologists take x-rays and administer nonradioactive materials into patients' bloodstreams for diagnostic purposes.

Application of Digital Radiography for the Detection and Classification of Pneumoconiosis

The role, responsibilities and status of the clinical medical physicist in AFOMP

Froedtert Hospital School of Radiology Curriculum Analysis Grid. Clinical Practice

SPOKANE COMMUNITY COLLEGE RADIOLOGY TECHNOLOGY MASTER PLAN OF EDUCATION

Professional Education in CBCT

Position Classification Standard for Diagnostic Radiologic Technologist Series, GS-0647

Comparing dose and image quality of a conventional 400 speed class film/screen system with a CR system by assessing pediatric chest examinations

Measuring Size and Distance in Digital Mammography

I-Max Touch Range. PAN / CEPH / 3D digital panoramic unit. Evolutive 3 in 1 panoramic unit

The convenient and compact CR system for veterinary practices

Patient-centered CT imaging: New methods for patient-specific optimization 1 of image quality and radiation dose

Medical imaging monitors specification guidelines

X-ray (Radiography) - Chest

Prepublication Requirements

Transcription:

KH Ng Transition from screen-film to digital radiography: Practical Advice Kwan-Hoong Ng, PhD, DABMP Department of Biomedical Imaging & Medical Physics Unit University of Malaya Kuala Lumpur

KH Ng 1. Introduction 2. Issues in Digital Imaging 3. Image Acquisition 5. Summary

1. Introduction KH Ng

KH Ng While digital techniques have the potential to reduce patient doses, they also have the potential to significantly increase them. - ICRP 93 Editorial

KH Ng This is a technology that is advancing rapidly and which will soon affect hundreds of millions of patients. If careful attention is not paid to the radiation protection issues of digital radiology, medical exposure of patients will increase significantly and without concurrent benefit. - ICRP 93 Editorial

KH Ng Most principles for dose reduction for screen-film radiography, especially justification, are relevant to digital systems However, in digital systems different scenarios apply for dose reduction and optimisation compared with screen-film radiography

Screen-Film Radiography Digital Radiography Recording Displaying Archiving Major limitation is film Digital detector Display Monitor PACS server Optimize each component independently KH Ng

KH Ng The diagnostic information provided by modern digital detectors can be equal or superior to conventional screen-film systems, with comparable patient doses. Digital imaging has practical technical advantages compared with film techniques, e.g. wide contrast dynamic range, post-processing functions, multiple image viewing options, and electronic transfer and archiving possibilities.

2. Issues in digital imaging KH Ng

KH Ng This is a technology that is advancing rapidly and which will soon affect hundreds of millions of patients. If careful attention is not paid to the radiation protection issues of digital radiology, medical exposure of patients will increase significantly and without concurrent benefit. [We are witnessing the consequences now] - ICRP 93 Editorial

KH Ng

Ng BMJ 2006 Digital imaging brings benefits but also demands changes in ways of working Digital x ray imaging involves issues cost and productivity need to acquire new skills radiation doses overuse image quality It is easy to delete digital images, and repeat exposures normally go undetected.

Different imaging tasks require different levels of image quality. e.g. the follow-up examination for a fracture does not require the same image quality as that required for its diagnosis. Ng BMJ 2006

Current issues with clinical digital radiography: Human Factors Inappropriate exposure Increase in number of examinations No (inadequate) collimation Image quality is not compatible with imaging task

Inappropriate exposure With digital systems, an overexposure can occur without an adverse impact on image quality. Overexposure may not be recognised by the radiologist or radiographer. In conventional radiography, excessive exposure produces a black film and inadequate exposure produces a white film, both with reduced contrast. In digital systems, image brightness can be adjusted post processing independent of exposure level.

Increase in number of examinations In several U.S. hospitals the number of examinations per in-patient day increased by 82% after a transition to film-less operation. Outpatient utilization (i.e. the number of examinations per visit) increased by 21% compared with a net decrease of 19% nationally at film-based hospitals. Reiner et al. Radiology. 2000 ;215(1):163-7

3. Image Acquisition Positioning Collimation & Electronic Masking Anatomic Side Makers Exposure Technique (kvp/mas) Anatomically Programmed Radiography (APR) & Exposure Technique Chart Automatic Exposure Control (AEC) Grid Shielding

Positioning

Positioning errors is the number one reason for repeat digital radiography examinations. Increase in exposure latitude in digital radiography leads to overall reduction in repeats. Cause of most repeat imaging has shifted to positioning errors. Inaccurate positioning of the part relative to the image receptor, along with a poorly collimated exposure field, often results in poor quality digital images.

For peadiatric patients - Immobilization is a critical component of positioning that helps to prevent retakes. A best practice use immobilization devices when needed prevent repeat exposures by appropriately positioning the patient.

