Gold Refining and Coin Manufacturing at the Royal Canadian Mint

Similar documents
X-RAY FLUORESCENCE SPECTROSCOPY IN PLASTICS RECYCLING

Universal Data Acquisition (UDA)

Observations on the determination of gold in gold alloys cupellation method (Fire Assay)

ROYAL CANADIAN MINT GOLD BAR MANUFACTURER IN OTTAWA

Multi-elemental determination of gasoline using Agilent 5100 ICP-OES with oxygen injection and a temperature controlled spray chamber

CHAPTER 21 ELECTROCHEMISTRY

Coating Thickness and Composition Analysis by Micro-EDXRF

Portable X-ray fluorescence Spectroscopy. Michael A. Wilson Research Soil Scientist USDA-NRCS National Soil Survey Center Lincoln, NE

UNITED STATES CONSUMER PRODUCT SAFETY COMMISSION DIRECTORATE FOR LABORATORY SCIENCES DIVISION OF CHEMISTRY 5 RESEARCH PLACE ROCKVILLE, MD 20850

Secondary Lead Smelting

OLIVÉR BÁNHIDI 1. Introduction

EDXRF of Used Automotive Catalytic Converters

Lead Testing and On Site Calibration for Water Testing Detection Range: 2 100ppb

X Ray Flourescence (XRF)

Elemental Analyses by ICP-AES

HOW CAST GOLD BARS ARE MANUFACTURED

Quartz Glass. Tubes and Rods

Standard Operation Procedure. Elemental Analysis of Solution samples with Inductively Coupled Plasma Mass Spectrometry

USE OF REFERENCE MATERIALS IN THE LABORATORY

SUS. Company Profile. Ulrich Nell, Feldstr.23, D Oberhausen, Tel. 0049(0)208/ Fax 0049(0)208/658536

Periodic Table, Valency and Formula

Instrumentation. (Figure 2)

Additional Lecture: TITRATION BASICS

ATOMIC ABSORTION SPECTROSCOPY: rev. 4/2011 ANALYSIS OF COPPER IN FOOD AND VITAMINS

Determination of Metals in a 3% Sodium Chloride (NaCl) Matrix by Axially-Viewed ICP-OES

NANOCOMPOSIX'S GUIDE TO ICP-MS MEASUREMENT

APPLICATION OF ICP-MS TECHNOLOGY FOR TRACE METALS ANALYSIS

100% ionic compounds do not exist but predominantly ionic compounds are formed when metals combine with non-metals.

Flow Injection Analysis

Analyses on copper samples from Micans

APPLICATION. b) The A&H Centre shall fulfill the following requirements regarding

ISTANBUL GOLD REFINERY

Sample preparation for X-ray fluorescence analysis

The NEW X-MET7000 Faster, lighter,

2. Write the chemical formula(s) of the product(s) and balance the following spontaneous reactions.

VHG Labs, Inc. Standards and Supplies for Spectrochemical Oil Analysis

Casting. Training Objective

Metals in Soil Analysis Using Field Portable X-ray Fluorescence

Assay Laboratory Quality Assurance

MILESTONE H E L P I N G C H E M I S T S

The LBMA Bullion Market Forum Shanghai, May 2011

RoHS Test Methods. Presented by: Bruce Peterson

Introduction to Elemental Analysis by ED-XRF Justin Masone Product Specialist 3 June 2015

LONDON GOOD DELIVERY BARS

BUREAU OF INDIAN STANDARDS

DMA-1 Direct Mercury Analyzer

Electrochemistry Voltaic Cells

Question Bank Electrolysis

RAND REFINERY SOUTH AFRICA

Using the PDF for material identification using elemental data. from XRF and SEM EDS.

08. SEK IDC CONNECTORS

Regional Environmental Protection Agency- Veneto Region LABORATORY REGIONAL DEPARTMENT

HPLC Analysis of Acetaminophen Tablets with Waters Alliance and Agilent Supplies

PRODUCT PROFILE. ELECTROLOY in partnership with FCT Asia Pte Limited. in manufacturing. Nihon Superior Lead Free Solder Bar.

