1 THE SOCIETY FOR Recording a standard 12-lead electrocardiogram An Approved Methodology February 2010 Review Date: February 2013
2 Introduction Practice varies across the Health Service predominantly due to historical circumstances, personal preferences and resources. Guidelines aim to reduce such variation by gathering and presenting the evidence for best practice, thereby improving the equity and quality of services. This guideline has been developed by the Society for Cardiological Science and Technology (SCST) and has been approved by the British Cardiovascular Society. The guideline is based on existing research and on the knowledge and experience of practitioners. Where research based evidence is not yet available consensus decisions have been made. As the research base is constantly growing, SCST will regularly review the guideline to ensure that it continues to reflect current best practice. Audit is a recognised way of changing and improving practice and will be used to revise, modify or update subsequent guidelines. Appendix C of the guideline contains a useful checklist intended for use as an audit tool. This checklist can be used to assess how well the guideline is being followed. Additionally, the audit checklist may be used to ensure that any local guidelines developed from this consensus document meet the essential requirements. Acknowledgements SCST is grateful to everybody who has contributed to and commented on the many draft versions of these guidelines. 2
3 Guidelines for recording a standard 12-lead electrocardiogram Equipment Electrocardiograph that meets or exceeds the requirements of IEC (2003) *. The device should be pre-programmed in accordance with American Heart Association (AHA) specifications 1,2. The low-frequency filter should be set no higher than 0.05Hz to avoid distortion of the ST segment ** The high frequency filter should be set no lower than 100Hz to prevent loss of high frequency information 3 Recommended recording bandwidths pre-stored in device set-up Auto mode 0.05Hz 100Hz or 0.5Hz 100Hz ** Manual or real time mode 0.05Hz 100Hz only NB. Mains (50Hz) filter - off When front panel filter buton selected Auto mode 0.5 Hz 40 Hz Manual or real time mode 0.05 Hz 40 Hz Disposable tab electrodes that meet or exceed the requirements of AAMI EC12-00 (2000) *** Skin preparation equipment (for example, razor, abrasive tape) * IEC (2003) - Particular requirements for safety, including essential performance, of recording and analysing single channel and multichannel electrocardiographs. This document establishes minimum safety and performance requirements for electrocardiographic (ECG) systems that are intended for use in the analysis of rhythm and of detailed morphology of complex cardiac complexes. ** Digital filter design allows for the low-frequency filter to be raised to 0.5Hz when recording in auto mode. However, ST-segment distortion may occur when this seting is used in manual mode. Fixing the low-frequency setting at 0.05Hz in the pre-set should prevent this error occurring 3. *** AAMI EC12-00 (2000) - This standard establishes minimum labeling, safety, and performance requirements for disposable electrodes used for diagnostic electrocardiography. 3
4 Training The operator should be competent in both the use of the electrocardiograph and in the recording of an electrocardiogram (ECG). Ideally, this will be in the form of a recognised qualification, awarded by the Society for Cardiological Science & Technology (Certificate or Associate membership) following the successful completion of the appropriate examination. As a minimum, personnel should provide evidence of having undertaken a practical assessment process that has been carried out by a competent practitioner holding the above award. It is essential that competence in the recording of an electrocardiogram is maintained and this should be reviewed on an annual basis. Procedure Prior to undertaking the procedure the following should be checked:- That electrocardiograph is safe and ready to use (date & time settings are correct). The patient area is clean and tidy. There is sufficient paper, electrodes, razors and skin preparation equipment. The identity of the patient should be confirmed and crosschecked with the request. Patient Preparation 2 Unrestricted access to the skin in the chest area, arms and lower legs is required to allow for correct placement of the electrodes. Once the electrodes are positioned and the connecting wires are appropriately attached, the patient should be covered with a gown to preserve his/her dignity during the procedure. Patients may feel uncomfortable about being touched on their upper torso. The ECG procedure requires sensitivity. Operators must make every effort to respect the sensitivities of patients and minimise their embarrassment. Operators must adhere to the Trust's chaperone policy and ensure that patients are made aware of the policy. The appearance of the electrocardiogram can vary depending on the position of the body at the time. Therefore, a recumbent position is recommended for the patient. Any variation to this position should be noted on the printed ECG. The bed or couch should be large enough to support the body and limbs. In order to achieve clinically accurate recordings with minimal artefact it is essential for patients to be comfortable and relaxed. This may be achieved through optimising the environmental conditions and providing sufficient explanation to the patient and / or carer. 4
