[ Care and Use Manual ] styragel column. Contents. I. Introduction. II. Installing the Column Bank. III. Preparing Solvent and Samples
|
|
|
- Noel Haynes
- 9 years ago
- Views:
Transcription
1 styragel column I. INTRODUCTION This manual covers the care and use of the Waters Styragel HR, HT, and HMW families of Gel Permeation Chromatography (GPC) columns. Please take a few moments to read this manual carefully. Follow the recommendations in this manual to prolong column life and enhance chromatographic reproducibility. This introduction describes the three families of Waters Styragel columns: Styragel HR Styragel HT Styragel HMW Waters Styragel columns are packed with high-performance, fully porous, highly cross-linked styrene-divinylbenzene copolymer particles. Their different characteristics allow you to choose the column optimally suited to your application. Styragel columns are shipped in three solvents: toluene, tetrahydrofuran (THF), and dimethylformamide (DMF). Contents I. Introduction a. Styragel HR b. Styragel HT c. Styragel HMW II. Installing the Column Bank a. Preparing the GPC/HPLC System b. Installing the Columns c. Repairing Damaged Compression Screw Assemblies d. Equilibrating the Column Bank III. Preparing Solvent and Samples a. Preparing the Solvent b. Changing Solvents c. Preparing the Sample IV. Using the Column a. Chromatography Guidelines b. Calibrating the Column V. Care and Maintenance a. Troubleshooting b. Storing the Column c. Efficiency Testing 1. Testing Instrument Band Spreading 2. Column Efficiency Test Styragel Columns 1
2 a. Waters Styragel HR Use Waters Styragel HR columns for the high-resolution analysis of lowmolecular weight polymers, oligomers and additives. Packed with 5 µm particles, Waters Styragel HR columns provide the high plate counts necessary for this type of analysis. b. Waters Styragel HT Use Waters Styragel HT columns for the analysis of polymers with mid-range molecular-weight distributions. They are the most versatile columns for molecular-weight analysis. Waters Styragel HT columns are packed with 10 µm particles to provide dependable performance over a wide range of temperatures and solvents. c. Waters Styragel HMW Waters Styragel HMW columns are designed for the molecular-weight analysis of ultrahigh-molecular-weight polymers. Their 20 µm particle size together with the nominally 10 µm HMW frit design prevents the breakdown of ultrahigh molecularweight polymers due to shear, which can occur with smaller particles. This manual covers both column sizes 7.8 x 300 mm columns, and the solvent efficient 4.6 x 300 mm columns. In sections with recommend flow rates or spare parts, the 4.6 mm column conditions, or spare part follow immediately after the 7.8 mm recommendations. II. INSTALLING THE COLUMN BANK This chapter describes: Preparing the GPC/HPLC system Installing the columns Repairing damaged compression screw assemblies Equilibrating the column a. Preparing the GPC/HPLC System Before attaching the columns in the flow path on a GPC/HPLC system, you must first prepare the system: 3. Flush the system to remove any microparticulates and old solvents. Flush the injector loop if applicable. Band spreading The connection tubing and fittings in any chromatographic system contribute to extra-column band spreading. Before installing the column, me sure your system instrument band spreading (see Section on Testing Instrument Band Spreading). If this test is not possible with your system, refer to your system operator s manual. Narrow-Bore Chromatography for 4.6 mm Solvent Efficient Columns The peak volume in a narrow-bore system is so small, it is critical to minimize band spreading. Use the shortest tubing possible for all connections. It is not necessary to use a microbore flow cell in your detector or to change your conventional HPLC system in any way. Use inch (0.25 mm) i.d. tubing throughout the system. b. Installing the Columns When connecting columns in series, use the inch (0.25 mm) i.d. U-shaped column-joining tube supplied with each column. Sequence of columns in a column bank Generally, the results of an analysis are independent of the sequence in which a column bank is arranged. However, to improve resolution and column life, arrange the columns in order of decreasing pore size, with the column with the largest pore size closest to the injector. This is recommended because: The columns with the larger pore sizes are more rugged and are better able to tolerate the accumulation of extraneous materials. The species with the highest molecular-weight in the sample contributes the most to the viscosity of the sample. If the largest species is separated first, the viscosity decreases more quickly, placing less strain on the column bank. In the case of ultrahigh MW polymers, there is less shear on the sample. 1. Directly connect the system injector to the detector by replacing the old columns with a zero-dead-volume union. 2. Convert the system to the solvent in which the columns have been stored. For a new column set, this is the shipping solvent. Styragel Columns 2
3 Installing columns in a column bank Figure 1. Ferrule and Compression Screw Assembly To install your columns: 1. Remove the end plugs from each column and save them. 2. Connect the first column to the injector outlet tubing. Ensure that solvent flow is in the direction shown by the arrow on the column label. Finger-tighten the fittings, then tighten the fittings with a wrench by another 1/4 turn. 3. Connect the next column to the previous column using the U-tubes supplied with the columns. Ensure that solvent flow is in the direction shown by the arrow on the column label. Thread the inlet and outlet fittings of the U-tube until finger tight, then tighten the fittings by another turn with a wrench. 4. Repeat step 3, until all columns are connected. 5. Connect the last column to the detector inlet tubing using steps 1 through 4. c. Repairing Damaged Compression Screw Assemblies To remove a damaged compression screw or a worn ferrule assembly: 1. Scribe the circumference of the tubing at the desired break point using a tube cutter or a file with a cutting edge. 2. Grasp the tubing on both sides of the scribe mark with clothcovered pliers (to prevent marring the tube surface). Gently work the tube back and forth until it separates at the scribe mark. Ensure that the tubing end is straight, open, and free of burrs. Ferrule inch I.D. Tubing (0.25 mm) inch gap (3.3 mm) The distance between the end of the ferrule and the end of the U-tube may differ for different column types. If you have used columns from another manufacturer, it may be necessary to reset the ferrules, or make up a new fitting (see Figure 1). Use inch (0.25 mm) i.d. tubing for all lines between the injector and detector. For Waters columns, the critical distance is inch (2.25 mm). d. Equilibrating the Column Bank Equilibrate the columns when you install them and when you use them after they have been stored. To equilibrate your column bank: 1. Set the pump flow rate at 0.0 ml/min, then turn on the pump. 2. Increase the flow rate by 0.1 ml/min at 15 second intervals until you reach the final flow rate. 3. Purge the columns with the shipping solvent until you obtain a stable baseline. 3. Slide the compression screw, followed by the ferrule (large end of the taper first), over the tube. Properly bottom the tubing in the fitting seat. If the tubing is not completely seated, the resulting dead volume can lead to poor chromatographic results. Styragel Columns 3
