Certificate of Analysis Standard Reference Material 2232

Similar documents
Certificate Standard Reference Material 1866b

FULL PAPER Standardization of PCM Characterization via DSC

RESULTS OF ICARUS 9 EXPERIMENTS RUN AT IMRA EUROPE

STANDARD CLEANING AND CALIBRATION PROCEDURE FOR TGA-50(H) AND TGA-51(H)

Table of content. L81/RITA high speed Thermo Balance. Quattro Dilatometer. L75/1250/B/S Macro Dilatometer. New air cooled furnace program

Standards Development For Differential Scanning Calorimetry

CALIBRATION OF A THERMISTOR THERMOMETER (version = fall 2001)

DETERMINATION OF THE HEAT STORAGE CAPACITY OF PCM AND PCM-OBJECTS AS A FUNCTION OF TEMPERATURE. E. Günther, S. Hiebler, H. Mehling

G8 GALILEO. Innovation with Integrity. High-End Melt-extraction Analyzer. Inert Gas Method

Phys222 W11 Quiz 1: Chapters Keys. Name:

DSC Differential Scanning Calorimeter

Measurement of Specific Heat Capacity Using Differential Scanning Calorimeter

Physical Properties of a Pure Substance, Water

E. K. A. ADVANCED PHYSICS LABORATORY PHYSICS 3081, 4051 NUCLEAR MAGNETIC RESONANCE

Precise temperature measurement above 1000 C using thermocouples

Carbon Dioxide and an Argon + Nitrogen Mixture. Measurement of C p /C v for Argon, Nitrogen, Stephen Lucas 05/11/10

LARGE MEASUREMENT RANGE MECHANICAL COMPARATOR FOR CALIBRATION OF LONG GAUGE BLOCKS

DETERMINATION OF TOTAL CARBON, TOTAL ORGANIC CARBON AND INORGANIC CARBON IN SEDIMENTS

16. Industrial Platinum Resistance Thermometers

Phase Transitions and Differential Scanning Calorimetry

IUCLID 5 COMPOSITION AND ANALYSIS GUIDANCE DOCUMENT: IRON ORES, AGGLOMERATES [EINECS NUMBER , CAS NUMBER ] IRON ORE PELLETS

Determination of the heat storage capacity of PCM and PCM objects as a function of temperature

Conductivity Sensor Calibrations to Meet Water Industry Requirements

GMP 11. Good Measurement Practice for Assignment and Adjustment of Calibration Intervals for Laboratory Standards

Chapter 4 and 5 solutions

Basic RTD Measurements. Basics of Resistance Temperature Detectors

Enthalpy of Combustion via Calorimetry

FXA Candidates should be able to : Define and apply the concept of specific heat capacity. Select and apply the equation : E = mcδθ

CALIBRATION PRINCIPLES

Conductivity measurement on pure water according to the recommendations of the USP Pharmacopoeia USP24-NF19

Method to Calculate Uncertainties in Measuring Shortwave Solar Irradiance Using Thermopile and Semiconductor Solar Radiometers

4 Thermomechanical Analysis (TMA)

Measurement of Charge-to-Mass (e/m) Ratio for the Electron

SHORT-TERM STABILITY TEST FOR URANIUM SOIL CANDIDATE A REFERENCE MATERIAL

This value, called the ionic product of water, Kw, is related to the equilibrium constant of water

SOLIDIFICATION. (a)formation of stable nuclei. Growth of a stable nucleus. (c) Grain structure

Experiment 6 ~ Joule Heating of a Resistor

Development of certified reference material of thin film for thermal diffusivity

Characterization of Electronic Materials Using Thermal Analysis

How to calibrate an RTD or Platinum Resistance Thermometer (PRT)

Product Information. Thermal Properties of Elvax Measured by Differential Scanning Calorimeter (DSC) Summary

