Comparison of a New Desktop Spirometer (Diagnosa ) with a Laboratory Spirometer

Similar documents
PULMONARY FUNCTION TESTS A Workshop on Simple Spirometry & Flow Volume Loops

Spirodoc. One Touch Laboratory for respiratory analysis two functional modes: doctor and patient. Spirometer with Touch Screen display

Forced vital capacity: maximal volume of air exhaled with maximally forced effort from a maximal inspiration.

Interpretation of Pulmonary Function Tests

Vmax Vyntus SPIRO. Advance your lung function testing experience

Twenty years of innovative solutions in Spirometry, Oximetry, Telemedicine

MicroLab. The world s most versatile. portable Spirometer

Spirodoc. 3D Laboratory for respiratory analysis two functional modes: doctor and patient. Four devices in one. Spirometer with Touch Screen display

SPIROMETRY FOR HEALTH CARE PROVIDERS Global Initiative for Chronic Obstructive Lung Disease (GOLD)

A Telehomecare monitoring project in Cystic Fibrosis

COURSE SYLLABUS RC 223 CLINICAL-3

Vitalograph Products and Services Guide

Spirometry Workshop for Primary Care Nurse Practitioners

RESPIRATORY VENTILATION Page 1

Point-of-care (POC) versus central laboratory instrumentation for monitoring oral anticoagulation

COVER PAGE FOR the PDF file of PULMONARY FUNCTION TESTING

CLINICAL NURSE LPN, RN, BSN,

Position Statement from the Irish Thoracic Society on the treatment of Idiopathic Pulmonary Fibrosis

THE EU PEAK FLOW METER

Sponsor Novartis Pharmaceuticals

Defending the Rest Basics on Lung Cancer, Other Cancers and Asbestosis: Review of the B-Read and Pulmonary Function Testing

Spirometry, the most frequently performed

Innovation in Spirometry Oximetry Telemedicine

Pulmonary Function Testing: Coding and Billing Issues

INVITATION FOR QUOTATIONS FOR SUPPLY OF EQUIPMENT THIS INVITATION DULY SIGNED ON ALL PAGES SHOULD BE ATTACHED WITH THE QUOTE

M.B.,Ch.B. (U.S.), B.Sc.Med.Sc.Hons.(Epidemiology) (U.S.), Dip. Occ. Health (U.S.) 1976 Matriculated, La Rochelle Girls High School, Paarl

PTE Pediatric Asthma Metrics Reporting Updated January 2015

MicroLab. Operating Manual

Spirometry Testing in Occupational Health Programs. Best Practices for Healthcare Professionals

NIOSH SPIROMETRY TRAINING GUIDE. December 1, 2003 UNIVERSITIES OCCUPATIONAL SAFETY AND HEALTH EDUCATIONAL RESOURCE CENTER

Airways Resistance and Airflow through the Tracheobronchial Tree

Revised Protocol: Criteria for Designating Substances as. Occupational Asthmagens on the AOEC List of. Exposure Codes

GRADE 11F: Biology 3. UNIT 11FB.3 9 hours. Human gas exchange system and health. Resources. About this unit. Previous learning.

A Guide to Performing Quality Assured Diagnostic Spirometry

Understanding Pulmonary Function Testing. PFTs, Blood Gases and Oximetry Skinny Little Reference Guide

Medicare Pulmonary Rehabilitation (PR) Benefit Frequently Asked Questions June 2010 (Latest Updates: December 18, 2013 and February 12, 2014)

An Overview of Asthma - Diagnosis and Treatment

COPD. Information brochure for chronic obstructive pulmonary disease.

Pulmonary Rehabilitation and Respiratory Therapy Services in the Physician Office Setting* Sam Birnbaum, BBA, CMPE; and Brian Carlin, MD, FCCP

This clinical study synopsis is provided in line with Boehringer Ingelheim s Policy on Transparency and Publication of Clinical Study Data.

