Accuracy and Precision Performance of the Accu-Chek Active System. Introduction. Method

Similar documents
Blood Glucose Monitoring: The Facts about Accuracy

Evaluation of Accuracy and User Performance of the TRUE METRIX Self-Monitoring Blood Glucose System *

Accuracy Test Omnitest 3 Blood Glucose Monitoring System. B. Braun Diabetes Care

Inform II. Quick Reference Guide. for Healthcare Professionals BLOOD GLUCOSE MONITORING SYSTEM

An important first step in identifying those at risk for Cardiovascular disease The Accutrend Plus system: from the makers of the ACCU-CHEK and

A comparison of blood glucose meters in Australia

Validation of measurement procedures

SKUP Scandinavian evaluation of laboratory equipment for primary health care

Analytical Test Method Validation Report Template

Applying Statistics Recommended by Regulatory Documents

Validation and Calibration. Definitions and Terminology

Accu-Chek Inform II system. Professional glucose testing for the wireless age

Content Sheet 7-1: Overview of Quality Control for Quantitative Tests

USING CLSI GUIDELINES TO PERFORM METHOD EVALUATION STUDIES IN YOUR LABORATORY

Blood Glucose Monitoring Test Systems for Prescription Point-of- Care Use Draft Guidance for Industry and Food and Drug Administration Staff

1. What is the critical value for this 95% confidence interval? CV = z.025 = invnorm(0.025) = 1.96

SureStep Pro Linearity Test Kit and Analytical Measurement Range Verification (AMR)

Quantifying measurement error from digital instruments

Basic Statistics and Data Analysis for Health Researchers from Foreign Countries

PROCEDURE NO. POC LBH. Printed copies are for reference only. Please refer to the electronic copy for the latest version.

Analytical Methods: A Statistical Perspective on the ICH Q2A and Q2B Guidelines for Validation of Analytical Methods

To avoid potential inspection

I know my value. CoaguChek XS Plus system Smart INR monitoring at your practice. 1 of 6

Method Validation/Verification. CAP/CLIA regulated methods at Texas Department of State Health Services Laboratory

Adequacy of Biomath. Models. Empirical Modeling Tools. Bayesian Modeling. Model Uncertainty / Selection

Guidance for Industry

Lab 2. Spectrophotometric Measurement of Glucose

510(k) SUBSTANTIAL EQUIVALENCE DETERMINATION DECISION SUMMARY

at The Valley Hospital (TVH) for Nursing Students/Nursing Instructors 2012

VALIDATION OF ANALYTICAL PROCEDURES: TEXT AND METHODOLOGY Q2(R1)

ANALYTICAL METHODS INTERNATIONAL QUALITY SYSTEMS

Measurement Systems Correlation MSC for Suppliers

Unit 31 A Hypothesis Test about Correlation and Slope in a Simple Linear Regression

A guidebook for people with diabetes

Experiment #1, Analyze Data using Excel, Calculator and Graphs.

Biochemistry Validation Form

Performance Variability of Seven Commonly Used Self-Monitoring of Blood Glucose Systems: Clinical Considerations for Patients and Providers

Test strips and meter designed for glucose self-measurement manufactured by Abbott. Report from an evaluation organised by SKUP

Hypertension and Diabetes Status Bangladesh Demographic and Health Survey

In vitro diagnostic reagent, calibrator and control material stability

510(K) SUMMARY FEB Submitter: Name of the device: * Common Name: Clinical Chemistry Analyzer (with optional ISE Module) Attachment #2

Core Fittings C-Core and CD-Core Fittings

US Diabetic Devices Industry Research Report: KenResearch

Diabetes Monitoring Diary

For In Vitro Diagnostic Use

SMF Awareness Seminar 2014

Blood Glucose Management

Introduction 1 The system 1 The meter 2 The display 3 The mode 3 The measurement 4 Coding the meter 4 How to obtain a drop of blood 6 Application of

FDA Public Hearing on Clinical Accuracy Requirements for Point of Care Blood Glucose Meters (BGMs)

11. Analysis of Case-control Studies Logistic Regression

for patient monitoring

Accu-Chek Connect Online diabetes management system. Setup Guide for Healthcare Providers

GUIDELINES FOR THE VALIDATION OF ANALYTICAL METHODS FOR ACTIVE CONSTITUENT, AGRICULTURAL AND VETERINARY CHEMICAL PRODUCTS.

