NSW Air Quality Monitoring Network: Quality Assurance Procedures



Similar documents
Environmental Management systems guidelines Risk-based licensing.

TransCity Legacy Way Air Monitoring Network

TR SUPPLEMENTARY REQUIREMENTS FOR THE ACCREDITATION OF CONTINUOUS AMBIENT AIR QUALITY MONITORING STATIONS

National Environment Protection (Ambient Air Quality) Measure

Air Monitoring Directive Summary of Feedback and Responses for Chapter 6 (Ambient Data Quality)

Quality System Training Module 4 Data Validation and Usability

The design of the Romanian national air quality monitoring network

UK Air Quality Monitoring Networks

Compliance Audit Handbook

Emission monitoring, reporting, and verification (MRV) for U.S. cap and trade programs. Jeremy Schreifels US EPA

The Ambient Air Monitoring System

Management of Environmental Risks and Emergencies. Compliance Audit Program

Enhanced calibration High quality services from your global instrumentation partner

MAC-ENC-PRO-057 AIR QUALITY MONITORING PROGRAM

Superseded by T MU AM PL v2.0

Which filter class for supply air is required in a typical HVAC system?

NOVEL STRATEGIES FOR ASSESSMENT OF AMBIENT AIR QUALITY USING GIS AND ONLINE POLLUTION MONITORING TOOLS I. Jaykumar 2, V. Hima Bindu 2, P.

AMBIENT AIR QUALITY MONITORING SYSTEM. Air Quality Monitoring... Easier. Faster. Better.

Memorandum of Understanding between the NSW Ministry of Health and Sydney Water Corporation

11.0 Instrument Equipment Testing, Inspection and Maintenance

INDUSTRIAL BOILER MACT FACTSHEET

ATTAINMENT PROJECTIONS

Assessment of Vaisala Veriteq vlog Validation System Compliance to 21 CFR Part 11 Requirements

MATHESON STACK EMISSION CALIBRATION PROGRAM. Experience the MATHESON Commitment to Supply Chain Excellence

A.1 Sensor Calibration Considerations

Oil and Gas Air Quality Regulations and Permitting

IMPLEMENTING ISO 14001:

Standardized Process for the Development of Military-Unique Emissions Factors

Stormwater harvesting

Tips for a Successful Diesel Retrofit Project

Basic Steps to Establish ISO/IEC Compliant Testing Laboratories

Basic Climatological Station Metadata Current status. Metadata compiled: 30 JAN Synoptic Network, Reference Climate Stations

ATTENTION FOR ONLINE MONITORING. In continuation to CPCB s earlier Web Advertisement, it is to further inform that All the

Tool Enhancement Assessment Framework

Calibration & Preventative Maintenance. Sally Wolfgang Manager, Quality Operations Merck & Co., Inc.

Continuous Emission Monitoring System (CEMS) Code

1.0 What Are the Purpose and Applicability of Performance Specification 11?

Guidance and Reporting Instructions on Temporary E-reporting for the MATS Rule

9 Things you need to know about Continuous Monitoring Systems in FDA-Regulated Environments

ISO/IEC QUALITY MANUAL

Laboratory Manual for Air Quality Sampling and Analysis

Guidelines on Resource Recovery Exemptions (Land Application of Waste Materials as Fill)

Using text messaging services:

WEATHER STATION FOR SOLAR FARM MONITORING

WARNING: You cannot rely on a printed version of this document to be current. Always check the DECCW intranet to ensure you have the latest version.

Welcome to chapter 10. The following chapter is called Equipment maintenance and monitoring. The author is Ronny Janssens.

Renewable Choice Energy

SMS Alarm Messenger. Setup Software Guide. SMSPro_Setup. Revision [Version 2.2]

Measurement and Verification Operational Guide

Sentry-MS. Cost Effective Perimeter & LDAR Monitoring. ...Precise, Species Specific. Open Path Multi-gas UVDOAS Analyzer. Ammonia.

