Establishing Quality Control Means and Standard Deviations for Hematology Instrumentation



Similar documents
Calibration Verification

STATS INTERLABORATORY QUALITY CONTROL PROGRAM

How to Verify Performance Specifications

Urinalysis Compliance Tools. POCC Webinar January 19, 2011 Dr. Susan Selgren

Content Sheet 7-1: Overview of Quality Control for Quantitative Tests

Individualized Quality Control Plan (IQCP) Frequently Asked Questions Date: May 5, 2015 (last updated 05/13/2016)

Quality Control of the Future: Risk Management and Individual Quality Control Plans (IQCPs)

XV.Quality Assurance in the Blood Bank

A Roadmap for Competency Assessment ASCLS-NJ Spring Seminar April 24, 2014

USING CLSI GUIDELINES TO PERFORM METHOD EVALUATION STUDIES IN YOUR LABORATORY

SureStep Pro Linearity Test Kit and Analytical Measurement Range Verification (AMR)

Validation and Calibration. Definitions and Terminology

2010 Laboratory Accreditation Program Audioconference. Accreditation Requirements for Waived Vs. Non-waived Tests

Clinical Laboratory Improvement Amendments (CLIA) How to Obtain a CLIA Certificate

Document Review: Centers for Medicare and Medical Service (CMS) What Do I Need to Do to Assess Personnel Competency?

Standards for Laboratory Accreditation

Laboratory Director Responsibilities

A Practical Guide to Internal Quality Control (IQC) for Quantitative Tests in Medical Laboratories

Repair Station Training Suspected Unapproved Parts

Prepublication Requirements

Procedure: Quality Assurance Policy Version 5 Quality Assessment Policy Version 3

Evaluating Customer Loyalty Solutions

PROCEDURES FOR HANDLING OOS RESULTS

Automation is Here. Now How Do You Handle All the Data? Swedish Covenant Hospital Lab. Life IS Good

Individualized Quality Control Plan

Bio-Rad Laboratories. QC data management solutions. Introduce Your Laboratory to a Whole New World of Unity Data Management Solutions

Quality Assurance for the Clinical Laboratory

Enterprise-level modeling solutions

Method Validation/Verification. CAP/CLIA regulated methods at Texas Department of State Health Services Laboratory

Medical Cannabis Laboratory Approval Program

CAPABILITY 12: Public Health Laboratory Testing

TEST METHOD VERIFICATION AND VALIDATION

How To Make A Global Cad/Cam Solution For A Large Company

8.0. Florida Department of Education. Contracted Services

The Importance of Following the PROTOCOL in Clinical Trials

WHITE PAPER. How to Launch a Data Governance Program: Practical Guidelines for Technology Implementers

Evaluating Laboratory Data. Tom Frick Environmental Assessment Section Bureau of Laboratories

USAGE AND PUBLIC REPORTING GUIDELINES FOR THE GREEN GRID S INFRASTRUCTURE METRICS (PUE/DCIE)

SAMPLE. Verification of Comparability of Patient Results Within One Health Care System; Approved Guideline (Interim Revision)

FASB s new qualitative goodwill impairment assessment Implications and opportunities

The Role of the Laboratory Medical Director: Responsibilities, Expectations

Ecology Quality Assurance Glossary

An Introduction to. Metrics. used during. Software Development

Below are some of the ways pathologists contribute to the six domains (listed in the order they appear on page 2 of the Strategy):

National Competency Standards for the Nurse Practitioner

Financial and Regulatory Reporting. Five Stages of Evolution The Critical Role of Financial Models

Calibration & Preventative Maintenance. Sally Wolfgang Manager, Quality Operations Merck & Co., Inc.

