LNG FLOW CALIBRATION FACILITIES
|
|
|
- Antony Holt
- 10 years ago
- Views:
Transcription
1 LNG FLOW CALIBRATION FACILITIES Oswin Kerkhof, VSL Mijndert van der Beek, VSL Peter Lucas, VSL Maria Mirzaei, VSL Gerard Blom, VSL Gertjan Kok, VSL Keywords: calibration, flow, flowmeter, custody transfer, measurement, uncertainty, accuracy, transport fuel, quality measurement, 1. Introduction Reliable, accurate and commonly agreed measurement methods are a first requirement for the trade of goods. For the large scale LNG distribution chain there is a commonly agreed measurement practice as described in various ISO standards and in the GIIGNL Custody transfer handbook. In comparison with other commodities like natural gas or gasoline the total uncertainty on measured energy is very high for LNG. The LNG industry is therefore committed to improve their measurement accuracies to reduce their financial risks and to optimize their mass and energy balances. As an extension of the traditional distribution chain for LNG a new market of professional consumers is developing with LNG as transport fuel. In this case there is no commonly agreed measurement practice and the metrological framework is not yet in place. This will be a showstopper for the roll-out of the LNG as transport fuel because laws will be enforcing protection of (professional) consumers. For small scale LNG the legal metrology framework will lay down requirements on traceability to national standards and will define maximum permissible measurement errors. This is not the case for large scale LNG where the industrial players are assumed to take care of their own business interests. For this reason the development of an LNG calibration facility is a crucial step and urgently required for the development of the small scale LNG distribution chain. The results obtained with this research facility will however impact the business of the large scale LNG distribution chain as well. The development of a primary calibration standard and a mid-scale calibration facility could be the stepping stones towards the development of larger scale calibration facilities in the more distant future. 2. Objectives Together with industry and other metrological institutes VSL has been working on the development of new LNG standards and calibration facilities for the past 3-5 years. The end goal is a facility available for R&D, testing and calibration of LNG flow metering and LNG quality measurement systems. This paper will show the roadmap for the development of this facility and the results obtained so far with the development of flow calibration standards. The first step is the development and building of a primary mass flow standard. In this paper, we present the results of the primary mass flow standard validation with liquid nitrogen. The second step will be the construction of a mid-scale mass flow standard, traceable to the primary standard. This project has newly started in Other elements of the roadmap are a calibration standard for volume flow measurement and an LNG composition standard. Those elements are to be realized in new R&D projects.
2 Figure 1 is showing the roadmap towards a metrological framework for LNG custody transfer with a prominent role for the use of flow metering. The mass flow metering route is a three step approach consisting of a primary standard, a mid-scale standard traceable to the primary standard and a largescale standard. The large scale standard may not be needed if the mid-scale standard leads to validated extrapolation schemes from small to higher flows and perhaps even from water to LNG. The volume flow route is needed to accomplish traceability for large scale ultrasonic flow meters. This requires the development and validation of cryogenic piston prover technology or the conversion from mass to volume flow through an accurate density measurement/calculation. Accurate and reliable measurement of LNG composition is needed for quality measurements and density calculations. A test and calibration standard will be of great value in validating various types and brands of measurement systems. These can be either the traditional sampling based systems or the new kind of Raman spectroscopy based systems. A primary density standard is being developed in order to produce reference data with improved accuracy. The new reference data can be used to validate and perhaps to improve the accuracy of the equations of state used to calculate the LNG density based on composition. Figure 1 Roadmap towards a sound metrological framework for LNG custody transfer. 3. Development of primary mass flow standard 3.1 Measurement principle The primary cryogenic mass flow calibration standard, like any primary standard, is designed to provide direct traceability to the SI units. In this case it provides traceability to the SI unit of mass, the kilogram. The primary standard is not to be used for regular calibrations of flow meters but for the periodic calibration of a limited set of flow meters, which are often referred to as master meters. Those meters can then be used to calibrate other flow meters in a specially designed calibration set-up. A schematic picture of the measurement set-up is displayed in Figure 2. It consists of a 1 m 3 storage tank, a pump, the flow meter under test and a 0,5 m 3 weight tank placed on a mass balance. The principle of operation is as follows: the pump generates a flow of cryogenic liquid that passes the meter under test (MuT) and is collected in the weighing tank resting on top of the mass balance. In order to obtain stable conditions at the position of the MuT (temperature, pressure) the flow is initially circulated through the MuT while bypassing the weighing tank. When the conditions are stable the flow
