[ method guidelines ] Method Guidelines. *OST: Oligonucleotide Separation Technology. Contents I. PRINCIPLES OF OLIGONUCLEOTIDE SEPARATIONS

Similar documents
[ Care and Use Manual ]

Guide to Reverse Phase SpinColumns Chromatography for Sample Prep

How To Use An Acquity Qda Detector

[ Care and Use Manual ]

ProteinChip Energy Absorbing Molecules (EAM)

A Complete Solution for Method Linearity in HPLC and UHPLC

Automated Method Development Utilizing Software-Based Optimization and Direct Instrument Control

The Theory of HPLC. Gradient HPLC

[ Care and Use Manual ]

ENrich SEC 70 ENrich SEC 650 High-Resolution Size Exclusion Columns Instruction Manual

[ Care and Use Manual ]

Protein Separation Technology

Purification of reaction mixtures using flash chromatography.

LC-MS/MS Method for the Determination of Docetaxel in Human Serum for Clinical Research

Oasis HLB Cartridges and 96-Well Plates

Extraction of Epinephrine, Norepinephrine and Dopamine from Human Plasma Using EVOLUTE EXPRESS WCX Prior to LC-MS/MS Analysis

Method Development for Size-Exclusion Chromatography of Monoclonal Antibodies and Higher Order Aggregates

WATERS QUANTITATIVE ANALYSIS solutions

UHPLC/MS: An Efficient Tool for Determination of Illicit Drugs

Application Note. USD 2995 (a) High Throughput Regeneration Study on MEP HyperCel Mixed-Mode Sorbent on ScreenExpert RoboColumns u

Fast, Reproducible LC-MS/MS Analysis of Dextromethorphan and Dextrorphan

ApplicationNOTE Introduction

Transforming a ChromLab Software 2-D Purification Template into an Automated Multidimensional (Multi-D) Purification Workflow An Instructional Guide

MEPS - Micro Extraction by Packed Sorbent Online SPE for GC and LC sample preparation - Extraction to injection in a single process

Determination of Pesticide Residues in Drinking Water Using Automated Solid-Phase Extraction and Gas Chromatography with Nitrogen Phosphorus Detection

Application Note. Purifying common light-chain bispecific antibodies using MCSGP. Summary

High-Throughput 3-D Chromatography Through Ion Exchange SPE

Application Note. Separation of three monoclonal antibody variants using MCSGP. Summary

The Use of Micro Flow LC Coupled to MS/MS in Veterinary Drug Residue Analysis

Analysis of Free Bromate Ions in Tap Water using an ACQUITY UPLC BEH Amide Column

CONFIRMATION OF ZOLPIDEM BY LIQUID CHROMATOGRAPHY MASS SPECTROMETRY

Thermo Scientific SOLAµ SPE Plates Technical Guide. Consistent excellence. for bioanalysis

Peptide synthesis, radiolabelling and radiochemical analysis

[ DISQUE DISPERSIVE SAMPLE PREPARATION EXTRACTION PRODUCTS ] QuEChERS. Simplifed

LUMEFANTRINE Draft proposal for The International Pharmacopoeia (October 2006)

An In-Gel Digestion Protocol

SIMULTANEOUS DETERMINATION OF TELMISARTAN AND HYDROCHLOROTHIAZIDE IN TABLET DOSAGE FORM USING REVERSE PHASE HIGH PERFORMANCE LIQUID CHROMATOGRAPHY

Application Note. Determination of Nitrite and Nitrate in Fruit Juices by UV Detection. Summary. Introduction. Experimental Sample Preparation

Analysis of Fat-Soluble Vitamin Capsules using UltraPerformance Convergence Chromatography (UPC 2 )

Materials for Pharmaceutical Manufacturing

[NUGENESIS SAMPLE MANAGEMENT ] AMPLE IMPROVING LAB EFFICIENCY, ANAGEMENT ACCELERATING BUSINESS DECISIONS. bigstock.com $69

