Transducer Mounting and Test Setup Configurations. Rick Bono The Modal Shop



Similar documents
Guide to adhesively mounting accelerometers

Decoding an Accelerometer Specification. What Sensor Manufacturer s Don t Tell You! David Lally VP Engineering PCB Piezotronics, Inc.

CHAPTER 3 MODAL ANALYSIS OF A PRINTED CIRCUIT BOARD

Time Domain and Frequency Domain Techniques For Multi Shaker Time Waveform Replication

Frequently Asked Questions Modal Shakers and Related Topics

Comparison of the Response of a Simple Structure to Single Axis and Multiple Axis Random Vibration Inputs

Similar benefits are also derived through modal testing of other space structures.

9000 Series Accelerometers Specifications

and Accelerometers

Product Overview 9100D

The use of Operating Deflection Shapes (ODS) to model the vibration of sanders and polishers HSL/2006/104. Project Leader: Author(s): Science Group:

The Universal DAQ Device. Connect and measure immediately!

Agilent N8973A, N8974A, N8975A NFA Series Noise Figure Analyzers. Data Sheet

QUICK START GUIDE FOR DEMONSTRATION CIRCUIT BIT DIFFERENTIAL ADC WITH I2C LTC2485 DESCRIPTION

Vibration Measurement of Wireless Sensor Nodes for Structural Health Monitoring

Cell Phone Vibration Experiment

MATRIX TECHNICAL NOTES

ACCELEROMETER SELECTION CONSIDERATIONS Charge and ICP Integrated Circuit Piezoelectric

FREQUENCY RESPONSE ANALYZERS

The Effect of Network Cabling on Bit Error Rate Performance. By Paul Kish NORDX/CDT

THERMAL ANEMOMETRY ELECTRONICS, SOFTWARE AND ACCESSORIES

Cable Analysis and Fault Detection using the Bode 100

AIR QUALITY SURVEILLANCE BRANCH ACCEPTANCE TEST PROCEDURE (ATP) FOR. Teledyne Advanced Pollution Instruments Model 400 E Ozone Analyzer AQSB ATP 002

Herschel PACS Test Procedure STM PACS-KT-PR-011 Issue: 1 Revision: -

Fiber Optic High Density ACC Sensor System Faser Optisches Beschleunigungs-Messsystem

Dayton Audio is proud to introduce DATS V2, the best tool ever for accurately measuring loudspeaker driver parameters in seconds.


MoveInspect HF HR. 3D measurement of dynamic processes MEASURE THE ADVANTAGE. MoveInspect TECHNOLOGY

Performing Amplifier Measurements with the Vector Network Analyzer ZVB

Vibration, Shock, and Acceleration

Sentinel I28 and Blackbelt Mass Flow Calibration and Verification

Frequency Response Function Extension Module for the Dynamix 2500 Data Collector

Continuous, remote monitoring and logging of electromagnetic fields

Application Note. Line Card Redundancy Design With the XRT83SL38 T1/E1 SH/LH LIU ICs

Fluid-Structure Acoustic Analysis with Bidirectional Coupling and Sound Transmission

ε: Voltage output of Signal Generator (also called the Source voltage or Applied

Microphone Test System

Research-Grade Research-Grade. Capture

EE 186 LAB 2 FALL Network Analyzer Fundamentals and Two Tone Linearity

MEMS mirror for low cost laser scanners. Ulrich Hofmann

99 Washington Street Melrose, MA Phone Toll Free Visit us at

Application Report. Propeller Blade Inspection Station

Operation Manual For NDJ-8S Digital Rotary Viscometer

Safakcan Tuncdemir 1, William M. Bradley *2. 1. Introduction

Impedance 50 (75 connectors via adapters)

The accelerometer designed and realized so far is intended for an. aerospace application. Detailed testing and analysis needs to be

Trigno/Vicon System Integration

DIGITAL DISPLACEMENT RISING STEP LOAD LRA/RSL TEST EQUIPMENT

Modal Testing Excitation Guidelines

Agilent PN RF Component Measurements: Amplifier Measurements Using the Agilent 8753 Network Analyzer. Product Note

PRODUCT DATA. Modal Test Consultant Types 7753 and 7753 A Operating Deflection Shapes Test Consultant Types 7765, 7765 A and 7765 B

Pre-Compliance Test Method for Radiated Emissions of Automotive Components Using Scattering Parameter Transfer Functions

Agilent N2717A Service Software Performance Verification and Adjustment Software for the Agilent ESA Spectrum Analyzers Product Overview

Physical Specifications (Custom design available.)

