GUIDE TO AUTOCLAVING PIONEERS IN STERILIZATION TECHNOLOGY

Size: px
Start display at page:

Download "GUIDE TO AUTOCLAVING PIONEERS IN STERILIZATION TECHNOLOGY"

Transcription

1 GUIDE TO AUTOCLAVING PIONEERS IN STERILIZATION TECHNOLOGY

2

3 What is Sterilization? Sterilization may be defined as the statistically complete destruction of all microorganisms including the most resistant bacteria and spores. This is a condition that is difficult to achieve and hard to prove. Whilst there are many chemicals, inorganic and organic, that kill microorganisms they may not be totally effective and can leave undesirable or toxic residues. Ultraviolet and Ionising radiations are also effective biocides, disrupting or modifying the DNA to prevent replication, but Ultraviolet will not produce the effective results and easy validation that moist heat (steam) sterilization can provide. If sterility is an absolute requirement then today s scientists turn, as their predecessors did, to steam. Microorganisms tend to become more active as the temperature of their surroundings rises, - most, but not all, die at above 80 o C. In the case of Prions the temperature and time requirements for deactivation are much higher. Steam molecules condense on cooler microorganisms, and transfer 2500 joules per gram of steam, very efficiently heating the microorganisms to the temperature at which they are destroyed. Other methods of heating suffer from the much lower heat transfer of hot dry gases and boundary layer effects, which can insulate and protect the microorganisms. For maximum effect the Steam must be saturated, and this condition, and the temperature and pressure of the steam are easily monitored, facilitating proof that sterilization has occurred. By employing Steam Sterilization techniques a high level of sterility can be achieved and the most popular piece of equipment used in laboratories and hospitals is the STEAM STERILIZER or AUTOCLAVE. How does an Autoclave work? The terms Autoclave and Sterilizer are interchangeable and mean the same thing. Autoclave is a term used mainly in laboratories whilst Sterilizer is used more commonly for medical and pharmaceutical applications. An effective Autoclave or Sterilizer must contain dry saturated steam. In order to achieve this air must be removed from both the load and the chamber; this can be accomplished in a number of ways. Air removal from high mass low load items (i.e. comprising mainly solid areas that contain little or no air pockets such as bottled media) will need little air removal and this can be facilitated by automatic air purging. Here air is allowed to leave the chamber through a vent as steam enters the chamber either from an integral source (upward displacement) or an external supply (downward displacement); this vent only closes when all of the air is removed from the chamber. This method can be further advanced by Freesteaming when the vent is allowed to stay open for a set length of time. The turbulent steam then passes through the vent forcing any trapped air out of the autoclave. For more stubborn loads that contain a number of air pockets such as wrapped instruments or fabrics, a more effective method of air removal is essential. By far the most effective way of air removal is a vacuum system, whereby a vacuum is achieved in the chamber before any introduction of steam, removing most of the air before freesteaming and/or vacuum pulsing. Once all of the air is removed from the load and chamber the temperature within the vessel will rise along with the pressure until the pre-selected temperature is reached. To achieve a typical sterilization temperature of 121 o C or more, requires the steam to be pressurized to at least 1.1 Bar G. Since pressures used are greater than 0.5 Bar G the Autoclave assembly is classified as a pressure system and must be designed to a strict engineering standard such as PED97/23EC/PD5500/ASME etc. The temperature within the chamber is reached in a number of ways: 1) Steam can be injected into the chamber via an internal steam source such as an integral stainless steel steam generator, which can be built within the Autoclave cabinet or can be supplied as a separate unit. 2) Some laboratories or hospitals have their own direct steam source, which would similarly be injected, into the chamber. 3) In some Autoclaves heaters are built into the base of the chamber and the water is heated until it boils and produces steam. The main advantage of using a steam generator (or the direct steam method) is that cycle times can be considerably faster as the steam is immediately available. In basic autoclaves the water is topped up manually by pouring water into the vessel whilst some autoclaves are connected to a direct water supply and the water level is maintained automatically. Some of Astell s smaller units are available with an integral water tank, which once filled, can run for up to 20 cycles repeatedly before being replenished. In CSSD applications it is a requirement that the steam is dry saturated steam with a known dryness and noncondensable gas content e.g. in the UK, Steam with values within the specification set out in HTM2010. When the desired temperature is achieved for the required time then the steam supply will cease either by shutting off the power to the heaters or by cutting off the steam supply. Thus the temperature and the pressure will gradually drop. In units with a vacuum system fitted the vacuum pump can be used to remove the steam (i.e where drying is required).

4 Typical Autoclave Cycles Typical temperature/time sterilization parameters are 115 o C for 30 minutes, 121 o C for 15 minutes and 134 o C for 3 minutes. However, overall cycle times will vary depending on load characteristics. Firstly it is important to realise that the time and the amount of steam that is needed for the load to reach Sterilization will vary greatly depending on the thermal capacity of the load in question. Two or Three 1 litre bottles of culture media will often need more steam than a dozen small containers, and the heating time will take longer. Likewise the time taken to achieve Sterilization once the temperature has been attained will depend on the type of microbiological organisms present. Bacteria and fungi are rapidly killed by moist heat but viruses and particular bacterial spores are very persistent, and may need long Sterilization times. The development and wide availability of reliable microprocessors means that most modern autoclaves are fully user programmable, allowing the Sterilization temperature to be set to suit the requirements of the load. The most accurate way of ensuring that Sterilization occurs when the correct temperature is reached is by Load Sensed Process Timing. Here the cycle is controlled via the temperature achieved in the centre of the load. A wandering thermocouple probe situated within the chamber is inserted into the load, or a load simulator and initiates the sterilization period once the probe reaches the programmed threshold temperature. This ensures that sterilization starts at the correct stage of the cycle. Following Sterilization and depressurization of the chamber it may be possible to open the autoclave immediately. However, when autoclaving bottled fluids, a cooling lock is normally fitted to comply with various safety regulations (e.g. in the UK: HSE PM73). This prevents the autoclave from being opened until the load temperature has fallen to typically 80 o C, thus reducing the risk of both the boiling over of fluids (e.g media) and the explosion of glass bottles. Types of Cooling Following Sterilization it is possible to have a variety of fast cooling options fitted which reduce the cool down period. Water Cooling Deluge Cooling This can often cut cycle times down by up to 50% and is used only in conjunction with sealed fluid loads. Water at Sterilization temperature is sprayed onto the load after the completion of the sterilization stage. The water being sprayed is passed through a heat exchanger cooled by cold water thus producing a controlled reduction in temperature throughout the load. This is a highly specialised method of cooling, and due to the volume of water used during the deluge stage it is only suitable for sealed bottled fluids. Water Fed Jacket A Jacket can be fitted to the autoclave chamber on both circular and square section units. Again cold water is introduced into the jacket at the end of the sterilization stage of the cycle using the chamber walls as a heat exchanger reducing temperature and resulting in improved cooling. As the jacket acts as a very efficient heat exchanger pressure reduction can be very fast using this method and therefore it is generally recommended that Air Ballast is used in conjunction with the jacket to reduce the possibility of fluid loss. Advanced Water Cooling This can only be fitted to machines that have been connected to a mains water supply. A series of cooling coils are fitted to the outside of the autoclave chamber. At the end of sterilization these coils are circulated with cold water and so results in a rapid decrease of temperature and improved cooling time. Fan Assisted/Air Cooling Assisted Air-Cooling Fans are fitted within the cabinet and circulate ambient air over the external chamber walls during the cooling stage of the cycle creating a controlled improvement to cooling times. Internal Convection Fan Assisted Cooling A fan or fans can be built within the chamber and ducted to create an efficient airflow. The fan is again activated during the cooling stage of the cycle. The fans create turbulence reducing cooling time greatly. Internal convection fan cooling is one of the most proficient ways of cooling the chamber. To ensure improved cooling time the optional jacket and/or advanced water-cooling is also needed. On loads using unsealed bottles air ballast is recommended. The importance of drying! It is important to take into account exactly what the autoclave will be used for before running a cycle. Some items such as fabrics & wrapped instruments require special attention as they will need a specific drying cycle to avoid them being left wet at the end of the cycle. The best way of ensuring that this does not happen is by having a pre and post vacuum as well as a steam heated jacket fitted to the chamber. The pre-vacuum is required to remove air from the load and the post vacuum will remove moisture left on the load. The drying will be enhanced with the addition of the steam heated jacket, which is filled with steam throughout the cycle; this heat prevents excessive condensation being formed on the chamber walls and maintains the temperature above the boiling point during drying so that water droplets boil off. The Heated jacket is essential for optimum drying performance.

