Fast Analysis of Natural Gas Using the Agilent 490 Micro GC Natural Gas Analyzer

Similar documents
Chemical Analysis. Natural Gas Analyzer. A Family of optimized GC Solutions. Gas Chromatography. think forward

Hands-on Workshop on Advanced Hydrocarbon Analysis July 23, 2015

Pilot Gasifier Producer Gas Analysis and Measurement Arrangements in. University of Applied Sciences

BTU ANALYSIS USING A GAS CHROMATOGRAPH

ADVANCEMENTS IN MICRO GAS CHROMATOGRAPHY (GC)

1.1 This test method covers the qualitative and quantitative determination of the content of benzene and toluene in hydrocarbon wax.

Chromatograph Maintenance and Troubleshooting Class # 5050

Schedule of Accreditation issued by United Kingdom Accreditation Service 2 Pine Trees, Chertsey Lane, Staines-upon-Thames, TW18 3HR, UK

Fundamentals of Gas Chromatography

PLOT Columns. Top: Bob Langford, GC Column Manufacturing Technician Bottom: Stephanie Sunner, Customer Service Representative

Strategies for Developing Optimal Synchronous SIM-Scan Acquisition Methods AutoSIM/Scan Setup and Rapid SIM. Technical Overview.

AC DHA Solutions. Detailed Hydrocarbon Analysis of Light Petroleum Streams & Light End in Crude Oils. ASTM D 5134, D 6729, D 6730, D 6733 Compliance

CHROMOCTANE O N LINE ANALYZER

Transformer Oil Gas Analysis

Get the Right GC for Your Industry

Micro GC Fusion Pre-Installation Checklist

23 The Thermal Conductivity Detector

Comparative Economic Investigation Options for Liquefied Petroleum Gas Production from Natural Gas Liquids

Fractional Distillation and Gas Chromatography

The Use of Hydrogen Carrier Gas for GC/MS

Fiscal Measurement Natural Gas

Process Gas Chromatograph PGC 9000 VC

THERMAL DESORPTION. Introduction and Principles. Focusing on Volatiles

EPA Method RSK 175 Dissolved Gasses in Water Matrices using the Teledyne Tekmar HT3 Headspace Analyzer. HT3 Application Note.

Addressing the World Helium Shortage For Gas Chromatography

Agilent 7697A Headspace Sampler Data Sheet

Natural Gas Processing Solutions Using the 700XA Gas Chromatograph

Facts about gas physical properties

IBP 2778_10 HIGH EFFICIENCY ON CO2 REMOVAL IN NATURAL GAS WITH UCARSOL SOLVENTS Thiago V. Alonso 1. Abstract. 1. Introduction

HEXANES. Insoluble in water, soluble in ether, alcohol, and acetone. Neutral to methyl orange (ph indicator) Not more than 0.

Hydrogen as an alternative to Helium for gas chromatography.

Process Instrumentation

LEAN LNG PLANTS HEAVY ENDS REMOVAL AND OPTIMUM RECOVERY OF LIGHT HYDROCARBONS FOR REFRIGERANT MAKE-UP

An Advanced Base Deactivated Capillary Column for analysis of Volatile amines Ammonia and Alcohols.

1. Standard conditions are in. Hg (760 mm Hg, psia) and 68 F (20 C).

Gas Chromatography. Let s begin with an example problem: SPME head space analysis of pesticides in tea and follow-up analysis by high speed GC.