Elemental Scientific. Stability. Elemental Scientific

DEVELOPMENT OF X-RAY FLUORESCENCE-BASED METHO- DOLOGY FOR QUALITY CONTROL OF ZARI USED IN SILK SARIS

Chapter 4 and 5 solutions

Portable GPS-XRF Protocols for Field Archaeology

Determining the Quantity of Iron in a Vitamin Tablet. Evaluation copy

How To Analyze Plasma With An Inductively Coupled Plasma Mass Spectrometer

The Limitations of Hand-held XRF Analyzers as a Quantitative Tool for Measuring Heavy Metal Pesticides on Art Objects. By Özge Gençay Üstün

BB-18 Black Body High Vacuum System Technical Description

FEATURES AND BENEFITS OF DIFFERENT PLATINUM ALLOYS. Kris Vaithinathan and Richard Lanam Engelhard Corporation

Company Introduction. Welcome to BEST. bestpcbs.com.

THE POSSIBILITIES OF RUNNER PLACEMENTS FOR CASTINGS PRODUCED BY SPIN CASTING INTO SILICON RUBBER MOULDS. Matej BEZNÁK, Martin BAJČIČÁK, Roland ŠUBA

B I N G O B I N G O. Hf Cd Na Nb Lr. I Fl Fr Mo Si. Ho Bi Ce Eu Ac. Md Co P Pa Tc. Uut Rh K N. Sb At Md H. Bh Cm H Bi Es. Mo Uus Lu P F.

1 TROY OUNCE GOLD INGOT

Lalor Project Base Metal Zones and Gold Mineralization January 2009

Metal Pressfitting Pipe Systems

How to manage wave solder alloy contaminations. Gerjan Diepstraten & Harry Trip Cobar Europe BV Balver Zinn

Using the Bruker Tracer III-SD Handheld X-Ray Fluorescence Spectrometer using PC Software for Data Collection

A Comprehensive Guide to the Krugerrand. Table of Contents

Specification for Electroless Nickel/ Electroless Palladium/ Immersion Gold (ENEPIG) Plating for Printed Circuit Boards

producer of zinc, aluminium, magnesium die casting and plastic moulding

X-Ray Fluorescence - Data Collection and Analysis

SCH 3UI Unit 2 Outline Up to Quiz #1 Atomic Theory and the Periodic Table

Adapter Catalog MOBILE METAL ANALYZER. SPECTRO isort and SPECTROTEST

Quantitative Analysis Software for X-Ray Fluorescence. XRF-FP is a full-featured quantitative analysis package for XRF

Nanoparticle Deposition on Packaging Materials by the Liquid Flame Spray

Basic Concepts of X-ray X Fluorescence by Miguel Santiago, Scientific Instrumentation Specialist

Amptek Application Note XRF-1: XRF Spectra and Spectra Analysis Software By R.Redus, Chief Scientist, Amptek Inc, 2008.

METHOD 9075 TEST METHOD FOR TOTAL CHLORINE IN NEW AND USED PETROLEUM PRODUCTS BY X-RAY FLUORESCENCE SPECTROMETRY (XRF)

Metal Ion + EDTA Metal EDTA Complex

CVD SILICON CARBIDE. CVD SILICON CARBIDE s attributes include:

An example: helium isotopes. An example: helium isotopes. Limits to Detection/Measurement. Lecture 14 Measurements II: Mass spectrometry

SILVERCORP INTERCEPTS 18.8 METRES TRUE WIDTH GRADING 368 G/T SILVER AND 1.78 PERCENT LEAD AT THE LM MINE WEST, YING MINING DISTRICT, CHINA

Metals and Non-metals. Comparison of physical properties of metals and non metals

Sample Analysis Design Isotope Dilution

47374_04_p25-32.qxd 2/9/07 7:50 AM Page Atoms and Elements

How To Use The Nexion 300

Application of ICP-MS as a Tool for Environmental Monitoring. Bill Spence Environmental Market Specialist

The volume of a penny will be calculated from its mass and density.

LC-MS/MS Method for the Determination of Docetaxel in Human Serum for Clinical Research

Appendix 5 Overview of requirements in English

Dimethylglyoxime Method Method to 6.0 mg/l Ni TNTplus 856

Highlights of selected high-grade mineralized zones exposed in drift tunnels include:

1332 CHAPTER 18 Sample Questions

Halogen Free: What, Why and How. Presented by : Jim Cronin Environmental Monitoring and Technologies, Inc.