5 Uses of Autoclaves Autoclaves can be used for numerous medical and laboratory applications in Market Sectors such as, Food and Dairy, Universities & Colleges, Agriculture & Horticulture Research, Healthcare (including Pathology), Industrial Research, Water & Environmental, Brewing, Biotechnology Industries and Pharmaceuticals Typical applications include: Media Preparation Sterilization of Fluids Instruments (Wrapped and Unwrapped) Utensils and Lab Equipment Fabrics and Textiles Media Preparation: Media is used to grow microorganisms under test conditions and a wide variety of gel and liquid media is used. Air removal from fluid/media loads is normally straightforward; however some media cycles can benefit from a short period of free steaming. Different types of media withstand different sterilizing temperatures, therefore any autoclave used for the Sterilization of media must be capable of sterilizing at a variety of temperatures. Cycle times must be short and the speed of processing is vital. A fast heat up and cool down is beneficial in the maintenance of the quality of the finished product and prevents the over cooking which leads to the deterioration in the quality of the growth medium. Cycle length should be set to as short as possible, and at the temperature specified by the media manufacturer. The cool down period however, needs to be a controlled process in order to prevent boil over and loss of media. The main method of controlling the sterilzation procedure of any cycle is to use a thermocouple probe in a dummy bottle within the chamber (Load Sensed Process Timing). Only when the probe reaches the correct temperature will the sterilization process begin. Load Sensed Process timing is used in conjunction with Profiled Overshoot Boost, this allows the temperature in the chamber to rise above the programmed level to allow the load to reach the correct temperature faster. The risk of any boil over can be greatly reduced by using Air Ballast which maintains chamber pressure during the cooling phase. 3000mbar 2100mbar 2000mbar 1000mbar Ombar START VAC POSITIVE PULSE STERILIZE COOLING DRYING AIR BREAK Sealed Fluids Cycle Graph Sealed Fluids: Some fluid loads may require to be sterilized in special sealed containers. Air purging is normally sufficient to remove air, and extended freesteaming should be unnecessary. As with media preparation the speed of processing is vital to the quality of the end product and often temperatures in excess of 121 o C will be used to speed up the Sterilization process. Again the cooling should be as short as possible. 3000mbar 2000mbar 1000mbar Ombar START VAC POSITIVE PULSE STERILIZE COOLING DRYING AIR BREAK Instruments: The Sterilization of instruments falls into two main categories: a) Unwrapped Instruments b) Wrapped Instruments Both types of instrument loads benefit from drying the end of the cycle. Unwrapped instruments can be dried by the heating of the surrounding air at the end of the cycle, whilst wrapped instruments require a more sophisticated drying system Post vacuum systems on small bench-top units will improve drying, whilst a jacketed pressure vessel is strongly recommended on larger units. 3000mbar 2000mbar 1000mbar Ombar Sealed Fluids Cycle Graph (including assisted cooling) START VAC NEGATIVE PULSE POSITIVE PULSE STERILIZE DRYING AIR BREAK Typical Glassware/Utensils (134ºC) Cycle Graph

6 Fabrics and Textiles: A variety of fabrics and textiles can be sterilized in an autoclave/sterilizer, e.g many hospitals will use an autoclave to sterilise gowns, instrument packs etc. As these are true porous loads they will require both pre and post vacuum systems, negative and positive steam pulsing and a fully heated jacket. Steam quality is important when operating Porous Load Autoclaves. Load Vacuum Assisted Sensed Vacuum & Assisted Cooling Process Pulsing Heated Cooling & Air Timing Jacket Ballast 3000mbar VAC NEGATIVE PULSE POSITIVE PULSE STERILIZE DRYING AIR BREAK Media x x ( 3) 2000mbar 1000mbar Ombar START Mixed Discard x x x Typical Fabrics Cycle Graph Discard Loads: There is an ever-increasing need for the safe Sterilization of contaminated material before disposal to ensure that it is microbiologically safe. The contaminated material can be of many types and in many forms. For example, all microbiological laboratory waste should be processed prior to disposal, as must all clinical samples e.g. blood, tissue etc. Loads consisting of such a wide variety of materials in a broad range of containers will present specific Sterilization problems. One of the main problems with such mixed discard loads is that they invariably contain small pockets of trapped air, which present particular air removal problems before steam penetration. Such loads are often contained in special plastic bags and comprise a variety of plastic containers; these are liable to melt and collapse before Sterilization, thus increasing the likelihood of trapped air. A pre-vacuum and steam pulsing feature is now a commonly requested option for autoclaves running frequent discard cycles. However an alternative to vacuum is a long period of freesteaming or Pulsed Freesteaming PAT. There are special discard (Morrison) containers available for this application. In certain circumstances a doubledoored autoclave may be required. The loading ( dirty or nonsterile) end is often separated from the unloading ( clean or sterile) end by a wall, which often contains an SPF seal to prevent bacteriological migration. In category III laboratories an effluent retention system would also be necessary. Fluid Discard Glassware Unwrapped Instruments Wrapped Instruments x x x x x x x ( 1) x x x ( 1) x x 3000mbar VAC POSITIVE PULSE STERILIZE DRYING AIR BREAK COOLING 2000mbar Fabrics x ( 2) x x 1000mbar Ombar START Typical Fluid Discard Cycle Graph Essential Recommended x Not Needed (1) If dry load is required (2) Vacuum & Heated Jacket required to ensure drying of Fabric Loads (3) To prevent excessive cycle time

