PROCEDURE COVERSHEET. Document Section: CHEMISTRY Author: JOANNE WONG Date Authored: JAN 2011. In Use Date: 2/1/2011 Discontinued Date:



Similar documents
Measuring ph in Water or CaCl2 Using a ph Meter. Compiled by Darren Murray, June 30, 2011 TABLE OF CONTENTS OVERVIEW 3

ph Measurement and Control

AIRMoN, ph and Conductivity Field Measurement

Measuring the Conductivity of Pure Water Using USP<645>

UltraBasic Benchtop ph Meter Operation Manual

Direct ISE Method Method to 1000 mg/l Na + Sodium ISE

To determine the equivalence points of two titrations from plots of ph versus ml of titrant added.

USEPA Electrode Method Method 8156

Experiment 17: Potentiometric Titration

The Determination of an Equilibrium Constant

Experiment 4 (Future - Lab needs an unknown)

Care and use guide. Contents. a handy solution made easy. page

15. Acid-Base Titration. Discover the concentration of an unknown acid solution using acid-base titration.

Milwaukee USER MANUAL. Milwaukee. Smart DO Meter PORTABLE DISSOLVED OXYGEN METER MODEL: SM600. Authorized Dealer: ISMIL600 11/01

ph units constitute a scale which allows scientists to determine the acid or base content of a substance or solution. The ph 0

Measurement by Ion Selective Electrodes

Acid Base Titrations

ExStik TM ph Waterproof Meters

SenTix 950 SenTix 980

Handbook. Thermo Scientific. ph Electrode Selection Handbook

Appendix C. Vernier Tutorial

Chem 1B Saddleback College Dr. White 1. Experiment 8 Titration Curve for a Monoprotic Acid

ph Measurements of Common Substances

Milwaukee USER MANUAL

User Guide. ExStik TM ph (PH100 & PH110) and ORP (RE300) Waterproof Series Pens. Patent Pending

Bluelab Soil ph Meter Instruction Manual

18 Conductometric Titration

Global Water Instrumentation, Inc.

Cadmium Reduction Method Method to 30.0 mg/l NO 3 N (HR) Powder Pillows or AccuVac Ampuls

ph and Acidity in Wine and Fruit Juice

AU680 Every Other Week or Every 3,000 Samples (ISE) Maintenance. AU680 Weekly (Analyzer and ISE) Maintenance

9. Analysis of an Acid-Base Titration Curve: The Gran Plot

Experiment 6 Titration II Acid Dissociation Constant

Chem101: General Chemistry Lecture 9 Acids and Bases

ph and Acidity in Tomato Sauce

(1) Hydrochloric acid reacts with sodium hypochlorite to form hypochlorous acid: NaOCl(aq) + HCl(aq) HOCl(aq) + NaCl(aq) hypochlorous acid

EXPERIMENT INTRODUCTION TO INDICATORS AND ACID-BASE TITRATIONS

Care and use guide. Contents. page

Color, True and Apparent

1. To perform a potentiometric titration of a sample of Liquid Drano with hydrochloric acid.

PART 1 - INTRODUCTION...

Nitrogen, Ammonia. Known Addition ISE Method 1 Method Minimum of 0.8 mg/l NH 3 N. Ammonia ISE. Test preparation. Instrument-specific table

Determination of calcium by Standardized EDTA Solution

Coating and Extraction of Honeycomb Denuders

Health Science Chemistry I CHEM-1180 Experiment No. 12 Acids, Bases, ph, Hydrolysis and Buffers (Revised 05/27/2015)

Analysis of Vitamin C Using Iodine. Introduction

DISSOLVED OXYGEN PROBE

CHM 130LL: ph, Buffers, and Indicators

ELECTROCHEMISTRY. Theory and Practice

MOISTURE (Karl Fischer, Buffered)

Acid Dissociation Constants and the Titration of a Weak Acid

MODEL 430 PORTABLE ph/conductivity METER OPERATING MANUAL

Color, True and Apparent

Dimethylglyoxime Method Method to 6.0 mg/l Ni TNTplus 856

BSTP SRF 81 Medical Emergency and Work Place Injury

Experiment 18: ph Measurements of Common Substances. Experiment 17: Reactions of Acids with Common Substances

Professional Plus. Calibration Tips

Dissolving of sodium hydroxide generates heat. Take care in handling the dilution container.

