Fiske Model 210 Micro-Osmometer Service Manual 2105SM Rev23 012010
Table of Contents 1. Introduction Specifications.....................7 General Description and Purpose.....8 Service Assistance.................8 Safety/Regulatory Information.......9 Symbol Conventions...............9 General Cautions..................9 Service & Maintenance Cautions....10 Hazardous Material Cautions.......11 Symbol Conventions..............11 FCC Requirements................11 Regulatory Notices...............12 2. General Overview General Overview................17 Systems Overview................18 Power Supply....................18 Keypad.........................18 Back Panel I/O Board Description...18 Application Board Description......18 Central Processor Unit (CPU Board Description...................18 210 Board Set Description..........18 Instrument Software Update........18 Replacement Parts................19 Supplies & Accessories............19 3. Maintenance Maintaining the Instrument.........23 Solenoid Cleaning Procedure.......25 4. Troubleshooting Introduction.....................29 Diagnostic Menu.................29 A/D Tests.......................29 Probe Bin Test...................29 Solenoid Test....................29 Display/Print Test................29 Key/Beeper Test..................29 Barcode Test....................29 Exit............................30 Troubleshooting Table.............31 5. Replacement Instructions Block Probe (110340)......21P110340 Central Processor Board and Application Board Set (210620R).....21P0620R Cooling Assembly (210335R/ 210336R)...............21P0335R Cooling Fan (210330R).....21P0330R Display Cable (210151R)....21P0151R Display Module (210800R)..21P0800R Fuse (70013R)...........21P70013R Keypad (210511R).........21P0511R NVRAM Battery (71027R)....71P27R Power Entry Assembly (210140R).......................21P0140R Power Supply (210950R)....21P0950R Sample Probe Cable Assembly (240725R)............21P240725R Sample Probe (242700).....21P242700 6. Printer User Manual 7. Schematics PCB100 Back Panel I/O Board Assembly Drawing PCB100 Back Panel I/O Board Schematic PCB140 Back Panel I/O Board Assembly Drawing PCB140 Back Panel I/O Board Schematic PCB115 Application Board Assembly Drawing PCB115 Application Board Schematic PCB117 Application Board Schematic PCB120 CPU 80C186 Assembly Drawing PCB120 CPU 80C186 Assembly Schematic PCB122 CPU 80C186 Assembly Schematic 210000 210 System Interconnect APPENDICES A: Event Record B: Symbol Definitions C: Product Disposal and Recycling D: Service Log Index 3 Fiske Micro-Osmometer Service Manual
Notes: Fiske Micro-Osmometer Service Manual 4
1. Introduction
Model 210 Micro-Osmometer Specifications Electrical: Voltage: 100 to 240 VAC Room Humidity: 5 to 80% relative humidity; (non-condensing) Frequency: 50 / 60 Hz Fuses (2): 250V time delay (Type T): 1.25-Amp Drift: Less than 1 digit per month plus 1 digit for every 9 F (5 C) ambient temperature change. Power: 60 Watts Start-up Time: 30 seconds from power on Memory Backup: integral lithium cell; 10 years typical Sample Volume: 20 µl Sample Capacity: single sample Test Time: 90 seconds per sample Dimensions inches centimeters Width: 12.0 30.5 Depth: 9.5 24.1 Height: 12.0 30.5 Readout: Units: Range: Linearity: Resolution: 2-line by 24 character vacuum fluorescent display mosm/kg H2O 0 to 2000 mosm/kg H2O <1% from straight line over calibration range 1 mosm/kg H2O Weight pounds kilograms Net: 13.6 6.2 Shipping: 20.5 9.1 Warranty: Certification: One-year limited warranty on workmanship and all parts except glass, plastic and parts warranted by their makers. Repeatability: ±2 mosm/kg H2O (1 σ) between 0 and 400 mosm/kg H2O ±0.5% (1 σ) between 400 and 2000 mosm/kg H2O Communications: Serial Port: DTE RS-232C Printer Port: IEEE 1284 Parallel Port Standard Barcode Port: Serial Storage Temperature: -40 to +113 F (-40 to +45 C) Refer to Regulatory Notices (see page 11) for applicable standards. Installation Class: Over-Voltage Category: Pollution Degree: 2 Moisture Protection: Class I II Ordinary (IPX0) Operating Temperature (ambient): 64 to 95 F (18 to 35 C) 7 Fiske Micro-Osmometer Service Manual
General Description and Purpose Fiske Osmometers are devices for the measurement of the osmotic concentration of aqueous solutions with freezing points between 0 and -6 C. Some examples are serum, urine, parenteral infusion solutions, laboratory control solutions and fixatives. Application information is available upon request. The Service Manual This Service Manual contains: instructions for performing repairs. guides for troubleshooting operational and mechanical problems. circuit system illustrations. information about replacement parts and service repairs. This manual is designed to assist service technicians, and does not imply a license to perform repairs without proper qualifications. Service Assistance To contact Fiske Associates Product Service: Call (800) 225-1002 (toll-free within the USA and Canada; after normal business hours, dial extension 2191) Call +US (781) 320-5656 (elsewhere) Or fax (781) 320-3669 When calling Fiske Associates, have the model and serial number from the label on the back of the instrument, and a description of the problem. Use a telephone as close to the instrument as possible in order to facilitate making recommended diagnostic checks. A service technician may assist in making minor repairs over the phone, providing you with recommended parts (or part numbers), or may issue an authorization (RMA) to ship the instrument for factory repair. The RMA procedure is as follows: 1. Contact Fiske Associates to obtain an RMA. 2. Pack and ship the instrument in its original carton. Do not send instrument supply items. 3. Prepay shipment to the factory. Fiske Associates cannot accept collect shipments. Insure the shipment or accept the damage risk. Fiske Micro-Osmometer Service Manual 8
Safety / Regulatory Information To reduce the risk of bodily injury, electric shock, fire, and damage to your instrument, please read and observe the following precautions. If the product is used in a manner not in accordance with the equipment design, operating instructions or manufacturer's recommendations, the operation of the product may be impaired to the extent that a safety hazard is created. Do not attempt to perform electrical work if you are not fully qualified. This manual is not a substitute for electrical training. Symbol Conventions The exclamation point within an equilateral triangle is intended to alert the user to the presence of important operating and maintenance (servicing) instructions in the literature accompanying this product. The lightning flash with arrowhead symbol within an equilateral triangle is intended to alert the user to the presence of uninsulated dangerous voltage within the product's enclosure that may be of sufficient magnitude to constitute risk of electric shock to persons. The static symbol within an equilateral triangle is intended to alert the user to the presence of internal components that could be damaged by static electricity. This static symbol is intended to alert the user to the presence of a specific component that could be damaged by static electricity. This symbol indicates the presence of alternating current (AC). This symbol indicates the presence of a fuse. This symbol indicates the presence of protective earth ground. This symbol indicates the power is ON. This symbol indicates the power is OFF. General Cautions This product should be operated only with the type of power source indicated on the product s electrical ratings label. Refer to the installation instructions included with the product. If the power cord provided is replaced for any reason or if an alternate cord is used, the cord must be approved for use in the local country. The power cord must be approved for the product s listed operating voltage and be rated at least 20% greater than the ampere ratings marked on the product s electrical ratings label. The cord end that connects to the product must have an IEC 60320 connector. Plug the product into an approved grounded electrical outlet. Do not disable the power cord s grounding plug. If an extension cord or power strip is used, make sure that the cord or strip is rated for the product, and that the total ampere ratings of all products plugged into the extension cord or strip do not exceed 80% of the cord s or strip s rating limit. Route power cords so that they will not be walked on, tripped on, or pinched by items placed upon or against them. Pay particu- 9 Fiske Micro-Osmometer Service Manual
lar attention to the plug, electrical outlet, and the point where the cord exits the product. Do not pull on cords and cables. When unplugging cords or cables, grasp the corresponding connector. Do not install or use this product in any area subject to extreme short-term temperature variations, or locations that exceed the specified operating environment temperatures. Never use this product in a wet area. To avoid injury or fire hazard, do not operate this product in an explosive atmosphere. Do not install or use the product on an unstable, non-level work surface. Do not operate this product with the covers removed or unsecured. Service & Maintenance Cautions Unplug the power cord prior to opening or removing covers, or else you may be exposed to electric shock, excessive temperatures, or mechanical hazards. Performing service or maintenance not detailed in the User s Guide, with or without this Service Manual, should only be done by a qualified service technician. Never restrict airflow into or out of the product. Occasionally, check the air vents for blockage. Wipe the exterior of the product with a soft, damp cloth as needed. Using cleaning products other than those specified, may discolor or damage the finish. If the product requires service for any of the following reasons, unplug the product from the electrical outlet and refer service to a qualified service technician. - The power cord, extension cord, power strip or power input module is damaged. - Liquid has been spilled into the interior of the product. - A foreign object has fallen into the product. - The product has been dropped or damaged by a falling object. - There are noticeable signs of overheating or a burning odor. - The product does not operate normally when you follow the operating procedures. - The main supply fuse(s) or any internal fuse(s) continually fail. A discharge of static electricity from contact with the human body or other conductor may damage system boards or static sensitive devices. Never perform internal maintenance without following recommended static protection procedures. The product is equipped with operator accessible fuses. If a fuse blows, it may be due to a power surge or failure of a component. Replace the fuse only once. If the fuse blows a second time, it is probably caused by failure of a component part. If this occurs, refer service to qualified service personnel. Always replace the fuse with one of the same rating, voltage, and type. Never replace the fuse with one of a higher current rating. When servicing the product, use only factory-specified parts. WARNING: When returning this product for service, or shipping this product to a second location, remove all hazardous specimens and decontaminate the product before packaging for shipment. If the product cannot be decontaminated, consult with your shipping agent on appropriate packaging and marking. Fiske Micro-Osmometer Service Manual 10
