Comparison of central corneal thickness measurements using Pentacam and ultrasonic pachymetry in post-lasik eyes for myopia



Similar documents
Long-term stability of the posterior cornea after laser in situ keratomileusis

Corneal Thickness Measurement with Optical and Ultrasonic Pachymetry

Comparison of Residual Stromal Bed Thickness and Flap Thickness at LASIK and Post-LASIK Enhancement in Femtosecond Laser-Created Flaps

Comparison to other anterior segment imaging devices

Keratoconus Detection Using Corneal Topography

PARALLEL TO THE DEVELOPMENTS OF SURGICAL

FIRST EXPERIENCE WITH THE ZEISS FEMTOSECOND SYSTEM IN CONJUNC- TION WITH THE MEL 80 IN THE US

Retreatment by Lifting the Original Laser in Situ Keratomileusis Flap after Eleven Years

THE BEST OF BOTH WORLDS Dual-Scheimpflug and Placido Reaching a new level in refractive screening

Total Corneal Power Estimation: Ray Tracing Method versus Gaussian Optics Formula PATIENTS AND METHODS

Customized corneal ablation and super vision. Customized Corneal Ablation and Super Vision

Diego Fernando Suárez Sierra, MD Fellow Cornea and Refractive Surgery Fellow Lens and Ocular Surface Vejarano Laser Vision Center

Life Science Journal 2014;11(9) Cross cylinder Challenging cases and their resultswith Nidek Quest (EC-5000)

IOL Power Calculation After Myopic LASIK. Hany Helaly, Lecturer of Ophthalmology, Faculty of Medicine, Alexandria University.

Alexandria s Guide to LASIK

Comparison Combined LASIK Procedure for Ametropic Presbyopes and Planned Dual Interface for Post-LASIK Presbyopes Using Small Aperture Corneal Inlay

Uniquely Safe. predictably better for our patients. enhancement, may be significantly reduced.

Ectasia after laser in-situ keratomileusis (LASIK)

Title: Anterior and Posterior Corneal Elevation after Orthokeratology, Standard and Customized LASIK Surgery

Refractive Surgery. Common Refractive Errors

Active Cyclotorsion Error Correction During LASIK for Myopia and Myopic Astigmatism With the NIDEK EC-5000 CX III Laser

Case Reports Post-LASIK ectasia treated with intrastromal corneal ring segments and corneal crosslinking

Common Co-management Questions

PATIENT CONSENT FOR LASER IN-SITU KERATOMILEUSIS (LASIK)

Curtin G. Kelley, M.D. Director of Vision Correction Surgery Arena Eye Surgeons Associate Clinical Professor of Ophthalmology The Ohio State

ALTERNATIVES TO LASIK

Excimer Laser Refractive Surgery

Corneal Power Measurements Using Scheimpflug Imaging in Eyes With Prior Corneal Refractive Surgery

refractive surgery a closer look

IntraLase and LASIK: Risks and Complications

Dr. Booth received his medical degree from the University of California: San Diego and his bachelor of science from Stanford University.

Customized corneal ablation can be designed. Slit Skiascopic-guided Ablation Using the Nidek Laser. Scott MacRae, MD; Masanao Fujieda

Validation of a New Scoring System for the Detection of Early Forme of Keratoconus

Consumer s Guide to LASIK

Page: 1 of 6. Corneal Topography/Computer-Assisted Corneal Topography/ Photokeratoscopy

Changes in central corneal thickness and refractive error after thin-flap laser in situ keratomileusis in Chinese eyes

INFORMED CONSENT FOR LASER IN-SITU KERATOMILEUSIS (LASIK)

INTRODUCTION MATERIAL AND METHODS

ALL-IN-ONE Optical Biometry, Dual Scheimpflug Tomography and Placido Topography

How do we use the Galilei for cataract and refractive surgery?