National Physical Laboratory Hampton Road Teddington Middlesex United Kingdom TW11 0LW

Lightweight, compact, ergonomic, safe POLARIS FID. Portable TOC Analyser for Stack Emissions

A VISION to On-Line SPE PTV GC MS Determination of Organic Micro Pollutants in Surface Water

Process gas chromatographs. Measurement made easy. Compliance Solutions for 40CFR60 Subpart Ja

Natural gas liquids recovery from gas turbine fuel

Siemens AG, Case Study December 2008

Gas Chromatography Liner Selection Guide

Preparation of Natural Gas Blends Used as Calibration Standards: Sources of Uncertainty and Best Preparation Practices

Background Information

MATHESON REFINERY PROGRAM for Specialty Gas Supply. Experience the MATHESON Commitment to Supply Chain Excellence

Gas Measurement and Analysis

The Characterization of Perfume Fragrances Using GC/MS, Headspace Trap and Olfactory Port

TECHNIQUES FOR NATURAL GAS SAMPLING A DISCUSSION OF FIELD METHODS FOR OBTAINING SPOT SAMPLES. Randall Messman SW Regional Sales Manager

Fast Continuous Online Analysis of VOCs in Ambient Air using Agilent 5975T LTM GC/MSD and Markes TD

TECHNICAL SPECIFICATION. StirLIN-1 Economy Liquid Nitrogen System

Lab 6: Paper Chromatography. Pages Pre-lab page 151 No Post lab Chromatogram must be turned in attached to lab report

Ferda ORHUN Mineral Research and Exploration Institute of Turkey

Regulator Options and Accessories

On-line Ozone Precursor Analyzers From the Leader in Thermal Desorption Technology

ANNEX I GAS MEASUREMENTS AND TESTING AT THE DELIVERY POINT. Table of Contents

Separation of Amino Acids by Paper Chromatography

N O T E S. Environmental Forensics. Identification of Natural Gas Sources using Geochemical Forensic Tools. Dispute Scenarios

SCION SQ Series GC-MS

G8 GALILEO. Innovation with Integrity. High-End Melt-extraction Analyzer. Inert Gas Method

Shimadzu Simulated Distillation Gas Chromatograph System C184-E030

ASimple Guide to Oil Refining

Impact Analysis for Software changes in OpenLAB CDS A.01.04

Micro Cryo-Trap Model 971

How To Run A Gas Plant

Dew Point Tester. Instruction Manual. CVS Regular Chiller Model A-2. CVS Regular Chiller Model A-2

Gas Separations: A Comparison of GasPro and Aluminum Oxide PLOT Columns for the Separation of Highly Volatile Compounds

Micro Motion 3098 Gas Specific Gravity Meter

SYNSPEC GC s ALPHA 115 and ALPHA 116

(a) Method 1 Sample and Velocity Traverses for Stationary Sources.

ECONOMICAL OPTIONS FOR RECOVERING NGL / LPG AT LNG RECEIVING TERMINALS

On-line Dissolved Gas Analysis

Open Cycle Refrigeration System

HIGH PRESSURE TECHNOLOGY HYDRAULICS PNEUMATICS TESTING EQUIPMENT

Pesticide Analysis by Mass Spectrometry

Natural Gas Information Contents

LESSONS LEARNED FROM THE API 14.1 GAS SAMPLING RESEARCH PROJECT. Darin L. George and Eric Kelner

CHEMISTRY GAS LAW S WORKSHEET

COMBUSTION. In order to operate a heat engine we need a hot source together with a cold sink

Identification of Arson Accelerants by Gas Chromatography

Quantitative Evaluation of Perfume Fleuressence Samples using the znose

Molar Mass of Butane

SENSITIVITY ANALYSIS OF A NATURAL GAS TRIETHYLENE GLYCOL DEHYDRATION PLANT IN PERSIAN GULF REGION

Thermal Conductivity Detector

Gas standards and measurements

Crude Assay Report. Crude Oil sample marked. "AL SHAHEEN CRUDE OIL" dated 10th April On Behalf Of. Maersk Oil Qatar AS

Lockout Specification BW Clip Maintenance-free single gas detector

MULTIPLE CHOICE. Choose the one alternative that best completes the statement or answers the question.

Bringing Innovations to Standards AC Analytical Controls by PAC All In One Biodiesel Analyzer

Certain specific properties of compressed gases make them highly useful in various research activities.

