12/15-Lead ECG Protocol approved October 2005 revised October 2008



Similar documents
The Basics of 12 Lead EKG s

Scott Hubbell, MHSc, RRT-NPS, C-NPT, CCT Clinical Education Coordinator/Flight RRT EagleMed

Podcast with Dr. Kossick

THE SOCIETY FOR. Recording a standard 12-lead electrocardiogram. An Approved Methodology

Hypoglycemia (Technician-Paramedic) Protocol revised October 2008

ELECTROCARDIOGRAPHY (I) THE GENESIS OF THE ELECTROCARDIOGRAM

12-Lead EKG Interpretation. Judith M. Haluka BS, RCIS, EMT-P

12 Lead ECGs: Ischemia, Injury & Infarction Part 2

Cardiac Arrest Pediatric Ventricular Fibrillation / Pulseless Ventricular Tachycardia Protocol revised October 2008

12 LEAD ECG ACQUISITION & TRANSMISSION FOR BLS PROVIDERS

Policy & Procedures. I.D. Number: 1142

Cardiac Arrest - Ventricular Fibrillation / Pulseless Ventricular Tachycardia Protocol revised October 2008

Biology 347 General Physiology Lab Advanced Cardiac Functions ECG Leads and Einthoven s Triangle

Coronary Artery Disease leading cause of morbidity & mortality in industrialised nations.

Activity 4.2.3: EKG. Introduction. Equipment. Procedure

BIPOLAR LIMB LEADS UNIPOLAR LIMB LEADS PRECORDIAL (UNIPOLAR) LEADS VIEW OF EACH LEAD INDICATIVE ECG CHANGES

Is it really so? : Varying Presentations for ACS among Elderly, Women and Diabetics. Yen Tibayan, M.D. Division of Cardiovascular Medicine

Operating Instructions

Tips and Tricks to Demystify 12 Lead ECG Interpretation

HOW TO USE THE LIFEPAK 1000 DEFIBRILLATOR 3

ACLS Chapter 3 Rhythm Review Instructor Lesson Plan to Accompany ACLS Study Guide 3e

Electrophysiology Introduction, Basics. The Myocardial Cell. Chapter 1- Thaler

Understanding the Electrocardiogram. David C. Kasarda M.D. FAAEM St. Luke s Hospital, Bethlehem

HEART MONITOR TREADMILL 12 LEAD EKG

Exercises for the Hip

ECG made extra easy. medics.cc

R Series Quick Reference Guide

Instytut Fizyki Doświadczalnej Wydział Matematyki, Fizyki i Informatyki UNIWERSYTET GDAŃSKI

Project 4.2.1: Heart Rate

Evaluation copy. Analyzing the Heart with EKG. Computer

Adult & Paediatric Resting Electrocardiography (ECG) Cardiac Sciences

Clinical Care Program

6/5/2014. Objectives. Acute Coronary Syndromes. Epidemiology. Epidemiology. Epidemiology and Health Care Impact Pathophysiology

Strengthening Exercises - Below Knee Amputation

Electrocardiography I Laboratory

Common Regional Nerve Blocks Quick Guide developed by UWHC Acute Pain Service Jan 2011

An ECG Primer. Quick Look. I saw it, but I did not realize it. Elizabeth Peabody

Osama Jarkas. in Chest Pain Patients. STUDENT NAME: Osama Jarkas DATE: August 10 th, 2015

Exercise, the Electrocardiogram, and Peripheral Circulation

RISK STRATIFICATION for Acute Coronary Syndrome in the Emergency Department

Section 8: Clinical Exercise Testing. a maximal GXT?

Greater Hartford EMS Education STEMI Workbook

Preventing Falls. Strength and balance exercises for healthy ageing

INTRODUCTORY GUIDE TO IDENTIFYING ECG IRREGULARITIES

123 Main St NY, New York ph: (202) fax: (202)

Low Back Pain: Exercises

The new generation in ECG interpretation

Body cavities. Body Planes

Clinical Reference Guide

ST Segment Elevation Nothing is ever as hard (or easy) as it looks

Diagnosis Code Crosswalk : ICD-9-CM to ICD-10-CM Cardiac Rhythm and Heart Failure Diagnoses

Electrocardiogram (ECG) Monitoring System using Bluetooth technology

Procedure 17: Cardiopulmonary Resuscitation

Deriving the 12-lead Electrocardiogram From Four Standard Leads Based on the Frank Torso Model

BASIC STANDARDS FOR RESIDENCY TRAINING IN CARDIOLOGY

General Guidelines. Neck Stretch: Side. Neck Stretch: Forward. Shoulder Rolls. Side Stretch

Electrocardiogram and Heart Sounds

THE INFLUENCE OF WALL PAINTING ON SHOULDER MUSCLE ACTIVITY AND HORIZONTAL PUSH FORCE

A Patient s Guide to Diffuse Idiopathic Skeletal Hyperostosis (DISH)

HOW TO CHECK YOUR LYMPH NODES

UW MEDICINE PATIENT EDUCATION. Aortic Stenosis. What is heart valve disease? What is aortic stenosis?