EXPERIMENT 4 The Periodic Table - Atoms and Elements

Transcription:

INTERNET Sub title will go here The Application of XRF to Gold Refining and Coin Manufacturing at the Royal Canadian Mint Michael W. Hinds, Ph.D.

The Business of the Mint Circulation Coins (Canada & Other Countries) Numismatic Coins (Sterling Silver & 22 Karat Gold ) Investor Coins (99.99% Gold and Silver) Gold Refining i Services 2

Ottawa 3

4

5

Winnipeg 6

Why XRF? Fast turnaround Minimal sample preparation Minimal labour Simple to use (for operators) Some models have an auto-sampler Determination of major, minor, and trace elements in one analysis Can be used for unknown materials 7

Overview of Applications Ag and P in Sterling Silver Au, Ag, Cu & Other Elements in Impure Gold and Silver to assist Fire Assay Ag and Cu in Chlorinated Gold as an Endpoint Indicator As, Cd, Hg, Se, & Te in Impure Gold and Silver coming into the Refinery Plated Layer Thickness in Circulation Coin Blanks 8

XRF Spectrometers Wavelength Dispersive XRF Panalytical l PW2404 (1999 2010) Panalytical Axios (2010) Energy Dispersive Spectro 200 InnvoX Omega Xpress (portable) 9

Sterling Silver Alloy Alloy: Ag 92.5%, P 700 ppm (max), Cu balance Used for Numismatic Coins Charges made from Ag, Cu, and P Alloy is continuously cast (point of sampling) Rolled, blanked, and stamped 10

Analysis Criteria Silver precision ± 0.0303 % RSD (n = 3) lower alloy limit 92.30% Phosphorous precision ± 5% RSD (n = 3) 700 ppm max. Copper Balance Analysis time < 5 minutes 11

Sample Handling Criteria Sample Handling Total sample analysis time < 5 minutes o 2 replicates & sample loading Minimal change to sample cutting (Concast) Minimal sample preparation (Assay) Autosampler to handle sample block (30mm) 3 12

Continuous Casting Furnace 13

Continuous Casting Furnace 14

Cutting a Sample 15

Sample Loading into XRF 16

XRF Parameters Sterling Ag Element Line X tal Detector Time (s) Ag K α LiF 200 Duplex 80 Cu K α LiF 200 Duplex 10 P K α Ge111 C Flow 20 Total Time 110 s X ray tube settings PW2404: 60 kv, 66 ma Axios 60 kv, 50 ma 17

Achieving Exceptional Precision Measure using the same cup Long Counting Time Keep X-ray tube at set power for 2 4 hr Sample Surface Preparation Control Environment Inside wall Air conditioning 18

Challenges for Ag 925 Samples Sample size to fill the aperture & fit the cup Constant sample size Sample labeling Smooth top - vacuum seal auto-sampler Sample flatness on analysed side Timely delivery of the samples 19

Sample Preparation Polishing (old): Analysed Side Lathe the face (flatness) or press flat 1st wet polish 400 mesh SiC paper (20 s) 2nd wet polish 125 mesh SiC paper (20 s) Opposite Side Wet polish 400 mesh SiC paper (20 s) Auto-sampler pick-up Milling Machine (new): Mill both surfaces Analysed side more carefully done 20

Ag 925 Standards Collected accepted and rejected samples Ag determined by Gay Lusac Method back titration (AgCl formation from excess Cl) P determined by solution ICP-AES 190.3 nm matrix matched standards Cu by difference 21

Ag Calibration: 85 94% Algorithm: De Jong (similar L-T) 22

P Calibration: 0 700 ppm Algorithm: De Jong (similar L-T) 23

Ag Precision: QC 32-96 (4000W) Time Period Precision Number of RSD Replicates 1 hour 0.02 % 10 8 hour 002% 0.02 20 7 days 0.03 % 98 24

Cu Precision: QC 32-96 (4000W) Time Period Precision RSD Number of Replicates 1 hour 0.06% 10 8 hour 0.06% 20 5 days 006% 0.06% 98 25

P Precision: QC 32-96 (330 ppm) Time Period Precision RSD Number of Replicates 1 hour 1.4% 10 8 hour 1.2% 20 5 days 2.6% 98 26

