RESTORING STRAUMANN IMPLANTS WITH LOCATOR ABUTMENTS



Similar documents
Straumann Dental Implant System. Fax Order Form Collection.*

The LOCATOR concept. Simplicity and versatility for prosthesis fixation

STEP-BY-STEP INSTRUCTIONS ON TEMPORARY ABUTMENTS. Straumann Temporary Abutments, VITA CAD-Temp

LOCATOR Implant Attachment System for XiVE, FRIALIT and ANKYLOS Implants

Order of Treatment STEP-BY-STEP INSTRUCTIONS TO PLACE LOCATOR ATTACHMENTS

LOCATOR BAR ATTACHMENT SYSTEM

LOCATOR ROOT ATTACHMENT SYSTEM

Edentulous patients. Straumann Dental Implant System

STEP BY STEP INSTRUCTIONS ON TEMPORARY ABUTMENTS. Straumann Temporary Abutments, VITA CAD-Temp

ball abutment overdenture: chairside pick-up using existing denture

More than a fixed rehabilitation.

BASIC INFORMATION ON THE STRAUMANN VARIOBASE ABUTMENT. Straumann Variobase Abutment

Straumann Dental Implant System. Implant Selection Guide.

Guidance for implant removal. Straumann Dental Implant System

PROSTHETIC OPTIONS FOR NARROW NECK IMPLANTS. Straumann Narrow Neck

PROSTHETIC PROCEDURE. for HG IMPLANT SYSTEM

GIVE YOUR PATIENTS THE FREEDOM TO EAT, SPEAK AND LAUGH AGAIN.

NARROW DIAMETER implant

Implant Parts. A Radford Heath Guide 1

Self-Aligning Lowest Profile

STEP-BY-STEP INSTRUCTIONS ON THE PROSTHETIC PROCEDURES. Straumann Anatomic IPS e.max Abutment

screw-retained crown using a PEEK temporary abutment

Implant Bar Overdenture Utilizing Locator Attachments

screw-retained single crowns using custom-cast abutments

Straumann Bone Level Tapered Implant Peer-to-peer communication

One Abutment - One Time

TRI-CAM Implant System

Restoration of a screw retained single tooth restoration in the upper jaw with Thommen Titanium base abutment.

Company information for patients. More than Swiss precision. A better quality of life.

Taking the Mystique out of Implant Dentistry. Dr. Michael Weinberg B.Sc., DDS, FICOI

Product Catalog. ERA Mini TM Dental Implant System

Contents. Cement retained restoration. Screw retained restoration. Overdenture retained restoration. TS Implant System. 70 ComOcta Gold Abutment

CATALOGUE DES IMPLANTS SUISSE 2014

Prosthodontist s Perspective

THE LOCATOR OVERDENTURE IMPLANT SYSTEM.

PROSTHETICS. Crown and bridge restorations with the synocta prosthetic system.

Gl GinGival level implant

ITI Small Diameter (SD) external hex 3.25 / 3.3 external hex. ITI Standard 4.8 internal non-octagon Morse taper 4.

Straumann. Time-tested

procedures & products NOBELESTHETICS including Procera

Cement-retained crowns and bridges with the solid abutment system. Straumann Solid Abutment Prosthetic System

AT THE RIGHT PRICE. ONE SIZE FITS ALL. FRICTION For a secure connection. BONEADAPT For immediate stability.

Multi-unit abutment bar overdenture

Care for your dental implants. Patient information on dental implant aftercare

GINGIVAL LEVEL implant

PRoduct catalog International Headquarters Straumann USA Straumann Canada talog PRoduct

Restorative Guidelines

Ridge Reconstruction for Implant Placement

Crown and Bridge Restorations. Straumann synocta Prosthetic System

Replacement of the upper left central incisor with a Straumann Bone Level Implant and a Straumann Customized Ceramic Abutment

Abutment fracture in a bridge supported by natural teeth and implants

páêçå~=aéåí~ä=`^al`^j=póëíéã på~åmçëí

Removable Partial Dentures 101 Back to the Basics. Luther A. Ison, CDT University of Minnesota School of Dentistry

PROSTHETICS. Crown and Bridge Restorations with the synocta Prosthetic System DENTAL IMPLANT SYSTEM

Prosthetic Manual. Implant System. Contact Us: Smarter thinking. Simpler design. Phone: Fax:

DESS. Screws. Tijuana Ventas: (664) /95 For all major implant systems!!