5 Skin preparation Skin preparation is often required to help produce an artefact-free and accurate ECG. Various methods are available, all of which are designed to minimise the skin-to-electrode impedance. For example, The removal of chest hair may be required to ensure adequate contact with the skin. Verbal consent should be obtained from the patient and a clean razor used which should be disposed of in a sharps bin immediately afterwards. Exfoliation may be required and should be undertaken with very light abrasion using either a paper towel, gauze swab or proprietary abrasive tape designed specifically for this purpose. On occasions the skin may require cleansing. A variety of methods exist ranging from washing with mild soap to cleaning with an alcohol wipe. However, care must be taken in patients with sensitive or broken skin. Electrode placement The following electrode sites should be correctly identified and the placement of the electrodes must conform to AHA recommendations 1,2. (NB. Each lead wire is generally colour coded to aid identification. However, the colour may vary depending on manufacturer. The colours listed are consistent with IEC (European) recommendations). Limb leads Evidence exists to demonstrate that moving the limb lead attachments away from the distal limbs alters the appearance of the ECG. This variation can invalidate the use of such recordings for many diagnostic purposes 4,5. To ensure consistency between recordings it is recommended that the electrodes are attached to both arms and legs, slightly proximal to the wrist and ankle. It is imperative that recordings from other sites are labelled accordingly so that the results are not confused with those obtained from standard sites. Right arm limb lead (RA, red) Left arm limb lead (LA, yellow) Left leg limb lead (LL, green) Right leg limb lead (RL, black) right forearm, proximal to wrist left forearm, proximal to wrist left lower leg, proximal to ankle right lower leg, proximal to ankle 5
6 Precordial (chest) leads Variation in the placement of the precordial electrodes produces diagnostically significant differences in the ECG 6. Previous studies have demonstrated that V1 and V2 are frequently placed too high and the left lateral leads too low 7,8. The correct anatomical positions for placing the precordial leads have been defined 9,10,11 and should always be used (see figure below). If any deviation from these positions is necessary then this must be clearly labelled on the printed ECG. The centre of the active surface area of the electrode should be aligned with the relevant anatomical landmark. Electrode V1 (C1) V2 (C2) V3 (C3) V4 (C4) V5 (C5) V6 (C6) Position Fourth intercostal space at the right sternal edge Fourth intercostal space at the left sternal edge Midway between V2 and V4 Fifth intercostal space in the mid-clavicular line Left anterior axillary line at same horizontal level as V4 Left mid-axillary line at same horizontal level as V4 & V5 Note: This is not a rib space. Do not include when counting. V2 Fourth intercostal space Immediately left of sternum V3 Mid-way between V2 and V4 V1 Fourth intercostal space Immediately right of sternum V5 Same horizontal level as V4 Anterior axillary line V4 Fifth intercostal space Mid clavicular line V6 Same horizontal level as V4 Mid-axillary line at right angles to the mid-clavicular line Standard ECG chest electrode positions CJE05/06 6
7 Locating chest electrode positions Care should be taken when counting the intercostal spaces down from the clavicle that the small space between the clavicle and the first rib is not mistaken for the first intercostal space. In order to avoid this common error the sternal angle (angle of Louis) should be used as the main reference point. This anatomical landmark denotes the position of the sternal angle at the manubriosternal joint. To locate the sternal angle, a finger should be run down the sternum, from the sternal notch at the top until a bony horizontal ridge, the sternal angle, is met. With the finger on this ridge, sliding down and to the side will locate the second intercostal space. Then count down to the third and fourth space. Locate the very edge of the sternum and place V1 there. Repeat this procedure on the left side to correctly position V2. Next the position for V4 should be located. This should be placed in the fifth intercostal space in line with the mid point of the clavicle. Once the V4 electrode has been correctly placed then the location for V3 can be identified, directly mid-way between V2 and V4. V5 and V6 are then positioned, taking care not to follow the line of the ribs, but to follow a horizontal line from V4. V5 is placed in line with the anterior axilla and V6 in line with the mid-axilla. The centre of the active surface area of the electrode should be aligned with the relevant anatomical landmark, i.e. the middle of the V1 electrode should be placed at the right sternal edge. When recording an ECG from female patients by convention the lateral chest electrodes (V4, V5 and V6) are placed beneath the left breast. Whilst it is acknowledged that there is emerging evidence to support the positioning of these electrodes over the breast without any attenuation of the signal 12,13 there are insufficient published data to support alteration of the widely adopted technique of placing V4 to V6 under the left breast. 7