Hazardous Material Cautions WARNING: Handle all biohazardous materials according to established good laboratory practices and follow your institution s exposure control plan. Persons handling human blood and body fluid samples must be trained in blood-borne hazards and observe universal precautions. Universal precautions is an approach to infection control, where all human blood and body fluids are treated as if known to be infectious. Use personal protective equipment such as gloves, gowns, etc., to prevent exposure. Store biohazardous materials in regulated waste containers and dispose of these materials in a safe and acceptable manner that is in compliance with all country, state and local requirements. If a biohazardous material is spilled on or inside the equipment, decontaminate the equipment using a 1% bleach solution, or as outlined by those policies and procedures established within your institution. To avoid injury or fire hazard, do not operate this product in an explosive atmosphere. Symbol Conventions This symbol indicates conformity to relevant European directives. This symbol indicates the product was tested to conform to relevant Canadian and U.S. safety standards by Intertek Testing Services NA, Inc. The ETL mark is approved in the United States as a Nationally Recognized Testing Lab (NRTL) by OSHA, and in Canada by the Standards Council of Canada. In Vitro Diagnostic Medical Device complying with EU Directive 98/79/EC. EXAMPLES ONLY This symbol, depending upon its location, indicates the assembly, component, and/or product are free of lead. In respect to its inclusion on a circuit component or assembly, the following must be adhered to when servicing. Failure to do so may result in defects caused by dissimilar metals. 1. Do not repair or modify using a lead-containing solder. 2. Do not repair or modify using any soldering tools that have been in contact with lead-containing solders. 3. Do not repair or modify with any lead-free solder other than that specified by the manufacturer or as indicated by the symbology of standard IPC-1066. 4. Do not repair or modify using any soldering tools that have been in contact with any other incompatible lead-free solders. 5. Contact factory service for further information. The number in this symbol indicates the type of lead-free solder used in the assembly of printed circuit boards, per standard IPC-1066. Repair or modification must be performed using the same or compatible type lead-free solder and matching tools. Failure to do so may result in defects caused by dissimilar metals. Contact factory service for further information. FCC Requirements WARNING: Changes or modifications to this unit not expressly approved by Advanced Instruments could void the user s authority to operate the equipment. This equipment has been tested and found to comply with the limits for a Class B 11 Fiske Micro-Osmometer Service Manual
digital device, pursuant to Part 15 of the FCC Rules. These limits are designed to provide reasonable protection against harmful interference in a residential installation. This equipment generates, uses, and can radiate radio frequency energy and, if not installed and used in accordance with the instructions, may cause harmful interference to radio communications. However, there is no guarantee that interference will not occur in a particular installation. If this equipment does cause harmful interference to radio or television reception, which can be determined by turning the equipment off and on, the user is encouraged to try to correct the interference by one or more of the following measures: - Re-orient or relocate the receiving antenna. - Increase the separation between the equipment and receiver. - Connect the equipment into an outlet on a circuit different from that to which the receiver is connected. - Consult the dealer or an experienced radio TV technician for help. Regulatory Notices This product has been designed and manufactured in accordance with U.S., Canadian, and European regulatory requirements as outlined below. Modifications made to this product that are not expressly approved in writing by the manufacturer will void the user s authority to operate this product, previously issued factory approvals, and the user s rights under the warranty. The distributor or dealer may have applied additional local, national, or international approvals to this product. Consult the distributor or dealer for more information and documentation. Connections to this product must be made with shielded cables. Use of non-shielded cables may violate RFI/EMI limits. Fiske Micro-Osmometer Service Manual 12
Regulatory approval type Description Regulatory approval type Description U.S. Safety Canadian Safety EC Declaration of Conformity - EMC EC Declaration of: Conformity - Low Voltage This product has been listed by ETL testing laboratories as being in compliance with the requirements of UL 61010-1, 1st Edition, "Electrical Equipment for Laboratory Use; Part 1: General Requirements". The "US" in the lower right of the ETL mark demonstrates this listing. This product has been listed by ETL testing laboratories as being in compliance with the requirements of CAN/CSA C22.2 No.1010.1-92, "Safety Requirements for Electrical Equipment for Measurement, Control and Laboratory Use; General Requirements"; Including Amendment Two. The "C" in the lower left of the ETL mark demonstrates this listing. This product meets the intent of Directive 89/336/EEC for Electromagnetic Compatibility. Compliance was demonstrated using the following standards, as listed in the Official Journal of the European Communities: Consult the Declaration of Conformance certificate shipped with the product for the latest update. EN 61326: 1998, Group 1, Class B, "Electrical Equipment for Measurement, Control, and Laboratory Use". This product meets the intent of Directive 2006/95/EC, the Low Voltage Directive. Compliance was demonstrated using the following standards as listed in the Official Journal of the European Communities: Consult the Declaration of Conformance certificate shipped with the product (if required) for the latest update. EN 61010-1 (1993), "Safety Requirements for Electrical Equipment for Measurement, Control and Laboratory Use - General Requirements"; Including Amendment Two (1995). EC Declaration of Conformity - WEEE EC Declaration of Conformity - IVD This product meets the intent of Directive 2002/96/EC as amended by 2003/108/EC for Waste Electrical and Electronic Equipment (WEEE). Consult the Declaration of Conformance certificate shipped with the product (if required) for the latest update This product meets the intent of Directive 2004/108/EC for In Vitro Diagnostic Medical Devices. Consult the Declaration of Conformance certificate shipped with the product (if required) for the latest update. FCC - Part 15 This device complies with Part 15 of the FCC Rules. Operation is subject to the following two conditions: (1) this device may not cause harmful interference, and (2) this device must accept any interference received, including interference that may cause undesired operation. Canadian ICES-003 RoHS This Class B digital apparatus complies with Canadian ICES-003. Cet appareil numérique de la classe B est conforme à la norme NMB-003 du Canada. This product meets the intent of Directive 2002/95/EC for the Restriction of Use of Certain Hazardous Substances. This product is exempt under the terms of Article 2, paragraph 1 due to its listing in Category 8 and/or 9 of Directive 2002/96/EC. 13 Fiske Micro-Osmometer Service Manual
Notes: Fiske Micro-Osmometer Service Manual 14
2. General Overview
General Overview A B F C N D M E L K J I H G A. Keypad (210511) B. Fluorescent Display Module (330800) C. Display Cable (210151) D. PC Board Set (210620) E. Printer Cable (210154) F. PCB Mounting Bracket (210125) G. Serial Cable (210153) H. Power Entry Assembly (210140) I. DC Power Harness (210150) J. Drive Harness (210152) K. Power Supply (210950) L. Instrument Foot (3D2123) M. Cooling Fan Assembly (330330) N. Sample Handling Assembly (210310) 17 Fiske Micro-Osmometer Service Manual
Systems Overview Power Supply - Part No. 210950 The power entry assembly interfaces the AC voltage presented to the instrument with the power supply. The power supply provides +15VDC to the main printed circuit board. The thermoelectric and solenoid operate at +15VDC. The fan operates at +12VDC. Keypad - Part No. 210511 The keypad provides a two-button interface between the user and the instrument. The second line of the display is used to indicate the function of each key. The software changes the button description on the second line of the display. Back Panel I/O Board Description - Part No. PCB100/PCB140 This board contains all external connections for the printer, RS-232, and barcode ports. Application Board Description - Part No. PCB115 All input/output operations are performed by this printed circuit board. The board provides local regulation for the analog and digital portions of the instrument. There are two 24 bit sigma-delta analog-to-digital converters (ADC) that interface the sample and block probes to the microprocessor. The board also provides the interface to the keypad, display, and Back Panel I/O Board. Central Processor Unit (CPU) Board Description - Part No. PCB120 The microprocessor is a 16 bit 80C186EB, a member of the 80X86 Intel family of microprocessors. The microprocessor runs at 32MHz and has an address space of 1 megabyte. The microprocessor executes the program stored in the 256 kilobytes of Flash EPROMs. The board uses a real-time clock with battery-backed RAM for storage of operating parameters, and contains 256 kilobytes of static RAM. 210 Board Set Description - Part No. 210620 This board set is made up of the following: PCB115, PCB120, and software. Instrument Software Update Your instrument software contained in two flash EPROMs on the main processor board may be updated via the RS-232 port. Such software updates are made available as new features and other improvements are added during the life cycle of the instrument. For information on the availability of software updates, contact Fiske Associates or an authorized distributor. For directions on performing the RS-232 port software upgrade, consult the documentation supplied with the upgrade package. Fiske Micro-Osmometer Service Manual 18