4 Flow rate and backpressure for 4.6 mm Solvent Efficient Columns For best resolution and maximum column life, do not allow the flow rate to exceed 0.3 ml/min or the backpressure to exceed 3.5 MPa (500 psi, 35 atm) per column. Please note: The flow rate recorded on the Certificate of Analysis, included with the column, may be higher than the guidelines provided. For maximum column life in your lab, please follow the flow and back pressure guidance we provide. When using the HR 0.5, HR 1, and HR 2 columns, increase the flow rate in 0.1 ml/min increments at 30-second intervals until you reach 0.3 ml/min. Defective columns One or more defective columns in a series may cause the entire set of columns to appear defective. One defective column can cause peak spreading that cannot be overcome by any number of good columns. See Table 4, Table 5, or Table 6 for column efficiency data. Initial efficiency tests Test your system and columns before the first analysis. Run a test sample using the recommended parameters for your system and columns, and record the results. These results serve as a baseline to compare future performance. See Section on Efficiency Testing, for the procedures to determine the efficiency of your system and columns. Save the chromatograms from these tests. For the column-efficiency test, record the retention times, system settings, and all experimental conditions so they may be reproduced exactly for future comparison. III. Preparing Solvent and Samples This chapter describes: Preparing the solvent Changing solvents Converting columns to high temperature solvents Preparing the sample a. Preparing the Solvent Use clean solvent for reproducible results and maintenance-free operation. Use solvents of LC-grade or better, filtered to remove micro particulate matter larger than 0.45 µm. Refer to Waters catalog, filtration section for filter choice and solvent compatibility chart. b. Changing Solvents Solvent compatibility Waters Styragel columns are shipped in the solvent of your choice: toluene, THF or DMF. Some applications require a different solvent. Changing solvents works best between compatible solvents. Refer to the table below for solvent compatibilities. Note: The use of highly aqueous mobile phases may damage the resin and is not recommended. Table 1. Solvent Conversion Table To convert to: To use Waters Styragel columns for high-temperature chromatography in solvents like trichlorobenzene (TCB) or orthodichlorobenzene (ODCB), you must convert the column to the selected solvent at elevated temperature. High-temperature conversion procedure To convert the column bank to high-temperature operation: 1. Convert the system to the column shipping solvent at room temperature. 2. Purge solvent efficient columns at 0.1 ml/min while gradually increasing the temperature to 90 C over a minimum of 3 hours. Solvent Efficient Columns Use columns shipped in: To convert to: Use columns shipped in: o-dichlorobenzene Toluene Trichlorobenzene Toluene Hexafluoroisopropanol THF Phenol/TCB Toluene N-Methyl pyrrolidone DMF y-butyrolactone THF m-cresol DMF (Purge solvent efficient columns at 0.1 ml/min while gradually increasing the temperature to 90 C over a minimum of 3 hours.) Set the system to 55 C when converting columns from THF to HFIP or Y _ Butyrolactone. Styragel Columns 4
5 3. With the system at 90 C, convert to the high-temperature solvent at 0.1 ml/min using 12 ml per column. Then purge the system for a minimum of four column volumes at 0.2 ml/min. When using a bank of columns, multiply the number of column volumes specified in the procedure by the number of columns being used. 4. Convert to the high-temperature solvent using a flow rate of 0.1 ml/min. Use at least 20 ml per column. 5. Increase the temperature to the final conditions over a minimum of four hours while continuing to purge the column at 0.1 ml/min. Never exceed 150 C. Increase the temperature to the final conditions over a minimum of four hours while continuing to purge the column at 0.1 ml/min. Never exceed 150 C. 6. Adjust the flow rate to the final operating conditions. The optimal flow rate is 0.3 ml/min. Adjust the flow rate to the final operating conditions. The optimal flow rate is 0.3 ml/min. Returning to room temperature To return the columns to room temperature, reverse the above procedure. Alternatively, set the flow rate to 0.1 ml/min and reduce the temperature by 10 C every 30 minutes. Restarting the column To restart the column, maintain a flow rate of 0.1 ml/min and increase the temperature to the desired temperature over 10 hours. Then program the desired operating flow rate. c. Preparing the Sample Good sample preparation prolongs column life and ensures reproducible results. Take into account factors such as the capacity of the column, sample viscosity, and the type and sensitivity of the detector. Remove micro particulates with a 0.45 µm filter. Refer to Waters catalog for filter choice and solvent compatibility chart. Reactive polymers Some reactive polymers (such as epoxies) may condition the column. Improve column life and reproducibility by dedicating columns to specific classes of reactive polymers. Sample concentration Sample concentration affects both viscosity and injection volume. While small sample amounts produce narrower peaks, viscous samples may require larger, more dilute samples. Table 2 lists the recommended concentration of sample for optimal results. High-molecular-weight polymers High-molecular-weight polymers are especially susceptible to viscosity problems. When analyzing high-molecular-weight polymers, use the concentrations indicated in Table 2. Run narrow-distribution polymers, such as polystyrene standards, with an injection volume of 50 µl per column (20 µl per column for solvent efficient columns) at a concentration of 0.02 percent. Polystyrene standards with molecular weights of approximately four million or greater become increasingly susceptible to degradation by shearing in solution. Shearing is indicated by molecular-weight distributions that are broader than expected. With proper handling, polymers with molecular weights as high as 20 million can be handled successfully. For maximum column life, avoid temperature cycling. Maintain operating temperature but reduce flow rate to 0.1 ml/min when columns are not in use. Styragel Columns 5
