Druck UPS-III Loop Calibrator

Similar documents
VOLTAGE/CURRENT CALIBRATOR ISO-TECH ILC-421

Druck PV 62x pneumatic/hydraulic pressure stations

DC Electronic Loads 8500 series

User's Guide. Integrating Sound Level Datalogger. Model Introduction

Manual Ranging MultiMeter

1. SAFETY INFORMATION

Druck DPI 104. GE Industrial Sensing. Digital pressure indicator User manual - K394

Model UT33A: OPERATING MANUAL Table of Contents (1)

Ground Resistance Clamp On Tester

Pocket Tach 99 (PT99) Non-Contact Tachometer

AutoRanging Digital MultiMeter

Auto-ranging Digital Multimeter INSTRUCTION MANUAL

OPERATION MANUAL. Pen type, separate electrode SOIL PH METER Model : PH-220S

User Guide. Temperature and Humidity Datalogger. Model 42280

Thermistor Thermometer MODEL NO

Differential Pressure Sensors for air and non-corrosive gases, with calibration certificate

Model UT713 OPERATING MANUAL

User Guide. Heavy Duty Dissolved Oxygen Meter. Model

DAB+ / FM Tuner Model: TU-201

IDEAL INDUSTRIES, INC. TECHNICAL MANUAL MODELS:

UT202A Operating Manual. Contents

Instruction Manual. 2in1 LAN Tester & Multimeter. Model: LA-1011

Wahl C50 Calibrator FEATURES

Model UT33B/C/D: OPERATING MANUAL Table of Contents (1)

User's Guide. Digital Psychrometer. Model RH390. F C WBT Dew Point MAX MIN

Milwaukee USER MANUAL. Milwaukee. Smart DO Meter PORTABLE DISSOLVED OXYGEN METER MODEL: SM600. Authorized Dealer: ISMIL600 11/01

SATEX SA Manual. Weegtechniek holland b.v. PRECISION BALANCE. Patroonsweg DA Zeewolde HOLLAND

Drayton Digistat +2RF/+3RF

Automation System TROVIS 6400 TROVIS 6493 Compact Controller

Contents. Document information

Digital I/O: OUTPUT: Basic, Count, Count+, Smart+

User Guide. 4 wire Earth Resistance Tester Model GRT300

Integrating Sound Level Meter and Datalogger

Loop Calibration and Maintenance

RDJ10RF/SET. Wireless room temperature controller with 24-hour time switch and LCD. Programmable, for heating systems

User Guide. Model Insulation Tester / Megohmmeter

Electrical data Nominal voltage AC/DC 24 V Nominal voltage frequency

SOUND LEVEL METER 94405

XP Series Balances Operation Manual

T0118 T2118 T3118. Instruction Manual

Hand-held thermometer Model CTH7000

PRODUCT PROFILE FEATURES

MILLENIUM INSTALLATION MANUAL NTR 735 A. Simple Automation Control Module (MAS)

VZ-M7 HDMI Field Monitor 7 Class HD LCD

MAKING MODERN LIVING POSSIBLE. living connect. Installation and User Guide. Danfoss heating

RS Stock No Instruction Manual RS-1340 Hot Wire Anemometer

User's Guide. Pinless Moisture Meter. Model MO257

Product Information Sheet UH28CS-12kV Hi Pot Tester

PREPARATION. INTERNAL CALIBRATION (No external weight needed.) SPAN CALIBRATION

Operating instructions Platform/floor scales

EVC40 EMERGENCY VOICE COMMUNICATION SYSTEM

FREQUENCY RESPONSE ANALYZERS

Daker DK 1, 2, 3 kva. Manuel d installation Installation manual. Part. LE05334AC-07/13-01 GF

Ground Resistance Clamp-On Tester

Handheld portable appliance testers

10 Multimeter. Users Manual

Druck DPI 142/150. GE Infrastructure Sensing. Pressure Indicator Calibration manual - K382

EPM3. Phase Sequence and Motor Rotation Tester. Users Manual

Remote Alarm Monitoring System

QUICK START GUIDE FOR DEMONSTRATION CIRCUIT BIT DIFFERENTIAL ADC WITH I2C LTC2485 DESCRIPTION

Programmable Room Thermostat 7 Day (5-2 Day) Models: / Power Supply: Battery / Mains

Instruction Manual FL

Proportional pressure regulators VPPE

AN500T, AN1000, AN1000T, AN1500, AN1500T AN2000, AN2000T

Multifunction Process Calibrator

Quick Start Guide , Rev DA March Rosemount 8714D (Calibration Standard) Magnetic Flowtube Simulator

e-4 AWT07MLED 7 Q TFT LCD MONITOR (LED Backlighted) USER MANUAL

Always there to help you. Register your product and get support at PD7001B. Question? Contact Philips.