Procedures & Products. NobelEsthetics

United States. Product Catalog 2015

Clinical and Laboratory Procedures for Fixed Margin Implant Abutments

SCD Case Study. Treatment Considerations for Implant Rehabilitation

ERA. Attachment Systems. A SNAP to Use and Maintain

MS SA Implant - Narrow Ridge MS SA Implant - Denture. MS Implant - Provisional MS KIT

Khaldoon G. Abu Afifeh, B.D.S, M. Sc in Prosthodontics

Standard Internal Hex

Restorative Manual Impression Transfer System

About DENTSPLY Implants

Tired of THE Gooey mess?

Regular C/X Prosthetics. Prosthetics

IMPRESSION TECHNIQUE AND MATERIALS

prosthetic technique manual

Straumann Cares. Process guide

Attachments And Their Use In Removable Partial Denture Fabrication

Improving Esthetics with Sequential Treatment Planning and Implant-Retained Dentures

platform shifting Nobel Biocare is shifting gingival beauty to the next level

While the prosthetic rehabilitation of

Patient information on dental implants. More than just beautiful teeth. A comprehensive guide to dental implants.

Tired of THe Gooey mess?

ANKYLOS. Syncone Concept. The Paradigm For Implant Supported Dentures. Redefining Success In Implant Dentistry

Renaissance of One-Piece Implants

Implants in your Laboratory: Abutment Design

Another Implant Option for Missing Teeth with Challenging Symmetry Patrick Gannon, DDS and Luke Kahng, CDT

The IMZ Implant System

ADA Insurance Codes for Laboratory Procedures:

BASIC INFORMATION ON THE SURGICAL PROCEDURES. Straumann Dental Implant System

The SKY Locator family

Smile Confidently With DIEM 2. Immediate Full Arch Rehabilitation

Washington State Dental Association s 2015 Pacific Northwest Dental Conference. Presents

Flexible dentures an alternate for rigid dentures? Volume 1 Issue 1

2016 Buy Up Dental Care Plan Procedure List

Straumann Dental implant system ONE SYSTEM ONE INSTRUMENT KIT ALL INDICATIONS*

INTRODUCTION. most popular stud attachment available to the dental profession to increase the retention of implant complete

Basic information on the surgical and prosthetic procedures. Straumann PURE Ceramic Implant

A New Beginning with Dental Implants. A Guide to Understanding Your Treatment Options

Restoration of Endodontically Treated Teeth

FASTON Terminals and Connectors

All-on-4 treatment concept with NobelSpeedy Groovy

Straumann Patient Pro. The plus for your business

How To Use A Cresco

Supervisors: Dr. Farhan Raza Khan

Transcription:

RESTORING STRAUMANN IMPLANTS WITH LOCATOR ABUTMENTS

Straumann is the industrial partner of the ITI (International Team for Implantology) in the areas of research, development, and education.

CONTENTS Product description 2 Full denture fabrication 3 Pick up with full denture reline or rebase 5 Retrofit LOCATOR to existing full denture 7 Determining the angulation of LOCATOR abutments 9 Selection of the retentive nylon inserts 9 LOCATOR core tool 9 Product overview 10 1

PRODUCT DESCRIPTION The self-locating design of the LOCATOR allows patients to easily seat their dentures. The LOCATOR male components are easily placed and removed with the LOCATOR core tool. Up to 40 divergence between two implants can be accommodated. Restorations are possible even in cases of limited space due to the small vertical dimension. Available for bone-level and tissue-level Straumann implant platforms. Indications The LOCATOR Implant Attachment System is designed for use with dentures or partial dentures retained in whole or in part by endosseous implants in the mandible or maxilla. Contraindications Not appropriate where a totally rigid connection is required. Use on a single implant with a divergence of more than 20 is not recommended. 2

FULL DENTURE FABRICATION Clinical photos courtesy of Dr. Robert C. Vogel The restorative platform of the implant should be above the gingival crest, and the shortest LOCATOR abutment providing a margin above the gingival crest should be utilized. The clinician may use the abutment retaining sleeve to carry the abutment to the mouth. The margin of the LOCATOR abutment should be positioned slightly above the height of the soft tissue. Hand tighten the abutment with the L OCATOR driver. Then torque the abutment to 35 Ncm using the Straumann ratchet with torque control device and LOCATOR driver. Place the impression copings on the LOCATOR abutments for the final impression. Take the final impression utilizing a border mold impression technique and place the female analogs into the impression copings. 3