6 Table 2. Recommended Sample Concentrations Molecular-weight Range Sample Concentration 0 to 25,000 <0.25% 25,000 to 200,000 <0.1 % 200,000 to 2,000,000 <0.05% Above 2,000,000 <0.02% Ultrahigh-molecular-weight polymers Shearing is also a factor to be considered with most ultrahigh-molecular-weight polymers. The effect of shearing may not be as easily observed as it is with narrow molecular-weight standards. Ultrahigh-molecular-weight polyolefin fractions are subject to possible shearing and incipient precipitation. Use especially dilute solutions of these fractions (for example, 0.02 percent) to minimize precipitation. IV. Using the Column This chapter describes: Chromatography guidelines Calibrating the column a. Chromatography Guidelines Gel permeation chromatography columns have a finite lifetime directly related to their care and use. Column life is reduced by contamination from samples and eluents, frequent solvent changeover, improper handling and storage, and temperature cycling. Guidelines for column use When using Waters Styragel columns, observe the following guidelines: For best resolution and maximum column life, do not exceed a flow rate of 1.0 ml/min or the backpressure exceed 3.5 MPa (500 psi, 35 atm) per column (corrected for system backpressure). Protect the column from rapid changes in pressure that can result from rapidly changing the composition of the solvent. When changing to a solvent with a different viscosity, it may be necessary to adjust the flow rate to stay below the operating backpressure specification of 3.5 MPa (500 psi, 35 atm) per column. Avoid precipitation by dissolving samples in the mobile phase. Always use high-quality HPLC solvents. Dedicate columns to specific applications. Frequent switching of samples and solvents accelerates column deterioration and loss of resolution. Using highly aqueous mobile phases may damage the resin and is not recommended. b. Calibrating the Column Whenever you replace a single column or a complete column bank, generate a new calibration curve to ensure the reproducibility of your application. Figure 2, Figure 3, and Figure 4 show typical calibration curves by column family. The calibration curves were obtained with polystyrene standards. Figure 2. Calibration Curves for Waters Styragel HR Columns Styragel HR Styragel HR Elution Volume (ml) Styragel HR 0.5 Styragel HR 1 Styragel Elution HR 2 Volume (ml) Styragel HR 3 Styragel HR 4Styragel HR 0.5 Styragel HR 5 Styragel HR 1 Styragel HR 6 Styragel Styragel HR 2 HR 4E Styragel HR 5E Styragel HR 3 Styragel HR 4 Styragel HR 5 Styragel HR 6 Styragel HR 4E Styragel HR 5E For solvent efficient columns, best resolution and maximum column life, do not let the flow rate exceed 0.3 ml/min or the backpressure exceed 3.5 MPa (500 psi, 35 atm) per column (corrected for system backpressure). Normal flow rate for these columns is 0.3 ml/min. Protect the column from vibration and mechanical shock. Minimize temperature cycling. Styragel Columns 6
7 Styragel HT Figure 3. Calibration Curves for Waters Styragel HT Columns Styragel HMW Figure 4. Calibration Curves for Waters Styragel HMW Columns Styragel HT Styragel HMW Elution Volume (ml) Styragel HT 2 Styragel HT 3 Styragel HT 4 Elution Volume (ml) Styragel HT 5 Styragel Styragel HT 6HT 2 Styragel Styragel HT 6E HT 3 Styragel HT 4 Styragel HT 5 Styragel HT 6 Styragel HT 6E Elution Volume (ml) Styragel HMW 2 Styragel HMW 6E Elution Styragel Volume HMW (ml) 7 Styragel HMW 2 Styragel HMW 6E Styragel HMW 7 V. Care and Maintenance This chapter describes: Troubleshooting Storing the column Efficiency testing a. Troubleshooting Table 3 describes specific problems, causes, and corrective actions. Table 3. Column Troubleshooting Problem Cause Corrective Action Buildup in system operating pressure Loss of resolution, broad peaks, low plate counts Inlet filter insert (from the first column only) plugged with particulates Injector and pump seal shedding Clogged tubing Filter inserts partially blocked Failing injector Insufficient equilibrium Column damaged Replace the inlet filter insert. Install a Styragel Guard column. Fit an in-line filter between the pump and the first column. Replace the tubing Replace or clean inlet and/or outlet filter inserts Repair the injector Continue equilibrium Replace column Styragel Columns 7
8 b. Storing the Column If you will be using the column again in less than 24 hours, special storage procedures are unnecessary. However, be sure that the columns never dry out. For longer storage periods, return the column to its box with the end plugs firmly in place for storage. Do not leave a column at elevated temperatures without solvent flow. For maximum column life, avoid temperature cycling. Maintain operating temperature and reduce flow rate to 0.1 ml/min when columns are not in use. To restart the column, maintain a flow rate of 0.1 ml/min and, if applicable, increase the temperature gradually over 10 hours. Then set the flow rate to the desired operating flow. c. Efficiency Testing Waters columns are tested for adherence to our specifications. Slight variations may occur depending on: The type and condition of equipment used The nature of the sample Instrument settings 1) Testing Instrument Band Spreading Test procedure The band spreading of a properly operating system should be less than 150 µl (75 µl for solvent efficient columns). To determine the band spreading of your system: 1. Remove the column(s) from the system. 2. Connect the inlet and outlet tubing with a zero dead-volume union. 3. Set the flow rate to 1.0 ml/min. 4. Use a fast chart speed to obtain a peak of easily measurable width. If you use a data system, set the sampling rate to at least 10 data points per second. You may need to adjust the detector sensitivity to keep the peak on scale. Figure 5. Method for Calculating Band Spreading Add calculation N= 25 (Vr/W)^2 Perform your own efficiency tests. To test both your system and each column in the system, carry out both of these tests: Instrument band-spreading test Column-efficiency test When to perform efficiency tests Perform efficiency tests each time you add columns or otherwise change your system. Then, as you use the system, conduct efficiency tests on a regular schedule. If problems occur during normal operation of the column, repeat the conditions for the initial efficiency tests and compare the results. Monitor instrument band spreading by performing a column efficiency test without the column in line. Resolve excessive band spreading before installing columns. Diluted/Distorted Sample Band 5. Inject the same sample as for a plate count determination (see Table 4, 5, or 6). 6. Measure the width of the resulting peak at 4.4% of the peak height to obtain a value in µl. Figure 5 illustrates the 5 sigma method, in which the width of the peak (w) is measured at 4.4% of the peak height (h). Styragel Columns 8