3 is diverted to the weighing tank using a set of fast switches. When the weighing tank is nearly full, the flow is diverted back. The integrated flow over the period between the start and stop time stamps is compared with the increase of weight on the mass balance. A correction is made for the weight of the vapor that is displaced and evaporated from the weighing tank during filling. This amount of vapor is measured using a specially designed gas flow meter in the vapor return line. The cryogenic liquid is kept under subcooled conditions to prevent two-phase flows. This is accomplished by pressurizing the system up to 3 bar(g) to raise the boiling point. After some time of operation the liquid temperature will have increased and be in equilibrium with the operating pressure. The liquid will be subcooled again by depressurizing the system, waiting until the liquid has cooled down and quickly pressurizing the system. Figure 2 Schematic picture of the primary LNG flow standard 3.2 Critical elements There are a couple of elements in the system which will influence the measurement result and uncertainty. The most critical ones are hysteresis in the mass balance system, displaced vapor correction, linepack volume corrections and time stamping. A qualitative description and some measures to minimize the influence of these effects are discussed below. The experimental results quantifying these effects are described in more detail in the results paragraph Hysteresis in the mass balance system As shown in Figure 2, the weighing tank has fixed connections to the outer world through the filling line at the bottom and the vapor return line at the top. This will automatically cause strain on the weighing system potentially leading to unpredictable effects. The variation of strain on the weighing tank is minimized by using flexible tubing and an L-shaped configuration (not shown in the schematic picture). The flexible tube however does expand as a function of pressure. This sets requirements on the stability of the operating pressure throughout the calibration sequence. A simple on/off pressure control system has been implemented to this purpose. These measures cannot completely avoid variations in the residual force applied on the weighing system, leading to hysteresis effects and occasional jumps in the balance response. A semi-automatic recalibration system has been developed to facilitate frequent recalibration of the weighing system. This allows comparing the linearity of the balance before and after the calibration. Displaced vapor correction
4 In the case that there is no evaporation and no pressure collapse, the theoretical vapor gas volume flow rate will be equal to the displaced liquid volume entering the weighing tank. Note that there would be no displaced vapor if the weighing tank were to be closed. Experiments with closing the vapor return line however showed that the pressure was not under control. This is due to the fact that the liquid is injected from the bottom of the tank instead of the top. The reason for that is that we need to push the liquid back to the storage tank by applying a pressure difference between the weighing tank and storage tank. The correction for the displaced vapor proved to be on the order of 1% which is significant compared to the target measurement uncertainty of 0.1% for the liquid flow rate. A vapor flow meter has been designed with a target measurement uncertainty of better than 5%. Details on this flowmeter are provided in the experimental results sections. Linepack volume corrections The linepack volume is defined as the liquid volume between the MuT and the entrance of the weighing tank. Any weight change of the linepack volume between the start and end of the calibration must be corrected. For example, if the density has decreased, this means that some extra mass has been moved into the weighing tank that is not being accounted for in the MuT however. In order to monitor the density variation the temperature is being measured both at the position of the filling line and at the MuT. Time stamping The calibration result is based on the comparison of the collected weight on the balance and the flow counted by the MuT. In the so-called dynamic or flying start-stop mode there is a constant flow that is passing through the MuT and being measured continuously. The counting must start exactly at the moment that the flow is diverted to the weighing tank and stops when the flow is diverted away from the weighing tank at the end. This requires accurate time stamping. The diverting is accomplished by closing and opening two fast switched valves simultaneously. The complete switching process takes some time (time window less than 0.05 s) during which the flow direction is not defined exactly. The switching valve design is such that they are making a symmetrical movement related to each other. However the crossing point repeatability is a few 0.01 s. The switching and time stamp challenge is avoided in the so-called static start-stop method. In this case the pump is switched off after stable temperature conditions have been obtained in the pre-run mode. The diverter switches are then positioned in filling mode and the pump is switched on. After the weighing tank is nearly filled the pump is switched off. The calibration interval is simply defined as the period between switching the pump on and off. The main disadvantage of the start-stop method is that the meter output has to cross potentially non linear regions over its range. A comparison between the results of the dynamic and start-stop method is shown in the results paragraph. 4. Results The calibration results in terms of the deviation of the mass flow meter under test (MuT) is defined as follows: In which the Reference Mass is calculated from: In which Mass Vapor = Vapor mass vented from the weighing tank during calibration time window.
