ph/redox Measurement MyPro CPM 431



Similar documents
R4D0-FB-IA* FieldBarrier. Features. Assembly. Function. Connection IN + - S OUT Spur 2. Spur 1. Zone 1. Zone 0

DMP 331. Industrial Pressure Transmitter for Low Pressure. Stainless Steel Sensor

Dimensions. LUC-M20 with F12 housing and process connection 2" approx. 86 AF60 G2 2 NPT Ø50. Additional dimensions see section dimensions.

ph/redox Measurement MycomS CPM153

FieldIT Head mounted Temperature Transmitter TF02/TF02-Ex

RHE12. Hazardous Area Coriolis Mass Flow Transmitter. Features. Applications. Benefits

Isolating Barriers for our sensors A5S1 in Ex-areas Series D461

Technical Data. Dimensions

Technical Data. General specifications Switching element function Rated operating distance s n 5 mm

System 6000 Electropneumatic Converter for Direct Current Signals Type 6116 i/p Converter

Electronic pressure switch with display Model PSD-30, standard version Model PSD-31, with flush diaphragm

Conductivity Sensor ConduMax W CLS 21

Technical Data. General specifications Switching element function Rated operating distance s n 15 mm

Technical Information RIA14

Pressure transmitter For general industrial applications Model A-10

Inductive sensor NI10-M18-Y1X-H1141

3DLevelScanner II TECHNICAL DATA SHEET

Digital inductive conductivity transmitter

Turbine Flow Sensor Series VTR

Liquisys M CCM223/253

Temperature sensor with integrated transmitter for industrial applications, MBT 3560

key- precision Tecum cuil Ludere ipsae. Possem Precision level detection When space and price matter Pointek CLS 100

PLUGS & RECEPTACLES. INNOVATIVE EXPLOSION PROTECTION by R. STAHL

Rosemount 333 HART Tri-Loop

How To Control A Power Supply With A Mini Transmitter And Switch (Power Supply)

Technical Data. General specifications Switching element function Rated operating distance s n 5 mm

Mobrey Ultrasonic. MSP900SH Level and MSP900FH Flow Transmitters. Product Data Sheet February 2015 IP2032, Rev BA

Plastic tube flowmeters Series PT/PS

Digital pressure gauge for industrial applications Model DG-10-S, Standard version Model DG-10-E, Enhanced version

Technical Data. Dimensions

SITRANS F flowmeters. SITRANS F M System information SITRANS F M electromagnetic flowmeters. 4/12 Siemens FI

Control- and feedback head for integrated mounting on Robolux valves Type 2036

DiwiTherm, Resistance Thermometers Model TR75, with digital Display Battery or Solar Powered

Safety instructions VEGAPULS PS62/63/66/67/68.GI/RI(*)****** IECEx BVS Ex t IIIC T Da, Da/Db, Db IP 66

Technical Data. General specifications Switching element function Rated operating distance s n 4 mm

SITRANS F flowmeters SITRANS F M. Transmitter MAG 5000/ /32 Siemens FI Overview

Compact pressure switch Flameproof enclosure Ex d Model PCA

IFC 070 Technical Datasheet

Design. Function. Transmitter. Operating and display panel. HART communication. Master

Linear Luminaire with LED Series EXLUX 6402/2

TEIP11 I/P signal converter for standard signals. Measurement made easy

S300 Series Dust Monitors

Product data sheet. JUMO MIDAS Pressure Transmitter. Sales No. Order code JUMO AS /

Automation System TROVIS 6400 TROVIS 6493 Compact Controller

Stainless steel process fittings

AT500 Magnetostrictive Level Transmitter. Compact magnetostrictive liquid level transmitter for direct insertion K-TEK Products

Power Supply and Indicator Unit Type Fig. 1 Type Mounting and Operating Instructions EB 9539 EN

2.12. page Products for Explosion hazardous Environment

Process pressure transmitter Model IPT-10, standard version Model IPT-11, flush diaphragm

SITRANS F flowmeters SITRANS F M

Inductive slot sensor

Ammeter / Voltmeter Series 8402, 8403, 8405, 8404

Differential Pressure Transmitter Model DPT-10

Ferrostat Speed Sensor Series DSF Explosion Proof Versions EEx

Retractable Housings. InTrac 781 / 784 Versatile and Rugged For the Toughest Process Conditions INGOLD