Replacement Parts Block Probe Replacement Kit......110340 PC Board Set..................210620R Cooling Assembly......210335R/210336R Cooling Fan Replacement Kit.....210330R Fluorescent Display Module Replacement Kit.............210800R Display Cable..................210151R 5/16" Hex Nut Driver............3LA820 Fuses (2 required): 1.25-Amp Time Delay Fuse.....70013R Keypad, Model 210.............210511R DC Power Harness...............210150 Drive Harness...................210152 Model 210 Service Manual........2105SM NVRAM Battery................71027R Power Entry Assembly...........210140R Power Supply..................210950R AC Ground Wire.................210156 Serial Cable.....................210153 Printer Cable....................210154 Sample Probe Cable Assembly....240725R Sample Probe Replacement Kit.....242700 Impactor Solenoid Assembly......3M2350 Spring.........................210227 Gasket.........................210306 Supplies & Accessories Test Kit (250 sample tubes and probe cleaners)...............110825 50 mosm/kg Calibration Standard (10 2-mL ampules)............00a005 850 mosm/kg Calibration Standard (10 2-mL ampules)............00a085 2000 mosm/kg Calibration Standard (10 5-mL ampules)............3la201 Accuref TM 290 Reference Solution (10 2-mL ampules)............00a030 Protinol Near-Serum Reference (3 levels)...................3ma028 Renol Urine Osmolality Controls (2 levels)....................3la085 Osmometer Linearity Set (5 levels)....................3la028 Thermal Printer with Interface Cable, Operation Manual, Thermal Paper Roll, and Printer Power Supply 100-120 VAC.........210555_NA 230 VAC.............210555_EU For current Thermal Printer 210555: Printer Paper (5 rolls)..........3d3835 For Dot Matrix Printer 210550: Printer Ribbon (3 cartridges).....330542 Printer Maintenance Kit........FL0425 Printer Paper (5 rolls)..........330830 20 µl Sampler..................240820 Sampler Tips (50 tips)............240821 Bar Code Scanner................330016 Serial Port Interface Cable..RS232-CABLE 3 Meters.....................330053 6 Meters.....................330056 9 Meters.....................330059 User s Guide......................2105 Service Manual.................2105SM How to order: To order parts, supplies and accessories, contact the Fiske Associates Customer Service. 800-225-1002 (toll-free within the USA and Canada) +US 781-320-5656 (elsewhere) +US 781-320-3669 (fax) 19 Fiske Micro-Osmometer Service Manual
Notes: Fiske Micro-Osmometer Service Manual 20
3. Maintenance
Maintaining the Instrument 1. Cleaning Air Vents Examine the air vents on the underside and rear of the instrument to ensure that they are unobstructed by dust or debris. 2. Cleaning the Instrument Exterior Use a mixture of warm water and soap. A mixture of Isopropyl Alcohol may also be used to further decontaminate the exterior. Do not use abrasive cleansers or scouring pads, as they may mar the surface. Do not allow liquid to enter the cooling chamber or other portions of the instrument. 23 Fiske Micro-Osmometer Service Manual
Notes: Fiske Micro-Osmometer Service Manual 24
Solenoid Cleaning Procedure Use this cleaning procedure if your samples are not freezing properly and you suspect that the solenoid impactor is not moving freely, perhaps due to the accumulation of sample residue (an unlikely event) or some other form of debris within the freezing chamber. C 4. Loosen the Solenoid retainer mounting screw (B) and rotate the retainer out of the way. Tools Needed: Phillips screwdriver. Instruction: Warning-Hazardous Voltage Warning-Internal components may be damaged by static electricity. CAUTION: Improper connections may cause damage to the instrument. 1. Place a dry chamber cleaner in the cooling chamber. 2. Turn off the power and unplug the instrument. Remove the five screws securing the rear cover, and remove the instrument cover. 3. Remove the four screws (A) holding the Sample Handling Assembly to the housing. Tip the unit onto the back panel and loosen the screw securing the shroud to the base. Return the unit to the upright position and remove the Sample Handling Assembly. A D 5. Withdraw the solenoid plunger (C) while leaving the solenoid body (D) in place. CAUTION: Care must be taken when removing the solenoid plunger, to not lose the spring or spring retainer. 6. Clean the smaller-diameter tip and the solenoid plunger with a 70% isopropanol solution. Do not use any abrasive for this cleaning procedure. 7. Dampen the wooden end of a cottontipped applicator with a 70% isopropanol solution, and insert it through the solenoid body into the smaller-diameter plunger hole until it reaches the chamber cleaner. Move the applicator in and out to scrub the sides of the hole. 8. Return the cleaned solenoid plunger to the solenoid body, secure the retainer, and remove the chamber cleaner. 9. Replace the Sample Handling Assembly and reassemble the unit by reversing steps 2-3. B 25 Fiske Micro-Osmometer Service Manual
10. Recalibrate the instrument. If you have difficulty with this procedure, or if the instrument does not function properly, please obtain Hot-Line service by calling 800-225-1002 (toll-free within the USA and Canada) or (+US) 781-320-5656 (elsewhere). Fiske Micro-Osmometer Service Manual 26
4. Troubleshooting
Introduction All troubleshooting must be performed by qualified service technicians who have basic electrical troubleshooting skills and the proper equipment (i.e., voltage meter). This manual does not imply a license to troubleshoot or perform repairs without proper qualifications. CAUTION: Some troubleshooting requires the instrument to be turned on while the cover is removed. Use extreme caution when operating the instrument without the cover. Hazardous voltages are present at the AC input and power supply. The troubleshooting chart in this section lists problems that may occur while operating the instrument. The left column describes the symptom, the middle column lists checks to determine what is causing the problem, and the right column suggests an action to resolve the problem. Follow the checks in the order they are listed. If you perform checks out of order, you risk replacing the wrong part. If, after performing all checks, there is still a problem, call the Fiske Associates Product Service Department. See Service Assistance (Section 1 - Introduction) for the appropriate phone number to call. Diagnostic Menu The Diagnostic Menu [DIAG] contains items that allow discrete testing of each component of the instrument. Enter the Diagnostic Menu by scrolling to it, and pushing the [DIAG] button. A description of the available tests follows. 1. A/D Tests A/D Tests are designed to check the performance of the cooling assembly, and the efficiency of heat transfer between the components of the cooling assembly. The result of an A/D Test gives a "duty cycle", expressed as a percentage indication of how much power the thermoelectric cooler uses to reach an operating temperature sensed by the block probe. An increasing duty cycle indicates deteriorating performance of the cooling assembly. Pressing the left button will change the set or selected cooling assembly temperature. Pressing the right button will change the display from resistance to temperature. The sample probe readings are also displayed. 2. Probe Bin Test The bin test is used to determine the resistance of the sample probe at a specific temperature. (-0.093 C is used for convenience, which corresponds to the 50 mosm/kg calibration standard.) When a new sample probe is installed, the bin number must be determined and entered into the instrument from the Setup Menu. With age or hard use, the resistance of a probe may change. It is therefore useful to recheck the bin test as part of a routine diagnostic check. 3. Solenoid Test The Solenoid Test will manually trigger the freeze solenoid. The solenoid should give an audible impact immediately on pressing the [START] button. 4. Display/Print Test This test sends a complete character set to the display, serial port and printer port. This test will identify any problems with these areas. 5. Key/Beeper Test Proper function of the keypad can be confirmed by this test. As each button is pressed, [*******] appears above the button and a beep is heard. 29 Fiske Micro-Osmometer Service Manual
6. Barcode Test This menu selection is for testing a barcode scanner. Scanned information will appear on the display, and a short beep will occur. Additional scans may be performed until the test is cancelled. Please note that the barcode scanner may also beep, if its beeper is not disabled. Exit Returns the user to the "Osmometer Ready" prompt. Fiske Micro-Osmometer Service Manual 30
Troubleshooting Table Symptom Check Suggested Action 1. Abrupt loss of power (no display) 2. "Block Probe Open?" 3."Fan Driver Failure" 4. "No Plateau Detected..." 5. "Parameter RAM Failed" or "New Software Version" Power availability. Temporarily unplug the power cord from its wall outlet and replace it with a known workable lamp or small appliance. Power connection. Unplug the power cord from the power cord socket. Measure the AC voltage supplied at the power outlet used for this instrument. Measure the AC voltage supplied at the power outlet used for this instrument. Unplug the power cord from the power cord socket. Use a small screwdriver to pry out the voltage selector/fuse holder. Visually check the two 5mm x 20mm fuses for a blown fuse. If there is any doubt, test the fuses with a continuity checker or ohmmeter or simply replace them. Also check the values marked on the fuses. Check for + 15VDC at supply output. Switch the instrument off, then on. Check the block probe by running the A/D tests. Check the block probe by using the ohmmeter. Probe should be approx. 2000 ohms at room temperature. Switch the instrument off, then on. Check the fan wire connection to the drive harness. Check the pipette s calibration for volume. Switch the instrument off, then on. Have the power restored. Make sure the power cord is firmly plugged into both the wall outlet and the socket on the back of the instrument. Replace blown or incorrect-value fuses with 1.25 Amp 250V time delay (T) fuses: If a fuse was blown and a replacement fuse blows too, contact our product service department for assistance. If the problem persists, contact our product service department for assistance. If the error message does not persist and other error messages are not displayed, ignore this message. If the problem persists, contact our product service department for assistance. If the problem persists, contact our product service department for assistance. Retry the test. Verify that the sample osmolality should be within the range limits for the instrument (i.e., within 0-2000 mosm/kg H O). 2 If the problem persists, contact our product service department for assistance. If the problem persists, contact our product service department for assistance. 6. "Recalibration Needed" Calibrate the instrument. Recalibrate. 31 Fiske Micro-Osmometer Service Manual