Machine Tool Inspection & Analyzer solutions

Development and optimization of a hybrid passive/active liner for flow duct applications

Key Features. Product Overview. Threaded Joint Rate Simulator

Power Amplifier Gain Compression Measurements

MP85 / MP85DP. Data sheet. Special features MP85 / MP85DP. B en

Reaction Torque Sensor

Measuring Temperature withthermistors a Tutorial David Potter

LMS Test.Xpress Powering personal productivity

Product Information. QUADRA-CHEK 3000 Evaluation Electronics For Metrological Applications

CALIBRATION PRINCIPLES

A FRAMEWORK FOR SMART TRANSDUCER INTERFACE SYSTEMS

DIRECTIONAL FIBER OPTIC POWER MONITORS (TAPS/PHOTODIODES)

Rapid Modal Testing System for Live Load Rating of Highway Bridges

Temperature Rise Measurements

CASE STUDY. Mazda Motor Corporation. Mazda Motor Corporation. Operational Modal Analysis. Japan. Automotive. PULSE, Software, Transducers

A Guide to Calibrating Your Spectrum Analyzer

EN54-16 Voice Alarm System PRODUCT DATASHEETS

Mounting instructions. Acceleration Transducer B12. B 26.B12.10 en

The Sonometer The Resonant String and Timbre Change after plucking

STUDY OF DAM-RESERVOIR DYNAMIC INTERACTION USING VIBRATION TESTS ON A PHYSICAL MODEL

Experimental Modal Analysis

SEISMIC MONITORING & PROTECTION SYSTEMS

Small, Low Power, 3-Axis ±3 g Accelerometer ADXL335

Alignment Laser System.

Fluid structure interaction of a vibrating circular plate in a bounded fluid volume: simulation and experiment

DDX 7000 & Digital Partial Discharge Detectors FEATURES APPLICATIONS

2. The Vector Network Analyzer

HD2030 [ GB ] Four channel vibration analyzer. Base screen: single axis measurements. Base screen: vector measurements.

THE SAS AUDIO ROUTING SWITCHER

Multiphase Flow - Appendices

Finite Element Model Calibration of a Full Scale Bridge Using Measured Frequency Response Functions

EFFECTS ON NUMBER OF CABLES FOR MODAL ANALYSIS OF CABLE-STAYED BRIDGES

ACE. Automated Soil CO 2 Exchange System. Soil flux: an important component of total carbon budget

Enhancing the SNR of the Fiber Optic Rotation Sensor using the LMS Algorithm

This datalogger is ideal for the Food Industry where process profiling is crucial to production efficiency and energy use optimisation.

Barcode option details & costing February 2008

The process components and related data characteristics addressed in this document are:

Agilent Test Solutions for Multiport and Balanced Devices

Structural Health Monitoring Tools (SHMTools)

A PC-BASED TIME INTERVAL COUNTER WITH 200 PS RESOLUTION

sigpod Press ASSEMBLY PRESS-FIT MONITORING SYSTEM BENEFITS FEATURES

MicroMag3 3-Axis Magnetic Sensor Module

Co-simulation of Microwave Networks. Sanghoon Shin, Ph.D. RS Microwave

ECE 435 INTRODUCTION TO THE MICROWAVE NETWORK ANALYZER

Performance Verification of GigaSPEED X10D Installations with Fluke Networks DTX 1800 CableAnalyzer

Advanced systems and services for Ground Vibration Testing Application for a research test on an Airbus A aircraft

Transcription:

Transducer Mounting and Test Setup Configurations Rick Bono The Modal Shop 1

Transducer Mounting Mechanical connection method Stud mount Adhesive mount Magnetic mount Press-fit friction mount Test parameter considerations Frequency range Mass loading 2