9 2) Column Efficiency Test The column-efficiency test described below may be used to calculate a theoretical plate count, which is a measurement of column efficiency within your system. Solvents It is not necessary to use the solvent that your column is shipped in to determine the efficiency of your column. Initial plate counts that are determined using different solvents will have different values. Test columns individually using your normal operating solvent. Test procedure To perform a column efficiency test: 1. Slowly increase the flow rate to 1.0 ml/min for the HR 0.5, HR 1, HR 2, and HR 4E columns, or to 2.0 ml/min for all other columns. 2. Adjust the detector to an attenuation that achieves a peak of 70 percent full scale (noise level <0.5 percent full scale). 3. Set the recorder chart speed to 50 mm/min, or the data system sampling rate to at least 5 data points per second. 4. Inject up to 20 µl of marker solution. Use a solution of up to 10 percent marker in solvent. Use up to 5 µl for solvent efficient columns. Refer to Table 4, Table 5, or Table 6 for the marker to use when performing the efficiency test. 5. Record the retention time, instrument settings, and column configuration so you can reproduce them exactly for comparison in the future. 6. Compute the plate count using the tangent method (see Figure 6). Use these results for comparison throughout the life of your column. 7. Replace any column that exhibits a plate count more than 30 percent below the original value. Table 4. Conditions and sample for 7.8 mm for Column Efficiency Test: HR Columns Column Type 1 Dicyclohexylphthalate Toluene THF DMF ml/min HR 0.5 Acetone Propylbenzene Ethyleneglycol 1.0 HR 1 Acetone Propylbenzene Ethyleneglycol 1.0 HR 2 Acetone Propylbenzene Ethyleneglycol 1.0 HR 3 ODCB Acetone Acetone 2.0 HR 4 ODCB Acetone Acetone 2.0 HR 4E Acetone Propylbenzene Ethyleneglycol 1.0 HR 5E DCHP 2 Acetone Acetone 2.0 Table 5. Conditions and Sample for 4.6 mm Column Efficiency Test: HR Columns Column Type Toluene THF DMF HR 0.5 ODCB 0.3 Acetone 0.3 Acetone 0.3 HR 1 ODCB 0.3 Acetone 0.3 Acetone 0.3 HR 2 ODCB 0.3 Acetone 0.3 Acetone 0.3 HR 3 ODCB 0.3 Acetone 0.3 Acetone 0.3 HR 4 ODCB 0.3 Acetone 0.3 Acetone 0.3 HR 4E ODCB 0.3 Acetone 0.3 Acetone 0.3 HR 5E DCHP Acetone 0.3 Acetone 0.3 Styragel Columns 9
10 Table 6. Conditions and sample for 7.8 mm for Column Efficiency Test: HT Columns1 Column Type Toluene THF DMF HT 3 ODCB Acetone Acetone 2.0 HT 4 ODCB Acetone Acetone 2.0 HT 5 ODCB Acetone Acetone 2.0 HT 6 DCHP Acetone Acetone 2.0 HT 6E DCHP Acetone Acetone 2.0 Table 8. Test Condition and Sample for Efficiency Test 7.8 mm HMW Columns Column Type Toluene THF DMF HMW 7 DCHP Acetone Acetone 2.0 HMW 6E DCHP Acetone Acetone 2.0 Table 7. Conditions and sample for 4.6 mm for Column Efficiency Test: HT Columns1 Column Type Toluene THF DMF HR 3 ODCB 0.3 Acetone 0.3 Acetone 0.3 HR 4 ODCB 0.3 Acetone 0.3 Acetone 0.3 HR 5 ODCB 0.3 Acetone 0.3 Acetone 0.3 HR 6 DCHP 0.3 Acetone 0.3 Acetone 0.3 HR 6E DCHP 0.3 Acetone 0.3 Acetone 0.3 Figure 6. Tangent Method for Calculating Column Efficiency N = 16 (tr/w) 2 t R = Elution distance from injection point to the peak apex on the chart paper. W= Peak width of marker measured where tangents drawn on peak intersect baseline (same units as trj. Styragel Columns 10
11 Sales Offices Austria and European Export (Central South Eastern Europe, CIS and Middle East) Australia Belgium Brazil Canada China Czech Republic Denmark Finland France Germany Hong Kong Hungary Norway Poland Puerto Rico Russia/CIS / Singapore Spain Sweden Switzerland Taiwan United Kingdom All other countries: Waters Corporation U.S.A India Ireland Italy Japan Korea Mexico The Netherlands Waters Corporation. Waters, The Science of What s Possible, and Styragel are trademarks of Waters Corporation. January 2011 WAT Rev 4 VW-PDF Waters Corporation 34 Maple Street Milford, MA U.S.A. T: F: Styragel Columns 11
[ Care and Use Manual ]
Analytical columns and cartridges i. Introduction Waters offers a large selection of analytical columns and cartridges packed with silica-based packing materials. Follow this instruction sheet to obtain
[ Care and Use Manual ]
The Waters GPC Cleanup System uses two Envirogel GPC Cleanup Columns to perform routine cleanup of environmental samples as specified by EPA Method 3640A, GPC Cleanup. The columns, 19 x 150 mm and 19 x
ENrich SEC 70 ENrich SEC 650 High-Resolution Size Exclusion Columns Instruction Manual
ENrich SEC 70 ENrich SEC 650 High-Resolution Size Exclusion Columns Instruction Manual Catalog numbers 780-1070 780-1650 Please read these instructions before you use ENrich SEC high-resolution size exclusion
[ Care and Use Manual ]
WATERS IC-PAK ION EXCLUSION COLUMNS Contents I. Introduction II. Installation AND EQUILIBRATING THE COLUMn a. Preparation b. Installing the Column c. Equilibration III. PREPARATION OF ELUENTS AND SAMPLES
INSTRUCTIONS 56-1190-98. Edition AC
Sephacryl S-100 High Resolution Sephacryl S-200 High Resolution Sephacryl S-300 High Resolution Sephacryl S-400 High Resolution Sephacryl S-500 High Resolution INSTRUCTIONS Sephacryl High Resolution chromatography
Waters Corporation. Waters 2690/5 USER & TROUBLESHOOTING GUIDE
Waters Corporation Waters 2690/5 USER & TROUBLESHOOTING GUIDE Contents 2690/5 Theory Setup procedures. Troubleshooting the 2690/5 User maintenance of the 2690/5 Spare Parts 2 2690/5 Theory 2690/5 Solvent
[ Care and Use Manual ]
PREP Calibration Mix DIOS Low i. Introduction Pre-packaged PREP Calibration Mixtures eliminate the need to purchase and store large quantities of the component calibration reagents, simplifying sample
ProteinChip Energy Absorbing Molecules (EAM)
ProteinChip Energy Absorbing Molecules (EAM) Instruction Manual Catalog #C30-00001, #C30-00002, #C30-00003, #C30-00004 For technical support, call your local Bio-Rad office, or in the US, call 1-800-4BIORAD
Gas Chromatography Liner Selection Guide