Lecture 35: Atmosphere in Furnaces

Series 2. Air Server. Options for UNITY 2. A robust, cryogen-free solution for round-the-clock on-line air monitoring

Chemistry 321, Experiment 8: Quantitation of caffeine from a beverage using gas chromatography

IB Chemistry. DP Chemistry Review

Multi-elemental determination of gasoline using Agilent 5100 ICP-OES with oxygen injection and a temperature controlled spray chamber

Validation of process gas systems

T U R B I N E G A S M E T E R

Transcription:

Fast Analysis of Natural Gas Using the Agilent 90 Micro GC Natural Gas Analyzer Application Note Author Micro Gas Chromatography, Hydrocarbon Processing, Natural Gas Analysis Remko van Loon, Agilent Technologies, Middelburg, the Netherlands Abstract During production and distribution of natural gas it is of high importance to determine its composition and calorific value because natural gas is bought and sold on its energy content. This application note shows the use of the Agilent 90 Micro GC Natural Gas Analyzer for the analysis of natural gas and the calculation of its heating value. With the 90 Micro GC, Agilent provides ideal solutions for laboratory, on-line and field use.

Introduction Natural gas mainly consists of methane and variable levels of other hydrocarbons and permanent gases such as oxygen, nitrogen, and carbon dioxide. Different sources of natural gas usually have similar composition but vary in concentration. Natural gas is traded on its energy content and therefore the analysis of the chemical composition and calorific value is of high importance for all stakeholders. That is where the 90 Micro GC based Natural Gas Analyzer can play a significant role. The 90 Micro GC Natural Gas Analyzers are shipped as a total solution; the analyzers are factory tuned, for optimal separation, and come with final test data, a complete method, a user manual, and a check-out sample. Easy-to-use software is available for the calculation of all required physical properties, such as heating value and Wobbe index, conform official methods from the American Society of Testing and Materials (ASTM), Gas Processors Association (GPA) and International Standards Organization (ISO). Natural Gas Analyzer setup Based on the 90 Micro GC, four Natural Gas Analyzers are available, depending on the composition of the natural gas and the compounds of interest. The configurations and analysis characteristics for all analyzers are shown in Table. Additional information for the configurations can be found in Natural Gas Analyzer Data Sheet []. The Natural Gas Analyzers are equipped with heated injectors and sample lines, both set to 0 C in the analyzer method, to eliminate any cold spots and prevent possible condensation of moisture, and to ensure the integrity of the sample is maintained throughout the sample flow path. Table shows multiple column channels are equipped with a back flush to vent option. To protect the CP-Molsieve A stationary phase and maintain the separation efficiency of the molecular sieve column, it is necessary to back flush carbon dioxide, moisture, and higher hydrocarbons. Moisture and carbon dioxide tend to adsorb quickly to the molecular sieve stationary phase change its chromatographic properties. This can result in retention shifts and loss of separation. Higher hydrocarbons will eventually elute, but will cause higher detector noise levels and would lead to reduced sensitivity; the back flush to vent functionality on the Molsieve A column channel prevents this from happening. On the PoraPLOT U and HayeSep A channels, the higher hydrocarbons, C and higher, are back flushed to vent. This prevents these late eluting components from interfering in the next analysis. Table. Agilent 90 Micro GC Natural Gas Analyzers Overview. Analyzer characteristics Natural Gas Analyzer A Natural Gas Analyzer A Extended Natural Gas Analyzer B Natural Gas Analyzer B Extended Micro GC cabinet Dual with channels Quad with channels Dual with channels Quad with channels Column channels installed HayeSep A 0 cm, without backflush HayeSep A 0 cm, with backflush PoraPLOT U 0 m, with backflush CP-MolSieve A 0 m, with backflush and retention time stability option CP-Sil CB 6 m, without backflush CP-Sil CB m, with backflush CP-Sil CB 6 m, without backflush PoraPLOT U 0 m, with backflush CP-Sil CB 8 m, without backflush CP-Sil CB 6 m, without backflush Analysis Hydrocarbons C-C9 Carbon dioxide, Air Hydrocarbons C-C Carbon dioxide, Air Hydrocarbons C-C9 Carbon dioxide, Air, Hydrogen sulfide Hydrocarbons C-C9 Carbon dioxide, Air, Hydrogen sulfide Permanent gases (N, O, He and H )