RI-215A Operator s Manual. Part Number: RK Revision 0 Released: 10/3/05

SVPRIBOR ALFA DSL. Users Manual

Power Supply and Indicator Unit Type Fig. 1 Type Mounting and Operating Instructions EB 9539 EN

HERZ-Thermal Actuators

Wireless 7 Day Programmable Room Thermostat

Frequency Converter Fv

User's Guide Integrating Sound Level Datalogger Model

Dear Customer! Thank you for choosing our product.

5320A Multifunction Electrical Tester Calibrator. Verify and calibrate electrical test tools with a single instrument

The table below lists the symbols used on the Clamp and/or in this manual. Important Information. See manual.

Model UT58A/B/C: OPERATING MANUAL

TECHNICAL DATASHEET #TD1404AX PWM CONTROLLED SOLENOID DRIVER

BlueSolar Pro Remote Panel For BlueSolar PWM-Pro charge controllers 12/24V 5, 10, 20, 30A Article number SCC

DPI 610/615 Series. Druck Portable. Pressure Calibrators. GE Sensing. Features. Setting the Standard for Portable. Pressure Calibrators

How To Use A Power Supply Unit (Upu)

The PACE5000 Modular Pressure Controller/Indicator

MAKING MODERN LIVING POSSIBLE. living connect. Installation and User Guide. Danfoss Heating Solutions

Hygro-Thermometer + InfraRed Thermometer Model RH101

CONTENTS. 1. Fitting the Battery

TX OPEN RS232/485 module

Technical Specifications: The specifications represent a particular hardware platform. Application-specific software is provided.

Electric Rotary Actuator - On/Off or programmable control actuator

DAB & FM RADIO Installation Manual

i410/i1010 AC/DC Current Clamp

Series /16 DIN Multi-Mode Bar Graph Display Timer TIMERS PRODUCT HIGHLIGHTS

OPERATING INSTRUCTIONS

DRTS 33. The new generation of advanced test equipments for Relays, Energy meters, Transducers and Power quality meters

Model UT60A: OPERATING MANUAL. Table of Contents. Page

FC6 FOOT CONTROLLER OWNER S MANUAL

J TB/TW Temperature limiter, temperature monitor

User's Manual. Research Isometric Transducer 0 to 5 grams and 0 to 50 grams

Transcription:

GE Sensing Druck UPS-III Loop Calibrator User manual -K0317

Approved Service Agents For the list of service centres visit our web site: www.gesensing.com Symbols This equipment meets the requirements of all relevant European safety directives. The equipment carries the CE mark. This symbol, on the instrument, indicates that the user should refer to the user manual. Do not dispose of this product as household waste. Use an approved organisation that collects and/or recycles waste electrical and electronic equipment. For more information: Contact us at www.gesensing.com

UPS III Loop Calibrator Introduction The Druck UPS III Series of loop calibrators can supply power (source mode) and produce readings (measure mode) to perform field calibrations on 2-wire devices. The set-up menu enables the user to source or measure in either voltage or current and to perform continuity tests. These user instructions include the operation, safety instructions and installation requirements for the loop calibrator. Specifications Accuracy Values in table below includes temperature effects 17 C to 27 C Outside these limits... 0.003%/ C(0.0015%/ F) Calibration Reference... 22 C ±1 C/RH45%±15% Mode Range Accuracy Remarks Source ma 0 to 24 ma* 0.01% rdg + 2 lsd V-max. 75V Source ma + 24V Measure ma Measure ma + 24V 0 to 24 ma* 0.01% rdg + 2 lsd R-max 1kΩ at 20 ma 0 to 24 ma* 0.01% rdg + 2 lsd V-max. 75V 0 to 24 ma* 0.01% rdg + 2 lsd R-measure 15Ω Measure V 0 to 60V* 0.02% rdg + 4 lsd R-measure 1MΩ Continuity <100Ω ** - 1 ma * Resolution 0.001 lsd least significant digits ** Audio + visual rdg reading Hart communications...menu selectable 220Ωloop resistor Operating Temperature... -10 C to 50 C (-14 F to 122 F) Storage Temperature... -20 C to 70 C (-4 F to 158 F) This loop calibrator meets the essential protection requirements of the relevant EEC directives. Conforms to... EN61010, EN 61326-1(1997)+ A1(1998) Electrical Power Supply... 4 x 1.5 V alkaline size AA or Universal power supply {see accessories} 1 K0317 Issue No. 3