FULL DENTURE FABRICATION (continued) Box and pour the master cast. The processing male is fixed into the record base for maximum stability. The processed full denture incorporating the denture cap with the black processing male is ready for finishing. Remove the black processing male from the finished full denture base with the male removal tool, and insert the replacement male of choice with the male seating tool. Upon insertion, check for pressure spots and adjust the occlusion. 4

PICK UP WITH FULL DENTURE RELINE OR REBASE The restorative platform of the implant should be above the gingival crest whenever possible. The clinician may use the abutment retaining sleeve to carry the abutment to the mouth. Hand tighten the abutment with the driver. Then torque the abutment to 35 Ncm using the Straumann ratchet with torque control device and LOCATOR driver. Place the denture cap with black processing male onto the LOCATOR abutments. Adjust the denture base to seat passively in occlusion without touching the processing cap. Apply impression adhesive to the internal aspect of the denture, and take a reline impression with the patient in occlusion utilizing a border mold impression technique. 5

PICK UP WITH FULL DENTURE RELINE OR REBASE (continued) Place the female analog into the black processing male, pour the master cast, and process the reline. Box and pour the master cast. Note: The black processing male must be secure on the female analog. After processing the reline or rebase, remove the black processing male with the male removal tool, and insert the replacement male of your choice with the male seating tool. Upon insertion, check for pressure spots and adjust the occlusion. 6

RETROFIT LOCATOR TO EXISTING FULL DENTURE Clinical photos courtesy of Dr. Robert C. Vogel Insert the LOCATOR abutments and torque to 35 Ncm. Place the white block-out spacer over the head of the LOCATOR abutment. Then insert a LOCATOR denture cap with black processing male into each LOCATOR abutment, leaving the white block-out spacer beneath it. Prepare the denture base to accommodate the protruding LOCATOR processing cap. Important: There must be NO contact between the denture and the processing cap prior to the curing process. 7

RETROFIT LOCATOR TO EXISTING FULL DENTURE (continued) Clinical photos courtesy of Dr. Robert C. Vogel Cold curing the processing male Use a lightcure acrylic resin or mix a permanent self-curing acrylic to bond the processing cap in the denture. Apply a small amount of acrylic resin to the recess of the denture base and around the processing cap. Insert the denture into the oral cavity. Once the denture is properly positioned, maintain the patient in full occlusion while the acrylic sets. It is important to maintain the denture in full contact with the underlying soft tissue during the set. Note: If the white block out spacer does not completely block out any undercuts it is important to use a suitable block out material to prevent acrylic from flowing into these areas. Once the acrylic has cured, remove the denture and discard the white block-out spacers. Remove any excess acrylic and finish the denture base before changing to the final replacement male. Remove the black processing male from the metal processing cap with the male removal tool. Note: LOCATOR, extra-light retention, extended range, 0 lbs males cannot be removed with the LOCATOR Core Tool. To remove the LOCATOR, extra-light retention, extended range, 0 lbs males use a sharp knife. Removal of the LOCATOR, extra-light retention, extended range, 0 lbs males with a sharp knife will require new retention inserts to be utilized. Use the male seating tool to insert a LOCATOR replacement male into the empty metal processing cap. The denture is ready for delivery to the patient. 8

