Care and use guide. Contents. a handy solution made easy. www.getbluelab.com. page



Similar documents
Care and use guide. Contents. a handy solution made easy. page

Care and use guide. Contents. page

Bluelab Soil ph Meter Instruction Manual

User Guide. ExStik TM ph (PH100 & PH110) and ORP (RE300) Waterproof Series Pens. Patent Pending

ExStik TM ph Waterproof Meters

Measuring ph in Water or CaCl2 Using a ph Meter. Compiled by Darren Murray, June 30, 2011 TABLE OF CONTENTS OVERVIEW 3

ph Measurement and Control

Milwaukee USER MANUAL. Milwaukee. Smart DO Meter PORTABLE DISSOLVED OXYGEN METER MODEL: SM600. Authorized Dealer: ISMIL600 11/01

Electrical Stability Tester, Model 23E Instruction Manual Manual No , Revision B Instrument No

ph Measurements of Common Substances

Phoenix Body Fat Scale Manual

Milwaukee USER MANUAL

SINGLE RANGE DS METERS Operation Instructions

AIRMoN, ph and Conductivity Field Measurement

YSI Meter Protocol: Specific Conductivity, ph, & Dissolved Oxygen

User Guide. Heavy Duty Dissolved Oxygen Meter. Model

Hygro-Thermometer + InfraRed Thermometer Model RH101

el-720 Carbon Dioxide Gas Leak Detector User s Guide

Household Acids and Bases

OPERATION MANUAL. Pen type, separate electrode SOIL PH METER Model : PH-220S

SenTix 950 SenTix 980

MODEL 970 DISSOLVED OXYGEN METER OPERATING MANUAL

User's Manual. Heavy Duty Dissolved Oxygen Meter Model

User's Guide. ExStik DO600. Dissolved Oxygen Meter RECALL. Dissolved Oxygen DO600

Care, Maintenance, and Storage of YSI 6-Series Probes and Sondes

USEPA Electrode Method Method 8156

MODEL 430 PORTABLE ph/conductivity METER OPERATING MANUAL

FARMEX HT-PRO OPERATING INSTRUCTIONS PORTABLE MOISTURE AND TEMPERATURE TESTER FOR BALED HAY

Bottom Loading Water Dispenser

Drop Counter Sensor Product Number: ENDRP-AD100

LASER DIAGRAM TABLE OF CONTENTS. SKY Technologies Inc. Laser Diagram. Technical Specifications. Laser Operation.

PROCEDURE COVERSHEET. Document Section: CHEMISTRY Author: JOANNE WONG Date Authored: JAN In Use Date: 2/1/2011 Discontinued Date:

Optical Power Meter. Specification & User Manual

6 H2O + 6 CO 2 (g) + energy

Direct Measurement Method Method to 20.0 mg/l (or 0 to 200% saturation) O 2 Clark-type Amperometric Sensor

Campus Services Hints and tips for residents

USER INSTRUCTIONS FOR GET PORTABLE 12k BTU AIR CONDITIONER MODEL No. GPACU12HR

Determining the Quantity of Iron in a Vitamin Tablet. Evaluation copy

Pro1030 USER MANUAL. English

Compact ph Measuring Instruments

Waters Corporation. Waters 2690/5 USER & TROUBLESHOOTING GUIDE

Instruction Manual Temp 4/5/6 Temperature Meter Series

Private Water Supplies Sampling Manual. A Field Guide

Nucleus Freedom BTE Rechargeable Battery (A25) and Nucleus Freedom BTE Recharger

USER MANUAL. Bottom Loading Bottled Water Dispenser SAVE THIS MANUAL FOR FUTURE USE. Model #

Refer to page 6 to page 19 for detailed instructions.