Health, Safety and Environment Policy

Comparison between FDA QSR and ISO 13485

CONFORMED COPY. Method Statement Helpdesk Services. Revision History. Revision Date Reviewer Status. 23 March 2007 Project Co Final Version

Ecotech Air Quality Monitoring Data Acquisition System

2) Relevance for environmental policy ) Data sources and reporting ) References at the international level... 6

Texas Commission on Environmental Quality Page 1 Chapter Control of Air Pollution from Nitrogen Compounds

Chromatograph Maintenance and Troubleshooting Class # 5050

Guide to the National Safety and Quality Health Service Standards for health service organisation boards

Quality Assurance for Hydrometric Network Data as a Basis for Integrated River Basin Management

Online Infrared Flue Gas Analyzer Gasboard 3000

FACT SHEET PROPOSED REVISIONS TO THE NATIONAL AMBIENT AIR QUALITY STANDARDS FOR SULFUR DIOXIDE

Information and Communications Technology (ICT) Steering Committee - Information Sheet

US Environmental Protection Agency Office of Pesticide Programs

REDUCING UNCERTAINTY IN SOLAR ENERGY ESTIMATES

Control Device Requirements Charts For Oil and Gas Handling and Production Facilities

Data Quality Assurance and Control Methods for Weather Observing Networks. Cindy Luttrell University of Oklahoma Oklahoma Mesonet

Software Verification and Validation

CATALYST Frequently Asked Questions

Canadian Technology Accreditation Criteria (CTAC) INDUSTRIAL ENGINEERING TECHNOLOGY - TECHNOLOGIST Technology Accreditation Canada (TAC)

Validation of process gas systems

Sector Development Ageing, Disability and Home Care Department of Family and Community Services (02)

A Guide for Local Authorities Purchasing Air Quality Monitoring Equipment

American Association for Laboratory Accreditation

Environment and Natural Resources Trust Fund 2017 Request for Proposals (RFP)

4. The role of fleets and fleet managers Role of REC?

Smart Integration of Wireless Temperature Monitoring System with Building Automation System

PROPOSAL AIR POLLUTION: A PROJECT MONITORING CURRENT AIR POLLUTION LEVELS AND PROVIDING EDUCATION IN AND AROUND SCHOOLS IN CAMDEN

London Underground Environment Strategy

Solar Radiation Measurement. Bruce W Forgan, WMO RAV Metrology Workshop, Melbourne, Novemberr 2011

/ BROCHURE / CHECKLIST: PCI/ISO COMPLIANCE. By Melbourne IT Enterprise Services

EVALUATING SOLAR ENERGY PLANTS TO SUPPORT INVESTMENT DECISIONS

Automotive Air Quality Sensors: industrial innovations to protect people s health

2014 Aamjiwnaang Air Monitoring Station Results. Ministry of the Environment and Climate Change April 19, 2016

The ITIL v.3. Foundation Examination

Indoor air quality monitoring made easy

Transcription:

NSW Air Quality Monitoring Network: Quality Assurance Procedures

Copyright State of NSW and Office of Environment and Heritage With the exception of photographs, the State of NSW and Office of Environment and Heritage are pleased to allow this material to be reproduced in whole or in part for educational and non-commercial use, provided the meaning is unchanged and its source, publisher and authorship are acknowledged. Specific permission is required for the reproduction of photographs. The Office of Environment and Heritage (OEH) has compiled this guideline in good faith, exercising all due care and attention. No representation is made about the accuracy, completeness or suitability of the information in this publication for any particular purpose. OEH shall not be liable for any damage which may occur to any person or organisation taking action or not on the basis of this publication. Readers should seek appropriate advice when applying the information to their specific needs. Every effort has been made to ensure that the information in this document is accurate at the time of publication. However, as appropriate, readers should obtain independent advice before making any decision based on this information. Published by: Office of Environment and Heritage NSW 59 Goulburn Street, Sydney NSW 2000 PO Box A290, Sydney South NSW 1232 Phone: (02) 9995 5000 (switchboard) Phone: 131 555 (environment information and publications requests) Phone: 1300 361 967 (national parks, climate change and energy efficiency information, and publications requests) Fax: (02) 9995 5999 TTY: (02) 9211 4723 Email: info@environment.nsw.gov.au Website: www.environment.nsw.gov.au Report pollution and environmental incidents Environment Line: 131 555 (NSW only) or info@environment.nsw.gov.au See also www.environment.nsw.gov.au ISBN 978 1 74359 787 3 OEH 2014/0754 October 2014