Electrochemistry Revised 04/29/15

ph Sensor Model 397 CAUTION SENSOR/PROCESS APPLICATION COMPATIBILITY WARNING ATEX DIRECTIVE Instruction Sheet PN 51A-397/rev.

Absorbance Spectrophotometry: Analysis of FD&C Red Food Dye #40

A Potentiometric Analysis of Fluoride Ion in Toothpaste

CALIBRATION OF A THERMISTOR THERMOMETER (version = fall 2001)

Care, Maintenance, and Storage of YSI 6-Series Probes and Sondes

Chemistry 119: Experiment 7. Potentiometric Titration of Ascorbic Acid in Vitamin C Tablets

Direct Measurement Method Method to 20.0 mg/l (or 0 to 200% saturation) O 2 Clark-type Amperometric Sensor

Understanding hydrogen ion concentration (ph)

Laboratory Evaluation Checklist - PSP

Preparation of frequently used solutions

Determining the Quantity of Iron in a Vitamin Tablet. Evaluation copy

URINE DIPSTICK Multistix 10SG

Determination of the Mass Percentage of Copper in a Penny. Introduction

This value, called the ionic product of water, Kw, is related to the equilibrium constant of water

Phosphorus Analyzer Procedure Last updated: A. Introduction

Lab 25. Acid-Base Titration and Neutralization Reactions: What Is the Concentration of Acetic Acid in Each Sample of Vinegar?

ph and CONDUCTIVITY ph and conductivity in fountain solution are necessary for plate humidification.

EXPERIMENT 20: Determination of ph of Common Substances

Ti-Pure TITANIUM DIOXIDE. DETERMINATION OF BROOKFIELD VISCOSITY, TiO 2 SLURRY METHOD: T WP

Electrical Conductivity of Aqueous Solutions

Juice Titration. Background. Acid/Base Titration

ACID-BASE TITRATIONS: DETERMINATION OF CARBONATE BY TITRATION WITH HYDROCHLORIC ACID BACKGROUND

STEP-BY-STEP INSTRUCTIONS FOR INVESTIGATIONAL USE. Rapid HCV Antibody Test FOR ORAQUICK RAPID HCV ANTIBODY TEST

Solubility Product Constants

EXPERIMENT # 3 ELECTROLYTES AND NON-ELECTROLYTES

POOL SUNSHINE IONIZER MODELS SPD, SPD2 AND SPS INSTALLATION INSTRUCTIONS

Relative Humidity Calibration Kit

ph Alkalinity of Water

Determining the Free Chlorine Content of Swimming Pool Water. HOCl H + + OCl. Evaluation copy

Household Acids and Bases

TITRATION CURVES, INDICATORS, AND ACID DISSOCIATION CONSTANTS

Care and use guide. soil ph meter. Contents. page

Calibration Verification

FT FIELD MEASUREMENT OF RESIDUAL CHLORINE

Evaluation copy. Titration of a Diprotic Acid: Identifying an Unknown. Computer

OneStep Fecal Occult Blood RapiDip InstaTest. Cat #

Experiment 2 Kinetics II Concentration-Time Relationships and Activation Energy

5.0 EXPERIMENT ON DETERMINATION OF TOTAL HARDNESS

6.3 SPECIFIC ELECTRICAL CONDUCTANCE

Properties of Acids and Bases

ph: Measurement and Uses

Transcription:

PROCEDURE COVERSHEET Procedure Title: ph (Urine, Body Fluid, Platelet Samples) Document Section: CHEMISTRY Author: JOANNE WONG Date Authored: JAN 2011 Laboratory Director: TED KURTZ, MD (PARNASSUS & CHINA BASIN) Distribution: Parnassus: CHEMISTRY In Use Date: 2/1/2011 Discontinued Date: Chemistry\Documents\Procedures\Autochem Misc\pH Meter Pg 1 of 5