Professional Plus. Calibration Tips

The Influence of Carbon Dioxide on Algae Growth

Measuring the Conductivity of Pure Water Using USP<645>

CQ-Series. User Manual. American Weigh Scales, Inc. CQ-350 (350x0.1g) CQ-500 (500x0.1g) Digital Pocket Scale . = + CALC AC CE % TARE OFF ON MC MR CALI

Handbook. Thermo Scientific. ph Electrode Selection Handbook

Deluxe Bottle Warmer

Direct ISE Method Method to 1000 mg/l Na + Sodium ISE

Measurement by Ion Selective Electrodes

Microwave Meter. Instruction Manual

Lead Testing and On Site Calibration for Water Testing Detection Range: 2 100ppb

Ground Resistance Clamp On Tester

Instruction Manual HI Refractometer for Refractive Index and Brix Measurements

PART 1 - INTRODUCTION...

ph Sensor Model 397 CAUTION SENSOR/PROCESS APPLICATION COMPATIBILITY WARNING ATEX DIRECTIVE Instruction Sheet PN 51A-397/rev.

POOL SUNSHINE IONIZER MODELS SPD, SPD2 AND SPS INSTALLATION INSTRUCTIONS

NewAir AC-10100E / AC-10100H Portable Air Conditioner Owner s Manual PLEASE READ AND SAVE THESE INSTRUCTIONS

Compact Water Quality Meter B-700 series

Instruction Manual ph 5+ ph 6+ Ion 6+ Technology Made Easy...

UltraBasic Benchtop ph Meter Operation Manual

EQUIPMENT CALIBRATION/MAINTENANCE

Calibration of Volumetric Glassware

NewAir AC-10000E, AC-10000H Portable Air Conditioner Owner s Manual PLEASE READ AND SAVE THESE INSTRUCTIONS

English. Symbols used to mark instructions...3. Congratulations...5 Getting the best results...5. Warnings...6 Operating Procedure...

Portable Air Conditioner. OWNER S MANUAL Read these instructions before use. Model: MF08CESWW. Voltage rating: 115V~60Hz Power rating : 800W

User's Guide. Pinless Moisture Meter. Model MO257

orlando OWNER S MANUAL

RBS detergents for laboratories

Operating Manual Remote Cost Control Monitor

Portable Air Conditioner. OWNER S MANUAL Read these instructions before use. Model: MN12CES / MN10CESWW

Nitrogen, Ammonia. Known Addition ISE Method 1 Method Minimum of 0.8 mg/l NH 3 N. Ammonia ISE. Test preparation. Instrument-specific table

ClikSTAR - Important facts about your new insulin delivery device.

Portable Air Conditioner. OWNER S MANUAL Read these instructions before use. Model: MM14CCS. Voltage rating: 115V~60Hz Power rating : 1400W

Please read and keep these instructions. Run water through your machine before first use

LCD BacPac. User Manual + Warranty Info

Use and Care Manual. Model CPA12KH AIR CONDITIONER

Atomic Alarm Clock model 13131W2

QuickStart Guide for The PeriCoach System

Appendix C. Vernier Tutorial

MAP COFFEE MIA CAPSULE COFFEE MACHINE INSTRUCTION MANUAL MODEL: CM4725. Read this manual thoroughly before using and save it for future reference

Instruction Manual. Image of SP-3015 & SP Important Safeguards. Automatic Dispensing Hot Water Pot with Reboil Function

User Guide Dissolved Oxygen Probes

The Drink-Aide The NEW look of independence

USER MANUAL WARNING! CONTENTS MODEL 1 SPECIFICATIONS READ ALL INSTRUCTIONS BEFORE PROCEEDING. Non-Programmable Single Stage Heat/Cool Thermostat

Dissolved Oxygen Sensors

RT-1261A Ear and Forehead Thermometer Owner s Manual

ELECTRIC POSITION for electric heat, then confirm with Fan Test below.

