ENrich Q ENrich S High-Resolution Ion Exchange Columns

Similar documents
ENrich SEC 70 ENrich SEC 650 High-Resolution Size Exclusion Columns Instruction Manual

UNO Q&S Continuous Bed Ion Exchange Column. Instruction Manual

RESOURCE Q, 1 ml and 6 ml RESOURCE S, 1 ml and 6 ml

Transforming a ChromLab Software 2-D Purification Template into an Automated Multidimensional (Multi-D) Purification Workflow An Instructional Guide

INSTRUCTIONS Edition AC

Aurum Ion Exchange Mini Kits and Columns. Instruction Manual

Affi-Prep Protein A Matrix Instruction Manual

ProteinChip Energy Absorbing Molecules (EAM)

Image Lab Software for the GS-900 Densitometer

Guide to Reverse Phase SpinColumns Chromatography for Sample Prep

[ Care and Use Manual ]

Bio-Gel P Polyacrylamide Gel Instruction Manual

Image Lab Software How to Obtain Stain-Free Gel and Blot Images. Instructions

Computer Specifications

Application Note. USD 2995 (a) High Throughput Regeneration Study on MEP HyperCel Mixed-Mode Sorbent on ScreenExpert RoboColumns u

Cell Sorting. S3e Cell Sorter Quick Guide

HiTrap Heparin HP, 1 ml and 5 ml

How To Install A Gel Doc Xr+ Camera On A Computer Or Mouse Or Tv (For A Powerbook) With A Zoom Lens (For An Ipa) With An Iphone Or Ipa (For Mac) With The Zoom Lens

Foreign Taxes Paid and Foreign Source Income INTECH Global Income Managed Volatility Fund

[ Care and Use Manual ]

Efficient Multi-Well Protein Purification Strategies

TECHNICAL BULLETIN. HIS-Select Nickel Affinity Gel. Catalog Number P6611 Storage Temperature 2 8 C

ProteoMiner Protein Enrichment Kits

[ Care and Use Manual ]

Data File. Sephadex G-25 media and pre-packed columns. Introduction. Sephadex G-25 Bead structure. Desalting/buffer exchange and gel filtration

Waters Corporation. Waters 2690/5 USER & TROUBLESHOOTING GUIDE

Determination of Calcium and Magnesium in Brine

Diafiltration: A Fast, Efficient Method for Desalting, or Buffer Exchange of Biological Samples

Method Development for Size-Exclusion Chromatography of Monoclonal Antibodies and Higher Order Aggregates

Oasis HLB Cartridges and 96-Well Plates

amplification tech Optical Design of CFX96 Real-Time PCR Detection System Eliminates the Requirement of a Passive Reference Dye

[ Care and Use Manual ]

Blue Sepharose 6 Fast Flow

Expression Proteomics // Tools for Protein Separation and Analysis. 2-D Electrophoresis. Tools for Rapid, High-Resolution Protein Separations

Installation Qualification/Operational Qualification Protocols and Instructions. Experion Automated Electrophoresis Station and Software

Thermo Scientific SOLAµ SPE Plates Technical Guide. Consistent excellence. for bioanalysis

Automated Method Development Utilizing Software-Based Optimization and Direct Instrument Control

[ Care and Use Manual ]

What is Eluent Generation?

Configuring DHCP for ShoreTel IP Phones

World Consumer Income and Expenditure Patterns

for IonSwift TM MONOLITH ANION CONCENTRATOR (MAC)

The Determination of Sugars in Molasses by High-Performance Anion Exchange with Pulsed Amperometric Detection

High-Throughput 3-D Chromatography Through Ion Exchange SPE

THE WORLD S LEADING CAR DESIGN MAGAZINE

His GraviTrap. GE Healthcare. Operation

HiTrap Chelating HP, 1 ml and 5 ml

Reporting practices for domestic and total debt securities

HiPer Ion Exchange Chromatography Teaching Kit

Size Exclusion Chromatography

Appendix 1: Full Country Rankings

OCTOBER Russell-Parametric Cross-Sectional Volatility (CrossVol ) Indexes Construction and Methodology

