GMP 11. Good Measurement Practice for Assignment and Adjustment of Calibration Intervals for Laboratory Standards



Similar documents
Calibration Service Selection Guide

Calibration Certificate

Calibration & Preventative Maintenance. Sally Wolfgang Manager, Quality Operations Merck & Co., Inc.

Gravimetric determination of pipette errors

User checks and maintenance of laboratory balances

Estimates of Uncertainty of the Calibration of Balances

Case study Realistic and best uncertainties with humidity instruments

Process Measurement Assurance Program For U.S. State Metrology Laboratories 1

CALIBRATION PRINCIPLES

Assessment of Accuracy and Precision

State Laboratory Program Workload Survey

Infrared Thermometer Calibration A Complete Guide

State Water Resources Control Board Environmental Laboratory Accreditation Program

calibrate confidence with Metrology Concepts: Understanding Test Uncertainty Ratio (TUR) Abstract Introduction White Paper

Determination of Moisture Content

Procedure for Equipment Calibration and Maintenance

OMCL Network of the Council of Europe QUALITY MANAGEMENT DOCUMENT

Temperature Calibration; Depths of Immersion

1 Quality Assurance and Quality Control Project Plan

ENVIRONMENTAL LABORATORY APPROVAL PROGRAM CERTIFICATION MANUAL

ALACC Frequently Asked Questions (FAQs)

Reaction of Magnesium with Hydrochloric Acid (Gas Laws) Chemicals Needed:

COPY. Revision History: Changes were made to reflect the current practice of only including the most recent changes in the revision history.

How to calibrate an RTD or Platinum Resistance Thermometer (PRT)

Experiment 1: Measurement and Density

Calibration of Volumetric Glassware

4.2 Bias, Standards and Standardization

Beamex. Calibration White Paper. How often should instruments be calibrated

Physical Properties of a Pure Substance, Water

A PRIMARY CALIBRATION SYSTEM FOR THE SUPPORT OF HIGH PERFORMANCE GAS FLOW TRANSFER STANDARDS

Ti-Pure TITANIUM DIOXIDE. DETERMINATION OF BROOKFIELD VISCOSITY, TiO 2 SLURRY METHOD: T WP

Figure 1. A typical Laboratory Thermometer graduated in C.

ACL 800 MEGOHMMETER OPERATION MANUAL

New York State Department of Health (NYSDOH) Bureau of Immunization Vaccine Storage and Handling Plan

IAS CALIBRATION and TESTING LABORATORY ACCREDITATION PROGRAMS DEFINITIONS

Beamex. Calibration White Paper. How often should instruments be calibrated?

Life & Times of a PM2.5 Filter: Quality Control in the Laboratory and Beyond

STANDARD OPERATING GUIDELINES: EVIDENTIAL BREATH ALCOHOL INSTRUMENT CALIBRATION

Calibration of temperature control and monitoring devices

3M CORPORATE METROLOGY SERVICES St. Paul, MN

Measurement and Calibration

Relative Humidity Calibration Kit

Measurement of RF Emissions from a Final Coat Electronics Corrosion Module

The Uncertainty of Reference Standards A Guide to Understanding Factors Impacting Uncertainty, Uncertainty Calculations, and Vendor Certifications

Calibration. How can you get lower costs for the calibration? What is traceability? What is measurement uncertainty? What is calibration?

The Basics of Fertilizer Calculations for Greenhouse Crops Joyce G. Latimer, Extension Specialist, Greenhouse Crops; Virginia Tech

NIST Calibration Services for Liquid Volume NIST Special Publication

CRM s and the Dilemma of the 2nd Source. Shawn Kassner, Sr Product Manager Tim Miller, Sr Organic Chemist Phenova, A Phenomenex Company

Quality Assurance Manual January 2008 Revision I Revision date of 01/17/13. Pi Tape Corporation. 344 North Vinewood St. Escondido CA USA

COMPLIANCE BY DESIGN FOR PHARMACEUTICAL QUALITY CONTROL LABORATORIES INSIGHT FROM FDA WARNING LETTERS

USE OF REFERENCE MATERIALS IN THE LABORATORY

A Determination of g, the Acceleration Due to Gravity, from Newton's Laws of Motion

ISO/IEC QUALITY MANUAL

Keysight Technologies Specifications Guidelines. Technical Overview

ISO Accreditation for a Standing Start Finish (SSF) Primary Flow Stand with Coriolis Flowmeters as Reference Standards.

1. The Determination of Boiling Point

How To Test A Titrette And Digital Burette

The Molar Mass of a Gas

TEST REPORT for the ARID TECHNOLOGIES VAPOR RECOVERY UNIT installed in LANTANA, FL

Custody Transfer Measurement. with the V-Cone Flowmeter

EXERCISE # 1.Metric Measurement & Scientific Notation

Glassware Calibration Guidelines Laura B. Secor and Dwight R. Stoll, 02/01/2012 Adapted from National Bureau of Standards Document

Hand-held thermometer Model CTH7000

Environment-Laboratory Ambient Conditions

Francis Mukalazi Stephen Fisheries Resources Department Uganda P.O. Box 4 Entebbe Tel francis@hafro.is or mukalazif@yahoo.