TABLE OF CONTENTS: LASER EYE SURGERY CONSENT FORM

Long-Term Outcomes of Flap Amputation After LASIK

Standardized Analyses of Correction of Astigmatism With the Visian Toric Phakic Implantable Collamer Lens

Ectasia Risk Scoring System (ERSS): Limitations and Variability in Predictive Value

Eye Associates Custom LASIK With IntraLASIK Correction Of Nearsightedness, Farsightedness, and Astigmatism Using IntraLase TM Technology

IOL Power Calculations for Postrefractive Surgery Eyes

Anterior Elevation Maps as the Screening Test for the Ablation Power of Previous Myopic Refractive Surgery

To date, several million patients have been treated worldwide. So why not discover the benefits The Eye Hospital can bring to your life.

Corneal intrastromal implantation surgery for the treatment of moderate and high myopia

Pentacam. The. The gold standard in anterior segment tomography. Supplement to February 2011

OCULUS PENTACAM BASIC CLASSIC HR

INFORMED CONSENT TO HAVE LASIK

Simple regression formula for intraocular lens power adjustment in eyes requiring cataract surgery after excimer laser photoablation

Your one stop vision centre Our ophthalmic centre offers comprehensive eye management, which includes medical,

Corneal flap thickness with the Moria M2 single-use head 90 microkeratome

Refractive Errors & Correction

Overview of Refractive Surgery

Alain Saad, MD, Alice Grise-Dulac, MD, Damien Gatinel, MD, PhD

LASIK. What is LASIK? Eye Words to Know. Who is a good candidate for LASIK?

VA high quality, complications low with phakic IOL

Epi-LASIK Surgery. The procedure of choice for surface ablation.

By Dr Waleed Al-Tuwairqi, MD Dr Omnia Sherif, MD Ophthalmology Consultants, Elite Medical & Surgical Center Riyadh -KSA.

LASIK SURGERY OUTCOMES, VOLUME AND RESOURCES

INFORMED CONSENT FOR LASER IN-SITU KERATOMILEUSIS (LASIK) USING INTRALASE TM BLADE-FREE TECHNOLOGY

Refractive Surgery Issue. Inlays and Presbyopia: On the Horizon P. 24. Crack a SMILE or Raise a Flap? P. 30. LASIK Xtra: Who Should Get It? P.

Keratorefractive Surgery for Post-Cataract Refractive Surprise. Moataz El Sawy

INTRODUCTION. Trans Am Ophthalmol Soc 2006;104:

INTRACOR. An excerpt from the presentations by Dr Luis Ruiz and Dr Mike Holzer and the Round Table discussion moderated by Dr Wing-Kwong Chan in the

For approximately two decades photorefractive keratectomy. Seven-Year Changes in Corneal Power and Aberrations after PRK or LASIK.

ReLEx smile Minimally invasive vision correction Information for patients

One of the exciting new research areas in laser

final corrected draft

The Efficacy of Multi-Zone Cross-Cylinder Method for Astigmatism Correction

LASIK: Clinical Results and Their Relationship to Patient Satisfaction

This article appeared in a journal published by Elsevier. The attached copy is furnished to the author for internal non-commercial research and

The Nurse/Technician Role Within the Emerging Ophthalmic Technology - OCTs/B-Scan

Factors Affecting Long-term Myopic Regression after Laser In Situ Keratomileusis and Laser-assisted Subepithelial Keratectomy for Moderate Myopia

Effect of Preoperative Keratometric Power on Intraoperative Complications in LASIK in 34,099 Eyes

How do you adjust for IOP in a LASIK patient?

The Shape of Posterior Corneal Surface in Normal, Post-LASIK, and Post epi-lasik Eyes

ICRS implantation with the Femto LDV laser in stabilized KC patients: 6 months results

Informed Consent for Refractive Lens Exchange (Clear Lens Replacement)

Management of Unpredictable Post-PRK Corneal Ectasia with Intacs Implantation

The Pentacam: Precision, Confidence, Results, and Accurate Ks!