Dissolved and precipitated oxalate

POROS CaptureSelect affinity columns for highspeed quantification of IgG Fc fusion proteins

UltraClean Soil DNA Isolation Kit

The Theory of HPLC. Gradient HPLC

41 T Korea, Rep T Netherlands T Japan E Bulgaria T Argentina T Czech Republic T Greece 50.

To install the I/O Module Board: 1. Remove AC power from decoder.

Corporate Office Von Karman Ave Suite 150 Irvine, California Toll Free: Fax:

Report on Government Information Requests

relating to household s disposable income. A Gini Coefficient of zero indicates

BT Premium Event Call and Web Rate Card

What Proportion of National Wealth Is Spent on Education?

How To Get A New Phone System For Your Business

Motion Graphic Design Census. 10 hrs. motiongraphicdesigncensus.org. 9 hrs.

GE Healthcare. Hydrophobic Interaction and Reversed Phase Chromatography. Principles and Methods

Determination of Pesticide Residues in Drinking Water Using Automated Solid-Phase Extraction and Gas Chromatography with Nitrogen Phosphorus Detection

Accuracy counts! SENSORS WITH ANALOG OUTPUT

MabSelect SuRe. GE Healthcare Life Sciences

EZ-Run Protein Gel Solution. EZ-Run Protein Standards. EZ-Run Gel Staining Solution. Traditional SDS-Page Reagents. Protein Electrophoresis

Pall Aria Integrated MF/RO Systems for Cost-Effective Treatment of Sea and Brackish Waters

6 Characterization of Casein and Bovine Serum Albumin

POROS CaptureSelect affinity columns for rapid, small-scale purification and sample preparation of recombinant proteins

Bio-Rad Laboratories. QC data management solutions. Introduce Your Laboratory to a Whole New World of Unity Data Management Solutions

Solid Phase Extraction Products PAGE: 1. Introduction of Solid Phase Extraction (SPE) Why Choose Nano-Micro Tech SPE

Dionex IonPac AS19-4µm

Supported Payment Methods

Report on Government Information Requests

DNA SPOOLING 1 ISOLATION OF DNA FROM ONION

Gel Filtration Standard

Supported Payment Methods

Accelerated Electrophoresis with Faster Run Times and Increased Sample Throughput

IMD World Talent Report. By the IMD World Competitiveness Center

LC-MS/MS Method for the Determination of Docetaxel in Human Serum for Clinical Research

[ Care and Use Manual ] styragel column. Contents. I. Introduction. II. Installing the Column Bank. III. Preparing Solvent and Samples

Purification of reaction mixtures using flash chromatography.

Protein Separation Technology

Global Economic Briefing: Global Inflation

Brochure More information from

NI USB-6008/6009 OEM USER GUIDE

Table 1: TSQM Version 1.4 Available Translations

Wizard DNA Clean-Up System INSTRUCTIONS FOR USE OF PRODUCT A7280.

Empore. 96-Well Solid Phase Extraction Plates. Technical Information. Instructions for Use. Product Characteristics. Product Description

Global Effective Tax Rates

The Role of Banks in Global Mergers and Acquisitions by James R. Barth, Triphon Phumiwasana, and Keven Yost *

Thermo Scientific Dionex AS-AP, AS-DV, and AS-HV Autosamplers. Automation, sample preparation. Autosamplers for Ion Chromatography

Quantum View Manage Administration Guide

Recent advances in the purification of IgM monoclonal antibodies

Quick Start Bradford Protein Assay

Building a Global Internet Company: Driving Traffic to Your Site. Benjamin Edelman Harvard Business School

Transcription:

ENrich Q ENrich S High-Resolution Ion Exchange Columns Instruction Manual Catalog numbers 780-0001 780-0003 780-0021 780-0023 Please read these instructions before you use ENrich Q and S high-resolution ion exchange media. If you have any questions or comments regarding these instructions, please contact your Bio-Rad Laboratories representative.

