DEAE Sephacel. Sephacel. GE Healthcare

Similar documents
RESOURCE Q, 1 ml and 6 ml RESOURCE S, 1 ml and 6 ml

Blue Sepharose 6 Fast Flow

INSTRUCTIONS Edition AC

HiTrap Heparin HP, 1 ml and 5 ml

Process-scale purification of monoclonal antibodies polishing using Capto Q

Sterile ReadyToProcess Hollow Fiber Cartridges Instructions for Use

UniFlux 10: hollow fiber conversion kit ( ) installation instruction

His GraviTrap. GE Healthcare. Operation

HiTrap Chelating HP, 1 ml and 5 ml

MabSelect SuRe. GE Healthcare Life Sciences

How to Download a GE Healthcare License File

soft ware change description

CU-950 Installation Guide

GE Healthcare. UNICORN 6.x. SQL Server 2008 configuration for UNICORN installations

GE Healthcare. UNICORN 5.2 vs Software Change Description

ReadyKart Mobile Processing Station Instructions for Use

Amersham High Molecular Weight Calibration Kit for native electrophoresis

Gain efficiency in your process development with ÄKTA avant

ENrich SEC 70 ENrich SEC 650 High-Resolution Size Exclusion Columns Instruction Manual

How To Test For Cleaning Efficiency With A Predictor 96 Well Filter Plate

TECHNICAL BULLETIN. HIS-Select Nickel Affinity Gel. Catalog Number P6611 Storage Temperature 2 8 C

Page finder. 1. Legal Handling Safety warnings and precautions Storage Expiry 4

Summary of GE Healthcare's Quality Management System (QMS) Covering BioProcess chromatography media, equipment, software, and spare parts

HisTrap HP, 1 ml and 5 ml

Bio-Gel P Polyacrylamide Gel Instruction Manual

HiPer Ion Exchange Chromatography Teaching Kit

GE Healthcare Life Sciences. Validation Services. Compliance support through life cycle management

UNICORN. Hardening Guide

UNICORN control software

HisTrap HP, 1 ml and 5 ml

UNICORN 7.0 software. Description. gelifesciences.com

Instrument services for WAVE Bioreactor systems

Guide to Reverse Phase SpinColumns Chromatography for Sample Prep

Data File. Sephadex G-25 media and pre-packed columns. Introduction. Sephadex G-25 Bead structure. Desalting/buffer exchange and gel filtration

Aurum Ion Exchange Mini Kits and Columns. Instruction Manual

GAMP5 - a lifecycle management framework for customized bioprocess solutions

AG 501-X8 and Bio-Rex MSZ 501(D) Mixed Bed Resin Instruction Manual. Bio-Rad Laboratories, 2000 Alfred Nobel Dr., Hercules CA LIT205 Rev B

Affi-Prep Protein A Matrix Instruction Manual

UNO Q&S Continuous Bed Ion Exchange Column. Instruction Manual

Purification of reaction mixtures using flash chromatography.

Size Exclusion Chromatography

For in vitro preparation of human mononuclear cells from peripheral blood, bone marrow, and umbilical cord blood. Not for in vitro diagnostic use.

Purification of GST-tagged Proteins

A novel AIEX chromatography medium (resin) to remove IgA and IVIG purification process

The Theory of HPLC. Gradient HPLC

Elevated temp up to 150 C. HiT

illustra MicroSpin G-25 Columns

Determination of Calcium and Magnesium in Brine

Transforming a ChromLab Software 2-D Purification Template into an Automated Multidimensional (Multi-D) Purification Workflow An Instructional Guide