Post LASIK Ectasia. Examination: Gina M. Rogers, MD and Kenneth M. Goins, MD

OCT-based IOL power calculation for eyes with previous myopic and hyperopic laser vision correction

...You Need to know about

How To Implant A Keraring

Our Commitment To You

1-1 INDIAN OIL CORPORATION - REFRACTIVE SURGERY CENTRE ( LASIK) SANKARA NETHRALAYA (JKCN BRANCH) NO 21, PYGROFTS GARDEN ROAD, CHENNAI 6

Insert to. January Comfort and confidence for all IOL calculations

Underestimation of Intraocular Pressure in Eyes After Laser In Situ Keratomileusis

Minimally Invasive Surgery: Femtosecond Lasers and Other Innovative Microsurgical Techniques

Case Report Laser Vision Correction on Patients with Sick Optic Nerve: A Case Report

Consent for LASIK (Laser In Situ Keratomileusis) Retreatment

Financial Disclosure. LASIK Flap Parameters IntraLase Microkeratome 6/9/2008. Femtosecond LASIK Flaps: What Could We Customize Yesterday?

CONSENT FOR INTRALASIK CORRECTION OF NEARSIGHTEDNESS, FARSIGHTEDNESS, AND ASTIGMATISM USING INTRALASE TM TECHNOLOGY BY AMJAD KHOKHAR, M.D.

Key Words: Aberration, LASEK, LASIK, Posterior corneal surface, Wavefront-guided ablation

KERATOCONUS IS A BILATERAL, ASYMMETRIC, CHRONIC,

Intraoperative Complications of Laser in Situ Keratomileusis in Yazd

Transcription:

Eur J Ophthalmol 2010; 20 ( 5) : 852-857 Original Article Comparison of central corneal thickness measurements using Pentacam and ultrasonic pachymetry in post-lasik eyes for myopia Hani S. Al-Mezaine 1, Saleh A. Al-Amro 1, Dustan Kangave 2, Saleh Al-Obeidan 1, Khaled M. Al-Jubair 3 1 Department of Ophthalmology, College of Medicine, King Saud University, Riyadh - Saudi Arabia 2 Research Department, College of Medicine, King Saud University, Riyadh - Saudi Arabia 3 Department of Ophthalmology, National Guard Hospital, Riyadh - Saudi Arabia Pu r p o s e. To compare central corneal thickness (CCT) measurements obtained using the Pentacam Scheimpflug system with those obtained using DGH ultrasound pachymetry (UP) in post laser in situ keratomileusis (LASIK) eyes for myopia. Me t h o d s. In a prospective study, measurement agreement was assessed in 143 eyes of 72 post-lasik patients using both the Pentacam and UP at the Eye Consultants Center in Riyadh, Saudi Arabia. Re s u l t s. The mean CCT was 522±42.2 µm with the Oculus Pentacam and 516.2±40.6 µm with UP. The Bland-Altman plot showed that the mean ± SD for the differences between the 2 devices was 5.8±13.6 μm, with 95% confidence interval limits ranging from 20.9 μm to 32.6 μm. A test of statistical significance indicated that the mean differences of 5.8±13.6 μm differed significantly from zero (p<0.001; Wilcoxon signed-rank test), thus indicating that the Pentacam measurements tended to overestimate CCT compared with UP. Analysis of regression showed a high correlation between the values obtained with both devices (r=0.947, p<0.001). Co n c l u s i o n s. In post-lasik myopic eyes, although a high correlation has been shown between Pentacam and UP measurements, Pentacam tends to overestimate CCT compared to UP. Pentacam probably cannot be used interchangeably with UP in post-lasik eyes for myopia. (Eur J Ophthalmol 2010; 20: 852-7) Key Words. Corneal thickness, Laser in situ keratomileusis, Pachymetry, Pentacam, Ultrasound Accepted: February 5, 2010 INTRODUCTION Attaining accurate central corneal thickness (CCT) measurements is important in the preoperative and postoperative management of corneal surgical procedures such as laser in situ keratomileusis (LASIK), a popular approach for the correction of refractive errors. Accuracy of measurements is particularly important in patients who have undergone previous LASIK with suboptimal outcomes and are being considered for an enhancement procedure. As postrefractive patients age, they are also more likely to experience glaucoma, cataracts, and degenerative corneal conditions. Therefore, having an accurate CCT measurement concerns not only surgeons who perform keratorefractive surgery but also any ophthalmologist who monitors and treats glaucoma, cataracts, and corneal disease states. Ocular ultrasound, which has been used over the last 30 years, is still recognized as the most commonly used method of accessing and quantifying ocular structures (1). Ultrasound pachymetry (UP) is an efficient and accurate way to measure corneal thickness and is considered the current gold standard (2). Its advantages include ease of 852

