Disclosure statement: Advances in Photon Counting Detectors in X-ray Imaging: Photon Counting Detectors for Mammography



Similar documents
Physics testing of image detectors

Quality Control of Full Field Digital Mammography Units

Scan Time Reduction and X-ray Scatter Rejection in Dual Modality Breast Tomosynthesis. Tushita Patel 4/2/13

Digital Mammography Update: Design and Characteristics of Current Systems

Concepts for High-Resolution Low-Dose CT of the Breast

Dose Measurement in Mammography; What are we measuring? David E. Hintenlang, Ph.D. DABR University of Florida

Q.A. Collectible. Sponsored by CRCPD s Committee on Quality Assurance in Diagnostic X-Ray (H-7)

BREAST CHARACTERISTICS AND DOSIMETRIC DATA IN X RAY MAMMOGRAPHY - A LARGE SAMPLE WORLDWIDE SURVEY

Radiography: 2D and 3D Metrics of Performance Towards Quality Index

Solid State Detectors = Semi-Conductor based Detectors

College on Medical Physics. Digital Imaging Science and Technology to Enhance Healthcare in the Developing Countries

for breast cancer detection

CONTENT SPECIFICATIONS FOR THE FLUOROSCOPY EXAMINATION

An Overview of Digital Imaging Systems for Radiography and Fluoroscopy

A Guide to Breast Imaging: The Latest Technology for Screening and Detecting Breast Cancer

SECTION 1: REQUIREMENTS FOR CERTIFICATES OF COMPLIANCE FOR CLASSES OF RADIATION APPARATUS

Mammography. What is Mammography?

EDXRF of Used Automotive Catalytic Converters

Role of the Medical Physicist in Clinical Implementation of Breast Tomosynthesis

Fundamentals of Digital Mammography: Physics, Technology and Practical Considerations Andrew P. Smith, Ph.D.

The Fundamentals of MTF, Wiener Spectra, and DQE. Motivation

Acquiring molecular interference functions of X-ray coherent scattering for breast tissues by combination of simulation and experimental methods

Avalanche Photodiodes: A User's Guide

The Principles of Contrast Mammography

Dose enhancement near metal electrodes in diamond X- ray detectors. A. Lohstroh*, and D. Alamoudi

Fundamentals of Cone-Beam CT Imaging

ENERGY PER PHOTOELECTRON IN A COINCIDENCE LIQUID SCINTILLATION COUNTER AS A FUNCTION OF ELECTRON ENERGY. Donald L. Horrocks

ORTEC AN34 Experiment 7 High-Resolution Gamma-Ray Spectroscopy

Amptek Application Note XRF-1: XRF Spectra and Spectra Analysis Software By R.Redus, Chief Scientist, Amptek Inc, 2008.

Radiation Detection and Measurement

Appendix A. An Overview of Monte Carlo N-Particle Software

Experimental study of beam hardening artefacts in photon counting breast computed tomography

Irradiation Field Size: 5cmX5cm 10cmX10cm 15cmX15cm 20cmX20cm. Focus-Surface Distance: 100cm. 20cm Volume of Ion Chamber : 1cmX1cmX1cm

A Digital Spectrometer Approach to Obtaining Multiple Time-Resolved Gamma-Ray. Spectra for Pulsed Spectroscopy. W. K. Warburton a

Infrared Thermography Not a Useful Breast Cancer Screening Tool

Mammography. For 35+ years, screenfilm has been the gold standard for breast cancer detection

Breast Imaging Made Brief and Simple. Jane Clayton MD Associate Professor Department of Radiology LSUHSC New Orleans, LA

1090 IEEE TRANSACTIONS ON NUCLEAR SCIENCE, VOL. 50, NO. 4, AUGUST Intelligent Gamma-Ray Spectroscopy Using 3-D Position-Sensitive Detectors

CT: Size Specific Dose Estimate (SSDE): Why We Need Another CT Dose Index. Acknowledgements

Activitity (of a radioisotope): The number of nuclei in a sample undergoing radioactive decay in each second. It is commonly expressed in curies