Estimating Weighing Uncertainty From Balance Data Sheet Specifications

11.I In-process control Authors: Dr. Christian Gausepohl / Paolomi Mukherji / Update 07

Calibration of Glassware

UNITED STATES CONSUMER PRODUCT SAFETY COMMISSION DIRECTORATE FOR LABORATORY SCIENCES DIVISION OF CHEMISTRY 5 RESEARCH PLACE ROCKVILLE, MD 20850

SUPPLIER Form #: 007 AUDIT QUESTIONNAIRE

ME Heat Transfer Laboratory. Experiment No. 7 ANALYSIS OF ENHANCED CONCENTRIC TUBE AND SHELL AND TUBE HEAT EXCHANGERS

Inventory Control. Inventory Control Tank & Silo Weighing

Understanding electrical conductivity

OVERVIEW OF STANDARD 7E & STANDARD 20

Millikan Oil Drop Experiment Matthew Norton, Jurasits Christopher, Heyduck William, Nick Chumbley. Norton 0

Phase Diagram of tert-butyl Alcohol

EQUIPMENT CALIBRATION/MAINTENANCE

Terms concerned with internal quality control procedures

Basic requirements for quality management in plant pest diagnosis laboratories

State Meat and Poultry Inspection (MPI) Program Laboratory Quality Management System Checklist

Testing Your Laboratory Balance

Calibration of Volumetric Glassware

ASSURING THE QUALITY OF TEST RESULTS

VERIFYING CALIBRATION OF WEIGHING DEVICES USED FOR LABORATORY TESTING

Experimental Uncertainties (Errors)

Validation and Calibration. Definitions and Terminology

Weight Calibration ComparatorPac TM Simple Calibration Efficient Workflow Reliable Results

Experiment 1: Colligative Properties

Apr 17, 2000 LAB MANUAL PARTICLE SIZE ANALYSIS OF SOILS AASHTO Designation T 88 (Mn/DOT Modified)

FederalDemocratic RepublicofEthiopia

Grammage of paper and paperboard (weight per unit area) (Revision of T 410 om-08) (underscores and strikeouts indicate changes from Draft 1)

American Association for Laboratory Accreditation

EVALUATION AND ACCREDITATION OF SUPPLIER TEST FACILITIES GP 10

US Environmental Protection Agency Office of Pesticide Programs

calibrate confidence with Risk Management: The Critical Role of Measurement and Calibration WHITE PAPER SERVICES

Guide to adhesively mounting accelerometers

EASIDEW PORTABLE HYGROMETER INSTALLATION, OPERATION AND MAINTENANCE MANUAL

DETERMINATION OF TOTAL CARBON, TOTAL ORGANIC CARBON AND INORGANIC CARBON IN SEDIMENTS

Transcription:

March 2003 GMP 11 Good Measurement Practice for Assignment and Adjustment of Calibration Intervals for Laboratory Standards 1. Introduction 1.1. Purpose Measurement processes are dynamic systems and often deteriorate with time or use. The design of a calibration system would be incomplete without some established means of determining how often to calibrate instruments and standards. A calibration performed only once establishes a one-time reference of uncertainty. Recalibration detects uncertainty growth and serves to reset values while keeping a bound on the limits of errors. A properly selected interval assures that an item will receive recalibration at the proper time. Proper calibration intervals allow specified confidence intervals to be selected and they support measurement traceability. The following practice establishes calibration intervals for standards and instrumentation used in measurement processes. 1.2. Prerequisites 1.3. Safety 2. Methodology 1.2.1. Calibration history for laboratory standards 1.2.2. Expected tolerance limits if applicable 1.3.1. No outstanding safety concerns 2.1. Summary Recommended calibration intervals are based on various examples of traceability as described in GMP 13. As data is collected and evaluated, the laboratory technical manager may adjust the calibration intervals to ensure that measurement results are not invalidated by the intervals selected. 2.2. Apparatus None. GMP 11 Page 1 of 6

March 2003 2.3. Procedure 2.3.1. Identification of Parameters Critical Parameters Components that contribute more than 25 % of a measurement s uncertainty are identified as critical parameters. To ensure an accurate evaluation of performance, calibration intervals are determined to meet a 99 % reliability target. Critical parameters are checked and defined below: Mass Critical Parameters (The example provided below corresponds to Option A in Mass Traceability in GMP 13.) Balance Performance Balance performance in Echelon 1 weighing processes is evaluated in every measurement series. An F-test ratio evaluates the observed standard deviation of the process against the accepted standard deviation of the process. The balance performance component is also tested with a check standard in each weighing series. The check standard value is evaluated with a t-test by compared the observed value to the accepted reference value. All other weighing processes have incorporated measurement control procedures and control charts that are evaluated as data is collected. Mass Standards Mass standards are dynamic with use. Wear, contamination and other factors can cause drift from accepted values. Thus, the following intervals have been set: Table 1. Calibration intervals for mass standards P1. kg + P1.. kg 48 NIST C1. kg + C1.. kg (alternating 2 years) 48 NIST P30 kg - P2 kg 12 Lab P500 g - P1 mg 6 Lab C500 g - C1 mg 6 Lab W25 kg - W1 mg 12 Lab P Pound Standards 24 Lab W Pound Standards 12 Lab P = primary; C = check/control; W = working standards GMP 11 Page 2 of 6