8 Recording In order to record a good quality ECG the patient must be relaxed and comfortable. If these conditions are not satisfied the ECG will record somatic muscle potentials as well as cardiac activity. Such interference will make the ECG more difficult to interpret. Some patients cannot relax fully because of pain from arthritis or may have a neurological condition such as Parkinson s disease causing tremor. Make them as comfortable as possible and annotate the trace with an appropriate explanation. On occasions it may be necessary to adapt the recommended ECG recording techniques. For example, wheelchair bound patients may need to remain in their chair during the recording process. Any variations to standard recording techniques must be highlighted on the trace, e.g. ECG recorded whilst patient in wheelchair. Patient details (name and a second unique identifier such as hospital number or date of birth) should be entered into the ECG machine, after it has been checked with the patient directly (or someone speaking on their behalf). Before recording the ECG, check that the patient s limbs are stil and appear relaxed. If the patient has clenched fists or stiff arms, or is moving his / her fingers, it will not be possible to obtain a high quality ECG. Press the appropriate button on the machine to initiate a recording (usualy labeled start or auto ). A 12-lead ECG and rhythm strip should be recorded at 25mm/s with a gain setting of 10mm/mV 1,2. The filter button should not be selected for this initial recording. If, despite efforts to relax the patient and make them comfortable, there is somatic muscle interference on the ECG, switch on the filter and repeat the recording. Use of the filter should be clearly identified on the final ECG. NB. The filter will reduce the interference. However, as it will also distort the ECG 3 it should only be used when absolutely necessary (after all attempts to eliminate the interference have failed). 8
9 If the ECG complexes are of high voltage then the gain should be adjusted (5mm/mV) to enable them to be measured accurately. Any alteration in the gain settings should be clearly marked on the tracing. If the rhythm is noted to be irregular then an additional rhythm strip (traditionally lead II) should be recorded for a minimum of 10 seconds. Any changes on the ECG that might require urgent medical attention should be identified and advice sought from a senior member of staff if necessary. If the patient has any cardiac symptoms at the time of recording, such as chest pain or palpitations then this should be noted on the tracing and brought to the immediate attention of a senior member of staff. If the ECG is technically correct and of good quality, ensure that it is fully and correctly labelled (patient identification information, relevant clinical details) then remove all of the electrodes from the patient and dispose as clinical waste. 9
10 Appendix A The recommended approach to recording the 12-lead ECG in the patient with dextrocardia The term dextrocardia refers to any situation where the heart is located within the right side of the chest rather than the left side. It may be associated with the condition situs inversus where all of the patient's organs are in a mirrorimage position. When the P wave in lead I is inverted (P axis > 90 ) the diagnosis of dextrocardia may be suspected, provided there is no technical error, such as reversal of the right and left arm connections. Poor R wave progression across the chest leads will support this ECG diagnosis. A second ECG should be recorded with the chest electrodes positioned on the right side of the chest using the same intercostal spacing and anatomical landmarks (right sided). This approach wil provide a true ECG representation. The limb lead complexes will continue to appear inverted, demonstrating the abnormal location of the heart. However, the re-positioned chest leads will now show the appropriate R wave progression. Clear annotation describing the repositioned electrodes should be handwritten onto the hard-copy (For example, V3R, V4R etc.) Both of the annotated ECGs should be retained for inclusion in the case notes. Note Swapping of the right and left limb leads wil normalize the appearance of the limb leads. If this approach is preferred it is imperative that the ECG is very clearly annotated to prevent the possibility of the dextrocardia being overlooked. 10
11 Appendix B Electrocardiography in childhood A patient and gentle approach is necessary for the recording of an ECG from infants (first year of life), and children, in order that an artefact-free electrocardiogram is obtained. With a few exceptions the recording method is closely similar to that described for adults. If possible, the recording is made with the subject supine, but the sitting position is allowed if this will prevent restlessness or distress. The choice of leads to be recorded will sometimes be determined by local preferences. The four limb leads are attached as previously described. Where possible, six precordial leads, V4R, V1, V2, V4, V5, and V6 are recorded. V3R and V7 may be recorded, by special request. (V3R and V4R are right sided leads placed in a mirror image position to V3 and V4. V7 is a posterior lead placed in the posterior axillary line at the same horizontal level as V4). For recording purposes a paper speed of 25mm/s is standard. The younger the child the faster the heart rate; where necessary, for accurate analysis of the electrocardiogram, a paper speed of 50mm/s is recommended. In terms of voltage, the standard is a vertical deflection on the electrocardiogram of 10mm representing 1mV. When QRS complexes in simultaneously recorded leads would be so large as to overlap, the deflection should be halved so that a deflection of 5mm would represent 1mV. Normal values for ECG intervals and waves, most divergent from adult values in the very young, gradually converge with physical growth. When an electrocardiogram is presented for analysis (diagnosis), any variation from the usual recording procedure must be clearly and indelibly stated. 11
