QuickAir - 5, - 10 Minute Emergency Escape Breathing Apparatus. Operating Manual



Similar documents
Your Portable Oxygen System. Use & Care Guide

Elgin Fire Department

INSPECTION, MAINTENANCE AND RECHARGE SERVICE MANUAL P/N 16303

COMPRESSED GASES. 1.2 The contents of each cylinder and container must be clearly identified (by tag or stamp) on the cylinder.

HEAD HARNESS & BUCKELS

SunMaxx Solar Filling Station Operating Instructions

SOLSTICETM International Leisure Products, Inc. 191 Rodeo Drive Edgewood, NY Tel (631)

Instruction Manual. Alfa Laval SB Membrane Sample Valve ESE02963-EN

Back Pack Sprayer. Operator's Manual MODELS MS - 40 MS - 50

STORED PRESSURE STAINLESS STEEL

Air-Oxygen Blender. Service Manual. Model No. PM5200 Series PM5300 Series (shown)

NITROUS TRANSFER PUMP INSTRUCTIONS

Compressed Gas Cylinder Program

Instructions For SF 6 Refill Kit

Iron Air Charge 2510-SXT Installation & Start-Up Guide

FIRE EXTINGUSIHER INSPECTION, MAINTENANCE, AND TESTING PROCEDURES

DON T FORGET TO REGISTER YOUR SPARE AIR ONLINE AT

Cleaning Instructions, Pilot Replacement and Valve Change. Model No.: Natural Gas

CARING FOR YOUR WATER HEATER

INSTALLATION AND OPERATION INSTRUCTIONS NOT FOR OXYGEN OR OTHER OXIDIZING GASES

3 WATT LED SPOTLIGHT Model No. SLM

PA94 Plus Basic. System Overview. Accessories/Options. PA94 Plus Basic Quick Release LDV Coupling PA94 Plus Basic Fixed LDV

FJ2. 2 Ton Trolley Floor Jack Assembly & Operating Instructions

User s Guide CARBON DIOXIDE CYLINDERS AND REGULATORS

Course Firefighter I. Unit II Safety and Orientation

Self-Contained Breathing Apparatus (SCBA) Program and Procedures for SCOTT AIR-PAK 75i

Reasons for reissue of this instruction sheet are provided in Section 5, REVISION SUMMARY.

MODEL G300 BRAKE BLEEDER

University of Tennessee Safety Guidelines

Cleaning Instructions for Burner, Pilot Assembly, and Emitter Screen Series:

Operating Instructions & Maintenance Guidelines for Precision Vaporizers

Ozone Generator Manual

Installation and Operating Instructions Installation Instructions for SS EPE-316L Series

Oxy-Fuel Gas Welding. Given a functional oxy-fuel gas unit, instruction and demonstration of use, each student will be able to:

Understanding Anesthetic Delivery Systems Dean Knoll, CVT, VTS (Anes.) Anesthesia Technician Supervisor University of Wisconsin Madison, WI May 2003

SALTER AIRE Plus COMPRESSOR Model 8350 / 8352 / Instruction Manual. Page 1 of 9

Dive Rite 200 & 300 Bar Isolator Manifold Service Manual

DTM04 TANK MONITOR DTM08 TANK MONITOR Dtm12 TANK MONITOR. Installation and Operation Manual

Spiromatic Face Mask and breathing valve User manual

San josé OWNER S MANUAL

U S E R M A N U A L. (CGA 870 Connection)

CHIPPING HAMMERS Tabor Place, Santa Fe Springs, California, CA 90670, U.S.A.

Product Manual. CVS Type 67AFR Filter Regulator. Introduction. Installation. Description

Air vacuum pump with r134a & r12/22

- Do not immerse cords, pump or any electrical wire or device in water or other liquid.

FIBERGLASS UNDERGROUND STORAGE TANKS Special Installation Instructions for New York City Gasoline and Diesel Motor Fuels

Failure to comply with the following cautions and warnings could cause equipment damage and personal injury.

