COOKBOOK. for. Aristarchos Transient Spectrometer (ATS)

Similar documents
SSO Transmission Grating Spectrograph (TGS) User s Guide

SKINAKAS OBSERVATORY. Astronomy Projects for University Students PROJECT THE HERTZSPRUNG RUSSELL DIAGRAM

Getting Started emacs.cshrc & :+/usr/local/classes/astr1030/astron/ source.cshrc cd /usr/local/classes/astr1030 idl .compile ccdlab2 ccdlab2 exit

Manual Analysis Software AFD 1201

Alpy 600 User Guide. Olivier Thizy François Cochard

To Begin Customize Office

Measuring the Doppler Shift of a Kepler Star with a Planet

Artemis Capture. User s Manual

The Concept(s) of Mosaic Image Processing. by Fabian Neyer

Planning Observations

Quick Start Guide NVR DS-7104NI-SL/W NVR. First Choice For Security Professionals

Detecting and measuring faint point sources with a CCD

Using Photometric Data to Derive an HR Diagram for a Star Cluster

Shimadzu UV-VIS User s Guide

Generative Drafting. Page DASSAULT SYSTEMES. IBM Product Lifecycle Management Solutions / Dassault Systemes

Extraction Heights for STIS Echelle Spectra

Meteor Spectroscopy First Steps

MY FIRST STEPS IN SLIT SPECTROSCOPY

Lacerta MGEN Quick Start Guide

Introduction to CCDs and CCD Data Calibration

4 Troubleshooting Your Projector

Artemis Capture. User s Manual. V. 2.1 August 2014

DAS202Tools v1.0.0 for DAS202 Operating Manual

product overview pco.edge family the most versatile scmos camera portfolio on the market pioneer in scmos image sensor technology

NVMS User Manual

High Resolution Imaging in the Visible from the Ground without Adaptive Optics: New Techniques and Results

Graphic Design. Background: The part of an artwork that appears to be farthest from the viewer, or in the distance of the scene.

INSTITUTE OF THEORETICAL PHYSICS AND ASTRONOMY MOLĖTAI ASTRONOMICAL OBSERVATORY THE 35/51 CM TELESCOPE OF THE MAKSUTOV SYSTEM (ACT-452) USER'S GUIDE

Quick Start Using DASYLab with your Measurement Computing USB device

XCal-View user manual

Flash MX Image Animation

ImagineWorldClient Client Management Software. User s Manual. (Revision-2)

Spectroscopy Using the Tracker Video Analysis Program

K Service Source. Video Conferencing. QuickTime Conferencing Kit, QuickTime Conferencing ISDN Kit

OPERATING INSTRUCTIONS Nikon TiE Deconvolution Microscope CCRB 1-220G

LCD Drawer User Manual

HR Diagram Student Guide

WEB TRADER USER MANUAL

QUIT MACSCHOOL AND DRAG YOUR SERVER ICONS TO THE TRASH OR HIT THE DISCONNECT BUTTON BESIDE THE SERVER ICON IN THE TOP SIDEBAR

The How Do Guide to AxioVision. Author: Brian Svedberg Image Analysis Specialist Carl Zeiss MicroImaging, Inc.

SMS (Server Management Software) Digital Video Recorder. User s Manual

MetaMorph Software Basic Analysis Guide The use of measurements and journals

SpecLine Offline Spectroscopy Analysis Software SPECLINE-P/SPECLINE-U. Installation and Operation Manual Document Number

User Guide LUXXOR VIDEO MICROSCOPE. 2 Luxxor Video Microscope Set Up

Contents. 1.2 Software Installation... 1

Using Excel (Microsoft Office 2007 Version) for Graphical Analysis of Data

Video surveillance camera Installation Guide

Labs in Bologna & Potenza Menzel. Lab 3 Interrogating AIRS Data and Exploring Spectral Properties of Clouds and Moisture

TopSpin Users Guide W. Dale Treleaven, Ph.D. Louisiana State University

Software Manual Ver 1.0 September 2015

Steps for using MagicConnect

Classroom Exercise ASTR 390 Selected Topics in Astronomy: Astrobiology A Hertzsprung-Russell Potpourri

DSLR Astrophotography

Ultraviolet-Visible (UV-Vis) Spectroscopy Background Information

This tutorial assumes that Visual3D has been installed and that a model has been created as described in Tutorial #1.