Answers to Your Questions About Blood Bank Proficiency, Competency and QC Follow Up to April 28, 2016 Webinar

Content Sheet 16-1: Introduction to Documents & Records

QUALITY CONTROL (QC) INFORMATION AND TROUBLESHOOTING GUIDE HEMATOLOGY

Control Charts and Trend Analysis for ISO Speakers: New York State Food Laboratory s Quality Assurance Team

Development and Validation of In Vitro Diagnostic Tests. YC Lee, Ph.D. CEO

FMEA and FTA Analysis

Level 1 Hematology Training Module For AcT Diff and AcT Diff 2

Laboratory Computer Systems

What is Contract Lifecycle Management Software?

DATASHEET. Xythos on Demand. Productivity and collaboration tools

Best Practices for Maintaining Quality in Molecular Diagnostics Gyorgy Abel, MD, PhD

Six Sigma Metric Analysis for Analytical Testing Processes

Chapter 2: Quality Assurance and Legal Issues

ADMINISTRATIVE MANUAL Policy and Procedure

NABL accreditation is a formal recognition of the technical competence of a medical testing laboratory

CLIA & Individualized Quality Control Plan (IQCP)

Equipment Qualification

LOAD & GO and make flow cytometry simplicity itself. the full spectrum of Flow Cytometry

ISO/IEC QUALITY MANUAL

ANCILLARY PROVIDER APPLICATION FOR PARTICIPATION PHYSICIANS HEALTH PLAN PO Box 30377, Lansing, MI

Document and Record Control in a Hospital Clinical Laboratory

The Re-emergence of Collective Investment Trust Funds

2009 LAP Audioconference Series. Laboratory Computer Systems

Unaccredited Point-of-Care Laboratory Testing Guideline for Physicians

Quality Requirements of a Point of Care Testing Service

Contents. Abstract...i. Committee Membership... iii. Foreword... vii. 1 Scope...1

Making Healthcare Easier: Benefits of an Integrated Electronic Medical Record and Practice Management System

IQCP GUIDELINES and TEMPLATE FOR GETTING STARTED

Medical Technologist Section Supervisor

College of American Pathologists Accreditation for Public Health Laboratories

JOINT COMMISSION INTERNATIONAL ACCREDITATION STANDARDS FOR. 2nd Edition

HIPAA and Network Security Curriculum

STANDARD OPERATING GUIDELINES: EVIDENTIAL BREATH ALCOHOL INSTRUMENT CALIBRATION

AN AUDIT QUESTIONNAIRE THAT EXAMINES SPECIFICALLY THE MANAGEMENT OF TECHNICAL ACTIVITIES CLAUSES IN ISO 15189

SAP integrates NX into Sophisticated Engineering

510(k) SUBSTANTIAL EQUIVALENCE DETERMINATION DECISION SUMMARY

Understanding Meaningful Use with a Focus on Testing the HL7 V2 Messaging Standards

Terms concerned with internal quality control procedures

USE OF REFERENCE MATERIALS IN THE LABORATORY

IAS CALIBRATION and TESTING LABORATORY ACCREDITATION PROGRAMS DEFINITIONS

Measurement Systems Correlation MSC for Suppliers

services NEW ZEALAND INSTITUTE OF ARCHITECTS INCORPORATED GUIDE TO ARCHITECTS SERVICES

information Records Management Checklist business people security preservation accountability Foreword Introduction Purpose of the checklist

CLSI TRAINING COURSES CATALOG

2. Contact Person: Garo Mimaryan, MS., RAC 7 Technical Regulatory Affairs Specialist III 3. Phone Number: (914)

Quality Monitoring Checklist

Why Disruptive Innovations Matter in Laboratory Diagnostics

MEDICAL PROTECTION SOCIETY PROFESSIONAL SUPPORT AND EXPERT ADVICE. Putting members first. Interested in a career as a Medicolegal Adviser with MPS?

