A METHOD OF PRECISE CALIBRATION FOR PIEZOELECTRICAL ACTUATORS



Similar documents
LOCATION DEPENDENCY OF POSITIONING ERROR IN A 3-AXES CNC MILLING MACHINE

Calibration of AFM with virtual standards; robust, versatile and accurate. Richard Koops VSL Dutch Metrology Institute Delft

Experiment 5. Lasers and laser mode structure

INTRODUCTION TO SCANNING TUNNELING MICROSCOPY

Nano Meter Stepping Drive of Surface Acoustic Wave Motor

Precision Miniature Load Cell. Models 8431, 8432 with Overload Protection

Preface Light Microscopy X-ray Diffraction Methods

Development of New Inkjet Head Applying MEMS Technology and Thin Film Actuator

STM and AFM Tutorial. Katie Mitchell January 20, 2010

Atomic Force Microscopy Observation and Characterization of a CD Stamper, Lycopodium Spores, and Step-Height Standard Diffraction Grating

Lecture 4 Scanning Probe Microscopy (SPM)

AFM-kit. Development of a kit for building from scratch an educational (but decently performing - i.e. usable) AFM.

ENS 07 Paris, France, 3-4 December 2007

Piezoelectric Scanners

Lecture 6 Scanning Tunneling Microscopy (STM) General components of STM; Tunneling current; Feedback system; Tip --- the probe.

Compact mobilized and low-cost scanning tunneling microscope for educational use

EXPERIMENT O-6. Michelson Interferometer. Abstract. References. Pre-Lab

Subminiature Load Cell Model 8417

5. Scanning Near-Field Optical Microscopy 5.1. Resolution of conventional optical microscopy

The Design and Characteristic Study of a 3-dimensional Piezoelectric Nano-positioner

Fabrication of Complex Circuit Using Electrochemical Micromachining on Printed Circuit Board (PCB)

It has long been a goal to achieve higher spatial resolution in optical imaging and

Radiant Dyes Laser Accessories GmbH

Calibration and Use of a Strain-Gage-Instrumented Beam: Density Determination and Weight-Flow-Rate Measurement

Department of Aerospace Engineering Indian Institute of Science Bangalore

Precision Work on the Human Eye

E_mail:

Analysis of Blind Microvias Forming Process in Multilayer Printed Circuit Boards

1 Introduction. 1.1 Historical Perspective

Precision Work on the Human Eye

Near-field scanning optical microscopy (SNOM)

MEASUREMENT OF END FACE GEOMETRY ON FIBER OPTIC TERMINI...2

FTIR Instrumentation

Experiment 5. Strain Gage Measurements

ENGINEERING METROLOGY

A Simple Fiber Optic displacement Sensor for Measurement of Light Intensity with the Displacement

METU DEPARTMENT OF METALLURGICAL AND MATERIALS ENGINEERING

Scanning Probe Microscopy

Vibration Suppression R&D at University of British Columbia. Tom Mattison UBC

Looking for the Origin of Power Laws in Electric Field Assisted Tunneling

ME 472 Engineering Metrology

International Journal of Engineering Research-Online A Peer Reviewed International Journal Articles available online

LINEAR ACTUATORS. Linear Actuators. Linear Actuators. Linear Actuators are Actuators that creates motion in a straight line, as contrasted

Non-Contact Vibration Measurement of Micro-Structures

NEAR FIELD OPTICAL MICROSCOPY AND SPECTROSCOPY WITH STM AND AFM PROBES

Basic principles and mechanisms of NSOM; Different scanning modes and systems of NSOM; General applications and advantages of NSOM.

4 SENSORS. Example. A force of 1 N is exerted on a PZT5A disc of diameter 10 mm and thickness 1 mm. The resulting mechanical stress is:

Experiment #5: Qualitative Absorption Spectroscopy

ATOMIC FORCE MICROSCOPY

Alternative Linear Motion Systems. Iron Core Linear Motors

A More Efficient Way to De-shelve 137 Ba +

Environmental Monitoring with Sensors: Hands-on Exercise

ELECTRON SPIN RESONANCE Last Revised: July 2007

Nanoscience Course Descriptions

INTERFERENCE OF SOUND WAVES

Self-Mixing Laser Diode Vibrometer with Wide Dynamic Range

SCANNING PROBE MICROSCOPY NANOS-E3 SCHOOL 29/09/2015 An introduction to surface microscopy probes

UNIVERSITY OF SOUTHAMPTON. Scanning Near-Field Optical Microscope Characterisation of Microstructured Optical Fibre Devices.