7 Monitoring the Sterilization process Verification has become increasingly important, as there is often a need to trace back through a series of results to ensure that a particular Sterilzation process has been carried out. There are a variety of options that are available for this; these can either be mounted externally i.e on an adjacent wall or shelf, or can be an integral part of the autoclave. Printers, Chart Recorders or Dataloggers are the preferred options. These instruments can provide either a hard copy record of exactly what the microprocessor is registering, or they can be completely independent from the microprocessor and record independent results of chamber and load conditions. Data Printers Integral Data Printer Built as an integral part of an autoclave this is the most popular and cost effective way of monitoring the cycle. When the cycle begins the printer records the starting time, cycle number, program number (or name) and machine type. The printer will print a complete record of cycle conditions, freesteam, sterilizing start and finish time and pressure readings at one-minute intervals. There is also provision for operator signature and batch number. Secureprint Data Printer This is a Freestanding Data- Printer independent of the microprocessor, which is designed to give an independent printer verification of the sterilizing function. Chamber temperature, pressure, load and time is displayed and printed with a dual speed papersave feature. (This printer will also operate with any non Astell Autoclave) Chart recorders Chart recorders are the traditional method of printing the Sterilization process. These show a graphical display of pressure and/or temperature come with either one, two or three pens, single pen recorders register chamber temperature, two pen recorders register chamber temperature and load temperature and the three pen recorders register chamber temperature and load temperature via two wandering probes, options also include measurement and display of pressure. The recorders can either come in strip or circular format. Circular Chart Recorders These are high performance microprocessor based recorders with an alphanumeric display to provide more information and easier operation. The recorders are available with up to 1, 2 or 3 pens. Reading accuracy is 1% of Span of the instrument. The programme system allows full control of ranging, plus control of the pen drive, lift and referencing, and display. The tactile membrane switches are mounted at the front, alongside the display, behind the door for security. Strip Chart Recorders These are microprocessor controlled strip chart recorders available with 1,2 or 3 coloured pens. Recording accuracy is 1% of span through the use of accurate stepper motors. The recorder incorporates a high definition alphanumeric display, indicating temperatures and pressure. Astell AutoSCRIBE Datalogger Typical printout from integral or Secureprint data printer. The Astell datalogger is a purpose designed, wall mounted datalogger with LCD display and keyboard. The datalogger is specially configured to record temperature and time parameters. The configuration of the AutoSCRIBE allows maximum storage by only recording during the operation cycle. The data is downloaded on demand to a laboratory P.C. for analysis and archiving.

8 Installation The satisfactory performance of any Autoclave requires that it is installed in accordance with the instructions supplied by the manufacturer. Some small benchtop autoclaves require little or no installation as they have an integral recirculation water tank; therefore the machine requires no plumbing to a mains water supply or drain. Where connection is required to the various services (i.e water, drain, power etc) Astell supplies full installation instructions and diagrams; an example of which is shown. (2) (4) 1) Machine Power Switch 2) Power Supply= Switched FUSED connection unit 3) Chamber Exhaust carrying steam & condensate 4) Vent to external atmosphere or extraction hood 5) Connection to Drain- 35mm Copper. Must be connected to drain with sealed plumbing connection no gaps for steam to escape 6) Building Drain This must flow freely under gravity away from machine;-minimum diameter 50mm. At no point must this slope uphill and there must be no dead legs, loops or u-bends. Between machine and the main drain system. 7) Chamber Drain for cleaning only. May be plumbed to Building drain if required. (3) (5) (1) (6) (7) Typical installation (Electrically heated model) Validation It has become increasingly important to comply with Quality Assurance procedures, which require validation that Sterilization has been achieved. Validation is best defined as the documented procedure of obtaining, recording and interpreting results to ensure that the machine will consistently carry out a defined process. This will involve carrying out multi-point temperature mapping to establish suitable loading configurations and cycle parameters to ensure that the process is controlled within predetermined limits. Astell Scientific is a UKAS accredited company. For further details contact our Service Department. TEMPERATURE/PRESSURE CORRELATION The table below shows Pressure and Temperature correlation for Saturated Steam, and the Maximum/Minimum acceptable Temperature display Values for a Correlation Accuracy of +/- 2%. It applies to ASTELL Sterilizers only. Pressure Temperature Temperature - Tolerance Bar Deg C Min Temp Max temp

9 Astell - Pioneers in Sterilization Technology Astell Scientific have been pioneers in the manufacture and validation of Scientific and Laboratory equipment for over 80 years. In the last decade the company has concentrated on becoming market leaders in Sterilization technology and is now one of the premier manufacturers in this field in Europe. A wide range of standard front and top loading microprocessor controlled Autoclaves are offered including circular section units from Litres and Square Section Sterilizers from litres. In recent years Astell has successfully developed its bespoke manufacturing facility and offers a wide range of non standard units in capacities up to 1000 litres. All of the units feature the secure micro-computer controller, which is user friendly and fully programmable. A variety of options and accessories can be supplied with all models to ensure optimum performance and user flexibility in the most demanding situations. These range from Data Printers, Load Sensed Process timing and a variety of cooling options to reduce cycle time. The Astell range can be catogorised as follows: NB. For further information on Astell s Autoclaves and Sterilizers please refer to our separate brochure: The Autoclave Range or visit our Website at Swiftlock Compact Autoclaves, 23, 40 & 60 Litres This range of circular section autoclaves from Astell offer guaranteed precision and versatility for a wide range of applications. Featuring many of the benefits associated with Astell s larger units these stylish yet economical models meet the demanding requirements of current Sterilization procedures. These are now available in the Classic and Autofill versions. Secure micro-computer control 316L stainless steel pressure vessels Temperature range C ( Bar) Super-quick New Swiftlock door mechanism Compact ergonomic designs Automatic cycle control Built in drying system (Autofill version) Integral reservoir and/or condensate bottle 10 Year Pressure Vessel Integrity Warranty Media holdwarm feature (Classic version)

10 Autoclaves, Swiftlock, 90 to 330 Litres A range of top loading and front loading microprocessor controlled autoclaves in volumes from 90 to 330 Litres. The easy open/close Swiftlock closure combines with a user-friendly control system to ensure simple operation for all applications, whilst a comprehensive range of safety features ensures peace of mind for the user. Cost Effective Full Heated Jacket Available for Optimum Drying Secure Microcomputer control PIN restricted user programming Single action closure Holdwarm for Media Preparation Timed freesteaming 316L Stainless Steel Vessels Cooling locks for the safe Sterilization of liquids Single or multi-program models available Temperature Range o C ( Bar) Optional Vacuum/AirBallast. 10 Year Pressure Vessel Integrity Warranty Autoclaves, Square Section (including models with Vacuum & Air Ballast) A comprehensive range of Square and Rectangular Section front-loading autoclaves, in standard volumes from 100 to 490 litres Maximum Space Utilisation Easy Loading PIN restricted user programming Fully interlocked sliding door. 316L Stainless Steel Vessels Temperature range o C ( Bar) Standard sectional jacket with water cooling. Optional Vacuum & Air Ballast Double door units available 10 Year Pressure Vessel Integrity Warranty The modular internal chamber dimensions of these units ensure maximum space utilisation for all Sterilization purposes. The models with vacuum facility are especially suitable for difficult discard loads, ensuring efficient air removal prior to Sterilization. All units now feature the New LogiColor Colour Touch Screen Controller. For further information on Astell s Autoclaves and Sterilizers please refer to our separate brochure: The Autoclave Range or visit our Website at Continuing efforts to improve products may necessitate alterations to specifications without notice.

11

12 Astell Scientific Limited, Powerscroft Road, Sidcup, Kent DA14 5DT United Kingdom T +44 (0) F +44 (0) E [email protected] Direct Sales: T +44 (0) /3/2 F +44 (0) Certificate No: Q536

Laboratory Line. Vertical and Bench-Top Life Science Autoclaves

Laboratory Line. Vertical and Bench-Top Life Science Autoclaves Laboratory Line Vertical and Bench-Top Life Science Autoclaves The Advanced Laboratory Line The Full Spectrum of Laboratory Sterilizers Sterilization in a laboratory environment has its unique requirements.