Thermal Storage Unit Using the Triple Point of Hydrogen

Temperature Calibration; Depths of Immersion

DETERMINING THE ENTHALPY OF FORMATION OF CaCO 3

BET Surface Area Analysis of Nanoparticles

15. Liquid-in-Glass Thermometry

Errors Due to Shared Leadwires in Parallel Strain Gage Circuits

THE WATER TRIPLE POINT CELL AN OPTIMAL REALIZATION

TIE-31: Mechanical and thermal properties of optical glass

Schedule of Accreditation issued by United Kingdom Accreditation Service High Street, Feltham, Middlesex, TW13 4UN, UK

Determination of Total Suspended Solids (TSS) and Total Volatile Solids (TVS) in Waters of Fresh/Estuarine/Coastal Waters

HOW ACCURATE ARE THOSE THERMOCOUPLES?

FP628. Nitrogen / Protein Determination by Combustion. Rev 1-6/8/11

Calibration of Dallas sensors

CHEM 105 HOUR EXAM III 28-OCT-99. = -163 kj/mole determine H f 0 for Ni(CO) 4 (g) = -260 kj/mole determine H f 0 for Cr(CO) 6 (g)

Analyses on copper samples from Micans

Concept for an Algorithm Testing and Evaluation Program at NIST

ORGANIC LABORATORY TECHNIQUES NEVER distill the distillation flask to dryness as there is a risk of explosion and fire.

Quality Assurance/Quality Control in Acid Deposition Monitoring

Analytical Testing Services Commercial Price List ManTech International Corporation January 2016

UNITED STATES CONSUMER PRODUCT SAFETY COMMISSION DIRECTORATE FOR LABORATORY SCIENCES DIVISION OF CHEMISTRY 5 RESEARCH PLACE ROCKVILLE, MD 20850

Effect of Magnesium Oxide Content on Final Slag Fluidity of Blast Furnace

Temperature Accuracy of Thermistors and RTDs Application Notes

Lecture 13: The Fictive and Glass Transition Temperatures

ELECTRON SPIN RESONANCE Last Revised: July 2007

EASIDEW PORTABLE HYGROMETER INSTALLATION, OPERATION AND MAINTENANCE MANUAL

Schedule of Accreditation issued by United Kingdom Accreditation Service 2 Pine Trees, Chertsey Lane, Staines-upon-Thames, TW18 3HR, UK

Phase coherency of CDMA caller location processing based on TCXO frequency reference with intermittent GPS correction

Differential Scanning Calorimetry DSC

THERMAL ANEMOMETRY ELECTRONICS, SOFTWARE AND ACCESSORIES

Related concepts Kohlrausch s law, equivalent conductivity, temperature dependence of conductivity, Ostwald s dilution law.

TRACEABILITY. Commercially available single-element solutions are used. of Single-Element Calibration Solutions

Chemistry 101 Generating Hydrogen Gas

Bomb Calorimetry. Example 4. Energy and Enthalpy

How do single crystals differ from polycrystalline samples? Why would one go to the effort of growing a single crystal?

By Authority Of THE UNITED STATES OF AMERICA Legally Binding Document

Answer, Key Homework 6 David McIntyre 1

MUHAMMAD ABDULLAH AL-AZZAZ COMPANY LIMITED P.O. BOX AL-KHOBAR KINGDOM OF SAUDI ARABIA

* This work is an official contribution of the National Institute of Standards and Technology and

Test 5 Review questions. 1. As ice cools from 273 K to 263 K, the average kinetic energy of its molecules will

Calibrating DC Shunts: Techniques and Uncertainties

The Analytical Balance

Reprint (R22) Avoiding Errors in UV Radiation Measurements. By Thomas C. Larason July Reprinted from Photonics Spectra, Laurin Publishing

Temp-485-Pt100. A temperature sensor (Pt100 or Pt1000) communicating over the RS-485 bus with a simple communication protocol

limit to percentage levels, a limited working range translates

TA INSTRUMENTS DIFFERENTIAL SCANNING CALORIMETER (DSC) Insert Nickname Here. Operating Instructions