PCB Component Placement Inspection

E190Q Lecture 5 Autonomous Robot Navigation

Industrial Robotics. Training Objective

Atomic Force Microscope and Magnetic Force Microscope Background Information

7/3/2014. Introduction to Atomic Force Microscope. Introduction to Scanning Force Microscope. Invention of Atomic Force Microscope (AFM)

Bending, Forming and Flexing Printed Circuits

8.2 Elastic Strain Energy

Physics 41, Winter 1998 Lab 1 - The Current Balance. Theory

Usage of Carbon Nanotubes in Scanning Probe Microscopes as Probe. Keywords: Carbon Nanotube, Scanning Probe Microscope

- particle with kinetic energy E strikes a barrier with height U 0 > E and width L. - classically the particle cannot overcome the barrier

Micro-Power Generation

PHYS 222 Spring 2012 Final Exam. Closed books, notes, etc. No electronic device except a calculator.

Introduction to Beam. Area Moments of Inertia, Deflection, and Volumes of Beams

Fraunhofer Diffraction

Safakcan Tuncdemir 1, William M. Bradley *2. 1. Introduction

DualBeam Solutions for Electrical Nanoprobing

Surface Profilometry as a tool to Measure Thin Film Stress, A Practical Approach. Gianni Franceschinis, RIT MicroE Graduate Student

Microscopy. MICROSCOPY Light Electron Tunnelling Atomic Force RESOLVE: => INCREASE CONTRAST BIODIVERSITY I BIOL1051 MAJOR FUNCTIONS OF MICROSCOPES

ATOMIC ABSORTION SPECTROSCOPY: rev. 4/2011 ANALYSIS OF COPPER IN FOOD AND VITAMINS

Scanning Near-Field Optical Microscopy for Measuring Materials Properties at the Nanoscale

Tungsten Tip Preparation

Realization of a UV fisheye hyperspectral camera

CREOL, College of Optics & Photonics, University of Central Florida

Nanometer-scale imaging and metrology, nano-fabrication with the Orion Helium Ion Microscope

Optimising plate girder design

Drive an Atomic Force Microscope using LabVIEW software

Table of content. L81/RITA high speed Thermo Balance. Quattro Dilatometer. L75/1250/B/S Macro Dilatometer. New air cooled furnace program

Nanoscale Resolution Options for Optical Localization Techniques. C. Boit TU Berlin Chair of Semiconductor Devices

Plate waves in phononic crystals slabs

MEASUREMENTS OF THRUST LOAD IN CABLE SYSTEMS SEMI-RIGIDLY INSTALLED IN DUCT / MANHOLE STRUCTURES

Laser diffuse reflection light scanner with background suppression. Dimensioned drawing

SENSORS. Miniature Sensors - S3Z. Advanced line of miniature Asian style of photoelectric sensors mm background suppression

FIBER LASER STRAIN SENSOR DEVICE

Three-dimensional figure showing the operation of the CRT. The dotted line shows the path traversed by an example electron.

Lab 9: The Acousto-Optic Effect

Helium-Neon Laser. Figure 1: Diagram of optical and electrical components used in the HeNe laser experiment.

DESIGN AND EVALUATION OF PROBE WITH THREE DEGREE- OF-FREEDOM FOR NON-DESTRUCTIVE TEST USING THREE- DIMENSIONAL FINITE ELEMENT METHOD

Photonic Hydrophones based on Coated Fiber Bragg Gratings

Interference. Physics 102 Workshop #3. General Instructions

Design and Implementation of a 4-Bar linkage Gripper

Ferrostat Speed Sensor Series DSF Explosion Proof Versions EEx

Transcription:

Uludağ Üniversitesi Mühendislik-Mimarlık Fakültesi Dergisi, Cilt 9, Sayı, 24 A METHOD OF PRECISE CALIBRATION FOR PIEZOELECTRICAL ACTUATORS Timur CANEL * Yüksel BEKTÖRE ** Abstract: Piezoelectrical actuators are very useful for making translations over distance less than µm. One of the most important applications of such piezo actuators is the scanning tunnelling microscopy and the related scanning probe microscopies because atomic scale distances have to be measured and controlled in scanning probe microscopies applications. Especially, tube shaped piezoelectrical actuators have been widely used for such purposes. In this study, a piezotube scanner is calibrated at three stages. First calibration was done by using interferometer. At the other two stages Scanning Tunnelling Microscope was used for measurement. Keywords: STM, piezoelectric, Calibration, Interferometer. INTRODUCTION Piezoelectric actuators (PZTs) are used widely to produce small, precise mechanical motions (Deck, 995). Piezo actuators are becoming increasingly popular for generating small displacement in mechanical constructions. By applying a voltage or current to a piezo it extends or retracts due to the piezoelectric properties of the ceramic material (Holman et al, 995). Piezoelectrics are classified as ferroelectric materials. Their gain depends sensitively on the material bulk polarization. PZT responds to a voltage variation, which in turn changes the gain and hence the length (Deck, 995). The requirements of a good scanner are (Park and Barrett, 993):. High resolution, 2. Orthogonality (Movement along the three axes should be independent), 3. Linearity (The amount of movement should linearly be proportional to the applied voltage) 4. Mechanical rigidity (A rigid scanner will have a high resonant frequency, which is desirable for vibration isolation and feedback performance) 5. Wide range There are several types of piezos which can be used for different positions, such as required extension range, control voltage, temperature range, etc. (Holman et al, 995). Piezoplates and piezotubes are widely used. The principles of the piezoelectric behaviour also are the basis of operation of the tube scanner. In this geometry the inner and outer surface of a tube of piezoelectric material are coated with a thin metal electrode. The outside electrode is separated into four parts, which are electrically isolated from each other (Bonnell, 993). The single tube piezo element has become a mechanical scanner for use in scanning probe microscope instruments. Some advantages of these piezo tube scanners are high mechanical resonance frequencies, ability to scan in three dimensions with a single element and ease of manufacture and use (Porter, 994). * ** Kocaeli University, Faculty of Arts and Sciences, Department of Physics, İzmit, 43 Kocaeli-TURKEY. Uludag University, Faculty of Arts and Science, Department of Physics, Görükle, 659 Bursa-TURKEY. 47

+ - (a) + - (b) Figure. Sheet shaped piezoelectric material. (a) There is no applied potential difference between electrodes (b) A certain potential difference between electrodes is applied. Upper and lower side of the silver coated piezoelectric material are electrically isolated as shown in Figure.a. When a positive potential is applied to the upper side of the material with respect to the underneath, the upper side will be extended and the under surface will be compressed. As aresult as seen Figure.b, it will be bended downward. B y A C x D z Figure 2. Cross section view of piezotube. MEASUREMENTS Piezotube s silver coated outer electrode, which cross-section is shown in Figure 2, is split into two symmetrical parts along piezotube axis in order to get displacement in three dimensions. The inner part of the cylinder is connected to the ground. A potential difference should be applied to the inner face of the cylinder in order to get a certain amount of displacement. Displacement of piezotube as follows; when a positive potential difference is applied to the part A of piezotube, this part will be extended and the other portions will be remain constant. As a result, free end of the piezotube will be bended in the +x direction. If a negative potential is applied at the same time to the C-section of piezotube with respect to the inner section, the amount of bending will be increased. In order to get a displacement in z-direction, a positive potential difference should be apllied to all. Accordingly, length of the piezotube will be extended in all parts. The total piezotube s displacement by applying potential difference to the electrodes of piezotube will be measured by an interferometer. PREPERATION and CALIBRATION of PIEZOTUBE In order to get four different electrodes from the outer part of the piezotube, the etching process was used. The etching process was made using 4 ml of %63. HNO 3 diluted with 6 ml of distilled water. 48

The device was set up for coarse calibration of piezotube as shown in the Figure 3. Where, Laser (λ=632 nm), 2 Michelson Interferometer, 3 Screen and phototransistor, 4 Amplifier 5 PC Lab Card (AD/DA converter) (PCL 88) and PC 5 PCL 88 + PC 2 3 4 Figure 3. A device that is count fringe, which is used for calibration of piezotube. y B mirror x C A z D Laser beam Figure 4. Light mirror and cross-section view of piezotube. A mirror is placed at the free end of piezotubes shown in Figure 4. This mirror was used instead of movable mirror in Michelson interferometer. 3 Volt potential difference was applied to the part A of piezotube. When the potential difference was increased, the free end (mirror) was bended. Therefore, mirror moved away from the laser source. When mirror was moving away, action of fringes at screen was red by phototransistor. Then incoming signal was gained by an amplifier and send to a PC. A diagram of amplifier circuit is shown at Figure 5. Figure 5. Amplifier circuit which gains signal coming from phototransistor. 49