12 Appendix C: AUDIT CHECKLIST Filter Settings 1. In Auto mode is the machine preset to a recording bandwidth of either: 0.05Hz 100Hz or 0.5Hz 100Hz? 2. In Manual or real time mode is the machine preset to a recording bandwidth of 0.05Hz 100Hz? 3. Is the mains filter (50Hz) set to of? Training 1. Has the person performing the ECG recording been appropriately assessed in this technique? Electrode placement 1. Has appropriate skin preparation been carried out prior to recording the ECG? 2. Are electrodes positioned as per the guidelines? 3. If electrodes are not positioned as per guidelines is the ECG annotated with an appropriate explanation? N/A Recording 1. Are the patient details present and correct? 2. Is the ECG free of artefact? Yes No 3. If artefact is present has an explanation been written on the ECG? N/A 4. Has the mains filter been used? 5. If the front panel filter has been selected has an explanation been writen on the ECG? N/A 6. Have the appropriate gain settings been used? 7. If the rhythm is irregular has an additional rhythm strip been recorded? N/A 12
13 References 1. Recommendations for standardization of leads and of specifications for instruments in electrocardiography and vectorcardiography. Kossman CE et al. Circulation 1967;35: Recommendations for standardization of leads and of specifications for instruments in electrocardiography and vectorcardiography. Pipberger HV et al. Circulation 1975;52; Recommendations for standardization and specifications in automated electrocardiography: Bandwidth and digital signal processing. Bailey et al. Circulation 1990;81; Evaluation of changes in standard electrocardiographic QRS waveforms recorded from activity-compatible proximal limb lead positions. Pahlm O et al. Am J Cardiol 1992;69: Invalidation of the resting electrocardiogram obtained via exercise electrode sites as a standard 12-lead recording. Sevilla DC et al. Am J Cardiol 1989;63: Positional and respiratory changes in precordial lead patterns simulating acute myocardial infarction. August T et al. Am Heart J 1958;55: Variability of precordial electrode placement during routine electrocardiography. Wenger W, Kligfield P. J Electrocardiol 1996;29: Electrocardiographic poor R wave progression III: the normal variant. Zema MJ et al. J Electrocardiol 1980;13: Joint Recommendations of the American Heart Association and the Cardiac Society of Great Britain and Ireland. Standardization of precordial leads. Am Heart J 1938;15: Committee of the American Heart Association for the Standardization of precordial leads. Supplementary report. Am Heart J 1938;15: Committee of the American Heart Association for the Standardization of precordial leads. Second supplementary report. J Am Med Assoc 1943;121: False positive ECG reports of anterior myocardial infarction in women. Colaco R et al. J Electrocardiol 2000:33; Precordial electrode placement in women. Macfarlane PW et al. Neth Heart J 2003;11:
Adult & Paediatric Resting Electrocardiography (ECG) Cardiac Sciences
Document Number # QH-GDL-387:2012 Adult & Paediatric Resting Electrocardiography (ECG) Cardiac Sciences Custodian/Review Officer: Chief Allied Health Officer Version no: 1.0 Applicable To: All health practitioners
Podcast with Dr. Kossick
Podcast with Dr. Kossick Interviewed by Western Carolina University Graduate Anesthesia Student Kristin Andrejco From the Head of the Bed [email protected] December 5, 2014 (33 min) EKG Lead
Clinical Guidelines by Consensus Recording a standard 12-lead electrocardiogram
Clinical Guidelines by Consensus Recording a standard 12-lead electrocardiogram An approved methodology by the Society for Cardiological Science & Technology (SCST) Issue Date: June 2014 Review Date: June
Policy & Procedures. I.D. Number: 1142
Policy & Procedures Title: CARDIAC (ECG) MONITORING I.D. Number: 1142 Authorization: [X] SHR Nursing Practice Committee Source: Date Revised: November 2012 Date Effective: October 2004 Scope: SHR - Acute
Understanding the Electrocardiogram. David C. Kasarda M.D. FAAEM St. Luke s Hospital, Bethlehem
Understanding the Electrocardiogram David C. Kasarda M.D. FAAEM St. Luke s Hospital, Bethlehem Overview 1. History 2. Review of the conduction system 3. EKG: Electrodes and Leads 4. EKG: Waves and Intervals
Biology 347 General Physiology Lab Advanced Cardiac Functions ECG Leads and Einthoven s Triangle
Biology 347 General Physiology Lab Advanced Cardiac Functions ECG Leads and Einthoven s Triangle Objectives Students will record a six-lead ECG from a resting subject and determine the QRS axis of the
HOW TO USE THE LIFEPAK 1000 DEFIBRILLATOR 3
3 How to Use the LIFEPAK 1000 Defibrillator HOW TO USE THE LIFEPAK 1000 DEFIBRILLATOR 3 This section provides an overview of information and instructions for using the LIFEPAK 1000 defibrillator. Modes
Deriving the 12-lead Electrocardiogram From Four Standard Leads Based on the Frank Torso Model
Deriving the 12-lead Electrocardiogram From Four Standard Leads Based on the Frank Torso Model Daming Wei Graduate School of Information System The University of Aizu, Fukushima Prefecture, Japan A b s
Evaluation copy. Analyzing the Heart with EKG. Computer
Analyzing the Heart with EKG Computer An electrocardiogram (ECG or EKG) is a graphical recording of the electrical events occurring within the heart. In a healthy heart there is a natural pacemaker in
BIPOLAR LIMB LEADS UNIPOLAR LIMB LEADS PRECORDIAL (UNIPOLAR) LEADS VIEW OF EACH LEAD INDICATIVE ECG CHANGES
BIPOLAR LIMB LEADS Have both a distinctive positive and negative pole. Lead I LA (positive) RA (negative) Lead II LL (positive) RA (negative) Lead III LL (positive) LA (negative) UNIPOLAR LIMB LEADS Have
Electrophysiology Introduction, Basics. The Myocardial Cell. Chapter 1- Thaler
Electrophysiology Introduction, Basics Chapter 1- Thaler The Myocardial Cell Syncytium Resting state Polarized negative Membrane pump Depolarization fundamental electrical event of the heart Repolarization
RESTING 12-LEAD ECG ELECTRODE PLACEMENT AND ASSOCIATED PROBLEMS. Prof. Macfarlane and Dr. Coleman, SCST Update 1995
RESTING 12-LEAD ECG ELECTRODE PLACEMENT AND ASSOCIATED PROBLEMS. Prof. Macfarlane and Dr. Coleman, SCST Update 1995 P.W. Macfarlane - Professor in Medical Cardiology, University of Glasgow Dept, of Medical
