Air Removal from the Sterilization Chamber



Similar documents
or

Steam Sterilization Cycles for Lab Applications

Biological Indicators and

Steam Sterilization and the 2007 Revision of PDA Technical Report 1

3 Attest Biological Monitoring System

AUTOCLAVE STEAM STERILIZERS

Comparison Study of the Growth Promotion Capabilities of Self Contained Biological Indicator Culture Medium in Extended Steam Sterilization Cycles

Because You Can t See Sterile.SM

Steam Sterilization Principles

Lab Exercise 3: Media, incubation, and aseptic technique

Sterilization of narrow channels: how to ensure effectiveness in daily routine? Dr.ir. Josephus Paulus Clemens Maria Van Doornmalen Gomez Hoyos

Laboratory Line. Vertical and Bench-Top Life Science Autoclaves

Medical Market Laboratory EMEA

Infection Control in the Dental Practice through Proper Sterilization

FACT SHEET : Using Autoclaves Safely

Application Note USTR 805. Steam Sterilization of Pall Filter Assemblies Utilizing Replaceable Filter Cartridges

THE BASICS OF STERILIZATION

Vertical and Benchtop Autoclaves for Life Sciences. Laboratory Line

STANDARD OPERATING PROCEDURES DIVISION OF COMPARATIVE MEDICINE UNIVERSITY OF SOUTH FLORIDA

Sterile Cleanroom Management

Sterilization A REVIEW OF THE BASICS. By Gayla Tilton and Myrna Kauffman 66 MANAGING INFECTION CONTROL

Medical Market Laboratory EMEA

Fact Sheet on Steam Sterilizers at Stanford University

Federal Wage System Job Grading Standard For Laboratory Working, Table of Contents

Penetration tests according to the ISO 11140

Calibration of Volumetric Glassware

Top 10 considerations when validating an autoclave

STEAM HEATING SYSTEM TROUBLESHOOTING GUIDE

1/22/2012. When is it a good time to learn about sterile processing in your facility? Continuing Education Contact Hours. Learning Objectives

STANDARD OPERATING PROCEDURES Safe Autoclave Operations

GUIDE TO AUTOCLAVING PIONEERS IN STERILIZATION TECHNOLOGY

Effective Use of Autoclaves

The Molar Mass of a Gas

KGI MEDICAL WASTE MANAGEMENT PLAN

EH&S. Sheet. Fact. Safe and Effective Use of Autoclaves. What are autoclaves? Factors for effective sterilization. Dry heat cycle - when to use

Effective Heat Sterilization in CO 2 Incubators

Sterilizers with extra class

Validation Guide. USTR 805j (5) Steam sterilization of Pall filter assemblies which utilize replaceable filter cartridges

Decontamination and Waste Management

The Supplier Perspective: Sterilization. Presented by: Thad Wroblewski STERIS Isomedix Services

TECHNIQUES FOR STEAM STERILIZING LABORATORY WASTE

ArtisanLink Staining System is an automated special stains slide

PORTABLE LABORATORY MODEL AVAILABLE!

Revised EHS Biosafety. 1 Select appropriate containers/bags for autoclaving.

OPERATOR MANUAL Amsco Century TM Medium Steam Sterilizer 26" x 26" (660 x 660 mm) Scientific. (12/16/05) P Rev. 0

TUTTNAUER OLD AUTOCLAVE OPERATING PROCEDURE

Sterilization methods and equipment Lab 1-2

BESTPRACTICE. HTM01-05 Q&A: You can now follow us on Twitter! We have launched our HTM01-05 questions and answers service on Twitter.

Bench Autoclave. Standard Operating Procedure. For Installation, Use and Maintenance

Oil Sight Glass BENEFITS

Model Voltage (V/Hz) Power (W) Capacity (L) Chamber size Outside size(mm) Weight (kg) STE-TEN 220/ Ø90x160mm W200xH385xD (Net:28)

AUTOCLAVES. Table Top. STE-TEN, 1 Liter Autoclave

DISINFECTION & STERILIZATION INFECTION CONTROL GUIDELINES SECTION 4 QUALITY MANAGEMENT IN STERILIZIATION

AUTOCLAVE PROGRAM. SOP Bio-006 FOR THE USE OF AUTOCLAVE FOR STERILIZATION OF MATERIALS AND BIOLOGICAL WASTE SOP

Partnership. Large Steam Autoclaves for the Life Sciences. Laboratory Line

Water Baths. with BioCote antimicrobial protection

STERILIZATION AND DISINFECTION

Laboratory & Scientific Line. Bulk Steam Sterilizers. Since

Assembly and Installation Procedures

Horizontal Sterilizer MU 012.B / I (09/07) Page 2/46. 1 Index... 2

Heat Transfer. Condensation. Revaporization

Managing Steam Sterilization Process By Patricia H. Nothum, RN, BSN, CRCST, CHL

Sterile E Online NetworkTutorials

LABOKLAV. Autoclaves Made in Germany.

