PARALLEL TO THE DEVELOPMENTS OF SURGICAL



From this document you will learn the answers to the following questions:

Where is the Orbscan I system from?

What does ACD stand for?

Where is the Oculus located?

Similar documents
Comparison to other anterior segment imaging devices

IOLMaster with Software Version V 3.0

THE BEST OF BOTH WORLDS Dual-Scheimpflug and Placido Reaching a new level in refractive screening

ALL-IN-ONE Optical Biometry, Dual Scheimpflug Tomography and Placido Topography

IOLMaster 500. Excellence in Biometry. Precision. Simplicity. Outcomes.

OCT-based IOL power calculation for eyes with previous myopic and hyperopic laser vision correction

Corneal Thickness Measurement with Optical and Ultrasonic Pachymetry

Comparison of central corneal thickness measurements using Pentacam and ultrasonic pachymetry in post-lasik eyes for myopia

Total Corneal Power Estimation: Ray Tracing Method versus Gaussian Optics Formula PATIENTS AND METHODS

EUROPEAN JOURNAL OF PHARMACEUTICAL AND MEDICAL RESEARCH

Insert to. January Comfort and confidence for all IOL calculations

Insert to October 2013 THE COMPLETE PICTURE. Experiences with the ALADDIN system. Sponsored by Topcon

Cataract Testing. What a Patient undergoes prior to surgery

Page: 1 of 6. Corneal Topography/Computer-Assisted Corneal Topography/ Photokeratoscopy

IOL Power Calculations for Postrefractive Surgery Eyes

Incision along Steep Axis

IOL Power Calculation After Myopic LASIK. Hany Helaly, Lecturer of Ophthalmology, Faculty of Medicine, Alexandria University.

How do we use the Galilei for cataract and refractive surgery?

Providing Optimal Optics For Your Astigmatic Cataract Patients. While the cornea remains relatively stable and prolate throughout life

OCULUS PENTACAM BASIC CLASSIC HR

FIRST EXPERIENCE WITH THE ZEISS FEMTOSECOND SYSTEM IN CONJUNC- TION WITH THE MEL 80 IN THE US

Corneal Power Measurements Using Scheimpflug Imaging in Eyes With Prior Corneal Refractive Surgery

Curtin G. Kelley, M.D. Director of Vision Correction Surgery Arena Eye Surgeons Associate Clinical Professor of Ophthalmology The Ohio State

Keratoconus Detection Using Corneal Topography

AcrySof IQ Toric IOL (SN6ATT) Surgeon Keys for Success & Acknowledgement

Customized corneal ablation and super vision. Customized Corneal Ablation and Super Vision

Long-Term Outcomes of Flap Amputation After LASIK

The Nurse/Technician Role Within the Emerging Ophthalmic Technology - OCTs/B-Scan

IOLMaster 500. Technical Data. Experience the essence of non-contact biometry

Simple regression formula for intraocular lens power adjustment in eyes requiring cataract surgery after excimer laser photoablation

Welcome to the Verisyse Seminar

What is the main target for all phaco surgeons?

COS Statement on Values for Uninsured Services in Canada

Corneal intrastromal implantation surgery for the treatment of moderate and high myopia

Kerry D. Solomon, MD, is Director of the Carolina Eyecare Research Institute at Carolina Eyecare Physicians in Charleston, S.C.

NEW. CIRRUS HD-OCT 5000/500 Advancing Smart OCT. Imaging Applications: Anterior Segment Glaucoma Retina. International Version

OCULUS Pentacam Pentacam HR

Supplement to October Master Your Toric Planning and Improve Refractive Outcomes With the LENSTAR LS 900. Sponsored by

Posterior Corneal Astigmatism: Is It Important? MP Weikert, MD Baylor College of Medicine October 23, 2011

ICRS implantation with the Femto LDV laser in stabilized KC patients: 6 months results

Your one stop vision centre Our ophthalmic centre offers comprehensive eye management, which includes medical,

All from one trusted partner:

OCT-guided Femtosecond Laser for LASIK and Presbyopia Treatment

Biometry for Refractive Lens Surgery

Ectasia after laser in-situ keratomileusis (LASIK)

HKSH Introduces New Femtosecond Laser Enhanced Refractive Cataract Surgery High Precision and Safety

Standardized Analyses of Correction of Astigmatism With the Visian Toric Phakic Implantable Collamer Lens

THE GUIDE TO REFRACTIVE LENS EXCHANGE SEE CLEARLY.