Contents Introduction 1 What is being measured in the NSW Air Quality Monitoring Network? 1 How can we be sure the data posted by OEH are reliable? 1 Why are there sometimes gaps in the data or no results on the website? 2 How is the quality of air quality monitoring data provided by the network assured? 3 How are air quality monitoring data validated? 3 Automated validation 3 Manual validation 5

Introduction The purpose of this guideline is to outline the quality assurance procedures used by the Climate and Atmospheric Science branch within the Office of Environment and Heritage (OEH) to ensure the validity and integrity of air quality data measured in the Air Quality Monitoring Network (AQMN). What is being measured in the NSW Air Quality Monitoring Network? Monitoring stations in the AQMN report on gaseous parameters (ozone, sulfur dioxide, carbon monoxide, oxides of nitrogen), fine particles (as PM 10 and PM 2.5 ), visibility, wind speed, wind direction, temperature and relative humidity. Figure 1 outlines the basic design of this system for collecting and reporting on NSW air quality monitoring data. Figure 1. NSW OEH air Quality Monitoring network and database system How can we be sure the data posted by OEH are reliable? It is important that the monitoring equipment used is appropriately accredited (e.g. compliance with local or international standards), properly installed and regularly calibrated and serviced to make sure it provides accurate and reliable data. NSW Air Quality Monitoring Network: Quality Assurance Procedures 1

The instruments used by OEH meet the relevant Australian Standards and international guidelines (e.g. United States Environmental Protection Agency Federal Equivalent Methods). OEH is accredited by NATA (the National Association of Testing Authorities) for the operation and maintenance of its air quality monitoring network and associated equipment. In accordance with the accreditation process, the monitoring network is periodically subjected to an independent review of operating, calibration and quality assurance and data validation procedures. Why are there sometimes gaps in the data or no results on the website? Because of the need for regular maintenance and calibration the AQMN is not online 100% of the time, although the network is operated to maximize online time. For gaseous instruments the target online time is 95%: about 5% of time is lost through routine maintenance and calibration. For instruments other than gaseous ones the online target is above 95%. Maintenance and calibration schedules comply with the requirements of the relevant Standards and operating procedures for servicing the equipment. Gaps in data can occur because of: 1. Scheduled maintenance and calibration in cases where the following are done on a regular basis: Gaseous instruments o Overnight: zero and span checks (1 hour) o Monthly: sample inlet cleaning; filter replacement; leak check (1 to 2 hours) o Quarterly: multi-point audit calibration to check linearity (1 to 2 hours) Particulate instruments o Quarterly: cleaning of sample head and inlet and replacement of filters; leak check (1 to 2 hours) o 6-monthly: flow audit (1 to 2 hours) o Annually: software (1 to 2 hours) and hardware (1 to 2 hours) calibration; zero stability check (at least 8 hours); site audit (1 to 2 hours) Meteorological instruments o Quarterly: cleaning, checking and alignment of wind direction/wind speed sensors (1 hour) o Annually: calibration of temperature and relative humidity sensors (1 to 2 hours). 2. On-site equipment failure or failure of communications between instrument and data logger (periodic): Data are lost permanently for either of these reasons; this results in gaps in the database and on the web. 3. On-site power outages (periodic): Data are lost when equipment is powered off during outages. 4. Telecommunications problems (periodic): Data are not loaded on to the web site during this period. However, these data are not lost but downloaded to the database and uploaded to the web once communication is re-established. 5. Website maintenance (periodic): Data are not lost, as they are stored locally until the website is up and running again. 2 NSW Air Quality Monitoring Network: Quality Assurance Procedures