Remote Observing with the 1.23m Telescope (Vers U.T.)

Alpy calibration User Guide. Olivier Thizy François Cochard

Metatrader 4 Tutorial

SE05: Getting Started with Cognex DataMan Bar Code Readers - Hands On Lab Werner Solution Expo April 8 & 9

Beginners Guide to CQG FX

Quick Start Guide. DVR DS-7200HWI-SH Series DVR. First Choice For Security Professionals

Lab 3: Introduction to Data Acquisition Cards

DUKANE Intelligent Assembly Solutions

DinoXcope User Manual

1. Introduction... 3 Introduction Deal command... 13

Users Manual Model # English

AST 4723 Lab 3: Data Archives and Image Viewers

User Tutorial on Changing Frame Size, Window Size, and Screen Resolution for The Original Version of The Cancer-Rates.Info/NJ Application

Central Management Software CV3-M1024

Spectrophotometry of Ap Stars

Network Video Recorder Quick Operation Guide

Plotting: Customizing the Graph

DVR4C Remote Viewer Operation Manual Table of Contents EN 3 1. OVERVIEW MINIMUM PC REQUIREMENTS INSTALLING THE PROGRAM...

C n o t n e t n e t n s

Remote Director. Apple imac G5. Monitor Proofing System using the I. MANUFACTURER INTRODUCTION. SWOP Application Data Sheet

GelAnalyzer 2010 User s manual. Contents

ACS CLIENT SOFTWARE USER MANUAL

The Applied Imaging Cytogenetic Workstation

Imaging Systems Laboratory II. Laboratory 4: Basic Lens Design in OSLO April 2 & 4, 2002

A Tutorial for 3D Point Cloud Editor

Using Image J to Measure the Brightness of Stars (Written by Do H. Kim)

Specifications for spectra eligible for BeSS

Plots, Curve-Fitting, and Data Modeling in Microsoft Excel

Mouse and Pointer Settings. Technical Brief

Petrel TIPS&TRICKS from SCM

OPERATION MANUAL. MV-410RGB Layout Editor. Version 2.1- higher

HCImage Live Getting Started Guide

Basic Operation and Function of the MSD Pro Data Plus. Best viewed in printed form. Click on File>Print.

TOPSPIN INSTRUCTIONS

ElliCalc. Software Manual

A remote diagnostics system for the MERLIN array. D. Kettle, N.Roddis Jodrell Bank Observatory University of Manchester Macclesfield SK11 9DL

High Resolution Spatial Electroluminescence Imaging of Photovoltaic Modules

University of Wisconsin Chemistry 524 Spectroscopic Applications (GFAA, ICP, UV/Vis, Fluorescence)

LUCIFER. LMS Version LMS User Manual. Document Number: LBT-LUCIFER-MAN 030. Issue Number: Issue 1.0

Creating the Ultimate Image: Introduction

GeoVision Setup. Once all the settings for Windows are completed and you have all the hard drives setup you can install GeoVision.

Chapter I Model801, Model802 Functions and Features

Computer Lane Monitor for Windows

1. Installing The Monitoring Software

Correcting the Lateral Response Artifact in Radiochromic Film Images from Flatbed Scanners

Transcription:

NATIONAL OBSERVATORY OF ATHENS Institute for Astronomy, Astrophysics, Space Applications and Remote Sensing HELMOS OBSERVATORY COOKBOOK for Aristarchos Transient Spectrometer (ATS) P. Boumis, J. Meaburn, E. Xilouris, J. Alikakos, A. Liakos, N. Nanouris, O. Giannakis October, 2015 Version 2.3

Panayotis Boumis, Emmanuel Xilouris, John Alikakos, Alexios Liakos, Nikolaos Nanouris, Omiros Giannakis, Institute for Astronomy, Astrophysics, Space Applications and Remote Sensing, National Observatory of Athens, Greece John Meaburn, Jodrell Bank Centre for Astrophysics, University of Manchester, UK 2015, Helmos Observatory, Institute for Astronomy, Astrophysics, Space Applications and Remote Sensing, National Observatory of Athens 2

