Improving ECG Quality



Similar documents
Innovative solutions in resuscitation therapy

Arrhythmia Monitoring Algorithm

Streamlining Case Management Workflow

HOW TO USE THE LIFEPAK 1000 DEFIBRILLATOR 3

Biology 347 General Physiology Lab Advanced Cardiac Functions ECG Leads and Einthoven s Triangle

Cardiac Conduction System (1) ECG (Electrocardiogram) Cardiac Conduction System (2) The ECG (1) The ECG (1) The ECG (1) Achmad Rizal BioSPIN

Deriving the 12-lead Electrocardiogram From Four Standard Leads Based on the Frank Torso Model

Podcast with Dr. Kossick

Policy & Procedures. I.D. Number: 1142

IntelliVue to go. IntelliVue MX40 wearable patient monitor

It Takes a Team. Quite Alarming! Implementation of Alarm Management Strategies to Reduce the Incidence of False Alarms 5/1/2015.

Alarm Management. Scope and Impact of the Problem. Expected Practice and Nursing Actions*

HEART MONITOR TREADMILL 12 LEAD EKG

INTRODUCTORY GUIDE TO IDENTIFYING ECG IRREGULARITIES

Always in touch. Philips PageWriter TC50 cardiograph

Electrocardiography I Laboratory

THE SOCIETY FOR. Recording a standard 12-lead electrocardiogram. An Approved Methodology

Living With Your Pacemaker

R Series Quick Reference Guide

The Electrocardiogram (ECG)

Heart of America Medical Center EMR, EMT, EMT-Intermediate & AEMT Protocols

BIPOLAR LIMB LEADS UNIPOLAR LIMB LEADS PRECORDIAL (UNIPOLAR) LEADS VIEW OF EACH LEAD INDICATIVE ECG CHANGES

BARD MEDICAL DIVISION UROLOGICAL DRAINAGE. Foley Catheter Care & Maintenance. Patient Education Guide

ST Segment Monitoring. IntelliVue Patient Monitor and Information Center, Application Note

Connected Care. Philips HeartStart connectivity and data management for EMS professionals

A new way of thinking. Philips Sparq regional anesthesia and pain medicine ultrasound system

Expand your view. IntelliVue MX600, MX700, and MX800 bedside patient monitors

Electrocardiogram (ECG) Monitoring System using Bluetooth technology

What you should know

Information for patients What is an EMG/Nerve Conduction Study?

Improve your image. ClearVue 350 ultrasound system. The print quality of this copy is not an accurate representation of the original.

TENS and Electrodes Guide

Thyroid Eye Disease. Anatomy: There are 6 muscles that move your eye.

ACCESSORIES LIFEPAK DEFIBRILLATOR. for use with the. Genuine accessories from Physio-Control, a division of Medtronic.

Safety FIRST: Infection Prevention Tips

Kendall DL Cable and Lead Wire System. A Disposable Cardiac Monitoring Solution Designed for Superior Clinical Performance. Safe, Simple, Secure

ELECTROCARDIOGRAPHY (I) THE GENESIS OF THE ELECTROCARDIOGRAM

You ll love the Vue. Philips IntelliVue Information Center ix

Exercise, the Electrocardiogram, and Peripheral Circulation

Wall Mount Installation and Use

Electrophysiology Introduction, Basics. The Myocardial Cell. Chapter 1- Thaler

Physiotherapy in Rheumatoid Arthritis. Information for patients Gina Wall Senior Physiotherapist

X-Plain Subclavian Inserted Central Catheter (SICC Line) Reference Summary

IntelliVue MP5/MP5T TRAINING GUIDE. IntelliVue Patient Monitor. MP5/MP5T Release G.0 with Software Revision G.0x.xx. Patient Monitoring

A simple solution for your complex patients. The market-leading servo ventilation device System One BiPAP autosv

The new generation in ECG interpretation

Radiation Therapy To the Arms or Legs

Products and services, maximizing defibrillator performance. Philips HeartStart OnSite Defibrillator supplies and accessories

ACCESSORIES BROCHURE LIFEPAK Defibrillators. Genuine Accessories from Physio-Control

Electrocardiogram and Heart Sounds

Protecting systems and patient privacy

Connect care for early intervention

Efficient Heart Rate Monitoring

OEM MAXNIBP Frequently Asked Questions

A Patient s Guide to the Calypso System for Breast Cancer Treatment

Stress Echocardiogram

X-Plain Foley Catheter Male Reference Summary

1 Meet Your AliveCor Heart Monitor

SimplyGo Quick-start guide

GE Healthcare. The heart of cardiology is connectivity The MUSE * v8 Cardiology Information System

Department of Emergency and Disaster Medicine Medical University of LODZ

The choice for quality ECG arrhythmia monitoring

Networked versatility. IntelliVue MP60 and MP70 patient monitors

COVIDIEN: Your solution for cardiology monitoring supplies.

