Keratoconus Detection Using Corneal Topography
|
|
|
- Preston Britton Wiggins
- 10 years ago
- Views:
Transcription
1 Keratoconus Detection Using Corneal Topography Jack T. Holladay, MD, MSEE, FACS ABSTRACT PURPOSE: To review the topographic patterns associated with keratoconus suspects and provide criteria for keratoconus screening. METHODS: Case study using maps from the NIDEK OPD-Scan II and OPD Station to highlight patterns seen in keratoconic corneas. RESULTS: Five criteria are listed for the detection of keratoconus: 1) apex of the cone is not centered at the 6-o clock semi-meridian, 2) cone should appear round on the tangential map, 3) keratometry diopters, 4) corneal thickness at the apex of the cone is approximately 30 µm thinner than the corresponding distance above the pupil center, and 5) topographic patterns are not symmetric. CONCLUSIONS: Standard corneal topography maps and simple clinical evaluation criteria are fundamental to detect keratoconus suspects and keratoconus. [J Refract Surg. 2009;25:S958-S962.] doi: / x K eratoconus is a progressive corneal disease characterized by central thinning and steepening of the corneal curvature. The progressive nature of this disease leads to increased myopia and irregular corneal astigmatism, which decrease visual acuity and visual quality. 1 Onset is often during the second decade of life onwards, typically when the patient is still socially and physically active. The reduced visual quality leads many patients with keratoconus to present at refractive surgery centers for alleviation of their symptoms via LASIK. Keratoconus and forme fruste keratoconus are contraindications for LASIK due to the high risk of postoperative ectasia. The incidence of keratectasia after LASIK is approximately 0.5%. 2 Early detection of forme fruste keratoconus and keratoconus is often performed by thorough topographic evaluation in conjunction with the clinical examination. Improvement in the ability to detect and diagnose suspicious to advanced keratoconus will enable the exclusion of patients at risk for corneal ectasia after corneal refractive surgery. Earlier detection of forme fruste keratoconus may lead to earlier intervention. The specificity and sensitivity of the various automated keratoconus screening software are still not adequate for broad clinical use. 3,4 Most of these automated detection programs cannot accurately classify suspicious topographies with an adequate degree of reliability. 3,4 It is precisely these suspicious cases that cause clinicians the most difficulty. To date, a thorough clinical and topographic examination by the ophthalmologist is still requisite for keratoconus detection. This article presents the use of the OPD- Scan II (Optical Path Difference Scanning System II; NIDEK Co Ltd, Gamagori, Japan) combined with clinical features to screen for keratoconus. From Baylor College of Medicine, Houston, Tex. The author is a consultant to NIDEK Co Ltd, Gamagori, Japan. Presented at the NIDEK NAVEX Seminar, World Congress of Ophthalmology; June 30, 2008; Hong Kong, China. Correspondence: Jack T. Holladay, MD, MSEE, FACS, Holladay Consulting Inc, PO Box 717, Bellaire, TX [email protected] S958
2 Figure 1. Schematic of axial topography including a sample topography of the left eye. Yellow denotes steeper curvature than green. CT axis = corneal topography axis Figure 2. Schematic of instantaneous topography including sample topography of the left eye (same as Figure 1). Yellow denotes steeper curvature than green. CT axis = corneal topography axis SCREENING FOR KERATOCONUS A number of topography maps are available on the OPD-Scan II. The most commonly used are the axial map, instantaneous map, and refractive map. Axial corneal topography measures all curvatures relative to the corneal topography axis (vertex normal). All radii are measured relative to this axis. Due to the common reference axis, small surface irregularities may not be visible (Fig 1). The instantaneous map, or local radius of curvature map, has no reference axis and only determines the local radius of curvature (instantaneous radius of curvature) at each point. It shows much more detail in the actual geometry of the cornea (Fig 2). The refractive map is similar to the axial map, but uses Snell s law with ray tracing to determine the power at each point on the cornea rather than the keratometric approximation (Gaussian optics) used for axial maps. The refractive map plots the refractive power of the corneal surface at each point. For example, if the cornea was spherical (of course it is not, it is a prolate ellipsoid) with a radius of 7.50 mm, the corneal power would be diopters (D) at all points using the keratometric index of refraction (1.3375). The map would look uniform in power at all points. If, however, Snell s law was used, the actual dioptric power would be found to increase when moving towards the periphery (spherical aberration). With the refractive power map, the increase in power would be apparent. The refractive map is especially useful when trying to correlate topographic power changes over the pupil with refractive changes in the eye. CASE EXAMPLE A case example of keratoconus is presented to highlight the criteria used for keratoconus detection that Journal of Refractive Surgery Volume 25 October (Suppl) 2009 S959