Improving ECG Quality
Improving ECG Quality Application Note Introduction Good ECG signal quality it is something every clinician strives to secure every time they monitor a patient s heart. This application note will look
ELECTROCARDIOGRAPHY (I) THE GENESIS OF THE ELECTROCARDIOGRAM
ELECTROCARDIOGRAPHY (I) THE GENESIS OF THE ELECTROCARDIOGRAM Scridon Alina, Șerban Răzvan Constantin 1. Definition The electrocardiogram (abbreviated ECG or EKG) represents the graphic recording of electrical
Monitoring EKG. Evaluation copy
Monitoring EKG Computer 28 An electrocardiogram, or EKG, is a graphical recording of the electrical events occurring within the heart. A typical EKG tracing consists of five identifiable deflections. Each
Exercise, the Electrocardiogram, and Peripheral Circulation
iworx Physiology Lab Experiment Experiment HH-3 Exercise, the Electrocardiogram, and Peripheral Circulation Note: The lab presented here is intended for evaluation purposes only. iworx users should refer
Activity 4.2.3: EKG. Introduction. Equipment. Procedure
Activity 4.2.3: EKG The following is used with permission of Vernier Software and Technology. This activity is based on the experiment Analyzing the Heart with EKG from the book Human Physiology with Vernier,
Scott Hubbell, MHSc, RRT-NPS, C-NPT, CCT Clinical Education Coordinator/Flight RRT EagleMed
Scott Hubbell, MHSc, RRT-NPS, C-NPT, CCT Clinical Education Coordinator/Flight RRT EagleMed Identify the 12-Lead Views Explain the vessels of occlusion Describe the three I s Basic Interpretation of 12-Lead
The Basics of 12 Lead EKG s
EMS Solutions Presents The Basics of 12 Lead EKG s NOTICE: You DO NOT Have the Right to Reprint or Resell this Publication. However, you MAY give this report away, provided you do not change or alter the
Electrocardiography I Laboratory
Introduction The body relies on the heart to circulate blood throughout the body. The heart is responsible for pumping oxygenated blood from the lungs out to the body through the arteries and also circulating
Marquette 12SL Algorithm. Connected Clinical Excellence
Marquette 12SL Algorithm Connected Clinical Excellence Clinical decision support for your ECG Since its introduction in 1980 the Marquette TM 12SL ECG analysis program has been consistently refined and
12-Lead EKG Interpretation. Judith M. Haluka BS, RCIS, EMT-P
12-Lead EKG Interpretation Judith M. Haluka BS, RCIS, EMT-P ECG Grid Left to Right = Time/duration Vertical measure of voltage (amplitude) Expressed in mm P-Wave Depolarization of atrial muscle Low voltage
Instytut Fizyki Doświadczalnej Wydział Matematyki, Fizyki i Informatyki UNIWERSYTET GDAŃSKI
Instytut Fizyki Doświadczalnej Wydział Matematyki, Fizyki i Informatyki UNIWERSYTET GDAŃSKI 12 Experiment 12 : Examination of the heart using ECG and PCG I. Background theory. 1. Construction and functioning
The Electrocardiogram (ECG)
The Electrocardiogram (ECG) Preparation for RWM Lab Experiment The first ECG was measured by Augustus Désiré Waller in 1887 using Lippmann's capillary electrometer. Recorded ECG: http://www.youtube.com/watch_popup?v=q0jmfivadue&vq=large
An ECG Primer. Quick Look. I saw it, but I did not realize it. Elizabeth Peabody
4 An ECG Primer Quick Look Cardiac Monitoring System - p. 64 ECG Paper - p. 73 Lead Polarity and Vectors - p. 77 Basic ECG Components - p. 79 Heart Rate and Pulse Rate - p. 91 Summary - p. 94 Chapter Quiz
ACLS Chapter 3 Rhythm Review Instructor Lesson Plan to Accompany ACLS Study Guide 3e
ACLS Chapter 3 Rhythm Review Lesson Plan Required reading before this lesson: ACLS Study Guide 3e Textbook Chapter 3 Materials needed: Multimedia projector, computer, ACLS Chapter 3 Recommended minimum
Lancashire Teaching Hospitals Cardio-Respiratory Department. Lead systems. Paula Hignett
Lancashire Teaching Hospitals Cardio-Respiratory Department Lead systems Paula Hignett Objectives Understand the terminology and theory of the 12 views of the heart with reference to: Unipolar and Bipolar
Range of Motion. A guide for you after spinal cord injury. Spinal Cord Injury Rehabilitation Program
Range of Motion A guide for you after spinal cord injury Spinal Cord Injury Rehabilitation Program This booklet has been written by the health care providers who provide care to people who have a spinal
Manual 5 Electrocardiography Assessment Visit 5
Manual 5 Electrocardiography Assessment Visit 5 April 1, 2011- Version 2 Study website - http://www.cscc.unc.edu/aric/ Electrocardiography Assessment Table of Contents 1. OVERVIEW... 3 2. FIELD CENTER
Project 4.2.1: Heart Rate
Project 4.2.1: Heart Rate Introduction Even before you were born, one of the first things your doctor did when you went for an office visit was listen to your heart. Your heart rate, the number of times
Diagnostic tests for syncope
Diagnostic tests for syncope Working together with individuals, families and medical professionals to offer support and information on syncope and reflex anoxic seizures www.stars.org.uk Registered Charity
The new generation in ECG interpretation
The new generation in ECG interpretation Philips DXL ECG Algorithm, Release PH100B The Philips DXL ECG Algorithm, developed by the Advanced Algorithm Research Center, uses sophisticated analytical methods
ECG made extra easy. medics.cc
ElectroCardioGraphyraphy ECG made extra easy Overview Objectives for this tutorial What is an ECG? Overview of performing electrocardiography on a patient Simple physiology Interpreting the ECG Objectives
EEG Cap Instructions
EEG Cap Instructions Supplies You will need to have the following supplies at hand before applying the cap, all of which can be found in the daily preparation supplies cabinet: NuPrep mildly abrasive skin
How can I be sure it will be performed to a high standard?