Al-Mezaine et al of the patient s eye on a computer screen, with the machine marking the pupillary edge and the corneal apex. The operator can manually focus and align the image. Arrows are displayed on the screen to guide the operator s alignment of the instrument in the horizontal, vertical, and anteroposterior axes. To reduce operator-dependent variables, the Pentacam s automatic release mode was used. In this mode, the instrument automatically determines when correct focus and alignment with the corneal apex have been achieved and then performs a scan in less than 2 seconds. The rotating camera captures up to 50 slit images of the anterior segment while minute eye movements are captured by the second camera and are corrected simultaneously. Each slit image consists of 500 true elevation points. The anterior surface of the cornea is calculated using mathematical software with no optical distortion, and, according to the manufacturer, the tear film has no effect on measurements. Single-point pachymetric measurements of the entire cornea are calculated from the front and back surfaces using ray tracing, with the calculation taking into account optical distortions. Because the center of the cornea is measured repeatedly during the rotational imaging process (in each of the images), a very precise determination of CCT measurement can be achieved. Single CCT measurement for each cornea using Pentacam was taken. All measurements were taken by a single observer. UP was determined using an A-scan ultrasonic pachymeter (DGH Technology, Inc.), which was precalibrated for all measurements. The ultrasonic velocity was set at 1,640 m/s. Topical anesthesia with benoxinate hydrochloride 0.4% (Novesin 0.4%) was used. CCT was measured with patients seated upright. A hand-held probe was aligned as perpendicular as possible on the central cornea. Five readings were obtained and averaged. The Pentacam readings were taken first to avoid any potential disturbance caused by the corneal contact involved in UP. All measurements were taken between 4:00 p m and 9:00 p m and were done by a single observer who was blinded to the Pentacam results, which were taken by another observer. The distributions of ultrasound and Pentacam measurements for CCT were compared using histograms. The differences between measurements of both methods were investigated by Wilcoxon signed rank test. The correlation between the measurements obtained using the 2 methods was investigated by computing the Pearson correlation coefficient. The Bland-Altman graph was used to study the agreement between the 2 devices in the studied vause, portability, and low cost. However, concerns about the possibility of patient discomfort, epithelial damage, and infection with this contact method exist. In addition, errors caused by indentation of the cornea have been reported (3). The Oculus Pentacam is a fast, noncontact method of measuring corneal thickness, which uses a rotating Scheimpflug camera that provides a 3-dimensional scanning of the anterior segment of the eye (4). From the images acquired, information regarding the anterior and posterior corneal topography, corneal pachymetry, anterior chamber depth, angle, and lens density can be evaluated. Because of video monitor control, precise positioning of the instrument s camera system is assisted by magnification, thus avoiding measurement bias caused by incorrect centration. The Pentacam has been widely used for topographic pachymetry in virgin corneas (5-11); it has been compared to UP, giving a thicker (5, 9-11), thinner (6-9), or similar central CT (6-8, 10). To our knowledge, few reports compared these modalities in post-lasik corneas (12-14). In this study, we compared the accuracy of CCT measurements using the Oculus Pentacam and UP in myopic patients who underwent LASIK in a relatively large sample of patients. PATIENTS AND METHODS The Oculus Pentacam (Oculus Inc., Wetzlar, Germany) and DGH-1000 ultrasonic pachymetry (DGH Technology Inc., San Diego, CA) were used to measure CCT in a total of 143 post-lasik myopic eyes of 72 patients who presented 6 weeks or more postoperatively at the Eye Consultants Center in Riyadh, Saudi Arabia. None of the patients had ocular diseases other than the corrected refractive error. We excluded from the study any patient with intraoperative or postoperative complications. All subjects were volunteers and had provided informed consent. LASIK procedures were performed by a single surgeon (K.M.J.). The corneal flaps were created using a gas turbine microkeratome (Moria, Antony, France). Laser photoablation was performed using a Nidek EC-5000 Excimer Laser Corneal Surgery System (Nidek, Gamagori, Aichi, Japan). When the Oculus Pentacam is used, the patient is seated with his or her chin on a chin rest and forehead against the forehead strap and is asked to stare straight ahead at a fixation target. The operator visualizes a real-time image 853