COPD and Asthma Differential Diagnosis

GCE AS/A level 1661/01A APPLIED SCIENCE UNIT 1. Pre-release Article for Examination in January 2010 JD*(A A)

Outcome of Drug Counseling of Outpatients in Chronic Obstructive Pulmonary Disease Clinic at Thawangpha Hospital

D-Facts. Vitamin D Competitive Evaluation

USING CLSI GUIDELINES TO PERFORM METHOD EVALUATION STUDIES IN YOUR LABORATORY

"Statistical methods are objective methods by which group trends are abstracted from observations on many separate individuals." 1

Tests. Pulmonary Functions

Impact of symptoms of anxiety and depression on COPD Assessment Test (CAT) scores

A comparison of blood glucose meters in Australia

6. MEASURING EFFECTS OVERVIEW CHOOSE APPROPRIATE METRICS

LUNG VOLUMES AND CAPACITIES

The economic burden of lung disease in Europe. Online supplement.

How To Qualify For EHR Stimulus Funds Under

Abnormalities Consistent with Asbestos-Related Disease Among Long-Term Demolition Workers

QUALITY CONTROL IN LUNG FUNCTION

Best Practices in Managing Patients With Chronic Obstructive Pulmonary Disease (COPD)

Chronic obstructive pulmonary

Lab #11: Respiratory Physiology

X X X a) perfect linear correlation b) no correlation c) positive correlation (r = 1) (r = 0) (0 < r < 1)

Validation and Calibration. Definitions and Terminology

ZEPHYRLIFE REMOTE PATIENT MONITORING REIMBURSEMENT REFERENCE GUIDE

CARDIOVIT AT lead display. The high-performance, multifunctional ECG wizard that meets highest demands and offers excellent value for money!

2.06 Understand the functions and disorders of the respiratory system

RESPIRATORY THERAPIST CLASSIFICATION SERIES

BEACON HEALTH STRATEGIES, LLC TELEHEALTH PROGRAM SPECIFICATION

Longitudinal Modeling of Lung Function in Respiratory Drug Development

Research Article Lung Function Tests in Sickle-Cell Patients in Benin City

National Learning Objectives for COPD Educators

Performance, ventilation, and oxygen consumption in three different types of exercise test in patients with

EUROPEAN LUNG FOUNDATION

Make your practice more efficient in a heartbeat.

This clinical study synopsis is provided in line with Boehringer Ingelheim s Policy on Transparency and Publication of Clinical Study Data.

The Radiologic Technology Program at Trenholm University

Medical Information Systems

epro Integration with Oracle Remote Data Capture

SARCOIDOSIS. Signs and symptoms associated with specific organ involvement can include the following:

Electronic patient diaries in clinical research

User Manual. User Manual Rev Issued on: 23/10/2009. Approved on: 23/10/2009. winspiropro user manual Rev 3.2 page 1 of 56

Coding Guidelines for Certain Respiratory Care Services July 2014

Maximal Flow at Functional Residual Capacity in Healthy Children From Birth to 7 Years, and Beyond

Current consensus guidelines, such as the Global. Office Spirometry Significantly Improves Early Detection of COPD in General Practice*

Accuracy Test Omnitest 3 Blood Glucose Monitoring System. B. Braun Diabetes Care

MIR Software Comparison Table. Multilanguage (18 Languages)

Flow Measurement with Respironics Flow Sensors

Pulmonary Rehabilitation in Chronic Obstructive Pulmonary Disease (COPD)

Measurement characteristics of two asthma symptom diary scales for use in clinical trials

Standardisation of spirometry

Value of Homecare: COPD and Long-Term Oxygen Therapy. A White Paper

Pre-lab homework Lab 6: Respiration and Gas exchange

From Computerized Patient Records to National Resource

Comparison of Different Exercise Tests in Assessing Outcomes of Pulmonary Rehabilitation

Patient and physician evaluation of the severity of acute asthma exacerbations

A Population Health Management Approach in the Home and Community-based Settings

Lung Function Increases With Increasing Level of Physical Activity in School Children

Bulletin Notifiable Occupational Diseases Information for Physicians

Time Course and Recovery of Exacerbations in Patients with Chronic Obstructive Pulmonary Disease

STATISTICAL METHODS FOR ASSESSING AGREEMENT BETWEEN TWO METHODS OF CLINICAL MEASUREMENT

Best Practices in Managing Patients With Chronic Obstructive Pulmonary Disease (COPD)

Life expectancy and years of life lost in chronic obstructive pulmonary disease: Findings from the NHANES III Follow-up Study

Original Contribution

Transcription:

Technical Note Respiration 2001;68:400 404 Received: October 6, 2000 Accepted after revision: February 24, 2001 Comparison of a New Desktop Spirometer (Diagnosa ) with a Laboratory Spirometer D.M. Maree E.A. Videler M. Hallauer C.H. Pieper C.T. Bolliger Lung Unit, Tygerberg Hospital, Department of Internal Medicine, University of Stellenbosch, Cape Town, South Africa Key Words Flow-volume loop W Spirometry W Telemedicine Abstract Background: The Diagnosa is a fully integrated system, able to determine spirometry, ECG, blood pressure and body composition. Real time data can be transferred via Internet to a remote receiving center. Objectives: The aim of this study was to perform biological testing of the spirometry component in subjects with normal and pathological pulmonary function. Methods: A group of 45 patients (mean age 43.3 years, 30 males) was tested on both the Diagnosa and the standard Jaeger Masterlab spirometer according to the guidelines of the American Thoracic Society. Three subgroups of 15 subjects each (normal spirometry, obstructive and restrictive airflow limitation) were selected. Results: All measurements performed with the Diagnosa (FVC, FIVC, FEV 1, PEF, FEF 25, FEF 50, FEF 75 ) correlated closely (r = 0.92 0.99) with those performed with the Jaeger spirometer and showed good limits of agreement (the largest difference between the two devices being 0.2 liter for FEV 1 ). Analysis of the 3 subgroups showed no difference for any parameters compared to the overall group. Electronic This study was supported by Cyberscope; M. Hallauer is employed by Cyberscope. transfer of all data was successful. Conclusions: The Diagnosa spirometer is comparable to a standard laboratory spirometer and can be used reliably for telemedicine purposes. Introduction Copyright 2001 S. Karger AG, Basel With the restraints of space in general practitioners rooms and industrial health clinics, as well as in epidemiology study laboratories, it would be ideal to have some important measurement devices incorporated in one casing. This has been implemented by a new system (Diagnosa; Cyberscope, Cape Town, South Africa) which contains 4 modules for separate measurements of spirometry, electrocardiography, body composition and blood pressure (fig. 1). Its software allows for secure electronic transmission of recorded results, which is ideal for use in telemedicine [1 3]. Many new spirometers are currently being developed and marketed worldwide, and hospital-based pulmonary function laboratories are ideal for biological testing of these apparatuses as these laboratories routinely encounter various types of airflow limitations. The aim of this study was to compare the spirometry component of the Diagnosa to a standard laboratory spirometer (Masterlab 4.0; Jaeger AG, Würzburg, Germany) in a group of patients exhibiting the whole range of flow limitations. ABC Fax + 41 61 306 12 34 E-Mail karger@karger.ch www.karger.com 2001 S. Karger AG, Basel 0025 7931/01/0684 0400$17.50/0 Accessible online at: www.karger.com/journals/res D.M. Maree A5 Laboratory, Department of Internal Medicine Tygerberg Hospital, Private Bag X Parow 7505 (South Africa)