Certified according to DIN EN ISO 9001 Technical Datasheet C-Flow Coriolis Mass Flow Meter

UM-X Field display for continous level sensors

Rosemount 333 HART Tri-Loop

Proximity Transmitters and Seismic Vibration Transmitters

Relative pressure transmitter type 520

Level Switch LBFS. Technical Data. ATEX data Internal inductivity Li < 10 μh Internal capacity Ci < 43 nf Barrier data. Sensor

Solenoid Valve Island Type 3965

Chicago Process Solution Summit: Liquid Analysis Made Easy

Safety instructions VEGAFLEX FX8*(*).FC/ O****A/H*****

Reaction Torque Sensor

Modular I/O System Analog and Digital Interface Modules

Certified according to DIN EN ISO 9001 Technical Datasheet TRICOR Series Mass Flow Meters

PD30ETB20xxIS. Photoelectrics, Background Suppression reflective with IR light. Main features. Description

Technical data. General specifications. Signal voltage V DC Signal duration. 1 s Input 2. Signal voltage. 1 s Analog output.

OPERATION MANUAL. Pen type, separate electrode SOIL PH METER Model : PH-220S

Duct Humidity Transmitter

COMBIMASS. Technical Data COMBIMASS eco-bio +

Electro-Magnetic Flowmeters COPA-XT 2-Wire DT43

Threaded Resistance Thermometers Model TR10-D, Miniature Design

VOLTAGE/CURRENT CALIBRATOR ISO-TECH ILC-421

Hand-held thermometer Model CTH7000

NIVOCONT CONDUCTIVE LEVEL SWITCHES LEVEL SWITCHES

TM7BDM8B expansion block - TM7 - IP67-8 DI/DO - 24V DC A - M8 connector

3/2-, 5/2- and 5/3-way Solenoid Valves for process pneumatics

JUMO BlackLine CR-GT/-EC/-GS Conductive 2-Electrode Conductivity Sensor

Online data sheet PET-1CB010N10SVMC PET PRESSURE SENSORS

Electronic Pressure Switch EDS 3000

3500/62 Process Variable Monitor

Bulk Flow Meter LB 442 P R O C E S S C O N T R O L

Thermometer in stainless steel with capillary and switch contacts


Software and Display instruments

High-precision valve control using intelligent positioners. sipart

Differential Pressure Sensors for air and non-corrosive gases, with calibration certificate

Weighing terminals IP69K GMP EHEDG NSF. Precision in stainless steel Easy to clean perfect hygiene

Flameproof Audible Signal 115 db (A) Series YA90

Technote. Frese OPTIMA Compact actuators DN10-DN32. Application. Features thermic actuators. Features motoric actuators. Approval.

APT3200 SMART PRESSURE TRANSMITTER. Application Areas: For Gauge And Absolute Pressure Measurement

TEIP11-PS I/P signal converter for standard signals. Measurement made easy

OEM Pressure Transmitters for heavy-duty applications, Type MBS 1200 and MBS 1250

TTF300 Field-mount temperature transmitter

Smartec S CLD134. Technical Information

Linear Luminaire for Fluorescent Lamps Series EXLUX 6001

WaterMAG. Electromagnetic Flowmeter for Water Applications. Model MCB10A OVERVIEW APPLICATIONS FEATURES. No. SS2-MCB

Level Plus. Magnetostrictive Liquid-Level Sensors with Temposonics Technology. M-Series Model MR Transmitter with Analog Output.

Transcription:

Technical Information TI 173C/07/en/06.02 50088309 ph/redox Measurement MyPro CPM 431 Two-wire transmitter for ph and redox with Hart and PROFIBUS-PA communication for use in Ex and non-ex areas Application MyPro CPM 431 is a transmitter for ph or redox measurement in all areas of process control and engineering. Compact design and versatile mounting options make MyPro a perfect match for any industrial environment: Ex applications Chemical and petrochemical industries Pharmaceutical industry Power plants Water conditioning Wastewater treatment. Your benefits High reliability is guaranteed by: comprehensive self-monitoring functions Sensor Check System SCS for ph and reference electrodes Versatility: Switchable between ph and redox Compact design: Smallest intelligent transmitter available Simple installation and versatile mounting; display and housing can be rotated Convenient operation via: Keypad Hand-held Hart terminal Commuwin II PROFIBUS-PA Keypad is protected underneath cover Two-level locking function protects configuration and calibration data 0 6 8 Quality made by Endress+Hauser ISO 9001