Same procedure is applied to the other parts of piezotube. An applied potential differences and amount of sliding of fringes are taken to a PC by using an AD/DA converter card. Graphs for each part are given at the Figure 6, 7, 8 and 9. In these graphs, the applied potential differences are represented in the horizontal axes and normalized numerical value of light intensity which is incoming to phototransistor are represented in the vertical axes. As seen in Figure 6, 7, 8 and 9, when 3 Volt potential differences are applied to the part A, B, C and D, four, three, four and three slides of fringe has seen respectively. He-Ne laser with 632 nm of wavelength is used in the Michelson interferometer. Therefore differences between two fringes is equal to 36 nm (half of wavelength) Normalized Intensity,9,8,7,6,5,4,3,2, 5 5 2 25 3 Potential differences (Volt) Normalized intensity,9,8,7,6,5,4,3,2, 5 5 2 25 3 Potential difference (Volt) Figure 6. Figure 7. For part A, the variation of normalised For part B, the variation of numerical value of normalised numerical value light intensity with applied potential difference. of light intensity with applied potential difference., Normalized intensity,9,8,7,6,5,4,3,2,, 5 5 2 25 3 Normalized intensity,9,8,7,6,5,4,3,2, 5 5 2 25 3 Potantial difference (Volt) Potantiel difference (Volt) 5 Figure 8. Figure 9. For part C, the variation of normalised For part D, the variation of normalised numerical value of light intensity numerical value of light intensity with applied potential difference. with applied potential difference. Average values of amount of differences between minima s of fringes are calculated from graphs. Then response of the piezotube to the applied voltage is found. Displacements per unit applied voltage are given below; 43.3 nm/v for part A, 3.29 nm/v for part B, 42.28 nm/v for part C, 33.57 nm/v for part D. Then, the applied potential differences between electrodes of piezotube are changed from to 3 Volts and response of the piezotube to the applied voltage is calculated; 43.22 nm/v for part A, 3.35 nm/v for part B, 42. nm/v for part C, 33.92 nm/v for part D.

According to these values, total displacements are given below: 85.23 nm/v for A and C electrodes, 6.9 nm/v for B and D electrodes. For measurement of amount of elongation for z-direction, a light mirror is placed in front of the piezotube. Then, to +3 volts potential difference is applied to all of electrodes. The amount of elongation is observed as 98.26 nm. So, The amount of elongation per unit applied voltage for z-direction is calculated as 6.687 nm/v. In order to study nanometer-scale devices, the initial and the important step is to fabricate regular nanometer-scale test structure.[gu 995] For the calibration of nanometer scale piezotube is used in Scanning Tuneling Microscope (STM) and image of the graphite structure in nanometer-scale is observed. Before the observation of graphite structure, test structure in micrometer scale is used for micrometer calibration. The STM image of the grating is given in the Figure. The Figure is arranged according to the data obtained and shown the Figure. µm 2µm 2µm 7µm Figure. Figure. Uncalibrated image of 2µm 2µm grating 2µm 2µm dimension s grating. From the Figure, new calibration results that are given below are obtained; 84.36 nm/v for electrodes A and C, 62.26 nm/v for electrodes B and D. And the final calibration is carried out by the graphite structure. A Simplicity reproducibility and reasonable theoretical understanding are compared a kind of test data obtained from surface for many STM investigations (Selloni et al,987). The STM image of graphite structure is given in the Figure 2. According to this figure; 75. nm/v for electrodes A and C, 55.4 nm/v for electrodes B and D. 5

.5 nm a a 2. nm Figure 2. STM image of graphite. DISCUSSION The STM is suitable for the determination of the voltage-displacement characteristics of the piezo scanners. We have determined a voltage-displacement characteristic and calculated actual displacements for different tube-shaped piezo actuators.the actual displacement depends significantly on the controlling regime. It has been shown, that the maximum displacement depends monotonously on the maximum of the applied voltage. In comparison to other calibration methods, the STM calibration is one of the precies method. The measurement of the voltage-displacement characteristics of tube piezo scanners for STM application show that the adjustment of the displacement depends on the scanning range. The proportionally constant between the displacement and the applied voltage has to be recalibrated to the actual total scanning voltage elongation. REFERENCES. MÜLLER, F. et al., 995. A Method for Contact-Free Calibration of Piezoelectrical Actuators, Rev. Sci. Instrum. 66 (9). 2. DECK, L., 995. Accurate, Repetitive Motion from Biased Piezoelectric Actuators, Rev. Sci. Instrum. 66 (5). 3. HOLMAN, A. E. et al., 995. Analysis of Piezo Actuators in Translation Constructions, Rev. Sci. Instrum. 66 (5). 4. PARK, S. and BARRETT, R. C., 993. Design Considerations for an STM System. Methods of Experimental Physics, Vol 27. 5. BONNELL, D, A. 993. Scanning Tunneling Microscopy and Spectroscopy. Chapter 2. 6. PORTER, T. L., 994. A Dual-Scan Piezo Element for Scanning Probe Microscopes. Rev. Sci. Instrum.65 (7). 7. SELLONI, A. et al., 988. Scanning Tunneling Spectroscopy of Graphite and Interlates, Physica Scripta. Vol. 38, 297-3. 52