The CP-Molsieve A is equipped with the retention time stability (RTS) option. This RTS option consists of additional in-line filters between the electronic gas control and the column module to ensure moisture and carbon dioxide free carrier gas. The use of the RTS option enables a more efficient back flush of carbon dioxide. This enhances column lifetime and, most importantly, leads to more stable retention times. The natural gas sample can be introduced to the 90 Micro GC either pressurized (maximum limit bar), through a Tedlar sampling bag using the internal sampling pump, or by using continuous flow sampling mode. When you need to analyze multiple streams on a single analyzer or you want to connect multiple calibration samples for automated calibration, the use of a stream selector valve is recommended. To expand the range of samples to Liquid Petroleum Gas (LPG) and Liquefied Natural Gas (LNG), the Micro-Gasifier provides controlled evaporation before the sample is introduced into the gas chromatographic injector for analysis. In addition, high-pressure gas samples can be reduced without creating cold spots, which prevents discrimination in the sample. Fast Natural gas analysis using the Natural Gas Analyzer A The first channel in the Natural Gas Analyzer A is equipped with a HaySep A column for separating methane from the composite air peak (nitrogen and oxygen). Carbon dioxide, ethane, and propane are analyzed on this column channel as well. Figure shows an example chromatogram for these compounds. For the analysis of the hydrocarbons from propane to n-nonane, a second column channel, equipped with a 6-meter CP-Sil CB column, is used. Figure a shows a chromatogram on the 6-meter CP-Sil CB for the separation until n-octane; n-hexane elutes in less than 60 seconds and n-octane in just over minutes. Propane is analyzed on both HayeSep A and CP-Sil CB column enabling the use of propane as a bridge component. The extended part of the chromatogram obtained with a 6-meter CP-Sil CB column, displayed in Figure b, shows the analysis of hydrocarbons until n-nonane. Column temperature 60 C helium, 60 kpa Injection time 0 ms. composite air peak. methane. carbon dioxide. ethane. propane 0 Zoom 00 Zoom 0 0 60 90 0 Figure. Chromatogram for nitrogen, carbon dioxide, and C C hydrocarbons on a HayeSep A column.

6 7 Column temperature 70 C helium, 0 kpa Injection time 0 ms 8 9. propane. i-butane. n-butane. neo-pentane. i-pentane 6. n-pentane 7. n-hexane 8. n-heptane 9. n-octane 0 Zoom 0 0 60 90 0 0 80 0 Figure a. Chromatogram for C C8 hydrocarbon using a 6-meter CP-Sil CB column. 9. n-octane 0. n-nonane 9 0 00 Zoom 60 0 80 0 00 Figure b. Chromatogram for C8 C9 hydrocarbons using a 6-meter CP-Sil CB column. Analysis up to n-dodecane with the Natural Gas Analyzer A Extended The extended version of the Natural Gas Analyzer A is used for the analysis of natural gas until n-dodecane. This extended analyzer is equipped with three column channels. First, a HayeSep A column channel is used for separation of composite air peak from methane, carbon dioxide ethane, and propane. This channel is equipped with back flush functionality ensuring that butanes and later eluting hydrocarbons are back flushed to vent. Figure shows an example for the HayeSep A channel, propane is eluting in less than minutes.