LASIK. Complications. Customized Ablations. Photorefractive Keratectomy. Femtosecond Keratome for LASIK. Cornea Resculpted

CONSENT FOR CATARACT SURGERY REQUEST FOR SURGICAL OPERATION / PROCEDURE AND ANAESTHETIC

ReLEx smile Minimally invasive vision correction Information for patients

Wavefront Analysis in Post-LASIK Eyes and Its Correlation with Visual Symptoms, Refraction, and Topography

Laser refractive surgery is becoming increasingly

Management of Astigmatism in Cataract Surgery

Making the Most of High-tech Biometry


The Digital Microscope Platform. Heads-up Surgery with Integrated Applications

5/24/2013 ESOIRS Moderator: Alaa Ghaith, MD. Faculty: Ahmed El Masri, MD Mohamed Shafik, MD Mohamed El Kateb, MD

CustomVue Treatments for Monovision in Presbyopic Patients with Low to Moderate Myopia and Myopic Astigmatism

Alexandria s Guide to LASIK

Common Co-management Questions

Wavefront technology has been used in our

THE EYES IN MARFAN SYNDROME

Customized corneal ablation can be designed. Slit Skiascopic-guided Ablation Using the Nidek Laser. Scott MacRae, MD; Masanao Fujieda

ABrand New Refraction Meas urement Instrument

Presbyopia modern surgical treatments. Johan Lønsmann Poulsen, Euroeyes

INTRODUCTION. Trans Am Ophthalmol Soc 2006;104:

Rediscover quality of life thanks to vision correction with technology from Carl Zeiss. Patient Information

Keratorefractive Surgery for Post-Cataract Refractive Surprise. Moataz El Sawy

Case Reports Post-LASIK ectasia treated with intrastromal corneal ring segments and corneal crosslinking

Comparison of Residual Stromal Bed Thickness and Flap Thickness at LASIK and Post-LASIK Enhancement in Femtosecond Laser-Created Flaps

Premium IOL Implantation Calculations in Post-LASIK Cataract Eyes Using ASCRS IOL Calculator

The Cornea, The Lens & Refractive Surgery. John S. Berestka, M.D. Timothy J. Ehlen, M.D. Matthew R. Ramsey, M.D. Stanley D. Walker, M.D.

Optimizing refractive outcomes

Active Cyclotorsion Error Correction During LASIK for Myopia and Myopic Astigmatism With the NIDEK EC-5000 CX III Laser

Comparison Combined LASIK Procedure for Ametropic Presbyopes and Planned Dual Interface for Post-LASIK Presbyopes Using Small Aperture Corneal Inlay

Trends in Ophthalmology a Market Perspective. Matthias Karl, Michael Kempe

Tucson Eye Care, PC. Informed Consent for Cataract Surgery And/Or Implantation of an Intraocular Lens

LASER VISION C ORRECTION REFRACTIVE SURGERY CENTER

VA high quality, complications low with phakic IOL

Post LASIK Ectasia. Examination: Gina M. Rogers, MD and Kenneth M. Goins, MD

SARAH PHYLLIPS. Now, a cataract procedure as unique as your eye itself. TABLE 1. Age: Depth (ACD) Axial LengTH

Informed Consent for Refractive Lens Exchange (Clear Lens Replacement)

The Pentacam: Precision, Confidence, Results, and Accurate Ks!

Supplement to January Sponsored by STAAR Surgical. New Outcomes:

Pentacam HR Criteria for Curvature Change in Keratoconus and Postoperative LASIK Ectasia

Corneal Refractive Power Estimation and Intraocular Lens Calculation after Hyperopic LASIK

Our Commitment To You

Update on Post- Refractive Surgery IOL Calculations

Fundus Photograph Reading Center

Transcription:

Comparison of Three Methods of Measuring Corneal Thickness and Anterior Chamber Depth WOLF BUEHL, MD, DANIJELA STOJANAC, MD, STEFAN SACU, MD, WOLFGANG DREXLER, MD, AND OLIVER FINDL, MD Accepted for publication Aug 22, 2005. From the Departments of Ophthalmology (W.B., D.S., S.S., O.F.) and Medical Physics (W.D.), Medical University of Vienna, Vienna, Austria. W.D. is a consultant for Carl Zeiss Meditec (Jena, Germany). Inquiries to Oliver Findl, MD, Department of Ophthalmology, Medical University of Vienna, Währinger Gürtel 18-20, 1090 Wien, Austria; fax: 43-1-40400-7881; e-mail: oliver.findl@meduniwien.ac.at PURPOSE: To compare three different methods of measuring corneal thickness (CT) and anterior chamber depth (ACD). DESIGN: Prospective clinical trial (Medical University of Vienna, Austria). METHODS: Central CT (CCT), CT at four peripheral points, and central ACD were measured in 88 eyes of 44 healthy subjects with the Pentacam (rotating Scheimpflug camera; Oculus, Wetzlar, Germany), Orbscan I (scanningslit topography system; Orbtek Inc, Salt Lake City, Utah, USA), and AC-Master (partial coherence interferometry; Zeiss Meditec, Jena, Germany), and the results were compared. RESULTS: The upper (lower) limits of agreement for CCT measurements were 7.9 ( 22.2) m between AC- Master and Pentacam, 17.6 ( 32.5) m between AC- Master and Orbscan, and 25.2 ( 25.9) m between Pentacam and Orbscan. Correlation was high between all three methods (r 0.94 to 0.97). The upper and lower limits of agreement for ACD were 0.174 ( 0.251) mm between AC-Master and Pentacam, 0.406 ( 0.004) mm between AC-Master and Orbscan, and 0.384 (0.095) mm between Pentacam and Orbscan. Correlation was high between the three methods (r 0.96 between Orbscan and Pentacam; others 0.92). Correlation was lower for the CT measurements at the four peripheral points. CONCLUSIONS: The CCT and ACD values obtained by Pentacam, Orbscan, and AC-Master measurements correlated well and showed few outliers. The two new systems (Pentacam, AC-Master) provide a reliable, easyto-use, noncontact method of measuring CCT and ACD. Larger differences occurred only when measuring peripheral CT values, especially between AC-Master and the other two methods. (Am J Ophthalmol 2006;141: 7 12. 2006 by Elsevier Inc. All rights reserved.) PARALLEL TO THE DEVELOPMENTS OF SURGICAL technique in cataract and refractive surgery, the accurate measurement of corneal topography, anterior chamber depth, thickness of the crystalline or artificial lens, and eye length has gained in importance. Until recently, ultrasound biometry has been a common method for measuring corneal thickness (CT) and anterior chamber depth (ACD). However, this method is operator dependent. 1,2 The most common method is applanation ultrasound, requiring corneal contact, which may lead to false results due to indentation of the cornea. 3 The measuring results also depend on the exact axial placement of the probe relative to the center of the cornea. Like all contact methods, it may be uncomfortable for the patient or even lead to damage of the corneal epithelium. 4 Thus, noncontact methods are preferred for biometry of the eye. 5,6 An accurate noncontact ocular biometry technique, based on the dual laser beam partial coherence interferometry (PCI) principle, has been developed in the past decade. 7 10 The PCI technology has been used for precise axial length measurements and resulted in the commercially available IOL Master (Carl Zeiss Meditec, Jena, Germany). However, the IOL Master uses a photographic (not PCI) technique for measuring ACD. 11 Therefore, the AC-Master (Zeiss Meditec) has been developed for PCI measurements of central corneal thickness (CCT) and ACD as well as lens thickness. The precision of this technique is in the micron region, and it is highly reproducible. 5,12 In addition, several other optical (non-pci) methods for imaging and measuring the corneal surface and the anterior chamber of the eye have been developed recently and are already commercially available. 13 15 One of these is the Pentacam (Oculus, Wetzlar, Germany), which uses a rotating Scheimpflug 16 camera to image the anterior segment of the eye. It is also a noncontact method, and it is 0002-9394/06/$32.00 2006 BY ELSEVIER INC. ALL RIGHTS RESERVED. 7 doi:10.1016/j.ajo.2005.08.048