Function and system design Measuring system In general a measuring system comprises: an electrode with an integrated temperature sensor Pt 100 an assembly for electrode installation in a pipeline or tank the corresponding measuring cable and the transmitter MyPro CPM 431. 4 5 1 2 3 Example of a measuring system 1 MyPro CPM 431 2 Supply and signal circuit, optionally with Hart or PROFIBUS-PA 3 Process assembly UniFit H CPA 442 4 Glass electrode CeraGel P CPS 71 5 Special measuring cable CPK 9 C07-CPM431xx-14-06-00-xx-001.EPS Self-diagnostics Electrode monitoring SCS MyPro continually checks the operational safety of the measuring point. It can identify 28 possible problem causes from these fields: Failure Service required Malfunction Warning Errors are signalled in the field via the display and simultaneously via the Hart interface, and optionally via an error current signal (22 ma). The Sensor Check System SCS monitors the ph and the reference electrodes for inaccurate measurement and total failure. SCS detects: Breakage of electrode glass Fine shorts in ph measuring circuit, also bridges due to moisture or soiling at terminals Soiling or blocking of reference electrode The following methods are employed: ph electrode resistance monitoring (alarm in case the impedance drops below a minimum threshold) Monitoring of reference electrode impedance (an alarm is issued when the defined threshold is exceeded). 2 Endress+Hauser

Input Measured variable Measuring range Input resistance Input current Ex version Hart Ex version PROFIBUS-PA ph, Redox, temperature ph: 2... 16 Redox: 1500... +1500 mv Temperature: 20... +150 C > 10 12 Ω (at nominal operating conditions) < 1.6 x 10 12 A (at nominal operating conditions) 0 Intrinsically safe supply and signal circuit, protection type EEx ib IIC T4 Max. input voltage: U i = 30 V Max. input current: I i = 100 ma Max. input power: P i = 750 mw Max. internal inductance: L ι = 200 µh Max. internal capacitance: C i = negligible Capacitance to PE: 5.3 nf 0 For connection to a supply circuit, protection type EEx ia IIC or EEx ib IIC Max. input voltage: U i = 24 V Max. input power: P i = 1.2 W or Max. input voltage: U i 17.5 V for connection to intrinsically safe PROFIBUS systems, specified by a supply voltage 17.5 V according to the FISCO model. Cable specification Without SCS: With SCS: max. cable length 50 m max. cable length 20 m Output Output signal Signal on alarm Load Transmission behavior Ex version 4... 20 ma, potential separated from sensor circuit 0.8... 1.2 ma peak to peak (Hart only) 22 ± 0.5 ma max. 600 Ω (depending on operating voltage and load) 230... 1100 Ω (Hart only) ph: adjustable, 2.0... 18 (error message if < 2) Redox: adjustable, 200... 3000 mv 0 Intrinsically safe sensor circuit, protection type EEx ia IIC T4 Max. Output voltage: U 0 = 12.6 V Max. Output current: I 0 = 37 ma Max. Output power: P 0 = 117 mw Max. external inductance: L 0 = 100 µh Max. external capacitance: C 0 = 50 nf Endress+Hauser 3

Power supply Electrical connection Detail A 12 30 V DC Detail A 1 2 3 4 20 ma Electrical connection CPM 431, terminal 3 not connected C07-CPM431ZZ-04-00-00-xx-001.EPS Electrode connection 11 GN WH ϑ YE BN ph/mv 12 13 14 15 16 17 Connection options: Symmetrical measurement with PMC (with potential matching pin A) Asymmetrical measurement without PMC (with potential matching pin A) BN GN WH YE PAL ϑ A ph Ref Connection scheme CPM 431 C07-CPM431ZZ-04-00-00-xx-002.EPS A = potential matching pin Supply voltage Without Hart : With Hart : 12... 30 V DC 13.5... 30 V DC Cable specification Power consumption max. cable profile: 2.5 mm 2, PE 4 mm max. 700 mw 4 Endress+Hauser