3 Figure 3. Refractive (left images) and instantaneous corneal (right images) topographies of A) a patient with keratoconus and B) a patient with contact lens warpage. includes evaluation of corneal topography. The following criteria are used to screen for keratoconus: 1. Use the instantaneous map to determine the apex of the cone. The apex of a keratoconic eye is usually not centered at the 6-o clock semi-meridian, but on either side. Contact lens warpage is almost always centered at 6 o clock. 2. The cone should appear round on the tangential map, whereas with contact lens warpage, it looks like a smile (arcuate in shape). 3. The keratometry values usually are D in keratoconus and flatter with contact lens warpage. 4. Measure the corneal thickness at the apex of the cone and at the corresponding distance above (opposite) the pupil center. In keratoconus, the apex will be approximately 30 µm (or more) thinner than the corresponding distance above the pupil, whereas in normal corneas with contact lens warpage, the lower cornea is rarely greater than 20 µm thinner. 5. The topographic patterns are rarely symmetric in keratoconus and almost always symmetric for contact lens warpage. Sample topographies are presented in Figure 3. The eye with keratoconus shows a cone that is not centered at 6 o clock whereas the eye with contact lens warpage has an arc-shaped pattern ( smile ) centered at 6 o clock (see Fig 3). Complementary clinical information is provided by the refractive wavefront maps and the modulation transfer function (MTF) from the OPD-Scan II (Fig 4). Although keratoconus and contact lens warpage both cause a reduction in the optical performance of the eye, the OPD map in diopters (rather than microns) provides much better information to the clinician than standard wavefront maps. A side-by-side comparison of the refractive wavefront map plotted in diopters (OPD map) and the traditional wavefront maps plotted in microns (wavefront higher order [HO]) shows little clinical information is garnered by evaluating the wavefront higher order maps (see Fig 4). The OPD map plots the change in refractive power across the pupil directly due to the cornea (approximately D in this case) (see Fig 4). The refractive gradient across the pupil due to the higher order aberrations is plotted in the OPD HO map. This map shows a pattern similar to the OPD map, indicating a major portion of the refractive error is due to the corneal distortion induced by the keratoconus (see Fig 4). The internal OPD map shows hyperopic and myopic areas bisecting the pupil indicative of the corneal distortion extending into the cornea below the front corneal surface (see Fig 4). The MTF graph plots the quality of the image across a range of spatial frequencies. In this case, the MTF indicates that correcting lower order aberrations alone will not have a significant effect on increasing visual performance across the range of spatial frequencies for daily living (see Fig 4). The range of clinical variability seen in practice means ambiguous cases will always arise. For example, a 30-year-old man presents with a suspicious topogra- S960
4 Keratoconus Screening/Holladay Figure 4. Wavefront measurement of an eye with keratoconus. OPD, OPD higher order (HO), and internal OPD maps are refractive wavefront maps plotted in diopters. The wavefront HO map is a wavefront map plotted in microns. MTF = modulation transfer function phy and a hot spot centered at the 4-o clock meridian. A repeat corneal topography is recommended to ensure accurate measurement followed by corneal pachymetry to determine whether a 30-µm difference exists as outlined in the criteria above. Subsequently, it is determined whether refractive change has occurred over the Journal of Refractive Surgery Volume 25 October (Suppl) 2009 past 5 years (approximately) consistent with increasing against-the-rule astigmatism. If corneal thickness is thinner and against-the-rule astigmatism has increased, this is considered indicative of keratoconus. Follow-up at 6-month intervals is warranted in such cases to determine the progression of this disease. S961
5 Corneal topography combined with clinical evaluation is required for the definitive diagnosis or exclusion of keratoconus. Evaluation of corneal curvature maps in addition to refractive wavefront maps and the five outlined criteria help determine the severity of the disease and its effects on visual performance. REFERENCES 1. Ertan A. Differentiating keratoconus and pellucid marginal degeneration. J Refract Surg. 2007;23: Randleman JB, Russell B, Ward MA, Thompson KP, Stulting RD. Risk factors and prognosis for corneal ectasia after LASIK. Ophthalmology. 2003;110: Fam HB, Lim KL. Corneal elevation indices in normal and keratoconic eyes. J Cataract Refract Surg. 2006;32: Chastang PJ, Borderie VM, Carvajal-Gonzalez S, Rostene W, Laroche L. Automated keratoconus detection using the EyeSys videokeratoscope. J Cataract Refract Surg. 2000;26: S962
Validation of a New Scoring System for the Detection of Early Forme of Keratoconus
10.5005/jp-journals-10025-1019 Alain Saad, Damien Gatinel ORIGINAL ARTICLE Validation of a New Scoring System for the Detection of Early Forme of Keratoconus Alain Saad, Damien Gatinel ABSTRACT Purpose:
How do we use the Galilei for cataract and refractive surgery?