Central corneal thickness measurements in post-lasik eyes Pentacam A Fig. 2 - A Bland-Altman plot for the Pentacam minus ultrasound differences. B Fig. 1 - Histogram of central corneal thickness, as measured by the Oculus Pentacam (A) and ultrasound pachymetry (B). riables, and the 95% limits of agreement were calculated. The mountain plot graph was used to assess the percentile distributions of negative and positive differences. Data are presented as ± standard deviation. The MedCalc statistical software version 9.3, Statistical Product and Service Solutions (SPSS) version 11.5, and BMDP 2007 software were used for the analysis. A p value less than 0.05 was considered significant. RESULTS CCT measurements were performed in a total of 143 (72 right, 71 left) post-lasik myopic eyes of 72 patients using UP and Pentacam. The mean postoperative period of measuring the CCT was 7.30±1.6 weeks (range, 6 16 weeks). The mean myopic spherical equivalent (SE) was 3.90±1.97 D (range, 9.75 to 0.75D). We used graphical and analytical methods to investigate the equivalence of measurement results obtained with the 2 methods. As shown in Figure 1, A and B, the ultrasound and the Pentacam measurements show close similarities in distribution as well as in central tendency and dispersion. The mean ± SD for the ultrasound distribution of CCT measurements was 516.2±40.6 μm as compared with 522.0±42.2 μm for the Pentacam CCT measurements. The Bland-Altman graph is shown in Figure 2. The mean ± SD for the differences was 5.8±13.6 μm, with 95% confidence interval (CI) limits ranging from 20.9 μm to 32.6 μm. The graph shows that, with the exception of a few outlier differences, the majority of differences between the Pentacam and the ultrasound measurements fell within the 95% confidence interval for the distribution of differences. A test of statistical significance indicated that the mean differences of 5.8±13.6 μm differed significantly from zero (p<0.001; Wilcoxon signed rank test), thus indicating that the Pentacam measurements tended to overestimate CCT compared with UP. The distribution of Pentacam ultrasound differences was further investigated using a mountain plot graph (Fig. 3). The graph should be symmetrical about a value of zero, along the horizontal axis, if the percentile distributions of negative and positive differences were similar. The graph 854

Al-Mezaine et al Pentacam Fig. 3 - A mountain plot demonstrating percentiles of differences in the distribution between the Pentacam minus ultrasound pachymetry values. Fig. 4 - Scatterplot graph of the central corneal thickness, as measured by the Pentacam versus ultrasound pachymetry. is slightly shifted to the right of zero, thus indicating that CCT measurements obtained by the Pentacam were often higher than those by UP. As shown in Figure 4, a very strong correlation (r=0.947; p<0.001) exists between measurements obtained using the 2 methods. DISCUSSION In this prospective case series, we compared CCT measurements between the Pentacam and UP in post-lasik eyes for myopia. Although the CCT measurements using the 2 devices were highly correlated, Pentacam significantly overestimated CCT compared to US pachymetry. Because biomechanical stability of the cornea after excimer laser surgery depends on residual corneal thickness, postoperative pachymetry is very important, especially in candidates for enhancement surgery. Overestimation of corneal thickness may increase the risk of corneal ectasia in ineligible patients. On the other hand, underestimation can result in exclusion of patients who may be candidates for safe refractive surgery (15). Previously, few studies evaluated CCT using the Pentacam and UP in eyes that had undergone LASIK (12-14). Ciolino et al (13) demonstrated that CCT measurements obtained using the Pentacam have good correlation and agreement with those obtained using UP; therefore, they suggested that the Pentacam can be used as a substitute for UP in post-lasik patients. In contrast, Ho et al (12) showed that the Pentacam significantly underestimated CCT compared with UP. In addition, Hashemi et al (14) demonstrated that Pentacam underestimated CCT in comparison to UP in a series of 38 eyes that had LASIK. Our study showed that the Pentacam significantly overestimated CCT measurements compared to UP in post-la- SIK eyes. We are not sure of the reason for the discrepancy between different reports. This discrepancy may have occurred because the ultrasound probe requires corneal touch it is well known that the applanation force can displace the 7- to 30-µm-thick tear film or that the probe compression may thin the epithelium, thus resulting in slightly thinner CCT when measured with UP (16, 17). In addition, UP includes the use of a local anesthetic, which may alter the corneal thickness (18-20). Moreover, the location where the ultrasonic probe is applied may vary in repeated measurements because there is no fixation target for controlling eye movement (21). Other possible causes of the inconclusive results might be explained by any of the following: pachymeters with low reproducibility, inexperienced observers, or no consideration of diurnal variation. Further studies are needed to address the causes of such discrepancies. 855