His Mag Sepharose excel. Ni Sepharose excel. HisTrap excel. gelifesciences.com

Pipette Tips PACKARD BIOSCIENCE MultiPROBE II COMPATIBLE

Fast Trak Training & Education

SODIUM CARBOXYMETHYL CELLULOSE

Oasis HLB Cartridges and 96-Well Plates

An In-Gel Digestion Protocol

ÄKTA system Training Guide

Recent advances in the purification of IgM monoclonal antibodies

Separation by Solvent Extraction

105 Adopted:

Smart Asset Management Services

Waters Corporation. Waters 2690/5 USER & TROUBLESHOOTING GUIDE

GE Healthcare s cell culture media, sera and process liquids

PrepTip. Reverse Phase PrepTip User Guide

Separation of Amino Acids by Paper Chromatography

Thermo Scientific HyperSep Solid Phase Extraction Method Development Guide

QUANTITATIVE AMINO ACID ANALYSIS. Aurélie Lolia Applications Manager, Biochrom Ltd

PAPER CHROMATOGRAPHY

SPE and HPLC. Dr Iva Chianella Lecturer in Analytical Chemistry Cranfield Health +44 (0)

Marmara Üniversitesi Fen-Edebiyat Fakültesi Kimya Bölümü / Biyokimya Anabilim Dalı PURIFICATION AND CHARACTERIZATION OF PROTEINS

8.9 - Flash Column Chromatography Guide

Ion Exchange Determination of Na+ by Displacement and Zn 2+ Using Preconcentration. Reading: Harris pp , 699

Gel Filtration Standard

Application Note. USD 2995 (a) High Throughput Regeneration Study on MEP HyperCel Mixed-Mode Sorbent on ScreenExpert RoboColumns u

Corning Spin-X UF Concentrator Selection and Use Guide

UNICORN MS SQL Server Integration

Diffusion and Fluid Flow

Glutathione Resin. User Manual. User Manual. Cat. Nos , , PT (071414)

Human serum albumin (HSA) nanoparticles stabilized with. intermolecular disulfide bonds. Supporting Information

Electrical Conductivity of Aqueous Solutions

Experiment 8 - Double Displacement Reactions

columns IonPac CS18 Cation-Exchange Column

Experiment SURFACE TENSION OF LIQUIDS. Experiment 1, page 1 Version of June 17, 2016

The Essential CHROMacademy Guide. Mobile Phase Optimization Strategies for Reversed Phase HPLC

Organic Chemistry Lab Experiment 4 Preparation and Properties of Soap

PART I SIEVE ANALYSIS OF MATERIAL RETAINED ON THE 425 M (NO. 40) SIEVE

Dow Consumer & Industrial Solutions TRITON CG-425 Alkyl Polyglucoside Surfactant

ÄKTA Laboratory-scale Chromatography Systems

CHT Ceramic Hydroxyapatite. Instruction Manual

High-throughput Process Development with PreDictor Plates

POROS CaptureSelect affinity columns for highspeed quantification of IgG Fc fusion proteins

Titrations. Acid-Base Indicators and Titration Curves. Shapes of Titration Curves. A titration curve is a graphical history of a titration

UltraClean Soil DNA Isolation Kit

Lewatit MP 62 is especially suitable for:

This value, called the ionic product of water, Kw, is related to the equilibrium constant of water

PowerFecal DNA Isolation Kit

GE Healthcare. Hydrophobic Interaction and Reversed Phase Chromatography. Principles and Methods

Troubleshooting Guide for DNA Electrophoresis

LUMEFANTRINE Draft proposal for The International Pharmacopoeia (October 2006)

Transcription:

GE Healthcare Instructions 71 7100 00 AC Ion Exchange DEAE Sephacel DEAE Sephacel is a weak anion exchanger based on beaded cellulose. The ion exchange group is diethylaminoethyl, which remains charged and maintains consistently high capacity over the entire working range, ph 2 9. DEAE Sephacel is macroporous and has an exclusion limit of approximately 1 000 000 daltons for globular proteins. Sephacel

Table of contents 1. Preparing the medium 3 2. Packing DEAE Sephacel 4 3. Using an adaptor 4 4. Equilibrating 5 5. Binding 6 6. Elution 6 7. Regeneration 6 8. Cleaning 7 9. Storage 7 10. Ordering information 7 p. 2