Tassimo The Art of the Perfect Cup

Important Safeguards

TEMPERATURE SENSOR USER MANUAL

Experiment 18: ph Measurements of Common Substances. Experiment 17: Reactions of Acids with Common Substances

Thermistor Thermometer MODEL NO

Instruction Manual. Window Wiper AFW900. (Read carefully before use)

CL90i Please read these instructions before operating the product. 3 - Beam Self-Leveling Cross Line Laser

Transcription:

Care and use guide Contents page Features Quick guide Before use To operate Important - ph pen probe care 4 Cleaning 5 Battery replacement 5 Hydration 6 Calibration 6 Error messages 7 Troubleshooting guide 7 Technical specifications 8 Bluelab Probe Care Kit - ph 8 Bluelab ph Probe KCl Storage Solution 8 Product guarantee 9 Contact details 9 a handy solution made easy www.getbluelab.com

Features Backlit LCD display Low battery indicator Hold reading function Fully waterproof Fully guaranteed for year Auto off function Double junction probe (not replaceable) Automatic Temperature Compensation (ATC) Successful calibration indicator Selectable units for C and F Quick guide Low battery warning Appears when batteries are low. Check mark/tick to indicate successful calibration Disappears 0 days after calibrating probe as a calibration due reminder. Power button / hold Short press to turn on. Short press to hold reading. Long press to turn off. Calibrate button See calibration section. Units button Hold until units flash then short press to change units. Screen will change back when no buttons have been pressed for seconds. Storage cap can be placed onto loop while being used. Storage cap The probe must not be allowed to dry out. Always place storage cap back onto probe tip after each use. Add -5 drops of KCl storage solution to the round hole (wetting seal) in the cap each week to keep the probe wet. See section.0 ATTENTION If it dries, it dies! Keep your probe tip wet at all times to avoid permanent damage

.0 Before use begins Hydrate in KCl storage solution for 4 hours before you start using the ph pen. See section 6.0 Calibrate the pen before you start using the pen. See section 7.0.0 To operate Turn pen on Press power button. To turn pen off Press and hold the power button until OFF is displayed. NOTE: The pen will automatically turn off after 4 minutes to conserve battery power. Measure ph Remove storage cap, place probe in solution and wait for reading to stabilize. Power button To hold reading If you want to hold the reading on the screen, short press the power button. To exit the hold function press the power button again. second alternating displays 4 To change temperature units Hold down the units button for seconds until the temperature units start fl ashing. Short press the units button again to cycle between F and C. To exit this mode don t press anything for seconds. NOTE: You can change units while in hold mode by holding down the units button. 5 Rinse and replace storage cap To ensure accurate ph readings always rinse the probe in clean fresh water before replacing the cap. Add -5 drops of KCl storage solution in to the wetting seal of the cap each week. The cap is tight to ensure a good seal and should click when on correctly.

.0 IMPORTANT - ph pen probe care ph probes do not last forever. They age through normal use and will eventually fail. The lifetime of a probe depends on the environment it is used in and the way it is treated. To ensure you receive a long life from your pen, please ensure you follow the guide below. Storing the ph pen When storing the ph pen, the probe tip must be kept moist. Add -5 drops of KCl storage solution into the wetting seal inside the storage cap once a week. Never use RO, Distilled or De-ionized water. Place the seal and cap over the probe tip. Long term storage For long term storage, remove the cap and place the pen upright in a plastic container. Cover the probe tip with KCl storage solution to ensure constant hydration. Check the container regularly and top up the KCl storage solution as needed. If the probe has been accidentally allowed to dry out: The probe must be hydrated for 4 hours in KCl storage solution (never use RO, Distilled or Deionised water). Following this; carry out a calibration to check if the probe has already suffered permanent damage. Temperature probe Wick Glass bulb Storage cap Wetting seal 4 DO NOT let the probe tip dry. IF IT DRIES IT DIES! DO NOT knock the pen; this will break its external glass bulb or internal glass tube. DO NOT touch the glass bulb with your fi ngers as this will contaminate the glass. DO NOT plunge a cold probe into a hot liquid (or visa versa) - sudden temperature changes can crack the glass and permanently damage the pen. DO NOT soak or rinse the ph probe in RO (Reverse Osmosis), Distilled or De-ionized water. Pure water changes the chemistry in the reference, causing the probe to die. DO NOT immerse in oils, proteins or suspended solids that will leave a coating on the glass bulb.