Hologic Selenia Dimensions C-View Software Module. October 24, 2012

Mammography: Dosimetry ACMP Annual Meeting Virginia Beach Saturday, May 2, 2009 Lawrence N. Rothenberg, Ph.D. Department of Medical Physics Memorial

3. What would you predict for the intensity and binding energy for the 3p orbital for that of sulfur?

Qualitätsmanagement-Handbuch

Application Note # EDS-10 Advanced light element and low energy X-ray analysis of a TiB 2 TiC SiC ceramic material using EDS spectrum imaging

Digital Breast Tomosynthesis QC Requirements

Performance testing for Precision 500D Classical R/F System

Radiographic Grid. Principles of Imaging Science II (RAD 120) Image-Forming X-Rays. Radiographic Grids

Update on ACR Digital Mammography QC Manual

Running head: COMPARISON OF AMORPHOUS SELENIUM AND AMORPHOUS 1

Introduction to Geiger Counters

Surveying and QC of Stereotactic Breast Biopsy Units for ACR Accreditation

(Amplifying) Photo Detectors: Avalanche Photodiodes Silicon Photomultiplier

Breast Density Legislation: Implications for primary care providers

Using the Bruker Tracer III-SD Handheld X-Ray Fluorescence Spectrometer using PC Software for Data Collection

Introduction. Digital Mammography Detector QC FFDM. What is FFDM QC and why is it. important? What to know before you start. Eric A. Berns, Ph.D.

Personalized Breast Screening Service

R/F. Efforts to Reduce Exposure Dose in Chest Tomosynthesis Targeting Lung Cancer Screening. 3. Utility of Chest Tomosynthesis. 1.

MAMMOGRAPHY GOALS AND OBJECTIVES

A VERSATILE COUNTER FOR CONVERSION MÖSSBAUER SPECTROSCOPY

Coating Thickness and Composition Analysis by Micro-EDXRF

Production of X-rays. Radiation Safety Training for Analytical X-Ray Devices Module 9

Annual MQSA Inspection Questions NMQAAC April 19, 2004

Name period AP chemistry Unit 2 worksheet Practice problems

ACR AAPM SIIM PRACTICE PARAMETER FOR DETERMINANTS OF IMAGE QUALITY IN DIGITAL MAMMOGRAPHY

Feasibility Study of Neutron Dose for Real Time Image Guided. Proton Therapy: A Monte Carlo Study

PHOTOELECTRIC EFFECT AND DUAL NATURE OF MATTER AND RADIATIONS

Gamma and X-Ray Detection

47374_04_p25-32.qxd 2/9/07 7:50 AM Page Atoms and Elements

Atomic and Nuclear Physics Laboratory (Physics 4780)

How To Compare Mammography To A 3D System

Update on MQSA and ACR Mammography Accreditation

CONSOLIDATED GUIDANCE ON STANDARDS FOR THE NHS BREAST SCREENING PROGRAMME

From lowest energy to highest energy, which of the following correctly orders the different categories of electromagnetic radiation?

X-ray absorptiometry of the breast using mammographic exposure factors: application to units featuring automatic beam quality selection

High-resolution bio-imaging with liquid-metal-jet x-ray sources

Recommendations for cross-sectional imaging in cancer management, Second edition

Contents. X-ray and Computed Tomography. Characterization of X-rays. Production of X-rays

2. John Dalton did his research work in which of the following countries? a. France b. Greece c. Russia d. England

Quality Control and Maintenance Programs

CHEM 1411 Chapter 5 Homework Answers

Chapter 15. Creating the Digital Image. Objectives. Key Terms

PERFORM-X DIGITAL X-RAY SYSTEM PERFORM-X. with ceiling mounted tube support PERFORM-X

The File-Card-Browser View for Breast DCE-MRI Data

Image Quality and Radiation Dose for Intraoral Radiography: Hand-Held Held (Nomad), Battery Powered