How do we use the Galilei for cataract and refractive surgery? Douglas D. Koch, MD Mariko Shirayama, MD* Li Wang, MD, PhD* Mitchell P. Weikert, MD Cullen Eye Institute Baylor College of Medicine Houston,
Customized corneal ablation can be designed. Slit Skiascopic-guided Ablation Using the Nidek Laser. Scott MacRae, MD; Masanao Fujieda
Slit Skiascopic-guided Ablation Using the Nidek Laser Scott MacRae, MD; Masanao Fujieda ABSTRACT PURPOSE: To present the approach of using a scanning slit refractometer (the ARK 10000) in conjunction with
Case Reports Post-LASIK ectasia treated with intrastromal corneal ring segments and corneal crosslinking
Case Reports Post-LASIK ectasia treated with intrastromal corneal ring segments and corneal crosslinking Kay Lam, MD, Dan B. Rootman, MSc, Alejandro Lichtinger, and David S. Rootman, MD, FRCSC Author affiliations:
Post LASIK Ectasia. Examination: Gina M. Rogers, MD and Kenneth M. Goins, MD
Post LASIK Ectasia Gina M. Rogers, MD and Kenneth M. Goins, MD October 6, 2012 Chief Complaint: Decreasing vision after laser- assisted in- situ keratomileusis (LASIK) History of Present Illness: This
The Pentacam: Precision, Confidence, Results, and Accurate Ks!
The Pentacam: Precision, Confidence, Results, and Accurate Ks! Produced under an unrestricted educational grant from Oculus, Inc. This monograph is based on an informational symposium presented at the
THE BEST OF BOTH WORLDS Dual-Scheimpflug and Placido Reaching a new level in refractive screening
THE BEST OF BOTH WORLDS Dual-Scheimpflug and Placido Reaching a new level in refractive screening GALILEI G4 Clinical Applications Corneal Implant Planning The comes with a licensable corneal inlay software
Customized corneal ablation and super vision. Customized Corneal Ablation and Super Vision
Customized Corneal Ablation and Super Vision Scott M. MacRae, MD; James Schwiegerling, PhD; Robert Snyder, MD, PhD ABSTRACT PURPOSE: To review the early development of new technologies that are becoming
Alain Saad, MD, Alice Grise-Dulac, MD, Damien Gatinel, MD, PhD
CASE REPORT Bilateral loss in the quality of vision associated with anterior corneal protrusion after hyperopic LASIK followed by intrastromal femtolaser-assisted incisions Alain Saad, MD, Alice Grise-Dulac,
KERATOCONUS IS A BILATERAL, ASYMMETRIC, CHRONIC,
Comparison of and Intacs for Keratoconus and Post-LASIK Ectasia MUNISH SHARMA, MD, AND BRIAN S. BOXER WACHLER, MD PURPOSE: To evaluate the efficacy of single-segment Intacs and compare with double-segment
CORNEAL TOPOGRAPHY. Refractive power of the eye
CORNEAL TOPOGRAPHY Anne Faucher, M.D., FRCS University of Toronto Refractive power of the eye Eye has 3 refractive elements: 1. Cornea 2. Lens 3. Axial length Cornea (air/tear film interface) provides
Ectasia after laser in-situ keratomileusis (LASIK)
Ectasia after laser in-situ keratomileusis (LASIK) 長 庚 紀 念 醫 院 眼 科 蕭 靜 熹 Post-LASIK ectasia A rare complication of LASIK Manhattan jury awarded a former investment banker a record $7.25 million for post-lasik
Irregular astigmatism:
Irregular astigmatism: definition, classification, topographic and clinical presentation Ming X. Wang, MD, PhD Clinical Associate Professor of Ophthalmology of University of Tennessee Director, Wang Vision
Map Interpretation Guide. Map Interpretation Guide. Software Version 5.2
Map Interpretation Guide Map Interpretation Guide Software Version 5.2 Map Interpretation Guide Software Version 5.2 contents Foreword... 4 Introduction... 5 Getting Started... 6 1 The Refractive Report...