Table of Contents Section 1: Characteristics of the ENrich Ion Exchange (IEX) Columns... 1 1.1 Introduction...1 1.2 The ENrich Separation Media... 1 1.3 Connection to the NGC and Other Chromatography Systems... 2 Section 2 : Use of the ENrich IEX Columns... 3 2.1 Preparation for Initial Use... 3 2.2 Buffer Selection... 3 2.3 Sample Preparation... 5 2.4 Elution Conditions... 5 Section 3 : Care of the ENrich IEX Columns... 7 3.1 Column Cleaning... 7 3.2 Bed Height Adjustment... 8 3.3 Frit Replacement... 9 3.4 Storage Conditions...12 Section 4 : Product Information...13

Section 1: Characteristics of the ENrich Ion Exchange Columns 1.1 Introduction ENrich prepacked ion exchange columns meet the needs of the biochromatographer for rapid and reproducible high-resolution separations of proteins and other biomolecules, including peptides and polynucleotides. Two column sizes (bed volumes of 1 and 8 ml) provide predictable scale-up of high-resolution separations without sacrificing capacity. 1.2 The ENrich Separation Media Each column contains a spherical, rigid, and highly porous polymeric support derivatized with the strongly basic N + (CH 3 ) 3 quaternary ammonium group or the strongly acidic SO 3 group. The 10 μm particle-size and narrow particle size distribution provide excellent resolution of biomolecules at high flow rates and with low backpressures. The hydrophilic ENrich media demonstrate extremely low non-specific binding of biomolecules accompanied by high recovery of biological activity. Stability The columns are stable over a 2 12 ph range, allowing easy cleaning and regeneration. The ENrich support is compatible with aqueous solutions of 6 M guanidine-hcl and 8 M urea. Nonionic detergents and organic solvents such as methanol, ethanol, and isopropanol may also be used. Ionic detergents can be used only if the detergents have the same charge as the support (that is, sodium dodecyl sulfate (SDS), a negatively charged detergent, should not be used with a positively charged anionic exchanger like ENrich Q). ENrich Instruction Manual 1

Table 1. Column characteristics. Column dimensions (diameter x height, mm) 5 x 50 10 x 100 5 x 50 10 x 100 Column volume, ml 1 8 1 8 Maximum protein capacity, mg Bovine serum albumin 130 1000 Human IgG 120 940 Average particle size, μm 10 10 10 10 Maximum operating pressure, psi 500 500 500 500 Recommended flow rate, ml/min* 1 4 1 4 * At room temperature. Viscosity may increase at lower temperatures, which will reduce the recommended flow rates. Q S 1.3 Connection to the NGC and Other Chromatography Systems The ENrich columns are fitted with 10-32 type female fittings on either end. Standard 10-32 fittings can be used to plumb the column to Bio-Rad s NGC chromatography system or other vendors systems. Adaptors are available for connection to systems that use 1/4-28 fittings (catalog # 750-0564). 2 ENrich Instruction Manual