Tamsulosin Hydrochloride Capsules

Develop a Quantitative Analytical Method for low (» 1 ppm) levels of Sulfate

RESOURCE Q, 1 ml and 6 ml RESOURCE S, 1 ml and 6 ml

Training Courses 2014 HPLC GC SPE

Analysis of Vegetable Oils by High Performance Liquid Chromatography

Aspects of industrial purification of peptides using large-scale chromatography. Lars Andersson and Jonas Persson

Determination of Fat in Dried Milk Products Using Accelerated Solvent Extraction (ASE)

SPE, LC-MS/MS Method for the Determination of Ethinyl Estradiol from Human Plasma

Thermo Scientific Prelude SPLC System FPO. Making LC/MS accessible. to clinical research and toxicology labs

Analysis of the Vitamin B Complex in Infant Formula Samples by LC-MS/MS

Intelligent use of Relative Response Factors in Gas Chromatography-Flame Ionisation Detection

Thermo Scientific HyperSep Solid Phase Extraction Method Development Guide

Simultaneous Quantitation of 43 Drugs in Human Urine with a Dilute-and-Shoot LC-MS/MS Method

Using Natural Products Application Solution with UNIFI for the Identification of Chemical Ingredients of Green Tea Extract

Comparison of the Reversed-Phase Selectivity of Solid Core HPLC Columns

Separation of Peptides from Enzymatic Digestion on Different Acclaim Columns: A Comparative Study

Strength in Technology HPLC COLUMNS. Product Guide

Foreign Taxes Paid and Foreign Source Income INTECH Global Income Managed Volatility Fund

Electrospray Ion Trap Mass Spectrometry. Introduction

Selective Testosterone Analysis in Human Serum by LC-FAIMS-MS/MS

Analytical Test Report

Determination of Calcium and Magnesium in Brine

Efficient Multi-Well Protein Purification Strategies

Reversed Phase High Presssure Liquid Chromatograhphic Technique for Determination of Sodium Alginate from Oral Suspension

HiPer Ion Exchange Chromatography Teaching Kit

Henry Shion, Robert Birdsall, Steve Cubbedge, and Weibin Chen Waters Corporation, Milford, MA, USA APPLICATION BENEFITS INTRODUCTION WATERS SOLUTIONS

SUCRALOSE. White to off-white, practically odourless crystalline powder

Dissolved and precipitated oxalate

Amino Acid Analyzer L-8900

This handbook presents the basic principles of reversed-phase HPLC for the

Application Note. Increasing the activity of monoclonal antibody isoforms by MCSGP. Summary

Simultaneous determination of aspartame, benzoic acid, caffeine, and saccharin in sugar-free beverages using HPLC

Diafiltration: A Fast, Efficient Method for Desalting, or Buffer Exchange of Biological Samples

Quantitative analysis of anabolic steroids in control samples from food-producing animals using a column-switching LC-HESI-MS/MS assay

QUANTITATIVE AMINO ACID ANALYSIS. Aurélie Lolia Applications Manager, Biochrom Ltd

Rapid Screening Method for Illicit Drugs, Using an Advanced Solid Core UHPLC Column and UHPLC System with MS/MS Detection

Image Lab Software for the GS-900 Densitometer

DEVELOPMENT OF PFCs HIGH SENSITIVITY ANALYSIS METHOD APPLIED RETENTION GAP TECHNIQUE WITH UPLC/MS/MS

Extraction of Cannabinoids in Marijuana and Edibles by QuEChERS

Supported Payment Methods

Accuracy counts! SENSORS WITH ANALOG OUTPUT

In-Depth Qualitative Analysis of Complex Proteomic Samples Using High Quality MS/MS at Fast Acquisition Rates

Gas Chromatography. Let s begin with an example problem: SPME head space analysis of pesticides in tea and follow-up analysis by high speed GC.

Supported Payment Methods

HPLC Analysis of Acetaminophen Tablets with Waters Alliance and Agilent Supplies

Pall in the Brewery (more value per hectolitre) DE Free Clarification FB1795

The Automated SPE Assay of Fipronil and Its Metabolites from Bee Pollen.