Measurement of Capacitance

Low-Level Contact Resistance Characterization

Objectives 200 CHAPTER 4 RESISTANCE

Thermal diffusivity and conductivity - an introduction to theory and practice

Measuring Temperature withthermistors a Tutorial David Potter

Content Sheet 7-1: Overview of Quality Control for Quantitative Tests

Coating Thickness and Composition Analysis by Micro-EDXRF

EXPERIMENT 1 (Organic Chemistry I)

4-Channel Thermometer / Datalogger

Effective: September 10, 2006 Vermont Attachment 1 to Rule Public Service Board Page 1 of 6

Chapter 8 Maxwell relations and measurable properties

KINETIC THEORY AND THERMODYNAMICS

vap H = RT 1T 2 = kj mol kpa = 341 K

Transcription:

National Institute of Standards & Technology Certificate of Analysis Standard Reference Material 2232 Indium DSC Calibration Standard Temperature and Enthalpy of Fusion This Standard Reference Material (SRM) is intended for use in the temperature and enthalpy of fusion calibrations of differential thermal analyzers (DTA), differential scanning calorimeters (DSC), and similar instruments. This SRM consists of a 1 g piece of indium metal sealed in an argon atmosphere in a mylar bag. Certified Temperature of Fusion: The certified value of 156.598 5 ºC is the temperature assigned to the melting point of pure indium [1]. The melting point is realized as the plateau temperature of the melting curve of the slowly melted, high purity indium. Temperature of Fusion: 156.598 5 ºC ± 0.000 34 ºC Based on samples tested, the temperature range of melting of bulk material is not expected to exceed 0.000 3 ºC. Temperatures of melting curve plateaus for samples of this material are expected to differ by not more than 0.000 1 ºC from each other and by not more than 0.000 34 ºC from the assigned temperature. The basis for assigning an expanded uncertainty (k = 2) of 0.000 34 ºC to the temperature of fusion of SRM 2232 is described in reference 2. Certified Enthalpy of Fusion: The certified enthalpy of fusion value at the DSC onset temperature was determined by performing duplicate measurements on seven SRM 2232 DSC samples. The DSC onset temperature has an expanded uncertainty (k = 2) of 0.046 ºC. Enthalpy of Fusion: 28.51 J g 1 ± 0.19 J g 1 The uncertainty was calculated as two times the square root of the sum of three components of variance. Component one was the standard deviation of the mean from 14 measurements of the enthalpy of fusion of indium. Component two was the standard deviation of the mean of 14 measurements of the enthalpy of fusion of SRM 2220 Tin that were used to calibrate the energy scale. Component three was the variance of the enthalpy certification measurements for SRM 2220 Temperature and Enthalpy of Fusion Tin the calibration standard. The variance for SRM 2220 was calculated as the square of one-half the stated uncertainty on the certificate. The basis for assigning an expanded uncertainty (k = 2) of 0.19 J g 1 to the enthalpy of fusion of SRM 2232 is described in reference 3. Expiration of Certification: The certification of this SRM is valid indefinitely within the measurement uncertainties specified, provided the SRM is used in accordance with the Notice and Warnings to Users section of this certificate. However, the certification is nullified if the SRM is damaged or contaminated. Maintenance of SRM Certification: NIST will monitor this SRM over the period of its certification. If substantive technical changes occur that affect the certification before the expiration of this certificate, NIST will notify the purchaser. Registration (see attached sheet) will facilitate notification. Temperature studies on fixed-point cells were performed by G.F. Strouse of the NIST Process Measurements Division. DSC studies were performed by D.R. Kirklin of the NIST Physical and Chemical Properties Division. Support aspects involved in the issuance of this SRM were coordinated through the NIST Standard Reference Materials Program by B.S. MacDonald of the NIST Measurement Services Division. Gregory J. Rosasco, Chief Physical and Chemical Properties Division James R. Whetstone, Chief Process Measurements Division Gaithersburg, MD 20899 Certificate Issue Date: 21 July 2005 See Certificate Revision History on Last Page Robert L. Watters, Jr., Chief Measurement Services Division SRM 2232 Page 1 of 7