I. PURPOSE ph is a measure of the acidity or alkalinity of a solution and can be represented by the equation: ph = -log[h + ] with H+ representing the concentration. When a precise ph on urine or other body fluid is desired, an electronic ph meter is used. The electronic ph meter is a sensitive electrometer designed to measure the hydrogen ion concentration in a liquid medium. It performs the measurement by indicating the voltage developed between two electrodes. These electrodes are connected to a galvanometer which in turn translates the result into a digital readout. The ph meter is equipped with a single ROSS combination electrode. Both the calomel and glass electrodes are in the same glass tubing, but separated physically. II. POLICY/SCOPE For use by licensed CLS staff only. III. TEST AVAILABILITY Test available 24 hours per day 7 days per week. IV. SPECIMEN REQUIREMENTS Red or black top or clean container. Sample stability- Room Temperature: 2 Hrs, Refrigerated: 24 Hrs A ph may be done on any body fluids in addition to urine. The specimen must be liquid; semi-formed or formed stool cannot be used. Since ph can change on storage, tests should be run immediately. Platelet samples may also be measured by this method. V. EQUIPMENT, REAGENTS AND SUPPLIES 1. ph Reference Buffer Solution (7.00). Purchased from Fisher, catalog number SB108-500. Stable at room temperature until expiration date on bottle. 2. ph Reference Buffer Solution (4.00). Purchased from Fisher, catalog number SB98-500. Stable at room temperature until expiration date on bottle. 3. Internal Filling solution (3M KCL). Purchased from Fisher, catalog number 13-641-755 (Orion # 810007). Stable at room temperature for 2 years once opened. 4. ph Electrode Storage Solution. Purchased from Fisher, catalog number 13-641-908 (Orion # 910001). Stable at room temperature for 2 years once opened. 5. HCl (0.1 M): Add 4.2 ml concentrated hydrochloric acid to 250 ml of deionized water in a 500 ml volumetric flask. QS to volume. Stable for 2 years at room temperature. VI. VII. WARNINGS AND PRECAUTIONS BIOHAZARD All products or objects that come in contact with human or animal body fluids should be handled, before and after cleaning, as if capable of transmitting infectious diseases. Wear facial protection, gloves, and protective clothing. Use bloodborne pathogen precautions when handling patient samples, calibrators or QC materials. CALIBRATION/ CALIBRATION VERIFICATION See Procedure below. VIII. QUALITY CONTROL 1. ph Buffer Solution (7.00), color yellow. Purchased from Fisher, catalog number SB107-500. Stable at room temperature until expiration date on bottle. Chemistry\Documents\Procedures\Autochem Misc\pH Meter Pg 2 of 5

2. ph Buffer Solution (4.00), color pink. Purchased from Fisher catalog number SB101-500. Stable at room temperature until expiration date on bottle. Always use fresh aliquots of buffers for controls. IX. PROCEDURE I. Standardization: The Denver UB-10 ph meter must be standardized each day of use. Use fresh buffers for this step. Once calibration has been performed, recalibration of the meter needs to be performed if the ph 4.00 and ph 7.00 control buffer solutions are out of the posted acceptable ranges. 1. Slide the filling hole cover open. 2. Immerse electrode in ph 4.00 buffer. Stir gently and allow the electrode to reach a stable value. (S will be displayed on left side of screen when stable.) 3. If necessary, press and release the mode button until the digital display indicates ph mode. 4. Clear existing buffers when doing a new standardization. Press the setup button. When the screen shows Clear, press the enter button to clear existing buffers. 5. Press standardize. The meter flashes the current buffer set and recognizes the flashing buffer. 6. When the signal is stable, the meter displays the percent slope of the electrode as 100% on the first buffer. Press enter to accept. 7. Remove the electrode from the ph 4.00 buffer, rinse with D.I. water, gently blot and place the electrode into the ph 7.00 buffer solution. Gently stir. Allow the electrode to stabilize. When this S box appears on the screen, press standardize. 8. The meter will display a slope reading that must be between 90 and 105 % slope to avoid a Slope Error. However, the slope reading must be between 92 and 102% to be acceptable due to the Ross electrode used. Record the slope value on the maintenance chart. Press enter to continue. Recalibrate the ph meter if the slope is not between 92 and 102%. 9. Slope Error indicates that the electrode is not working properly. Measurements causing Slope Error are not accepted, used or stored by the meter. Press enter to continue. II. ph of Unknowns: 1. For the first sample of the day, calibrate the electrode as described in previous section. 2. Prior to reading the ph of the unknown sample, read the ph 7.00 and 4.00 control buffers as unknowns. If either buffer is out of the acceptable range, recalibrate the electrode before reading the unknown sample. Record all results on the ph worksheet. 3. Rinse the electrode with deionized water and gently absorb off the excess water with a gauze sponge. 4. Place the electrode in the solution. 5. When the display is stable, record sample ph. Rinse the electrode with deionized water. 6. Place the electrode in the ph Electrode Storage Solution. 7. Slide the filling hole cover closed. Chemistry\Documents\Procedures\Autochem Misc\pH Meter Pg 3 of 5