PRACTITIONER S FITTING GUIDE. Keratoconus Irregular Cornea Post Graft
PRACTITIONER S FITTING GUIDE Keratoconus Irregular Cornea Post Graft Four lens designs to fit all corneal shapes... One simple systematic approach to fitting Featuring Easy-to-fit using a simple systematic
Management of Unpredictable Post-PRK Corneal Ectasia with Intacs Implantation
Management of Unpredictable Post-PRK Corneal Ectasia with Intacs Implantation Mohammad Naser Hashemian, MD 1 Mahdi AliZadeh, MD 2 Hassan Hashemi, MD 1,3 Firoozeh Rahimi, MD 4 Abstract Purpose: To present
Page: 1 of 6. Corneal Topography/Computer-Assisted Corneal Topography/ Photokeratoscopy
Section: Surgery Effective Date: July 15, 2015 Last Review Status/Date: June 2015 Page: 1 of 6 Corneal Topography/ Photokeratoscopy Description Computer-assisted topography/photokeratoscopy provides a
Excimer Laser Refractive Surgery
Excimer Laser Refractive Surgery In the field of ophthalmology has achieved great technological advances and, undoubtedly, the most representative have focused on refractive surgery, which aims to eliminate
7/24/2014 ORBSCAN. Dr. HARITHA
ORBSCAN Dr. HARITHA Cornea is the most powerful refractive element, contributing 43D of total 60D power of eye. Anterior surface elliptical, HD-11.5mm VD-10.6mm. Posterior surface- circular,avg D- 11.5mm
ALL-IN-ONE Optical Biometry, Dual Scheimpflug Tomography and Placido Topography
ALL-IN-ONE Optical Biometry, Dual Scheimpflug Tomography and Placido Topography GALILEI G6 Clinical Applications High confidence premium IOL selection The GALILEI G6 offers a link to the ray-tracing software
SCHWIND CAM Perfect Planning wide range of applications
SCHWIND CAM Perfect Planning wide range of applications ORK-CAM PresbyMAX PALK-CAM PTK-CAM 2 SCHWIND CAM the system solution The latest version of the modular SCHWIND CAM represents an even more efficient
Pentacam. The. The gold standard in anterior segment tomography. Supplement to February 2011
Supplement to February 2011 The Pentacam The gold standard in anterior segment tomography. Produced under an educational grant from Oculus Optikgeraete GmbH. THE PENTACAM The Pentacam: An Impressive Technology
Wavefront-guided Custom Ablation for Myopia Using the NIDEK NAVEX Laser System
Wavefront-guided Custom Ablation for Myopia Using the NIDEK NAVEX Laser System Jan Venter, MD ABSTRACT PURPOSE: To determine the predictability, effi cacy, safety, and stability of LASIK using custom ablation
How To Implant A Keraring
Corneal Remodeling Using the Keraring A variety of thicknesses, arc lengths, and optical zone sizes allows tailoring of the procedure to the individual patient. BY DOMINIQUE PIETRINI, MD; AND TONY GUEDJ
final corrected draft
Archived at the Flinders Academic Commons http://dspace.flinders.edu.au/dspace/ This is the author s final corrected draft of this article. It has undergone peer review. Citation for the publisher s version:
ABrand New Refraction Meas urement Instrument
PW ABrand New Refraction Meas urement Instrument The TOPCON KR-9000PW WAVEFRONT ANALYZER incorporates comprehensive measurement of the eye's total optical system which has not been achieved until now,
Total Corneal Power Estimation: Ray Tracing Method versus Gaussian Optics Formula PATIENTS AND METHODS
Cornea Total Corneal Power Estimation: Ray Tracing Method versus Gaussian Optics Formula Li Wang, 1 Ashraf M. Mahmoud, 2 Betty Lise Anderson, 3 Douglas D. Koch, 1 and Cynthia J. Roberts 2 PURPOSE. To evaluate
Course # Intra Corneal Ring Segments Contact Lens Management of Irregular Astigmatism
Course # 772 Intra Corneal Ring Segments Contact Lens Management of Irregular Astigmatism Intra Corneal Ring Segments Contact Lens Management of Irregular Astigmatism Financial Disclosure I do not own
Cornea and Refractive Surgery Update
Cornea and Refractive Surgery Update Fall 2015 Optometric Education Dinner Sebastian Lesniak MD Matossian Eye Associates Disclosures: None Bio: Anterior Segment and Cornea Surgery Fellowship Wills Eye
Incision along Steep Axis