Collimation & Electronic Masking

No collimation Or too broad the collimation size Post-exam collimation (cropping)

No collimation Apparently Perfect radiographs of the paranasal sinuses with post exam cropping Courtesy Dr. Evelyn Ho, SDMC

No collimation I cant believe it! A series of radiographs which were supposed to be for the paranasal sinuses (yellow collimation lines) but instead practically the whole head was x-rayed! Courtesy Dr. Evelyn Ho, SDMC

No collimation What happens to collimation? Lumbar spine radiography - poor collimation practices after implementation of digital technology Zetterberg and Espeland Brit J Radiol, 84 (2011), 566 569

No collimation Lumbar spine radiography - poor collimation practices after implementation of digital technology Digital Analog (Film) Irradiated field outside ROI Total field size 61.7% 42.4% 791 cm 2 541 cm 2 Zetterberg and Espeland Brit J Radiol, 84 (2011), 566 569

No collimation A survey of 450 technologists by the American Society of Radiologic Technologists (ASRT) revealed that half of the respondents used electronic cropping after the exposure Pediatric Radiology, 2011, 41:5, 602-610

No collimation Radiation Protection in Paediatric Radiology Radiation protection in digital radiography IAEA 28 Poor collimation Large part of the body is irradiated Not seen on digitally cropped image

The Art of Collimation Some machine has automatic collimation to detector size not practical for smaller size patient (e.g. Asia). Omit the use or disable this feature. Learn to use anatomical landmarks for collimation guide Always refer to old x-ray image when doing follow-up examination, helps to define collimation size Correct placement of anatomical markerhelps to reduce collimation size

The Dilemma of Collimation Radiographers have dilemma in tight collimation- 1. Fear of cutting the ROI- hence: Basic rule better to be large than cut the ROI. 2. The gap between theory and practice e.g. Lumbar spine In theory: ROI to include T12 to S1 In practice: Doctor may want to see from lower thoracic spine to coccyx.

The Dilemma of Collimation Radiographers have developed the practice (habit) to include more anatomical structures (especially in trauma cases) to avoid repeat or retake radiographs due to in accurate/ less precise request from doctors.

Anatomical Side Marker Most radiographic technique books emphasise that it is mandatory to place an anatomical side marker for every radiograph. Radiographers are taught the correct way of placing the marker in practice.

Collimation & anatomical marker Sometimes due to the ineffective placement of anatomical side marker, collimation size is compromised to include the marker. Strategically place the anatomical side marker can help to improve collimation.

Collimation & anatomical marker R R R

Standardized Exposure Technique Department cannot depend solely on vendors to set exposure standards Setting policies and protocols ensure consistency in image quality and dose optimization Radiographer with sound knowledge should work together with vendor during automatic exposure setting.

AEC Radiographers must have sound knowledge of how AEC works to be effective in dose optimization. Selection of correct sensor field Accurate positioning of ROI to sensor field Take note on examination part- smaller than sensor field should not use AEC Total knee replacement- try not to use AEC over exposure may happen if metal part is placed over sensor field.

AEC The wide exposure dynamic range of such systems may have the disadvantage that, if the X-ray generator AEC develops a fault or the output calibration drifts, the dose increase/decrease may not be identified readily. Also, the wide exposure dynamic range means there is significant potential for the initial set-up of such systems not to be optimised.

AEC Calibration Physicists have to be sure that the AEC is calibrated accordingly to digital system and not based on screen/film To properly exposure digital radiographs, recalculate the kvp correction curve to AEC system to response correctly to the receptor characteristics is necessary.

Anatomically Programmed Radiography & Exposure Technique Charts

Anatomically Programmed Radiography (APR) A system of preprogrammed exposure technique settings that is organized by position and procedure and set through the control panel of the radiography unit. APR settings provide recommendations for small, medium and large adult patient sizes and include a combination of AEC and manual exposure technique settings.

Exposure Technique Charts To standardize exposure techniques according to patient size, procedure and position. provide consistent and accurate radiation exposure to the image receptor. Establish department standards and eliminates confusion and concern regarding appropriate use of kvp, ma, grid use and SID. Allow radiologists and radiographers to work together to set an acceptable level of radiation exposure that provides diagnostic quality images within the ALARA principle.

Exposure Technique Charts The charts don t take the place of radiographers carefully assessing individual patient pathology, condition and unusual circumstances because the charts are designed for the average or typical patient. Exposure technique charts should be monitored and revised continuously to ensure these techniques are producing diagnostic images within the ALARA principle.

A thorough exposure technique chart includes these variables for each x-ray tube: Backup exposure time or mas (if set). Source-to-image receptor distance (SID). kvp. Focal spot size. ma (if set). Use of a grid and the grid ratio. AEC detector(s). Acceptable exposure indicator range.

A best practice in digital radiography is to use exposure technique charts that are continuously improved and applicable to a wide range of patient sizes.

Exposure (sensitivity) Index Each image should ideally have an associated number to indicate the level of exposure to the detector. Currently all digital systems have a exposure (sensitivity) index which is related to detector exposure. Once digital radiography systems are in use, the constancy of applied exposure factors should be monitored on a regular basis.

List of terms for exposure indices for various digital systems and their relationship to traditional dose measure (in µgy). In the 2 nd column the proposal for an international standardisation is detailed (courtesy of Ulrich Neitzel, Philips Medical Systems, Hamburg). Uffmann et al E J Radiol 72 (2009) 202 208

Am J Roentgenol. 2012 Dec;199(6):1337-41 Both the IEC (IEC standard 62494-1) and the AAPM (AAPM Task Group 116) have developed similar standards for monitoring exposure in digital radiography to eliminate proprietary and confusing terminology. Radiologists and technologists will need to learn three new terms - exposure index, target exposure index, and deviation index - to understand the new standards.

Grid Digital imaging is more sensitive to lowlevel radiation exposure, thus the need to use antiscatter grids. It results in increase in patient dose. Grids are appropriate for anatomy that is > 10 cm thick & for kvp settings of > 70.

Shielding Creative shielding

Shielding

KH Ng Educate, Educate, EDUCATE Train, Retrain, Train, RETRAIN Collimation is urgently needing attention

KH Ng Thanks to Lai-Kuan Chan, MSc (Health Science) Napapong Pongnapang, PhD For their advice