CHAPTER 7 SCINTILLATION COUNTING

PHOTONS - ATTENUATION

Evaluation of a metal artifact reduction algorithm in CT studies used for proton radiotherapy treatment planning

Lectures about XRF (X-Ray Fluorescence)

GAMMA-RAY SPECTRA REFERENCES

MODERN ATOMIC THEORY AND THE PERIODIC TABLE

Diagnostic x-ray imaging relies on the attenuation of

An Innovative Method for Dead Time Correction in Nuclear Spectroscopy

MDCT Technology. Kalpana M. Kanal, Ph.D., DABR Assistant Professor Department of Radiology University of Washington Seattle, Washington

Scattered radiation in projection X-ray mammography and digital breast tomosynthesis

OBJECTIVES By the end of this segment, the community participant will be able to:

Ionizing Radiation, Czech Republic, CMI (Czech Metrology Institute)

Transcription:

AAPM 51 st ANNUAL MEETING Anaheim, CA June 7, 009 Advances in Photon Counting Detectors in X-ray Imaging: Photon Counting Detectors for Mammography Disclosure statement: -Founder Sectra Mamea -Shareholder Sectra Mats Danielsson Royal Institute of Technology Stockholm, Sweden md@mi.physics.kth.se To Count Photons What does it mean? Work in progress - FDA approval pending for all devices described in this presentation Energy X-ray photon Threshold 1 Threshold Threshold 3 Result: -Number of X-rays -Pulse height give the color for each X-ray -No electronic noise (regardless of pixelsize and dose!) Noise 1 3 Time

Integrating Current: Today s praxis in X-ray Imaging Mammography Energy X-ray photons 1 Result: -Area of -Indirect estimate of number of photons Time 100 M women examined every year Mortality in breast cancer reduced 30-50% Glandular tissue sensitive to radiation High requirements on spatial and contrast resolution Low kvp Relatively low count rate Object relatively still Photon Counting Mammography Today In the basement Sectra mammography started 9 years ago.. Sectra Silicon detectors -00 systems in 15 countries XCounter Gas detectors Research and development Both are scanned-slit

First Image Jan 001 Sectra Photon Counting Mammography Today Development of Sectra Photon Counting Mammography XCounter Photon Counting Mammography with Gas Detector First model 003 (D0) heavy and large Current model 007(L30) light and slim Prototype at Karolinska Sjukhuset, Stockholm Outline of technology 1

Crystalline Silicon Detector Sectra Photon Counting Mammography Scanned Slit Efficient Scatter Rejection Sectra Silicon Detector Geometry Geometry for photon counting (diamonds) increases system DQE with 40% compared to typical grid X-rays Dead area Active Detector Volume 8 degrees Dead area * M. Åslund et al. Med Phys, vol. 33, pp. 933-940, 006

Silicon for Photon Counting Silicon for Photon Counting Detector Response to High Energy X-rays Cross sections for interaction Intensity Compton edge Photo absorption peak Silicon CdZnTe Energy

Summary of Crystalline Silicon Properties for Mammography Mobility electrons 1350 cm /Vs Mobility holes 480 cm /Vs Fano factor 0.1 Signal 5500 electron-holes pairs for 0 kev x-ray Noise (dark current) ~ 1 electron Breakdown voltage 3 10 5 V/cm Fluorecence No Generation Dead time is already a major industry standard and easily available in large production volumes Measured Effect of Dead Time Measured Effect of Dead Time Deadtime per event 189 ns Area of operation Deadtime at 0 khz 0.4% Deadtime at 00 khz 3.7% R.L. Lucke Rev. Sci. Instrum. 47 (6):766 (1976) E. Fredenberg et al. SPIE 758 (009) E. Fredenberg et al. SPIE Physics of Medical Imaging 009

How to avoid problems with charge sharing: Co-incidence Logic Charge sharing Effect of charge sharing DQE + clinical data *E Fredenberg et al, Submitted to Nuclear Instr Meth A

