KFF/KFG Low-Volume Flow Meter User Instructions



Similar documents
KOBOLD VKP FLOWMETER/SWITCH. User Instructions

FLR 1000 & 1000-BR Series Flow Sensors & Meters For Gases

How To Use A Flowmeter

Series DFMT Digital Paddlewheel Flowmeter

IFC 070 Technical Datasheet

Turbine Flow Sensor Series VTR

Submersion/Insertion Toroidal Sensor

POSITIVE DISPLACEMENT FLOWMETER - IM012P (PULSE)

VOLTAGE/CURRENT CALIBRATOR ISO-TECH ILC-421

ATL Fuel Level Sender Probes

Turbine Flowmeter Vision 1000 Flowrates 0,1-2,5L/min. Turbine Flowmeter Vision 2000 Flowrates 0,5-35L/min

CONTOIL DFM 8EDM. Table of contents. Mounting and operating instructions

Operating Instructions for Plastic Flow Meter. Model: KSK

CRITICAL MATERIALS HANDLING INVUE INTEGRATED FLOW CONTROLLERS NT6510, NT6520

Electro-Magnetic Flowmeters COPA-XT 2-Wire DT43

Instructions Manual. Electromagnetic sensor Series FLOMID FX. instrumentation for fluids. R-MI-FlomidFX Rev.: 0 English version

OEM-EP Pressure Controllers

Series 6000 Torque measured metal bellow coupling

T U R B I N E G A S M E T E R

MTI10 Insertion and MTL10 In-line Thermal Mass Flowmeter and Temperature Transmitter

FMA 3100/3100ST/3300/3300ST Series Thermal Mass Flow Sensors and Meters

Model 111 Mass Flow Instruments

BRAZED PLATE INSTALLATION MANUAL

Art.S001-S002 SOLAR MODULE WITH DELIVERY AND RETURN CONNECTIONS

Toroidal Conductivity Sensor

Sanitary and CIP Toroidal Sensor

Series 7500 Torque Sensor

ELECTROMAGNETIC FLOWMETER COMPACT

Duct Humidity Transmitter

Digital inductive conductivity transmitter

Non-invasive CLAMP-ON ultrasonic flowmeter for liquids

EASIDEW PORTABLE HYGROMETER INSTALLATION, OPERATION AND MAINTENANCE MANUAL

Description. Dimensions. Features. precision works better

Stealth Electronic Manual ZTM1010 Installation & Maintenance. For Oval Gear Meter Models:

Certified according to DIN EN ISO 9001 Technical Datasheet C-Flow Coriolis Mass Flow Meter

Electric Rotary Actuator - On/Off or programmable control actuator

Series 510 Submersible Level Transmitters

4 Posiflow Proportional Solenoid Valves

3½ DIGIT VOLTMETER MODULE LED- / LCD- SERIES

Flow Sensors. Convenient, Reliable and Economical Flow Measurement. THORNTON Leading Pure Water Analytics

Oil and Coolant Circulating Heating System. Model - OCSM

WE-350 Series ¼ Turn Electric Actuator

Wafer Check Valve ASAHI AV VALVES. Contents. User s Manual. (1) Be sure to read the following description of our product warranty 1

Electronic Flow Switches

SKP Series SKP15U.. Gas Valve Actuator with Safety Shutoff Function

BM SERIES WAVE COMPACT 6.3 GHZ CW SERIES PULSE RADAR LEVEL INSTRUMENTS (6.3 GHz)

Product data sheet. JUMO MIDAS Pressure Transmitter. Sales No. Order code JUMO AS /

Flow switches Series AD/VH

DSD 70 Dual Channel Hall Effect Speed Sensor for Railway Applications, compliant with EN 50155

Uniquely designed and engineered specifically for the pulp and paper industry.

For Multi-Parameter Meters see mvx

Ergo-Pro Single Line Solar Station Installation and Operating Instructions

Operating instructions

Submersion/Insertion Toroidal Sensor

Hydraulic Control Solutions

289 SERIES FLOW METER TRANSMITTERS

400B SERIES DIAPHRAGM PUMP

CORONA E / FLYPPER ELECTRONIC METER MULTI-JET

Certified according to DIN EN ISO 9001 Technical Datasheet KVO Series Vortex Flow Meter

Introduction to Process Control Actuators

DMP 331. Industrial Pressure Transmitter for Low Pressure. Stainless Steel Sensor

red-y smart series product information Thermal Mass Flow Meters and Controllers for Gases

OEM Manual MODEL 2350 ELECTRONIC DUAL CYLINDER SCALE

TMS TANK MANAGEMENT SYSTEM

Typical Aluminum GFM Mass Flow Meter

SWITCH! DESCRIPTION. ENVIRONMENTAL SPECIFICATIONS Operating Temperature: 0 C to +40 C SPECIFICATION SHEET:

Electronic Pressure Switch EDS 300

Mass Flow Meters & Controllers

Technical data. General specifications. Signal voltage V DC Signal duration. 1 s Input 2. Signal voltage. 1 s Analog output.