Toric IOL An option or a must? ~ 15% cataract surgical patients >1.5 D Options: spectacles, CLs, Incision along steep axis, LRI, AK, toric IOL, Excimer Laser or a combination Walter J. Stark, MD Professor
IOL Power Calculations for Postrefractive Surgery Eyes
eyetube.net The Creator s Forum: IOL Power Calculations for Postrefractive Surgery Eyes Olsen Formula By Thomas Olsen, MD, PhD The calculation of IOL power in an eye that has undergone corneal refractive
Single-Segment and Double-Segment INTACS for Post-LASIK Ectasia. Received: 8 Mar. 2013; Accepted: 8 Oct. 2013
ORIGINAL ARTICLE Single-Segment and Double-Segment INTACS for Post-LASIK Ectasia Hassan Hashemi 1, Ali Gholaminejad 1, Kazem Amanzadeh 1, Maryam Hashemi 2, and Mehdi Khabazkhoob 3 1 Noor Ophthalmology
The effect of corneal wavefront aberrations on corneal pseudo-accommodation
The effect of corneal wavefront aberrations on corneal pseudo-accommodation Li Wang, MD, PhD Elizabeth Yeu, MD, Douglas D. Koch, MD Cullen Eye Institute Baylor College of Medicine, Houston, TX Background
Rediscover quality of life thanks to vision correction with technology from Carl Zeiss. Patient Information
Rediscover quality of life thanks to vision correction with technology from Carl Zeiss Patient Information 5 2 It was really w Vision defects: Light that goes astray For clear vision the eyes, cornea and
Corneal Power Measurements Using Scheimpflug Imaging in Eyes With Prior Corneal Refractive Surgery
Corneal Power Measurements Using Scheimpflug Imaging in Eyes With Prior Corneal Refractive Surgery Jack T. Holladay, MD, MSEE, FACS; Warren E. Hill, MD, FACS; Andreas Steinmueller, MSc ABSTRACT PURPOSE:
Dr. Booth received his medical degree from the University of California: San Diego and his bachelor of science from Stanford University.
We've developed this handbook to help our patients become better informed about the entire process of laser vision correction. We hope you find it helpful and informative. Dr. Booth received his medical
Corneal intrastromal implantation surgery for the treatment of moderate and high myopia
TECHNIQUE Corneal intrastromal implantation surgery for the treatment of moderate and high myopia Albert Daxer, MD, PhD I describe a corneal intrastromal implantation technique that uses a new type of
IOL Power Calculation After Myopic LASIK. Hany Helaly, Lecturer of Ophthalmology, Faculty of Medicine, Alexandria University.
IOL Power Calculation After Myopic LASIK Hany Helaly, Lecturer of Ophthalmology, Faculty of Medicine, Alexandria University. SUPERVISORS Prof. Dr. Mohammad El-Hifnawy Professor of Ophthalmology Faculty
Kerry D. Solomon, MD, is Director of the Carolina Eyecare Research Institute at Carolina Eyecare Physicians in Charleston, S.C.
I think the ideal diagnostic technology for all of us would be a device where we could take a measurement, make an adjustment based on the patient s history, including past surgery, and come up with an
The Evolution of the Optical Zone in Corneal Refractive Surgery. Bruce Drum, Ph.D.
The Evolution of the Optical Zone in Corneal Refractive Surgery. Bruce Drum, Ph.D. FDA, Division of Ophthalmic and ENT Devices, Rockville, MD Disclaimer This presentation represents the professional opinion
The Pentacam: Offering a Clearer View
Supplement to Sponsored by an educational grant from Oculus, Inc. February 2008 The Pentacam: Offering a Clearer View Densitometry software for cataracts Topography for premium-channel cataract surgery
Life Science Journal 2014;11(9) http://www.lifesciencesite.com. Cross cylinder Challenging cases and their resultswith Nidek Quest (EC-5000)
Cross cylinder Challenging cases and their resultswith Nidek Quest (EC-5000) Gamal Mostafa Abo El Maaty, Mohamed Elmoddather, Mahmoud Ibrahem Ghazy, Mohamed Al-Taher Ophthalmology Department, Faculty of
AVOIDING VISUAL COMPLAINTS WITH PREMIUM IOLS Total corneal power must be evaluated to achieve success with multifocal and toric IOLs.
Supplement to October 2013 ALL-IN-ONE The most complete solution for refractive and cataract surgery. CONTENTS TECHNOLOGICAL ADVANCES OF THE GALILEI All you need for daily clinical practice, in one device.