Symptom Check Suggested Action 7. Results not repeatable (too scattered) 8. "Sample Did Not Freeze..." 9. "Sample Freeze Error..." 10. "Sample Pre-freeze..." 11. "Sample Probe Open?" Was the pipette clean for each test? Is the pipette tip replaced every 500 tests? Is the sampler properly calibrated? Has solenoid become loose? Impacts Occurred Was the sample configuration correct? Refer to the instructions supplied with the pipette. Is the probe configuration correct? No Impacts Check the impactor solenoid by running the solenoid test. Was the sample configuration correct? Refer to the instructions provided with the pipette. Check the plunger tip (piston) condition. Is the block probe bin # correct? Was the pipette clean? Was the sample probe clean? Is the probe configuration correct? Switch the instrument off, then on. Check the sample probe by running the A/D tests. Check the probe by using an ohmmeter. Probe should be approx. 2000 ohms at room temperature. Replace the pipette tip or replace and calibrate the plunger tip (piston) according to the instructions supplied with the pipette. Calibrate the pipette according to the instructions supplied. Impacts Occurred Retry the test. Check/correct the probe bin # s. No Impacts Retry the test. Please refer to the solenoid cleaning procedure in this manual. If the problem persists, contact our product service department for assistance. Retry the test. Replace the pipette tip or replace and calibrate the plunger tip (piston) according to the instructions supplied with the sampler. Check/correct the block probe bin number. Replace the sample probe. If the problem persists, contact our product service department for assistance. Clean the pipette and sample probe, and then retest. Check/correct the probe bin numbers. If the problem persists, contact our product service department for assistance. If the error message does not persist and other messages are not displayed, ignore this message. If the problem persists, contact our product service department for assistance. Fiske Micro-Osmometer Service Manual 32
Symptom Check Suggested Action 12. Standards Reversed? Please Repeat..." 13. "T E Driver Failure" 14. "Test Time-out Error" 15. Other error messages 16. Date and time are not maintained when instrument is powered off Calibration standards. Switch the instrument off, then on. Check connections to cooling assembly with ohmmeter. Are the probe bin numbers correct? Was the sample probe completely down during the test? - - - - - - Recalibrate with correct standards. If the problem persists, contact our product service department for assistance. Retry the test. If the problem persists, contact our product service department for assistance. Check/correct the block probe bin number. If the problem persists, contact our product service department for assistance. Turn the instrument power off and on. If the problem persists, contact our product service department for assistance. Contact our product service department for assistance. 33 Fiske Micro-Osmometer Service Manual
Notes: Fiske Micro-Osmometer Service Manual 34
5. Replacement Instructions
Block Probe Replacement Kit 110340 Reference: Use this instruction with replacement part 110349. Tools Needed: Phillips screwdriver, 5/16 hex nut driver. Warning-Hazardous Voltage Warning-Internal components may be damaged by static electricity. 3. Remove the four screws (B) holding the Sample Handling Assembly to the housing. Tip the unit onto the back panel and loosen the screw securing the shroud to the base. Return the unit to the upright position and remove the Sample Handling Assembly. B CAUTION: Improper connections may cause damage to the instrument. Instruction: 1. Turn off the power and unplug the instrument. Remove the five screws securing the rear cover, and remove the instrument cover. 2. Disconnect the block probe (A) from the master circuit board. 4. Remove the screw and then the cable clamp (C) that secures the block probe cable to the heatsink. A C D For additional information, please contact the Fiske Customer Service Center (U.S. 1-800-225-1002, outside North America +US 1-781-320-5656), or by e-mail at: orders@fiskeassociates.com. For international customers, please contact your authorized Fiske Associates distributor. 21P110340 Rev0 (2105 Service Manual) Page 1 of 2
5. Notice the length of probe cable as you remove it from the block. Cut any tie wraps used to create the service loop, and then remove the block probe cable from the Cooling Assembly. 6. Insert the tip of the syringe (filled with thermal grease) all the way into the block probe hole (D). Inject grease into the hole while slowly withdrawing the tip, so that there are no air pockets. Fill the hole with grease. (The thermal grease should surround the probe when it is inserted.) 7. Insert the new block probe cable into the block probe hole until it will not go any further. Insert slowly so the probe is not damaged, allowing the thermal grease to be displaced by the probe. Once fully inserted, hold the probe cable in place until excess grease has stopped oozing out and the block probe remains in place. Wipe away any excess grease. 8. Note the block bin number located on the block probe cable (see step 9). Reassemble the unit by reversing steps 1-4. 9. Set the new block bin number. 10. Recalibrate the instrument. 21P110340 Rev0 (2105 Service Manual) Page 2 of 2
Cooling Assembly Replacement 210335R (For Instruments with Serial Suffix A through C) 210336R (For Instruments with Serial Suffix D or Greater) Reference: Use this instruction with replacement part 210335/210336. Tools Needed: Phillips screwdriver, needlenose pliers (if available), alignment fixture F21001. Warning-Hazardous Voltage 3. Remove the four screws (C) holding the Sample Handling Assembly to the housing. Tip the unit onto the back panel and loosen the screw securing the shroud to the base. Return the unit to the upright position and remove the Sample Handling Assembly. C Warning-Internal components may be damaged by static electricity. CAUTION: Improper connections may cause damage to the instrument. Instruction: 1. Turn off the power and unplug the instrument. Remove the five screws securing the rear cover, and remove the instrument cover. 2. Disconnect the sample probe cable (A) and the block probe cable (B) from the PC Board set. F 4. Disconnect the solenoid cable (D) and the thermoelectric cable (E) from the drive harness. A B D E For additional information, please contact the Fiske Customer Service Center (U.S. 1-800-225-1002, outside North America +US 1-781-320-5656), or by e-mail at: orders@fiskeassociates.com. For international customers, please contact your authorized Fiske Associates distributor. 21P0335R Rev1 (2105 Service Manual) Page 1 of 2
5. Remove the Dress screw (F) that attaches the sample probe cover to the probe mount. Pull the probe cover forward to remove it. G H I 6. Unplug the sample probe cable from the two-terminal connector (G) at the top of the probe housing (needle-nose pliers may facilitate this step). Note: Do not pull the connector by its cable. 7. Remove the two Phillips probe-mounting screws and washers (H), and pull the sample probe up and out of the head. Place the sample probe cover and the sample probe aside, to be reassembled later. CAUTION: The sample probe tip is very fragile. Extra care must be taken to prevent breakage when out of the instrument. 9. Note the block probe bin number located on the block probe cable (see step 13). 10. Secure the probe alignment fixture in the probe mount and lower the fixture into the well of the new Cooling Assembly. Hold the fixture in place and tighten the mounting screws to the heatsink brackets. Test for proper alignment by sliding the probe fixture in and out of the well, and adjust if necessary. 11. Remove the probe alignment fixture and remount the sample probe. 12. Replace the Sample Handling Assembly and reassemble the unit by reversing steps 1-6. 13. Set the new block probe bin number and recalibrate the instrument. 8. Remove the four 6-32 screws (I) that mount the Cooling Assembly to the 210 heat sink brackets. Remove the old Cooling Assembly from the instrument and mount the new one to the heatsink brackets, being careful not to dislodge the solenoid plunger and spring during this procedure. 21P0335R Rev1 (2105 Service Manual) Page 2 of 2
Reference: Use this instruction with replacement part 330330. Tools Needed: Phillips screwdriver, 5/16 nut driver or wrench. Cooling Fan Replacement Kit 210330R Warning-Hazardous Voltage Warning-Internal components may be damaged by static electricity. CAUTION: Improper connections may cause damage to the instrument. Instruction: B C 1. Turn off the power and unplug the instrument. Remove the five screws securing the rear cover and remove the instrument cover. 2. Remove the four screws holding the Sample Handling Assembly (A) to the housing. Tip the unit onto the back panel and loosen the screw securing the shroud to the base. Return the unit to the upright possition and remove the Sample Handling Assembly from the housing. A 3. Disconnect the fan from the drive harness (B). 4. Remove the two screws (C) and replace the old fan with the new fan, making sure the airflow arrow is pointing up, and the wires and connector are towards the rear of the machine. NOTE Some units may have four (4) screws (C) threaded into four hex nuts, which are located on the bottom of the unit. 5. Replace the Sample Handling Assembly and reassemble the unit by reversing steps 1-3. For additional information, please contact the Fiske Customer Service Center (U.S. 1-800-225-1002, outside North America +US 1-781-320-5656), or by e-mail at: orders@fiskeassociates.com. For international customers, please contact your authorized Fiske Associates distributor. 21P0330R Rev1 (2105 Service Manual) Page 1of 1
Display Cable Replacement 210151R Reference: Use this instruction with replacement part 210151. Tools Needed: Phillips screwdriver. Warning-Hazardous Voltage Warning-Internal components may be damaged by static electricity. CAUTION: Improper connections may cause damage to the instrument. Instruction: 1. Turn off the power and unplug the instrument. Remove the five screws securing the rear cover, and remove the instrument cover. 2. Disconnect the ribbon Display Cable from the PC board set (A) and the display module (B), noting pin 1 orientation of the cable and connector. 3. Replace the Display Cable and reassemble the unit by reversing steps 1-2. B A For additional information, please contact the Fiske Customer Service Center (U.S. 1-800-225-1002, outside North America +US 1-781-320-5656), or by e-mail at: orders@fiskeassociates.com. For international customers, please contact your authorized Fiske Associates distributor. 21P0151R Rev0 (2105 Service Manual) Page 1