Accuracy: Ag (n=5 replicates) Sample XRF Gay Lusac Sig. Diff. 277 925.6 ± 0.21 925.5 ± 0.4 No 282 923.4 ± 0.15 924.2 ± 0.5 No 287 925.7 ± 014 0.14 926.22 ± 034 0.34 No 2A 921.3 ± 0.23 920.8 ± 0.35 No 4B 938.7 ± 0.15 939.2 ± 0.25 No 27

P contamination vacuum oil QC Chart of P in QCAg32-96 for 2002 430 410 390 370 350 330 310 290 270 250 Re-polish 12-Au g 17-Au g 22-Au g 27-Au g 1-Se p Concenn tratio n (fin en e ss) 6-Se p 11-Se p 16-Se p Date (dd-mm) P UL Target LL 28

Measuring Ag Alloy Coin Blanks One contract had blanks with diameter less than 27 mm Must use 12 mm sample aperture Lower counts from lower surface area Must increase counting time 29

Blank compared to sample cups 30

27 vs 12 mm Comparison Parameter 27 mm 12 mm Ag counts, kcps 800 120 Effective Diameter, 25 10 mm Effective Area, mm 2 491 78.5 Ratios 27 vs 12 mm Area 6.25 Count 6.67 31

Precision: 12 vs 27 mm cups Ag 90% alloy; n =10 Parameter 12 mm 27 mm Ag Counting Time, s 120 (20 bkgd) Precision, %RSD 003 0.03 004 0.04 002 0.02 003 0.03 80 P Counting Time, s 30 20 (10 bkgd) (10 bkgd) Precision, % RSD 2 2 32

Analysis of Impure Gold Settlement of Gold Deposit After melt mass Determine Au and Ag by Fire Assay Fire Assay: Gravimetric High precision and accuracy Industry standard XRF used to streamline Fire Assay 33

Overview of Fire Assay Gold Sample Button Fire cupelled in Pb Au and Ag Assay Fire cupelled with Inquarted Ag Ag dissolved Au 34

Traditional Fire Assay Gold Sample Button Fire Au & Ag Inquartation Visual Assay Fire Au 35

Fire Assay with XRF Fire Assay Button Fire Au & Ag Estimate Au & Ag other elements XRF Assay Fire Au 36

XRF for Ag Inquartation (FA) Measurement of Au and Ag used to add Inquartation Ag for the Assay Fire Ag Inq = ((2.5*Au)-Ag)/2 Ag Inq must be within ± 50 mg Ag Minimal sample and sample prep Press small dip sample Polish 3 minute analysis time Elements: Au, Ag, Cd, Cu, Fe, Ni, Pb, Pt, Pd, Te, Zn 37

XRF for FA: Limitations Small sample size 12 mm aperture Automation required: large number cups Fast response e Personnel with minimal training 4000 W power (60 kv and 66mA) Must also have a 6 mm aperture program Some samples break apart 38

INQ12 Au Calibration (40 100%) 41

INQ12 Ag Calibration (0-100%) 42

Au & Ag Analytical Results Standard: INQ12QC2, n= 38 determinations over 3 weeks Units: parts per thousand or fineness Au Ag Average 624.3 85.6 Std Dev 0.7 1.7 RSD 0.1% 2% Certified Values 623.8 84.7 Difference 0.5 0.9 % Error 0.1% 1% 43

Au & Ag Analytical Results Standard: QC60847, n= 5 determinations Units: parts per thousand or fineness Au Ag Average 644.5 73.5 Std Dev 0.8 0.1 RSD 0.1% 0.1% Certified Values 646.0 75.5 Difference -1.5-2.0 % Error -0.2% -3% 44

Observations Good analytical results from XRF due to: Sampling molten metal (induction melt) XRF application Background, overlap corrections Calibration Standards Produced in-house Specially cast or use normal samples Rigorously analysed Fire Assay Fire Assay ICP-AES Ag, Pd, & Pt (Cu & Pb) ICP-AES FAAS 45

Overview XRF very useful to refining and coin operations: Timely results Analytical performance Accuracy (standards) Precision Ease of use Minimal sample preparation XRF requires investment in: Standard manufacturing & analysis Method development 46

Acknowledgements Charlie Wu Alexander Seyfarth Maggi Loubser Bruno Vrebos Royal Canadian Mint Jon Forrest Paul Laframboise Robin Entwistle James Willis and Andy Duncan 47