Reference temperature 25 C Performance characteristics Measured value resolution Maximum measured error a ph: 0.01 ph Redox: 1 mv Temperature: 0.1 C ph: max. 0.2% of measuring range Redox: max. 0.2% of measuring range Temperature: max. 1 C Repeatability a ph: Redox: Temperature: 0.1% of measuring range 0.1% of measuring range 0.1% of measuring range Zero drift range Glass electrode 7.0: ph 5.7... 8.3 Glass electrode 4.6: ph 3.32... 5.92 Antimony electrode: ph 1.0... 3.0 Slope adaption Glass electrode 4.6 and 7.0: Antimony electrode: 45... 65 mv/ph 25... 65 mv/ph Electrode offset redox Temperature compensation range ±200 mv 20... +150 C Temperature offset ±20 C Installation Installation instructions 104 160 137 53 118 227 97 CPM 431: wall mounting C07-CPM431ZZ-11-00-00-xx-001.EPS a) acc. to DIN IEC 746 part 1, at nominal operating conditions Endress+Hauser 5

254 177 CPM 431: flange mounting bracket (with CPM431-xxx4xx in scope of delivery) C07-CPM431ZZ-11-00-00-xx-002.EPS 84 Ø6 77 Ø60 77 CPM 431: pipe mounting (pipe DN 60) C07-CPM431ZZ-11-00-00-xx-003.EPS 53 8 82 Ø 30 200 102 165 CPM 431: pipe mounting (pipe DN 30... 200) C07-CPM431ZZ-11-00-00-xx-004.EPS 6 Endress+Hauser

Installation in Ex environment 0 Measuring system in Ex environment C07-CPM431ZZ-11-00-00-xx-006.TIF 1 2 3 4 Sensor in Ex version, e.g. CPS 71G Intrinsically safe sensor circuit EEx ia MyPro CPM 431 Supply and signal circuit EEx ib (4... 20 ma) 5 6 7 8 Active barrier, e.g. preline RN 221 Signal line for Hart / PROFIBUS (0/4... 20 ma) Commuwin II via Hart or PROFIBUS-PA Hand-held Hart terminal Environment Ambient temperature range Ambient temperature limit 10... +55 C 20... +60 C (non-ex version) 15... +55 C (Ex version) Storage temperature 25... +80 C Electromagnetic compatibility Interference emission and interference immunity acc. to EN 61326: 1997 / A1: 1998 Ingress protection IP 65 Relative humidity 10... 95%, non-condensing Mechanical construction Dimensions Weight Materials H x W x D: 227 x 104 x137 mm max. 1.25 kg Housing: GD-AISI 10 Mg, plastic-coated Endress+Hauser 7

Human interface Display elements LC display, rotable Display MyPro C07-CPM431ZZ-19-00-00-xx-001.EPS Operating elements Operation via keypad of MyPro (A) There are two levels for operation: Level 1: Viewing of active settings Error diagnostics Current output settings Calibration Level 2: other parameters, e.g. switching between measured variables Operation via Hart or PROFIBUS-PA (B and C) Hand-held terminal (Hart only) Plain text menu guidance Operation via Commuwin II (Hart and PROFIBUS communication) Clear overall view in the form of a matrix Grafic user interface Documentation Offline-programming 8 Endress+Hauser

HART Handterminal 1 Offline 2 Online 3 Transfer 4 Frequenzg erät 5 Extras I O MyPro CPM 431 Overview operation A + C F F > 3 s B C Operation of MyPro A Operation via keypad B Operation via hand-held Hart terminal C Operation via Commuwin II via Hart or PROFIBUS-PA C07-CPM431ZZ-19-00-00-xx-002.EPS Certificates and approvals 4 approval Declaration of conformity The product meets the legal requirements of the harmonised European standards. Endress+Hauser confirms compliance with the standards by affixing the 4 symbol. Ex approval CSA IS NI Cl.I, II, III, Div. 1&2, Group A-G FM IS NI Cl.I, II, III, Div. 1&2, Group A-G EEx ia/ib IIC T4, ATEX II (1)2G Endress+Hauser 9