Notes: 1. Always use fresh buffers for calibration. 2. Check electrode slope daily by performing a two buffer calibration. Slope should be 92 to 102%. 3. To view the slope of current calibration/standardization, press the setup button twice. Press the setup button four more times to return to the measuring mode. 4. Only use ROSS Internal Filling Solution, Orion cat. No. 810007, or 3M KCl. 5. Remove the filling hole cover during measurement and slide it closed during storage. 6. Between measurements, rinse electrode with deionized water and then with the next solution to be measured. 7. Stir all buffers and samples. 8. Avoid rubbing or wiping electrode bulb, to reduce chance of error due to polarization. 9. For troubleshooting see Denver Instruments instruction manual. 10. Weekly Electrode Maintenance 1) Inspect the electrode for scratches, cracks, salt crystal build-up, or membrane/junction deposits. 2) Rinse off any salt build-up with deionized water, and remove any membrane/junction deposits. 3) Drain the reference chamber and refill it with fresh ROSS Filling Solution (3M KCl). To maintain an adequate flow rate, the level of filling solution must cover the end of the coil. 4) Check lot numbers of all four ph buffers. Record on reagent log sheet if different. 11. Monthly Electrode Cleaning Soak the electrode in 0.1 M HCl for half an hour. Drain the reference chamber, flush with fresh ROSS Filling Solution or 3M KCl and refill the chamber. Soak the electrode in storage solution for at least one hour before use. X. RESULTING/REPORTABLE RANGE Reportable Range: ph 2.00 to ph 10.00 All results are recorded on the worksheet. The worksheet code is PH. Report urine, body fluid and platelet sample ph s to two decimal places. Report ph values less than 2.00 as <2.00 and values above 10.00 as >10.00. If the sample is >2 hours old at room temperature, append ETC STAQ (Stability period exceeded. Result questionable) to the result. XI. XII. EXPECTED VALUES (N/A) LIMITATIONS OF PROCEDURE (N/A) Chemistry\Documents\Procedures\Autochem Misc\pH Meter Pg 4 of 5

XIII. SPECIFICITY/INTERFERENCES Electrode efficiency and each buffer change depending on the temperature of the solution. The table below is representative of most buffers. Due to the insignificant change of the ph at room temperature, an Automatic Temperature Compensation (ATC) probe will not be used and temperature compensation will not take effect. U.S Standard Buffer ph 4.00 ph 7.00 ph 10.00 0 o C 4.005 7.13 10.34 5 o 4.003 7.10 10.26 10 o 4.001 7.07 10.19 15 o 4.002 7.05 10.12 20 o 4.003 7.02 10.06 25 o 4.008 7.00 10.00 30 o 4.010 6.99 9.94 35 o 4.020 6.98 9.90 40 o 4.030 6.97 9.85 50 o 4.061 6.97 9.78 XIV. ALTERNATE METHODS Beckman model 3500 digital ph meter (in Hematology, room M524) XV. REFERENCES 1. Denver instrument UltraBasic Benchtop ph Meter operational manual. 2. Orion Ross ph Electrode (8103BN) instruction manual. Chemistry\Documents\Procedures\Autochem Misc\pH Meter Pg 5 of 5