Case Study in Test Method Validation - ASTM F1886 (2004), Visual Examination Todd Dehmer CPP Senior Packaging Engineer St. Jude Medical, AFD Division March 3-5, 2009 www.healthpack.net 1
Presentation Overview Visual Inspection Usage ASTM F1886 Overview Reason for Validation Process for Validation On-going Validation 2
Visual Inspection Usage Many Medical Device Manufactures are referencing ASTM F1886 as the visual inspection method they use to verify the sterile barrier integrity of their packaging. Are the MDM s referencing this Standard accurately? As specified per ASTM F1886: Should be validated to meet the precision and bias of ASTM F1886. Should be validated in accordance with AAMI/ANSI/ISO 11607-1:2006 and -2:2006? Work within your own quality department to determine applicability. 3
Background Per ISO 11607 Part 1: 4.4 Test Methods 4.4.1 All test methods used to show compliance with this part of ISO 11607 shall be validated and documented. NOTE Annex B contains a list of suitable test methods (such as ASTM F1886). 4.4.2 The test method validation shall demonstrate the suitability of the method used. The following elements shall be included: establishment of a rationale for the selection of the appropriate test for the packaging system; establishment of acceptance criteria; NOTE Pass/fail is a type of criteria; determination of test method repeatability; determination of test method reproducibility; and establishment of test method sensitivity for integrity tests. 4
ASTM F1886 Overview Scope: To determine channels in a package seal down to a width of 75μm (0.003 in.) with a 60-100% probability. To visually detect channel defects applicable to flexible and rigid packages with at least one transparent side. 5
ASTM F1886-04 Overview Summary: To provide a qualitative visual inspection method to evaluate the appearance characteristics of unopened, intact seals in order to determine the presence of defects that may affect the integrity of the package. 6
ASTM F1886-04 Overview Significance and Use Seal attributes can be linked directly to a number of variables in process parameters, equipment, or material, as well as environmental. Visual seal characteristics and defects can provide evidence of sterile package integriy and production sealing problems. 7
ASTM F1886-04 Overview Lighting and Procedure Lighting arrangements to give approximately 540 lumens/m 2 (50fc) of white light or daylight. Visual acuity shall be such that the inspection of the seal may be performed at a distance of 12 to 18 in. 8
ASTM F1886 Overview Precision and Bias A round robin study was conducted for ASTM F 1886. The internal test method should be shown to be equivalent to or better than ASTM F 1886 round robin results to differentiate acceptable package seals from specific types of defects. 9
ASTM F1886 Overview Precision and Bias - continued To accomplish this, samples with the same defect categories as those documented in the ASTM method can be produced as well as a quantity of negative control samples with no channels. The defect categories used were 0.003 (75μm) channel leaks and 0.005 (125μm) channel leaks. 10
When would this method be used? As an internal method to validate SBS sealing processes and new SBS configurations. As an internal method to evaluate SBS after package performance testing. As a process monitoring/control tool. Evaluating seal integrity after an aging study on an SBS. Note: ASTM F1886 is an FDA Recognized Consensus Standard 11
Need for Validation Why is it necessary to validate this test? Requirement per ISO 11607 Current internal visual inspection method may not have provided any formal training or attempt to validate. Provides better packaging quality. 12
Process for Validation Test Prerequisites Acquire 0.003 & 0.005 diameter wire to create channel defects. Determine sample category to use for validation (such as pouches or sealed trays). Sample has to have one transparent side. Determine quantity of each sample category to use during validation (this can be based on internal statistical requirements and significance). 13
Process for Validation Test Prerequisites - continued Have properly installed, calibrated, and maintained sealing equipment (bar sealers/tray sealers). Confirm lighting requirements are met per ASTM F1886 where the internal visual inspection of defects will be performed (measure via a light meter or similar device). Determine who will be assembling the samples. Determine who will be examining the samples. 14
Process for Validation Create a test method validation plan to include: Plan information Plan approvals Test Method Data Validation Planning Validation Targets Sampling Plan Preparation Records Training, Examination and Data Collection Plan Analyze the data and prepare a Validation Report Approval Section 15