Basic research methods. Basic research methods. Question: BRM.2. Question: BRM.1

Control Charts and Trend Analysis for ISO Speakers: New York State Food Laboratory s Quality Assurance Team

Samsung POINT OF CARE Systems - Cardiac/Acute Care Biomarkers LABGEO IB Clinical Chemistry Analytes LABGEO PT

USE OF ARIMA TIME SERIES AND REGRESSORS TO FORECAST THE SALE OF ELECTRICITY

PRINCIPLE. REF (200 tests/cartridge) REF (400 tests/cartridge) ANNUAL REVIEW Reviewed by: Date. Date INTENDED USE

How To Run Statistical Tests in Excel

Self-Monitoring of Blood Glucose in Noninsulin- Treated Diabetes

Point-of-care (POC) versus central laboratory instrumentation for monitoring oral anticoagulation

HEMATOCRIT/HCT AND CALCULATED HEMOGLOBIN/HB

ICH Topic Q 2 (R1) Validation of Analytical Procedures: Text and Methodology. Step 5

Sample Size Planning, Calculation, and Justification

Report reference guide. mmol/l. One. solution EMR. No fuss

Evaluating Current Practices in Shelf Life Estimation

BLOOD GLUCOSE METER / GLYCOMÈTRE. User s Manual Manuel de l'utilisateur

AP Physics 1 and 2 Lab Investigations

Self-Monitoring Of Blood Glucose (SMBG)

Data Management Implementation Plan

Practice Advisory Blood Glucose Meter Accuracy Issued September 16, 2013

Medtronic MiniMed Solutions Pumps & Meters Software. Model MMT-7311 Version 7.0

Introduction to Minitab and basic commands. Manipulating data in Minitab Describing data; calculating statistics; transformation.

Assay Migration Studies for In Vitro Diagnostic Devices Guidance for Industry and FDA Staff

Blood Glucose Monitoring Systems

QUALITY ASSURANCE and QUALITY CONTROL DATA STANDARD

Title: Fingerstick Glucose by Accu-Chek Inform

How To Understand And Understand The Concept Of Intermediate Hyperglycaemia

Basic Lessons in Laboratory Quality Control

Design of Experiments for Analytical Method Development and Validation

CHAPTER THREE COMMON DESCRIPTIVE STATISTICS COMMON DESCRIPTIVE STATISTICS / 13

Statistical Rules of Thumb

Supplementary appendix

X X X a) perfect linear correlation b) no correlation c) positive correlation (r = 1) (r = 0) (0 < r < 1)

Data Analysis on the ABI PRISM 7700 Sequence Detection System: Setting Baselines and Thresholds. Overview. Data Analysis Tutorial

The Economic Impact and Cost-Effectiveness of Glucose Monitoring

Carbohydrate Ratio Correction Target Correction Factor

1.0 XRF Analyses Performance Evaluation

OFFICE OF CATHOLIC SCHOOLS DIOCESE OF ARLINGTON DIABETES MEDICAL MANAGEMENT PLAN Page 1 of 5 TO BE COMPLETED BY PARENT OR GUARDIAN

Qualification Study CHO 360-HCP ELISA (Type A to D)

Validating Methods using Waters Empower TM 2 Method. Validation. Manager

Interpretation of Somers D under four simple models

Statistics Review PSY379

Getting Started 5. Using the TRUEmanager 7. Patient Management 10. Reports 43. Database 57. Options 64

Transcription:

Accuracy and Precision Performance of the Accu-Chek Active System I. ACCURACY The accuracy of the system was assessed per the ISO 15197. Introduction The purpose of this study was to determine the accuracy of the Accu-Chek Active blood glucose system using one Accu-Chek Active blood glucose strip lot. Method Capillary blood from subjects diagnosed with diabetes was obtained at one external diabetes clinic. These results were compared to reference values obtained by using the hexokinase reference method. Testing was performed using Accu-Chek Active glucose test strips. Two Accu-Chek Active blood glucose meters were assigned for testing with one lot of blood glucose test strips. Per the ISO standard, blood glucose meter results must lie within the following ranges for each meter tested as stated in the table below: % Samples Glucose concentration (mmol/l) Glucose concentration (mg/dl) 5 < 2.8 < 50 15 > 2.8 4.3 > 50 80 20 > 4.3 6.7 > 80 120 30 > 6.7 11.1 >120 200 15 > 11.1 16.6 > 200 300 10 > 16.6 22.2 > 300 400 5 > 22.2 > 400 100 values (100 specimen) were used in the analysis per glucose meter, 2 meters were used. Glucose concentration was artificially altered for samples at less than 2.8 mmol/l (50 mg/dl) and greater than 22.2 mmol/l (400 mg/dl).

Results The Accu-Chek Active blood glucose test lot was analyzed by regression analysis according to Bablok/Passing and is summarized in the following table for lot number 228835. N Slope Intercept Correlation Slope CI Int. CI (mmol/l) (mmol/l) 200 0.984-0.135 0.998 (0.974, 0.993) (-0.202, -0.071) (mg/dl) (mg/dl) 200 0.984-2.4 0.998 (0.974, 0.993) (-3.64, -1.28) The results demonstrate excellent correlation with all values near the optimum value of 1.000. The following figure illustrates the regression graph according to Bablok/Passing. The capillary data for Accu-Chek Active strip lot # 228835 was analyzed by regression analysis according to Bablok/Passing is summarized as follows: for the Accu-Chek Active blood glucose monitoring system, the regression shows a slope of 0.984 with a 95 % confidence interval of (0.974, 0.993). The intercept is -0.135 mmol/l (-2.4 mg/dl). The data presented demonstrate excellent correlation with a value of 0.998 with 1.000 as the optimum value. No data were excluded. Roche Diagnostics Diabetes Care Method-Comparison: Regression System Capillary Blood Glucose vs Hexokinase Capillary Blood Glucose 35 30. 25 System [mmol/l] 20 15 n = 200 r = 0.998 tau = 0.962 BP (V) Y = -0.135 + 0.984 *X 45 º Y = 1. *X 10 5 0 0 5 10 15 20 25 30 35 Hexokinase [mmol/l]

The following tables illustrate the bias for the Accu-Chek Active system using test strip lot number 228835. Results less than 4.2 mmol/l (75 mg/dl) Within ± 0.28 mmol/l (Within ± 5 mg/dl) Within ± 0.56 mmol/l (Within ± 10 mg/dl) Within ± 0.83 mmol/l (Within ± 15 mg/dl) 23 / 34 (68 %) 32 / 34 (94 %) 34 / 34 (100 %) Results greater than or equal to 4.2 mmol/l (75 mg/dl) Within ± 5 % Within ± 10 % Within ± 15 % Within ± 20 % 116 / 166 (70 %) 160 / 166 (96 %) 164 / 166 (99 %) 166 / 166 (100 %) The minimum acceptable accuracy for results produced by a glucose monitoring system shall be as follows: Ninety-five percent (95 %) of the individual glucose results shall fall within ± 0.83 mmol/l (15 mg/dl) of the manufacturer s measurement procedure at glucose concentrations less than 4.2 mmol/l (75 mg/dl) and within ± 20 % of glucose concentrations greater than or equal to 4.2 mmol/l (75 mg/dl). The Accu-Chek Active glucose monitoring system meets the ISO 15197 requirements for accuracy. All of the 200 samples (100 %) were within the minimum acceptable performance criteria. II. INTERMEDIATE PRECISION Introduction The purpose of this study was to determine the intermediate precision of the Accu-Chek Active blood glucose system using one Accu-Chek Active blood glucose strip lot and three levels of Accu-Chek Active control solution. Intermediate precision is defined as precision under conditions where test results are obtained with the same method on identical test items in the same location, but where other variables such as operators, equipment, calibration, environmental conditions and/or time intervals differ.