Column temperature 90 C helium, 0 kpa Injection time 0 ms Backflush time 0 s. composite air peak. methane. carbon dioxide. ethane. propane 0 Zoom 00 Zoom 0 0 60 90 0 Figure. Chromatogram for HayeSep A column with backflush. The second channel, equipped with a -meter CP-Sil CB column with back flush functionality, is used to analyze C to C hydrocarbons; the chromatogram is shown in Figure. N-hexane and higher hydrocarbons are back flushed to vent. Column temperature 60 C helium, 0 kpa Injection time 0 ms Backflush time s. propane. i-butane. n-butane. neo-pentane. i-pentane 6. n-pentane 6 0 Zoom 0 0 0 0 Figure. Chromatogram for C to C hydrocarbons on a -meter CP-Sil CB.

A third column channel, equipped with a 8-meter CP-Sil CB column, is used to analyze the higher hydrocarbons from n-hexane to dodecane; n-dodecane elutes in approximately 0 seconds. A natural gas sample sample until n-decane, demonstrated in Figure a, is analyzed in less than minutes. Figure b displays a hydrocarbon gas mixture from n-hexane until n-docecane, typical analysis time is 0 seconds. Analysis of natural gas including hydrogen sulfide using the Natural Gas Analyzer B When your natural gas analysis needs to include hydrogen sulfide, the 90 Micro GC Natural Gas Analyzer B is the analyzer of choice. This analyzer uses a PoraPLOT U column channel for the separation of methane from the composite air peak (nitrogen and oxygen). This column is also used for the analysis of carbon dioxide, ethane, and propane. The chromatogram in Figure 6 shows an example of natural gas on the PoraPLOT U column; total analysis is done in approximately 60 seconds. For the analysis of hydrogen sulfide, the stainless steel tubing in the PoraPLOT U channel and the sample inlet of the Micro GC have an UltiMetal deactivation layer, which results in an inert sample flow path and excellent peak shape ensuring correct analysis of hydrogen sulfide even at ppm level. Hydrocarbon analysis from propane until n-nonane for the Natural Gas Analyzer B is done with a second channel equipped with a 6-meter CP-Sil CB. This column is identical to the one used for the Natural Gas Analyzer A. The chromatograms for this channel are displayed in Figure a and b. Column temperature 0 C helium, 00 kpa Injection time 0 ms Zoom. n-hexane. n-heptane. n-octane. n-nonane. n-decane 0 0 60 90 0 Figure a. Analysis of natural gas on an 8-meter CP-Sil CB.. n-hexane. n-heptane. n-octane. n-nonane. n-decane 6. n-undecane 7. n-dodecane 6 7 0 60 0 80 0 Figure b. Analysis C7 C hydrocarbon mix on an 8-meter CP-Sil CB. 6

Column temperature 60 C helium, 7 kpa Injection time 0 ms Backflush time 7 s 6. composite air peak. methane. carbon dioxide. ethane. hydrogen sulfide 6. propane 0 Zoom 00 Zoom 6 0 60 7 Figure 6. Chromatogram for natural gas on the PoraPLOT U column channel. Permanent gas analysis using Natural Gas Analyzer B Extended The Extended version of the 90 Micro GC Natural Gas Analyzer B is equipped with an additional CP-MolSieve A column channel for the analysis of permanent gases in your natural gas sample. Helium carrier gas on this channel enables the separation and quantification of oxygen and nitrogen, an example is shown in Figure 7 (top part). When you need to analyze helium, neon, or hydrogen as well, the use of argon instead of helium as carrier gas is required. The bottom part of Figure 7 shows a chromatogram for the molecular sieve column running with argon as carrier gas. To have the flexibility to change the carrier gas for only the molecular sieve column to argon, this channel is connected to a separate carrier gas inlet at the rear of the micro GC. Sample Helium carrier gas Column temperature 80 C helium, 00 kpa Injection time 0 ms Backflush time s Zoom 6. helium. neon. hydrogen. oxygen. nitrogen 6. methane Sample Argon carrier gas Column temperature 80 C argon, 00 kpa Injection time 0 ms Backflush time s 0 Zoom 6 0 0 60 7 Figure 7. Chromatograms for the analysis of permanent gases on the CP-MolSieve A column channel. 7