4.0 Cleaning To ensure accurate readings the probe needs to be rinsed in water before replacing the storage cap and cleaned regularly using the following instructions. Rinse probe tip under fresh water. Fill small container with clean water. Add a small amount of Bluelab ph Probe Cleaner or mild detergent (dish washing liquid). 4 Gently stir the probe tip in the mixture. Ensure that you do not knock the pen on the side of the container as this may cause damage to the glass probe. Rinse well under fresh running water to remove all traces of the detergent mixture. If the probe requires removal of heavy contamination: Gently brush around the glassware with a few drops of Bluelab ph Probe Cleaner or mild detergent (dishwashing liquid) and a soft toothbrush. Detergent 5 6 Rinse well under fresh running water to remove all traces of the detergent mixture. Calibration of the probe is required after every clean. See the ph calibration in section 7.0. Place storage cap back onto the probe. 5 5.0 Battery replacement The ph pen is powered with x AAA alkaline battery. Do not use rechargeable batteries. A low battery warning is indicated by a battery symbol appearing on the screen. Only remove the battery cap when the batteries require changing. Battery life is expected to be 50 hours. To remove old battery Undo battery cap fasteners. Remove battery cap and tip out the old battery. Check for corrosion Flat batteries may leak and cause corrosion. Check battery contacts and the battery for any sign of corrosion. Battery contacts should be cleaned fi rst if corrosion is found before proceeding to step. Fit new battery Insert the new batteries positive (+) end down into the body. Waterproof seal 4 Ensure waterproof battery cap seal is clean Seal will fail if any dirt is present. 5 Replace battery cap Tighten fasteners on battery cap until there is no space left between the cap and body. This ensures the unit remains 00% waterproof.

6.0 Hydration Hydrate the ph pen tip in Bluelab ph Probe KCl Storage Solution before first use and after cleaning to improve the reading response speed. Never use RO, Deionized or Distilled water. Pure water changes the ions, causing the probe to die. Remove the storage cap. Place the ph pen upright in a small plastic container. Add enough Bluelab ph Probe KCl Storage Solution to submerge the probe tip. Leave to soak for at least 4 hours. After hydration, always calibrate the ph pen to ensure accuracy. See section 7.0. KCl Storage Solution 7.0 Calibration ph calibration is required before first use to ensure that the first reading is accurate. Calibration is also required when: The check mark/tick has disappeared from the LCD screen (0 days after last calibration) The reading is different from what you expected After cleaning and hydration After changing the batteries ph 7.0 and ph 4.0 solutions are required for calibration. You may also calibrate using ph 7.0 and ph 0.0 solutions if your readings will normally be higher than 7.0 ph. Excluding first use, YOU MUST CLEAN the probe before calibrating. See section 4.0. Hydrate the probe before fi rst use and after cleaning, see section 6.0 After hydration, rinse probe in fresh water and place in ph 7.0 solution. Wait for the reading to stabilize. Hold until Press the cal button until CAL is displayed. Release button. When CAL 7 is displayed, point calibration is complete. ph 7.0 4 Rinse probe in fresh water and place it in either ph 4.0 or ph 0.0 solution (use ph 0.0 solution if you expect to measure above 7.0 ph). Wait for reading to stabilize. Hold until 6 5 Press the cal button until CAL 4 or CAL 0 is displayed CAL 4 or CAL 0 should be displayed (depending on what solution you are calibrating in). The check mark/tick is displayed when a point (or ) calibration is completed. NOTE: For a three point calibration repeat the steps using ph 7.0, ph 4.0 then ph 0.0 solution. ph 4.0 or ph 0.0