Section 2: Use of the ENrich Columns 2.1 Preparation for Initial Use The columns are supplied in a storage solvent of 20% ethanol in water. Prior to initial use and after extended storage periods, each column should be conditioned as described in steps 1 4. Always use HPLC-grade reagents and be sure to filter and degas solvents. During these four steps do not exceed more than 50% of the recommended flow rates (see Table 1). 1. Wash with 5 column volumes of water. 2. Wash with 5 column volumes of low ionic strength equilibration buffer (such as 20 mm buffer salt). 3. Wash with 5 column volumes of high ionic strength limit buffer (such as 1.0 M NaCl). 4. Wash with 5 column volumes of low ionic strength equilibration buffer. The column may now be further equilibrated in the start buffer at the desired flow rate. 2.2 Buffer Selection Table 2 lists commonly used buffers for ion exchange chromatography. The buffers are specific to the type of ion exchange. Therefore, it is important not to use anionic buffers with the ENrich Q, which would interact with the anionic exchange group on the support. The choice of whether to use an anion or cation exchanger is determined mainly by the isoelectric point (pi) and the relationship between ph and the activity/stability of the protein(s) of interest. When the type of ion exchanger is determined, the ph-activity relationship also determines the choice of buffer. As a general rule, the chosen buffer should be used within ±1.0 ph unit of its pka value. This permits use of the lowest possible buffer concentration while maintaining maximum buffering capacity. We recommend a minimum buffer concentration of 20 mm. ENrich Instruction Manual 3

Table 2. Buffers for ion exchange chromatography. Anion Exchange ENrich Q ph Buffer MW pka @ Counter-Ion pka/ C Range 25 C 5.0 6.0 Piperazine 86.1 5.7 Cl-/HCOO 0.015 5.0 6.0 L-histidine 155.2 6.15 CI 5.8 7.2 Bis-Tris 209.2 6.5 CI 0.017 6.4 7.3 Bis-Tris propane 282.3 6.8,9 CI 7.3 8.3 Triethanolamine 149.2 7.8 CI-/CH3COO 0.02 7.6 8.6 Tris 121.1 8.1 CI 0.031 8.4 8.8 Diethanolamine 105.1 8.9 CI 0.025 9.0 9.9 Ethanolamine 61.1 9.5 CI 0.029 9.8 10.3 1,3-diaminopropane 74.1 10.5 CI 0.023 Cation Exchange ENrich S ph Range Buffer MW pka @ 25 C Counter-Ion pka/ C 3.6 4.3 Lactic acid 90.1 3.8 Na + 4.2 5.2 Citric acid 192.1 3.1 Na + 5.5 6.7 MES 195.2 6.1 Na + 0.011 6.1 7.5 PIPES 302.4 6.8 Na + 0.009 6.5 7.9 MOPS 209.3 7.2 Na + 0.006 6.7 7.6 Phosphate 120/142 7.2 Na + 0.003 6.8 8.2 TES 229.2 7.4 Na + 0.002 6.8 8.2 HEPS 238.3 7.5 Na + 0.014 7.4 8.8 Tricine 179.2 8.1 Na + 0.021 Always use buffer components of the highest purity available, as UV-absorbing impurities may cause baseline disturbances and interfere with the detection of protein peaks. Always filter and degas buffers. 4 ENrich Instruction Manual

2.3 Sample Preparation Proper adjustment of the sample ph and ionic strength is critical for consistent and reproducible chromatography. For best results, the sample should be exchanged into the start buffer or diluted to the start buffer s concentration. Buffer exchange can be performed using Bio-Spin 6 and Micro Bio-Spin 6 columns, Econo-Pac 10DG desalting columns, Bio-Gel P-6DG gel, or the Bio-Scale Mini desalting cartridges with Bio-Gel P-6. The choice of product depends on sample volume. Centrifuge or filter the sample (0.2 0.45 μm filter) to remove particulates. Application of turbid or lipid-containing samples may reduce the column lifetime. Sample Load The maximum sample load for each column is shown in Table 1. This amount may vary somewhat depending on the actual sample composition. We do not recommend overloading the column as both resolution and column lifetime will decrease. For larger loads, perform several chromatographic runs with a reduced load. Ideally, samples should be bound in a concentrated zone at the top of the column. Higher sample loads produce a broad application zone in which components with less charge are displaced by more highly charged components. This may result in a shift of certain peaks to an earlier elution position in the gradient. The recommended sample load is approximately 20% of the maximum. 2.4 Elution Conditions Separations by ion exchange are typically accomplished by increasing the salt concentration of the eluent either as a step or as a continuous gradient. Sodium chloride (NaCl) and potassium chloride (KCl) are the most common elution salts. For many separations, varying the ph of the elution buffer in addition to its salt concentration may be advantageous. ENrich Instruction Manual 5