Chemical Indicators & Accessories from Albert Browne International Ltd. Assurance through Technology

Model Voltage (V/Hz) Power (W) Capacity (L) Chamber size Outside size(mm) Weight (kg) STE-TEN 220/ Ø90x160mm W200xH385xD (Net:28)

The Power of Innovation. For better Sterilization. Pass-through autoclaves: Systec D-Series 2D Horizontal front loading bench-top autoclaves: Systec

Steripro cgmp. Steam Sterilizers

Experiment 3: Extraction: Separation of an Acidic, a Basic and a Neutral Substance

Thermo Scientific VARIOKLAV Modular Laboratory Autoclaves Versatile, space-saving performance for steam sterilization applications

Vacuum Tube Solar Water heater Non-pressure. User manual

2. Laboratory equipment

OPERATOR MANUAL Lab 250 and Lab 110 Life Sciences Small Sterilizers. (11/13/06) P Rev. 0

Autoclave Safety. Autoclaves are sterilizers using high pressure and high temperature steam. The potential safety risks for the operators are:

Total Heat Versus Sensible Heat Evaporator Selection Methods & Application

Microbiological Evaluation of the STI Series 2000 Medical Waste Treatment Process

FVA THIRD-GENERATION VERTICAL SATURATED-STEAM STERILIZERS

A GUIDELINE FOR THE SAFE USE OF AUTOCLAVES UNIVERSITY OF OTTAWA. Environmental Health and Safety Service

Laboratory- and sterilisation equipment HMC EUROPE

Laboratory- and sterilisation equipment HMC EUROPE

Autoclave Testing Service, Inc.

OPERATION OF THE AUTOCLAVES

the stainless steel water-heater for your woodstove USER S MANUAL

Laboratory & Scientific Line. Horizontal Steam Sterilizers

Autoclave Quality Control Policy

TMI-Orion CERAMICS TILES BRICKS VALUE PROPOSITION HIGH TECHNOLOGY DATA LOGGING SOLUTIONS.

Pure Solid Compounds: Molecules held together in rigid formations by intermolecular forces.

#002. Flash Sterilization. S t u d y G u i d e. The Hot Issues of. A Continuing Education Study Guide presented by

Fundamental Autoclave Techniques

IDENTIFICATION BROCHURE FOR Central RESIDENTIAL WET SPRINKLERS

Standard laryngoscope blade assemblies Directions for use

Subpart 1. Installation. All plumbing systems must be. installed and tested according to this chapter and chapter 4715,

STANDARD OPERATING PROCEDURE

STEAM STERILIZERS. DLOV & DLOM series. more than fifty years of experience in sterilization. DE LAMA s Quality Management System is certified:

Guidance on safe use of Autoclaves

BSL3 and BSL4 Autoclaves. Laboratory Line

Medical Waste Autoclaves. Medical Waste Autoclaves

Refrigeration Installation 101

PAPER CHROMATOGRAPHY

Transcription:

TM Volume 5, Number 2 March 2008 Garrett Krushefski Scientific & Technical Services Manager Air Removal from the Sterilization Chamber I recently received a question from a customer asking why we state in the EZTest instructions for use to Place the EZTest biological indicators in a horizontal position When sterilizing by steam, it is imperative to first remove ambient air not only from the chamber but also from the devices being sterilized. Failure to do so will impede steam contact with all surfaces which can result in non-sterile items being used from a seemingly acceptable sterilization cycle. There are two general means for removing ambient air from the sterilization chamber: dynamic air removal or gravity-displacement. Dynamic-air-removal sterilization cycles can be either a prevacuum cycle, in which case air is actively removed by a series of pressure and vacuum phases prior to admission of steam, or by steam-flush-pressure pulsing (SFPP) in which case a series of steam flushes and pulses of pressure are employed. Gravity-displacement is a passive means of air removal in which incoming steam displaces or pushes residual air out through the drain of the chamber. ANSI/AAMI ST79:2006 notes, The dynamic-air-removal cycle is generally preferred to a gravity-displacement cycle because of more efficient air removal, a shorter exposure time at higher temperatures, and a vacuum drying phase, resulting in an overall reduction in cycle time. While the above ANSI/AAMI note is accurate, do not think that you must start upgrading and replacing your gravity-displacement sterilizers with new machinery. Gravity displacement is still an effective means of sterilization as long as one employs a few precautions. Such precautions are not necessary in a dynamic-air-removal process because the air is actively sucked-out of the chamber and load. Without the aid of active air removal, one needs to facilitate removal of air from the load when using gravity-displacement systems. This is achieved by positioning items in such a manner that does not create any potential air traps. To illustrate this point, start

with a simple example such as a standard laboratory beaker. Of the three orientations shown in Figure 1, which are appropriate for a gravity-displacement steam sterilizer? Figure 1 A C B NOTE: In all of the photos non-packaged items are shown for visual clarity. In actual practice these items would be in sterilization pouches, allowing for post-exposure protection from contamination. The answer is (B & C). The ambient air that is contained within the beaker is heavier (i.e. more dense) than the steam that will be introduced into the chamber. In the upright orientation (A), the steam will not be able to displace or push the ambient air out of the beaker because the air has no means of escape (i.e. the beaker does not have a drain ). Note that the orientation shown for (B) does allow the air to drain from the beaker. The next example is an Erlenmeyer flask which is a bit trickier. Review the four orientations shown in Figure 2 and decide which are the most appropriate for a gravitydisplacement cycle.