Keypad Replacement 210511R Reference: Use this instruction with replacement part 210511. Tools Needed: Phillips screwdriver, 5/16 hex nut driver. Instruction: Warning-Hazardous Voltage Warning-Internal components may be damaged by static electricity. CAUTION: Improper connections may cause damage to the instrument. 1. Turn off the power and unplug the instrument. Remove the five screws securing the rear cover and remove the instrument cover. 2. Beginning at the top and moving clockwise, disconnect the following cables from the PC Board Set: A. Keypad Interface cable B. Display cable C. Sample Probe cable D. Block Probe cable E. Printer cable F. Serial cable G. DC Power Harness H. Drive Harness 3. Remove the two Sems screws (I) from the left-side top and bottom of the board set, and undo the board from the snap studs (J) on the right-side top and bottom. Remove the PC board set. 4. Remove the hardware from the ground tab stud (K). A B I H G C D J F E For additional information, please contact the Fiske Customer Service Center (U.S. 1-800-225-1002, outside North America +US 1-781-320-5656), or by e-mail at: orders@fiskeassociates.com. For international customers, please contact your authorized Fiske Associates distributor. 21P0511R Rev0 (2105 Service Manual) Page 1 of 2
5. Pull the Keypad off the front of the unit. 6. Carefully pull the Keypad grounding tab and ribbon cable through the case slots from the front of the unit (L). 7. Install the replacement Keypad and reassemble the unit by reversing steps 1-6. Note: Before pressing the Keypad firmly into place, check alignment of the display window and keypad with the display panel door. 8. Run Keypad Test to verify that the keypad is functioning properly. K L 21P0511R Rev0 (2105 Service Manual) Page 2 of 2
Fluorescent Display Module Replacement Kit 210800R Reference: Use this instruction with replacement part 330800. Tools Needed: Small Phillips screwdriver. Warning-Hazardous Voltage Warning-Internal components may be damaged by static electricity. CAUTION: Improper connections may cause damage to the instrument. Instruction: 1. Turn off the power and unplug the instrument. Remove the five screws securing the rear cover, and remove the instrument cover. 2. Remove the four screws and hardware from the Fluorescent Display Module (A). 3. Disconnect the ribbon cable from the Display Module, noting pin 1 orientation of the cable and connector (B). 4. Replace the Display Module and reassemble the unit by reversing steps 1-3. A B For additional information, please contact the Fiske Customer Service Center (U.S. 1-800-225-1002, outside North America +US 1-781-320-5656), or by e-mail at: orders@fiskeassociates.com. For international customers, please contact your authorized Fiske Associates distributor. 21P0800R Rev0 (2105 Service Manual) Page 1
Fuse Replacement 70013R Reference: Use this instruction with replacement part 70013. Note: Use caution when removing the fuse holder from the power entry module. Tools Needed: Flat-bladed screwdriver. Warning-Hazardous Voltage Warning-Internal components may be damaged by static electricity. 3. Dislodge the old fuses and reinstall the new fuses by placing them in the fuse holder brackets. 4. Place the fuse holder back into the power entry module and close the door. CAUTION: Improper connections may cause damage to the instrument. Instruction: 1. Turn off the power and unplug the instrument. 2. Insert a flat-bladed screwdriver between the fuse holder and the power entry module. Carefully pry open the fuse holder door (A) and remove the fuse holder from the power entry module. A For additional information, please contact the Fiske Customer Service Center (U.S. 1-800-225-1002, outside North America +US 1-781-320-5656), or by e-mail at: orders@fiskeassociates.com. For international customers, please contact your authorized Fiske Associates distributor. 21P70013R Rev1 (2105 Service Manual) Page 1 of 1
NVRAM Battery Replacement 71027R Reference: Use this instruction with: Advanced Instruments Inc. 4250 Cryoscopes 3250 Osmometers 3320 Osmometers 2020 Osmometers 4-digit year Model 3300 Osmometers Intel-based 3D3 Osmometers Intel-based 4D3 Cryoscopes Fiske Associates 210 Osmometers Intel-based MK-3 Osmometers Intel-based MK-2 Cryoscopes Tools Needed: Small flat-bladed screwdriver, static grounding (earthing) wrist strap (included). CAUTION: Unplug the power cord prior to opening or removing covers, or else you may be exposed to electric shock, excessive temperatures, or mechanical hazards. Performing this service or maintenance should only be done by a qualified service technician. A discharge of static electricity from contact with the human body or other conductor may damage system boards or static sensitive devices. Never perform internal maintenance without following recommended static protection procedures. Instruction: 1. Turn off the instrument power and remove the power cord from the rear of the instrument. 2. Open the instrument cover. 3. Attach the static grounding strap according to the instructions on its packaging. 4. Locate the processor Printed Circuit Board (PCB120, PCB125, or PCB620) and the NVRAM chip: If using PCB120, locate U14. If using PCB125 locate U13. If using PCB620, remove the processor board from the application board and locate U15 on the back side. 5. Once you have located the NVRAM chip, locate the small notch and insert the small screwdriver, as described in figure 1. 6. Remove the cover (powercap) and set it aside. Figure 1 For additional information or technical assistance, please contact Advanced Instruments Hot-Line Service Center. (U.S. 1-800-225-4034, outside North America +US 1-781- 320-9000. After normal business hours, dial extension 2191.) 71P27R Rev2 (2025/2105/3255/3305/3325/ 4D35/MK05 Service Manuals) Page 1 of 2
7. Remove the replacement powercap from the anti-static bag and install it on the component base, as described in figure 2. 8. Re-install the processor board, if removed. 9. Restore power to the instrument and verify correct power-up sequence without any LOW BATTERY warnings. 11. Turn off the power and wait ten minutes. Turn the instrument back on and verify that the correct time has been maintained. 12. Turn off the power and restore the instrument cover. 13. Reset all stored parameters (e.g., block and sample bin numbers, plateau modes, etc.) and recalibrate the instrument. 10. Set the date and time, following the instructions in the user s guide. Figure 2 71P27R Rev2 (2025/2105/3255/3305/3325/ 4D35/MK05 Service Manuals) Page 2 of 2
PC Board Set Replacement 210620R Reference: Use this instruction with replacement part 210620R. Tools Needed: Cross-point screwdriver, pliers, static grounding (earthing) wrist strap (included). Warning-Hazardous Voltage CAUTION: Improper connections may cause damage to the instrument. CAUTION: A discharge of static electricity from contact with the human body or other conductor may damage system boards or static-sensitive devices. NEVER UNPACK, TOUCH OR HANDLE ANY PCB WITHOUT WEARING A GROUNDING (EARTHING) STRAP TO MINIMIZE YOUR STATIC DISCHARGE. Instruction: PLEASE NOTE: The processor printed circuit board holds all the custom setup parameters in memory. Replacement of this board will require that you transfer your settings to the new board after installation. If your current board is still bootable, please write down your sample and block bin numbers, instrument serial number, and feature lockout information. These and other setup parameters will need to be entered after completing the installation. 1. Turn off the power and unplug the instrument. Remove the five screws securing the rear cover and remove the instrument cover. A B I H G C D K J F E For additional information, please contact the Fiske Customer Service Center (U.S. 1-800-225-1002, outside North America +US 1-781-320-5656), or by e-mail at: orders@fiskeassociates.com. For international customers, please contact your authorized Fiske Associates distributor. 21P0620R Rev2 (2105 Service Manual) Page 1 of 2
2. Beginning at the top and moving clockwise, disconnect the following cables from the board set: A. Keypad Interface cable B. Display cable C. Sample Probe cable D. Block Probe cable E. Printer cable F. Serial cable G. DC Power Harness H. Drive Harness 3. Remove the two 6-32 Sems screws (I) from the left-side top and bottom of the board set, and undo the board from the snap studs (J) on the right-side top and bottom. 4. Locate dip switch SW1 (K) on the new board set and verify that switch positions 1-4 are in the off position. 5. Replace the new board set and reassemble the unit by reversing steps 1-3. 6. After completing reassembly, apply power to the instrument and enter the SETUP menu. Transfer the information and custom settings from your original board set, as noted at the start of this document. 21P0620R Rev2 (2105 Service Manual) Page 2 of 2
Power Entry Assembly Replacement 210140R Reference: Use this instruction with replacement part 210140. Tools Needed: Phillips screwdriver, 5/16 (#6-32) nut driver. Instruction: Warning-Hazardous Voltage Warning-Internal components may be damaged by static electricity. CAUTION: Improper connections may cause damage to the instrument. 1. Turn off the power and unplug the instrument. Remove the five screws securing the rear cover and remove the instrument cover. 2. Disconnect the serial cable (A) and the printer cable (B) from the Power Entry Assembly. (In the photo enlargement, below, the cables are disconnected and folded up and out of the way for viewing purposes.) 3. Remove the two screws that mount the Power Entry Assembly (C) and remove the assembly to the extent allowed by the service loops. 4. Disconnect the AC Wiring Harness (D) from the power supply and remove the power entry ground wire (bottom wire) from the ground stud (E). Note: This ground wire is at the bottom of the stud, and should be remounted in the same order. 5. Remove and replace the old Power Entry Assembly with the new one and reassemble the unit by reversing steps 1-4. A B D C E For additional information, please contact the Fiske Customer Service Center (U.S. 1-800-225-1002, outside North America +US 1-781-320-5656), or by e-mail at: orders@fiskeassociates.com. For international customers, please contact your authorized Fiske Associates distributor. 21P0140R Rev0 (2105 Service Manual) Page 1