Transcription:

Establishing Quality Control Means and Standard Deviations for Hematology Instrumentation For informational purposes only. WHO REQUIRES IT AND WHAT THEY REQUIRE Regulatory agencies require the use of Quality Control (QC) materials to assess the validity of results on patient specimens. Laboratories are required to ensure that the basic QC practices used in their facilities meet or exceed the regulations put forth by these agencies. Streck Technical Services technicalservices@streck.com

REGULATORY BODIES REQUIRE THE USE OF QUALITY CONTROL The Centers for Medicare & Medicaid Services (CMS) regulates all laboratory testing (except research) performed on humans in the U.S. through the Clinical Laboratory Improvement Amendments (CLIA). For laboratories subject to CLIA-88, for quantitative procedures, two control materials of different concentrations should be tested at least once each day that patient specimens are assayed. For each QC procedure employed, the laboratory must have appropriate QC ranges. CLIA deems it good laboratory practice for the individual laboratory to establish its own means and ranges. U.S.-based regulations state laboratories must establish their own means and ranges for their instruments. However, offi cials at the Division of Laboratory Services (the federal agency that administers the CLIA regulations) have stated, We are aware that for smaller laboratories, performing a limited volume of testing, this may not be practical, and they may use the manufacturer s suggested ranges, provided these are suitable for the methodology and instrument used by the laboratory. The CLIA regulation at 493.1256(d)(10) requires a laboratory to establish or verify the criteria for acceptability of all control materials. When control materials providing quantitative results are used, statistical parameters (for example, mean and standard deviation) for each batch or lot number of control materials must be defi ned and available. The laboratory may use the stated value of a commercially assayed control material provided the stated value is for the methodology and instrumentation employed by the laboratory and is verifi ed by the laboratory. The key phrase is smaller laboratories performing a limited volume of testing. These laboratories, which would be exempt from establishing their own means and ranges, would operate only on a limited basis. They would be running more QC samples than patient samples. The number of laboratories meeting these requirements would be small. For the majority of laboratories, establishing lab specifi c means and ranges is required. The published assay range is a range of means. The published assay means and ranges of a commercial hematology control are considered a range of means. The expected ranges listed represent estimates of variation due to different laboratories, instrument calibration, maintenance, reagents and operator technique. The laboratory s calculated means should fall within the ranges listed on the published assay. All QC results for an individual laboratory do not need to fall within the published assay range. International consensus is reached on the establishment of lab-specifi c means. In June 2010, the Clinical Laboratory and Standards Institute (CLSI), an international educational organization that promotes the development and use of standards and guidelines within healthcare communities, released document H26-A2, Validation, Verifi cation and Quality Assurance of Automated Hematology Analyzers, Approved Standard - Second Edition. This document reinforces the importance of establishing laboratory-specifi c quality control means and ranges. Establishing Quality Control Means and Standard Deviations for Hematology Instrumentation I 2

THE JOINT COMMISSION The Joint Commission on Accreditation of Health Care Facilities (JCAHO), in a Q&A item on their web site, stated: The standards require each laboratory to establish its own control ranges through repetitive testing. However, there is an allowance to use manufacturer ranges when the following conditions are met: 1. The stated values correspond to the method and instrument used by the laboratory, and 2. The mean obtained by the laboratory refl ects the manufacturer s stated mean, and 3. The Laboratory Medical Director assures the range is narrow enough to detect clinically signifi cant error. Determination of QC protocols is ultimately the responsibility of the Laboratory Medical Director. Manufacturer ranges may also be implemented if a test is used so infrequently that calculation of valid statistics is not possible. In settings where there is a high reproducibility (precise instrumentation, limited testing personnel), the laboratory s own calculated standard deviation (SD) may be small. When compared with the manufacturer ranges, a laboratory may fi nd that the range spans more than the commonly used +/-2 SD. Using the laboratory s calculated +/-2 SD may produce unnecessarily narrow ranges, causing the testing personnel to frequently repeat QC and investigate when the control performs outside the laboratory s range, but within the manufacturer s range. Alternatively, the full manufacturer range may be too broad to promote the detection of clinically signifi cant error. Selection of the appropriate range is a balance between these two ends of the spectrum. CAP REQUIREMENTS The College of American Pathologists (CAP) Hematology and Coagulation checklist states that at least two different analytic levels of control must be analyzed and recorded during each 24 hours of the analyzer s use. For each QC procedure employed, the laboratory must have appropriate QC ranges. For example, expected recovery ranges for commercial control materials are NOT the same as betweenrun SD ranges and are probably too wide for daily QC of a single instrument. The laboratory should calculate its own imprecision statistics for each instrument. The lab should also consider the following, per the survey checklist: HEMATOLOGY 20035 PHASE II Are tolerance limits (numeric and/or non-numeric) fully defi ned and documented for all hematology and coagulation control procedures? NOTE: The goal is to have scientifi cally valid, logical action limits for quality control procedures that promptly alert the technologist of the need for immediate evaluation of the particular assay, including initiation of corrective action, before release of patient results. HEM.25870 PHASE II If commercially ASSAYED controls are used for CBC instruments, do control values correspond to the methodology and have target values (mean and QC ranges) been verifi ed or established by the laboratory? Establishing Quality Control Means and Standard Deviations for Hematology Instrumentation I 3