FREQUENTLY ASKED QUESTIONS (About ultrasound scans at Derriford) What is an Ultrasound Scan? How does it work? What are the dangers? What are the advantages? Who will perform my scan? How can I be sure
Measuring central venous pressure
Elaine Cole Senior lecturer ED/Trauma, City University Barts and the London NHS Trust 1 Learning outcomes That the clinician can: Describe the sites of central venous catheterisation Understand why central
Remote Working Policy for Juniper Users
Remote Working Policy for Juniper Users 1 Introduction The Local Authority supports Remote Access to facilitate managers and employees to maximise their time and resources, and to provide anytime, anywhere
Operating instructions Diffuse reflection sensor. OJ50xx 701396 / 01 07 / 2004
Operating instructions Diffuse reflection sensor OJ50xx 7096 / 0 07 / 004 Contents Preliminary note. Symbols used Function and features Installation. Installation of the supplied mounting fixture 4 4 Electrical
stretches and exercises
stretches and exercises The enclosed sheets contain stretches and exercises which can be used to delay and minimise the development of contractures and deformities occurring in children with Duchenne muscular
510(k) Summary May 7, 2012
510(k) Summary Medicalgorithmics 510(k) Premarket Notification 510(k) Summary May 7, 2012 1. Submitter Name and Address Medicalgorithmics LLC 245 West 107th St., Suite 11A New York, NY 10025, USA Contact
2 ECG basics. Leads and planes. Leads. Planes. from different perspectives, which are called leads and planes.
558302.qxp 3/14/12 10:47 PM Page 12 2 ECG basics One of the most valuable diagnostic tools available, an electrocardiogram (ECG) records the heart s electrical activity as waveforms. By interpreting these
Fundamentals of Signature Analysis
Fundamentals of Signature Analysis An In-depth Overview of Power-off Testing Using Analog Signature Analysis www.huntron.com 1 www.huntron.com 2 Table of Contents SECTION 1. INTRODUCTION... 7 PURPOSE...
The Value of High Fidelity Electrocardiography Using the Cathode Ray Oscillograph and an Expanded Time Scale
The Value of High Fidelity Electrocardiography Using the Cathode Ray Oscillograph and an Expanded Time Scale By PAUL H. LANGNER, JR., M.D. The cathode ray is practically free from inertia and therefore
HOW TO CHECK YOUR LYMPH NODES
HOW TO CHECK YOUR LYMPH NODES What are the aims of this leaflet? This leaflet has been written to help you to understand more about how to check your lymph nodes yourself. It tells you why you should be
Operating Instructions
Operating Instructions Vision 5L Digital Holter Recorder Document Number: 010-1646-00 Revision: H Bothell, Washington 98021-8969 (608)-764-1919 Caution: U.S. Federal law restricts this device to sale by
INSTRUCTIONS FOR USE. Read this information before you start. Important things to know about your SmartJect autoinjector. Important things to remember
INSTRUCTIONS FOR USE SIMPONI (SIM-po-nee) (golimumab) SmartJect autoinjector If your doctor decides that you or a caregiver may be able to give your injections of SIMPONI at home, you should receive training
Move and position individuals in accordance with their plan of care
Move and position individuals in accordance with their plan of care Page 1 of 21 Level 2 Diploma in Health and Social Care Unit HSC 2028 Tutor Name: Akua Quao Thursday 12 th September 2013 Release Date:
HEART MONITOR TREADMILL 12 LEAD EKG
2 HEART MONITOR TREADMILL 12 LEAD EKG Portable ambulatory monitoring system Continuously records electrical activity of the heart for 24 hours or more Also known as ambulatory electrocardiographic monitor
This is how quickly your brain forgets things
Brain gym starters This is how quickly your brain forgets things Your brain is a muscle Like every other muscle it needs exercise Brain gym is one way to exercise your mind so that you can learn better.
Instructions to User
Dear Users, User Manual for PC-80D Easy ECG Monitor Instructions to User Thank you very much for purchasing our product. Please read the manual very carefully before using this device. These instructions
Signal-averaged electrocardiography late potentials
SIGNAL AVERAGED ECG INTRODUCTION Signal-averaged electrocardiography (SAECG) is a special electrocardiographic technique, in which multiple electric signals from the heart are averaged to remove interference
An Overview of Transradial Patient Set-up. Susan R. Cooney RN,BSN,CCRN Durham VAMC Cardiac Catheterization Lab Duke University School of Nursing
An Overview of Transradial Patient Set-up Susan R. Cooney RN,BSN,CCRN Durham VAMC Cardiac Catheterization Lab Duke University School of Nursing Advantages of the Transradial Approach from a Nursing Perspective!!
Range of Motion Exercises
Range of Motion Exercises Range of motion (ROM) exercises are done to preserve flexibility and mobility of the joints on which they are performed. These exercises reduce stiffness and will prevent or at
Cardiac Conduction System (1) ECG (Electrocardiogram) Cardiac Conduction System (2) The ECG (1) The ECG (1) The ECG (1) Achmad Rizal BioSPIN
ECG (Electrocardiogram) Cardiac Conduction System (1) Achmad Rizal BioSPIN ARL-EL4703-Instrumentasi Biomedis 2 Cardiac Conduction System (2) The ECG (1) ARL-EL4703-Instrumentasi Biomedis 3 ARL-EL4703-Instrumentasi
Understand nurse aide skills needed to promote skin integrity.
