Gas Chromatography with FID

Similar documents
Chemistry 321, Experiment 8: Quantitation of caffeine from a beverage using gas chromatography

1.1 This test method covers the qualitative and quantitative determination of the content of benzene and toluene in hydrocarbon wax.

Determination of Caffeine in Beverages HPLC-1

Identification of Arson Accelerants by Gas Chromatography

Fractional Distillation and Gas Chromatography

ACETALDEHYDE and ISOVALERALDEHYDE (Gas Chromatography)

Background Information

NANYANG TECHNOLOGICAL UNIVERSITY. School of Civil and Environmental Engineering. CV2701: Laboratory 2A. Laboratory Manual. For

Waters Corporation. Waters 2690/5 USER & TROUBLESHOOTING GUIDE

6 H2O + 6 CO 2 (g) + energy

In this experiment, we will use three properties to identify a liquid substance: solubility, density and boiling point..

AIRMoN, ph and Conductivity Field Measurement

Distillation Experiment

Understanding Analytical Chemistry (Weighing, Mixing, Measuring and Evaluating)

Paper Chromatography: Separation and Identification of Five Metal Cations

EPA Method RSK 175 Dissolved Gasses in Water Matrices using the Teledyne Tekmar HT3 Headspace Analyzer. HT3 Application Note.

Determining the Quantity of Iron in a Vitamin Tablet. Evaluation copy

Solubility Curve of Sugar in Water

Dissolved and precipitated oxalate

Laboratory procedures in polymer chemistry

With proper calibration, the upper limit may be extended as needed.

Calibration of Glassware

GC METHODS FOR QUANTITATIVE DETERMINATION OF BENZENE IN GASOLINE

UNIVERSITY OF MINNESOTA DULUTH DEPARTMENT OF CHEMICAL ENGINEERING ChE ABSORPTION OF CARBON DIOXIDE INTO WATER

Project 5: Scoville Heat Value of Foods HPLC Analysis of Capsaicinoids

On-line Dissolved Gas Analysis

Expectations for GC-MS Lab

Enzyme Action: Testing Catalase Activity

The Determination of an Equilibrium Constant

ATOMIC ABSORTION SPECTROSCOPY: rev. 4/2011 ANALYSIS OF COPPER IN FOOD AND VITAMINS

Experiment 12- Classification of Matter Experiment

COMMON LABORATORY APPARATUS

CONFIRMATION OF ZOLPIDEM BY LIQUID CHROMATOGRAPHY MASS SPECTROMETRY

Environmental Water Testing: Surface Water, Groundwater, Hard Water, Wastewater, & Seawater

Determining Equivalent Weight by Copper Electrolysis

Osmosis. Evaluation copy

Mixtures and Pure Substances

Molar Mass of Butane

Determination of a Chemical Formula

EnAlgae SOP: Biogas analysis

Lead Testing and On Site Calibration for Water Testing Detection Range: 2 100ppb

Validation and Calibration of Analytical Instruments a D.Gowrisankar, b K.Abbulu, c O.Bala Souri, K.Sujana*

This study was performed by Patrick Sullivan and staff at TestAmerica Analytical Testing Corp., Westfield, MA, USA

The Determination of an Equilibrium Constant

18 Conductometric Titration

EXPERIMENT 10: TITRATION AND STANDARDIZATION

Sugar or Salt? Ionic and Covalent Bonds

NWTPH-HCID. Hydrocarbon Identification Method for Soil and Water

The Molar Mass of a Gas

Acid Base Titrations

Investigation M3: Separating Mixtures into Component Parts

Determination of Haloacetic Acids and Dalapon in Drinking Water by SPE and GC/ECD*

Catalase. ***You will be working with hot water, acids and bases in this laboratory*** ****Use Extreme Caution!!!****

Purification of reaction mixtures using flash chromatography.

GC/MS Analysis of Vanillin and Ethyl Vanillin in Distilled Spirits

[ Care and Use Manual ]

ISOLATION OF CAFFEINE FROM TEA

23 The Thermal Conductivity Detector

Halogen Free: What, Why and How. Presented by : Jim Cronin Environmental Monitoring and Technologies, Inc.