PrepTip. Reverse Phase PrepTip User Guide

Chemical analysis service, Turner s Green Technology Group

IOOF QuantPlus. International Equities Portfolio NZD. Quarterly update

HPLC Basics, Equipment, and Troubleshooting

High sensitivity assays using online SPE-LC-MS/MS -How low can you go? Mohammed Abrar Unilabs York Bioanalytical solutions, York, UK

What is Eluent Generation?

amplification tech Optical Design of CFX96 Real-Time PCR Detection System Eliminates the Requirement of a Passive Reference Dye

Pipette Tips PACKARD BIOSCIENCE MultiPROBE II COMPATIBLE

Thermo Scientific Dionex Chromeleon 7 Chromatography Data System Software

The Essential CHROMacademy Guide. Mobile Phase Optimization Strategies for Reversed Phase HPLC

[ Care and Use Manual ]

Transcription:

Method Guidelines *OST: Oligonucleotide Separation Technology columns are based on Waters second generation of hybridsilica BEH Technology particles and can be effectively used for the lab scale purification and analysis of detritylated synthetic oligonucleotides using ionpair, reversedphase chromatography. This document provides useful method guidelines for the effective use of this column chemistry for this group of compounds. Contents I. PRINCIPLES OF OLIGONUCLEOTIDE SEPARATIONS II. III. IV. SAMPLE PREPARATION RECOMMENDED MOBILE PHASES RECOMMENDED INJECTOR WASH SOLVENT V. GENERAL CONSIDERATIONS IN DEVELOPING SEPARATIONS VI. ANALYSIS OF MODIFIED OLIGONUCLEOTIDES VII. PURIFICATION CONSIDERATIONS

I. P RINCIPLES OF OLIGONUCLEOTIDE SEPARATIONS Separations of detritylated synthetic oligonucleotides on an XBridge OST C 18 column are based on ionpair, reversedphase chromatographic principles (IPRPLC). As shown in Figure 1, the ionpairing additive in the mobile phase is adsorbed on a hydrophobic sorbent and provides for chargetocharge interactions with negative charges contained on the oligonucleotide backbone (e.g., phosphate groups). Figure 2: Separation of a 15 60mer Deoxythymidine Ladder on Figure 1: Proposed Mechanism of IPRPLC for Synthetic Oligonucleotide Separations + + + + TEA As a result, an efficient chargebased (lengthbased) oligonucleotide separation is achieved (Figure 2). Gradient elution using an acetonitrile or methanol eluent displaces both ionpairing agent and the oligonucleotides from the sorbent surface. Separation selectivity and resolution decreases with increasing oligonucleotide length (Figure 2) making the separation of long oligonucleotides challenging. Modified oligonucleotides such as phosphorothioates and 2O alkyl modified species are also more difficult to analyze. Special mobile phase may be required (see Section III, Recommended Mobile Phases). chain PO group on Oligo chain + Sample Injected: Approximately 100 pmoles of a detritylated 15 60mer oligonucleotide ladder diluted in 0.1 M TEAA Waters, 2.5 µm (2.1 x 50 mm) Mobile Phases: A: 0.1 M TEAA, B: Acetonitrile / 0.1M TEAA, 20/80, v/v 0.2 ml/min Column Temp.: Gradient delay: 0.45 ml Gradient: 40 to 62.5% B in 30 minutes (812.5% acetonitrile, 0.15% acetonitrile per minute) 260 nm, 5 scans per second Two commonly used ionpairing agents for oligonucleotide applications are triethyl ammonium and dimethylbutyl ammonium ions. The final ph of these mobile phases containing either of these ionpairing reagents is adjusted by the addition of Acetic Acid, or in some cases, Hexafluoroisopropanol (HFIP). These mobile phases are volatile making them suitable for LCMS applications. The ability to adequately resolve synthetic oligonucleotide mixtures by ionpair, reversedphase chromatography is significantly affected by the particle size of the material contained in an efficiently packed column (see Figure 3). Consequently, columns are efficiently packed with 2.5 micron material to maximize detritylated oligonucleotide component resolution. In order to improve oligonucleotide separation efficiency and speed, elevated separation temperature (e.g. ) is recommended. Elevated temperature will also reduce operating LC System back pressure. 2