Unit B Resident Care Skills Essential Standard NA5.00 Understand nurse aide s role in providing residents hygiene, grooming, and skin care. Indicator Understand nurse aide skills needed to promote skin
Official reprint from UpToDate www.uptodate.com 2013 UpToDate
Official reprint from UpToDate www.uptodate.com 2013 UpToDate Patient information: Breast cancer screening (The Basics) Written by the doctors and editors at UpToDate What is breast cancer screening? Breast
Information for patients What is an EMG/Nerve Conduction Study?
Information for patients What is an EMG/Nerve Conduction Study? Department of Neurophysiology We must obtain your consent for any procedure beforehand. Staff will explain all the risks, benefits and alternatives
Efficient Heart Rate Monitoring
Efficient Heart Rate Monitoring By Sanjeev Kumar, Applications Engineer, Cypress Semiconductor Corp. Heart rate is one of the most frequently measured parameters of the human body and plays an important
INTRODUCTORY GUIDE TO IDENTIFYING ECG IRREGULARITIES
INTRODUCTORY GUIDE TO IDENTIFYING ECG IRREGULARITIES NOTICE: This is an introductory guide for a user to understand basic ECG tracings and parameters. The guide will allow user to identify some of the
Paediatric Intensive Care Unit Nursing Procedure: Care of Arterial Lines.
Paediatric Intensive Care Unit Nursing Procedure: Care of Arterial Lines. Definition: Arterial Line Placement of an indwelling arterial catheter for the purpose of continuous monitoring of intra arterial
INTRODUCTION TO CLINICAL PRACTICE AND CLINICAL SKILLS 2nd Year MEDICAL YEAR 2009/2010
INTRODUCTION TO CLINICAL PRACTICE AND CLINICAL SKILLS 2nd Year MEDICAL YEAR 2009/2010 POINT OF CARE TESTING This session is designed to teach you the principals of point of care testing. This is common
Information for patients and nurses
Information for patients and nurses Rocket IPC Pleural Catheter Indwelling Catheter Rocket Indwelling Pleural Catheter (IPC) Contents Contact Information...03 What s in the Rocket Dressing Pack and Bottle
How To Stretch Your Body
Exercise Module A New Leaf Choices for Healthy Living University of North Carolina at Chapel Hill 2007 Center for Health Promotion and Disease Prevention Physical Activity Exercises for Keeping Active
INTRODUCTION TO POSITIONING. MODULE 3: Positioning and Carrying
MODULE 3: Positioning and Carrying Blankets, pillows and towels, sandbags, carpet on the floor, a doll with floppy limbs that can be positioned i.e. not a hard plastic doll, display materials. Icebreaker
CLINICAL PROTOCOL THE MALNUTRITION UNIVERSAL SCREENING TOOL (MUST)
CLINICAL PROTOCOL THE MALNUTRITION UNIVERSAL SCREENING TOOL (MUST) RATIONALE The Malnutrition Universal Screening Tool (MUST) is now recommended best practice within the Trust as a way to screen patients
Biopac Student Lab Lesson 6 ELECTROCARDIOGRAPHY (ECG) II Introduction. Rev. 06132012
42 Aero Camino, Goleta, CA 93117 www.biopac.com Biopac Student Lab Lesson 6 ELECTROCARDIOGRAPHY (ECG) II Introduction Rev. 06132012 Richard Pflanzer, Ph.D. Associate Professor Emeritus Indiana University
Requirements for Provision of Outreach Paediatric Cardiology Service
Requirements for Provision of Outreach Paediatric Cardiology Service Dr Shakeel A Qureshi, Consultant Paediatric Cardiologist, Evelina Children s Hospital, London, UK On behalf of British Congenital Cardiac
How to read the ECG in athletes: distinguishing normal form abnormal
How to read the ECG in athletes: distinguishing normal form abnormal Antonio Pelliccia, MD Institute of Sport Medicine and Science www.antoniopelliccia.it Cardiac adaptations to Rowing Vagotonia Sinus
Electrocardiogram and Heart Sounds
Electrocardiogram and Heart Sounds An introduction to the recording and analysis of electrocardiograms, and the sounds of the heart. Written by Staff of ADInstruments Introduction The beating of the heart
Stress Echocardiogram
Purpose: Heart First Guidelines Stress Echocardiogram To outline the guidelines and protocols for performing Stress Echocardiograms as established by the American College of Cardiology References: 1. Stress
Bulging Disc Back Pain Relief at Home Using Lumbar Traction Techniques.
Bulging Disc Back Pain Relief at Home Using Lumbar Traction Techniques. A no-nonsense Guide Written By Tommy Hoffman, P.T. 1 Copyright 2009 by Tommy Hoffman All rights reserved under the Pan-American and
The science of medicine. The compassion to heal.
A PATIENT S GUIDE TO ELECTROPHYSIOLOGY STUDIES OF THE HEART The science of medicine. The compassion to heal. This teaching booklet is designed to introduce you to electrophysiology studies of the heart.