DQE for all different mammography detector materials (Se,CsI, Si)* Sectra Tomosynthesis DQE 0.7 0.6 0.5 0.4 0.3 Kodak CR 850 EHR-M Konica Regius 190 Fuji CR Profect GE Senographe DS Lorad Selenia Sectra MDM Tomo 0. 0.1 0.0 0 1 3 4 5 6 Spatial frequency [mm -1 ] Note: DQE does not incorporate effects of energy weighting which benefits photon counting (10-15%)** 7 * Monnin et al. Med. Phys., vol. 34 (3), pp. 906-914, 007 ** R.N. Cahn et al. Med. Phys., vol. 6 (1), pp. 680-683, 1999 8 9 10 Elimination of electronic noise give low dose extra advantage for photon counting D Clinical Studies dose, sensistivity and specificity B. Heddson, K. Rönnow, M. Olsson, D. Miller Digital versus screen-film mammography: A retrospective comparison in a population-based screening program European Journal of Radiology, Volume 64, Issue 3, Pages 419-45 (007) -75% dose reduction, increased cancer detection rate at same callback rate Future applications: measuring the energy and using the spectral information M. Wallis Evaluation and clinical assessment of digital mammography screening using a Sectra Microdose full field digital x-ray unit and Sectra breast imaging PACS NHSBSP REPORT 0601 MAY 006 UNITED KINGDOM -70% dose reduction, same cancer detection rate and call-back rate

Numerous Microvessels in Stained Section Experimental evidence suggests that the growth of a tumor beyond a certain size requires angiogenesis N Weidner et al. New England J. Med. 1991;34:1 8. Dromain et al. AJR 006; 187:58 537 Typical MRI Images Contrast-enhanced Digital Mammography Maximum-Intensity Intensity-Projection (MIP) contrast-enhanced T1-weighted coronal image Multicentric invasive ductal carcinoma with extensive DCIS component (S:t Goran Hospital, Stockholm Sweden) Diekmann et al: 14-mm invasive ductal carcinoma(occult on mammography)

Challenges with Temporal Subtration CEM 1. Compression may interfere with blood flow. Motion blurring because of long time between exposures (5-10 min) 3. Compromise in image quality of standard mammogram since the kvp is to high (40-50 kvp) Dual Energy Contrast Enhanced Mammography (DE-CEM) 1. High Energy Exposure 40-45 kvp, 00 ma, eg Rh anode, heavy filtering. Low Energy Exposure 30 kvp, 140 ma, eg Mo anode 3. Combine the high and low energy images into one image Subtract the tissue, enhance Iodine signal Advantages: Less motion blur, compression not a problem Spectral Imaging Lewin et al. Radiology, 003 14 Previous example: dual exposures Spectra after breast 18 Photon-counting: single exposure Spectrum after breast 1 16 Fluence (10 6 phot/cm ) 10 8 6 4 Rh W Fluence(10 6 ph./cm ) 14 1 10 8 6 4 Iodine K-edge 0 0 10 0 30 40 Energy (kev) 0 10 15 0 5 30 35 40 45 Energy (kev) Non-overlapping spectra better tissue cancellation Concentrated around K-edge improved I-contrast No motion blurring, easy to implement

Breast MRI v. Spectral Mammography Spectral Mammography -a Future Alternative to Breast MR? Mammography is first with photon counting and thus a technology leader in medical x-ray imaging -Same diagnostic information? -Faster (no motion) -Improved spatial resolution -Similar appearance as normal mammograms -Contrast to noise ratio better for MRI? -10 years more of clinical experience with breast MR Easier in combination with biopsy? -Cancer outside of the breast will not be detected money, money, money Phantom study 1-9 mm deep iodine containers in 15 mm PMMA Anatomical clutter: 35 mm PMMA and oil Previous study with a single detector line: E. Fredenberg et al. SPIE (008)