Contents 1. Starting up the system... 4 2. Grating Operation....... 2.1 Computer control..... 2.2 Manual control...... 2.3 Explanation of Grating operation... 3. CCD Operation. 3.1 CCD operation (Maxim DL software).. 3.2 Exposure..... 3.3 Shutdown procedure... 7 7 8 8 9 9 11 11 4. Closing down the system.. 12 5. Wavelength Calibration.. 12 6. Standard stars.. 14 7. Signal to Noise ratio 16 3

1. Starting up the system 1. Go to the TCC Cabinet (Dome building) and check that the power of the CCD camera is plugged-in (there is an indication ATS CCD on the appropriate plug which is inside the TCC cabinet at the bottom (Fig. 1). Fig. 1 The ATS CCD plug (white color) inside the TCC Cabinet 2. Select on the Telescope operation menu the SFM Positioning (Fig. 2) Fig. 2. The telescope operation menu for SFM positioning 4

Select the ATS instrument position (Fig.3) Fig. 3. The SFM positioning of ATS 3. To select the appropriate instrument offsets for ATS, do the following steps: Select the Options menu on the Telescope operation menu (Fig.4) Fig. 4. The telescope operation menu for telescope options Select the instrument offsets menu (Fig.5). Put the following values: On the x instrument offset à + 0 1 38 and on the y instrument offset à +0 2 32. 5

Fig. 5. The instrument offsets menu 4. To activate the instrument offsets select the Instrument Offsets on the Telescope menu (Fig.6) Fig. 6. The telescope operation menu for instrument offsets 5. Correct to the appropriate focus value using M2 / Absolute positioning (this value varies between -0.3 and -1.5 and has to be found the best one for each time of operation) 6

2. Grating Operation The systems that are needed in order to operate ATS in full are: (a) the mechanical console, (b) the graphical user interface and (c) the CCD software. gratings (grooves/mm) (arcsec) spectral range resolution dispersion centered wavelength Table 1. Parameters at the Aristarchos f/8 focus (f.l. = 17714 mm) option 1 (RED) option 2 (BLUE) option 3 1200 10 5736.9-7071.8 Å 1.3 Å 103 Å mm -1 6441.6 Å 1200 10 4309.8-5768.8 Å 1.4 Å 95 Å mm -1 5074.1 Å 600 10 4009.0-7257.0 Å 3.2 Å 245 Å mm -1 5691.5 Å 2.1 Computer control All positions are fixed so the user only needs to move to a specific position (grating). In order to set the instrument to a specific grating, the user must follow the steps below: 1. Turn on the ATS computer. 2. Switch the mechanical console to the computer control option (Fig. 7). 3. Turn on the mechanical console (Fig. 7). 4. Open the graphical user interface (grating) by double clicking (Fig. 8). 5. Depending on the grating (G1, G2 or G3) click the Set option and then Go. The gratings will start to move onto the linear mechanical rail until the selected one will go to the appropriate position. A red line at the bottom of the graphical user interface always shows always that position and the movement. During the movement there is an indication which shows if the movement is forward or reverse (the black circle becomes red). 6. When the led in the circle below the selected position (G1, G2 or G3) becomes red then the grating is on the appropriate position and the instrument is ready for use. Fig. 7. The mechanical console of ATS 7

Fig. 8. Graphical user interface to operate ATS 2.2 Manual control In case of a computer failure, the gratings can be moved and positioned by using the mechanical console (Fig. 7). To do that the user must follow the steps below: 1. Turn on the ATS computer. 2. Switch to the handset control option on the mechanical console (Fig. 7). 3. Turn on the mechanical console (Fig. 7). 4. Depending on the grating (G1, G2 or G3) push the Motor button on the handset console. The gratings will start to move onto the linear mechanical rail until the selected one goes to the appropriate position. Then stop pushing the Motor button. A red line below G1, G2, G3 at the mechanical console always shows the position and the movement. There are also two switches on the handset console where the forward or reverse motion can be selected. 5. When the light in the circle below the selected position (G1, G2 or G3) becomes red on the mechanical console, then the grating is at the appropriate position and the instrument is ready for use. 2.3 Explanation of Grating operation Some explanations concerning the operation of the ATS gratings are given below. There are a number of options that can be used. These are: G1: Position of first grating (Red) G2: Position of second grating (Blue) G3: Position of third grating Reverse: Moves back the gratings support on the trail Forward: Moves front the gratings support on the trail Stop: Stops the movement of the gratings support on the trail 8