Figure 3: Effectiveness of Waters BEH Technology HybridSilica C 18 Particle Size on Deoxythymidine Ladder Separations Figure 4: Impact of Ionpairing System on Separation of a 1030mer Heterooligonucleotide Ladder Sample Injected: Mobile Phases: Column Temp.: Gradient delay: Gradient: Approximately 100 pmoles of detritylated 15 60mer crude oligonucleotide ladder diluted in 0.1 M TEAA Waters BEH HybridSilica C 18 particles (2.1 x 50 mm) A: 0.1 M TEAA, B: Acetonitrile / 0.1M TEAA, 20/80, v/v 0.2 ml/min 0.45 ml 40 to 62.5% B in 30 minutes (812.5% acetonitrile, 0.15% acetonitrile per minute) 260 nm, 5 scans per second Sample: 3 exonuclease Mobile phases: Column Temp.: Gradient delay: 20 mer: TCC CTA GCG TTG AAT TGT CC 25 mer: TCC CTA GCG TTG AAT TGT CCC TTA G 30 mer: TCC CTA GCG TTG AAT TGT CCC TTA GCG GGT Ladder was prepared by hydrolyzing detritylated 20, 25, and 30mer oligonucleotides with a Waters, 2.5 µm (4.6 x 50 mm) Upper chromatogram: 0.1 M TEAA with acetonitrile gradient; Lower chromatogram: 16.3 mm TEA 400 mm HFIP with methanol gradient 1.0 ml/min 0.45 ml 260 nm, 5 scans per second In addition to ionpairing, a hydrophobic reversedphase mechanism also takes place in the oligonucleotide separation. The residual interaction of nucleobases has an impact on overall retention and separation selectivity, especially when using Triethylammonium Acetate (TEAA) ionpairing mobile phases. Separation of N and N1mers may be either enhanced or suppressed by the sequence contribution. More potent ionpairing systems such as Triethylammonium ion with Hexafluoroisopropanol counter ion provide for more regular chargebased separations (Figure 4). II. SAMPLE PREPARATION 1. Dissolve the detritylated synthetic oligonucleotide sample in Mobile Phase A (e.g., 0.1 M TEAA). For example, a 0.05 0.2 µmole scale synthesis can be prepared in 0.1 ml of 0.1 M TEAA. Proportionately larger or smaller volumes of 0.1M TEAA are required when dissolving samples from different scale syntheses. Due to the nature of gradient separations, relatively large volumes of sample (in low organic strength eluent) can be injected and concentrated onto the head of the column before beginning the gradient elution program. 2. Samples must be completely in solution and free of particulates before injecting onto the column. Remove all particles from the sample (Controlled Pore Glass Synthesis Support, etc.), which may block the inlet column frit, increase the operating pressure, and shorten the column life time. Sample contamination with high concentration of salts and/or detergents may also interfere with analysis. 3