Ordering information Product structure Type of certificate A Variant for non-ex area H EEx ia/ib IIC T4, ATEX II (1)2G O FM IS NI C1.I, II, III, Div. 1&2, Group A-G S CSA IS NI C1.I, II, III, Div. 1&2, Group A-G Y Special version acc. to customer s specification Power supply cable entry 1 Cable gland Pg 13,5 3 Cable entry M 20 x 1,5 5 Cable entry NPT ½" 7 Cable entry G ½ 8 PROFIBUS-PA M12 plug 9 Special version acc. to customer s specification Electronics, communication, display A 4... 20 ma, Hart, without display B 4... 20 ma, Hart, LC display C PROFIBUS-PA, without display D PROFIBUS-PA, LC display Y Special version acc. to customer s specification Accessories 1 No accessories 2 For wall and pipe mounting DN 60 3 For wall and pipe mounting DN 30... DN 200 4 With flange mounting bracket 9 Special version acc. to customer s specification Factory parameter configuration P ph, measuring range ph -2... 16 R Redox, measuring range ±1500 mv Y Special version acc. to customer s specification Cable, sensor connection A Without cable B With 1m cable, GSA plug-in (without Pt 100) D With 2 m cable, GSA plug-in (without Pt 100) F With 2 m cable, TOP 68 / ESA / HDA plug-in G With 1 m cable, TOP 68 / ESA / HDA plug-in K With Y-form cable gland, without cable CPM 431- complete order code 10 Endress+Hauser

Accessories Sensors! Note! CeraGel P CPS 71 C07-CPS71xxx-21-05-00-xx-001.EPS CeraGel P CPS 71, ph/redox electrode with double chamber reference system and integrated bridge electrolyte; Ordering depending on specification, s. Technical Information CeraLiquid P CPS 41, ph/redox electrode with ceramic diaphragm and liquid KCl electrolyte; Ordering depending on specification, s. Technical Information All sensors are available as Ex sensors. CeraLiquid P CPS 41 C07-CPS41xxx-21-05-00-xx-001.EPS Assemblies FlowFit P CPA 240 C07-CPA240xx-21-07-00-xx-001.EPS C07-CPA140xx-21-07-00-xx-001.EPS DipFit P CPA 140 C07-CPA471xx-21-07-00-xx-001.EPS CleanFit P CPA 471 C07-CPA442ZF-21-0-00-xx-001.EPS UniFit H CPA 442 FlowFit P CPA 240, Flow assembly for processes with high requirements of PVDF or stainless steel DipFit P CPA 140, Immersion assembly with flange connection for processes with high requirements CleanFit P CPA 471, Compact retractable assembly for installation in tanks and pipelines UniFit H CPA 442, Process assembly for foodstuffs, biotechnology and chemicals with EHEDG and 3A certificate Special measuring cable Buffer solutions Active barrier with power supply CPK 1, for electrodes with GSA plug-in head CPK 9, for electrodes with ESA or HDA plug-in head ph buffer solution ph 4.0 100 ml; order no. CPY2-0 ph buffer solution ph 4.0 1000 ml; order no. CPY2-1 ph buffer solution ph 7.0 100 ml; order no. CPY2-2 ph buffer solution ph 7.0 1000 ml; order no. CPY2-3 ph buffer solution ph 9.2 100 ml; order no. CPY2-4 ph buffer solution ph 9.2 1000 ml; order no. CPY2-5 Redox buffer solution +220 mv, ph 7.0 100 ml; order no. CPY3-0 preline RN 221 (non-ex) preline RN 221 Z (Ex) Endress+Hauser 11

Documentation CeraGel P CPS 71/72, Technical Information, TI 245C/07/en; order no. 51505837 CeraLiquid P CPS 41/42/43, Technical Information TI 079C/07/en; order no. 50059346 FlowFit P CPA 240, Technical Information TI 179C/07/en; order no. 50088970 UniFit H CPA 442, Technical Information TI 306C/07/en; order no. 51507254 DipFit P CPA 140, Technical Information TI 178C/07/en; order no. 50088968 CleanFit P CPA 471, Technical Information TI 217C/07/en; order no. 51502596 ph measuring cable CPK 1-12, Technical Information TI 118C/07/en; order no. 50068526 Active barrier preline RN221, Technical Information TI 073R/09/en; order no. 51001410 Endress+Hauser GmbH+Co. KG Instruments International P.O. Box 2222 D-79574 Weil am Rhein Germany Tel. (07621) 975-02 Tx 773926 Fax (07621) 975 345 e-mail: info@ii.endress.com Internet: http://www.endress.com 11.01 TI 173C/07/en/06.02 50088309 Printed in Germany / FM+SGML 6.0 / DT