PATIENT GUIDE. Care and Maintenance Drainage Frequency: Max. Drainage Volume: Dressing Option: Clinician s Signature: Every drainage Weekly

Model MS-41BP 4G/ 15L. WARNING Carefully Read These Instructions Before Use R0514

Please read and keep these instructions. Run water through your machine before first use

Respiratory Safety and PPE Module 7. Special Warehouse Worker Hazards in Structural Steel Fabricating and Supply Companies

Fluid Regulator. Instructions Parts List L LOW PRESSURE

=.03% CO2 10/4/2012. Carbon Dioxide (CO2) Safety. Carbon Dioxide (CO2) Safety. Carbon Dioxide (CO2) Safety. Carbon Dioxide (CO2) Safety

Product Name: Beamer 2000

THC 85 INDUSTRIAL / COMMERCIAL SPACE HEATER

SERVICE MANUAL NO for AMEREX HALOTRON I HAND PORTABLE CLEAN AGENT FIRE EXTINGUISHERS (FOR NON-RESIDENTIAL APPLICATIONS)

Reduced Pressure Principle Assembly 3 Valve Procedure

Important: Please read these instructions carefully and completely before starting the installation. TITAN Fuel Tanks

Rail Vehicle Systems IP-241. Rev 07 3/9/ en. Installation and Maintenance Guide. Slack Adjuster Model KRD-482-E and KRD-482-R P/N ,

PC1131 Electric Air Compressor

Hydrostatic Testing of Aviation Components

MANI-FLEX AutoSwitch

Brookings Fire Dept. SCBA Specifications

Cylinder HAZMAT & Fill Station Operations Compliance. PSI/PCI, Inc. William L. (Bill) High & Mark A. Gresham

Cleaning Instructions for Burner, Pilot Assembly, and Emitter Screen. Series: LP Gas

PROCEDURES PREVENTION OF ACCIDENTAL IGNITION

AIR HYDRAULIC RIVETER

E500 Sound Isolating Earphones

Recharge Procedure for Telemark TVP-2000 and TVP-3500

MODEL J SERIES CONVERTER

INSTALLATION INSTRUCTIONS Fan Coil Replacement Coil Kit EBX & EBXX

The troubleshooting guide is designed to provide a logical starting point to locate general problems that could

G26 Corp L11 Revised LENNOXOEMPARTS.COM. Service Literature G26 SERIES UNITS

The Breather Box. High Performance Grade-D Breathing Air Filtration system. BB50-cO

MSA Air-Line Respirators

National- Spencer Inc.

Compressor Service & Maintenance Manual

Tech. Services: (800) Fax: (800) Order Entry: (800) Fax: (800) Internet Address:

A Safe Workplace A Workplace Safety and Health Manual for Your Community Section: III-C Subject: G-9

THE UNIVERSITY OF MAINE COMPRESSED GAS CYLINDERS FOR SCUBA DIVING

CYLINDER REFILL AND INSPECTION

Bench Autoclave. Standard Operating Procedure. For Installation, Use and Maintenance

Continuous-Flow Supplied Air Respirator

ENGINE FUEL FUEL FILTER... FUEL HEATER... INJECTOR... SUPPLY PUMP... COMMON RAIL... FUEL PRESSURE LIMITTER...

Powers Controls Free Energy Band TH 193 HC Heating/Cooling Room Thermostat

DIAMOND Retractable Rodding Robot Model SPRAYROD-R

BINKS MODEL 460 LIGHTWEIGHT AUTOMATIC SPRAY GUN

Guidance for Those Who Use and Store Gas Cylinders

English. Symbols used to mark instructions...3. Congratulations...5 Getting the best results...5. Warnings...6 Operating Procedure...

COMPRESSED GASES REVISION DATE: 1/15/08

Compressed Gas Cylinder Safety

Cylinder Leak- Down Tester

ON TRACK Reprocessing Inservice / Competency for GI Endoscopes (EVIS, EXERA, EXERA II)

TITAN Fuel Tanks. INSTALLATION INSTRUCTIONS G e n e r a t i o n V

7500 Series. Important: Keep these user instructions for reference.

(1) Acceptable means having been inspected by the Commissioner or his/her representative and approved for the intended usage.