Master of Engineering in Electronic Instrumentation and Control. Submitted By. SARAH ALI HASAN Roll No: 801251019
Novel Algorithm to Differentiate Astigmatism from Keratoconus A Dissertation submitted in partial fulfilment of the requirements for the award of degree of Master of Engineering in Electronic Instrumentation
Wavefront Analysis in Post-LASIK Eyes and Its Correlation with Visual Symptoms, Refraction, and Topography
Analysis in Post-LASIK Eyes and Its Correlation with Visual Symptoms, Refraction, and Topography Maria Regina Chalita, MD, 1 Sai Chavala, MD, 1 Meng Xu, MS, 2 Ronald R. Krueger, MD, MSE 1 Purpose: To evaluate
MAke A difference in someone s life
MAke A difference in someone s life Fitting guide En del av Multilens contents Introduction 1 Applications 2 characteristics 2 Fitting philosophy 3 Fitting process 4 Ideal fit 5 Comfort vs lens awarness
Refractive Errors. Refractive Surgery. Eye Care In Modern Life. Structure of the Eye. Structure of the Eye. Structure of the Eye. Structure of the Eye
Structure of the Eye Eye Care In Modern Life Dr. Dorothy Fan Department of Ophthalmology & Visual Sciences September 2007 Information age > 90% of sensory input Blindness is one of the most fearful disabilities
Eye Care In Modern Life
Eye Care In Modern Life Dr. Dorothy Fan Department of Ophthalmology & Visual Sciences November 2009 [email protected] Structure of the Eye Information age > 90% of sensory input Blindness is one of
The future of laser refractive surgery is exciting
The Cornea is Not a Piece of Plastic Cynthia Roberts, PhD Editorial The future of laser refractive surgery is exciting with the potential for ever-improved postoperative visual performance. In the past,
Surgical Advances in Keratoconus. Keratoconus. Innovations in Ophthalmology. New Surgical Advances. Diagnosis of Keratoconus. Scheimpflug imaging
Surgical Advances in Keratoconus Keratoconus Ectatic disorder 1 in 1,000 individuals Starts in adolescence & early adulthood Uncertain cause 20% require corneal transplant Innovations in Ophthalmology
Refractive Surgery. Evolution of Refractive Error Correction
Refractive Surgery Techniques that correct for refractive error in the eye have undergone dramatic evolution. The cornea is the easiest place to place a correction, so most techniques have focused on modifying
Looking for Keratoconus
07.31.06 Screening for LASIK: Tips and Techniques Surgeons share helpful strategies for identifying which patients are good candidates and which could be trouble. Christopher Kent, Senior Editor Screening
INTRODUCTION. Trans Am Ophthalmol Soc 2006;104:402-413
ORBSCAN II ASSISTED INTRAOCULAR LENS POWER CALCULATION FOR CATARACT SURGERY FOLLOWING MYOPIC LASER IN SITU KERATOMILEUSIS (AN AMERICAN OPHTHALMOLOGICAL SOCIETY THESIS) BY Henry Gelender MD ABSTRACT Purpose:
OCULUS PENTACAM BASIC CLASSIC HR
OCULUS PENTACAM BASIC CLASSIC We focus on progress Many Patients have Apprehensions before an Examination. > The eye is an extremely sensitive area Loss or impairment of sight always produces anxiety.
Comparison Combined LASIK Procedure for Ametropic Presbyopes and Planned Dual Interface for Post-LASIK Presbyopes Using Small Aperture Corneal Inlay
Comparison Combined LASIK Procedure for Ametropic Presbyopes and Planned Dual Interface for Post-LASIK Presbyopes Using Small Aperture Corneal Inlay Minoru Tomita, MD, PhD 1,2 1) Shinagawa LASIK, Tokyo,
OCT-based IOL power calculation for eyes with previous myopic and hyperopic laser vision correction
American Academy of Ophthalmology Annual Meeting Orlando, Florida, 21-25 October 2011 OCT-based IOL power calculation for eyes with previous myopic and hyperopic laser vision correction Weeks Professor
Vision Glossary of Terms
Vision Glossary of Terms EYE EXAMINATION PROCEDURES Eyeglass Examinations: The standard examination procedure for a patient who wants to wear eyeglasses includes at least the following: Case history; reason
Refractive surgeons routinely face the challenge of
Tomographic Screening for Ectasia Susceptibility Analysis must go beyond corneal curvature and central thickness. By Renato Ambrósio Jr, MD, PhD; Isaac Ramos, MD; Fernando Faria Correia, MD; and Michael
Dr. Hijab Mehta - MS, DOMS, FCPS Dr. S. Natarajan DO Dr. Hitendra Mehta MS
Customized Ablation Dr. Hijab Mehta - MS, DOMS, FCPS Dr. S. Natarajan DO Dr. Hitendra Mehta MS such as the patient s pupil size, in dim and bright light, corneal diameter and thickness are important considerations
refractive surgery a closer look
2011-2012 refractive surgery a closer look How the eye works Light rays enter the eye through the clear cornea, pupil and lens. These light rays are focused directly onto the retina, the light-sensitive
415.922.9500 www.pacificvision.org
Issue 046 September 2012 efocus Excellence in Co-Managed Care P A C I F I C V I S I O N I N S T I T U T E Life in Focus 415.922.9500 www.pacificvision.org Last month, we held a dinner with the Bay Area
The pinnacle of refractive performance.