3. CCD operation (Maxim DL software) 3.1 Exposure Start Maxim DL and open the Camera Control window by clicking on the 9). toolbar button (Fig. Fig. 9. Start up window of Maxim DL Move the Camera Control window to a convenient location on the screen. To do this, point the mouse at the title bar, click and hold the left mouse button, and drag. The Setup tab should appear first; if it is not displayed, click on the Setup tab near the upper left corner of the window. Click Connect (Fig. 10). Fig. 10. Camera Control window Set the Coolers On from the respective submenu (Fig. 10 - right side & Fig. 11) to activate the cooling system, and then click the associated Cooler button (Fig. 10 - left side) to set the temperature setpoint. Set the temperature setpoint to -20 degrees C and click OK. Wait until the temperature reaches the appropriate value. 9

Fig. 11. Coolers buttons Select Expose tab (Fig. 12). The Subframe option must be unchecked. This will ensure that you get full image. Turn off Binning by setting the X Binning value to 1 and the Y Binning value (if enabled) to Same. This will give you the highest resolution image. Set the Readout Mode if enabled to Normal. Check the Options menu to see if additional Camera Settings are available. If so, you may wish to check the settings of the camera. Under the Options menu, select No Calibration. If you wish to perform calibration (dark subtraction, flat fielding, etc.) as soon as each image is taken, you must first set up calibration frames using the Process menu Set Calibration command. Then select Full Calibration. Simple Auto-dark can also be used. Fig. 12. Expose tab on Camera Control Set Frame type to Light (normal image frame). Type in the desired exposure time in the Seconds field. Click Start and an image will appear (Fig. 13). If you like, you can save the image to disk. 10

Fig. 13. CCD image 3.2 Shutdown Procedure Warm up the CCD cooler before closing Maxim DL, unless you plan to restart the software. This can be triggered using the Warm Up button on the Setup tab (Fig. 14 & Fig. 11). When the camera is warmed up (i.e. its temperature is close to the ambient s temperature), then press the Coolers Off and Disconnect buttons and close the software. Fig. 14. Warm Up the CCD 11

4. Closing down the system 1. Check that the CCD camera is disconnected at the ATS computer and the Maxim DL software is closed. If not, then follow the instructions given on section 3.2. 2. Close down the graphical user interface. 3. Turn off the mechanical console. 4. Turn off the ATS computer. 5. Go to the TCC Cabinet (Dome building) and unplug the power of the CCD camera (there is an indication ATS CCD on the appropriate plug which is inside the TCC cabinet at the bottom. 5. Wavelength Calibration The calibration of the three gratings was made in the Optical Laboratory using a CuAr arc lamp. The arc produced for each grating can be seen below. The coefficients of each polynomic order used for the wavelength calibration for each grating are given in Table 2. Fig. 15. Arc for the Red 1200 groove/mm grating centered on 6441.6 Å. 12

Fig. 16. Arc for the Blue 1200 groove/mm grating centered on 5074.1 Å. Fig. 17. Arc for the 600 g/mm grating centered on 5691.5 Å. Table 2. The coefficients of the polynomials used for the wavelength calibration λ = λ 0 + Α*px + B*px 2 + C*px 3 + D*px 4 + E*px 5 where px = pixel value (0-1023) and λ = wavelength Red grating Blue grating Blue-red grating Value error Value error Value error λ 0 5736.877 0.87784 4309.794 0.47064 4008.963 11.76908 Α 1.43924 0.01373 1.55323 0.0045 3.42126 0.19224 Β -1.05E-04 5.81E-05-1.07E-04 1.13E-05-9.69E-05 0.00104 C -3.27E-08 8.74E-08-1.71E-08 7.63E-09-7.99E-07 2.39E-06 D 7.18E-12 4.27E-11 0 0 1.25E-09 2.45E-09 E 0 0 0 0-5.93E-13 9.24E-13 13

6. Standard stars In the following plots are shown the spectra of standard stars (luminosity class V) taken with the ATS using all gratings. A few famous stellar spectral lines (e.g. Balmer lines, MgI, NaI) are also indicated. In Table 3 are given the names of the standards, their spectral types, and their magnitudes. Fig. 18. Spectra of standard stars obtained with the ATS using the Red Grating. Fig. 19. Spectra of standard stars obtained with the ATS using the Blue Grating. 14