Stretching the Low Back THERAPIST ASSISTED AND CLIENT SELF-CARE STRETCHES FOR THE LUMBOSACRAL SPINE

EMS SKILL CARDIAC EMERGENCY: AUTOMATED EXTERNAL DEFIBRILLATION (AED)

Myofit Massage Therapy Stretches for Cycling

2 ECG basics. Leads and planes. Leads. Planes. from different perspectives, which are called leads and planes.

Cardiac Rehab Program: Stretching Exercises

IMGPT: Exercise After a Heart Attack N. RICHMOND ST (Located next to Fleetwood HS) Why is exercise important following a heart

Coccydynia. (Coccyx Pain) Information for patients. Outpatients Physiotherapy Tel:

Objectives. The ECG in Pulmonary and Congenital Heart Disease. Lead II P-Wave Amplitude during COPD Exacerbation and after Treatment (50 pts.

Acute Coronary Syndrome. What Every Healthcare Professional Needs To Know

MET: Posterior (backward) Rotation of the Innominate Bone.

ECG Findings. IV Access. 12 Lead Interpretation: STEMI and NSTEMI. ACLS Acute Coronary Syndrome Chest Pain Suggestive of Ischemia.

iworx Sample Lab Experiment HN-2: Stretch Receptors and Reflexes

Pilates for Kyphosis A BASI Pilates program designed to help correct thoracic kyphosis

Living Room Bodyweight Workout Week 1 March or jog in place for 1 min to increase heart rate and lubricate joints.

Introduction to Electrocardiography. The Genesis and Conduction of Cardiac Rhythm

Otago Exercise Program

Name: Age: Resting BP: Wt. kg: Est. HR max : 85%HR max : Resting HR:

Basic Stretch Programme 3. Exercise Circuit 4

Exercises after breast or upper body lymph node surgery

CHAPTER 3: BACK & ABDOMINAL STRETCHES. Standing Quad Stretch Athletic Edge (650)

Anatomy & Physiology 120. Lab #7 Muscle Tissue and Skeletal Muscles

Pre - Operative Rehabilitation Program for Anterior Cruciate Ligament Reconstruction

Section III. Heart Mate II (2 nd Generation) III.41.1 EMT-BASIC EMT-INTERMEDIATE EMT-PARAMEDIC

Anatomy and Physiology: Understanding the Importance of CPR

Knee Conditioning Program. Purpose of Program

NEEDLE THORACENTESIS Pneumothorax / Hemothorax

Improving ECG Quality

A PATIENT S GUIDE TO CARDIAC CATHETERIZATION

Tired, Aching Legs? Swollen Ankles? Varicose Veins? An informative guide for patients

Signal-averaged electrocardiography late potentials

Don t. Hamstrings. Calf Muscles. both legs 2-3 times. stretch is felt in the back of the calf. Repeat with both legs 2-3 times.

The Thorax - Anterior and Lateral Chest Wall

The Power Plate is the innovative, time saving and results driven way to improve your fitness and well being.

Mission: Lifeline Recommendations for Criteria for STEMI Systems of Care

X-ray (Radiography), Chest

Take a few minutes for yourself and incorporate some Office Yoga into your daily routine.

Aquatic Exercises: Upper Body Strengthening

Fact sheet Exercises for older adults undergoing rehabilitation

Info. from the nurses of the Medical Service. LOWER BACK PAIN Exercise guide

Transcription:

approved October 2005 revised October 2008 Preamble Rapid diagnosis of acute myocardial infarction is essential to initiating appropriate treatment and improving outcomes. In selected practice environments prehospital ECG s may facilitate emergency department treatment or may facilitate primary triage to appropriate cardiac care centers. Requirements 1) Fully licensed Technician-Paramedic. 2) Certification in 12/15-lead protocol by the medical director. Indications Chest pain suggestive of cardiac ischemia (See appendix) Procedure 1. The EMS personnel identifies the medical indications for 12 lead ECG and explains the procedure to the patient. 2. The EMS personnel ensures that the 12 lead ECG patient cable includes both the 4 lead and 6 lead sections. 3. The EMS personnel inserts the 12 lead ECG patient cable into the ECG connector on the defibrillator/monitor. 4. The EMS personnel snaps ECG electrodes onto the 10 lead patient cable. 5. The EMS personnel places the patient supine with the head slightly elevated and ensures that the patient's arms and legs are well supported. 6. The EMS personnel exposes the chest and limb areas where electrodes will be placed.