Physical Dimensions... 77 x 129 x 24 mm(3 x 5 x 1 ) Weight... 275 grams (9.7 oz.) Terminals...4 mm sockets {gold plated} Case... High impact ABS Relative Humidity... 0 to 90% Safety This symbol, on the loop calibrator, indicates that the user should refer to the user guide or manual. Batteries Remove batteries from the loop calibrator immediately when discharged and before storage. Dispose of batteries in accordance with local regulations and battery manufacturers' instructions. When storing and transporting batteries make sure they cannot be short circuited. Power Supply The power supply for this loop calibrator can be the internal non-rechargeable batteries or the external Universal power supply unit (see accessories). Battery life The display shows > 75 hours in measure mode > 18 hours at 12 ma (source mode) with low battery power. Battery Replacement Unscrew and remove the securing screw from the battery panel. Replace the batteries, check the polarity of the batteries. Refit and secure the battery panel. Accessories Assy 305 Test lead set 191-129 Power supply, Universal, 100-240 V a.c. 47-63Hz 38016 Carrying case 38023 Protective rubber boot K0317 Issue No. 3 2

OPERATION Keys The key switches the calibrator on and off. Press and hold for 2 seconds. The key changes the measure or source operating mode. Pressing the keys makes menu selections, sets numerical values and controls step and ramp functions (up/down). The select advanced functions shown on the bottom of the display. When no key is pressed for 10 minutes, the calibrator times out and switches off. To disable this automatic time out, select autpower down in the set-up menu. 3 K0317 Issue No. 3

Operating Modes Pressing switches the instrument on and the display shows the start-up sequence. Pressing shows the information screen: F1, at this time, the display F1 Information Serial number VERSION CAL. DATE HART BATTERY Pressing F2, at this time, the display shows the set-up screen: F2 Contrast F2 adjust Resolution F2 adjust Range F2 4-20 ma/0-20ma Source unit F2 ma/% Hart F2 ON/OFF Aut power down F2 ON/OFF Decimal F2., / Access code F2 menu Calibration F2 menu The calibrator can be used in two modes measure or source. Measure mode The display shows the measured value; depending on the settings made in set-up and advanced settings: F1 When measuring current pressing enables linear or flow, F2 pressing enables ma or % (value of 4 to 20 ma or 0 to 20 ma). F2 When measuring voltage pressing changes the resolution between 0.00V and 0.000V. To measure continuity the displays shows an open or closed switch symbol with an audible signal on switch closure. K0317 Issue No. 3 4

Connect the loop calibrator to the device to be tested: and Measure ma Press the mode key and select [Measure ma]. External power supplies Vmax = 60 V for the loop. The calibrator measures the current flow of the loop. Closed loop current measurement from transmitter test terminal. * test terminal Measure ma with 24 V Press mode key and select [Measure ma and 24V]. The calibrator supplies 24 V for the loop, maximum 24 ma 5 K0317 Issue No. 3

Measure Volts Press mode key and select [Measure V], measure range 60V, maximum impedance 1 Mohm. Continuity Test Press mode key and select [Continuity Test]. Pressing switches the audible signal on/off. K0317 Issue No. 3 6

Source Mode The display shows the source value in ma or % value of 4 to 20 ma or 0 to 20 ma, linear or flow depending on the settings made in set-up and advanced settings. Source ma Press mode key and select [Source ma]. The calibrator supplies maximum output of: 24 ma; Vmax = 60; receiver input Rmax = 1kOhm. Source ma with 24V Press mode key and select [Source ma and 24V]. The calibrator supplies maximum loop power of: 24 V and 24 ma. 7 K0317 Issue No. 3