3. To remove particulates the sample may be filtered with a 0.2 μm membrane. Be sure that the selected membrane is compatible and does not dissolve with the selected Mobile Phase diluent. Contact the membrane manufacturer with solvent compatibility questions. An alternative method of particulate removal involves centrifugation for 20 minutes at 8,000 rpm, followed by the transfer of the supernatant liquid to an appropriate vial. III. RECOMMENDED MOBILE PHASES The most common ionpair mobile phase for synthetic oligonucleotide separations is based on Triethylammonium Acetate (TEAA). This mobile phase can be prepared by titrating Glacial Acetic Acid aqueous solution with Triethylamine (TEA). Note: To maximize column life, it is ESSENTIAL that all prepared OST Mobile Phases be filtered through a solvent compatible, 0.45 µm membrane and contained in bottles that are clean and particulate free. TEAA 1L of 0.1 M TEAA may be prepared as follows: 1) Perform work in a hood. 2) Add 5.6 ml of glacial Acetic Acid into 950 ml of water and mix well. 3) Slowly add 13.86 ml of TEA. 4) The ph should be adjusted to ph 7 +/ 0.5 by careful addition of Acetic Acid. 5) Adjust final volume to 1 L with water. Alternatively, premixed TEAA can be used [(e.g., Sigma 1 M TEAA (part no. 90357)]. Mix 100 ml with 900 ml of water to prepare 1 L of 0.1 M TEAA mobile phase. Alternative ionpairing reagents are recommended for improved separation of phosphorothioates or when performing LCMS analyses. An ionpairing mobile phase based on Triethylamine (TEA) and Hexafluoroisopropanol (HFIP) as the buffering acid produces an efficient eluent system for improved separations involving these application types. As indicated below, two ionpairing systems are useful. For routine detritylated oligonucleotide applications, aqueous buffer consisting of 8.6 mm TEA and 100 mm HFIP is effective. For applications such as those involving the separation of Grich oligonucleotides, it is advisable to use aqueous buffer consisting of 15 mm TEA and 400 mm HFIP (ph 7.9). TEAHFIP System 1 1L of 8.6 mm TEA / 100 mm HFIP is prepared as follows: 1) Perform work in a hood 2) Add 10.4 ml of HFIP (16.8 g) into 988.4 g of water and mix well. 3) Slowly add 1.2 ml of TEA. 4) The ph is approximately 8.3 +/ 0.1. TEAHFIP System 2 1 L of 15 mm TEA / 400 mm HFIP is prepared as follows: 1) Perform work in a hood 2) Add 41.56 ml (67.17 g) of HFIP into 956.36 g of water and mix well. 3) Slowly add 2.08 ml (1.52 g) of TEA. 4) The ph of final buffer is approximately 7.9 +/ 0.1. IV. RECOMMENDED INJECTOR WASH SOLVENTS Between analyses, the HPLC system injector seals should be washed. A 90% Water / 10% Acetonitrile injector wash solvent is recommended. V. GENERAL CONSIDERATION IN DEVELOPING SEPARATIONS Separation of detritylated synthetic oligonucleotides by ionpair, reversedphase chromatography uses very shallow gradients. With both TEAA and TEAHFIP ionpairing systems, a rate of strong eluent change between 0.10.25 % Acetonitrile (or Methanol) per minute is recommended. However, the formation of shallow gradients can place performance demands on LC pumps and mixers that can compromise the quality of the separation. Consequently, it is strongly advised that Mobile Phase B formulation contain a premix blend of aqueous and organic solvents (e.g., Mobile Phase A= 0.1 M TEAA and Mobile Phase B = Acetonitrile / 0.1M TEAA, 20/80, v/v) to minimize potentially inadequate solvent mixing that can compromise component resolution. 4