Mechanical Mounting: Impact on Frequency Range Hand Probe Dual Rail Magnet Flat Magnet Mounting Pad Adhesive Mount Stud Mount Sensitivity Deviation (db) ~ Ref. 100 Hz +40 +30 +20 +10 0-10 -20 1.0 10 100 1000 10 000 100 000 Log Frequency (Hz) 3

Stud Mount Transducers Best frequency response characteristics just like the manufacturer s cal labs Apply silicon grease at mating surface Requires surface preparation Proper torque recommended 4

Adhesive Mounting Supplies 5

Adhesive Mount Transducers Cyanoacrylate (superglue) Instant adhesive; strong, but still removable Gel vs liquid depends upon surface flatness Excellent frequency response characteristics 6

Adhesive Mount Transducers Petro wax (bees wax) Ultra convenient and simple Good for short term testing only Frequency response characteristics highly dependent upon surface prep and amount 7

Adhesive Mount Transducers Hot glue Allows attachment to poorly-mated surfaces Good for short term to mid term testing Frequency response characteristics poor, but generally good enough for modal apps 8

Adhesive Mount Transducers Dental cement / fast-cure epoxy Allows attachment to poorly-mated surfaces Pseudo-permanent attachment for reference transducer at shaker input location Use disposable mounting pad with stud 9

Magnetic Mount Transducers Extremely convenient High attraction forces allow for reasonable high frequency characteristics Available in dual-rail style for attachment to curved surfaces 10

Press-fit Mount Transducers Extremely convenient and efficient Designed specifically for low frequency (<1000 Hz) laboratory modal applications Cable base mounts adhesively, modal sensor mechanically attaches using electrical pins 11

Mounted Accelerometer Frequency Response Calibration Mounted Accelerometer Frequency Response 300.0 250.0 Cyanoacrylate Sensitivity (mv/g) 200.0 150.0 100.0 Hot Glue Putty Petrowax Stud (handtight w/out grease) Stud (w/ grease) 50.0 0.0 0 2000 4000 6000 8000 10000 12000 Frequency (Hz) 12

Mass Loading Considerations Acquire FRF with a single accelerometer Mount a second accelerometer next to the first and re-acquire FRF Compare for measurable differences 13

Test Setup Considerations Understand goals/reasons for performing experimental modal analysis Troubleshooting or failure analysis Finite element model verification Finite element model correction Component substructure / system modeling 14

Test Setup Considerations Recognize the 4 primary assumptions of experimental modal analysis Observability Time Invariance (Stationarity) Linearity Maxwell s Reciprocity 15

Observability Assumption Response DOF must have adequate spatial resolution to represent the modes of interest Graphics from Agilent Application Note 243-3 16

Observability Assumption First bending beam with seven accelerometer measurement points 17

Observability Assumption If data acquired only at endpoints bending is not observable 18

Observability Assumption 19

Observability Assumption Forcing function(s) applied at input location(s) must adequately excite the modes of interest AVOID NODES! Graphic from Agilent Application Note 243-3 20

Modally-Tuned Impact Hammers as Pre- Test Tool for Evaluating Structures... 21

for Optimizing Reference Locations and Ensuring Observability Good Input Location Freq Resp 2:1 Freq Resp 2:1 40 dbg/lb 82798009.DAT 82798229.DAT Mag (db) Bad Input Location -60 0 Hz 1 khz 22

for Determining Optimal Frequency Range Assure adequate spatial resolution to observe: - Important, dominant modes - Necessary modal density Empty Freq Resp 2:1 40 dbg/lb Mag (db) -60 0 Hz 1.6 khz 23

for Testing Boundary Conditions Rule of Thumb: 5-10x separation between rigid body and flexible modes Graphic from Agilent Application Note 243-3 24

Time Invariance Assumption Test article (and its boundary conditions) must exhibit stationarity Parameter estimation algorithms assume consistent global modal properties throughout data set Environmental changes during data acquisition cause shifts in stiffness/damping properties resulting in measurable shifts in resonant frequencies Roving accelerometers to acquire data set results in variable mass loading on test article 25