NOTE: Most commercial controls have expected recovery ranges for each parameter, provided by the manufacturer. The mean of such ranges may not be the exact target value in a given laboratory. Each laboratory should assign its own initial target value, based on initial analysis of the material; this target value should fall within the recovery range supplied by the manufacturer, but need not exactly match the package insert mean. The laboratory should establish specifi c recovery ranges that accommodate known changes in product attributes, assuming that calibration status has not changed. ESTABLISHING MEANS AND STANDARD DEVIATIONS Streck supports and suggests CLSI Document C24-A3, Statistical Quality Control for Quantitative Measurement Procedures: Principles and Defi nitions, Approved Guideline - Third Edition, for establishing each laboratory s individual mean target values and ranges, and CLIA Title 42 part 493 subpart K, Quality Systems for Nonwaived Testing, Section 493.1256, Standard Control Procedures. CLSI document C24-A3, Section 8.6, Setting Control Limits, and Section 8.6.1, Values for the Mean and Standard Deviation, outline these procedures. Section 8.6.2, Assayed Control Materials, states: If assayed control materials are used, the values stated on the assay sheet provided by the manufacturer should be used only as guides in setting the initial control limits for testing new control materials. Actual values for the mean and standard deviation must be established by serial testing in the laboratory. The observed mean should fall within the range published by the manufacturer. External Quality Assessment (EQA) and peer-comparison programs provide useful measures of the means and SDs observed in other laboratories. Establishing laboratory-specifi c means and standard deviations is not a diffi cult process. Here is a brief procedure for establishing your own mean and Standard Deviation (SD): 1. Analyze the control a minimum of 20 times across several days. 2. Take the average of these runs. 3. This average should be within the range slated on the assay sheet and will be considered as the new mean. 4. Calculate a two Standard Deviation range from your results. 5. Incorporate this SD range around your new mean and monitor throughout the dating of the product. 6. The mean and SD values should be periodically recalculated during the life of the new lot. Most hematology instruments have quality control fi les that will calculate these means and ranges automatically. Peer comparison programs such as Streck s STATS are helpful in providing historical lab SD calculations. Establishing Quality Control Means and Standard Deviations for Hematology Instrumentation I 4