Gas Chromatography Liner Selection Guide Peter Morgan, Thermo Fisher Scientific, Runcorn, Cheshire, UK Technical Note 20551 Key Words Liner, focus Abstract The liner serves an important function in allowing
Image Lab Software for the GS-900 Densitometer
Image Lab Software for the GS-900 Densitometer Quick Start Guide Catalog # 170-9690 Bio-Rad Technical Support For help and technical advice, please contact the Bio-Rad Technical Support department. In
[ Care and Use Manual ]
ACQUITY UPLC CSH Columns Contents I. Getting started a. Column Connectors b. Column Installation c. Column Equilibration d. ecord Installation e. Initial Column Efficiency Determination f. VanGuard Pre-Columns
Transforming a ChromLab Software 2-D Purification Template into an Automated Multidimensional (Multi-D) Purification Workflow An Instructional Guide
Transforming a ChromLab Software -D Purification Template into an Automated Multidimensional (Multi-D) Purification Workflow An Instructional Guide Instructional Guide Chromatography Bulletin 675 ChromLab
A Complete Solution for Method Linearity in HPLC and UHPLC
Now sold under the Thermo Scientific brand A Complete Solution for Method Linearity in HPLC and UHPLC Frank Steiner, Fraser McLeod, Tobias Fehrenbach, and Andreas Brunner Dionex Corporation, Germering,
How To Install A Gel Doc Xr+ Camera On A Computer Or Mouse Or Tv (For A Powerbook) With A Zoom Lens (For An Ipa) With An Iphone Or Ipa (For Mac) With The Zoom Lens
Molecular Imager Gel Doc XR+ Installation Guide To install your Gel Doc XR+, perform these seven steps in the following order: 1. Unpack the Gel Doc XR+ system. 2. Install Image Lab software. 3. Set up
[NUGENESIS SAMPLE MANAGEMENT ] AMPLE IMPROVING LAB EFFICIENCY, ANAGEMENT ACCELERATING BUSINESS DECISIONS. bigstock.com $69
[NUGENESIS SAMPLE MANAGEMENT ] AMPLE ACCELERATING BUSINESS DECISIONS IMPROVING LAB EFFICIENCY, ANAGEMENT bigstock.com $69 [ N U GENESIS SAMPLE MANAGEMENT ] MANAGING SAMPLES SHOULD NEVER BE A BARRIER TO
Rheodyne Manual Sample Injector Valves
Rheodyne Manual Sample Injector Valves Rotor Seal Shaft End Rotor Seal Groove Stator Face Stator Stator Face Seal Rotor Seal Grooves Position A Rotor Seal Grooves Position B Position C Flow paths showing
Thermo Scientific Dionex Chromeleon 7 Chromatography Data System Software
Thermo Scientific Dionex Chromeleon 7 Chromatography Data System Software Frank Tontala, Thermo Fisher Scientific, Germering, Germany Technical Note 708 Key Words Chromeleon Chromatography Data System,
WATERS QUANTITATIVE ANALYSIS solutions
More sensitivity. More speed. What more can you ask for in a complete quantitative analysis solution? WATERS QUANTITATIVE ANALYSIS solutions THE CHALLENGE OF QUANTITATIVE ANALYSIS Tandem mass spectrometry,
1.1 This test method covers the qualitative and quantitative determination of the content of benzene and toluene in hydrocarbon wax.
Standard Method for Analysis of Benzene and Toluene Content in Hydrocarbon Waxes by Headspace Gas Chromatography EWF METHOD 002/03 (Version 1 Reviewed 2015) 1 Scope 1.1 This test method covers the qualitative
INTRODUCTION. The principles of which are to:
Taking the Pain Out of Chromatographic Peak Integration Shaun Quinn, 1 Peter Sauter, 1 Andreas Brunner, 1 Shawn Anderson, 2 Fraser McLeod 1 1 Dionex Corporation, Germering, Germany; 2 Dionex Corporation,
Installing a Capillary Column into an ECD and a Micro ECD
Installing a Capillary Column into an ECD and a Micro ECD Procedure: Preparing capillary columns You must prepare your capillary column before installation. Proper preparation assures that the column end
ApplicationNOTE Introduction
Introduction Honey, like other foods, is subject to strict quality control measures before it can be sold commercially. It is a natural product, produced solely by bees from flower and tree pollen. Human
802.11ac Power Measurement and Timing Analysis
802.11ac Power Measurement and Timing Analysis Using the 8990B Peak Power Analyzer Application Note Introduction There are a number of challenges to anticipate when testing WLAN 802.11ac [1] power amplifier
Application Note. USD 2995 (a) High Throughput Regeneration Study on MEP HyperCel Mixed-Mode Sorbent on ScreenExpert RoboColumns u
Application Note USD 2995 (a) High Throughput Regeneration Study on MEP HyperCel Mixed-Mode Sorbent on ScreenExpert RoboColumns u Summary A regeneration study of MEP HyperCel mixed-mode sorbent was conducted
Successful HPLC Operation. A Troubleshooting Guide Version 1.1
c h r o m a t o g r a p h y Successful HPLC Operation A Troubleshooting Guide Version 1.1 Analyze Detect Measure Control Table of Contents 1. Introduction 4 1.1 Purpose 4 1.2 Content 4 2. Troubleshooting
Foreign Taxes Paid and Foreign Source Income INTECH Global Income Managed Volatility Fund
Income INTECH Global Income Managed Volatility Fund Australia 0.0066 0.0375 Austria 0.0045 0.0014 Belgium 0.0461 0.0138 Bermuda 0.0000 0.0059 Canada 0.0919 0.0275 Cayman Islands 0.0000 0.0044 China 0.0000
Fractional Distillation and Gas Chromatography
Fractional Distillation and Gas Chromatography Background Distillation The previous lab used distillation to separate a mixture of hexane and toluene based on a difference in boiling points. Hexane boils
Diafiltration: A Fast, Efficient Method for Desalting, or Buffer Exchange of Biological Samples
Scientific & Technical Report PN 33289 Diafiltration: A Fast, Efficient Method for Desalting, or Buffer Exchange of Biological Samples By Larry Schwartz, Senior Technical Manager, Pall Life Sciences Diafiltration
Image Lab Software How to Obtain Stain-Free Gel and Blot Images. Instructions
Image Lab Software How to Obtain Stain-Free Gel and Blot Images Instructions Table of Contents Obtaining Stain-Free Gel images...1 5 Obtaining Stain-Free Blot Images...6 9 If the Gel Doc XR+ or ChemiDoc
relating to household s disposable income. A Gini Coefficient of zero indicates
Gini Coefficient The Gini Coefficient is a measure of income inequality which is based on data relating to household s disposable income. A Gini Coefficient of zero indicates perfect income equality, whereas
1.5 GHz Active Probe TAP1500 Datasheet
1.5 GHz Active Probe TAP1500 Datasheet Easy to use Connects directly to oscilloscopes with the TekVPI probe interface Provides automatic units scaling and readout on the oscilloscope display Easy access
Exceptional Accuracy and Durability for Tough Industrial Environments INDUSTRIAL FLOOR SCALES. VERTEX Industrial Floor Scales Heavy-Duty Design
VERTEX Industrial Floor Scales Heavy-Duty Design LOCATE HI-RES Exceptional Accuracy and Durability for Tough Industrial Environments INDUSTRIAL FLOOR SCALES METTLER TOLEDO VERTEX Floor Scales Heavy-Duty
To install the I/O Module Board: 1. Remove AC power from decoder.