DETERMINING THE ANGULATION OF LOCATOR ABUTMENTS Snap the LOCATOR parallel post on to the LOCATOR abutment. Use the angle measurement guide behind the parallel post to determine the angulation of the implant. Choose the appropriate nylon LOCATOR replacement male according to the angulation measured and the desired retention. Caution: For the safety of the patient, tie dental floss through the hole of the angle measurement guide to prevent aspiration. LOCATOR parallel post (048.199V4) LOCATOR angle measurement guide (048.200) SELECTION OF THE RETENTIVE NYLON INSERTS The LOCATOR retentive nylon inserts can be used to restore an implant with up to 10 of divergence (20 between implants). Use of multiple LOCATOR abutments may necessitate the use of less retentive inserts such as the 3.0 lbs (light retention) pink-colored retentive nylon insert (048.191V4) or 1.5 lbs (extra light retention) blue-colored retentive nylon insert (048.192V4) for easier removal of the prosthesis by the patient. The LOCATOR extended range replacement males can accommodate a divergent implant between 10 and 20 (40 between implants). The LOCATOR extended range replacement male (green, 048.193V4) offers more retention for divergent implants (3.0 4.0 lbs). The retentive holding force is dependent upon the degree of angulation. The extra light retention extended retentive nylon insert (grey, 048.181V4) provides the least retention possible (0 lbs). LOCATOR CORE TOOL Male removal tool Male seating tool LOCATOR core tool (046.415) assembled LOCATOR core tool (046.415) disassembled * LOCATOR abutment driver *The LOCATOR abutment driver is not needed for Straumann implants. The LOCATOR driver (046.416 or 046.417) should be used instead. See page 15 for more information. 9

Regular Neck Ø 4.8 mm Art. No. Article Dimensions Material LOCATOR abutments 048.175 RN LOCATOR abutment height 1.0 mm titanium alloy/tin 048.176 RN LOCATOR abutment height 2.0 mm titanium alloy/tin 048.177 RN LOCATOR abutment height 3.0 mm titanium alloy/tin 048.178 RN LOCATOR abutment height 4.0 mm titanium alloy/tin 048.179 RN LOCATOR abutment height 5.0 mm titanium alloy/tin 048.180 RN LOCATOR abutment height 6.0 mm titanium alloy/tin RN = Regular Neck = Titanium nitride coated Important RN components can only be used with Straumann Standard, Standard Plus and Tapered Effect implants with a Ø 4.8 mm Regular Neck. 10

Wide Neck Ø 6.5 mm Art. No. Article Dimensions Material LOCATOR abutments 048.183 WN LOCATOR abutment height 1.0 mm titanium alloy/tin 048.184 WN LOCATOR abutment height 2.0 mm titanium alloy/tin 048.185 WN LOCATOR abutment height 3.0 mm titanium alloy/tin 048.186 WN LOCATOR abutment height 4.0 mm titanium alloy/tin 048.187 WN LOCATOR abutment height 5.0 mm titanium alloy/tin WN = Wide Neck = Titanium nitride coated Important WN components can be used only with Straumann Standard, Standard Plus and Tapered Effect implants with a Ø 6.5 mm Wide Neck. 11

Narrow CrossFit Connection Art. No. Article Dimensions Material LOCATOR Abutments 022.2502 NC LOCATOR abutment height 2.0 mm titanium alloy/tin 022.2503 NC LOCATOR abutment height 3.0 mm titanium alloy/tin 022.2504 NC LOCATOR abutment height 4.0 mm titanium alloy/tin 022.2505 NC LOCATOR abutment height 5.0 mm titanium alloy/tin 022.2506 NC LOCATOR abutment height 6.0 mm titanium alloy/tin NC = Narrow CrossFit = Titanium nitride coated All final abutments are torqued into the implant at 35 Ncm. Important NC prosthetic components can only be used with Straumann Bone Level implants, Ø 3.3 mm, Narrow CrossFit. 12

Hybrid Dentures, Regular CrossFit Connection LOCATOR Abutment Art. No. Article Dimensions Material LOCATOR Abutments 022.4501 RC LOCATOR abutment height 1.0 mm titanium alloy/ TiN 022.4502 RC LOCATOR abutment height 2.0 mm titanium alloy/ TiN 022.4503 RC LOCATOR abutment height 3.0 mm titanium alloy/ TiN 022.4504 RC LOCATOR abutment height 4.0 mm titanium alloy/ TiN 022.4505 RC LOCATOR abutment height 5.0 mm titanium alloy/ TiN 022.4506 RC LOCATOR abutment height 6.0 mm titanium alloy/ TiN RC = Regular CrossFit = Titanium nitride coated All final abutments are torqued into the implant at 35 Ncm. Important RC prosthetic components can only be used with Straumann Bone Level implants, Ø 4.1 and Ø 4.8 mm, Regular CrossFit. 13