Introducing! The pinnacle of refractive performance. REFRACTIVE SURGERY sets a new standard in LASIK outcomes More than 98% of patients would choose it again. 1 It even outperformed glasses and contacts
Introducing TOPOGRAPHY-GUIDED REFRACTIVE SURGERY
Sponsored by Introducing TOPOGRAPHY-GUIDED REFRACTIVE SURGERY Results of the T-CAT Phase III Clinical Trial TOPOGRAPHY-GUIDED REFRACTIVE SURGERY Topography-Guided Custom Ablation Treatments (T-CAT) with
Refractive errors are caused by an imperfectly shaped eyeball, cornea or lens, and are of three basic types:
Tips on Lasik Eye Surgery If you re tired of wearing glasses or contact lenses, you may be considering Lasik eye surgery one of the newest procedures to correct vision problems. Before you sign up for
Highlights from the POWER FORUM a live discussion held during the 2011 meeting of the American Academy of Ophthalmology
Larry Patterson, MD Moderator Arturo Chayet, MD Highlights from the POWER FORUM a live discussion held during the 2011 meeting of the American Academy of Ophthalmology Damien Gatinel, MD Mitchell A. Jackson,
Alexandria s Guide to LASIK
Alexandria s Guide to LASIK A Community Service Project sponsored by: Wallace Laser Center Your Guide To A Successful LASIK Procedure The word LASIK is actually an acronym for Laser Assisted In-Situ Keratomileusis.
Simple regression formula for intraocular lens power adjustment in eyes requiring cataract surgery after excimer laser photoablation
J CATARACT REFRACT SURG - VOL 32, MARCH 26 Simple regression formula for intraocular lens power adjustment in eyes requiring cataract surgery after excimer laser photoablation Samuel Masket, MD, Seth Everett
5/24/2013 ESOIRS 2013. Moderator: Alaa Ghaith, MD. Faculty: Ahmed El Masri, MD Mohamed Shafik, MD Mohamed El Kateb, MD
ESOIRS 2013 Moderator: Alaa Ghaith, MD Faculty: Ahmed El Masri, MD Mohamed Shafik, MD Mohamed El Kateb, MD 1 A systematic approach to the management of Keratoconus through the presentation of different
TRUSTED LASIK SURGEONS. Eye Conditions Correctable by Refractive Surgical Procedures
Eye Conditions Correctable by Refractive Surgical Procedures How does the eye focus? Light rays are focused on to the retina (where the image is relayed to the brain) by the cornea and the lens of the
Common visual problems in older LASIK patients
丘 子 宏 LASIK 手 術 後 的 視 覺 Visual acuity:the measurement of high contrast Snellen acuity but not other functions under different condition Quality of vision: measure the visual functions in variable condition
Custom-Q Presbyopic LASIK
Custom-Q Presbyopic LASIK Dr. Arthur Cheng Consultant Ophthalmologist, Hong Kong Sanatorium & Hospital Honorary Clinical Associate Professor, The Chinese University of Hong Kong Hong Kong Sanatorium &
PRK Wavefront Guided idesign Photorefractive Keratectomy
PRK Wavefront Guided idesign Photorefractive Keratectomy What is PRK? PRK (photorefractive keratectomy) is the same laser procedure as LASIK. Like LASIK it involves the use of the cool energy of an Excimer
Insert to October 2013 THE COMPLETE PICTURE. Experiences with the ALADDIN system. Sponsored by Topcon
Insert to October 2013 THE COMPLETE PICTURE Experiences with the ALADDIN system Sponsored by Topcon THE COMPLETE PICTURE Advantages of The Aladdin How this all-in-one optical biometer edged out the IOLMaster
New topographic custom ablation procedure for treating irregular astigmatism post keratoplasty with high frequency (1 KHz) excimer laser.