More information

Steam Sterilization Cycles for Lab Applications

Steam Sterilization Cycles for Lab Applications Steam Sterilization Cycles for Lab Applications Presented by Gary Butler STERIS Life Sciences August 2009 Early Steam Sterilizers Koch Upright Sterilizer First Pressurized Sterilizer First built by Pasteur-Chamberlain

More information

CONTENTS. Swiftlock Compact Units 23-60 litres Including Benchtop and Top Loading Models. Pages 1-2

CONTENTS. Swiftlock Compact Units 23-60 litres Including Benchtop and Top Loading Models. Pages 1-2 CONTENTS Swiftlock Compact Units 23-60 litres Including Benchtop and Top Loading Models Pages 1-2 Swiftlock Top Loading Units 90-130 litres For all standard laboratory applications Page 3 VACUUM Swiftlock

More information

THE AUTOCLAVE RANGE PIONEERS IN STERILIZATION TECHNOLOGY. Featuring the NEW 'LogiColor' Full Colour Touch Screen Controller

THE AUTOCLAVE RANGE PIONEERS IN STERILIZATION TECHNOLOGY. Featuring the NEW 'LogiColor' Full Colour Touch Screen Controller THE AUTOCLAVE RANGE PIONEERS IN STERILIZATION TECHNOLOGY Featuring the NEW 'LogiColor' Full Colour Touch Screen Controller Contents Pages 1-3 New Swiftlock Compact Autoclaves Pages 4-6 New Swiftlock Autoclaves

More information

Vertical and Benchtop Autoclaves for Life Sciences. Laboratory Line

Vertical and Benchtop Autoclaves for Life Sciences. Laboratory Line Vertical and Benchtop Autoclaves for Life Sciences Laboratory Line Designed for Your Laboratory Applications Advanced Laboratory Autoclaves Tuttnauer laboratory autoclaves have been designed to provide

More information

LABOKLAV. Autoclaves Made in Germany. www.shp-steriltechnik.de

LABOKLAV. Autoclaves Made in Germany. www.shp-steriltechnik.de LABOKLAV Autoclaves Made in Germany www.shp-steriltechnik.de 2 LABOKLAV STEAM STERILIZERS Steam sterilization is the sterilization method with the widest range of applications efficient, safe, reproducible

More information

Sterilization methods and equipment Lab 1-2

Sterilization methods and equipment Lab 1-2 Sterilization methods and equipment Lab 1-2 PHT 434 Sterilization Sterilization a process that by which all viable M.O are removed or destroyed, based on a probability function. Sterilization concept Sterilization

More information

Fundamental Autoclave Techniques

Fundamental Autoclave Techniques Fundamental Autoclave Techniques Version 3, September 2010 Veronika Tatarinoff GSBmE University of NSW Brandy Nelson University of Kentucky The following information is directed at aseptic/sterile surgical,

More information

Touchclave-Lab... the revolutionary new range of square-section laboratory sterilizers from LTE Scientific

Touchclave-Lab... the revolutionary new range of square-section laboratory sterilizers from LTE Scientific Other Products and Services from LTE Scientific LTE Scientific Ltd. LTE Cylindrical Chamber Autoclaves 40 to 380 litres capacity LTE Touchclave Systems Autoclaves from 150 to 10,000 litres capacity Kill

More information

Large Steam Sterilizers for the Life Sciences. Laboratory Line

Large Steam Sterilizers for the Life Sciences. Laboratory Line Large Steam Sterilizers for the Life Sciences Laboratory Line Large Steam Sterilizers for the Life Sciences Sterilization in a laboratory environment has its unique requirements. The types of loads that

More information

Partnership. Large Steam Autoclaves for the Life Sciences. Laboratory Line

Partnership. Large Steam Autoclaves for the Life Sciences. Laboratory Line Partnership Large Steam Autoclaves for the Life Sciences Laboratory Line Large Steam Autoclaves for the Life Sciences Sterilization in a laboratory environment has its unique requirements. Tuttnauer laboratory

More information

AUTOCLAVE STEAM STERILIZERS

AUTOCLAVE STEAM STERILIZERS AUTOCLAVE STEAM STERILIZERS (BENCH TOP) Image: HS-1606VD Hanshin Bench Top Steam Sterilizer Small and easy to operate Three preset sterilization cycles Adjustable cycle presets Rapid heat up times Option

More information

Partnership. Medical Waste Autoclaves. Medical Waste Autoclaves

Partnership. Medical Waste Autoclaves. Medical Waste Autoclaves Partnership Medical Waste Autoclaves Medical Waste Autoclaves Safe, Cost-Effective, Medical Waste Sterilization Keep your environment safe from Medical Waste with high performance Tuttnauer Medical Waste

More information

LTE Scientific Ltd. Touchclave-Lab... the revolutionary range of square-section laboratory sterilizers from LTE Scientific

LTE Scientific Ltd. Touchclave-Lab... the revolutionary range of square-section laboratory sterilizers from LTE Scientific LTE Scientific Ltd. Touchclave-Lab... the revolutionary range of square-section laboratory sterilizers from LTE Scientific About LTE Scientific Contents Touchclave-Lab the revolutionary range of sterilizers

More information

Laboratory- and sterilisation equipment HMC EUROPE

Laboratory- and sterilisation equipment HMC EUROPE Laboratory- and sterilisation equipment HMC EUROPE PROFILE HMC EUROPE Laboratory- and sterilisation equipment Consistent marketing of technologies and the conversion of new ideas create product ranges,

More information

Sterilizers with extra class

Sterilizers with extra class Sterilizers with extra class ClassB Ready for the future The European standard EN 13060 specifies the performance requirements and test methods for small steam sterilizers. These are used for hygiene and

More information

Medical Waste Autoclaves. Medical Waste Autoclaves

Medical Waste Autoclaves. Medical Waste Autoclaves Medical Waste Autoclaves Medical Waste Autoclaves Safe, Cost-Effective, Medical Waste Sterilization Treat Medical Waste with high performance Tuttnauer Medical Waste autoclave systems designed for small,

More information

Guidance on safe use of Autoclaves

Guidance on safe use of Autoclaves Safety Office Reviewed: Dec 2013 Reviewed: DD Month Year Guidance on safe use of Autoclaves 1. Hazards & Uses There are several different types of hazard associated with the use of autoclaves. The main

More information

Thermo Scientific VARIOKLAV Modular Laboratory Autoclaves Versatile, space-saving performance for steam sterilization applications

Thermo Scientific VARIOKLAV Modular Laboratory Autoclaves Versatile, space-saving performance for steam sterilization applications Thermo Scientific VARIOKLAV Modular Laboratory Autoclaves Versatile, space-saving performance for steam sterilization applications Part of Thermo Fisher Scientific Thermo Scientific Varioklav Laboratory

More information

BSL3 and BSL4 Autoclaves. Laboratory Line

BSL3 and BSL4 Autoclaves. Laboratory Line BSL3 and BSL4 Autoclaves Laboratory Line BSL3 and BSL4 Autoclaves In today s society the need for safe bio-containment systems is increasingly important. This is especially true in the sterilization process

More information

Fact Sheet on Steam Sterilizers at Stanford University

Fact Sheet on Steam Sterilizers at Stanford University Fact Sheet on Steam Sterilizers at Stanford University By: Environmental Quality and Water Efficiency Group, Stanford University; December 2013 Jennifer Fitch, Marty Laporte, Julia Nussbaum, Adam Kern,