As illustrated in Figures 5 through 7, these analyses were performed with the following mobile phases: Mobile Phase A: 0.1 M TEAA Mobile Phase B: Acetonitrile (ACN) containing 0.1 M TEAA, 20:80 (v:v) The 0.1% ACN / min gradient change from an initial 5 to 10% Acetonitrile concentration over 50 minutes was programmed as specified in Table 1: Table 1 Time % A % B Actual Acetonitrile (ACN) Concentration 0 min 75 25 5% 50 min 50 50 10% Example: For the initial 5% Acetonitrile concentration: Initial %B = desired ACN % / Volume Fraction of ACN in Mobile Phase B. So, initial %B = 5% / 0.2 = 25% For the final 10% Acetonitrile concentration: Final %B = desired ACN % / Volume Fraction of ACN in Mobile Phase B. So, final %B = 10% / 0.2 = 50% With TEAA mobile phases, the unmodified oligonucleotides elute within a 710 % ACN gradient window. However, C and G rich oligonucleotide sequences are generally less retained (i.e., elute within a 58% ACN gradient window) than A and T rich sequences (i.e., elute within a 811% ACN gradient span). When using a shallow gradient, the total length of analysis for an unknown sample sequence may be excessive. Use of a fast scouting gradient with a 1% ACN per minute change is recommended in such cases. Information gathered from this scouting analysis can then be used to create a more appropriate and time efficient set of gradient conditions for the particular sample. Gradient slope has a direct impact on the achievable oligonucleotide component resolution (along with the type of ionpairing agent, sequence, and oligonucleotide modification). Steeper gradients (e.g., 1% ACN change per minute on a 4.6 x 50mm column at a 1.0 ml/min flow) are recommended for labeled oligonucleotides or for short, 515 mer sequences. Separation of longer sequences are typically performed using more shallow gradient slopes (e.g. 0.15% ACN change per minute on a 4.6 x 50mm column at a 1.0 ml/min flow). The organic solvent concentration at initial sample loading conditions has to be well chosen. If the initial organic solvent strength is too high, then some desired oligonucleotide sequences may be unretained. In the other extreme, when the gradient starts with too low an organic concentration, the analysis is excessively long without the benefit of enhanced component resolution. A suitable gradient separation method can be approximated from the oligonucleotide base (C, G, A, and T) composition. The initial gradient is typically adjusted while keeping the gradient slope constant. Table 2: Suggested Gradient Conditions for NonStandard Detritylated Synthetic Oligonucleotide Sequences Gradient 1 [Standard oligonucleotides (1)] Gradient 2 [High GC content or short oligonucleotides (2)] Gradient 3 [High AT content or long oligonucleotides (3)] Initial % ACN 7.00% 5.25% 7.50% Final % ACN 10.75% 9.00% 12.50% Gradient Length(4) 15 min 15min 20min 1: Standard oligonucleotides: 10 30mers 2: Short oligonucleotides: Less than 10mer 3: Long oligonucleotides: 30 60mers 4: Assuming use of a 2.1 x 50mm column at a flow of 0.2 ml/min and a separation temperature of. The retention of single and dual dyelabeled oligonucleotides is dictated by the nature of label. For example, the retention of 25 mer oligonucleotide increases according to the type of label attached as follows: no label<6fam< <TAMRA<TET<HEX<Cy3. VI. ANALYSIS OF MODIFIED OLIGONUCLEOTIDES columns are suitable for analysis of unmodified as well as modified detritylated oligodeoxyribonucleotides and oligoribonucleotides. Phosphorothioate and 2 O alkyl modified oligonucleotides can also be analyzed with IPRPHPLC method. However, these full length oligonucleotide products are usually more difficult to resolve from their shorter length failure sequences. The recommended ionpair system for phosphorothioate oligonucleotide analysis is TEAHFIP (see Recommended mobile phases). An example of a 25mer phosphorothioate oligonucleotide analysis is shown in Figure 5. 5