Coating and Extraction of Honeycomb Denuders

Freezing Point Depression: Why Don t Oceans Freeze? Teacher Advanced Version

Determination of the enthalpy of combustion using a bomb calorimeter TEC. Safety precautions

Direct Measurement Method Method to 20.0 mg/l (or 0 to 200% saturation) O 2 Clark-type Amperometric Sensor

15. Acid-Base Titration. Discover the concentration of an unknown acid solution using acid-base titration.

ANALYSIS OF FOOD AND NATURAL PRODUCTS LABORATORY EXERCISE

EUROPEAN COMMISSION DIRECTORATE-GENERAL TAXATION AND CUSTOMS UNION TAX POLICY Excise duties and transport, environment and energy taxes

Inc. Wuhan. Quantity Pre-coated, ready to use 96-well strip plate 1 Plate sealer for 96 wells 4 Standard (liquid) 2

HS 1003 Part 2 HS 1003 Heavy Metals Test

Chromatography...Oh Yeah!

Experiment 3: Extraction: Separation of an Acidic, a Basic and a Neutral Substance

STANDARDIZATION OF A SODIUM HYDROXIDE SOLUTION EXPERIMENT 14

Recovery of Elemental Copper from Copper (II) Nitrate

Making Biodiesel from Virgin Vegetable Oil: Teacher Manual

The basic division explosives for forensic purposes is shown in Fig. 1

Vitamin C quantification using reversed-phase ion-pairing HPLC

HEXANES. Insoluble in water, soluble in ether, alcohol, and acetone. Neutral to methyl orange (ph indicator) Not more than 0.

PAPER CHROMATOGRAPHY

Measurement and Calibration

Process Instrumentation

Name Lab #3: Solubility of Organic Compounds Objectives: Introduction: soluble insoluble partially soluble miscible immiscible

Experiment 5: Column Chromatography

Experiment 2 Kinetics II Concentration-Time Relationships and Activation Energy

Determining the Free Chlorine Content of Swimming Pool Water. HOCl H + + OCl. Evaluation copy

1. The Determination of Boiling Point

Physical Properties of a Pure Substance, Water

Chloride and Salinity

Analysis of Liquid Samples on the Agilent GC-MS

Reaction of Magnesium with Hydrochloric Acid (Gas Laws) Chemicals Needed:

Direct ISE Method Method to 1000 mg/l Na + Sodium ISE

105 Adopted:

Lab 6: Paper Chromatography. Pages Pre-lab page 151 No Post lab Chromatogram must be turned in attached to lab report

SOLUBILITY OF A SALT IN WATER AT VARIOUS TEMPERATURES LAB

CRYOCOOLER CONTAMINATION STUDY. S.W.K. Yuan, D.T.Kuo, and A.S. Loc. BEI Technologies, Cryocooler Group Sylmar, CA ABSTRACT

Milwaukee USER MANUAL

Nitrogen, Ammonia. Known Addition ISE Method 1 Method Minimum of 0.8 mg/l NH 3 N. Ammonia ISE. Test preparation. Instrument-specific table

Extraction of Caffeine from Energy Drinks

METHOD 8020A AROMATIC VOLATILE ORGANICS BY GAS CHROMATOGRAPHY

Lecture Chromo-3: Gas Chromatography. CHEM 5181 Fall 2004 Mass Spectrometry & Chromatography. Jessica Gilman and Prof. Jose-Luis Jimenez CU-Boulder

Transcription:

Gas Chromatography with FID Introduction Gas chromatography is an instrumental method for the separation and identification of chemical compounds. Chromatography involves a sample (or sample extract) being dissolved in a mobile phase (which may be a gas, a liquid or a supercritical fluid). The mobile phase is then forced through an immobile, immiscible stationary phase. The phases are chosen such that components of the sample have differing solubilities in each phase. A component that is quite soluble in the stationary phase will take longer to travel through it than a component that is not very soluble in the stationary phase but very soluble in the mobile phase. As a result of these differences in mobilities, sample components will become separated from each other as they travel through the stationary phase. The diagram of the separation process is shown on the Figure 1. Figure 1.