specifically designed to calculate a three-dimensional model of the anterior segment, including data for corneal topography (also of the posterior corneal surface), CT (pachymetry), ACD measurements, and measurements of lens opacity and lens thickness. An already established instrument for analysis of cornea and anterior chamber, which does not make use of the Scheimpflug principle, is the Orbscan (I) scanning-slit topography system (Orbtek Inc, Salt Lake City, Utah, USA). 1,17 It uses a horizontally moving slit beam to produce multiple slit images of the anterior segment and provides data for (anterior and posterior) corneal topography and ACD. However, little is known about the reliability of the different measuring techniques. In particular, it is not entirely clear whether the results of the different methods (PCI, scanning-slit, Scheimpflug method) are comparable and whether they can be used interchangeably. The purpose of this study was to compare CT and ACD measurements by the AC-Master with those of the Pentacam and the Orbscan in a number of healthy subjects. METHODS EIGHTY-EIGHT EYES OF 44 YOUNG VOLUNTEERS WERE INcluded in this prospective study. The subjects had to be healthy except for myopia/hyperopia and/or astigmatic ametropia. Exclusion criteria were other ocular diseases such as glaucoma and retinal pathology, or any history of ocular surgery, including (refractive) laser surgery. The volunteers were informed about the purpose of the study and had to give informed consent before inclusion. The study was performed in adherence to the Declaration of Helsinki. Ethics committee approval was not required for this study. First, the refraction of each eye was determined with an automated refractometer (Zeiss Meditec, Jena, Germany). Contact lenses had to be removed at least 12 hours before the measurements. Both eyes of every subject were measured with the Pentacam, the Orbscan, and finally the AC-Master. Thirty minutes were scheduled for the examination time for each subject. All examinations were performed by the same experienced examiner. The Pentacam system consists of the Pentacam itself and a personal computer. The software is almost fully automated. After the patient s data are entered, the program changes to imaging mode. The subject sits in front of the camera and places the chin on a chin rest. He or she is asked to fixate on a fixation target in the center of the camera. The examiner sees a real-time image of the patient s eye on the computer screen. The image must be focused and centered manually by moving the Pentacam in the respective directions. Markings on the screen indicate the direction the operator should move the joystick of the camera. As soon as the image is perfectly aligned, the patient is asked not to move and to keep his or her eye open, and the scan is started. The rotating camera takes multiple (25 or 50, depending on the user settings) slit images of the anterior eye segment in approximately 2 seconds with 500 true elevation points incorporated in each slit image. For this study, the 25 images per scan option was chosen. After completing a scan, the software calculates a three-dimensional image of the anterior segment, including anterior and posterior corneal surface and lens surface as well as lens opacities. The quality of the lens data depends on the pupil size; of course, only the part of the lens visible in the pupillary aperture can be examined with the Scheimpflug camera. In this study, we did not use mydriatic eyedrops because we were mainly interested in central axial biometric data. The center of the cornea is measured in each of the single images of a scan. It is therefore possible to calculate very precise values for the CCT and the (central) ACD. In the Pentacam, ACD is defined as the distance between the posterior surface of the cornea and the anterior surface of the lens. One high-quality scan per eye was saved on hard disk. The CCT and central ACD readings were noted for each eye. Additionally, peripheral CT measurements in the 3, 6, 9, and 12 o clock positions at a distance of 1.5 mm from the corneal center (corresponding to a circle around the center with a diameter of 3 mm) were noted. Similar to the Pentacam, the Orbscan, the horizontally moving camera, produces multiple slit-lamp images of the eye to calculate a mathematical model of the cornea and the anterior lens surface with computer software. A realtime image of the eye is visible on the screen and has to be centered and aligned manually. The operator uses the reflections of the slit-lamp light as orientation for the correct alignment. Then the subject is asked to keep the eyes open and not to move while the scan is started. Artifacts due to eye movement during the measurement lead to an error message, and the scan has to be repeated. Again, we noted CCT, central ACD (which is again defined as distance between posterior corneal surface and anterior lens surface), and four additional values for peripheral CT (again in the horizontal and vertical meridians of the cornea, at 1.5 mm distance from the corneal center). The AC-Master measures CT, ACD, and lens thickness from a single point of the cornea (usually the center) in one step by the PCI method. The white-to-white distance is determined from an image of the iris before performing biometry. The principle of the dual-beam version of PCI has been reported previously. 7,8 The method requires that measurements are performed along the optical axis (except for peripheral CT measurements). In the AC-Master, the patient fixates on a fixation target visible on a liquid crystal display, which serves to control the direction of gaze of the eye. The ideal position of the fixation target is calculated by the AC-Master, but can be adjusted manu- 8 AMERICAN JOURNAL OF OPHTHALMOLOGY JANUARY 2006