Figure 2 D G E F It is easy to see that (F) or (G) is the best option and (D) must be avoided. But what about (E)? This orientation is also problematic because there is a zone within the Erlenmeyer flask where ambient air will become trapped and steam contact to the surfaces in that zone will be impeded (see outline of dashed red zone in Figure 3). A very effective way to demonstrate the effects of trapped air is to run a standard sterilization cycle with three flasks oriented as shown in D, E and F (see Figure 3 below). Place an EZTest biological indicator at each location marked in the photo and incubate the units after exposure. The EZTest from locations 1 and 2 will show surviving spores while the EZTest from location 3 will have no growth. This is a perfect example to show that the parametric data for the cycle (time, temperature, and pressure readings) will indicate acceptable cycle performance but the biological indicator indicates non-sterility.

Figure 3 D E F 3 1 2 When conducting a validation several years ago, I performed the above recommended test in a 65-minute exposure. The penetration thermocouples (with attached EZTest BIs) recorded accumulated Fo values in excess of 80 minutes even from locations 1 and 2. The physical data indicated a lethal delivery that was well in excess of what was needed to kill the spore challenge, yet the EZTest from locations 1 and 2 were positive while location 3 was negative. How could the spores survive 80 Fo? The reason is the prevailing condition at locations 1 and 2 was NOT saturated steam; rather, it was 121-123 C dry heat. Without an escape route, the less-dense steam molecules could not displace the heavier ambient air and air pockets (dashed red zones in Figure 3) become trapped in the flasks. The penetration thermocouples recorded temperatures of 121 to 123 C for most of the 65-minute dwell time, but unlike the spores in the biological indicator, the thermocouples can not differentiate between dry heat conditions and saturated steam. When reviewing the cycle data, time, temperature and pressure were all within acceptable levels, thus indicating acceptable cycle performance, yet the spores survived. The biological results indicated non-sterility and did not agree with the physical data which indicated sterilization was achieved. We refer to this condition as a catastrophic sterilization failure. SPORES DON T LIE! Air pockets within the materials being sterilized can not be detected by any instrumentation available today. The only system available to detect these pockets is the biological indicator. Air pockets occur in the worst case to sterilize locations in the

product or load. All standards direct the user to challenge the worst case locations with biological indicators. As a final example, consider the lengths of tubing shown in Figure 4. When oriented as shown in (G) multiple air pockets will form at various low spots along the length of the tube. One must carefully coil the tube in a helix shape (H) to facilitate air removal and steam penetration. Figure 4 G H Using colored water, one can illustrate the multiple low spots that exist in (G) where air pockets will form. For Figure 5, both tubes were filled with purple water and placed on the perforated stainless tray. All of the colored water drained from (H). Compare that to the amount and location of water that was unable to escape the (G) configuration. The colored water allows one to visually discern where the trapped air will form pockets and thus, the areas in the tube where steam will not penetrate.

Figure 5 G H The water visualization method is a valuable tool that can be used when training personnel to use steam sterilizers. One must position items in the chamber such that if they were filled with water, the water would completely drain from the item. If any water were retained, the positioning is not suitable (as shown for the Erlenmeyer flask in Figure 6).

Figure 6 The colored water is unable to drain from the flask and shows where trapped air will collect and impede steam contact. Now back to the original question, Why is it necessary to place the EZTest biological indicators in a horizontal position? The EZTest, by virtue of its design is an excellent detector of trapped air. The EZTest itself is not unlike the Erlenmeyer flask in that it too can harbor ambient air which would impede steam contact with the spores on the inoculated paper strip. Thus, just like the beaker shown in (B), this is the correct orientation for the EZTest BI. Is it necessary to use the horizontal orientation in a dynamic-air-removal sterilization cycle? In dynamic-air-removal cycles, one is relying on the prevacuum or SFPP to remove air from the chamber, the load and the EZTest (which may or may not be in a test pack). Regardless, it is always a good idea to use the horizontal orientation for your EZTest as it does help avoid inadvertent trapping of air, even in dynamic-air-removal systems.

Finally, consider Figure 7; can you see the flaw in (I) and why (J) is the preferred set-up? I J Figure 7 Recall the earlier statement, the EZTest itself is not unlike the Erlenmeyer flask in that it too can harbor ambient air which would impede steam contact with the spores on the inoculated paper strip. In both (I & J) above, the Erlenmeyer flask is inverted to the extent that if filled with water, all of the water would drain from the flask. HOWEVER, in (I) the EZTest is in a slight cap-up orientation and if the EZTest biological indicator were filled with water, it would be trapped inside the EZTest. In (J) a slight cap-down orientation prevails (by virtue of the inverted positioning of the Erlenmeyer flask) and thus we have a drain which allows the ambient air from inside the EZTest to escape the unit. SPORES DON T LIE Please email us with topics you would like to see addressed in Spore News.