Our laboratory setup is presented in Figure 2. Figure 2. After the separation of the compounds, Flame Ionization Detector (FID) is used to identify each of them and determine their mass. The simple diagram of the FID is shown on the Figure 3. Figure 3. The effluent from the column is mixed with hydrogen and air, and ignited. Organic compounds burning in the flame produce ions and electrons that can conduct electricity through the flame. A large electrical potential is applied at the burner tip, and a collector electrode is located above the flame. The current resulting from the pyrolysis of any organic compounds is measured. FIDs are mass sensitive rather than concentration sensitive; this gives the

advantage that changes in mobile phase flow rate do not affect the detector's response. The FID is a useful general detector for the analysis of organic compounds; it has high sensitivity, a large linear response range, and low noise. It is also robust and easy to use, but it destroys the injected sample. After detection, a signal is sent to the recording device. You will see the development of the curve on the computer screen (Figure 4.) Figure 4. The time between sample injection and an analyte peak reaching a detector at the end of the column is termed the retention time (t R ). Each analyte in a sample will have a different retention time. The time taken for the mobile phase to pass through the column is called t M. A GC can separate the compounds, but cannot identify them itself. By calibrating GC you can find out at what time various organic compounds are being detected. The area under the curve (automatically calculated by the computer) may be expressed in terms of concentration of the pollutant, by running some calibration standards at known concentration. You will need: Laboratory procedure - 4 calibration samples of water contaminated with Toluene and MTBE. Concentrations are 1 ppm, 10 ppm, 25 ppm and 50 ppm - Your test samples from other experiments - Safety equipment (gloves, glasses, etc.) - 100 ml Beaker with 30-50 ml Acetone - 100 ml Beaker with 50 ml distilled water - GC Syringe ( 10 µl) - SRI GC with FID detector

- Notebook - 100-250 ml Wastewater beaker Make sure that you read and understand the lab procedure. Follow the safety procedure for the work in the lab. Turn GC on. Let it warm up for about 15 minutes Open up the valves on two rightmost reservoirs with compressed air and Hydrogen. Ignite the flame in FID by pushing and holding Flame Ignite bottom on GC. You can see if it works by bringing a stainless steel surface (or a mirror) into the contact with FID outlet. The surface will get foggy. Make sure that the temperature is programmed to increase from 40 to 90 degrees. Rinse the syringe in acetone and then in distilled water several times. Be very careful. The syringe is very fragile and very expensive. Start testing lower concentration of one-component liquids first. Then go to higher concentrations, change to other compound and then to mixtures. Insert the needle into the testing vial through the lid. Don t open the vial. Withdraw some liquid. Discard this liquid in the wastewater beaker. Insert needle into vial again. Slowly withdraw 4 µl of liquid. Make sure that you don t have any air bubbles inside the syringe. The volume must be precise. Make sure that GC thermometer indicates it is at about 40 o C. Start a new file. Click on 0 bottom on the screen of the computer for Autozeroing. Press space on the keyboard to start the run. Inject testing fluid into Injector 2. Wait 6-7 minutes. Review the output curve. Click results at the bottom of the screen. Write down type of the pollutant tested, concentration of the sample and area under the curve calculated by the computer. Save the file in the new folder. Repeat the procedure for other samples. When finished, close valves on air and hydrogen reservoirs. Turn GC off. Using Excel, draw the calibration curve. Calibration curve: Prepare 4 vials with 25 ml DI (de-ionized) water using a graduated cylinder. The stock solution is water saturated with toluene or MTBE. For toluene, the water is at a concentration of 546 mg/l; for MTBE, the water is at a concentration of 54,000 mg/l. Add toluene and MTBE solutions to obtain the following concentrations: 10, 25, 50 and 100 mg/l. (1 ppm = 1 mg/l)

The required volume can be calculated according to the equation: C 1 * V 1 = C 2 * V 2 (Where C = concentration, V =volume.) For example to get a 10 mg/l toluene solution; Solving for V1; 546 mg/l * V1 = 10 mg/l* 25 ml V1= 0.458 ml = 458 µl of toluene saturated water is needed. The final volume should be 25ml, so you have to get this amount of water (i.e. 458 µl) out from the vial with pipette, and then add this amount of toluene water back into the vial. Cap & shake well to make sure it is a homogeneous solution. References www.umd.umich.edu www.shu.ac.uk