Power Supply Replacement 210950R Reference: Use this instruction with replacement part 210950. Tools Needed: Phillips screwdriver, pliers. Instruction: Warning-Hazardous Voltage Warning-Internal components may be damaged by static electricity. CAUTION: Improper connections may cause damage to the instrument. 1. Turn off the power and unplug the instrument. Remove the five screws securing the rear cover and remove the instrument cover. 2. Disconnect the serial cable (A) and the printer cable (B) from the Power Entry Assembly. (In the photo enlargement, below, the cables are disconnected and folded up and out of the way for viewing purposes.) 3. Remove the two screws that mount the Power Entry Assembly (C) and remove the assembly to the extent allowed by the service loops. 4. Remove the bottom two screws (D) and associated hardware on the power supply. Pry the power supply from the two top snap-top studs (E). 5. Disconnect the ground wire (F), the AC wiring harness (G), and the DC power harness (H) from the power supply. 6. Remove and replace the old Power Supply with the new one and reassemble the unit by reversing steps 1-5. E A B C H F D G For additional information, please contact the Fiske Customer Service Center (U.S. 1-800-225-1002, outside North America +US 1-781-320-5656), or by e-mail at: orders@fiskeassociates.com. For international customers, please contact your authorized Fiske Associates distributor. 21P0950R Rev0 (2105 Service Manual) Page 1
Sample Probe Replacement Kit 242700 (For Instruments with Serial Suffix A) Reference: Use this instruction with replacement part 242700A. Tools Needed: Small Phillips screwdriver, needle-nose pliers (if available). Instruction: Model 210 Probe Plug Retaining Clip Installation Diagram RETAINING CLIP PREVENTS PLUG FROM BACKING OFF PINS MAKE SURE THAT CLIP IS BETWEEN THE RED AND BLACK LEADS 1. Turn off the power and unplug the instrument. 2. Remove and save the dress screw (A) that attaches the probe cover to the probe mount. Pull the probe cover forward to remove it. C D Probe Mount CAPTURE CLIP BETWEEN LOCK WASHER AND THE NYLON WASHER E B Note: cable. Do not pull the connector by its A 4. Remove and save the front probe mounting screw and nylon washer (D). Pull the probe being replaced, up and out of the probe mount. Remove and save the spacer (E). CAUTION: The sample probe tip is very fragile. Extra care must be taken to prevent breakage when out of the instrument. 3. Remove and save the rear probe mounting screw and nylon washer (B). If present, remove and save the retainer clip that wraps around the back of the sample probe cable connector (C), then unplug the sample probe cable connector from the two-terminal connector at the top of the probe housing (needle-nose pliers may facilitate this step). 5. Clean the top surface of the probe mount. Replace the spacer (E). Being careful to avoid damage to its thermister tip, mount the new probe using the front probe mounting screw and nylon washer. Tighten the screw enough to prevent the probe from moving up and down, but still allowing the probe to move from side to side. For additional information, please contact the Fiske Customer Service Center (U.S. 1-800-225-1002, outside North America +US 1-781-320-5656), or by e-mail at: orders@fiskeassociates.com. For international customers, please contact your authorized Fiske Associates distributor. 21P242700 Rev2 (2105 Service Manual) Page 1 of 4
6. Plug the sample probe cable connector (C) firmly into the two pins at the top rear of the sample probe housing. Replace the nylon washer at the rear probe mounting hole. 7. Guide the end of the retaining clip between the red and black leads at the back of the sample probe cable connector. Position the ring end of the clip on top of the nylon washer, then reinstall the rear probe mounting screw. Again, tighten the screw enough to prevent the probe from moving up and down, but still allowing the probe to move from side to side. Refer to the illustration for proper orientation of the retaining clip. 8. Place a sample tube in the sample well. Move the sample probe head so that the sample probe is inserted into the tube. Apply firm downward pressure on the top of the sample probe to automatically center it in the sample tube and well. Tighten the front probe mounting screw. 9. Using a finger to restrain the retaining clip from turning, tighten the rear probe mounting screw. Move the head up and down, while observing the sample tube. When properly centered, the probe should slide out of the sample tube without pulling it up, and then should slide into the sample tube without pushing it to the side. If necessary, loosen screws and repeat step 7 to properly center the sample tube. 10. A factory-determined sample probe bin number is marked on the probe (F). Compare the bin numbers on the new and the old sample probes. If the bin numbers are not the same, record the new bin number in the Service Log, located in the back of the Service Manual. 11. Raise the sample probe head and replace the probe cover and its mounting screw. 12. Make sure an empty sample tube is in the sample-cooling well and lower the sample probe head all the way down. 13. Plug in the instrument and turn on the power. Select the option to Run Diagnostics. 14. If the bin number on the new sample probe is the same as the bin number on the old sample probe, it is not necessary to reset the probe configuration. If the bin number on the new sample probe is not the same as the bin number on the old sample probe, enter the Setup menu and select the Set Probe Bin # s option. Set the sample probe bin number, accordingly. 15. Recalibrate the instrument. F 21P242700 Rev2 (2105 Service Manual) Page 2 of 4
Sample Probe Replacement Kit 242700 (For Instruments with Serial Suffix B or Greater) Reference: Use this instruction with replacement part 242700A. C Tools Needed: Small Phillips screwdriver, needle-nose pliers (if available). Instruction: 1. Turn off the power and unplug the instrument. 2. Remove and save the dress screw (A) that attaches the probe cover to the probe mount. Pull the probe cover forward to remove it. F B E D 3. Remove and save the probe mounting screws and nylon washers (B). A 4. Pull the sample probe (F) being replaced up, then unplug the sample probe cable connector (C) from the two-terminal connector at the top of the sample probe housing (needle-nose pliers may facilitate this step). Once disconnected, finish removing the probe from the probe mount. Remove and save the spacer (E). Note: Do not pull the connector by its cable. CAUTION: The sample probe tip is very fragile. Extra care must be taken to prevent breakage when out of the instrument. For additional information, please contact the Fiske Customer Service Center (U.S. 1-800-225-1002, outside North America +US 1-781-320-5656), or by e-mail at: orders@fiskeassociates.com. For international customers, please contact your authorized Fiske Associates distributor. 21P242700 Rev2 (2105 Service Manual) Page 3 of 4
5. Clean the top surface of the probe mount. Replace the spacer (E). Being careful to avoid damage to its thermistor tip, begin to insert the tip through the hole in the probe mount. Plug the sample probe cable connector (C) into the two pins at the top rear of the sample probe housing, then complete mounting the new sample probe in the probe mount using the probe mounting screws and nylon washers. Tighten the screws enough to prevent the sample probe from moving up and down, but still allowing the probe to move from side to side. 6. Make sure the wires from the sample probe cable are down in the slot provided in the probe mount. 7. Place a sample tube in the sample well. Move the sample probe head so that the sample probe is inserted into the tube. Apply firm downward pressure on the top of the sample probe to automatically center it in the sample tube and well. Tighten both probe mounting screws. 8. Move the head up and down, while observing the sample tube. When properly centered, the probe should slide out of the tube without pulling it up, and then should slide into the sample tube without pushing it to the side. If necessary, loosen the probe mounting screws and repeat step 7 to properly center the sample probe. 10. Raise the sample probe head and replace the probe cover and its mounting screw (A). Make sure that the wire from the sample probe cable is captured within the slot in the probe mount. 11. Make sure an empty sample tube is in the sample cooling well and lower the sample probe head all the way down. 12. Plug in the instrument and turn on the power. Select the option to Run Diagnostics. 13. If the bin number on the new sample probe is the same as the bin number of the old sample probe, it is not necessary to reset the probe configuration. If the bin number on the new sample probe is not the same as the bin number on the old sample probe, then enter the Setup menu and select Set Probe Bin # s option. Set the sample probe bin number, accordingly. 14. Recalibrate the instrument. 9. A factory-determined sample probe bin number is marked on the probe (F). Compare the bin numbers on the new and old sample probes. If the bin numbers are not the same, record the new bin number in the Service Log, located in the back of the Service Manual. 21P242700 Rev2 (2105 Service Manual) Page 4 of 4
Sample Probe Cable Assembly Replacement 240725R Reference: Use this instruction with replacement part 240725. B Tools Needed: Phillips screwdriver, 5/16 hex nut driver. Warning-Hazardous Voltage Warning-Internal components may be damaged by static electricity. Instruction: CAUTION: Improper connections may cause damage to the instrument. 1. Turn off the power and unplug the instrument. Remove the five screws securing the rear cover, and remove the instrument cover. D 3. Remove the four screws (B) holding the Sample Handling Assembly to the housing. Tip the unit onto the back panel and loosen the screw securing the shroud to the base. Return the unit to the upright position and remove the Sample Handling Assembly. A C 2. Disconnect the sample probe (A) from the application board. 4. Remove the screw and then the cable clamps (C) in three places on the Sample Handling Assembly that secure the sample probe cable to the assembly. For additional information, please contact the Fiske Customer Service Center (U.S. 1-800-225-1002, outside North America +US 1-781-320-5656), or by e-mail at: orders@fiskeassociates.com. For international customers, please contact your authorized Fiske Associates distributor. 21P240725R Rev0 (2105 Service Manual) Page 1 of 2
5. Remove the Dress screw (D) that secures the probe cover to the probe mount and remove the probe cover. CAUTION: The sample probe tip is very fragile. Extra care must be taken to prevent breakage when out of the instrument. 6. Being careful to avoid damage to the probe s thermister tip, disconnect the probe cable connector (E) from the sample probe. 7. Replace the old sample probe cable with the new sample probe cable. 8. Reassemble the unit by reversing steps 1-6. 9. Recalibrate the instrument. E 21P240725R Rev0 (2105 Service Manual) Page 2 of 2
6. Printer User Manual Model 210550 Only See manuals supplied with other printer models