Operating and Maintenance Instructions

IMPORTANT SAFETY INFORMATION

THE NEWCASTLE UPON TYNE HOSPITALS NHS FOUNDATION TRUST SAFE USE AND STORAGE OF LIQUID NITROGEN AND SOLID CARBON DIOXIDE (DRY ICE)

Transcription:

QuickAir - 5, - 10 Minute Emergency Escape Breathing Apparatus Operating Manual

Table of Contents 1 General... 2 1.1 Purpose... 2 1.2 Description... 2 2 Operation... 3 2.1 Pre-donning Inspection... 3 2.1.1 Cylinder Pressure Check... 3 2.2 Donning the QuickAir... 3 2.3 End of Use... 4 3 Maintenance... 5 3.1 Cleaning/Disinfecting... 5 3.1.1 Pneumatics/Case... 5 3.1.2 Hood/Manifold... 5 3.2 Inspection... 6 3.2.1 Hood... 6 3.2.2 Carrying Bag... 6 3.2.3 Cylinder Pressure... 6 3.2.4 Filling Cylinders with Air... 6 3.3 Storage... 7 3.4 Maintenance Schedule... 7 3.4.1 Weekly Procedures... 7 3.4.2 Monthly Procedures... 8 3.4.3 Yearly Procedures... 8 4 Spare Parts and Accessories... 9 Draeger Safety, Inc. For correct and effective use of the apparatus and to avoid hazards, it is essential to read the following recommendations and to act accordingly. IMPORTANT NOTICE 1 Before using this apparatus, carefully read the operating manual. 2 The apparatus should be inspected by qualified and trained personnel before and after each use. 3 The apparatus is intended only for the purposes specified in the operating manual or for purposes confirmed by Draeger Safety. 4 Only original Draeger Safety replacement parts should be used for maintenance and repairs. Maintenance and the replacement of user replaceable parts should only be performed by Draeger Safety trained technicians or authorized Service Centers. 5 Responsibility for the reliable function of the apparatus passes to the owner or operator, in all cases, where the apparatus has not been maintained or repaired by persons as specified above, or where it has been used in a manner which does not conform to the intended use. 6 All federal, local and workplace recommendations, regulations and laws governing the use of technical equipment should be observed. 7 If you require any additional information on the use or replacement of this unit call Draeger Safety at 800-922-5518 or fax at 800-922-5519. 1 QuickAir

1 General 1.1 Purpose The Draeger Safety QuickAir Emergency Escape Breathing Apparatus is intended for use as a respiratory protection device in hazardous or oxygen deficient atmospheres. The primary mode of operation is intended for escape and emergency purposes only. 1.2 Description The QuickAir is a constant-flow self-contained escape breathing apparatus. It is available in 5 and 10 minute durations. The QuickAir consists of an aluminum cylinder and valve assembly which is protected by a flexible PVC carrying case. The carrying case is flame, mildew and UV resistant and is designed to be accessible in emergency situations whether the user is right or left handed. The QuickAir can be worn around the neck or diagonally over the shoulder and chest. The clear PVC hood is contained in the easily accessible pouch. The hood offers full visibility and good chemical resistance. An elastic collar is provided, on the hood, for comfort. A constant flow of air is delivered from the cylinder to the hood by the regulator and hose assembly. An initial flow of approximately 40 lpm is delivered to the hood when the cylinder valve is completely open. Also available is a high flow five minute unit with a flow rate of approximately 70 lpm for emergency situations that demand a higher work rate in order to exit to a safe area. The QuickAir will provide air to the user for 5 or 10 minutes depending on whether he is using the 5 or 10 minute version. Therefore, the escape route should be planned, by the safety administrator, prior to the escape situation to ensure the user will have enough time to reach a safe area before the air supply of the escape unit is depleted. 1 7 2 3 1 Hood 2 Elastic Collar 3 PVC Carrying Case 4 Cylinder 5 Regulator Assembly 6 Valve Assembly 7 Diffuser 6 5 4 Figure 1 QuickAir 2

2 Operation Draeger Safety, Inc. 2.1 Pre-donning Inspection NOTE The QuickAir should be adjusted to a comfortable position before an emergency situation to facilitate a rapid escape. 2.1.1 Cylinder Pressure Check Check weekly to ensure the cylinder is "FULL" by reading the pressure gauge. The indicator needle should be in the green region on the gauge. 2.2 Donning the QuickAir QuickAir is designed for emergency escape purposes only. Place the strap over the neck (Figure 2a) or diagonally over one shoulder and the chest (Figure 2b). Ensure the cylinder valve is facing forward for ease of access. Pull back velcro tab (Figure 3). Figure 2a Figure 2b Figure 3 3 QuickAir