TABLE 1. Comparison of Central Corneal Thickness (CCT) and Central Anterior Chamber Depth (ACD) Values Between Three Methods of Measuring Corneal Thickness Characteristic AC-Master, Pentacam AC-Master, Orbscan Pentacam, Orbscan Mean paired difference SD CCT ( m) 7.1 7.7 7.5 12.8 0.3 13.0 ACD (mm) 0.038 0.108 0.201 0.105 0.239 0.047 Maximum difference CCT ( m) 27.0 55.0 61.0 ACD (mm) 0.448 0.473 0.540 Median absolute difference CCT ( m) 8.0 9.0 7.0 ACD (mm) 0.046 0.215 0.225 Correlation coefficient (r) CCT ( m) 0.974 0.947 0.940 ACD (mm) 0.921 0.924 0.964 Lower limit of agreement CCT ( m) 22.2 32.5 25.9 ACD (mm) 0.251 0.004 0.095 Upper limit of agreement CCT ( m) 7.9 17.6 25.2 ACD (mm) 0.174 0.406 0.384 ally. To measure along the optical axis, the four Purkinje images that result from reflections of the intraocular interfaces must be superimposed during the measurement. This can be controlled by the observer on the real-time image of the patient s eye shown on the computer screen and, if necessary, adjusted by moving the fixation target. Each measurement consists of one or preferably more PCI A-scans (approximately 1 second per scan). With the AC-Master, CT can also be measured peripherally in four meridians at distances of 1.5, 3, and 4.5 mm from the center. The patient is then asked to look at additional fixation lights incorporated in the equipment. In the AC-Master, ACD is defined as the distance between the anterior surface of the cornea and the anterior surface of the lens. Thus, CCT must be subtracted from that result before comparing the ACD values with the results of Pentacam or Orbscan. In this study, we measured CCT and central ACD of both eyes of all subjects. Additionally, peripheral CT was measured at the 3, 6, 9, and 12 o clock positions at a distance of 1.5 mm from the corneal center. For each measurement, several scans were performed (at least 20 for the central measurements and at least five for each peripheral point) and the mean values calculated by the AC-Master software were noted. The precision of measurements was assessed by using the standard deviation of all consecutive scans in each eye. This is our standard procedure for AC-Master measurements; however, the standard deviation is usually smaller than 2 m for CCT measurements and smaller than 11 m for ACD measurements, and a single scan would be sufficient in most cases. 18 All data were entered into a Microsoft Excel spreadsheet. Paired differences were calculated for all comparisons (AC-Master vs Pentacam, AC-Master vs Orbscan, and Pentacam vs Orbscan) and are presented as mean and maximum values, standard deviation, coefficient of variation, median absolute differences, and upper and lower limits of agreement (ie, 95% Wald confidence intervals of the differences). 19,20 The level of statistical significance and the correlation between the three methods were calculated by the Wilcoxon matchedpairs signed rank test and Pearson correlation coefficients. Bland-Altman plots were used to find a potential dependency between differences and means of two measurements. 19 P values of less than.05 were considered to be statistically significant. Statistical analysis was performed by Microsoft Excel and SPSS 12.0 software (SPSS Inc, Chicago, Illinois, USA). RESULTS THE MEAN AGE OF THE 44 SUBJECTS WAS 23.6 YEARS (range, 21 32 years). The mean spherical equivalent was 2.15 D (range, 8.5 to 3.0 D). 13. There were seven contact lens wearers who removed the lenses at least 12 hours before the study. Both eyes of all subjects (88 eyes) could be measured with the three instruments. Mean total examination time was 25 minutes per subject (approximately 15 minutes for Pentacam and Orbscan, and 10 minutes for the AC-Master measurements including scan evaluation). The Pentacam turned out to be the quickest examination, followed by the Orbscan topographer. Because measurements of the peripheral CT with the AC-Master must be performed step by step VOL. 141, NO. 1 THREE METHODS OF MEASURING CORNEAL THICKNESS 9