Dual Energy Subtraction Phantom study Logarithmic subtraction with energy bins: ln I = wln n S = ln I - ln I 1 lo - ln n hi = w( d m ª wdm eff 1 lo,1 eff lo - d m - dm eff hi eff lo, + K ) + ( d m S = 0 for proper choice of w (if d 1 = d ) Cost of increased quantum noise? eff hi,1 - d m ) eff 1 hi, Results with the full system: 40 kv, 3 mm Al, 1.13 mgy, ~10 s 3 mg/ml > 6 containers visible in the subtracted image Phantom study Detector model* Results with the full system: 40 kv, 3 mm Al, 1.13 mgy, ~10 s 6 mg/ml 9 containers visible in the subtracted image Quantum efficiency, Intrinsic energy resolution of silicon Charge sharing: simulated** Pile-up: dead time measured to 47 ns Anti-coincidence logic: ~0% leakage, ~150 ns AC window chance coincidence Spread of thresholds: measured from threshold scans *E Fredenberg et al, Submitted to Nuclear Instr Meth A **M Lundqvist et al, IEEE Trans Nucl Sci, 000

Detector model Threshold scans Energy Resolution Detector model verification Measurements of the PMMA-iodine phantom Image Noise - quantum noise dominates at high f - anatomical noise at low f

Generalized DQE* Detectability index MTF GDQE = q (NPS + NPS MTF ) MTF, NPS quantum measured NPS anatomical modeled as 1/f 3.9 0 quantum *Richard et al, Med Phys (006) anatomical Contrast + GDQE + task function detectability index d GDQE F = C du d v ~. times improvement compared to the absorption image Can we use the new technology in mammography screening? Theory of Photon Counting Challenge No contrast agent acceptable (?) except for high risk groups But, spectral x-ray imaging may work anyway to improve the results in dense breasts r 4 r m ª C1 Z f PE ( E) + C Zf KN ( E), 14A 4 443 14A 4 443 photoelectric effect Compton scattering f PE µ E -3, discontinuities at absorption edges f KN is the Klein-Nishina cross section Alvarez et al. Phys. Med. Biol. 1976 76)

Composition of the Female Breast (% of wet weight) Elements in the Female Breast Tissue Density (g/cm3) Hydrogen Carbon Nitrogen Oxygen Ash (S P K Ca) Fat - 11.9 76.4 0. 11.5 0.1 Skin 1.09 9.8 17.8 5.0 66.7 0.7 Adipose Tissue 0.93 11. 61.9 1.7 5.1 0.1 Gland 1.04 10. 18.4 3. 67.7 0.5 R. Hammerstein et al. Radiology 130:485-491 (1979) Element Normal Tissue (median µg/g) Al 13 8 Br 35 Ca 94 653 Malign Tissue (median µg/g) Cl 10140 155 Co 1.19 1.95 Cs 0.51 1.1 Fe 94 194 K 60 14400 Na 900 1330 Rb 11.5 8.5 Zn 3.5 57 15 paired samples Kwan-Hong et al. Differentiation of Elemental Composition of Normal and Malignant Breast Tissue by Instrumental Neutron Activation Analysis Appl. Radiat. Isot. (4) 1993, p 511-516 Additional Elements in the Female Breast The Highrex Project Element Silicon Iodine Gadolinium Breast implants Contrast agent (X-ray) Contrast agent (MR, X-ray) Clinical Participants Addenbrook s Hospital, Cambridge, UK (M. Wallis) S:t Goran s Hospital, Stockholm, Sweden (K. Leifland) Health Unit of Pistoia, Florence, Italy (M. Rosselli del Turko) Charite Hospital, Berlin, Germany (F. Diekmann) Munster University Hospital, Munster, Germany (W. Heindel) Arcades, Marseille, France (B. Seradour) Aim Test 3D and Spectral Mammography (with and without contrast) versus standard mammography and breast MRI, enriched populations of 3*00 women Status units deployed with 3D imaging (around 00 cases collected) 1 unit deployed with spectral detector

Conclusions For mammography photon counting is out there in clinical reality (except for US) Low radiation dose advantage Spectral information most valuable for large objects (> a few mm) Promising results for contrast mammography with Iodine Spectral information may be used to increase cancer detection Spectral information easy to implement as add-on information in mammography at almost no cost to be used when desired Thank you!