NOTICE AND WARNING TO USERS Handling of SRM: Any handling procedures of this high purity material are subject to introduce contamination. Every possible effort should be made to maintain the purity of this SRM. SOURCE AND ANALYSIS 1 Source of Material: The indium metal (Lot S2739) for this SRM was obtained from Arconium (Providence, RI). The indium is of high purity, with the total of all elements that affect the melting point temperature being less than 0.1 mg/kg. Temperature of Fusion Measurements: The thermal tests for the certification of this SRM were performed on a fixed-point cell prepared in a manner similar to that described in reference 4. The cell contains approximately 1190 g of indium obtained from the randomly-selected 1 g pieces of indium of lot S2739. The freezing points were prepared using the recommended induced inner freeze method. With the metal completely melted, the furnace was set at approximately 3 ºC below the freezing point temperature. After supercooling and recalescence had been observed with a 25.5 Ω standard platinum resistance thermometer (SPRT) in the cell, the thermometer was removed and two fused silica glass rods, each initially at room temperature, were inserted successively into the well for about three minutes each to induce freezing of a thin mantle of solid metal around the well. The thermometer was then reinserted into the cell and the recording of readings was begun. After equilibrium was established, the temperature of the plateau on the freezing curve was found to vary no more than 0.000 03 ºC during the first 50 % of the duration of the freeze. A typical freezing curve obtained under such conditions is shown in Figure 1 (the region of supercooling and recalescence is not shown, as the curve begins after the reinsertion of the thermometer); a sample of the data is plotted at greater resolution in Figure 2. After the metal was slowly and completely frozen in the above manner, the furnace was set at approximately 1 ºC above the freezing point temperature to slowly melt over a time period of approximately 10 hours. Thermometer readings were recorded continuously until the melting was complete. A typical melting curve obtained under such conditions is shown in Figure 3. Following the freezing and melting curve measurements, the plateau temperature of a freezing curve of the test cell was compared directly with that of the standard indium freezing point cell of the Platinum Resistance Thermometer Calibration Laboratory, using a 25.5 Ω SPRT. The method of direct comparison is described in detail in reference 5. During the freezing and melting curve measurements, an environment of inert argon gas at 101 325 Pa was maintained in the cells. The electronic measurement equipment included an ASL F18 resistance ratio bridge, operating at a frequency of 30 Hz, and temperature-controlled Tinsley 5685A 100 Ω reference resistor. This reference resistor was maintained at a temperature of 25.00 ºC ± 0.010 ºC. Freezing curve and melting curve measurements were made with an excitation current of 1 ma. Direct comparison measurements were made using a 25.5 Ω SPRT at low exitation currents of l ma and 2 ma to allow for the analysis of the results at zero-power dissipation. A computercontrolled data acquisition system was used to acquire the ASL F18 bridge readings through the use of an IEEE-488 bus. Enthalpy of Fusion Measurements: Seven indium samples, 5 mg to 15 mg each, cut from a l g ingot of SRM 2232 and seven tin samples, 9 mg to 14 mg each, cut from a 6.25 cm 2 sheet of SRM 2220 were used in the DSC measurements, which were performed on a Perkin Elmer DSC 7 Differential Scanning Calorimeter. The temperatures of fusion for tin and indium were used to calibrate the temperature axis. The enthalpy of fusion of tin was used to calibrate the measured energy (measured area). After linearization of the temperature axis, the temperatures and enthalpies of fusion were measured for the set of seven SRM 2220 tin samples used in the calibration of the instrument. Duplicate measurements were made on seven SRM 2232 indium samples to determine the certified values and the observed variability due to the instrument and sample configuration. A typical DSC melting curve of the indium is shown in Figure 4. 1 Certain commercial equipment, instruments or materials are identified in this certificate to adequately specify the experimental procedure. Such identification does not imply recommendation or endorsement by the National Institute of Standards and Technology, nor does it imply that the materials or equipment identified are necessarily the best available for the purpose. SRM 2232 Page 2 of 7