Figure 5: Analysis of a Digested 25mer Phosphorothioate Oligonucleotide Figure 6: Analysis of a 20mer Oligodeoxythymidine Crude Synthesis Mixture Sample: Mobile phase: Column Temp.: Gradient delay: Gradient: Detritylated 25mer phosphorothioate oligonucleotide mix (CTC TCG CAC CCA TCT CTC TCC TTC T) digested with 3 exonuclease, 2.5 µm (2.1 x 50 mm) A: 15 mm TEA with 400 mm HFIP B: methanol 0.2 ml/min 0 ml 15 to 20% B in 20 minutes (0.25% methanol per minute) 260 nm, 2 scans per second Sample: Mobile phase: Column Temp: Gradient Delay: Gradient: ~600 pmol of a detritylated 20mer, ~18 pmol of 19mer, ~4.5 pmol of 17mer was injected on column., 2.5 µm (2.1 x 50mm) A: 0.1 M TEAA, B: Acetonitrile / 0.1M TEAA, 20/80 (v/v) 0.2 ml/min 0 ml 35 to 50% B in 30 minutes (710% acetonitrile) 260 nm, 2 scans per second Peptide nucleic acids (PNA) can also be analyzed using columns. The ionpairing system recommended for analysis of PNA is similar to those used for peptide analysis (0.1% Trifluoroacetic Acid or Formic Acid). UV detection of eluted oligonucleotide peaks is often performed at 260 nm. Injection of 50 pmol of detritylated oligonucleotide sample on a 2.1 x 50mm column yields relatively abundant peaks. Limits of quantitation (LOQ) vary with the type of oligonucleotide, LC system and detector; LOQ generated on Waters 2996 PDA detector equipped with micro UV cell is approximately 1 pmol (2.1 x 50mm ). The Limit of detection (LOD) estimate is shown in Figure 6. VII. PURIFICATION CONSIDERATIONS columns are designed for laboratory scale oligonucleotide purifications and analyses. Sufficient amount of isolated material suitable for molecular biology and other experiments can be prepared in a single injection. For example, a 4.6 X 50 mm XBridge OST C 18 column can suitably purify approximately 20200 nmoles of sample in a single injection. It is important to understand that column overloading results in a peak broadening and that some earlier eluting impurities may coelute with the component of interest. With a proper heartcutting technique, a good purity of the target oligonucleotide can be obtained without significant yield sacrifice (Figure 7). Chromatographers frequently develop a separation on the analytical scale before moving to preparative work. The steps required to optimize the analytical separation involve: 1) Selecting the appropriate column packing material and mobile phase. 2) Determining the optimal flow rate, gradient during and separation temperature. 3) Determining the amount of material that can be satisfactorily loaded and separated on the analytical scale column. 6

Once the separation has been optimized, one begins preparing for the preparative separation. The steps to successfully scale a separation from an analytical to a preparative column, containing the same packing material composition, are detailed below Step A: Calculate the flow rate for use on the Preparative column. Preparative Column Flow Rate = Analytical Column Flow Rate x (Diameter of Prep Column) 2 (Diameter of Anal Column) 2 Step B: To get similar chromatography, the gradient elution profile should be created on both columns using the same number of column volumes. When the analytical and preparative columns are of the same length, as is recommended for this application, then the gradient duration should be the same. Note: This assumes use of the same flow rate linear velocity for both runs as calculated above. For preparative runs, it is also important to note that an initial gradient delay is required to allow the entire sample to load onto the head of the column prior to beginning chromatography. * Custom Column ** Values are only approximate and vary depending on detritylated oligonucleotide length, base composition, and heartcutting fraction collection method used Figure 7 shows the separation of 90 nmoles of a detritylated 30 mer deoxythymidine crude reaction mixture on a 4.6 x 50mm XBridge OST C 18 column. The collection interval is suggested by the lines. Due to partial column overloading, the N 1, N 2... impurities are partially displaced and elute earlier than expected. With the proper hearthcutting technique, 9598% purity is typically achieved for 1535 mer oligonucleotides at this purification scale. Figure 7: Purification of a Detritylated 30mer Deoxythymidine Sample Step C: The last calculation involves determining how much sample can be loaded on the preparative column. This calculation compares the relative volumes of the two columns assuming that both columns are the same length as recommended for this application. Preparative Column Sample Load = Previously Determined Analytical Column Sample Load x (Diameter of Prep Column) 2 (Diameter of Anal Column) 2 Sample: Mobile phase: Column Temp.: Gradient delay: Gradient: Crude detritylated 30mer oligothymidine, 200 nmole dissolved in 100 µl of mobile phase A, 45 µl was injected on column, 2.5 µm (4.6 x 50mm) A: 0.1M TEAA with 400 mm HFIP B: Acetonitrile/0.1M TEAA, 20/80 (v/v) 1.0 ml/min 0 ml (compensated) 35 to 65% B in 24 minutes (713% ACN, 0.25% ACN per minute) 260 nm, 2 scans per second Table 3: Column Selection Guide for Detritylated Oligonucleotide Purification Column (mm) Approx Mass Load (µmoles)** Flow Rate (ml/min) 2.1 x 50 0.04 0.2 4.6 x 50 0.20 1.0 10.0 x 50 1.00 4.5 19.0 x 50* 4.00 16.0 30.0 x 50* 9.00 40.0 50.0 x 50* 25.00 110.0 Table 4: Ordering Information Description Particle Size Pore Size Dimension Part No. 2.5 μm 135Å 2.1 x 50 mm 186003952 2.5 μm 135Å 4.6 x 50 mm 186003953 2.5 μm 135Å 10.0 x 50 mm 186003954 Custom 186003955 7