Radiotherapy to the chest wall

What Can I Do about Atrial Fibrillation (AF)?

Throughout this reference summary, you will find out what massage therapy is, its benefits, risks, and what to expect during and after a massage.

Introduction. Planned surgical procedures

Monitoring EKG. Evaluation copy

Rest ECG

Cardiac Arrest VF/Pulseless VT Learning Station Checklist

ENGLISH...3 DEUTSCH...13 FRANÇAIS...23 ESPAÑOL...33 ITALIANO...43 PORTUGUÊS...53

Activity 4.2.3: EKG. Introduction. Equipment. Procedure

Central Venous Catheter (CVC) Sterile Dressing Change - The James

GE Healthcare. Rest assured. MAC 5500 resting ECG analysis system

Easy to be used, they allow simple but specific and optimal clinical performances both in emergency cases and for elective uses.

Bringing Order Out of Chaos

12-Lead EKG Interpretation. Judith M. Haluka BS, RCIS, EMT-P

Use and Care Manual. Model CPA12KH AIR CONDITIONER

Level 1, Summer Street ORANGE NSW 2800 Ph: Fax:

Selecting Microwave/RF Cable Assemblies for Reliable Performance Over Time. White Paper

Electrophysiology study (EPS)

Section Two: Arterial Pressure Monitoring

Introducing the first 12-lead for the BLS team.

Regarding Insurance Reimbursement

Trilogy accessory guide

Patient Information Guide Morpheus CT Peripherally Inserted Central Catheter

Fetal monitoring on the move, from a hospital to in-home setting

Congratulations! Thank you!

Emergency Scenario. Chest Pain

Tachyarrhythmias (fast heart rhythms)

Cardiac Catheterization Lab Procedures

Feature Vector Selection for Automatic Classification of ECG Arrhythmias

High-intensity performance. IntelliVue MP90 patient monitor

Provided by the American Venous Forum: veinforum.org

The Radiotherapy Department Radiotherapy to the breast Information for patients

S.T.O.R.M Foreskin Restoration (Staggered Tapered O-Ring Method)

Transcription:

Improving ECG Quality Application Note Introduction Good ECG signal quality it is something every clinician strives to secure every time they monitor a patient s heart. This application note will look at why ECG signal quality is important and discuss how to improve the stability and quality of the signal as part of cardiac monitoring with electrodes, whether the patient is in an EMS or hospital environment. Importance of ECG Quality Acquisition of quality ECG data is critical in providing accurate and timely diagnosis and patient treatment. Poor ECG quality can: result in the monitor s inability to recognize the R-wave, thus affecting arrhythmia analysis and the ability to perform procedures such as synchronized cardioversion and transcutaneous pacing, inhibit reading a 12-Lead accurately, and result in false alarms and unnecessary troubleshooting, potentially delaying treatment.

Improving ECG Quality ECG Quality Improvement Back in 1984, in Critical Care Nursing, an article discussing ECG artifact stated: Improvement in electrode preparation techniques and a better understanding of the sources of artifact can enhance equipment performance, resulting in improved patient assessment, more effective utilization of nursing time, and reduced operating costs. 1 This statement still holds true today despite advancements in monitoring devices. In other words, poor electrode preparation can result in a poor waveform, no matter how advanced the device is. To prepare the skin for electrode placement, dry, dead epidermal layers of skin must be removed, along with any natural oils and dirt that impede electrical flow and thus create a resistance to signal quality. Though skin preparation should be well known by clinicians involved with ECG monitoring, it is worth reviewing the steps and related benefits. Figure 1 Epidermis Cross Section 4 So, how can you improve ECG quality and subsequently patient assessment and care? It involves: Good skin preparation, Use of quality electrodes, Proper electrode application, Good electrode-to-patient contact, Artifact elimination, and/or Proper lead selection. Let's look at each option more closely. Skin Preparation corneum lucidum granulosum mucosum germinativum Nerve fibrils Skin is a poor conductor of electricity and may create artifact that distorts the ECG signal. In fact, artifact from the outer layer of the skin (epidermis) is more troublesome than other types of artifact because: it is difficult to filter electronically and its amplitude is often larger than the ECG signal. 2 Unfortunately, clinical studies have shown that the electrode-skin interface is frequently overlooked as a major source of artifact affecting many electrophysiologic recordings. 3 Also, in a survey at the 1998 AAMI Annual Meeting and AACN NTI, only 17% of the participants said they had protocols in place that required skin preparation prior to electrode placement on a patient. 3 Although skin preparation can add a little time, the extra effort may reduce the time spent dealing with poor electrode-to-patient contact and ECG artifact. 1. Shave or cut hair from electrode sites since excessive hair prevents good electrode contact. 2. Clean each site thoroughly with soap and water, a non-alcohol wipe, or 4 x 4 to improve electrical flow. 3. Dry the skin vigorously to increase capillary blood flow to the tissues. NOTE: Though cleaning and drying the skin may be difficult in emergency situations, it can improve electrode adhesion and trace quality. Also, it is recommended not to use alcohol for cleaning purposes because it dries the skin and may diminish electrical flow. 4. Use an ECG skin prep pad or paper (such as Philips' M4606A ECG Skin Prep Paper), with its abrasive fine sandpaper finish, to: remove part of the stratum corneum (epidermis outer layer, see Figure 1 for epidermis cross section) to allow electrical signals to travel to the electrode and scratch the stratum granulosum (epidermis middle layer) to reduce motion artifact. 2