New topographic custom ablation procedure for treating irregular astigmatism post keratoplasty with high frequency (1 KHz) excimer laser. G. COLONNA M.D., G. Lorusso M.D., S. Santoro M.D. ESCRS Berlin
Corneal Collagen Cross-Linking (CXL) With Riboflavin
Dr. Paul J. Dubord, MD, FRCSC Clinical Professor Department of Ophthalmology and Visual Sciences University of British Columbia Patient Information Guide Corneal Collagen Cross-Linking (CXL) With Riboflavin
Insert to. January 2011. Comfort and confidence for all IOL calculations
Insert to January 2011 Comfort and confidence for all IOL calculations Advanced Optical Biometry and Keratometry with LENSTAR LS 900 CONTENTS Lenstar and Ray-Tracing Calculations...3 By Jaime Aramberri,
Update on Post- Refractive Surgery IOL Calculations
Update on Post- Refractive Surgery IOL Calculations Hawaiian Eye 2014 Kauai Rhonda G Waldron MMSc, COMT, CRA, ROUB, RDMS Diagnostic Echographer, Senior Associate in Ophthalmology Emory Eye Center Atlanta,GA
Comparison of Corneal Power and Intraocular Lens Power Calculation Methods after LASIK for Myopia. Abstract
Comparison of Corneal Power and Intraocular Lens Power Calculation Methods after LASIK for Myopia Seyed Mohammad Reza Taheri, MD 1 Azita Kheiltash, MD, MPH 2 Hassan Hashemi, MD 1,3 Abstract Purpose: To
The Galilei Dual Scheimpflug Analyzer incorporates multiple diagnostic modalities into one device to map the cornea and anterior segment.
Supplement to March 2009 The Galilei Dual Scheimpflug Analyzer incorporates multiple diagnostic modalities into one device to map the cornea and anterior segment. Sponsored by an unrestricted educational
Overview of Refractive Surgery
Overview of Refractive Surgery Michael N. Wiggins, MD Assistant Professor, College of Health Related Professions and College of Medicine, Department of Ophthalmology Jones Eye Institute University of Arkansas
ADVANCED IOL POWER CALCULATIONS. Jack T. Holladay, MD, MSEE, FACS
Jack T. Holladay, M.D., M.S.E.E., F.A.C.S. ADVANCED IOL POWER CALCULATIONS Jack T. Holladay, MD, MSEE, FACS I. Formulas and Measurements A. Variables Used to Predict ACD 1. Binkhorst 2-1981 - AL 2. Holladay
ORIGINAL ARTICLES. Anastasios John Kanellopoulos, MD; Perry S. Binder, MS, MD
ORIGINAL ARTICLES Management of Corneal Ectasia After LASIK With Combined, Same-day, Topographyguided Partial Transepithelial PRK and Collagen Cross-linking: The Athens Protocol Anastasios John Kanellopoulos,
Cataract Testing. What a Patient undergoes prior to surgery
Cataract Testing What a Patient undergoes prior to surgery FINANCIAL DISCLOSURE I have no financial interest or relationships to disclose What do most Technicians find to be the most mundane yet very important
Topographically-guided Laser In Situ Keratomileusis to Treat Corneal Irregularities
Topographically-guided Laser In Situ Keratomileusis to Treat Corneal Irregularities Michael C. Knorz, MD, Bettina Jendritza, MD Objective: To evaluate the predictability and safety of topographically guided
Phacoemulsification: Considerations for Astigmatism Management Jason P. Brinton, MD and Thomas A. Oetting, MS, MD June 10, 2011
Phacoemulsification: Considerations for Astigmatism Management Jason P. Brinton, MD and Thomas A. Oetting, MS, MD June 10, 2011 Introduction Patient expectations for cataract surgery have increased significantly
Refractive Errors & Correction
Super Vision 6 Chapter Lasik is currently the most sophisticated procedure for correction of refractive errors. Lasik is an acronym for Laser Assisted Insitu Keratomileusis. In Lasik, Excimer laser is
Ectasia Risk Scoring System (ERSS): Limitations and Variability in Predictive Value
Ectasia Risk Scoring System (ERSS): Limitations and Variability in Predictive Value Jay S. Pepose, M.D., Ph.D. Director, Pepose Vision Institute Professor of Clinical Ophthalmology Washington University
PATIENT CONSENT FOR LASER IN-SITU KERATOMILEUSIS (LASIK)
INTRODUCTION: You have been diagnosed with myopia (nearsightedness) or hyperopia (farsightedness) with or without astigmatism, or astigmatism alone. Myopia is a result of light entering the eye and focusing