Method Ten Accu-Chek Active blood glucose meters were assigned to this study. One blood glucose test strip was removed from each vial and placed into the designated meter. The test strips were dosed with control solution and the process was repeated for each meter over ten days for each assigned level of control solution. Results For each control solution, daily for a period of ten days ten different determinations were made on ten Accu-Chek Active blood glucose meters. From the resulting 100 determinations, the mean value was calculated and is shown below. Nonparametric methods were used to calculate a confidence interval for the mean standard deviation (SD). The following table lists the intermediate precision results using Accu-Chek Active control solutions and Accu-Chek Active blood glucose test strips: Results less than 4.2 mmol/l (75 mg/dl) Control Solution Level Mean Mean SD 95 %Confidence (mmol/l) (mmol/l) (mmol/l) 1 2.8 0.18 (0.16, 0.22) (mg/dl) (mg/dl) (mg/dl) 1 51 3.3 (2.9, 3.9) Results greater than 4.2 mmol/l (75 mg/dl) Control Solution Level Mean Mean SD 95 %Confidence Mean CV (mmol/l) (mmol/l) (mmol/l) (%) 2 8.6 0.19 (0.17, 0.23) 2.3 3 17.0 0.32 (0.28, 0.37) 1.9 (mg/dl) (mg/dl) (mg/dl) (%) 2 155 3.5 (3.1, 4.1) 2.3 3 306 5.8 (5.1, 6.7) 1.9

III. REPEATABILITY Introduction The purpose of this study was to determine the repeatability of the Accu-Chek Active blood glucose system using one Accu-Chek Active blood glucose strip lot. Repeatability is defined as precision under conditions where independent test results are obtained with the same method on identical test items in the same location by the same operator using the same equipment within a short interval of time. Method Ten Accu-Chek Active blood glucose meters were assigned to this study. One lot of Accu-Chek Active blood glucose test strips was tested. The glucose in a venous blood sample was allowed to degrade and concentrated glucose solution was added to this blood to achieve varying blood glucose concentrations. Once the manipulated blood sample had achieved stability, testing was performed on each of the ten blood glucose meters and the results were recorded. All blood testing occurred in one day. The test strips were dosed with blood and the process was repeated for each meter ten times (n = 100) for each assigned level of spiked venous blood. Results The following table lists the repeatability results using manipulated venous blood: Results less than 4.2 mmol/l (75 mg/dl) Mean Mean SD 95 %Confidence (mmol/l) (mmol/l) (mmol/l) 2.2 0.14 (0.12, 0.16) (mg/dl) (mg/dl) (mg/dl) 39 2.5 (2.2, 2.9)

Results greater than 4.2 mmol/l (75 mg/dl) Mean Mean SD 95 %Confidence Mean CV (mmol/l) (mmol/l) (mmol/l) (%) 4.6 0.17 (0.14, 0.19) 3.6 7.9 0.18 (0.15, 0.21) 2.3 9.7 0.19 (0.17, 0.22) 1.9 17.9 0.28 (0.24, 0.33) 1.6 (mg/dl) (mg/dl) (mg/dl) (%) 82 3.0 (2.6, 3.5) 3.6 142 3.2 (2.8, 3.8) 2.3 175 3.4 (3.0, 3.9) 1.9 323 5.0 (4.4, 5.9) 1.6 IV. CONCLUSION The Accu-Chek Active system meets the accuracy requirement for the ISO 15197 standard. ACCU-CHEK and ACCU-CHEK ACTIVE are trademarks of Roche. 2006 Roche Diagnostics. All rights reserved. 04788575001-0206