TABLE 2. Mean Corneal Thickness (CT) Values Measured With Three Systems at Four Peripheral Points Mean Peripheral CT SD ( m) Point* Pentacam Orbscan AC-Master 3 o clock 551 35 549 41 532 33 6 o clock 549 34 557 34 530 33 9 o clock 551 34 552 39 532 32 12 o clock 568 34 568 38 537 31 *Measured at 1.5 mm from the center of the cornea. FIGURE. Bland-Altman plots comparing mean anterior chamber depth (ACD) and ACD difference between AC-Master and Pentacam (top), AC-Master and Orbscan (middle), and Orbscan and Pentacam (bottom). ACD values in millimeters. in different directions of gaze, these were the most time-consuming. Mean SD CCT was 534.6 33.4 m with the Pentacam, 535.0 37.8 m with the Orbscan, and 527.5 31.9 m with the AC-Master. Mean SD ACD was 3.35 0.28 mm with the Pentacam, 3.12 0.27 mm with the Orbscan, and 3.32 0.24 mm with the AC-Master. The differences between the results of the three instruments were all statistically significant (P.01), except for CCT between Pentacam and Orbscan (P.81). Mean and maximum paired differences, median absolute differences, correlation coefficients, and upper and lower limits of agreement between the three methods are shown in Table 1. Bland-Altman plots of the paired ACD differences against the mean values are shown in the Figure. We found no dependency between paired differences and mean values in ACD (or CCT) measurements. CCT correlated best between Pentacam and AC-Master (r 0.97). ACD correlated best between Pentacam and Orbscan (r 0.96). CCT values were higher with the Pentacam than with the AC-Master in 80% of cases (median difference: 7.0 m) and higher with the Orbscan than with the AC-Master in 76% of cases (median difference: 7.0 m). The Orbscan ACD values were lower than the Pentacam results in all cases (median difference: 0.23 mm) and lower than the AC-Master results in 95% of cases (median difference: 0.22 mm). There was no marked offset between measurements of right and left eyes. The same statistics were performed for the peripheral CT measurements (four points in the 3, 6, 9, 12 o clock positions at 1.5 mm distance from the center of the cornea). Mean CT values SD at those points are shown in Table 2. Mean paired differences and median absolute differences between the three methods are shown in Table 3. The values correlated best between Pentacam and AC-Master (r 0.93 0.94). However, the median differences between AC-Master and Pentacam (Orbscan) were significantly larger than those of the CCT measurements (median differences between 17.0 and 30.0 m), whereas median differences between Pentacam and Orbscan did not differ so greatly (0.3 to 5.8 m). The highest difference was 86.0 m between AC-Master and Orbscan at the 6 o clock position. 10 AMERICAN JOURNAL OF OPHTHALMOLOGY JANUARY 2006