Table 1. Medium characteristics Type of ion exchanger: Total ionic capacity: Available capacity*: Bead structure: Bead size range: Mean particle size: Max linear flow rate**: Weak anion 0.11 0.14 mmol/ml medium 10 mg Thyroglobulin (MW 669 000)/ml drained medium 160 mg HSA (MW 68 000)/ml drained medium Beaded cellulose 40 160 μm 100 μm 50 cm/h ph working range: 2 9 ph stability*** long term: 2 12 short term: 2 12 Chemical stability: All commonly used aqueous buffers, 1.0 M acetic acid, 1.0 M NaOH, 8 M Urea, 8 M guanidine hydrochloride, ethanol, methanol etc. Physical stability: Negligible volume variation due to changes in ph or ionic strength. Autoclavable: In 0.1 M NaCl at 121 C for 30 min. * The available capacity was estimated in 0.5 M Tris, ph 8.3. The elution buffer contained 2 M NaCl. ** Linear flow rate = volumetric flow rate (cm 3 /h) column cross-sectional area (cm 2 ) *** The ranges given are estimates based on our knowledge and experience. Please note the following: ph stability, long term refers to the ph interval where the medium is stable over a long period of time without adverse effects on its subsequent chromatographic performance. ph stability, short term refers to the ph interval for regeneration and cleaning. 1. Preparing the medium DEAE Sephacel is supplied pre-swollen in 24 % ethanol. Prepare a slurry by decanting the 24 % ethanol solution and replace it with starting buffer in a ratio of 75 % settled medium to 25 % buffer. The starting buffer should not contain agents which significantly increase the viscosity. The column may be equilibrated with viscous buffers at reduced flow rates after packing is completed. p. 3

2. Packing DEAE Sephacel 1. Equilibrate all material to the temperature at which the chromatography will be performed. 2. De-gas the medium slurry. 3. Eliminate air from the column dead spaces by flushing the end pieces with buffer. Make sure no air has been trapped under the column net. Close the column outlet with a few centimeters of buffer remaining in the column. 4. Pour the slurry into the column in one continuous motion. Pouring the slurry down a glass rod held against the wall of the column will minimize the introduction of air bubbles. 5. Immediately fill the remainder of the column with buffer, mount the column top piece onto the column and connect the column to a pump. 6. Open the bottom outlet of the column and set the pump to run at the desired flow rate. This should be at least 133 % of the flow rate to be used during subsequent chromatographic procedures. However, the maximum flow rate, see Table 1, is typically employed during packing. Note: If you have packed at the maximum linear flow rate, do not exceed 75 % of this in subsequent chromatographic procedures. 7. Maintain the packing flow rate for 3 bed volumes after a constant bed height is reached. 3. Using an adaptor Adaptors should be fitted as follows: 1. After the medium has been packed as described above, close the column outlet and remove the top piece from the column. Carefully fill the rest of the column with buffer to form an upward meniscus. 2. Insert the adaptor at an angle into the column ensuring that no air is trapped under the net. 3. Make all tubing connections at this stage. There must be a bubble-free liquid connection between the column and the pump and column and the sample application system. p. 4

4. Slide the plunger slowly down the column so that the air above the net and in the capillary tubings is displaced by eluent. Valves on the inlet side of the column should be turned in all directions during this procedure to ensure that air is removed. 5. Lock the adaptor in position on the medium surface. Open the column outlet and start the eluent flow. Pass eluent through the column at the packing flow rate until the bed is stable. Re-position the adaptor on the medium surface as necessary. 4. Equilibrating Before starting a run, make sure that the ion-exchange bed has reached equilibrium. This is done by pumping binding buffer through the column until the conductivity and/or ph of the effluent is the same as that of the in-going solution. The column is now equilibrated and ready for use. Table 2. Suggested buffers for use with DEAE Sephacel. Buffer Counter ion Concentration pka (25 C) N-methylpiperazine Cl - 20 mm 4.8 piperazine Cl - 20 mm 5.7 HCOO - L-histidine Cl - 20 mm 6.2 bis-tris Cl - 20 mm 6.5 bis-tris propane Cl - 20 mm 6.8 triethanolamine Cl - 20 mm 7.8 CH 3 COO - Tris Cl - 20 mm 8.2 N-methyldiethanolamine 2- SO 4 50 mm 8.5 Cl - CH 3 COO - diethanolamine Cl - 20 mm at ph 8.4 8.9 50 mm at ph 8.8 1,3-diaminopropane Cl - 20 mm 8.6 ethanolamine Cl - 20 mm 9.5 piperazine Cl - 20 mm 9.7 1,3-diaminopropionate Cl - 20 mm 10.5 p. 5