EVIDENCE-BASED CONTROL RANGES When control limits are properly calculated and implemented, the lab can be confi dent in its procedure to detect measurement errors at a high rate with a low false rejection rate, based on the performance characteristics of their particular method. The best response to an out-of-control result is to investigate the entire procedure to identify possible root cause(s), defi ne fail-safe solutions to eliminate the cause(s), and prevent the same problem from occurring in the future. The worst response would be to repeat the control until it lands within range as this becomes testing the controls. Once the lab has established an initial mean and range, as described in the previous section of this paper, it is important to re-evaluate the criteria after a full lot of data, and continue to evaluate on an ongoing basis and to ensure that the control ranges are appropriate for the particular test method. Applying too narrow limits or allencompassing SD ranges for all test methods may lead to false alarms, repeat control testing, which leads to waste of control and tech time and reduced lab productivity. Applying too broad ranges may lead to accepting too many results that really should be evaluated and to potentially missing errors. Evidence-Based Control Ranges are: Not assay sheet limits or stand-alone rules Built using performance goals, bias and precision of the test method Specifi c to the instrument or method, level of control, and parameter Designed to identify issues with an instrument or method The best Quality Control Rule approach to monitor performance defi nes the relevant clinical requirements for quality testing while integrating a high error detection rate and low false rejection rate. LOT-TO-LOT CORRELATIONS CLSI H26-A2 states that new lots of control material should be assayed in parallel with the current lot in use before the expiration of the current lot. Here is a brief procedure to transition to a new lot: 1. Set up new QC fi les for each control level of the new lot. 2. Verify the new lot by running each level of control three times in its respective fi le. Ensure that the mean values of the three runs fall within the ranges published on the manufacturer assay, without any regard to matching any published mean value. 3. Run each level twice a day for 3-5 days to calculate new mean values for each measurand. 4. Compare the calculated mean values for each level to the range specifi ed on the manufacturer assay sheet. 5. If the calculated mean is within range, enter it as the expected mean. Once the values are updated in the QC fi les, these fi les are used for the remainder of the dating period. Establishing Quality Control Means and Standard Deviations for Hematology Instrumentation I 5

OTHER QUALITY CONTROL SITUATIONS Some hematology parameters, such as MCV, will increase slightly over time. CAP gives guidance on how to accommodate this: Most commercial controls have expected recovery ranges for each parameter, provided by the manufacturer. The mean of such ranges within the recovery range supplied by the manufacturer, but need not exactly match the package insert mean. The laboratory should establish specific recovery ranges that accommodate known changes in product attributes, assuming that calibration status has not changed. Means of parameters that change over time may be handled differently. If it is known that the MCV rises 2 units over the life of the control, it is acceptable to raise the mean by half this change to accommodate the known rise. For example, if the initial mean is calculated at 84 and the historical 2 SD is 4, the lab could establish a mean of 85, with a range of 81-89. This will allow the values to start below the mean, rise through the mean and fi nish above the mean. The cumulative mean calculated at the end of the product life should be 85. QUALITY CONTROL FOR ANALYZERS WITH TWO SAMPLE MODES OR PATHWAYS Analyzers with multiple sample modes require additional QC. The following recommendation for quality control of two sample modes is taken from the JCAHO web site Q&A forum: Since there are two distinct sample pathways, QC is required for each sample mode according to the parameters established in the hematology standards. This would mean (for JCAHO purposes) performance of at least one control every eight hours of patient testing and performance of at least two levels of commercial controls every 24 hours of patient testing. The following are CAP requirements for analyzers using different pathways: Two levels of control at every 24 hours of patient testing for analyzers that perform body fl uid cell count using the same pathways as CBC testing. For analyzers using different pathways for CBC and body fl uid cell counts, two levels of control at every 24 hours of patient testing is required for each pathway, i.e., two sets of control (it is as if there are two separate analyzers). Establishing Quality Control Means and Standard Deviations for Hematology Instrumentation I 6

SUMMARY Establishing laboratory means and ranges does not need to be a tedious task. REFERENCES Regulatory agencies require the use of Quality Control (QC) materials to assess the validity of results on patient specimens. This is achieved by assaying and evaluating at least two levels of control material every 24 hours. Most laboratories are required to establish their own means and standard deviations for the parameters tested and reported on patients. The published assay range for a given control is the range in which a laboratory s mean must fall to be considered acceptable. The determination of mean and range does not have to be a diffi cult process. Most hematology analyzers and Laboratory Information Systems (LIS) have software to accomplish this task fairly effortlessly. 1. Clinical and Laboratory Standards Institute - www.clsi.org 2. Clinical Laboratory Improvement Amendments - www. cms.gov/clia 3. College of American Pathologists - www.cap.org 4. The Joint Commission - www.jointcommission.org Establishing Quality Control Means and Standard Deviations for Hematology Instrumentation I 7