Starting position: Lying with knees bent up and feet flat on floor/bed about 12" (30cms) apart
Exercise 3 Bridging Starting position: Lying with knees bent up and feet flat on floor/bed about 12" (30cms) apart Tighten your buttocks, then raise them off the floor to form a bridge, then hold. Then
Rehabilitation after shoulder dislocation
Physiotherapy Department Rehabilitation after shoulder dislocation Information for patients This information leaflet gives you advice on rehabilitation after your shoulder dislocation. It is not a substitute
The Super 7 For Tennis Elbow
Exercises for Tennis elbow Tennis elbow recovery time varies with each person and may take several weeks. Recovery may be faster and more successful when a combination of local injections (e.g. steroids
Achieving Independence
Bard: Intermittent Self-Catheterization A Guide to Self-Catheterization Achieving Independence Introduction This brochure is provided by Bard, a leading provider of urology products since 1907. The best
CARDIAC REHABILITATION HOME EXERCISE ADVICE
CARDIAC REHABILITATION HOME EXERCISE ADVICE Introduction Exercise is a very important as part of your rehabilitation and your future health. You will need to exercise at least three times per week, minimum,
Section Two: Arterial Pressure Monitoring
Section Two: Arterial Pressure Monitoring Indications An arterial line is indicated for blood pressure monitoring for the patient with any medical or surgical condition that compromises cardiac output,
Epinephrine Auto Injector Interim Policy (Amended March 12, 2008)
Epinephrine Auto Injector Interim Policy (Amended March 12, 2008) Pursuant to the authority conferred by N.J.S.A. 26:2K-47.1, et seq., the Department of Health and Senior Services (the Department) shall
Wall Mount Installation and Use
Wall Mount Installation and Use 1. Introduction The Wall Mount works the same way for all Touchfire cases - the one or two row magnet strip on the back of the case is magnetically attracted to the Wall
Introduction to Electrocardiography. The Genesis and Conduction of Cardiac Rhythm
Introduction to Electrocardiography Munther K. Homoud, M.D. Tufts-New England Medical Center Spring 2008 The Genesis and Conduction of Cardiac Rhythm Automaticity is the cardiac cell s ability to spontaneously
Transcutaneous Electrical Nerve Stimulation Device LUMI-TENS
Transcutaneous Electrical Nerve Stimulation Device LUMI-TENS Operation Manual Read Before Using LUMI-TENS-INS-LAB-RevA08 TABLE OF CONTENTS INTRODUCTION TO TENS INDICATIONS AND CONTRAINDICATIONS WARNINGS
trust clinical guideline
CG23 VERSION 1.0 1/7 Guideline ID CG23 Version 1.0 Title Approved by Transient Loss of Consciousness Clinical Effectiveness Group Date Issued 01/01/2013 Review Date 31/12/2016 Directorate Authorised Staff
Patient Instruction Manual
Patient Instruction Manual Preventice Services, LLC and Preventice Technologies, Inc., are independent, wholly owned subsidiaries of Preventice Solutions, Inc. Welcome to BodyGuardian Verité Your physician
Exercises after breast or upper body lymph node surgery
Exercises after breast or upper body lymph node surgery Surgery may limit the strength and movement of your arm. These exercises are helpful if you have had: breast surgery with or without lymph node removal
Hemiplegic shoulder pain/shoulder subluxation
UPPER LIMB NEUROMUSCULAR ELECTRICAL STIMULATION: Electrode positions Please note that the polarity (red and black leads) can be altered according to your clinical reasoning. The area in which you want
SC9: NON SURGICAL LASERS / IPLS/ LIGHT
DEFINITIONS SC9: NON SURGICAL LASERS / IPLS/ LIGHT The conditions in this part refer specifically to Class 3B lasers, Class 4 lasers and Intense Light Source systems (ILS) used for non-surgical cosmetic
Introduction 1 The system 1 The meter 2 The display 3 The mode 3 The measurement 4 Coding the meter 4 How to obtain a drop of blood 6 Application of
Introduction 1 The system 1 The meter 2 The display 3 The mode 3 The measurement 4 Coding the meter 4 How to obtain a drop of blood 6 Application of the blood 7 Procedure to test glucose 7 Procedure to
New 7/1/2015 MCFRS 1
New 7/1/2015 MCFRS 1 The providers will summarize the need for this change from an epinephrine auto injector The provider will define the proper dosage of epinephrine for the adult and pediatric patient
Interference in the ECG and its elimination
Interference in the ECG and its elimination General points The signal voltages in ECG recording are known to be very small and are in the millivolt range, i.e. they have amplitudes of only a few thousands
Passive Range of Motion Exercises
Exercise and ALS The physical or occupational therapist will make recommendations for exercise based upon each patient s specific needs and abilities. Strengthening exercises are not generally recommended
THE SKELETAL SYSTEM FUNCTIONS OF THE SKELETAL SYSTEM
THE SKELETAL SYSTEM The skeleton is the body s bony framework which consists of 206 bones. The bones are made up of water(45%), calcium and phosphorous(35%) and other organic materials(20%). The calcium
Instructions for In-Lab Sleep Study Procedures
Instructions for In-Lab Sleep Study Procedures Please refer to the font of this booklet or email for the test you have been scheduled for Description Procedure Code Standard PSG (Polysomnogram) 95810 Split
Heart of America Medical Center EMR, EMT, EMT-Intermediate & AEMT Protocols
USE OF THE LUCAS DEVICE TO PROVIDE EXTERNAL CHEST COMPRESSIONS IN PATIENTS SUFFERING CARDIAC ARREST Purpose This procedure describes the appropriate methods to apply, operate, and discontinue the LUCAS