7. The EMS personnel places the electrodes of the limb leads as follows: Right arm and left arm leads Right and left shoulder Right leg and left leg leads - Preferably midway between the ankle and the knee In situations of muscular artifact on limb leads, the EMS personnel moves the leg electrodes into the position used for cardiac arrest (right and left hip). 8. The EMS personnel places the precordial or chest leads as follows: NOTE: The EMS personnel can refer to the lead placement diagram for chest lead placement. V1 - In the 4th intercostal space at the right sternal border, not more than 3 cm from the sternal border V2 - In the 4th intercostal space at the left sternal border, not more than 3 cm from the sternal border V3 - In a straight line, midway between V2 and V4 (locate V4 first) V4 - In the 5th intercostal space in the left midclavicular line V5 - At the same level and in a straight line from V4 in the left anterior axillary line V6 - At the same level and in a straight line from V4 and Vs in the left midaxillary line 9. When placing lead V4 on a female, the EMS personnel lifts the breast up and places the electrode on the chest wall, not on top of the breast. 10. The EMS personnel presses the "12 lead" button on the front panel. The screen may display "enter patient age" at which point the EMS personnel enters the patient's age using the selector knob or uses the default age of 40 years. The monitor may require the patient's age because the criterion for ECG interpretation is different for patients less than 17 years than adult patients. 11. The EMS personnel asks the patient to keep still with their muscles as relaxed as possible and to breath normally during the 12 lead ECG recording which takes approximately 10 seconds. 12. The EMS personnel is aware that when the "12 lead" button is pressed that the monitor automatically adjusts to diagnostic mode and standardizes the ECG to record a 1 mv signal. The monitor records all leads simultaneously and "Acquiring ECG" appears on the display during the recording.

13. If there are technical problems, e.g., lead off or artifact (muscular or electrical) during the 12 lead ECG recording, the monitor will automatically indicate on the display that there is a problem. The EMS personnel checks the contact of the ECG electrodes with the patient's skin, checks that all leads are connected, and attempts to reduce muscular or electrical interference as much as possible. If a 12 lead ECG is required despite the presence of artifact, the EMS personnel must press the "12 lead" button again to obtain a 12 lead recording when the display indicates ECG technical problems or ECG noise due to artifact. NOTE: The resulting 12 lead ECG will not provide an interpretive diagnosis. If patient motion and resulting artifact ceases, the 12 lead will resume as normal. 14. If certified to interpret 12-lead ECG s, the EMS personnel checks the 3 channel 12 lead ECG (tracing) for its quality and ECG evidence of acute myocardial infarction (AMI). 15. The EMS personnel is aware that the 15 lead ECG is used to enhance the traditional 12 lead ECG with leads reflecting the right ventricle and posterior heart. The indications for 15 lead ECG are: - Cardiac chest pain greater than 15 minutes with normal 12 lead ECG - 12 lead ECG showing ST depression in V1 and V2 with prominent R waves - 12 lead ECG showing signs of acute inferior Ml 16. To attain the additional 3 leads for a 15 lead ECG, the EMS personnel removes the patient cables from precordial electrode positions V4, V5 and V6, leaving V1, V2, V3, and the limb leads intact. The EMS personnel then places three ECG electrodes in the following positions: V4R - 5th intercostal space right midclavicular line V8-5th intercostal space midscapula on the left V9-5th intercostal space immediately left of the vertebral column The EMS personnel attaches the patient cables for V1, V2, and V3 as follows: V4 cable for V4R position \/5 cable for V8 position \/6 cable for V9 position 17. The EMS personnel presses the 12 lead ECG button to obtain the recording. 18. The EMS personnel labels the second 12 lead recording, identifying V4 as the V4R tracing, V5 as the V8 tracing, and V6 as the V9 tracing.

Documentation Requirements 1. The EMS personnel writes or inputs the patient name and incident number on the 12 lead ECG tracing. NOTE: The date and time of 12 lead ECG is automatically indicated on the tracing. The EMS personnel documents the ID number of the 12 lead tracing on the patient care report, in case the tracing becomes detached from the patient care report. Certification Requirements 1. Attend in-depth classes and lectures on pulmonary edema, including anatomy, physiology, and pathophysiology of the cardiorespiratory system. 2. Attend in-depth classes in the operation of the appropriate monitor and performance of 12/15-lead ECG. 3. Pass a written and practical examination. Recertification Requirements 1. Review class and recertification is done every 12 months. 2. A record will be kept to document all cases where this protocol is used. Decertification 1. Decertification is at the discretion of the Medical Director or the Provincial Medical Director, Emergency Medical Services, Manitoba Health & Healthy Living. Quality Assurance Requirements 1. Appropriate quality assurance policies must be in place. The Medical Director or designate must review all instances where this protocol is used. As a minimum, the following must be assessed: i) appropriateness of implementation ii) adherence to protocol iii) any deviation from the protocol iv) corrective measures taken, if indicated 2. Yearly statistics for protocol use compiled and forwarded to Emergency Medical Services, Manitoba Health & Healthy Living.

Appendix Factors indicating that chest pain is consistent with cardiac ischemia - dull central chest pain - radiation to arms/neck/jaw - dyspnea - diaphoresis - nausea/vomiting Note: Patients may present with atypical symptoms different from those listed above. Absence of the above symptoms does not rule out cardiac ischemia.