Remove hood from pouch (Figure 4). Don hood by stretching the elastic collar open with both hands and pulling the hood over the head (Figure 5). IMMEDIATELY open cylinder valve completely. A rush of air should begin to flow into the hood (Figure 6). CAUTION If air does not flow to hood, DO NOT use respirator! Figure 4 Figure 5 Figure 6 Adjust the hood so that the diffuser/hose assembly is in front of the nose and mouth of the user. Exit to a safe area (Figure 7). 2.3 End of Use Remove hood from head by stretching the elastic collar open with both hands and pull hood off of head. Close cylinder valve completely. Remove unit by lifting strap over the head. NOTE After each use clean and disinfect the apparatus. Figure 7 QuickAir 4

3 Maintenance Draeger Safety, Inc. 3.1 Cleaning/Disinfecting 3.1.1 Pneumatics/Case Wipe off all surfaces with a mild soapy solution. Rinse thoroughly. Carefully wipe off all surfaces with a disinfectant solution of Airkem 33 or equivalent disinfectant. Rinse thoroughly. Dry completely. 3.1.2 Hood/Manifold Before cleaning and disinfecting, disassemble the manifold from the hood. Loosen the manifold from the inside of the hood with a spanner wrench (Figure 8) and unscrew the manifold halves. Figure 8 Immerse the entire hood in a solution of Airkem 33 or equivalent disinfectant. After disinfecting, rinse parts thoroughly in clear water and air dry completely. Reassemble by securing the manifold to the housing using the spanner wrench. CAUTION Care should be taken to ensure that the holes in the manifold are directed towards the top of the hood. 5 QuickAir

3.2 Inspection 3.2.1 Hood QuickAir 6 After each use, inspect the hood for any tears, cracking or holes. Remove from service if there is any sign of degradation. (Section 4, Spare Parts and Accessories) 3.2.2 Carrying Bag After each use, inspect the carrying bag for any signs of degradation. Ensure that all hardware is intact and in good working condition. Inspect the velcro to make sure it is free of lint or dirt that could prevent it from adhering. Remove from service if there is any sign of degradation. (Section 4, Spare Parts and Accessories) 3.2.3 Cylinder Pressure Recharge or replace the cylinder with a Draeger Safety approved cylinder charged to full service pressure. Use only cylinders with current hydrostatic test pressure dates. 3.2.4 Filling Cylinders with Air Safe and effective use of compressed air cylinders requires the user to adhere to the following guidelines: 1. Do Not over pressurize the cylinder. The rated cylinder service pressure is identified on the cylinder. The rated service pressure is either stamped into the shoulder of the cylinder or marked on the cylinder label. 2. When refilling the cylinder, use only Type 1, Grade D or better breathing air as identified in the Compressed Gas Association (CGA) Commodity Specification for Air, Pamphlet G-7.1. This pamphlet is available from the Compressed Gas Association, Inc., P.O. Box 6208, Falls Church, VA 22016. Additionally, the water vapor content of the breathing air should be less than 50 mg/m 3 for cylinders. 3. Periodically, and prior to refilling, check to ensure the cylinder hydrostatic retest date is within the prescribed time period. 4. Prior to refilling, inspect the cylinder for damage. Criteria for determining the extent of damage is provided in CGA pamphlet C-6.2. 5. Do not refill the cylinder at an excessive rate. A general industry standard is a rate less than 300 psi per minute. 6. After filling the cylinder with breathing air to the rated service pressure, check the following areas for leakage using a solution of soap and water (Leaktek; available from Draeger Safety): cylinder to cylinder valve connection cylinder valve burst disc assembly cylinder valve pressure gauge If leakage is detected, the cylinder should be removed from service and repaired by a qualified trained technician.

7. Check the cylinder pressure as indicated on the cylinder valve pressure gauge at regularly scheduled intervals. For stored apparatus, the check intervals should be weekly. For situations requiring frequent use, the pressure should be checked daily. If a loss in pressure is observed, remove the cylinder from service. 8. Always check the cylinder pressure immediately prior to use. 9. Do not modify the cylinder or any of its related components including safety related devices. 10. Always store compressed air cylinders in a full or empty condition. Storage of partially full cylinders can result in less than expected service life for the apparatus and also in faulty operation of pressure relief devices in the event of or exposure to excessive temperatures. WARNING If breathing air moisture content is greater than recommended levels the water can form ice particles and reduce or block the breathing air flow. The water content of breathing air in any container must be checked. 3.3 Storage Before storing the apparatus, fold the hood/hose assembly and place into the pouch. Close the velcro strap. Store the apparatus ready for use. A QuickAir storage case and wall mount kit are available. Test the units monthly even when they have not been used. The maintenance schedule must be followed to keep the apparatus in good working condition. 3.4 Maintenance Schedule Frequency Weekly Monthly Yearly Every 5 years Service ensure cylinder is full function testing component testing hydrotest aluminum cylinders 3.4.1 Weekly Procedures Ensure the indictor needle is in the green region "FULL" on the gauge. 7 QuickAir