5. Binding The most common procedure is to let the molecules of interest bind to the ion exchanger and allow the others to pass through. However, in some cases it may be more useful to bind contaminants and let the molecules of interest remain in the flow through. For adsorption, it is critical to choose a buffer with an appropriate ph. Please refer to Table 2. The ionic strength of the buffer should be kept low, so as not to interfere with sample binding. The recommended operating ph is within 0.5 ph units of the buffer s pka and at least one ph unit above the isoelectric point (pi) of the molecule of interest. 6. Elution Desorption is achieved using either an increasing salt gradient (continuous or step-wise) or an decreasing ph gradient (continuous or step-wise). 7. Regeneration Depending of the nature of the sample, regeneration is normally performed by washing with a high ionic strength buffer (e.g. 1 2 M NaCl) and/or decreasing ph, followed by re-equilibration in binding buffer. In some applications, substances such as denaturated proteins or lipids do not elute in the regeneration procedure. These can be removed by cleaning procedures described below. p. 6

8. Cleaning Remove ionically bound proteins by washing the column with 0.5 1 bed volume of a 2 M NaCl solution. Remove precipitated proteins, hydrophobically bound proteins and lipoproteins by washing the medium with 1 bed volume of a 0.1 M NaOH solution followed by binding buffer until free from alkali. Strongly hydrophobically bound proteins, lipoproteins and lipids can be removed by washing the medium with up to 70 % ethanol or 30 % isopropanol. Alternatively, wash the medium with 2 bed volumes of detergent in a basic or acidic solution. Use for example, 0.1 0.5 % non-ionic detergent (e.g. Triton X-100) in 0.1 M acetic acid. After treatment with detergent always remove residual detergents by washing with 5 bed volumes of 70 % ethanol. After cleaning the medium, re-equilibrate the ion-exchanger according to the recommendations above. 9. Storage Store the medium at 4 8 C in 20 % ethanol. 10. Ordering information Product Pack size Code No. DEAE Sephacel 500 ml 17-0500-01 p. 7

www.gehealthcare.com/protein-purification GE Healthcare Bio-Sciences AB Björkgatan 30 751 84 Uppsala Sweden GE Healthcare Bio-Sciences AB Björkgatan 30 SE-751 84 Uppsala Sweden GE Healthcare Europe GmbH Munzinger Strasse 5 D-79111 Freiburg Germany GE Healthcare UK Ltd Amersham Place Little Chalfont Buckinghamshire, HP7 9NA UK GE Healthcare Bio-Sciences Corp 800 Centennial Avenue P.O. Box 1327 Piscataway, NJ 08855-1327 USA GE Healthcare Bio-Sciences KK Sanken Bldg. 3-25-1, Hyakunincho Shinjuku-ku, Tokyo 169-0073 Japan Sephacel and Drop Design are trademarks of GE Healthcare companies. GE, imagination at work and GE monogram are trademarks of General Electric Company. Triton is a trademark of Union Carbide Chemicals and Plastics co. All goods and services are sold subject to the terms and conditions of sale of the company within GE Healthcare which supplies them. GE Healthcare reserves the right, subject to any regulatory and contractual approval, if required, to make changes in specifications and features shown herein, or discontinue the product described at any time without notice or obligation. Contact your local GE Healthcare representative for the most current information. 2006 General Electric Company All rights reserved. GE Healthcare Bio-Sciences AB, a General Electric Company. Elanders Tofters 2006 71-7100-00 AC