More information

EH&S. Sheet. Fact. Safe and Effective Use of Autoclaves. What are autoclaves? Factors for effective sterilization. Dry heat cycle - when to use

EH&S. Sheet. Fact. Safe and Effective Use of Autoclaves. What are autoclaves? Factors for effective sterilization. Dry heat cycle - when to use Please post or circulate Fact heet nvironment, ealth and afety Information for the Berkeley Campus No. 33 Revised 04/04/11 afe and ffective Use of Autoclaves Autoclaves are easy to use but can pose a safety

More information

Laboratory & Scientific Line. Bulk Steam Sterilizers. Since

Laboratory & Scientific Line. Bulk Steam Sterilizers. Since Laboratory & Scientific Line Bulk Steam Sterilizers Since 1925 Bulk Steam Sterilizers for Laboratory & Scientific Applications The Tuttnauer line of bulk autoclaves are designed to handle the most critical

More information

Laboratory- and sterilisation equipment HMC EUROPE

Laboratory- and sterilisation equipment HMC EUROPE Laboratory- and sterilisation equipment HMC PROFILE HMC Laboratory- and sterilisation equipment Consistent marketing of technologies and the conversion of new ideas create product ranges, that are tailor-made

More information

Horizontal Steam Sterilizers. Healthcare Line. Since

Horizontal Steam Sterilizers. Healthcare Line. Since Horizontal Steam Sterilizers Healthcare Line Since 1925 Horizontal Steam Sterilizers for Healthcare Applications The Tuttnauer line of healthcare sterilizers are designed for hospital operating rooms,

More information

Laboratory & Scientific Line. Horizontal Steam Sterilizers

Laboratory & Scientific Line. Horizontal Steam Sterilizers Laboratory & Scientific Line Horizontal Steam Sterilizers Horizontal Steam Sterilizers for Laboratory & Scientific Applications Sterilization in a laboratory environment has its unique requirements. The

More information

STEAM STERILIZERS. DLOV & DLOM series. more than fifty years of experience in sterilization. DE LAMA s Quality Management System is certified:

STEAM STERILIZERS. DLOV & DLOM series. more than fifty years of experience in sterilization. DE LAMA s Quality Management System is certified: STEAM STERILIZERS DLOV & DLOM series DE LAMA s Quality Management System is certified: more than fifty years of experience in sterilization UNI EN ISO 9001:2000 - Cert. n 0351 UNI EN ISO 13485 - Cert.

More information

The Power of Innovation. For better Sterilization. Pass-through autoclaves: Systec D-Series 2D Horizontal front loading bench-top autoclaves: Systec

The Power of Innovation. For better Sterilization. Pass-through autoclaves: Systec D-Series 2D Horizontal front loading bench-top autoclaves: Systec The Power of Innovation. For better Sterilization. Pass-through autoclaves: Systec D-Series 2D Horizontal front loading bench-top autoclaves: Systec D-Series t h e a u t o c l a v e c o m p a n y Systec

More information

Model Voltage (V/Hz) Power (W) Capacity (L) Chamber size Outside size(mm) Weight (kg) STE-TEN 220/50 1100 1 Ø90x160mm W200xH385xD450 30 (Net:28)

Model Voltage (V/Hz) Power (W) Capacity (L) Chamber size Outside size(mm) Weight (kg) STE-TEN 220/50 1100 1 Ø90x160mm W200xH385xD450 30 (Net:28) Table Top STE-TEN STE-TEN, 1 Liter Autoclave It adopts B-wclass (pred-13060), with 3-times pre-vacuum preceding vacuum drying. Residual humidity is 0.2, vacuum measurement reaches -0.7bar. It sterilizes

More information

FVA THIRD-GENERATION VERTICAL SATURATED-STEAM STERILIZERS

FVA THIRD-GENERATION VERTICAL SATURATED-STEAM STERILIZERS The key difference between a small vertical or bench-top sterilizer and industrial sterilizer is performance. Small saturated-steam sterilizers with vertical or horizontal chambers were devised many years

More information

Tabletop Line. Automatic and Semi-Automatic Autoclave Series

Tabletop Line. Automatic and Semi-Automatic Autoclave Series Tabletop Line Automatic and Semi-Automatic Autoclave Series Tuttnauer Your Sterilization & Infection Control Partner Company Profile For over 80 years, Tuttnauer s sterilization and infection control products

More information

STANDARD OPERATING PROCEDURES Safe Autoclave Operations

STANDARD OPERATING PROCEDURES Safe Autoclave Operations Department of Environmental Health & Safety Biosafety & Biosecurity STANDARD OPERATING PROCEDURES Safe Autoclave Operations The purpose of this document is to provide standard operating procedures for

More information

Infection Control in the Dental Practice through Proper Sterilization

Infection Control in the Dental Practice through Proper Sterilization Midmark Corporation 60 Vista Drive P.O. Box 286 Versailles, Ohio 45380-0286 937-526-3662 midmark.com Infection Control in the Dental Practice through Proper Sterilization In today's world, the need for

More information

FVA/A1 saturated-steam sterilizer

FVA/A1 saturated-steam sterilizer FVA/A1 where technology innovation and design come together FVA/A1 saturated-steam sterilizer www.fedegariautoklaven.ch Fedegari's new FVA/A1 series of automatic vertical sterilizers has been expressly

More information

Vakulab PL. All-rounder for laboratory and research. Long life time and short cycle times. Space saving and ergonomical

Vakulab PL. All-rounder for laboratory and research. Long life time and short cycle times. Space saving and ergonomical Vakulab PL All-rounder for laboratory and research Long life time and short cycle times Space saving and ergonomical Model diversity and tailored to specific application MMM. Protecting human health. Vakulab

More information

ORGANIC SAMPLE PREPARATION

ORGANIC SAMPLE PREPARATION ORGANIC SAMPLE PREPARATION W W W.LA BT E C H S R L.CO M WSPE MANUAL VACUUM MANIFOLD SPE Process control of the flow rate is critical to guarantee reproducible extractions. Differently then any other systems,

More information

The Power of Innovation. For better Sterilization. Autoclaves: Systec V-Series

The Power of Innovation. For better Sterilization. Autoclaves: Systec V-Series The Power of Innovation. For better Sterilization. Autoclaves: Systec V-Series t h e a u t o c l a v e c o m p a n y Systec V-Series: Vertical autoclaves. These autoclaves don't just seem to be different;

More information

Model Voltage (V/Hz) Power (W) Capacity (L) Chamber size Outside size(mm) Weight (kg) STE-TEN 220/50 1100 1 Ø90x160mm W200xH385xD450 30 (Net:28)

Model Voltage (V/Hz) Power (W) Capacity (L) Chamber size Outside size(mm) Weight (kg) STE-TEN 220/50 1100 1 Ø90x160mm W200xH385xD450 30 (Net:28) Table Top STE-TEN STE-TEN, 1 Liter Autoclave It adopts B-class (pred-13060), with 3-times pre-vacuum preceding vacuum drying. Residual humidity is 0.2, vacuum measurement reaches -0.7bar. It sterilizes

More information

Autoclave Safety. Autoclaves are sterilizers using high pressure and high temperature steam. The potential safety risks for the operators are:

Autoclave Safety. Autoclaves are sterilizers using high pressure and high temperature steam. The potential safety risks for the operators are: Autoclave Safety Purpose: Sterilization refers to the complete killing of all living organisms, including spores. Common sterilization techniques include the application of wet heat, dry heat, chemicals,

More information

Incubator with mechanical convection

Incubator with mechanical convection BF series 400 Incubators and cooling incubators Incubator with mechanical convection The BINDER incubator of the BF series with mechanical convection is suitable for all gentle incubation applications,

More information

Rodwell. Autoclave Range. Rodwell Scientific Instruments: UK Manufactured since 1965. 40-770 litres, vertical & horizontal loading

Rodwell. Autoclave Range. Rodwell Scientific Instruments: UK Manufactured since 1965. 40-770 litres, vertical & horizontal loading Rodwell Rodwell Scientific Instruments: Autoclave Range UK Manufactured since 1965 40-770 litres, vertical & horizontal loading Why specify a Rodwell Autoclave? Make the smart choice Autoclaves are often

More information

Application Note USTR 805. Steam Sterilization of Pall Filter Assemblies Utilizing Replaceable Filter Cartridges

Application Note USTR 805. Steam Sterilization of Pall Filter Assemblies Utilizing Replaceable Filter Cartridges Application Note USTR 805 Steam Sterilization of Pall Filter Assemblies Utilizing Replaceable Filter Cartridges 0 Table of Contents Table of Contents 1. Important Recommendations...1 1.1 Installation

More information

Revised 4-3-2012 EHS Biosafety. 1 Select appropriate containers/bags for autoclaving.

Revised 4-3-2012 EHS Biosafety. 1 Select appropriate containers/bags for autoclaving. Requirements for Decontamination by Autoclaving Summary: Steam sterilization (autoclaving) is one of the most effective methods for decontaminating biohazardous material. Autoclaves use saturated steam

More information

18 LITER VOLUME STERILISATION MATERIALS TESTING MEDIA PREPARATION

18 LITER VOLUME STERILISATION MATERIALS TESTING MEDIA PREPARATION 18 LITER VOLUME STERILISATION MATERIALS TESTING MEDIA PREPARATION Applications THE WIDE RANGE OF USING CERTOCLAV AUTOCLAVES MICROBIOLOGY / BIO-CHEMISTRY Autoclave culture mediums Refuse sterilisation Efficiently

More information

Steripro cgmp. Steam Sterilizers

Steripro cgmp. Steam Sterilizers Steripro cgmp Steam Sterilizers Steripro cgmp Steam Sterilizers BMT worldwide QUALITY, RELIABILITY, INNOVATIO N BMT USA is a member of the MMM Group, one of the world s leading suppliers of sterilization

More information

FACT SHEET : Using Autoclaves Safely

FACT SHEET : Using Autoclaves Safely CSULA Environmental Health and Safety Biosafety Office FACT SHEET : Using Autoclaves Safely Most science research laboratories on campus require the use of autoclaves. The primary purpose of the autoclave

More information

Specifications. Amaro 5000 & Steri 21 MD009/04 EN. A. J. COSTA (IRMÃOS), LDA. Desde 1953 ao Serviço Médico-Hospitalar

Specifications. Amaro 5000 & Steri 21 MD009/04 EN. A. J. COSTA (IRMÃOS), LDA. Desde 1953 ao Serviço Médico-Hospitalar Specifications Amaro 5000 & Steri 21 1 1. GENERAL INFORMATION 1.1. Technical Introduction A. J. Costa (Irmãos), Lda. steam sterilizers are designed to cover a large field of applications for hospitals

More information

Top 10 considerations when validating an autoclave

Top 10 considerations when validating an autoclave Top 10 considerations when validating an autoclave Validating an autoclave is a daunting and time-consuming task. This white paper details the tricks, tips and traps to such a validation project from how

More information

AUTOCLAVES. Table Top. STE-TEN, 1 Liter Autoclave

AUTOCLAVES. Table Top. STE-TEN, 1 Liter Autoclave Table Top STE-TEN, 1 Liter utoclave STE-TEN It adopts B-wclass (pred-13060), with 3-times pre-vacuum preceding vacuum drying. Residual humidity is 0.2, vacuum measurement reaches -0.7bar. It sterilizes

More information

Monitoring the autoclaving process in the pharmaceutical industry

Monitoring the autoclaving process in the pharmaceutical industry Application Description AD/RandC/006-EN Monitoring the autoclaving process in the pharmaceutical industry - Provides independent verification and validation monitoring of the autoclaving process - Enables

More information

Lab Exercise 3: Media, incubation, and aseptic technique

Lab Exercise 3: Media, incubation, and aseptic technique Lab Exercise 3: Media, incubation, and aseptic technique Objectives 1. Compare the different types of media. 2. Describe the different formats of media, plate, tube etc. 3. Explain how to sterilize it,

More information

Energy Efficiency. Bars & RestauRants. www.energia.ie

Energy Efficiency. Bars & RestauRants. www.energia.ie Energy Efficiency Bars & RestauRants www.energia.ie 2 Your chance to reduce your business energy usage by as much as 20%! 20% is a significant figure and reducing your energy bill by this amount could

More information

UNDERSTANDING HOT WATER SYSTEMS IN THE HOME

UNDERSTANDING HOT WATER SYSTEMS IN THE HOME UNDERSTANDING HOT WATER SYSTEMS IN THE HOME This advice guide is part of a series of free guides produced by the Association of Plumbing & Heating Contractors Ltd. which provide consumers with essential

More information

German engineering under high pressure

German engineering under high pressure German engineering under high pressure Over 25 years of experience as a manufacturer of sterilizers, autoclaves and lyophilization systems for laboratories, production and research Production-scale autoclaves

More information

Heat Transfer. Condensation. Revaporization

Heat Transfer. Condensation. Revaporization Objectives Wrapping a Towel Around the Problem of Wet Packs Identify causes of wet packs in steam sterilization Review suggested documentation that can assist in solving wet pack problems Why are Wet Packs

More information

LAKE COUNTY ANIMAL SERVICES EXPANSION SECTION 15459

LAKE COUNTY ANIMAL SERVICES EXPANSION SECTION 15459 SECTION 15459 PART 1 - GENERAL 1.1 RELATED DOCUMENTS: A. Drawings and general provisions of Contract, including General and Supplementary Conditions and Division-1 Specification sections, apply to work

More information

Clinic Washer-Disinfectors. Washer-Disinfector Line

Clinic Washer-Disinfectors. Washer-Disinfector Line Clinic Washer-Disinfectors Washer-Disinfector Line Easy-to-Use Washer-Disinfectors for Clinics The Clinic Range of Washer Disinfectors provides superior cleaning and disinfection combined with innovative

More information

Incubators, cooled incubators, drying ovens, sterilizers, climatic chambers

Incubators, cooled incubators, drying ovens, sterilizers, climatic chambers Incubators, cooled incubators, drying ovens, sterilizers, climatic chambers Incubators, cooled incubators, drying ovens, sterilizers, climatic chambers Laboratory incubators (CL) and drying ovens (SL)

More information

KGI MEDICAL WASTE MANAGEMENT PLAN

KGI MEDICAL WASTE MANAGEMENT PLAN 1revised 2012 I. Facility Information KGI MEDICAL WASTE MANAGEMENT PLAN a. Contact Person: Barbara Erwin Director Research Operations/Chemical and Biological Laboratory Safety/Chair. 909-607-0160 i. Facility:

More information

Heating Water by Direct Steam Injection

Heating Water by Direct Steam Injection Heating Water by Direct Steam Injection Producing hot water by direct steam injection provides a solution where large volumes of hot water at precise temperatures are required, and where energy and space

More information

Steam Sterilization and the 2007 Revision of PDA Technical Report 1

Steam Sterilization and the 2007 Revision of PDA Technical Report 1 Steam Sterilization and the 2007 Revision of PDA Technical Report 1 14 November 2007 Presented By: Mike Finger (Tunnell Consulting) Don Drew (Abbott Bioresearch Center) Agenda: Introduction Sterilization

More information

Professional Certification in Biological Waste Management Examination Content, Sample Questions & References

Professional Certification in Biological Waste Management Examination Content, Sample Questions & References Professional Certification in Biological Waste Management Examination Content, Sample Questions & References The IFBA s Professional Certification (PC) in Biological Waste Management identifies individuals

More information

FILTRATION SOLUTIONS PhARmAceUTIcAL manufacturing FILTRATION SPecIALISTS

FILTRATION SOLUTIONS PhARmAceUTIcAL manufacturing FILTRATION SPecIALISTS FILTRATION SOLUTIONS Pharmaceutical Manufacturing filtration specialists Delivering quality filtration products As one of Europe s leading manufacturers of process filters, Amazon Filters is able to offer

More information

Eurovacuum EV Series. Single Stage Oil Sealed Rotary Vane Pump

Eurovacuum EV Series. Single Stage Oil Sealed Rotary Vane Pump Eurovacuum EV Series Single Stage Oil Sealed Rotary Vane Pump Eurovacuum EV Series Single Stage Oil Sealed Rotary Vane Pumps Eurovacuum Company is offering its various products to meet industrial vacuum

More information

Retort Control in the Food Industry using the C360

Retort Control in the Food Industry using the C360 Product Application Guide AG/RCP 007_2 Retort Control in the Food Industry using the 11.35 Retransmission Output to Recorder Current to Pressure Converter 108.6 Current to Pressure Converter Steam 11.82

More information

clean and quiet dental

clean and quiet dental clean and quiet dental clean air for your dental clinic JUN-AIR introduced the first compressor almost 50 years ago and started the development of the technology and design, which has made JUN-AIR compressors

More information

globe reactions made easy Process perfection

globe reactions made easy Process perfection globe reactions made easy Process perfection Benefits More reactions in less time: The Globe oil drain unit and quick clamp allow single hand vessel changes in seconds Space saving: 50ml, 100ml, 250ml,

More information

Hospital Sterilizers. T-Max Line

Hospital Sterilizers. T-Max Line Hospital Sterilizers T-Max Line Hospital Sterilizers Sterilization is a complex process. Like any complex process, many things can go wrong. A Hospital depends on its CSSD: Central Sterile Services Department.

More information

Silicone oil is as the medium and the drying effect is uniform. The Temperature error is less than or equal 1.

Silicone oil is as the medium and the drying effect is uniform. The Temperature error is less than or equal 1. Characteristic: The process of pre-freezing and drying finishes in situ. The freeze dryer is equipped with observation windows and the drying process is intuitive. Silicone oil is as the medium and the

More information

Subpart 1. Installation. All plumbing systems must be. installed and tested according to this chapter and chapter 4715,

Subpart 1. Installation. All plumbing systems must be. installed and tested according to this chapter and chapter 4715, 4658.4500 PLUMBING SYSTEMS; NEW CONSTRUCTION. Subpart 1. Installation. All plumbing systems must be installed and tested according to this chapter and chapter 4715, the Minnesota Plumbing Code. Subp. 2.

More information

Effective Heat Sterilization in CO 2 Incubators

Effective Heat Sterilization in CO 2 Incubators Effective Heat Sterilization in CO 2 Incubators Volume 4, Number 3 Key Words Key Words Class 100 air Contamination control Contamination elimination HEPA filtration Incubator Heat Sterilization White Paper

More information

Energy Efficiency HOSPITALITY. www.energia.ie

Energy Efficiency HOSPITALITY. www.energia.ie Energy Efficiency HOSPITALITY www.energia.ie Your chance to reduce your business energy usage by as much as 20%! 20% is a significant figure and reducing your energy bill by this amount could make a real

More information

OT 40L / 90L. Laboratory Steam Sterilizers

OT 40L / 90L. Laboratory Steam Sterilizers OT 40L / 90L Laboratory Steam Sterilizers OT 40L / 90L LABORATORY STEAM STERILIZERS OT 40L / OT 90L laboratory steam sterilizers are specially designed for the steam sterilization of liquid and culture

More information

AUTOCLAVE PROGRAM. SOP Bio-006 FOR THE USE OF AUTOCLAVE FOR STERILIZATION OF MATERIALS AND BIOLOGICAL WASTE SOP

AUTOCLAVE PROGRAM. SOP Bio-006 FOR THE USE OF AUTOCLAVE FOR STERILIZATION OF MATERIALS AND BIOLOGICAL WASTE SOP ENVIRONMENTAL AND EMERGENCY MANAGEMENT ENVIRONMENTAL HEALTH AND SAFETY 175 Cabot St. Wannalancit Suite 311 Lowell MA 01854 Ruth Medina, Ph.D. Tel: 978-934-2778 Senior Biosafety Specialist/Biosafety Officer

More information

REGULATORS 175/153 Series

REGULATORS 175/153 Series 175/153 Series 175 HEATING 153 COOLING Watson McDaniel reserves the right to change the designs and/or materials of its products without notice. 2002 Watson McDaniel Company Revised 7/2002 175 / 153 Screwed

More information

Mechanical Seal Piping Plans

Mechanical Seal Piping Plans Mechanical Seal Piping Plans Single Seals plans 01, 02, 03, 11, 13, 14, 21, 23, 31, 32, 41 Dual Seals plans 52, 53A, 53B, 53C, 54, 55 Quench Seals plans 62, 65A, 65B, 66A, 66B Gas Seals plans 72, 74, 75,

More information

STERILIZATION AND DISINFECTION

STERILIZATION AND DISINFECTION STERILIZATION AND DISINFECTION Importance of hand washing shown by Semmelweis STERILIZATION A physical or chemical process that destroys or eliminates all forms of microbial life, including spores. A satisfactory

More information

SECTION 11 78 13 MORTUARY REFRIGERATORS

SECTION 11 78 13 MORTUARY REFRIGERATORS PART 1 - GENERAL 1.1 DESCRIPTION SECTION 11 78 13 MORTUARY REFRIGERATORS SPEC WRITER NOTE: Delete between //---// if not applicable to project. Also delete any other item or paragraph not applicable in

More information

Product-friendly and fast...

Product-friendly and fast... Product-friendly and fast......vacuum Heating and Drying Ovens V V T Optimum drying results under exceptional conditions Drying in a vacuum... The Advantages No doubt, there are various ways of drying

More information

Thermo Scientific Heratherm Microbiological Incubators. Safe, easy and efficient

Thermo Scientific Heratherm Microbiological Incubators. Safe, easy and efficient Microbiological Incubators Safe, easy and efficient Microbiological Incubators your samples We are obsessed with sample integrity. Each of our new laboratory microbiological incubators is designed with

More information

AIR CONDITIONING EFFICIENCY F8 Energy eco-efficiency opportunities in Queensland Foundries

AIR CONDITIONING EFFICIENCY F8 Energy eco-efficiency opportunities in Queensland Foundries AIR CONDITIONING EFFICIENCY F8 Energy eco-efficiency opportunities in Queensland Foundries Hot tips and cool ideas to save energy and money! Air conditioning units or systems are often used by foundries

More information

Systec Laboratory Autoclaves Systec H-Series. Horizontal floor-standing autoclaves. Systec H-Series 2D. Pass-through autoclaves.