SunMaxx Solar Filling Station Operating Instructions

- CONSTRUCTIONS 300 SECTIONAL VIEW 300 MECHANICAL SEAL DIMENSIONS AND WEIGHT 400 PUMP 400 PACKING 401

Product Information. Gateway For Connecting EnDat Encoders to PROFIBUS-DP

KNC Model 3666 Automatic Pressure Calibration System

Electronic Pressure Switch EDS 3000

Models SFP33 & SFP32 Pressure Transmitters

PART 1 - INTRODUCTION...

VP23, 3-way proportional pressure control valves Seat valve with μp-driven pressure control

Fluid Level Gauge Fluid Level Sensor Temperature Switch FSA / FSK / TS

Physical Specifications (Custom design available.)

Dissolved Oxygen Sensors

HG600/HG800. Hydro-Power Unit. Installation and Operation Guide

DXPE*J /215 ED. DIRECtIONAL CONtROL VALVE PILOt OPERAtED, WItH OBE AND FEEDBACK. Q max (see performance table) SERIES 30

1115 4G SERIES GOVERNOR ma ANALOGUE DIGITAL SPEED SETTING

THE THERMAL FLOW METER, A GAS METER FOR ENERGY MEASUREMENT

K-Type Thermocouple Sensor User s Guide

2/ These vacuum switches are used in measuring ranges between -1 and 0 bar.

WaterMAG. Electromagnetic Flowmeter for Water Applications. Model MCB10A OVERVIEW APPLICATIONS FEATURES. No. SS2-MCB

PRODUCT PROFILE FEATURES

ABB Drives. User s Manual HTL Encoder Interface FEN-31

Grant Agreement No SFERA. Solar Facilities for the European Research Area SEVENTH FRAMEWORK PROGRAMME. Capacities Specific Programme

Modular I/O System Analog and Digital Interface Modules

COMBIMASS. Technical Data COMBIMASS eco-bio +

713 Open Channel Flowmeter and Transmitter. All-in-one package for open channel flow measurement.

Plastic tube flowmeters Series PT/PS

FIXED DISPLACEMENT HYDRAULIC VANE PUMPS BQ SERIES

Mono-Block II Installation & Operation Manual

V47 Series Temperature Actuated Modulating Water Valves

TECHNICAL DATASHEET #TD1404AX PWM CONTROLLED SOLENOID DRIVER

FIXED DISPLACEMENT HYDRAULIC VANE PUMPS BQ SERIES

MCR-VDC-UI-B-DC. Voltage Transducer for DC Voltages. INTERFACE Data Sheet _en_01. 1 Description

Transcription:

KFF/KFG Low-Volume Flow Meter User Instructions KFG_KFF_manual_0304

1. General Information Please read this service manual carefully before unpacking and readying the unit for operation, and follow the instructions precisely as described herein. These devices may only be installed, used and maintained by skilled personnel who are familiar with this service manual and can observe applicable regulations regarding industrial safety and accident prevention. 2. Contents 1. General Information... Page 2 2. Contents...Page 2 3. Suggested Application...Page 3 4. Operating Principle... Page 3 5. Unit Check-up...Page 4 6. Mechanical Connection...Page 4 7. Electrical Connection...Page 5 8. Electrical Commissioning...Page 6 9. Mechanical Commissioning...Page 6 10. Technical Data...Page 7 11. Dimensions... Page 8 Manufactured and Marketed by: Kobold Instruments Inc. 1801 Parkway View Drive Pittsburgh, PA 15205 Tel: 412-788-2830 Fax: 412-788-4890 Page 2 03/04

3. Suggested Application The KFF and KFG series flow meters are used for measuring small liquid or gas flows. These units are configured as follows: Analog output In order to transmit the measured flow results, an analog output (0-5 V) is provided. A 500 Hz. max. pulse output is available on some models. Flow sensor with analog output Integrated sensor with LCD digital display (3-digit) The KFF meters may only be used for measurement of low-viscosity, clear fluids (<10 cst). Measurement of liquids with higher viscosity can result in significant measurement errors. The KFG meters are calibrated at standard conditions in air (14.7 PSIG, 68 F). Notes for KFG-1 series: Non-conventional media, pressures or temperatures may require a user calibration on site, by means of a comparison unit or volumetric test. Attention! The media to be measured must not be contaminated. In particular, large particles can cause jamming or destruction of the rotor. 4. Operating Principle KOBOLD flow meter KFG and KFF are being used to measure small and very small flow volumes. The medium to be measured can be in gaseous form (KFG) or in liquid form (KFF). These units function according to Pelton principle, meaning that the flowing medium will set a paddle wheel in rotation. The rotary movement is translated into pulses via photo cells. An electronic integrator uses these pulses to generate a flow proportional, analog output signal (0-5 V). 03/04 Page 3