Model 210550 External Printer Specifications Character Buffering: Print Method: Character Matrix: 1.5K standard 8K optional Approximately 9,500 byte capacity with option installed. Impact dot matrix 6 x 8 dot matrix Print Speed: 38 lines per minute for 24 column 33 lines per minute for 32 and 40 column Paper: Table top: 2.25"W x 2.75"D; 0.44" I.D. Panel mount: 2.25"W x 1.25"D Power: 1.5 Watts (idle) 10 Watts (while printing) Character Spacing: 24 column: 12.8 characters/inch 32 column: 17 characters/inch 40 column: 21 characters/inch DC Voltage: 9-12 VDC 100mA idle, 1500mA with 100% printing 2.9A peak with 100% printing Line Feed Spacing: 7.6 lines per inch...character mode 9.1 lines per inch...graphic mode External Dimensions: 4"W x 4.5"L x 2"H Operating Temp: 5 C to 40 C 41 F to 104 F For additional information, please contact the Fiske Customer Service Center (U.S. 1-800-225-1002, outside North America +US 1-781-320-5656), or by e-mail at: orders@fiskeassociates.com. For international customers, please contact your authorized Fiske Associates distributor. 210P550 Rev0 (210550) Page 1 of 7
Installation Please follow the precautions listed below when setting up your printer. They are designed to help you keep your printer working at its best. 5. Slide the paper through the slot between the paper compartment and the printer compartment. You can slide the paper in about one-quarter inch before it stops. Plug your printer into an appropriate grounded outlet. Place your printer on a flat hard surface, like a tabletop. Keep your printer out of direct sunlight. Connect the centronics cable to the 36-pin centronics connector on the printer, and then connect the DB25 cable to the female DB25 connector on the back panel of the instrument. Connect the DIN cable to the printer power jack on the back panel of the instrument, and then to the DC input on the printer. Installing the Paper 1. Remove the printer cover by pressing on the groove pattern to pop the front edge up. Lift off the cover. 2. Press the rocker switch to the left. The light will go off. 6. While holding the paper in place, press the rocker switch to the paper feed position. The printer will activate, and a rubber roller will pull the paper into the printer compartment. Hold the switch in the paper feed position until the paper emerges from the top of the printer mechanism. 7. When an inch of paper has emerged from the top of the printer, release the paper feed button and pull the paper through the printer until several inches are exposed. 3. Unroll several inches of the paper. 4. Cut a straight edge on the paper roll if it is jagged. This will facilitate inserting the paper into the printer. 8. Slide the paper through the slot in the printer cover. 210P550 Rev0 (210550) Page 2 of 7
2. Push down on the embossed grooves on one or both sides of the printer cover until the printer cover tilts. 9. Push the back of the printer cover down and into place. 10. Press the front of the printer cover down to lock in place. 3. Lift off the printer cover. Push down on the right side of the ribbon cartridge where it is marked PUSH, and remove the cartridge. 4. Install the new cartridge. Be sure the ink cartridge is inserted firmly to prevent weak or irregular printing. The cartridge must be properly seated and aligned for best printing. Changing an Ink Ribbon When printing becomes faint or difficult to see, the ribbon may be dried out or in need of replacement. Before replacing the cartridge, try advancing the ribbon to a new section by holding down the paper feed switch for several seconds. 5. Turn the cartridge clockwise (marked by an arrow) to stretch the ribbon. If the print impression is still weak, purchase a new ribbon cartridge (Product #330541) and install it using the following instructions. 1. Turn the printer offline by pressing the left side of the rocker switch. The light should go out. 6. Replace the cover. Replace the paper. 210P550 Rev0 (210550) Page 3 of 7
Printer Test and Setup The printer can be tested and setup using the steps described in this section. However, please note that the printer has been setup to work with your instrument, as shipped. No setup changes are necessary for correct operation. Before attempting to change your printer settings, be sure to disconnect the data cable from the instrument. Printer Test Follow these steps to test the printer: 1. Unplug the printer at the outlet or at the back of the printer. 2. Press and hold the right side of the rocker switch as you plug in the printer. The printer will print out a list of the current configuration settings, and then run a continuous print test. To stop the print test, press either side of the rocker switch. Below is a sample of what is printed when you do the print test. Setup Menu Follow these steps to access the Setup Menu: 1. Unplug the printer at the outlet or at the back of the printer. 2. Press and hold the left side of the rocker switch as you plug in the printer. The printer will advance the paper. 3. After the paper advance has stopped, count 3-5 seconds and release the switch. The following is printed: *** SETUP MENU *** CONFIGURE... [NEXT/OK] If you wait less than three or more than five seconds, Ready... is printed and you will have to repeat steps 1 and 2 to access the Setup Menu. 4. Press NEXT (left side of rocker switch) repeatedly to display the Setup Menu options. The list of menu options will repeat as you continue pressing NEXT. Following is a list of available Setup Menu options: *** PRINTER TEST *** Parallel Version B122XL COLS=32 FONT=5X8 INVERT=NO MAG=NONE BUFFER: 1.5K INT RAM: OK EXT RAM: OK ROM: OK EEPROM: OK! #$%& ()*+,-./0123456789:;< = >?@ #$%& ()*+,-./0123456789:;<=>?@A Ready... *** SETUP MENU *** CONFIGURE... CUSTOM... RESET SEQ# [NEXT/OK] [NEXT/OK] [NEXT/OK] These options for customizing your printer are explained below. 210P550 Rev0 (210550) Page 4 of 7
CONFIGURE... The first menu item reads CONFIGURE... [NEXT/OK] [NEXT/OK] is a visual reminder that pressing the left side of the rocker switch will go to the next menu item, and that pressing the right side of the rocker switch will select the current option. Press OK (right side) to access the Configuration Menu. Mag Magnification refers to the size of printed type from your printer. Choices are: NONE Double Wide Double High Double Wide/High After choosing a magnification, the printer will print Ready... to show that the printer is out of the Configuration Menu and the Setup Menu, and is ready to print. *** SETUP MENU *** CONFIGURE... [NEXT/OK] *** CONFIGURATION MENU *** *** CONFIGURATION MENU *** LOAD DEFAULTS [NEXT/OK] COLS=32 [NEXT/OK] INVERT=NO [NEXT/OK] FONT=5X8 [NEXT/OK] MAG=NONE [NEXT/OK] Ready... CUSTOM... The second Setup Menu item reads CUSTOM... [NEXT/OK] Press OK (right side) to access the Custom Menu. Configuration parameters are described below. Cols Use this parameter to set the number of characters (columns) per line. Choices are 24, 32, or 40 (the larger numbers represent a smaller, more condensed appearance). *** SETUP MENU *** CONFIGURE... CUSTOM... ****** CUSTOM MENU ****** PRINT CUSTOM SETUP [NEXT/OK] [NEXT/OK] [NEXT/OK] Press OK (right side) to print the current Custom setup. A sample printout follows. Invert Use this parameter to change the orientation of text in your printouts. Choose YES if you want inverted (upside down) text, or NO if you want non-inverted (right-side up) text. Font Choose from a 5 x 5 or a 5 x 8 dot matrix print pattern. The 5 x 5 dot matrix pattern produces only uppercase (capital) letters. 210P550 Rev0 (210550) Page 5 of 7
*** SETUP MENU *** CONFIGURE... CUSTOM... ****** CUSTOM MENU ****** PRINT CUSTOM SETUP AUTO SEQ=YES ZERO=0 POUND SIGN=# _(Underscore) BUSY INVERT=NO ONLINE/OFFLINE=YES EXT CH SET=NO PRINT READY=YES Ready... [NEXT/OK] [NEXT/OK] [NEXT/OK] [NEXT/OK] [NEXT/OK] [NEXT/OK] [NEXT/OK] [NEXT/OK] [NEXT/OK] [NEXT/OK] [NEXT/OK] Customization parameters are described below. Auto Seq# NO - Don t autoprint sequence number after CR (carriage return). YES - Autoprint sequence number after CR. Zero Choose how you want the zero character to look in your printouts -- with or without a slash through the character. Pound Sign Choose to show pound as # symbol, or as the British Sterling pound symbol. _Underscore Choose which symbol the same ASCII code will print -- an _underscore, or a left arrow. Busy Invert NO - Voltage will be in a high state until the unit is busy, then voltage level goes low. YES - Voltage will be in a low state until the unit is busy, then voltage level goes high. Online/Offline YES - Enables the rocker switch to turn the printer offline. NO - Ext Ch Set Disables the Online/Offline option. The option to use the extended character set is available only when 8 data bits are chosen. YES - Allows you to use hexadecimal numbers above 80 (true only for 8 data bits). NO - Print Ready Disables the Extended Character Set option. YES - Prints Ready... upon power up. NO - Disables printing Ready... WARNING: If you choose NO, hold down the left side of the rocker switch for 4-6 seconds to access the Setup Menu. Begin timing when you plug in the unit and the red light comes on. The paper feed motor does not run upon powerup when Ready... is disabled. RESET SEQ#... The third Setup Menu item reads RESET SEQ#... NEXT/OK] This menu option allows you to reset the sequence number, which is the number of print transactions since the last reset. 210P550 Rev0 (210550) Page 6 of 7
To reset the sequence number, press OK (right side). The sequence number will be reset to zero and the printer will print Ready..., showing it is no longer in the Setup Menu and that the printer is ready to print. To skip resetting the sequence number to zero, press NEXT (left side). CONFIGURE... is printed. Unplug and replug the printer to return to printing mode. Ready... is printed. 210P550 Rev0 (210550) Page 7 of 7
7. Schematics
THIS DESIGN, DRAWING, AND PRINTS ARE THE PROPERTY OF FISKE ASSOCIATES. REPRODUCTION OF ANY KIND IS PROHIBITED EXCEPT WITH SPECIFIC PERMISSION OF FISKE ASSOCIATES. PCB100 Back Panel I/O Board Assembly Drawing (For Serial Suffix None, A, or B)
THIS DESIGN, DRAWING, AND PRINTS ARE THE PROPERTY OF FISKE ASSOCIATES. REPRODUCTION OF ANY KIND IS PROHIBITED EXCEPT WITH SPECIFIC PERMISSION OF FISKE ASSOCIATES. PCB100 Back Panel IO Board Schematic (For Serial Suffix None, A, or B)
THIS DESIGN, DRAWING, AND PRINTS ARE THE PROPERTY OF FISKE ASSOCIATES. REPRODUCTION OF ANY KIND IS PROHIBITED EXCEPT WITH SPECIFIC PERMISSION OF FISKE ASSOCIATES. PCB140 Back Panel I/O Board Assembly Drawing (For Serial Suffix C and Later)
THIS DESIGN, DRAWING, AND PRINTS ARE THE PROPERTY OF FISKE ASSOCIATES. REPRODUCTION OF ANY KIND IS PROHIBITED EXCEPT WITH SPECIFIC PERMISSION OF FISKE ASSOCIATES. PCB140 Back Panel IO Board Schematic (For Serial Suffix C or Later)
THIS DESIGN, DRAWING, AND PRINTS ARE THE PROPERTY OF FISKE ASSOCIATES. REPRODUCTION OF ANY KIND IS PROHIBITED EXCEPT WITH SPECIFIC PERMISSION OF FISKE ASSOCIATES. PCB115 Application Board Assembly Drawing