Installation Data (Catalog Number 2755-NB0, -NB1, -NB2) This document shows how to install the following components in the 2755-DSxx ➀ and 2755-DDxx ➀ Bar Code Decoders. I/O Module Board (Catalog No. 2755-NB1
Accuracy counts! SENSORS WITH ANALOG OUTPUT
Accuracy counts! SENSORS WITH ANALOG OUTPUT OTHER APPLICATIONS: KEY ADVANTAGES: Distance measurement Positioning Profile detection Deformation monitoring Vibration monitoring Process monitoring Detection
Supported Payment Methods
Supported Payment Methods Global In the global payments market, credit cards are the most popular payment method. However, BlueSnap expands the payment selection by including not only the major credit
Report on Government Information Requests
Report on Government Information July 1 - December 31, 2014 apple Apple takes our commitment to protecting your data very seriously and we work incredibly hard to deliver the most secure hardware, software
Method Development for Size-Exclusion Chromatography of Monoclonal Antibodies and Higher Order Aggregates
Method Development for Size-Exclusion Chromatography of Monoclonal Antibodies and Higher Order Aggregates Paula Hong and Kenneth J. Fountain Waters Corporation, 34 Maple St., Milford, MA, USA APPLICATION
MERCER S COMPENSATION ANALYSIS AND REVIEW SYSTEM AN ONLINE TOOL DESIGNED TO TAKE THE WORK OUT OF YOUR COMPENSATION REVIEW PROCESS
MERCER S COMPENSATION ANALYSIS AND REVIEW SYSTEM AN ONLINE TOOL DESIGNED TO TAKE THE WORK OUT OF YOUR COMPENSATION REVIEW PROCESS MERCER S COMPENSATION ANALYSIS AND REVIEW SYSTEM www.imercer.com/cars Mercer
Troubleshooting Common HPLC Problems. http://www.hplc1.com/shodex/english/dd.htm
Troubleshooting Common HPLC Problems http://www.hplc1.com/shodex/english/dd.htm 300 mv 0.00 5.00 1 2 3 Minutes 4 1. Fucose 2. Galactosamine 3. Glucosamine 4. Galactose 5. Glucose 6. Mannose 5 6 20.00 HPLC
Supported Payment Methods
Sell Globally in a Snap Supported Payment Methods Global In the global payments market, credit cards are the most popular payment method. However, BlueSnap expands the payment selection by including not
Report on Government Information Requests
Report on Government Information January 1 - June 30, 2015 Apple takes our commitment to protecting your data very seriously and we work incredibly hard to deliver the most secure hardware, software and
Determination of Calcium and Magnesium in Brine
Application Note Determination of Calcium and Magnesium in Brine INTRODUCTION To prevent membrane poisoning, new membrane technology in chlor-alkali cells requires feed brine that is relatively free of
How to Measure 5 ns Rise/Fall Time on an RF Pulsed Power Amplifier Using the 8990B Peak Power Analyzer
How to Measure 5 ns Rise/Fall Time on an RF Pulsed Power Amplifier Using the 8990B Peak Power Analyzer Application Note Introduction In a pulsed radar system, one of the key transmitter-side components
Chapter 28: High-Performance Liquid Chromatography (HPLC)
Chapter 28: High-Performance Liquid Chromatography (HPLC) Scope Instrumentation eluants, injectors, columns Modes of HPLC Partition chromatography Adsorption chromatography Ion chromatography Size exclusion
Agilent 87421A/87422A Power Supply
Agilent 87421A/87422A Power Supply Technical Overview Designed specifically for Agilent Technologies microwave system amplifiers Bias cable permits remote placement Compact size for easy system integration
Keysight Technologies Using Fine Resolution to Improve Thermal Images. Application Note
Keysight Technologies Using Fine Resolution to Improve Thermal Images Application Note Introduction Digital imaging opened up new possibilities and has been very much part of our lives since social media
ÄKTA system Training Guide
ÄKTA system Training Guide Explorer 00 / 0 amersham bioscience amersham bioscience amersham bioscience amersham bioscience ÄKTAexplorer ÄKTAexplorer Buffers Injector Box amersham bioscience Monitor Conductivity
Computer Specifications
Computer Specifications Specifications Imaging Systems Bulletin 6126 These tables list current specifications for Bio-Rad imaging systems. Refer to the following pages for information about each imaging
AutoSpec Premier. Are you? Analyzing dioxins and related compounds? Required to satisfy regulatory requirements?
AutoSpec Premier Are you? Analyzing dioxins and related compounds? Required to satisfy regulatory requirements? Looking for sensitivity, selectivity and dynamic range? The AutoSpec Premier is the latest
Report on Government Information Requests
Report on Government Information July 1 - December 31, 2015 Apple takes our commitment to protecting your data very seriously and we work incredibly hard to deliver the most secure hardware, software and
PXI and AXIe Modular Instrumentation
PXI and AXIe Modular Instrumentation M9018A PXIe Chassis M9502A AXIe Chassis M9505A AXIe Chassis U4002A Digital Test Console Chassis Tested Computer List Technical Note DISCOVER the Alternatives......