3.4.2 Monthly Procedures QuickAir 8 Function Tests Check function of the apparatus by cracking the cylinder valve open for a few seconds. You should hear a rush of air going into the hood. Close the valve tightly. Inspect the unit for any signs of degradation. Remove the unit from service if it fails to meet the function tests. 3.4.3 Yearly Procedures Component Tests Equipment Escape Unit Test Kit P/N 4055044 1/2" wrench NOTE The SCBA Test Kit, P/N 4052350 and medium pressure adapter, P/N 4054919 may be used in place of the Escape Unit Test Kit, P/N 4055044 to perform the component tests. All repairs or replacement of parts must be performed by a Draeger Safety approved technician or an authorized service center. High Pressure Test Remove the regulator/hose assembly from the cylinder valve and attach the high pressure gauge to the cylinder valve. Open cylinder valve and note the pressure reading on the gauge. The readings on the cylinder gauge and the test gauge should be the same. Close the cylinder valve and release the pressure in the system with the relief valve control located on the test pressure gauge. Remove cylinder from service if discrepancies in the cylinder pressure occur. Medium Pressure Test Disconnect the regulator from the compressed air cylinder. Disconnect the regulator barb from the regulator. (Use a 1/2" wrench). Replace the regulator barb with the medium pressure test pigtail. Connect medium pressure gauge to safety-type plug-in coupling on the test pigtail. Connect the regulator to the compressed air cylinder. Open the cylinder valve. Read the medium pressure on the test gauge. The pressure should be between 85-120 psi. If the medium pressure is not within range, replace the regulator. Close the cylinder valve. Open the purge valve on test pressure gauge, relieve pressure and close the valve. Disconnect the test pressure gauge and test pigtail. Reassemble the apparatus line (regulator barb) to the regulator. NOTE When releasing the pressure, ensure that face and hands are clear of vent port.

4 Spare Parts and Accessories Description Part Number QuickAir - 5, Complete 4054953 QuickAir - 10, Complete 4054954 QuickAir - H5, Complete 4054966 QuickAir 5 2216, Complete 4055438 Hood, QuickAir 4055051 Regulator, Emergency Escape 4055052 Regulator, High Flow, Emergency Escape 4055053 O-ring, Ballnose, QuickAir 4055054 Carrying Bag, 5 Minute 4055055 Carrying Bag, 5 Minute, High Flow 4055056 Carrying Bag, 5 Minute/2216 4055434S Carrying Bag, 10 Minute 4055057 Hose Retainer Assembly 4055058 Diffuser, QuickAir 4055059 5 min/3000 psi, Cylinder/Valve Assembly 4055060 10 min & 5 min HF/3000 psi, Cylinder/Valve Assembly 4055061 5 min/2216 Cylinder/Valve Assembly 4055437 Storage Case, QuickAir 10 Minute/5 Minute High Flow 4055041 Storage Case, QuickAir 5 Minute 4055042 Wall Mount Kit 4055000 Escape Unit Test Kit 4055044 Spanner Wrench 1628089 Cylinder Fill Adapter 4052794 Leaktek Soap Solution 4054084 Airkem 33 4054085 For complete illustrations and parts listings, consult Draeger Safety, Inc., customer service or your local distributor. 9 QuickAir

Notes QuickAir 10

Notes Draeger Safety, Inc. 11 QuickAir

These instructions for use apply only to the unit with this serial number: If no serial number appears in the box above, these instructions were provided for general information only and are not meant to be used with a particular device. Draeger Safety, Inc. 101 Technology Drive P.O. Box 120 Pittsburgh, PA 15230-0120 TELEPHONE: (412) 787-8383 CUSTOMER SERVICE TEL: 800-922-5518 CUSTOMER SERVICE FAX: 800-922-5519 http://www.draeger-usa.com P/N 4054955 - Rev. 4 - ERMA - F001