Reporting tools for the physical properties of natural gas The results for all individual components are sent from the chromatography data software of choice (EZChrom Elite, OpenLAB EZChrom, or OpenLAB Chemstation) to optional EZReporter software to calculate a wide range of physical properties like, calorific value, relative density, compressibility, and Wobbe index, see Figure 8 (left). These key parameters are important to determine the commercial value of the natural gas. EZReporter supports reports in accordance with official methods ASTM D88, ISO 6976, and GPA 7. The reports can be printed locally or exported to a file for further use in a laboratory information management system (LIMS). The software includes functionality to select raw analysis amounts and calculated key parameters for monitoring and historical trend analysis. Upper and lower warning limits can be given to these monitor parameters to better visualize the results from your natural gas stream. Some examples are given in Figure 8 (middle and right). Conclusion Micro GC Natural Gas Analyzer is a genuinely better solution for analyzing your natural gas stream. Whether in the lab, on-line/at-line, or in the field, the Measure Anywhere Micro GC provides natural gas analysis in a matter of seconds. The Natural Gas Analyzer A analyzer combined with a HayeSep A and 6-meter CP-Sil CB column channel is used for the analysis of natural gas. This analyzer will separate methane from air and can analyze up to n-nonane. Carbon dioxide is also analyzed. Total analysis time depends on the hydrocarbons in the sample; up to n-heptane is done in approximately 90 seconds, n-nonane elutes just under 00 seconds. When you want to analyze until n-dodecane in natural gas, the Natural Gas Analyzer A Extended is required. The 6-meter CP-Sil CB column channel is replaced by two different CP- Sil CB channels. A short CP-Sil CB (-meter) will analyze from propane to the pentanes; hexane and higher will be back flushed to vent. The second CP-Sil CB channel, with an 8-meter column, is used for analysis of hexane up to n-dodecane. Figure 8. EZReporter, sample results with calculated physical properties (left), parameter monitoring (middle), and trend analysis (right). 8

The Natural Gas Analyzer B, equipped with a PoraPLOT U and a 6-meter CP-Sil CB CB column channel, provides fast analysis of natural gas, from the separation of air and methane, carbon dioxide, and hydrocarbons up to n-nonane. This analyzer setup is designed for the analysis of hydrogen sulfide. The stainless steel sample inlet of the systsm is deactivated using an UltiMetal treatment resulting in excellent peak shape for hydrogen sulfide. If more detailed analysis of the permanent gases in the natural gas sample is required, the extended version of the Natural Gas Analyzer B is the system of choice. This analyzer is equipped with an additional CP-MolSieve A column enabling the separation of oxygen and nitrogen, using helium as carrier gas. When this analyzer uses argon as carrier gas, helium and hydrogen can be detected as well. The 90 Micro GC Natural Gas Analyzers are factory tuned, including the appropriate settings for the back flush times. The Agilent Natural Gas Analyzers are shipped with final test data, optimized analytical method, Natural Gas Analyzer User Manual, and a check out sample kit to have all information available upon installation. The analyzer hardware together with your chromatography data system (CDS) of choice provides an easy-to-use and powerful system. The EZReporter software is linked to Agilent CDS, resulting in automatic calorific value/btu calculations and reports according to American Society of Testing and Materials (ASTM D88), Gas Processors Association (GPA 7), and International Standards Organization (ISO 6976). For more information about the 90 micro GC Natural Gas Analyzer or other Micro GC solutions, visit our website at www.agilent.com/chem/microgc. References. 99-00EN; Agilent 90 Micro GC Natural Gas Analyzers; Data Sheet; April 0. For More Information These data represent typical results. For more information on our products and services, visit our Web site at www.agilent.com/chem. 9

www.agilent.com/chem/microgc Agilent shall not be liable for errors contained herein or for incidental or consequential damages in connection with the furnishing, performance, or use of this material. Information, descriptions, and specifications in this publication are subject to change without notice. Agilent Technologies, Inc., 0 Printed in the USA April 6 0 99-07EN