Gradient Volumes and Salt Concentrations As a starting point for developing a separation, we recommend using the ENrich column with a simple linear gradient over 15 column volumes (15 ml). Suggested protocol for ENrich 5 x 50 ion exchange columns: Use a flow rate of 1.0 ml/min. Equilibrate the column with 5 column volumes (5 ml) of equilibration buffer. Then apply the sample. Following sample application, wash unbound proteins from the column with 3 column volumes (3 ml) of equilibration buffer A. For elution, use a gradient volume of 15 column volumes (15 ml) to an NaCl concentration of 0.5 M. Follow this segment of the gradient by stepping the salt concentration to 1.0 M (100% B) and then hold at 1.0 M for 3 ml before re-equilibrating the column with 3 ml of start buffer A. This gradient is shown schematically in Figure 1. When an initial separation has been performed and the elution position of the protein of interest determined, the gradient composition and volume are adjusted to achieve maximum resolution. Normally, a gradient volume of 10 to 20 ml per ml of column bed volume is sufficient. The slope of the gradient will affect resolution. A steep gradient will result in relatively small peak volumes but short peak-to-peak distances. A shallower gradient normally gives greater resolution but peak volumes are larger. Fig. 1. Schematic gradient for separation on an ENrich IEX column. Volume, column volume. 6 ENrich Instruction Manual

Section 3: Care of the ENrich Ion Exchange Column 3.1 Column Cleaning Careful preparation (especially filtration) of the sample and the buffers will maintain the column performance and lifetime. Normally, washing with 1.0 M NaCl or KCl will remove most bound components. However, if there is a significant decrease in column performance (increasing backpressures or a significant drop in resolution), then a more extensive cleaning protocol such as that described in steps 1 6 should be used. Always reverse the flow during this procedure so tightly bound substances at the top of the column are removed quickly. During this operation do not exceed more than 50% of the recommended flowrates (see Table 1). 1. Wash with 2 column volumes of 2.0 M NaCl or KCl followed by 3 column volumes of water. 2. Wash with two 100 μl injections of 1.0 M NaOH followed by 3 column volumes of water. 3. Wash with two 100 μl injections of 50% acetic acid followed by 3 column volumes of water. 4. If lipid contamination is suspected, wash with 1 column volume of 20% ethanol followed by 3 column volumes of water. 5. Wash with 2 column volumes of 2.0 M NaCl or KCl, or the salt containing the desired counter-ion. 6. Wash with 3 column volumes of equilibration buffer. ENrich Instruction Manual 7

3.2 Bed Height Adjustment Under certain conditions of buffer composition, high flow rates or long-term use, the resin bed may compress, creating a void between the frit and the top of the bed. Normally, the void can be eliminated by turning the adjusting nut clockwise until the frit just touches the top of the bed (Figure 2). If the bed compresses at high flow rates, stop the pump and loosen the top fitting, then use the adjusting nut to remove the void, retighten the top fitting and resume pumping buffer. If the bed has compressed from long use, the top frit should be replaced as a precaution. Fig. 2. Turning the adjusting nut raises or lowers the adaptor. 8 ENrich Instruction Manual

3.3 Frit Replacement The top frit may need to be replaced after extensive column use or if increasing backpressures are noticed. Always try cleaning the column in the reverse direction (as described in Section 3.1) before replacing the frit. A frit kit is available which contains a frit removal tool, 2 O-rings and 2 frits. Figure 3 shows a column diagram to assist in the replacement of the top frit. Fig. 3. Column diagram. 1. Start buffer through the column at a slow flow rate (0.5 ml/min or less). 2. Remove the lower end tubing and fitting from the column. Firmly hold the bottom of the column over a sink or container. 3. Loosen the lock nut turning it clockwise (Figure 4). Fig. 4. Loosen the lock nut by turning it clockwise. ENrich Instruction Manual 9