Systec Laboratory Autoclaves Systec H-Series. Horizontal floor-standing autoclaves. Systec H-Series 2D. Pass-through autoclaves. Fotolia Layout Systec Laboratory Autoclaves Systec H-Series. Horizontal floor-standing autoclaves. Systec H-Series 2D. Pass-through autoclaves. Performance and competence. Experience counts We focus on

More information

STERILIZERS, LABORATORY DRYING OVENS

STERILIZERS, LABORATORY DRYING OVENS TS9026 TS9053 TS9135 TS9430 TS 9000 SERIES: STERILIZERS, LABORATORY DRYING OVENS Series TS9000 consists of four different cabinets available in sizes from 26 litres to 430 litres. Interior housing and

More information

Steam Sterilization Principles

Steam Sterilization Principles Reprinted from PHARMACEUTICAL ENGINEERING THE OFFICIAL TECHNICAL MAGAZINE OF ISPE NOVEMBER/DECEMBER 2013, VOL 33, NO 6 Copyright ISPE 2013 www.pharmaceuticalengineering.org facilities and equipment by

More information

VACUUM STERILIZER SWIFTLOCK SECURE-TOUCH+ TOUCHSCREEN. FRONT-LOADING Versions. 80 to 300 Litre. Instruction Manual

VACUUM STERILIZER SWIFTLOCK SECURE-TOUCH+ TOUCHSCREEN. FRONT-LOADING Versions. 80 to 300 Litre. Instruction Manual mxn738-p01ed a B&R #5-10 Vacuum basic SECURE-TOUCH + SWIFTLOCK STERILISER 1 SWIFTLOCK SECURE-TOUCH+ TOUCHSCREEN VACUUM STERILIZER FRONT-LOADING Versions 80 to 300 Litre Instruction Manual Powerscroft Rd

More information

Cooking at the Speed of light!

Cooking at the Speed of light! Cooking at the Infrared Cooking & Colouring Infrabaker is a modular infrared continuous cooking system developed by Infrabaker International. The machine is designed to cook and/or put colour on a wide

More information

Kompressoren. Adsorption Dryer AD

Kompressoren. Adsorption Dryer AD Kompressoren Adsorption Dryer AD Adsorption Drying: why? Modern industries require compressed air that is increasingly filtered with low dew point and condensate. Today, the equipment is more sophisticated

More information

Validation Guide. USTR 805j (5) Steam sterilization of Pall filter assemblies which utilize replaceable filter cartridges

Validation Guide. USTR 805j (5) Steam sterilization of Pall filter assemblies which utilize replaceable filter cartridges Validation Guide USR 805j (5) Steam sterilization of Pall filter assemblies which utilize replaceable filter cartridges CONENS 1. Important Recommendations 4 1.1 Installation 4 1.2 Integrity esting 4 1.3

More information

COMBIMASS. Technical Data COMBIMASS eco-bio +

COMBIMASS. Technical Data COMBIMASS eco-bio + COMBIMASS Technical Data THE SYSTEM COMBIMASS The field transmitters of the COMBIMASS eco series are suitable for gas flow measurement and cover a wide range of different applications. The instruments

More information

Decontamination and Waste Management www.biosecurity.sandia.gov

Decontamination and Waste Management www.biosecurity.sandia.gov Decontamination and Waste Management www.biosecurity.sandia.gov SAND No. 2006-3684C Sandia is a multiprogram laboratory operated by Sandia Corporation, a Lockheed Martin Company, for the United States

More information

USERS GUIDE. LOGIC Combi 24, 30, 35. For installation guide see reverse of book

USERS GUIDE. LOGIC Combi 24, 30, 35. For installation guide see reverse of book USERS GUIDE LOGIC Combi 24, 30, 35 For installation guide see reverse of book When replacing any part on this appliance, use only spare parts that you can be assured conform to the safety and performance

More information

LyoBeta. R+D Freeze-Dryers

LyoBeta. R+D Freeze-Dryers LyoBeta R+D Freeze-Dryers LyoBeta series A Responsible Choice New demands in R+D centers are related to scale-up feasibility and quality of procedures and research methods. FDA investigators consider that

More information

Assembly and Installation Procedures

Assembly and Installation Procedures Assembly and Installation Procedures for Pall Pharmaceutical Grade Capsule Assemblies 1. Introduction The following procedures must be followed for the installation of Pall pharmaceutical grade capsule

More information

Desuperheater Online Program Sizing Guidance

Desuperheater Online Program Sizing Guidance Local regulations may restrict the use of this product to below the conditions quoted. In the interests of development and improvement of the product, we reserve the right to change the specification without

More information

USERS GUIDE. LOGIC Heat 12, 15, 18, 24, 30. For installation guide see reverse of book

USERS GUIDE. LOGIC Heat 12, 15, 18, 24, 30. For installation guide see reverse of book USERS GUIDE LOGIC Heat 12, 15, 18, 24, 30 For installation guide see reverse of book When replacing any part on this appliance, use only spare parts that you can be assured conform to the safety and performance

More information

Cal Water 1961 Petra Lane, Placentia, CA 92870 (800) CAL-WATER (800) 225-9283 FAX: (714) 792-0794 http://www.cal-water.com

Cal Water 1961 Petra Lane, Placentia, CA 92870 (800) CAL-WATER (800) 225-9283 FAX: (714) 792-0794 http://www.cal-water.com Industrial Water Purification MAINTAINING A STERILE HIGH PURITY WATER SYSTEM Introduction Maintaining a microbe free deionized water system is like trying to maintain a vacuum, and nature abhors a vacuum.

More information

Multiple flasks. Easy to adapt. Stuart range of heating blocks

Multiple flasks. Easy to adapt. Stuart range of heating blocks Heating Blocks The Stuart range of heating blocks have been developed in conjunction with DrySyn to provide the best possible solution for heating round bottom flasks. Combined with the Stuart range of

More information

Hydraulic control unit series 0086-372-01

Hydraulic control unit series 0086-372-01 Technical Product Information No. 1290 EN Hydraulic control unit series 0086-372-01 Contents Page About this Technical Product Information (TPI) 2 The ORTLINGHAUS numbering system 2 About the product 3

More information

Link-Up. Solid Fuel Association Guide to Central Heating. www.solidfuel.co.uk

Link-Up. Solid Fuel Association Guide to Central Heating. www.solidfuel.co.uk Solid Fuel Association Guide to Central Heating Link-Up Combining solid fuel with other central heating technologies using various methods ranging from a simple link-up to a fully integrated thermal store

More information

ECOCIAT. Domestic hot water heat recovery unit

ECOCIAT. Domestic hot water heat recovery unit Heat recovery unit Domestic hot water High energy efficiency with R410A Compact and quiet Scroll compressors Brazed-plate heat exchangers Heating Heat recovery ENVIRONMENTALLY HFC R410A PROTECTION DE FRIENDLY

More information