Advanced Options in a Source mode Press the key and select ma Source or ma Source & 24V. MODE Use and (Enter) to select the function. Press the key (Advanced) and the display shows: Linear simulates linear transmitters. Flow simulates flow transmitters. Valve simulates valve control signals. Use and (Enter) to select the Advanced option: Advanced Step Auto-step Span Check Ramp 25% steps for linear and flow - fixed values for valve. The same as step with a timed step interval. Step between 4 (or 0) ma and 20 ma. Automatic ramp between 4 (or 0) ma and 20 ma. Note: Ramp function not available for valve selection. Use to quit. The display returns to the selected source mode with the advanced setting available. Mode ma source ma source & 24V select F2 Enter F1 Advanced Lin Flow Step Auto-step Time (M:SS) Valve F2 select Span Check Ramp F2 select Adjust +1/ -1 second F2 Time (M:SS) Adjust +1/ -1 second F2 K0317 Issue No. 3 8

Operation of Advanced Options Press the key to switch the advanced setting on and off: e.g. on or off Press or to: step the output up or down. step the span check maximum or minimum start the ramp. Press then to start: continuous auto-step. or continuous ramp cycle. Hart Application V ma ma ( 24V ) COM +24.000 ma +100.0% Advanced % Hart communicator 220 menu selectable loop resistor This application allows ma measure and source modes to be used through the Hart communicator. 9 K0317 Issue No. 3

blank page K0317 Issue No. 3 10

Maintenance Return the loop calibrator to an authorised repair centre for any repairs, it cannot be repaired on-site. To keep the loop calibrator accurate a calibration check should be carried out once per year. Cleaning Clean the loop calibrator case with a moist, lint-free cloth and weak detergent. 11 K0317 Issue No. 3

Battery Replacement Only use the battery type listed on page one. Unscrew and remove the securing screw from the battery panel. Replace the batteries, check the polarity of the batteries. Refit and secure the battery panel. K0317 Issue No. 3 12

Calibration Instructions General The instrument is supplied by the manufacturer, complete with calibration certificate(s). A calibration period of 12 months is recommended. The actual calibration interval depends on instrument usage and the total measurement uncertainty acceptable for the specified application. The UPS-III is a very precise measuring instrument and the test equipment and conditions of test must be suitable for the type of work. The calibration check and calibration adjustment should be carried out in a controlled environment by a calibration technician*. The manufacturer offers a comprehensive and, if required, UKAS accredited calibration service. * A calibration technician must have the necessary technical knowledge, documentation, special test equipment and tools to carry out the calibration work on this equipment. Calibration Equipment The following tables give the accuracy requirements for the calibration equipment and the UPS-III. Calibration requires a stable temperature of 21 ±1 C (70 ±2 F). UPS-III measure mode Applied ma Table 1 ma measure Permitted UPS-III error (ma) Calibrator error (ma) 0 0.002 0 4 0.002 0.00014 12 0.002 0.00030 20 0.002 0.00046 13 K0317 Issue No. 3

Table 2 V measure Applied V Permitted UPS-III error (mv) Calibrator error (mv) 0 0.004 0.00040 20 0.004 0.00014 40 0.005 0.00064 50 0.005 0.00070 UPS-III source mode Applied ma Table 3 ma source Permitted UPS-III error (ma) Calibrator error (ma) 0 0.002 0 4 0.002 0.00012 12 0.002 0.00011 20 0.002 0.00015 Calibration Check 1.Connect the UPS-III to the electrical calibrator. Switch on the electrical calibrator and allow it to thermally stabilise. 2.Switch on the UPS-III and allow the instrument to thermally stabilise. 3.Set the UPS-III to ma measure, adjust the electrical calibrator to apply the first value in the table 1. Record the reading of the UPS-III. 4.Repeat step 3 for all the values in the table 1. 5.Compare the recorded values and the applied values. If the difference is greater than the permitted error, the instrument requires a calibration adjustment. 6.Repeat this procedure for V measure (table 2) and ma source (table 3). K0317 Issue No. 3 14

Calibration Adjustment 1.Connect the UPS-III to the electrical calibrator. Switch on the electrical calibrator and allow it to thermally stabilise. 2.Switch on the UPS-III and press, within two seconds to select Calibration. Enter the access code [9410 factory setting] and allow the instrument to thermally stabilise. 3.Select the parameter required for calibration. Use the display menu to select the calibration values. After a successful calibration enter the new calibration date. 15 K0317 Issue No. 3

K0317 Issue No. 3 16

17 K0317 Issue No. 3

K0317 Issue No. 3 18