[ EMPOWER 3 SOFTWARE ] MORE CAPABILITIES FOR YOUR LAB, MORE VALUE FOR YOUR ENTERPRISE
MORE CAPABILITIES FOR YOUR LAB, MORE VALUE FOR YOUR ENTERPRISE Empower 3 Software offers single-window control of a wide range of analytical instruments: Waters HPLC, ACQUITY UPLC, ACQUITY UPC, 2 and integrated
Building a Global Internet Company: Driving Traffic to Your Site. Benjamin Edelman Harvard Business School
Building a Global Internet Company: Driving Traffic to Your Site Benjamin Edelman Harvard Business School Mumbai, India January 20, 2012 About me Assistant Professor, Harvard Business School Teaching:
Application Note. Determination of Nitrite and Nitrate in Fruit Juices by UV Detection. Summary. Introduction. Experimental Sample Preparation
Application Note Determination of Nitrite and Nitrate in Fruit Juices by UV Detection Category Food Matrix Fruit Juice Method HPLC Keywords Ion pair chromatography, fruit juice, inorganic anions AZURA
Pall in the Brewery (more value per hectolitre) DE Free Clarification FB1795
Pall in the Brewery (more value per hectolitre) DE Free Clarification FB1795 The Pall World Pall Corporation is a world leader in the design, production and supply of filters, membranes and systems for
DRAM Memory Modules. Pentium (Use gold-plated SIMMs) Memory Module
Instructions D Memory Modules (For 6180 Industrial Computers with Pentium and Pentium II Processors) This document describes how to add Memory to the 6180 Computer processor board. Topics include: Available
Thermo Scientific Prelude SPLC System FPO. Making LC/MS accessible. to clinical research and toxicology labs
Thermo Scientific Prelude SPLC System FPO Making LC/MS accessible to clinical research and toxicology labs Specifically designed to make the power of LC/MS accessible Analysis of target compounds in complex
Get the benefits of Norgren s unique range of Online services
Get the benefits of Norgren s unique range of Online services Make your job easier and save time - everything you need to select, design and purchase Norgren pneumatics is in one convenient location, available
Keysight Technologies How to Choose your MAC Lever. Technical Overview
Keysight Technologies How to Choose your MAC Lever Technical Overview Introduction Atomic force microscopy (AFM) is a sub-nanometer scale imaging and measurement tool that can be used to determine a sample
How To Get A New Phone System For Your Business
Cisco Phone Systems Telemarketing Script Cold Call 1. Locate Contact: Name listed Owner General Manager / Office Manager Chief BDM (Business Decision Maker) Note: Avoid talking to IT since this is not
AT-S39 Version 1.3 Management Software for the AT-8024 and AT-8024GB Fast Ethernet Switches. Software Release Notes
AT-S39 Version 1.3 Management Software for the AT-8024 and AT-8024GB Fast Ethernet Switches Supported Platforms Software Release Notes Please read this document before you begin to use the AT-S39 management
Determination of Fat in Dried Milk Products Using Accelerated Solvent Extraction (ASE)
Application Note 340 Determination of Fat in Dried Milk Products INTRODUCTION Many extraction techniques for the determination of fat in food are labor-intensive or require long extraction times. The Roese-Gottlieb
Determination of Pesticide Residues in Drinking Water Using Automated Solid-Phase Extraction and Gas Chromatography with Nitrogen Phosphorus Detection
Determination of Pesticide Residues in Drinking Water Using Automated Solid-Phase Extraction and Gas Chromatography with Nitrogen Phosphorus Detection Application Note 1097 Liu Qian, Zheng Hongguo and
Cell Sorting. S3e Cell Sorter Quick Guide
Cell Sorting S3e Cell Sorter Quick Guide STARTUP Checking Fluidics STARTUP There are three bulk fluid containers (Figure 1) that may require attention before starting the system. If fluidics need to be
Dissolved and precipitated oxalate
Accepted 2005 Process liquors from bleach plants Dissolved and precipitated oxalate Using Ion Chromatography 0 Introduction In bleach plants of pulp mills with a high degree of system closure, there is
LC-MS/MS Method for the Determination of Docetaxel in Human Serum for Clinical Research
LC-MS/MS Method for the Determination of Docetaxel in Human Serum for Clinical Research J. Jones, J. Denbigh, Thermo Fisher Scientific, Runcorn, Cheshire, UK Application Note 20581 Key Words SPE, SOLA,
Purification of reaction mixtures using flash chromatography.
Purification of reaction mixtures using flash chromatography. This technical note details the use of ISOLUTE Flash chromatography columns for the purification of reaction mixtures. What is flash chromatography?
Quantum View Manage Administration Guide
2010 United Parcel Service of America, Inc. UPS, the UPS brandmark and the color brown are trademarks of United Parcel Service of America, Inc. All rights reserved. Quantum View Manage Administration Guide
PCI Express Probes for Agilent E2960B PCI Express Analysis Systems
PCI Express for Agilent E2960B PCI Express Analysis Systems Version: 1.2 Superior Signal Probing options to address diverse designs and form factors As an industry leader in the innovation of non-intrusive,
Gel Filtration Standard
Gel Filtration Standard Instruction Manual Catalog # 151-1901 Table of Contents Section 1 Introduction... 1 1.1 Instructions... 1 1.2 Recommended Volume of Standard... 2 1.3 Shelf Life... 5 1.4 Storage...
Thermo Scientific SOLAµ SPE Plates Technical Guide. Consistent excellence. for bioanalysis
Thermo Scientific SOLAµ SPE Plates Technical Guide Consistent excellence for bioanalysis SOLAµ - delivering reproducible low volume extractions. Everytime! Thermo Scientific SOLAµ plates are designed for
Reporting practices for domestic and total debt securities
Last updated: 4 September 2015 Reporting practices for domestic and total debt securities While the BIS debt securities statistics are in principle harmonised with the recommendations in the Handbook on
Agilent 8762/3/4A,B,C Coaxial Switches
Agilent 8762/3/4A,B,C Coaxial Switches Technical Overview High performance switches for microwave and RF instrumentation and systems Agilent Technologies offers a ver satile line of multiport coaxial switches.