5. Unit Check-up These devices are checked before shipment and sent away in perfect condition. Should the damage to a device be visible, we recommend a thorough inspection of the delivery packing. In case of damage, please inform your parcel service/ forwarding agent immediately, since they are responsible for damages occuring during transit. Scope of delivery Standard scope of delivery applies: Flow meter Connection cable with plug: KF-7000 Instruction manual 6. Mechanical Connection Before Installation KFG: Please ensure that the actual flow volume is in agreement with the measuring range of the device. The reference number of measuring range can be read from the item number label. Note: Continuously exceeding the measuring range can cause damage to bearings. Make sure that the maximum operational pressure and temperature are not exceeded. Ensure that there is no remaining packing material inside the unit. The calibration of the unit is performed in horizontal position; the item label faces upwards. With other mounting positions, a measured-value deviation of 0.25 %/ C at 50% nominal flow should be taken into consideration. KFF/KFG: Mount the measuring device on desired position. Two mounting flanges are supplied for this purpose. Connect unit to your piping system. Do not remove the meter fittings as the calibration may change. Attention:When tightening fittings, it is necessary to work with two suitable wrenches. A transfer of torque on device's housing while tightening the fittings can result in the destruction of the measuring device. Check the sealing of all connections. Page 4 03/04

7. Electrical Connections Attention: Ensure that the voltage level of your supply system corresponds with the voltage level required by the measuring device. (+12.5 VDC ± 2 VDC) Make sure that the electrical supply lines are not active. Connect the ends of supplied connection cable KF-7000 with your system, according to the wiring diagram shown below. The green wire is used only if the unit has an optional pulse output. Attention: An incorrect connection can lead to damage of the unit electronics. Connect the cable plug with the corresponding socket to the measuring device. The plug and socket are indexed, so that an incorrect connection is impossible. KF-7000 Connection Cable Wiring 1 - Black (- V Power/Signal gnd.) 2 - White (0-5VDC Output) 3 - Red (+ V Power in) 1000 mm 6±1 40±3 03/04 Page 5

8. Electrical Commissioning KFF, KFG The unit is supplied ready for operation. The electronics are adjusted and calibrated to the sensor. By turning the potentiometer, located on the side of the electronics, the analog output (0-5 V) can be adjusted by using a reference meter if necessary. Page 6 03/04

9. Mechanical Commissioning To avoid pressure spikes, flow medium should enter the unit slowly. Attention: Pressure spikes generated by water hammer or caused by magnetic valves, ball valves or similar devices, can lead to destruction of the unit. During operation it must be ensured that the sensor is completely filled with flow media. Note: Large air bubbles in the measurement chamber can cause error indications and may destroy the bearings and impeller. 10. Technical Data Accuracy: Linearity: Repeatability: Mounting Position: Temperature Range: Maximum Pressure: Analog Output: Supply Voltage: ± 3 % F.S. ± 3 % F.S. KFF: ± 0.2 % F.S. KFG: ± 1 % F.S. Horizontal 32-120 F KFF: Ryton:100 PSI, Brass:500 PSI, PTFE: 60 PSIG KFG: 40 PSIG 0-5 VDC 12.5 VDC ± 2 VDC, regulated Power Consumption: 35 ma (typical) Wetted Materials All Units: See Current Datasheet 03/04 Page 7

11. Dimensions KFF/KFG-1... 1/8 " KFF/KFG-1... 1/4" KFF/KFG-1... 3/8"-1/2" KFF/KFG-3... Page 8 03/04

KFF,KFG 12. Cautionary Information PLEASE READ THE FOLLOWING GENERAL FLOW METER/MONITOR WARNINGS BEFORE ATTEMPTING INSTALLATION OF YOUR NEW DEVICE. FAILURE TO HEED THE FOLLOWING INFORMATION MAY RESULT IN EQUIPMENT FAILURE AND POSSIBLE SUBSEQUENT PERSONAL INJURY. Inspect the instrument for damage upon arrival. Cracked, fractured, or otherwise damaged instruments must not be put into use, since the device is weakened to an unknown extent. Under NO circumstances must the maximum tolerances of flow, pressure, temperature, or supply voltage be exceeded. The maximum tolerances of the device have been determined using water, air, and/or oil. If using other media, especially corrosive media, it is critically important that the user determine chemical compatibility with our instruments. KOBOLD Instruments Inc. cannot accept responsibility for failure and consequences resulting from the use of media other than water, air, mineral oil or nitrogen. The sudden cessation of fluid flow causes what is typically called water hammer. Water hammer is a transient force caused by the transfer of momentum of a flowing fluid to the piping system when the flow of fluid is suddenly stopped (i.e. pump trip or valve closure). Water hammer can cause fluid pressure surges, which could cause the flow measuring device's pressure limit to be exceeded, resulting in equipment failure and possible personal injury. These pressure surges can be particularly harmful if the flow instrument is empty since there would be no backpressure in the device. To avoid these pressure surges, the fluid lines should remain full and flow should be introduced and isolated from the system slowly. Freezing of fluid in the instrument must be avoided since the resultant expansion will damage the flowmeter and make it unsafe for use. 03/04 Page 9