Central corneal thickness measurements in post-lasik eyes The reproducibility of optical devices such as Pentacam and Orbscan that rely on measurements of scattered reflected light beams through the corneal tissues might be questionable when the corneal medium is not clear or has optical interference such as cases post corneal surgery. Recently, Matsuda et al (22) showed that the CCT measurements with Pentacam compared to Orbscan were statistically stable in postoperative corneas and corneas with haze. They hypothesized that with Orbscan, the camera, which faces the cornea, catches scattered reflected light beams as it moves to the right and left. As the angle of illumination changes, the distance at which the instrument captures images changes. As a result, the perfect focal distance cannot be achieved for both the anterior and posterior cornea. Corneal haze causes additional distortion and lowers the brightness of the posterior edge of the cornea, resulting in a lower quality posterior surface edge. In comparison, the Scheimpflug camera of the Pentacam catches rotating, scattered reflected light beams which are set at the front of the cornea. The camera itself rotates about an imaginary line through the center of the cornea, keeping a stable angle with the central corneal axis, thereby fulfilling the Scheimpflug principle. The Scheimpflug camera system provides the advantages of a higher depth of focus and a sharp, undistorted picture. Thus, Pentacam is more accurate at measuring CCT compared to Orbscan. In conclusion, the results from our study are based on a reasonably large sample of post- LASIK eyes in which CCT was measured using Pentacam (a noncontact optical pachymeter) and UP. Although both methods showed very strong correlation, Pentacam significantly overestimated CCT compared to UP. Considering the systematic differences (different technological methods and operating techniques) between both tools that resulted in different values, we believe that CCT measurements cannot simply be substituted between both modalities and Pentacam cannot be used interchangeably with UP in post-lasik eyes for myopia. ACKNOWLEDGEMENTS The authors thank Howaida Bakili, Bassam Abbara, and Renelyn Quizon for help with data processing. The authors report no proprietary interest or financial support. Address for correspondence: Hani S. Al-Mezaine, MD Department of Ophthalmology College of Medicine King Saud University P.O. Box 230387 Riyadh 11321 Kingdom of Saudi Arabia almez2001@hotmail.com REFERENCES 1. Murphy GE, Murphy CG. Comparison of efficacy of longest, average, and shortest axial length measurements with a solid-tip ultrasound probe in predicting intraocular lens power. J Cataract Refract Surg 1993; 19: 644-5. 2. Marsich MW, Bullimore MA. The repeatability of corneal thickness measures. Cornea 2000; 19: 792-5. 3. Solomon OD. Corneal indentation during ultrasonic pachymetry. Cornea 1999; 18: 214-5. 4. Holmén JB, Ekesten B, Lundgren B. Anterior chamber depth estimation by Scheimpflug photography. Acta Ophthalmol Scand 2001; 79: 576-9. 5. Al-Mezaine HS, Al-Amro SA, Kangave D, et al. Comparison between central corneal thickness measurements by oculus Pentacam and ultrasonic pachymetry. Int Ophthalmol 2008; 28: 333-8. 6. O Donnell C, Maldonado-Codina C. Agreement and repeatability of central thickness measurement in normal corneas using ultrasound pachymetry and the OCULUS Pentacam. Cornea 2005; 24: 920-4. 7. Lackner B, Schmidinger G, Pieh S, et al. Repeatability and reproducibility of central corneal thickness measurement with Pentacam, Orbscan, and ultrasound. Optom Vis Sci 2005; 82: 892-9. 8. Barkana Y, Gerber Y, Elbaz U, et al. Central corneal thick- 856