Length Critical Parameters March 2003 Length Comparison Performance The measurement performance of each length calibration is evaluated with a check standard (when performing tape-to-tape comparison). Initial Length Intervals Length standards are dynamic with use. Wear, contamination and other factors can cause drift from accepted values. The following intervals have been set due to these factors: Table 2. Calibration intervals for length standards 100 ft Tape #1 60 NIST 100 ft Tape #2 60 NIST 25 ft or 7 m Tape 60 NIST 18 in Steel Rule 120 NIST Length Bench 24 (if used or moved) Lab Volume Critical Parameters (Example shown corresponds to Option A for volume calibration in GMP 13.) Volume Comparison Performance The measurement performance of a volume transfer calibration is evaluated in each use with a repeatability check. Use of check standards is the preferred method for evaluating the measurement process over time. Traceability for volume standards may be established through gravimetric calibrations using traceable mass standards. Initial Volume Intervals Volume standards are dynamic with use. Wear, contamination and other factors can cause drift from accepted values. Calibration intervals are as follows: GMP 11 Page 3 of 6

Table 3. Calibration intervals for volume standards March 2003 P100 gal standard 60 NIST P25 gal standard * 60 Lab P5 gal standard * 12 Lab Glassware- Autopipetes 5 L to 100 ml 120 Lab *Gravimetric calibration for volumes 5 gallon or smaller, and all slicker plate standards. Laboratory must be qualified for performing gravimetric calibrations. Volume transfer is acceptable above 5 gallon. *May be a slicker plate type. None are hand-held, dump style, test measures. Temperature Critical Parameters Temperature Comparison Performance The measurement performance of each temperature comparison calibration is evaluated with a check standard and can be verified periodically using triple point cells, melting point cells, and ice baths (using documented procedures). Initial Intervals Temperature standards are dynamic with use. Shock, contamination and other factors can cause drift from accepted values. Recalibration intervals are as follows: Table 4. Calibration intervals for temperature standards 25.5 ohm SPRT 36 NIST 100 ohm PRT s 12 Lab Standard Thermistor 12 Lab Check Standards 6 Lab Liquid-in-glass standards 6 Lab Secondary Parameters Components that contribute less than 25 % but more than 1 % of a measurement s uncertainty are identified as secondary parameters. Secondary parameters are assigned calibration intervals designed to meet a 95 % reliability target. Secondary parameters are defined below: Mass and Gravimetric Volume Secondary Parameters GMP 11 Page 4 of 6

March 2003 Environmental Measurement Equipment Table 5. Calibration intervals for environmental equipment Item Initial Calibration Interval (months) Source Barometer 12 NIST Hygrometer 24 NIST Thermometer 12 Lab Length Secondary Parameters - No secondary parameters. Volume Secondary Parameters Water Temperature Measuring Device Table 6. Calibration intervals for volume secondary parameters Thermometer 12 Lab Temperature Secondary Parameters - No secondary parameters. 3. Calculations 3.1. Initial Intervals 3.1.1. Assignment of Initial Intervals Assignment of initial intervals is based on these recommendations. Subsequent intervals may be adjusted based on analysis of Check standard data that is recorded on control charts. The initial interval is equivalent to the amount of time that passes before a metrologist makes the first observation of a result lying outside the warning limits of the chart when the investigation yields no apparent correctable cause. Warning limits are established at ± 2 standard deviations of the measurement process around the accepted value of the check standard. 3.1.2. Absence of Control Charts If no control charts are available, the laboratory s Technical Manager will assign the initial interval based on engineering evidence, manufacturer s specifications, NIST recommendations, and experience. GMP 11 Page 5 of 6

3.2. Adjustment of Intervals March 2003 3.2.1. Intervals will be adjusted when determined to be necessary by the laboratory s Technical Manager. 3.2.2. The intervals will be adjusted by taking the following factors into consideration as appropriate: 4. Assignment of Uncertainty calibration history; measurement assurance data; data for the population of similar equipment; NIST recommendations; statistical analysis methods; and manufacturer s recommendations. The uncertainty associated with the establishment of calibration intervals is not included as a part of the uncertainties associated with the respective measurement SOPs. See SOP 29 for calculating uncertainties for standards and the instruments. GMP 11 Page 6 of 6