6. Data validation: After collection, all data are subjected to routine validation performed by accredited air quality scientists from OEH s Quality Systems and Reporting (QSR) unit. On occasion, the data may be invalidated as a result of the inspection of calibration data, data records, instruments and site-operation logs. How is the quality of air quality monitoring data provided by the network assured? The Quality Systems and Reporting (QSR) unit is part of the Climate and Atmospheric Science branch. QSR is responsible for maintaining the validity of data collected from the AQMN. Maintenance and operation of instrumentation are the primary responsibilities of the air quality monitoring field officers, but QSR staff are responsible for ensuring that all data are flagged appropriately in the AQMN database for eventual reporting. Data flagged as valid must meet the goals set by the policies outlined in the Climate and Atmospheric Science quality manual. They must be: Precise: The data conform to set guidelines for precision. Precision is tested through daily calibration checks. Accurate. The data are accurate. Accuracy is determined by regular multi-point calibration and depends on the linearity of the instrument response. Representative. The data collected from each site are typical with respect to time of day, siting and meteorological conditions. Comparable. The data are comparable to data collected by other institutions using similar collection protocols. How are air quality monitoring data validated? The data validation procedure is outlined below and also shown in Figure 2. Automated validation Data validation step 1 Air quality data are retrieved from instrumentation by an on-site data logger and stored on site as raw data files. Instruments are also set up to provide alarm information if the instrument operates outside standard control parameters. The data logger detects this alarm information and flags the associated raw data as invalid. Data validation step 2 The raw data files are transmitted from the monitoring sites to a central location. Raw data are then loaded into the air quality database; a database loading program takes a raw hourly record and (on the basis of overnight calibration data) may correct and/or reflag this record, storing it in the database with a flag denoting the record as production data. The database loading program performs correction and reflagging according to a set of validation rules stored in the database. NSW Air Quality Monitoring Network: Quality Assurance Procedures 3

Figure 2. Data correction and flagging protocol. WDR, wind direction; WSP, wind speed; SD1, sigma theta 4 NSW Air Quality Monitoring Network: Quality Assurance Procedures

Once the data have been corrected and flagged, plausibility checks and value checking rules are applied to the data still flagged as valid to this point. Plausibility checks are applied on a site and parameter basis and may include: checks for outliers above or below pre-set thresholds (e.g. ozone values >20 pphm [parts per hundred million] or temperature values below 10 C are flagged as invalid) step change checks where data values show abrupt changes from one hour to the next and maintain the level of change. Value checking rules include: Internal consistency checking for NO X, NO 2 and NO values: if one of these is invalid, all are marked as invalid. If (NO X tolerance NO + NO 2 ) or (NO X + tolerance NO + NO 2 ), then all three values are invalidated. (Note: The tolerance value allows for instrumental uncertainty.) If collocated PM 2.5 > PM 10 + tolerance, then both are invalidated for further investigation. Manual validation Data validation step 3 If, after all validation and data quality checks have been done, data need to be invalidated, this must be done by using applications that provide an audit trail of all affected data within the database. Production data are the only records in the database that can be edited by QSR staff. If a production record is edited, a mandatory transaction record is created in the database; this record contains information on the user who edited the record, the date of the transaction, the time range of the data edited, the site and parameter, the old data value and the old status flag. It also contains comments as to why the change was made. Data validation step 4 Data irregularities are identified from data requests and data analysis reports or from stakeholder feedback. NSW Air Quality Monitoring Network: Quality Assurance Procedures 5