Al-Mezaine et al ness measurement with the Pentacam Scheimpflug system, optical low-coherence reflectometry pachymeter, and ultrasound pachymetry. J Cataract Refract Surg 2005; 31: 1729-35. 9. Buehl W, Stojanac D, Sacu S, et al. Comparison of three methods of measuring corneal thickness and anterior chamber depth. Am J Ophthalmol 2006; 141: 7-12. 10. Ucakhan OO, Ozkan M, Kanpolat A. Corneal thickness measurements in normal and keratoconic eyes: Pentacam comprehensive eye scanner versus noncontact specular microscopy and ultrasound pachymetry. J Cataract Refract Surg 2006; 32: 970-7. 11. Fujioka M, Nakamura M, Tatsumi Y, et al. Comparison of Pentacam Scheimpflug camera with ultrasound pachymetry and noncontact specular microscopy in measuring central corneal thickness. Curr Eye Res 2007; 32: 89-94. 12. Ho T, Cheng AC, Rao SK, et al. Central corneal thickness measurements using Orbscan II, Visante, ultrasound, and Pentacam pachymetry after laser in situ keratomileusis for myopia. J Cataract Refract Surg 2007; 33: 1177-82. 13. Ciolino JB, Khachikian SS, Belin MW. Comparison of corneal thickness measurements by ultrasound and Scheimpflug photography in eyes that have undergone laser in situ keratomileusis. Am J Ophthalmol 2008; 145: 75-80. 14. Hashemi H, Mehravaran S. Central corneal thickness measurement with Pentacam, Orbscan II, and ultrasound de- vices before and after laser refractive surgery for myopia. J Cataract Refract Surg 2007; 33: 1701-7. 15. Prisant O, Calderon N, Chastang P, et al. Reliability of pachymetric measurements using Orbscan after excimer refractive surgery. Ophthalmology 2003; 110: 511-5. 16. Tam ES, Rootman DS. Comparison of central corneal thickness measurements by specular microscopy, ultrasound pachymetry, and ultrasound biomicroscopy. J Cataract Refract Surg 2003; 29: 1179-84. 17. Nissen J, Hjortdal JØ, Ehlers N, et al. A clinical comparison of optical and ultrasonic pachymetry. Acta Ophthalmol (Copenh) 1991; 69: 659-63. 18. Herse P, Siu A. Short-term effects of proparacaine on human corneal thickness. Acta Ophthalmol (Copenh) 1992; 70: 740-4. 19. Asensio I, Rahhal SM, Alonso L, et al. Corneal thickness values before and after oxybuprocaine 0.4% eye drops. Cornea 2003; 22: 527-32. 20. Nam SM, Lee HK, Kim EK, Seo KY. Comparison of corneal thickness after the instillation of topical anesthetics: proparacaine versus oxybuprocaine. Cornea 2006; 25: 51-4. 21. Gordon A, Boggess EA, Molinari JF. Variability of ultrasonic pachymetry. Optom Vis Sci 1990; 67: 162-5. 22. Matsuda J, Hieda O, Kinoshita S. Comparison of central corneal thickness measurements by Orbscan II and Pentacam after corneal refractive surgery. Jpn J Ophthalmol 2008; 52: 245-9. 857

Copyright of European Journal of Ophthalmology is the property of Wichtig Editore and its content may not be copied or emailed to multiple sites or posted to a listserv without the copyright holder's express written permission. However, users may print, download, or email articles for individual use.