TABLE 3. Comparison (Differences) of Corneal Thickness Values at Four Peripheral Points Between Three Methods Point* Mean Paired CT Difference (mean absolute difference) ( m) AC-Master, Pentacam AC-Master, Orbscan Pentacam, Orbscan 3 o clock 19.6 (17.0) 16.9 (17.5) 2.7 (12.0) 6 o clock 19.3 (17.3) 26.8 (24.5) 7.5 (9.5) 9 o clock 19.7 (18.5) 20.8 (19.5) 1.1 (9.0) 12 o clock 31.3 (29.8) 31.2 (30.0) 0.0 (10.0) *Measured at 1.5 mm from the center of the cornea. DISCUSSION ACCURATE MEASUREMENT OF CT AND ACD HAS BECOME important with the rapidly increasing popularity of corneal refractive surgical procedures and with the implantation of new (phakic) intraocular lenses. In this study, we have compared three different noncontact optical methods of measuring CT and ACD. Two of these instruments, the AC-Master and the Pentacam, have been made commercially available only quite recently. The Orbscan is an already established method and was made commercially available several years ago. However, the two new methods, which are based on different optical measuring techniques, have potential advantages. These new instruments have already been shown to have a high repeatability and interexaminer reproducibility 18 (Lackner B, unpublished data, presented at the 19th meeting of the DGII [DGII Deutschsprachige Gesellschaft für Intraokularlinsen-Implantation und Refraktive Chirurgie (German Speaking Society for Intraocular Lens Implantation and Refractive Surgery)], February 2005). Still, it is not clear whether the results of different systems match well and can be used interchangeably. Significant differences in CCT values between different systems have already been shown in previous studies for example, between ultrasound and Orbscan pachymetry. 5 Partial coherence interferometry was shown to be the method with the least intraobserver or interobserver variability for CCT measurements compared with ultrasound pachymetry and the Orbscan system. 6 In this study, we used a prototype version of the AC-Master. Although CCT measurements are highly accurate with this method, measurement of peripheral CT is more sensitive because it depends on the subject s direction of gaze and the ability to fixate on the corresponding fixation light in front of the AC-Master. We found higher standard deviations ( 5 m) between repeated scans in the corneal periphery. These results are therefore less reliable and reproducible than the CCT values. Differences between AC-Master values and Pentacam and Orbscan values were highest for these peripheral points. Whether this is a result of erroneous measurements with the AC-Master or of the different measuring techniques, resulting in a bias between central and peripheral measurements, is uncertain. The Pentacam is an easy-to-use biometry system for analysis of the anterior segment. Measurements with the Pentacam proved to be very simple, user-independent and quick (approximately 2 seconds per scan). Measurements with the Orbscan turned out to be a little more complicated and time-consuming than with the Pentacam because measurements had to be repeated more often, at least with our setup. Apart from a few outliers, from a practical point of view, the limits of agreement between the three methods show acceptable ranges for both CCT and ACD. CCT values tend to be lower with the AC-Master than with the other two methods (mean, 7.3 m), whereas ACD values are lower with the Orbscan than with Pentacam and AC-Master (mean, 0.22 mm). As mentioned above, differences between AC-Master and the other two methods were considerably larger for the peripheral CT measurements (mean, 16 to 31 m). However, the differences between Orbscan and Pentacam remained low. Contrarily, correlation was higher between Pentacam and AC-Master than between Pentacam and Orbscan for the peripheral CT values. In this study, we measured only normal corneas of healthy subjects. Therefore, we do not have data concerning the agreement between the three methods when measuring corneas with pathological alterations, or postoperative corneas. The differences between the three systems might be larger in such cases. In summary, both the AC-Master and the Pentacam proved to be excellent noncontact methods for measuring CCT and (central) ACD in the healthy eye. Considering the different measuring techniques of the two novel devices and the Orbscan, the measurement differences between the three systems were quite small. Bigger differences appeared only when measuring peripheral CT values that might be related to the fact that the AC-Master was primarily designed to measure centrally along the optical axis with high precision and requires different directions of gaze for measuring peripheral CT values. REFERENCES 1. Marsich MW, Bullimore MA. The repeatability of corneal thickness measures. Cornea 2000;19:792 795. 2. Miglior S, Albe E, Guareschi M, et al. Intraobserver and interobserver reproducibility in the evaluation of ultrasonic pachymetry measurements of central corneal thickness. Br J Ophthalmol 2004;88:174 177. 3. Solomon OD. Corneal indentation during ultrasonic pachometry. Cornea 1999;18:214 215. VOL. 141, NO. 1 THREE METHODS OF MEASURING CORNEAL THICKNESS 11