ECG Quality Improvement Using an abrasive has been documented to reduce skin resistance and artifact (with little probability of skin irritation) by Medina et al. 5 with just one stroke of fine sandpaper. Another study showed a 334 kω reduction in average impedance (from 354 to 20 kω), in part by using fine sandpaper material. 6 Abrasion also minimizes inertial and relative motion problems in the stratum granulosum, and rids the skin of oils and other debris that impede electrical flow. 7 Keep in mind that the stratum corneum can regenerate itself in as little as 24 hours, thus compromising electrode-to-patient contact. So, repeated skin abrasion should be considered during long-term monitoring. Though time is critical for the clinician who first treats a patient, a few more seconds of skin preparation using an abrasive can potentially improve ECG analysis and diagnosis, and subsequent treatment. Quality Electrodes Use of quality electrodes is an important part of the overall monitoring process. When selecting electrodes, keep the following guidelines in mind: Electrodes exist in different shapes and sizes, are made up of different materials, and are recommended for different clinical applications (e.g., ambulatory ECG, critical care unit, etc.). Use only electrodes specified by the manufacturer of the monitor/defibrillator you are using. Philips data sheet entitled Why buy Philips ECG Electrodes? can help make the electrode selection task easier. All electrodes selected should be of the same brand and type to help minimize noise. Adhere to all instructions for use found on the electrode packaging (e.g., Use by date). Electrodes that are packaged together in large quantities should be used shortly after the packaging is open or the packaging should be tightly re-sealed since extended exposure to air will dry out electrodes prematurely and reduce their adhesive and conductive properties. Consider writing the date and time that packaging is opened to help track electrode quality. If monitoring for extended periods of time, change electrodes periodically, based on manufacturer s specifications. Electrode Application Proper electrode application can help maximize ECG signal quality. Consider the following steps the next time you apply electrodes: When possible and appropriate, explain the electrode application procedure to the patient to decrease anxiety and increase relaxation. Attach the lead wire to each electrode before applying the electrode for patient comfort. For 3-lead and 5-lead, place electrodes as shown in Figures 2 and 3. For 10-lead, place electrodes on flat, fleshy parts of the arms and legs, avoiding bony areas and major muscles if possible, as shown in Figure 4. This minimizes muscle and motion-related artifact and maximizes the ECG signal strength. Apply the electrode by pressing around the entire edge of the electrode. Avoid pressing directly on the electrode center since it spreads the gel out and may create air pockets that contribute to noise. NOTE: If you are using multifunction electrode defibrillator pads, you may need to reposition ECG electrodes to allow for correct pad placement to facilitate pacing or defibrillation therapy. 3

Improving ECG Quality Figure 2 3-Lead Electrode Placement Figure 4 10-Lead Electrode Placement RA LA LL Figure 3 5-Lead Electrode Placement RA I LA II avr V avf avl III Electrode-to-Patient Contact Poor electrode-to-patient contact can result in poor signal quality and relative waveforms, as well as false alarms. Poor contact could result from to poor skin preparation, dried electrode gel, and/or defective ECG cables or lead wires. 1 Keep in mind the following guidelines to ensure good contact: RL LL Once attached, electrodes should not move in any way. Check electrodes periodically for firm adhesion and complete contact. If an electrode moves easily, the electrode connection is too loose and may need to be changed. Tape down the lead wire about 3-4 inches away from the electrode, with slack in the wire to prevent it from pulling on the electrode. 4