Fig. 20. Spectra of standard stars obtained with the ATS using the third Grating. Table 3. The standard stars list Name m V Spectral type HIP12640 5.74 A0 HIP 93443 6.45 A4 HIP 12332 5.45 A7 HIP 102253 5.6 A8 HIP 194 5.71 F0 HIP 19205 5.3 F1 HIP 5833 5.7 F5 HIP 102611 7.6 F9 Sky (Sun) G2 HIP 7733 7.72 G8 HIP 110606 8.07 K4 15

7. Signal to Noise ratio The signal-to-noise ratio (SNR) for ATS is estimated as a function of wavelength λ for a series of measurements of spectroscopic standard stars (with absorption only lines). The sample stars were collected from the Gemini observatory database; they were observed in the summer of 2015 and covered a wide range of spectral types. The sensitivity of SNR was then evaluated on the following three factors: (a) the spectral range, as provided in the three available gratings (BR, B, and R), (b) the exposure time, and (c) the apparent magnitude of a specific star. Estimating the ATS performance among the three available gratings The reduction process was carried out through suitable IDL routines in a bias-corrected framework, including the following steps: (i) dark (and bias) subtraction from the raw images, (ii) bias subtraction from the dark images, (iii) outliers filtering (to deal with hot or damaged pixels), (iv) pixel integration in the space direction (direction perpendicular to dispersion), and (v) robust polynomial representation of the stellar (and the sky background) continuum. The SNR was then estimated via the formula:, where N cont (λ) and N sky (λ) are the integrated intensity in the space direction of the stellar continuum and the sky background, respectively, along the whole spectral range of each grating. N dark is the dark noise accumulating intensity at the respective pixel series, while p is the number of the integrated pixels, f is the binning factor (here we worked with the highest pixel resolution, so f = 1), m is the number of stacked images (here we worked with a single only image, so m = 1) and σ read is the readout noise of the CCD in the working mode. All values of the aforementioned quantities were transformed to the number of electrons recorded from the CCD through the gain conversion of each mode. Note that no attempt was made to remove the telluric lines by flat-fielding our raw measurements. The sky background contribution proved to be too low to affect the SNR values (less than 2%), suggesting that sky measurements could be discarded from an observing program of bright stellar targets. The ATS performs better in the BR grating and closer to the maximum continuum of the program star. The spectral type is crucial for the ATS efficiency in the B and R gratings; as anticipated, the R mode behaves better than the B mode toward lower temperatures (Fig. 21). 16

Fig. 21. SNR sensitivity on the spectral range. SNR values are depicted for a A8V (upper) and a G8V (lower) standard star. 17

Estimating the ATS performance among various exposure times and stellar magnitudes To estimate the SNR for a star with an apparent magnitude m and an exposure of time t, we extend the aforementioned analysis by employing the Pogson law (to deal with the brightness) and by taking advantage of the linear response of a CCD camera (to deal with the time) with respect to our instrumental values of a standard star with the same spectral type. To simplify the procedure, we restrict our computations to the central visual wavelength of 5400 A. Assuming that m 0 is the magnitude of the reference star, the SNR is then calculated according to the formula below:, where n cont is the integrated intensity in the space direction of the stellar continuum divided by the exposure time of the reference star, while n dark is the corresponding rate of dark noise accumulating intensity at the respective pixel series. HIP 93443 (A4V, m 0 = 6.45 mag) was selected as reference star and the BR grating was set as the working mode; many trial exposures were available at different focal positions and offsets giving the opportunity to choose the one of the highest SNR value. However, this does not exclude the possibility of an even more successful offset, and as a result, the inferred values should be considered as a lower limit of ATS efficiency. The analysis shown that ATS can provide hourly exposure spectra of high quality (SNR > 100) for stars brighter than m = 13 mag, while the range 13-15 mag is still considered admissible (SNR > 20). The performance of ATS hardly reaches the SNR = 5 for a star with m = 17 mag, suggesting that this magnitude refers to the ATS lower brightness limit (Fig. 22). Note that the sky background was considered negligible and it was omitted from the overall procedure. 18

(a) (b) Fig. 22. SNR sensitivity on the stellar brightness and the exposure time (upper). SNR values are depicted for stars lying in the range 11-15 mag and various exposure times less than 1 hr. Magnitude-exposure time curves for a wide SNR range are also displayed (lower). ATS is inefficient to deal with stars fainter than m = 17 mag. 19