Fig. 1. A patient is shown performing a flow-volume loop maneuvre with the spirometer of the Diagnosa system. Materials and Methods Test Device The spirometry component of the Diagnosa records the flow-volume loop test; the data are digitized by means of a 12-bit analogueto-digital converter with a sampling speed of 200 Hz. The Diagnosa measures flow rates using an orifice plate pneumotachometer and the flow time curve is processed by a standard desktop computer. The following parameters are derived from the curve: forced inspiratory vital capacity (FIVC), FVC, FEV 1, FEV 1 /FVC, peak expiratory flow rate (PEF), peak inspiratory flow rate (PIF) and the forced expiratory flows at 25, 50 and 75% of FVC (FEF 25, FEF 50 and FEF 75, respectively). FEV 1 is determined by back extrapolation as recommended by the American Thoracic Society (ATS) [5]. The results of the best curve, after all ATS criteria [5] are met, are displayed as an absolute value corrected for BTPS [4] and as a percentage of predicted normal. The program of Diagnosa requests a full calibration every day before any flow time curves can be measured. The testing procedure consists of maneuvers at different flow rates with a calibrated 3-liter calibration syringe. The Diagnosa has an added feature of transmitting patient data via the Internet from the test site to the patient s general practitioner, treating physician or a respiratory consultant and confidentiality is maintained by removing the patient s identity. Patients Subjects eligible for this study were routine patients from the respiratory out-patient clinic of Tygerberg Hospital, where routine pulmonary function testing is performed. The device was tested on 45 subjects. These subjects were subdivided into three groups of 15 subjects each (normal spirometry: mean age 35.1 years, range 19 61, 9 males; obstructive airflow limitation (stratified as follows: 9 with chronic obstructive airway disease and 6 with bronchial asthma): mean age 44.7 years, range 19 70, 11 males; restrictive airflow limitation: mean age 50.2 years, range 28 80, 10 males). Subjects were only drafted into the project if they had not previously had any flowvolume determination on any of the two test apparatuses, this eliminated any bias in favor of one of the apparatuses. Trial Design The Diagnosa was compared to a worldwide commercially available standard spirometer, the Jaeger Masterlab 4.0. The Diagnosa used in this study was a standard model with all the extra attachments for the other components of the system. Respiratory technologists tested the subjects according to the guidelines of the ATS [5]. To achieve a balanced design, each subject performed 3 acceptable maneuvers alternatively on each device, making up a total of 6 maneuvers. Each subject randomly selected which spirometer he or she wanted to use first, thereby allowing the learning effect to be evenly distributed between both instruments. All tests had to be sent to the manufacturer via the Internet; the manufacturer immediately returned the results by facsimile, enabling the investigator to verify the accuracy of the system. Comparison of Spirometers Respiration 2001;68:400 404 401

Fig. 2. Correlation between Diagnosa and Jaeger for: FCV (r = 0.99), FEV1 (r = 0.99), FIVC (r = 0.99) and PEF (r = 0.98) of all 45 subjects. Regression lines are shown. Statistical Analysis The curves from both devices were assessed for acceptability and reproducibility according to the standards established by the ATS. The best value for the following parameters from all acceptable curves for both devices were compared: FVC, FEV 1, FIVC and PEF. FEF 25, FEF 50 and FEF 75 were also compared but from the test with the highest sum of FVC and FEV 1, on each device. In order to test whether the measurements of the Diagnosa correlated with those of the Jaeger, the correlation coefficient (r) between the data obtained with both apparatuses was calculated. A significance level of!5% was chosen. The mean of the differences was analyzed using the paired t test. To further elucidate whether the measurements of both spirometers could be used interchangeably, the mean of the differences between the indices measured by both devices was calculated as described by Bland and Altman [6]. Results The Diagnosa was stable with respect to the accuracy of the measurements throughout the day after calibration of the system. All tests recorded met ATS acceptability criteria for both devices. The ATS reproducibility criteria (FEV 1 and FVC of the two best curves of each device differing by!5% or!200 ml) were met by 37 (82%) of the subjects on the Jaeger and by 39 (87%) subjects on the Diagnosa. The Diagnosa correlated very well with the Jaeger (r values for all parameters 0.92 0.99) (table 1) for the following parameters: FVC, FEV 1, FIVC, PEF, FEF 25, FEF 50 and FEF 75. Figure 2 shows the correlation between the Diagnosa and the Jaeger measurements for FVC, FEV 1, FIVC and PEF for all subjects. Figure 3 shows the plots of the differences between the readings of the two spirometers against their mean, as described by Bland and Altman [6], for FVC, FEV 1, FIVC and PEF. Analysis of the different subgroups (normal, obstructive and restrictive) for correlation and differences between the readings of the apparatuses did not show worse results for a particular type of airflow limitation in comparison to the overall results. Significant differences between the means were found for FVC, FEV 1, FEF 25 and FEF 75 using the paired t test, with all measurements reading higher with the Jaeger, with the exception of PEF and FEF 75 (table 1). All 45 subjects results were transmitted to the manufacturer via the Internet and immediately returned via facsimile; all 45 results were received without any inaccuracies. Discussion The results of our study show that the Diagnosa spirometer correlated very well with the standard spirometer for all measured parameters. Although the difference between the means (Diagnosa versus Jaeger) was significant for 4 parameters, they were small in absolute values, 402 Respiration 2001;68:400 404 Maree/Videler/Hallauer/Pieper/Bolliger