8.0 Error messages The following error messages appear for the following reasons. Temperature under range ph over range Temperature over range ph under range ph calibration failed Hardware error 9.0 Troubleshooting guide Trouble Reason Correction Glassware not clean Clean glassware, hydrate and calibrate. Drift - readings slowly varying Displays similar ph reading in all buffers no matter what the buffer value is Wick contaminated or blocked Glassware aged Glassware broken Buffers inaccurate Soak probe in KCl storage solution for 4 hours and retest. Do not measure proteins or oils with this unit. Replace unit. Replace unit. Replace unit. Replace buffers. 7 Unsuccessful calibration Noisy - readings jumping Displays ph 7 for all buffers Incorrect sample reading following successful calibration Glassware not clean Glassware aged (glassware will not clean) Clean glassware. Replace unit. Probe not hydrated Soak probe in KCl storage solution for 4 hours and retest. Contact zone not immersed Broken glassware Ground loop (often occurs in process systems) Lower pen into solution at least cm/. Replace unit. Verify by removing the sample from its environment and measuring in a glass beaker. May require electrical circuitry to be checked in system. Wick blocked Soak probe in KCl storage solution for 4 hours and retest. Do not measure proteins or oils with this unit. Replace unit.

0.0 Technical specifications Measurement range Resolution 0.0-4.0 ph 0. ph Accuracy (at 5 C / 77 F) ± 0. ph Temperature compensation Automatic Operating temperature 0-50 C, - F Calibration Manual calibration, or points Units ph, F and C Power source x AAA alkaline battery Bluelab Probe Care - ph The instrument is only as accurate as the probe is clean! Probe cleaning is one of the most important parts of owning and operating any Bluelab meter, monitor or controller. If the probe is contaminated (dirty) it affects the accuracy of the reading displayed. Bluelab Probe Care Kit - ph contents: Cleaning instructions inside box lid 500ml ph4.0 and ph7.0 calibration solutions Plastic cups Bluelab ph Probe Cleaner Toothbrush (probe cleaning instrument) Bluelab ph Probe KCl Storage Solution The perfect solution to store and hydrate your Bluelab ph products. Bluelab ph Probe KCl Storage Solution is designed to increase response time and maximize the life of Bluelab ph pens and ph probes. For best results, use the KCl solution to store the ph pen/ probe after use and hydrate monthly. Instructions are on the label of the bottle. Use Bluelab ph Probe KCl Storage Solution with: 8 Bluelab ph Pen Bluelab Soil ph Pen Bluelab ph Probes Bluelab Soil ph Probes

Bluelab ph Pen product guarantee Bluelab Corporation Limited guarantees this product for a period of year ( months) from the date of sale to the original purchaser. The product will be repaired or replaced, should it be found faulty due to component failure, or faulty workmanship. The faulty product should be returned to the point of purchase. The guarantee is null and void should any internal parts or fixed external parts be tampered with or altered in any way, or should the unit have been incorrectly operated, or in anyway be maltreated. This guarantee does not cover reported faults which are shown to be caused by any or all of the following: contaminated measuring tip (see instruction manual for cleaning instructions), broken glassware or drying of the ph probe glassware, flat or damaged batteries or batteries that have been incorrectly inserted, or damaged battery contacts or connections caused by incorrect battery replacement or ingress of moisture from incorrect positioning of the battery cap and waterproof seal. NO RESPONSIBILITY will be accepted by Bluelab or any of its agents or resellers should any damage or unfavourable conditions result from the use of this product, should it be faulty or incorrectly operated. Register your guarantee online at www.getbluelab.com Limitation of Liability Under no circumstances shall Bluelab Corporation Limited be liable for any claims, losses, costs and damages of any nature whatsoever (including any consequential loss) that result from the use of, or the inability to use, these instructions. To watch instruction videos, visit our online video library: vimeopro.com/bluelab/videos If you need assistance or advice - we re here to help you. Phone: +64 7 578 0849 Fax: +64 7 578 0847 Email: support@getbluelab.com Looking for specifi cations or technical advice? Visit us online at www.getbluelab.com Bluelab Corporation Limited 8 Whiore Avenue, Tauriko Industrial Park Tauranga 0, New Zealand 9 Instruction Manual PENPH_V0_05 Copyright 0, all rights reserved, Bluelab Corporation Limited