How To Build A Powerline Box Powerline Generator
Quick Reference Guide DYNAMIC Series Dynamic Series is a broad product family that provides connector solutions ranging from signal level circuitry through power circuit connectivity in a ruggedized, industrialized
Hydraulic. Manifold ON/OFF ISOLATION CONTROL FOR TEST SYSTEMS AND HYDRAULIC ACTUATORS WHAT MOVES YOUR WORLD. Rev.-, October 2011
Hydraulic SERVICE Manifold ON/OFF ISOLATION CONTROL FOR TEST SYSTEMS AND HYDRAULIC ACTUATORS Rev.-, October 2011 WHAT MOVES YOUR WORLD INTRODUCTION Whenever the highest levels of motion control performance
Make the invisible visible! SENSORS WITH EXCELLENT BACKGROUND SUPPRESSION
Make the invisible visible! SENSORS WITH EXCELLENT BACKGROUND SUPPRESSION photoelectric sensors Thanks to its vision for innovation and technological enhancement, Contrinex keeps on setting new standards
BT Premium Event Call and Web Rate Card
BT Managed Event and BT Self-Managed Event (also referred to as Express, Plus and Premium) Conference Bridge and Call for Booked Audio Conferencing Services will comprise the following for each phone-conference:
HiTrap Heparin HP, 1 ml and 5 ml
Instructions 71-7004-00 AU HiTrap affinity columns HiTrap Heparin HP, 1 ml and 5 ml HiTrap Heparin HP is a prepacked ready to use, column for preparative affinity chromatography. The special design of
Report on Government Information Requests
Report on Government Information Requests January - June, Apple takes our commitment to protecting your data very seriously and we work incredibly hard to deliver the most secure hardware, software and
ST13005. High voltage fast-switching NPN power transistor. Features. Applications. Description
High voltage fast-switching NPN power transistor Datasheet production data Features Low spread of dynamic parameters Minimum lot-to-lot spread for reliable operation Very high switching speed Applications
Working Holiday Maker visa programme report. 31 December 2014
Working Holiday Maker visa programme report 31 December 2014 Contents Page About this report 1 Enquiries 1 Definition of terms 2 Background to the Working Holiday Maker programme 3 Recent developments
Evaluating System Suitability CE, GC, LC and A/D ChemStation Revisions: A.03.0x- A.08.0x
CE, GC, LC and A/D ChemStation Revisions: A.03.0x- A.08.0x This document is believed to be accurate and up-to-date. However, Agilent Technologies, Inc. cannot assume responsibility for the use of this
Model 43AP Pneumatic Controller, Style B
Instruction MI 011-476 January 1980 Model 43AP Pneumatic Controller, Style B Installation and Operation Model 43AP Controller continuously detects the difference between a process measurement and its set
Installation Qualification/Operational Qualification Protocols and Instructions. Experion Automated Electrophoresis Station and Software
Installation Qualification/Operational Qualification Protocols and Instructions Experion Automated Electrophoresis Station and Software Validation Kit (Catalog #700-7051) Security Edition Software (Catalog
PXI and AXIe Modular Instrumentation
PXI and AXIe Modular Instrumentation M9018A PXIe Chassis M9502A AXIe Chassis M9505A AXIe Chassis Tested Computer List Technical Note DISCOVER the Alternatives...... Agilent MODULAR Products OVERVIEW This
Field Safety Notice Urgent Medical Device Correction 2955842-03-13-2015-003-C Cloudy Appearance and Potential Tears on System Drapes
Field Safety Notice Urgent Medical Device Correction 2955842-03-13-2015-003-C Cloudy Appearance and Potential Tears on System Drapes Introduction and Reason for Field Action Dear da Vinci Customer, This
ChromQuest 5.0 Chromatography Data System
ChromQuest 5.0 Chromatography Data System User Guide for the Surveyor LC CHROM-97257 Revision A March 2008 2008 Thermo Fisher Scientific Inc. All rights reserved. Surveyor and ChromQuest are trademarks
ß-CYCLODEXTRIN SYNONYMS
ß-CYCLODEXTRIN Prepared at the 44th JECFA (1995), published in FNP 52 Add 3 (1995) superseding specifications prepared at the 41st JECFA (1993), published in FNP 52 Add 2 (1993). Metals and arsenic specifications
What Proportion of National Wealth Is Spent on Education?
Indicator What Proportion of National Wealth Is Spent on Education? In 2008, OECD countries spent 6.1% of their collective GDP on al institutions and this proportion exceeds 7.0% in Chile, Denmark, Iceland,
Guide to Reverse Phase SpinColumns Chromatography for Sample Prep
Guide to Reverse Phase SpinColumns Chromatography for Sample Prep www.harvardapparatus.com Contents Introduction...2-3 Modes of Separation...4-6 Spin Column Efficiency...7-8 Fast Protein Analysis...9 Specifications...10
NOMADIX, INC Nomadix Premium Support Policy
This End User Support Policy ( Policy ) sets forth the terms under which Nomadix, Inc., ( Nomadix ) upon Customer s payment of applicable fees and successful installation, will provide technical support
White Paper. 10 Reasons to Choose Image-based Barcode Readers
White Paper 10 Reasons to Choose Image-based Barcode Readers Introduction Image-based barcode readers are rapidly replacing laser scanners in a wide range of industries including food, beverage, consumer
amplification tech Optical Design of CFX96 Real-Time PCR Detection System Eliminates the Requirement of a Passive Reference Dye
amplification tech note 6047 Optical Design of CFX96 Real-Time PCR Detection System Eliminates the Requirement of a Passive Reference Dye Liz Jordan and Richard Kurtz, Gene Expression Division, io-rad
Diode, Schottky SMB/DO-214AA. Page <1> 24/04/08 V1.1. Dimensions Inches (Millimeters)
Features: For surface mounted application. Metal to silicon rectifier, majority carrier conduction. Low forward voltage drop. Easy pick and place. High surge current capability. Plastic material. Epitaxial
Analysis of Vegetable Oils by High Performance Liquid Chromatography
Analysis of Vegetable Oils by High Performance Liquid Chromatography Using Evaporative Light Scattering Detection and Normal Phase Eluents Andrew J. Aubin, Cecilia B. Mazza, Donald A. Trinite, and Patricia
PCI Express 1.0 (2.5 GT/s) Protocol Test
PCI Express 1.0 (2.5 GT/s) Protocol Test Configuration and Pricing Guide Gen1 today, ready for Gen2 in the future As PCI Express Gen2 becomes more mature in the market, you must decide whether to continue
NetFlow Feature Acceleration
WHITE PAPER NetFlow Feature Acceleration Feature Description Rapid growth in Internet and intranet deployment and usage has created a major shift in both corporate and consumer computing paradigms. This
OCTOBER 2010. Russell-Parametric Cross-Sectional Volatility (CrossVol ) Indexes Construction and Methodology
OCTOBER 2010 Russell-Parametric Cross-Sectional Volatility (CrossVol ) Indexes Construction and Methodology SEPTEMBER 2010 Russell-Parametric Cross-Sectional Volatility (CrossVol) Indexes Construction
The Theory of HPLC. Gradient HPLC
The Theory of HPLC Gradient HPLC i Wherever you see this symbol, it is important to access the on-line course as there is interactive material that cannot be fully shown in this reference manual. Aims
MOLECULAR DRAG PUMPS MDP SERIES
MOLECULAR DRAG PUMPS MDP SERIES MOLECULAR DRAG PUMPS MDP SERIES If you need: to replace a rotary vane pump with a dry pump for an application at 10-3 a turbopump that can operate up to 10 a pump that can