4. Raise the adaptor a few millimeters by slowly by turning the adjusting nut counterclockwise (Figure 5). Fig. 5. Raise the adaptor by turning the adjusting nut counterclockwise. 5. Unscrew the upper retainer. Let it rest on the bottom retainer (Figure 6). Fig. 6. Remove the retainer. 6. Slowly pull out the upper adaptor from the glass column. Allow the buffer flow to continue in order to maintain the integrity of the top of the bed. 10 ENrich Instruction Manual

7. Stop the pump. Plug the bottom of the column. Set the column upright in a beaker. Remove the tubing and fitting from the top of the adaptor. 8. Remove the frit from the adaptor by hooking one end of the frit removal tool/tweezer into the frit in a sideways motion with slight downward pressure. 9. Push the new frit into the end of the adaptor. 10. Replace the O-rings if they appear worn or torn. If the O-rings are replaced, wet them with buffer before the next step. 11. Add a few drops of buffer to the top of the resin bed. Insert the adaptor and push it down to the bed. Some buffer should flow out of the top of the column. 12. Screw on the upper retainer (Figure 7). Fig. 7. Reattach and hand-tighten the retainer. ENrich Instruction Manual 11

13. Lower the adaptor to the top of the bed by turning the adjusting nut clockwise (Figure 8). Fig. 8. Lower the adaptor by turning the adjusting nut clockwise. 14. Tighten the lock nut by turning it counterclockwise (Figure 9). Fig. 9. Hand-tighten the lock nut by turning it counterclockwise. 3.4 Storage Conditions Prior to long-term storage, the column should be cleaned as previously described and then washed with 3 column volumes of 20% ethanol at a flow rate that is 50% of the recommended flow rate (see Table 1). This will prevent microbial growth. Store the column in a safe place at room temperature or 4 C. Never allow the column to freeze. 12 ENrich Instruction Manual

Section 4: Product Information Catalog Number Product Description 780-0001 ENrich Q 5 x 50 Column 780-0003 ENrich Q 10 x 100 Column 780-0021 ENrich S 5 x 50 Column 780-0023 ENrich S 10 x 100 Column 780-0091 ENrich 5 Frit Kit, includes 2 frits, 1 frit remover, 2 O-rings 780-0093 ENrich 10 Frit Kit, includes 2 frits, 1 frit remover, 2 O-rings 780-0008 1/16 10-32 Male Fittings, 2/pk ENrich Instruction Manual 13

Bio-Rad Laboratories, Inc. Life Science Group 10027324 Rev E US/EG Web site www.bio-rad.com USA 800 424 6723 Australia 61 2 9914 2800 Austria 01 877 89 01 Belgium 09 385 55 11 Brazil 55 11 3065 7550 Canada 905 364 3435 China 86 21 6169 8500 Czech Republic 420 241 430 532 Denmark 44 52 10 00 Finland 09 804 22 00 France 01 47 95 69 65 Germany 089 31 884 0 Greece 30 210 9532 220 Hong Kong 852 2789 3300 Hungary 36 1 459 6100 India 91 124 4029300 Israel 03 963 6050 Italy 39 02 216091 Japan 81 3 6361 7000 Korea 82 2 3473 4460 Mexico 52 555 488 7670 The Netherlands 0318 540666 New Zealand 64 9 415 2280 Norway 23 38 41 30 Poland 48 22 331 99 99 Portugal 351 21 472 7700 Russia 7 495 721 14 04 Singapore 65 6415 3188 South Africa 27 861 246 723 Spain 34 91 590 5200 Sweden 08 555 12700 Switzerland 026 674 55 05 Taiwan 886 2 2578 7189 Thailand 1800 88 22 88 United Kingdom 020 8328 2000 Sig 1213