Sales Offices Australia Waters Australia Pty. Limited Tel: 2 9933 1777 Fax: 2 9898 1455 Denmark Waters A/S Tel: 45 46 59 8080 Fax: 45 46 59 8585 Italy Waters S.P.A. Tel: 39 02 274 211 Fax: 39 02 250 1827 Singapore Waters Asia Ltd. Tel: 65 6273 1221 Fax: 65 6273 6116 Austria and European Export (Central South Eastern Europe, CIS and Middle East) Tel: 431 877 18 07 Fax: 431 877 18 08 Belgium Waters S.A.N.V. Tel: 32 2 726 1000 Fax: 32 2 726 1100 Brazil Waters Technology Tel: 55 11 5094 3788 Fax: 55 11 5093 6413 Canada Waters Limited Tel: 800 252 4752 Fax: 905 678 9237 China Waters Technologies (Shanghai) Limited Tel: 8621 6495 6999 Fax: 8621 6495 1999 CIS/Russia Tel: +7 495 727 4490 Fax: +7 495 336 7000 Czech Republic Tel: 42 02 617 11384 Fax: 42 02 617 11386 Finland Waters Finland Tel: (09) 5659 6288 Fax: (09) 5659 6282 France Waters S.A Tel: (33) 1 30 48 72 00 Fax: (33) 1 30 48 72 01 Germany Waters GmbH Tel: 49 6196 400600 Fax: 49 6196 4006010 Hungary Tel: 36 1 350 5086 Fax: 36 1 350 5087 Hong Kong Waters China Ltd. Tel: 852 29 64 1800 Fax: 852 25 49 6802 India Waters India Pvt. Ltd. Tel: 91 80 2 837 1900 Fax: 91 80 2 839 2157 Ireland Waters Chromatography Ireland Ltd. Tel: 353 1 448 1500 Fax: 353 1 448 1510 Japan Nihon Waters K.K. Tel: (81) 3 3471 7191 Fax: (81) 3 3471 7118 Korea Waters Korea Limited, Tel: (82) 2 820 2700 Fax: (82) 2 820 2730 Mexico Waters S.A. DE C.V. Tel: 5255 5200 1860 Fax: 5255 5524 9376 The Netherlands Waters Chromatography B.V. Tel: +31 (0)7650 87 200 Fax: +31 (0)7650 87 280 Norway Waters AS Tel: 47 63 84 60 50 Fax: 47 63 84 60 51 Poland Tel: (48) 22 833 4400 Fax: (48) 22 833 0987 Puerto Rico Waters Technologies Corporation Tel: 787 747 8445 Fax: 787 747 8448 Spain Waters Cromatografía, S.A. Tel: 34 93 600 93 00 Fax: 34 93 600 93 60 Sweden Waters Sverige AB Tel: 46 8 555 11500 Fax: 46 8 555 11550 Switzerland Waters AG Tel: 41 62 889 2030 Fax: 41 62 889 2059 Taiwan Tel: 886 2 2543 1898 Fax: 886 2 2543 1918 United Kingdom Waters U.K. Ltd. Tel: 44 208 238 6100 Fax: 44 208 207 7070 All Other Countries Waters Corporation 34 Maple Street Milford, MA 01757 USA Tel: 508 478 2000 800 252 4752 Fax: 508 872 1990 http://www.waters.com Waters, XBridge, BEH Technology and BioAlliance are trademarks of Waters Corporation. 2008 Waters Corporation. April 2008 715001476 Rev B SC Waters Corporation 34 Maple Street Milford, MA 01757 U.S.A. T: 1 508 478 2000 F: 1 508 872 1990 www.waters.com