Process for Validation Plan information Internal document number/identification/revision Plan approvals Author(s) of Validation Plan and others (such as quality representative) per your internal requirements Test Method Data Test method description (Documented Visual Inspection procedure of pouch seals or tray seals) Test method identification (ASTM F 1886) 16
Process for Validation Validation Planning The scope/purpose of the validation Description of what your trying to accomplish Validation Targets Repeatability Reproducibility Validation (target of what needs to be met to conform to ASTM F 1886) 17
Process for Validation Sampling Plan Document quantity to be tested for each category (0.003, 0.005, & no defect negative controls) 18
Process for Validation Preparation Records Sample numbering (what labeling/marking method to use to determine which samples have certain channel sizes and no defect samples) Equipment Equipment settings (process number and revision) Environmental Conditions during assembly 19
Process for Validation Preparation Records - continued Tooling (as required) Part and lot numbers of the SBS components used to assemble the samples Person(s) assembling the samples Date parts were assembled Person pre-judging the parts (if applicable/required) 20
Process for Validation Training, Examination and Data Collection Plan Instruction provided to examiners (such as a review of ASTM F 1886) and existing internal visual inspection procedure Data Collection Plan Administering the examinations 21
Process for Validation Analyze the data and prepare a Validation Report Organize and analyze the data collected from the examination Make determination of the validation status based upon the data Document the results in a Validation Report including all applicable data After a Validation Plan is complete its time to get started 22
Process for Validation Getting started follow thru with your Validation plan Document your equipment used (internal designation or otherwise, calibration date, etc.) Confirm equipment settings meet internal manufacturing process requirements Document environmental conditions during assembly Tooling used Document part numbers, lot information, etc. Person(s) assembling the samples Date of assembly Person pre-judging the parts (if applicable) 23
Process for Validation Getting started follow thru with your Validation plan - continued Document your Package Sample Defect & Identification Key 24
Process for Validation Prepare samples per validation plan Sample Preparation Randomly place the wire between the location to be sealed to create the defect (channel). 25
Process for Validation Sample Preparation Example photo of how defect areas were designated. 26
Process for Validation Sample Preparation Follow your Identification Key and label sample appropriately. Depending on your sample category (trays, pouches, etc) you will seal up samples for the 0.003, 0.005, and no defect. 27
Administer Visual Examinations Perform visual examination (confirm adequate lighting per ASTM F 1886) When successfully completed it will provide evidence that an operator can perform the visual inspection and will be considered a validated operator. 28
Analyze the data Based on the results, the personnel who examined the samples should demonstrate the ability to repeat the results of the round robin study done by ASTM 29
Analyze the data Trays Personnel who examined tray/lid samples should have identified 98% of no defects or better. Personnel who examined defects of the 0.003 channels should have determined there to be defects 60% of the time. Personnel who examined defects of the 0.005 channels should have determine there to be defects 83% of the time. 30
Analyze the data Pouches Personnel who examined pouch samples should have identified 97% of no defects or better. Personnel who examined defects of the 0.003 channels should have determined there to be defects 69% of the time. Personnel who examined defects of the 0.005 channels should have determine there to be defects 96% of the time. 31
Internal visual inspection This presentation walked through the validation process for ASTM F1886. This standard can be used for the following and are examples: Manufacturing Procedure Workmanship Specification Work within your own quality department to determine applicability. 32
References - ASTM F1886-98(2004); Standard Test Method for Determining Integrity of Seals for Medical Packaging by Visual Inspection. - ANSI/AAMI/ISO 11607-1:2006; Packaging for terminally sterilized medical devices Part 1: Requirements for materials, sterile barrier systems, and packaging systems. - ANSI/AAMI/ISO 11607-2:2006; Packaging for terminally sterilized medical devices Part 2: Validation requirements for forming, sealing, and assembly processes. 33
Thank You Questions? 34