Fig. 3. Plots of the differences between Diagnosa and Jaeger against their mean for: FVC, FEV 1, FIVC and PEF. The mean difference ( ) and upper and lower limits of agreement (B 2 SD; WWWWWW) are indicated. Table 1. Comparison of measurements between Diagnosa and a standard spirometer (Jaeger Masterlab 4.0) Parameters Diagnosa Jaeger Correlation coefficient Differences 1 Limits of agreement FVC, liters 3.02 (1.10) 3.12 (1.40) 0.99 0.10* 0.12 0.32 FIVC liters 3.12 (1.02) 3.16 (1.10) 0.99 0.04 0.06 0.45 FEV 1 liters 2.17 (0.97) 2.37 (1.06) 0.99 0.20* 0.29 0.38 PEF, liters Ws 1 6.44 (2.56) 6.47 (2.72) 0.98 0.03 1.15 1.21 FEF 25, liters Ws 1 4.94 (2.47) 5.16 (2.73) 0.98 0.12* 0.87 1.30 FEF 50, liters Ws 1 2.73 (1.53) 2.82 (1.76) 0.96 0.09 0.92 1.11 FEF 75, liters Ws 1 0.99 (0.69) 0.87 (0.73) 0.92 0.12* 0.70 0.48 All data are presented as mean B SD (in parentheses). 1 Mean Diagnosa reading mean Jaeger reading. * p! 0.05 (paired t test). the largest being 0.2 liter for FEV 1. This difference is acceptable for clinical practice, all the more so that the real value of the measurement lies somewhere between the 2 measurements. Furthermore, analysis of the limits of agreement for all 7 parameters showed that only 2 out of 45 (4%) measurements lay outside 2 standard deviations, which is generally accepted as a criterion of good agreement. The Diagnosa was also tested for its reliability in electronically transmitting the test results via the Internet from the test site to the manufacturer s offices. Additionally, the capability of Diagnosa of storing and transmitting secure patient results is useful in clinical trials and epidemiological studies. In an industrial medicine setting or in epidemiological surveys, measurement of flow-volume loops on site is expected to produce superior data to the often-prescribed and unsupervised use of PEF measurements, which have been shown to be unreliable [7, 8]. Another spirometer with telemedicine features, which we tested previously, Comparison of Spirometers Respiration 2001;68:400 404 403

was different in its concept in that the device was handed out to individuals who would directly report to a remote receiving center [3, 9]. The Diagnosa as a multifunctional unit is designed for use in a general practice or industrial set-up. The data would be screened and secondarily transmitted by the practice personnel and not by the patient. In conclusion, the Diagnosa is deemed comparable to standard laboratory spirometry for the determination of forced flow-volume curves in clinical practice. The quality of the results can be assessed on location in real time and where required, expert advice can be sought immediately via the Internet. References 1 Abboud S, Bruderman I: Assessment of a new transtelephonic portable spirometer. Thorax 1996;51:407 410. 2 Bruderman I, Abboud S: Telespirometry; a novel system for home monitoring of asthmatic patients. Telemed J 1997;3:127 133. 3 Izbicki G, Abboud S, Jordan P, Perruchoud AP, Bolliger CT: A comparison of a new transtelephonic portable spirometer with a laboratory spirometer. Eur Respir J 1999;14:209 213. 4 Turney SZ, Blumenfeld W: Heated Fleisch pneumotachometer: A calibration procedure. J Appl Physiol 1973;34:117 121. 5 American Thoracic Society: Standardization of spirometery, 1994 update. Am J Respir Crit Care Med 1995;152:1107 1136. 6 Bland JM, Altman DG: Statistical methods for assessing agreement between two methods of clinical measurement. Lancet 1986;i:307 310. 7 Quirce S, Contreras G, Dybuncio A, Chan- Yeung M: Peak expiratory flow monitoring is not a reliable method for establishing the diagnosis of occupational asthma. Am J Respir Crit Care Med 1995;152:1100 1102. 8 Verschelden P, Cartier A, L Archevêque J, Trudeau C, Malo J-L: Compliance with and accuracy of daily self-assessment of peak expiratory flows (PEF) in asthmatic subjects over a three-month period. Eur Respir J 1996;9:880 885. 9 Bolliger CT, Wyser C, Perruchoud AP: Telemedicine in the postinterventional surveillance of patients with central airway obstruction. Eur Respir J 1998;12(suppl 28):266S. 404 Respiration 2001;68:400 404 Maree/Videler/Hallauer/Pieper/Bolliger