ECG Quality Improvement Artifact Elimination An artifactual signal is anything on an ECG that is not caused by the electrical currents generated by the heart. 2 This signal can come from sources such as: electrical interference (also known as low frequency noise) originating from the 60Hz current which supplies power to electrical wall outlets, or from a remote device such as a cell phone, muscle tremor, patient movement, or loss of electrode contact. 8 It is difficult to deal with electrical interference since it can t be filtered without compromising the ECG complex because of its similarity to the ECG signal frequency, so it is best to monitor away from other equipment, ensuring cable and lead wires do not cross the power cables of other equipment or vent tubing. To reduce muscle tremor and patient movement, attempt to warm a shivering patient or make them more comfortable in a reclined position, if possible, rather than adjust a filter setting. 2 Finally, continually check lead wire-to-electrode connection and electrode-to-patient s skin adhesion to ensure ECG quality and prevent false alarms, as mentioned in the previous section. Lead Selection It is important to select a suitable lead that shows the largest amplitude and cleanest signal so that a QRS complex and R-wave, in particular, can be accurately detected by the monitor. Consider the following lead selection guidelines: The QRS complex should be tall and narrow (recommended amplitude > 0.5 mv). The R-wave should be above or below the baseline (but not biphasic). The P-wave should be smaller than (1/5) R-wave height. The T-wave should be smaller than (1/3) R-wave height. To prevent detection of P-waves or baseline noises as QRS complexes, the minimum detection level for QRS complexes is set at 0.15 mv, according to AAMI-EC 13 specifications. If the ECG signal is too weak, you may get false alarms for asystole. In many monitors, increasing the gain (size) of the ECG does not change the signal to the monitor. Therefore, the signal should meet the above criteria without changing the gain to ensure the monitor can accurately analyze the waveform. 5

Improving ECG Quality Conclusion Clinicians depend on the monitors they use every day to provide accurate and useful information. When it comes to ECG quality, the electrode type, electrode application, and skin preparation are factors that play an important role in sending a good ECG signal to the monitor for analysis. The goal is to eliminate as much resistance, interference, and/or artifact as possible to ensure stable electrophysiologic recordings for better analysis, diagnosis, and treatment of the patient s condition. The results should be a reduction in: troubleshooting ECG quality, time delay to patient care, and frustration on the part of the clinician. References 1 Smith, M. Rx for ECG monitoring artifact. Critical Care Nurse, 1984: 4: 64-66. 2 Odma, S., Oberg, P. Movement-induced potentials in surface electrodes. Medical Engineering and Computing, 1982: 20: 159-166. 3 Adams-Hamoda, M., Caldwell, M., et al. Factors to consider when analyzing 12-lead electrocardiograms for evidence of acute myocardial ischemia. American Journal Crit Care, 2003: 12: 9-18. 4 Gray, Henry. Anatomy of the Human Body, 20th ed. Philadelphia: Lea & Febiger, 1918. 5 Medina, V., Clochesy, J.M., Omery, A. Comparison of electrode site preparation techniques. Heart Lung: J Crit Care, 1989: 18: 456-460. 6 Oster, C.D. Improving ECG trace quality. Biomedical Instrumentation & Technology, 2000: 34: 219-222. 7 Millar, S., Sampson, L.K., Soukup, S.M. AACN procedure manual for critical care. Philadelphia: WB Saunders, 1985. 8 Mirvis, D.M., Berson, A.S., Goldberg, A.L. Instrumentation and practice standards for electrocardiographic monitoring in special care units. Circulation, 1989: 79: 464-471. Philips Healthcare is part of Royal Philips Electronics On the web www.philips.com/heartstart By e-mail healthcare@philips.com By fax +31 40 27 64 887 By postal service Philips Healthcare 3000 Minuteman Road Andover, MA 01810-1085 Asia Tel: +852 2821 5888 Europe, Middle East, and Africa Tel: +49 7031 463 2254 Latin America Tel: +55 11 2125 0744 North America Tel: +425 487 7000 1 800 285 5585 (USA only) 2008 Koninklijke Philips Electronics N.V. All rights are reserved. Reproduction in whole or in part is prohibited without the prior written consent of the copyright holder. Philips Healthcare reserves the right to make changes in specifications or to discontinue any product at any time without notice or obligation and will not be liable for any consequences resulting from the use of this publication. Published Sept. 2008, Edition 1 Printed in the USA 453564119681 *453564119681* *1* 6