Time Invariance Assumption DATA CONSISTENCY DATA CONSISTENCY DATA CONSISTENCY i.e. acquire entire data set simultaneously (single snapshot ) or at least as fast as possible 26

Time Invariance Assumption Test methodology to achieve best data consistency Simultaneous MIMO/SIMO testing Automated bankswitching Manual bankswitching Roving accelerometers Impact testing 27

Benefits of Bank-Switching No. of Test Configurations No. of Test Configurations No. of Test Configurations Simultaneous, 288 ch PreSetup Time Roving, 112 ch PreSetup Time Bank-switch, 288 to 112 ch PreSetup Time Acquisition Time Acquisition Time Acquisition Time Cost Estimate 1 9 hrs 17 min 60% 2 17 min 3 17 min 4 17 min Total Time Allotted Cost Estimate 1 3 hrs 6 hrs 17 min 40% 2 6 hrs 17 min 3 6 hrs 17 min 4 6 hrs 17 min Total Time Allotted Cost Estimate 1 9 hrs 5 min 100% 2 5 min 3 5 min 4 5 min Total Time Allotted 9 hrs 20 min 28 hrs 8 min 10 hrs 8 min 28

Bank-Switching Example Inputs: 2 vertical, 1 lateral, 1 skewed 29

Bank-Switching Example Response points: 17 patch panels, each bank of 16 accelerometers 30

Bank-Switching Example Bank-switch patches of data (3 x 96 ch) into smaller data acquisition system 31

Modular Cabling / Patch Panel System for Clean Setup Eases setup troubleshooting Eliminates messy rat s nest of cables Economical multi-conductor cabling 32

Time Invariance Assumption Roving accelerometers results in inconsistent global resonant frequencies due to variable mass loading on test structure Graphic from Agilent Application Note 243-3 33

Linearity Assumption Input and output characteristics remain proportional within measurement range Confirm using precisely controlled inputs from shaker(s) across a range force levels Impact testing technique poorly suited when dealing with nonlinear test structures 34

Electrodynamic Modal Shakers as Excitation Source for MIMO Allows best control of input forcing function to optimize frequency content and signal-tonoise ratio Through-hole armature greatly simplifies setup attachment to test structure 35

Through-Hole Armature Eases Setup Traditional shakers with tapped armature connection leave little tolerance since setup has tapped connection at both ends 36

Reciprocity Assumption Maxwell s Theory of Reciprocity states that FRF matrix is symmetric FRF between input A and output B is the same as output A and input B Confirm using multiple shaker locations and impedance heads for driving point measurement 37

Impedance Heads for Verifying Reciprocity Assumption Accelerometer built into preload stud of force transducer Freq Resp 2:1 Freq Resp 2:1 30 dbg/lb Mag (db) -70 0 Hz 500 Hz 38

Other Pre-Test Considerations Free Boundary Conditions Shock Cord Foam Rubber Air Suspension 39

Other Pre-Test Considerations Fixed Boundary Conditions Realistic Boundary Conditions Match Impedance(s) at Boundaries Mass Loaded Boundary Conditions 40

Other Pre-Test Considerations Transducer selection Single axis vs triaxial package Sensitivity, measurement range & resolution Frequency range & mass 41

Transducer Electronic Data Sheet (TEDS, IEEE 1451.4) Identifies transducer (type, serial number, location) Stores calibration data Automates book keeping, reducing errors To ICP Supply TEDS Memory 4ma -5 V Uses reverse bias scheme to access digital memory 42

Other Transducer Setup Considerations Use PDA scanner with bar-coded TEDS transducers to ease bookkeeping 43

Final Channel Setup Definition Combine Data From Geometry, PDA, and TEDS Complete Test Set-up Information Defined in Universal Files Virtual Channel Table (1807) Channel Table (1808) Geometry (15) Parameter Stored In TEDS Stored In PDA Stored On Host (PC) Calibration X Model / Serial No. X X Direction X Node No. X X Meas. Ch. X Geometry X 44

Thank you for your time Any questions? 45