LOCATOR Components Art. No. Article Dimensions Material LOCATOR components 048.189V2 LOCATOR male processing package, contents: denture cap (Ø 5.5 mm, height 2.5 mm) with black processing male (height 1.9 mm), block-out spacer (048.196), white replacement male (048.190), pink replacement male (048.191), blue replacement male (048.192) titanium/ nylon 048.190V4 LOCATOR replacement male, white, 0 10 *, 5 lbs, 2.27 kg** height 1.7 mm nylon 048.191V4 LOCATOR replacement male, pink, light retention, 0 10 *, 3 lbs, 1.36 kg** height 1.7 mm nylon 048.192V4 LOCATOR replacement male, blue, extra-light retention, 0 10 *, 1.5 lbs, 0.68 kg** height 1.7 mm nylon 048.182V2 LOCATOR male processing package, extended range, contents: denture cap (Ø 5.5 mm, height 2.5 mm) with black processing male (height 1.9 mm), block-out spacer (048.196), green replacement male (048.193), orange replacement male (048.188), red replacement male (048.194) titanium/ nylon 048.193V4 LOCATOR replacement male, green, extended range, 10 20 *, 3 4 lbs, 1.36 1.82 kg** height 1.7 mm nylon 048.188V4 LOCATOR replacement male, orange, light retention, extended range, 10 20 *, 2 lbs, 0.91 kg** 048.194V4 LOCATOR replacement male, red, extra-light retention, extended range, 10 20 *, 0.5 1.5 lbs, 0.23 0.68 kg** 048.181V4 LOCATOR replacement male, grey, extra-light retention, extended range, 0 lbs. height 1.7 mm height 1.7 mm height 1.7 mm nylon nylon nylon 048.218V4 LOCATOR processing spacer POM 046.413V4 LOCATOR abutment holder sleeve PSU V2 = Pack of 2 V4 = Pack of 4 * = For the correction of angle divergences ** = Retention force 14 Important RC prosthetic components can only be used with Straumann Bone Level implants, Ø 4.1 and Ø 4.8 mm, Regular CrossFit.

LOCATOR Components Art. No. Article Dimensions Material LOCATOR components 025.0120-04 BL LOCATOR female analog length 10.0 mm, diameter 4.0 mm 048.198V4 TL LOCATOR female analog length 10.0 mm, diameter 5.0 mm aluminum aluminum 048.197V4 LOCATOR impression coping height 4.0 mm aluminum housing with LDPE insert 048.196V20 LOCATOR block-out spacer thickness 0.4 mm Teflon 048.195V4 LOCATOR black processing male height 1.9 mm LDPE Auxiliary parts 046.415 LOCATOR core tool length 100.0 mm stainless steel 046.416 LOCATOR driver, short, for ratchet length 15.0 mm stainless steel 046.417 LOCATOR driver, long, for ratchet length 21.0 mm stainless steel 048.199V4 LOCATOR parallel post length 8.0 mm LDPE 048.200 LOCATOR angle measurement guide length 5.0 mm, width 15.0 mm stainless steel 046.414 LOCATOR core tool tip LDPE = Low Density Polyethylene -04 = Pack of 4 V4 = Pack of 4 V20 = Pack of 20 LOCATOR is a registered trademark of Zest Anchors, Inc., USA. Straumann is the sole worldwide distributor of LOCATOR abutments for the Straumann Dental Implant System. Teflon is a registered trademark of E. I. du Pont de Nemours and Company. Manufacturer Zest Anchors, LLC. Escondido, CA 92029 USA Distributor Institut Straumann AG 4002 Basel Switzerland 15

NOTES 16

www.straumann.com International Headquarters Institut Straumann AG Peter Merian-Weg 12 CH-4002 Basel, Switzerland Phone +41 (0)61 965 11 11 Fax +41 (0)61 965 11 01 Straumann USA Straumann USA, LLC 60 Minuteman Road Andover, MA 01810 Phone 800/448 8168 978/747 2500 Fax 978/747 2490 www.straumannusa.com Dolder is a registered trademark from Prof. Eugen Dolder, former director, School of Dentistry of Zurich. Straumann Canada Straumann Canada Limited 3115 Harvester Road, Suite 100 Burlington, ON L7N 3N8 Phone 800/363 4024 905/319 2900 Fax 905/319 2911 www.straumann.ca Institut Straumann AG, 2011. All rights reserved. Straumann and/or other trademarks and logos from Straumann that are mentioned herein are the trademarks or registered trademarks of Straumann Holding AG and/or its affiliates. All rights reserved. Straumann products are CE marked CALIT118 8/11