THIS DESIGN, DRAWING, AND PRINTS ARE THE PROPERTY OF FISKE ASSOCIATES. REPRODUCTION OF ANY KIND IS PROHIBITED EXCEPT WITH SPECIFIC PERMISSION OF FISKE ASSOCIATES. PCB115 Application Board Schematic, Rev. 0
THIS DESIGN, DRAWING, AND PRINTS ARE THE PROPERTY OF FISKE ASSOCIATES. REPRODUCTION OF ANY KIND IS PROHIBITED EXCEPT WITH SPECIFIC PERMISSION OF FISKE ASSOCIATES. PCB115 Application Board Schematic, Rev. 2
THIS DESIGN, DRAWING, AND PRINTS ARE THE PROPERTY OF FISKE ASSOCIATES. REPRODUCTION OF ANY KIND IS PROHIBITED EXCEPT WITH SPECIFIC PERMISSION OF FISKE ASSOCIATES. PCB120 CPU 80C186 Assembly Drawing
THIS DESIGN, DRAWING, AND PRINTS ARE THE PROPERTY OF FISKE ASSOCIATES. REPRODUCTION OF ANY KIND IS PROHIBITED EXCEPT WITH SPECIFIC PERMISSION OF FISKE ASSOCIATES. PCB120 CPU 80C186 Assembly Schematic, Rev. 0
THIS DESIGN, DRAWING, AND PRINTS ARE THE PROPERTY OF FISKE ASSOCIATES. REPRODUCTION OF ANY KIND IS PROHIBITED EXCEPT WITH SPECIFIC PERMISSION OF FISKE ASSOCIATES. PCB120 CPU 80C186 Assembly Schematic, Rev.1
THIS DESIGN, DRAWING, AND PRINTS ARE THE PROPERTY OF FISKE ASSOCIATES. REPRODUCTION OF ANY KIND IS PROHIBITED EXCEPT WITH SPECIFIC PERMISSION OF FISKE ASSOCIATES. 210000 210 System Interconnect, Rev. 0 (For Serial Suffix None, A, and B)
THIS DESIGN, DRAWING, AND PRINTS ARE THE PROPERTY OF FISKE ASSOCIATES. REPRODUCTION OF ANY KIND IS PROHIBITED EXCEPT WITH SPECIFIC PERMISSION OF FISKE ASSOCIATES. 210000 210 System Interconnect, Rev. 2 (For Serial Suffix C and Later)
Appendix A Event Record
Event Record The Event Record displays the last actions of the instrument, and is used when servicing the instrument. The output should look like: # Code Description Data Date/Time =========================================== 1: 0 State Change 2 [10/31/2006 11:02:11 AM ] 2: 0 State Change 9 [10/31/2006 11:02:14 AM ] 3: 4 Diagnostic Test 5 [10/31/2006 11:02:26 AM ] 4: 2 Bar Code Scan 0 [10/31/2006 11:02:33 AM ] 5: 5 Error 212 [10/31/2006 11:02:35 AM ] 6: 5 Error 207 [10/31/2006 11:33:51 AM ] 7: 0 State Change 7 [10/31/2006 11:33:53 AM ] 8: 8 Utility Routine 0 [10/31/2006 11:33:55 AM ] 9: 8 Utility Routine 2 [10/31/2006 11:33:57 AM ] 10: 5 Error 207 [10/31/2006 11:33:58 AM ] End Event Record Download Complete INSTRUMENT STATES CODE: 0 DATA DESCRIPTION 0 RESET 1 ERROR 2 READY 3 STANDBY 4 TEST 5 CALIB 6 CALI_TEST 7 UTIL 8 SETUP 9 DIAG 10 BINTEST Event Record output consists of: # A sequential row number from 1 to 100. Code: A number used to identify the table containing the data definition. Description: A text field describing the code table. Data: Date/Time: A numerical descriptor that, when located in the appropriate code table (see below), identifies the action recorded. The date and time the event was recorded. The code, data, and error numbers presented in the Event Record can be decoded, using the following tables. SWITCHES CODE: 1 DATA 0 NONE 1 RIGHT 2 LEFT 4 TEST 8 SUPERVISOR DESCRIPTION NOTE: Value reported is a combination of above switches. BAR CODE SCANS CODE: 2 DATA DESCRIPTION NOTE: No subcategories. Fiske Micro-Osmometer Service Manual
PARAMETER EDIT COMMANDS (instrument must be in factory mode), subset of code-3 data-6 CODE: 3 DATA DESCRIPTION Solenoid Freeze Parameters 0 A On time 1 B Off Time 2 C Impacts Solenoid Buzz Parameters 3 D On Time 4 B Off Time 5 C Impacts Temperatures 6 G Supercool 7 H Post-Freeze 8 I TE Cooler Off 9 J Freeze Difference 10 K Buzz Difference 11 L Coldest Calibration 12 M Slope 13 N Intercept 14 Y High Pt. Coefficient 15 O No. of Calibration Samples 16 P Reference Solution A 17 Q Reference Solution B 18 Z Reference Solution C Probes 19 R Block Bin 20 S Sample Bin Misc. 21 U Plateau Mode 22 V Serial Number NOTE: The shared CODE:3 is a bug, recorded as #123 on 11/6/2006. Care should be taken when decoding any serial port logs after a code-3 data-6 sequence. SERIAL COMMANDS (instrument must be in factory mode) CODE: 3 DATA 0? Help DIAGNOSTIC ROUTINES CODE: 4 DATA 0 A/D TEST DESCRIPTION 1 ~ Reset Instrument 2 = Instrument Status 3 B Beep 4 C Add Comment to File 5 D Toggle Curve Trace On/Off 6 E Edit Parameters 7 H Block Temp (H+xxxxx mc) 8 I Impact (Ixx ms) 9 L Change Language 10 M A/D Channel Utilities 11 P Print Test 12 S Set Date/Time 13 T Show Date/Time 14 # Clear Parameters 15 1 Parameter Contents 16 2 Dump Results 17 3 Dump Events 18 4 Statistics 19 5 Manual Calibration 20 6 Memory Contents 21 0 Burn Parameters 22 Extra Carriage Returns DESCRIPTION 1 PROBE BIN TEST 2 SOLENOID TEST 3 DISPLAY/PRINTER TEST 4 KEYPAD/BEEPER TEST 5 BAR CODE TEST Fiske Micro-Osmometer Service Manual
ERRORS CODE: 5 DATA DESCRIPTION LEVEL 1 ERRORS 0 System Error: NMI 1 System Error: Trap 2 System Error: ESC 3 Memory Allocation Error LEVEL 2 ERRORS 100 Fan Driver Failure 101 TE Driver Failure 102 Solenoid Driver Failure 103 Block Probe Open? 104 Block Probe Failure 105 Sample Probe Open? 106 Sample Probe Failure 107 Cooling System Error 108 Event Record Lost LEVEL 3 ERRORS 200 Sample Pre-Freeze 201 Sample Freeze Error 202 Sample Did Not Freeze 203 No Plateau 204 Test Cancelled 205 Calibration Cancelled 206 Test Time-Out Error 207 Recalibration Needed 208 Standards Reversed? 209 Calibration Out of Range 210 Reset Probe Config. 211 Parameter RAM Failed 212 No Parameters in RAM 213 Count Error: Parameter 214 ROM Serial Number Error 215 Baudrate Error 216 Key Input Timeout 217 Lift Sample Probe 218 A/D Init Failure 219 A/D High Filter Error ERRORS (continued) CODE: 5 DATA SETUP ROUTINES CODE: 6 DATA DESCRIPTION 0 FEATURE LOCKOUT 1 SET BIN NUMBERS 2 SET KEY BEEP 3 SET TIME 4 SET SERIAL RATE 5 SET SERIAL NUMBER 6 SET LANGUAGE SETTING CHANGE CODE: 7 DATA UTILITY ROUTINES CODE: 8 DATA DESCRIPTION 0 RECALL RESULTS 1 STATISTICS 2 EVENT RECORD 3 ASSISTANCE DESCRIPTION NOTE: No subcategories. EVENT_TEST CODE: 9 DATA DESCRIPTION NOTE: Data is the final result of a successful test. CALIB ROUTINE CODE: 10 DATA DESCRIPTION 220 A/D Low Filter Error 221 A/D Cal Mode Error 222 Bar Code Cancelled 223 Diagnostics Cancelled 224 Low Battery 225 Out of Range DESCRIPTION NOTE: No subcategories. Fiske Micro-Osmometer Service Manual
Notes: Fiske Micro-Osmometer Service Manual
Appendix B Symbol Definitions
Symbol Definitions The following symbols may appear in product literature, or on intruments or product packaging. On-Off Caution Hot Surface Feed Dangerous Voltage Interrupt Lifting Hazard Test Calibrator Start Content Stop Control Record Review Negative Control Setup Positive Control Calibration Flammable Cancel; Delete Fragile Functional Arrow Irritant Printer Keep Dry Enter Date Manufactured RS232 Sterile Bar Code Non-Sterile Attention Serial Number Fiske Micro-Osmometer Service Manual
Solution Use By; Expiration Date Do Not Open Top Authorized Representative Handle With Care Part Number Toxic x Sufficient for [x] Tests Use Blade To Open Open Here Do Not Re-Use Low Fluid Level For In Vitro Diagnostic Use Keep Hands Clear European Conformity Latex-Free Temperature Limit Diluent See Instructions See Instructions for Temperature Guidelines Lot Number Potential Puncture Hazard Manufacturer Electronic Equipment - Dispose of Properly Fiske Micro-Osmometer Service Manual
Appendix C Product Disposal and Recycling
Product Disposal and Recycling International concern about environmental pollution resulting from improper disposal of products and materials at the end of their useful life has resulted in an increase in legislation to control the methods and procedures used to handle waste electrical and electronic equipment. While the regulatory status in some regions of the world has progressed to the point where formal legislation is already in effect, many other regions are in the process of creating similar legislation or adopting that already in existence in other areas. The result in the years ahead will be more stringent control over disposal of products and recycling of their components once they are withdrawn from use. and restrictions in effect to ensure proper disposal, and available locations where products can be sent. Contact Fiske Associates Hot-Line Service: 800 225-1002 (toll-free within the USA and Canada; after normal business hours, dial extension 2191) +US 781-320-5656 (elsewhere) 781-320-0811 (fax) Service personnel will provide you with contact information for local disposal, or instructions for returning the product directly to Fiske Associates. Since regulations governing the disposal of your instrument and accessories may vary depending upon your geographic location, the following guidelines are provided to assist you in identifying the options available to you once the decision has been made to replace or dispose of this product: Contact the supplier who sold you the product. Whether this was Fiske Associates itself, or one of its authorized dealers, this supplier should be knowledgeable about the national and local regulations governing disposal and recycling of products in your area. In some cases, this supplier may be legally obligated to accept the product from you and arrange for proper disposal or recycling with no further involvement on your part. Alternately, the supplier can provide you with specific instructions for actions that you can take to dispose of the product properly. Contact your local government agency responsible for waste collection and disposal. They can identify procedures Fiske Micro-Osmometer Service Manual
Notes: Fiske Micro-Osmometer Service Manual
Appendix D Service Log
Service Log To call for Service: (781) 320-5656 or (800) 225-1002 FAX (781) 320-0811 Model Number 210 Serial Number Symptoms Repairs, Replacements or Problems or Changes Serviced by
Symptoms Repairs, Replacements or Problems or Changes Serviced by
Index
Index A A/D Tests..........................29 Application Board Description..........18 B Back Panel I/O Board Description.......18 Barcode Test........................29 C Central Processor Unit (CPU) Board Description.................18 Conventions......................9, 11 D Diagnostic Menu....................29 Display/Print Test....................29 E Event Record................Appendix A G General Description and Purpose.........8 General Overview...................17 I Instrument Software Update............18 Introduction.........................7 K Keypad............................18 Key/Beeper Test.....................29 M Maintaining the Instrument............23 P Power Supply.......................18 Probe Bin Test......................29 R Replacement Parts...................19 S Safety..............................9 Service Assistance....................8 Software Update.....................18 Solenoid Cleaning Procedure...........25 Solenoid Test.......................29 Specifications........................7 Systems Overview...................18 T 210 Board Set Description.............18 Troubleshooting Table................31 Fiske Micro-Osmometer Service Manual