4. Kawana K, Tokunaga T, Miyata K, et al. Comparison of corneal thickness measurements using Orbscan II, noncontact specular microscopy, and ultrasonic pachymetry in eyes after laser in situ keratomileusis. Br J Ophthalmol 2004;88:466 468. 5. Rainer G, Petternel V, Findl O, et al. Comparison of ultrasound pachymetry and partial coherence interferometry in the measurement of central corneal thickness. J Cataract Refract Surg 2002;28:2142 2145. 6. Rainer G, Petternel V, Findl O, et al. Central corneal thickness measurements with partial coherence interferometry, ultrasound, and the Orbscan system. Ophthalmology 2004;111:875 879. 7. Drexler W, Findl O, Menapace R, et al. Partial coherence interferometry: a novel approach to biometry in cataract surgery. Am J Ophthalmol 1998;126:524 534. 8. Findl O, Drexler W, Menapace R, et al. High precision biometry of pseudophakic eyes using partial coherence interferometry. J Cataract Refract Surg 1998;24:1087 1093. 9. Hitzenberger CK. Optical measurement of the axial eye length by laser Doppler interferometry. Invest Ophthalmol Vis Sci 1991;32:616 624. 10. Vogel A, Dick HB, Krummenauer F. Reproducibility of optical biometry using partial coherence interferometry: intraobserver and interobserver reliability. J Cataract Refract Surg 2001;27:1961 1968. 11. Lam AK, Chan R, Pang PC. The repeatability and accuracy of axial length and anterior chamber depth measurements from the IOLMaster. Ophthalmic Physiol Opt 2001;21:477 483. 12. Drexler W, Baumgartner A, Findl O, et al. Submicrometer precision biometry of the anterior segment of the human eye. Invest Ophthalmol Vis Sci 1997;38:1304 1313. 13. Modis L, Langenbucher A, Seitz B. Scanning-slit and specular microscopic pachymetry in comparison with ultrasonic determination of corneal thickness. Cornea 2001;20: 711 714. 14. Suzuki S, Oshika T, Oki K, et al. Corneal thickness measurements: scanning-slit corneal topography and noncontact specular microscopy versus ultrasonic pachymetry. J Cataract Refract Surg 2003;29:1313 1318. 15. Koranyi G, Lydahl E, Norrby S, Taube M. Anterior chamber depth measurement: a-scan versus optical methods. J Cataract Refract Surg 2002;28:243 247. 16. Holmen JB, Ekesten B, Lundgren B. Anterior chamber depth estimation by Scheimpflug photography. Acta Ophthalmol Scand 2001;79:576 579. 17. Liu Z, Huang AJ, Pflugfelder SC. Evaluation of corneal thickness and topography in normal eyes using the Orbscan corneal topography system. Br J Ophthalmol 1999;83:774 778. 18. Sacu S, Findl O, Buehl W, et al. Optical biometry of the anterior eye segment: inter- and intra-examiner reliability of the ACMaster. J Cataract Refract Surg, in press. 19. Bland JM, Altman DG. Measuring agreement in method comparison studies. Stat Methods Med Res 1999;8:135 160. 20. Bland JM, Altman DG. Statistical methods for assessing agreement between two methods of clinical measurement. Lancet 1986;2:307 310. REPORTING VISUAL ACUITIES The AJO encourages authors to report the visual acuity in the manuscript using the same nomenclature that was used in gathering the data provided they were recorded in one of the methods listed here. This table of equivalent visual acuities is provided to the readers as an aid to interpret visual acuity findings in familiar units. Table of Equivalent Visual Acuity Measurements Snellen Visual Acuities 4 Meters 6 Meters 20 Feet Decimal Fraction LogMAR 4/40 6/60 20/200 0.10 1.0 4/32 6/48 20/160 0.125 0.9 4/25 6/38 20/125 0.16 0.8 4/20 6/30 20/100 0.20 0.7 4/16 6/24 20/80 0.25 0.6 4/12.6 6/20 20/63 0.32 0.5 4/10 6/15 20/50 0.40 0.4 4/8 6/12 20/40 0.50 0.3 4/6.3 6/10 20/32 0.63 0.2 4/5 6/7.5 20/25 0.80 0.1 4/4 6/6 20/20 1.00 0.0 4/3.2 6/5 20/16 1.25 0.1 4/2.5 6/3.75 20/12.5 1.60 0.2 4/2 6/3 20/10 2.00 0.3 From Ferris FL III, Kassoff A, Bresnick GH, Bailey I. New visual acuity charts for clinical research. Am J Ophthalmol 1982;94:91 96. 12 AMERICAN JOURNAL OF OPHTHALMOLOGY JANUARY 2006

Biosketch Wolf BUEHL, MD Department of Ophthalmology Medical University of Vienna Waehringer Guertel 18-20, Vienna 1090, Austria wolf.buehl@meduniwien.ac.at phone: 43-1-40400-7944 Date of birth: 24.5.1974 2000: PhD Thesis ( Fully automated analysis of PCO ) 2000 2002: Research fellow in Ophthalmology, Medical University of Vienna Since 2002: Clinical fellow in Ophthalmology, Medical University of Vienna Primary research interests: Cataract surgery Intraocular lenses Posterior capsule opacification (PCO) Biometry Digital imaging of the anterior segment VOL. 141, NO. 1 THREE METHODS OF MEASURING CORNEAL THICKNESS 12.e1