0.5 Change in fixed point cell temperature, mºc 0.0-0.5-1.0-1.5 freeze 1, 14 hours freeze 2, 13 hours freeze 3, 10 hours -2.0 0.0 0.1 0.2 0.3 0.4 0.5 0.6 0.7 0.8 0.9 1.0 Fraction of sample frozen in In 96-6 Figure 1. The freezing curves of SRM 2232 Indium DSC Calibration Standard using the induced inner freeze preparation technique. SRM 2232 Page 3 of 7

0.25 Change in fixed point cell temperature, mºc 0.00-0.25-0.50-0.75 freeze 1, 14 hours freeze 2, 13 hours freeze 3, 10 hours -1.00 0.0 0.1 0.2 0.3 0.4 0.5 0.6 0.7 0.8 0.9 1.0 Fraction of sample frozen in In 96-6 Figure 2. The freezing plateau regions of Figure 1 at greater resolution. SRM 2232 Page 4 of 7

4 Change in fixed point cell temperature, mºc 3 2 1 0-1 -2-3 melt 1, 10 hours melt 2, 12 hours melt 3, 9 hours -4 0.0 0.1 0.2 0.3 0.4 0.5 0.6 0.7 0.8 0.9 1.0 Fraction of sample melted in In 96-6 Figure 3. Three melting curves of SRM 2232 Indium DSC Calibration Standard following a slow freeze. Each melt followed the respective slow freeze of Figure 1. SRM 2232 Page 5 of 7

Figure 4. Typical DSC melting curve for SRM 2232 Indium DSC Calibration Standard. SRM 2232 Page 6 of 7

REFERENCES [1] Preston-Thomas, H.; The International Temperature Scale of 1990 (ITS-90); Metrologia, Vol. 27, pp. 3 10, p. 107 (1990). [2] Strouse, G.F.; Tew, W.L.; Assessment of Uncertainties of Calibration of Resistance Thermometers at the National Institute of Standards and Technology; NISTIR 5319; National Institute of Standards and Technology, U.S. Department of Commerce: Gaithersburg, MD, p. 16 (1994). [3] ISO; Guide to the Expression of Uncertainty in Measurement; ISBN 92-67-10188-9, 1st ed.; International Organization for Standardization: Geneva, Switzerland (1993); see also Taylor, B.N.; Kuyatt, C.E.; Guidelines for Evaluating and Expressing the Uncertainty of NIST Measurement Results; NIST Technical Note 1297; U.S. Government Printing Office: Washington, DC (1994); available at http://physics.nist.gov/pubs/. [4] Furukawa, G.T; Riddle, J.L; Bigge, W.R.; Pfeiffer, E.R.; Standard Reference Materials: Application of Some Metal SRMs as Thermometric Fixed Points; NBS Special Publication 260 77, U.S. Government Printing Office: Washington, DC, p. 140 (1982). [5] Mangum, B.W.; Pfeiffer, E.R.; Strouse, G.F.; (NIST); Valencia-Rodriguez, J.; (CENAM); Lin, J.H.; Yeh, T.I.; (CMS/ITRI); Marcarino, P.; Dematteis, R.; (IMGC); Liu, Y.; Zhao, Q.; (NIM); Ince, A.T.; Gakirolu, F.; (UME); Nubbemeyer, H.G.; Jung, H.J.; (PTB); Intercomparisons of Some NIST Fixed-Point Cells with Similar Cells of Some Other Standards Laboratories; Metrologia, Vol. 33, pp. 215 225 (1996). Certificate Revision History: 21 July 2005 (This revision reflects editorial changes); 01 March 1999 (Original certificate date). Users of this SRM should ensure that the certificate in their possession is current. This can be accomplished by contacting the SRM Program at: telephone (301) 975-6776; fax (301) 926-4751; e-mail srminfo@nist.gov; or via the Internet at http://www.nist.gov/srm. SRM 2232 Page 7 of 7