5 Mass Linepack Mass LNG It is calculated according to: = The change in trapped liquid mass between MuT and weighing tank. = Accumulated cryogenic liquid in the weighing tank during the test time window. In which: AvgReading τ 0, τ 1 = Average of all balance readings just before and after the test time window that fulfill the requirements defined below. Number of readings is mostly between 3 and 20 = time at start and stop moment respectively. The system is designed to run in dynamic or flying start-stop mode, although it is also suitable for static start-stop mode. The static mode is not applicable for some flow meter types. The balance readings just before τ 0 and after τ 1 are valid under the following prerequisites: 1. The difference between subsequent balance readouts (updated once per second) is less than kg; 2. The difference between the actual linepressure in the system and the linepressure setpoint is less than 30 mbar. (Linepressure was set at 3 bar gauge and controlled by a on/off control system). The average is calculated from a minimum of 3 balance readings that fulfilled these criteria. During the pre-run, a small purge batch is generated to flush out any gas bubbles in the linepack volume. The minimum purge amount is equal to the linepack volume and is approximately 0.03 cubic meter. The following prerequisites need to be fulfilled as well in order to produce a valid calibration result: 1. The temperature of the cryogenic liquid should be lower than the saturation temperature at the process pressure condition. The subcool margin is calculated from 2. The process temperature variations (expressed as standard deviation) should be smaller than 1 C; 3. The reading of the ultrasonic meter (installed upstream of the MuT) should be smooth without spikes. Spikes are indications of vapor bubbles in the process. ; 4.1 Physical characteristics of a calibration run In Figure 3, a number of measurands are plotted against time. From the figure we can see what is happening: A small pre-run test shows that there were still some bubbles in the lines. This can be observed from the spiky output of the USM (ultrasonic meter); Bubble formation seems logic at the start: the liquid temperature just crosses the saturation temperature. After a few moments the Sub cool margin reaches a safe level up to -4 C during the calibration run (dashed line). After an initial filling line purge of approximately 20 kg, the balance was loaded with 140 kg LIN during the actual calibration run; Apparently the flow rate in the pre-run loop is 0.5 kg/s higher than during the calibration run. This is caused by the fact that the pre-run piping configuration has a different pipeline resistance
6 compared to the calibration loop. Keep in mind that the calibration loop also consists of the vapor return line (including a gas flow meter and several valves) contributing to the total calibration flow resistance. The control valve at the weighing tank entrance can be used for tuning the line resistance to get a smooth transition from pre-run flow to calibration flow. Measurements however show that a smooth transition is not essential for this type of mass flow meter under test given the fast response of the meter, the linearity and the long test time; During the calibration run, the flow rate keeps varying a little. The USM and CFM (coriolis flow meter) are trending in the same way. The variation is caused by the pressure variations in the system. The pipeline resistance in the vapor return line is dependent on the gas pressure and thus affects the working point of the LNG pump. The unstable equilibrium between evaporation, saturation and condensation in the weighing tank causes the smallest movements of the liquid to disturb temperature stratifications leading to pressure collapse. The pressure control system will immediately respond by adding Nitrogen gas into the system. The balance reading is carried out at the points were all criteria earlier mentioned are fulfilled. These reading moments are indicated with red crosses in the graph; In the case that there is no evaporation and no pressure collapse, the theoretical vapor gas volume flow rate will be equal to the displaced liquid volume entering the weighing tank. The assumption that the vapor return gas temperature is equal to the saturation temperature in the tank at 3 bar(g) leads to a density of the nitrogen vapor of 16.7 kg/m 3 and a liquid density of kg/m 3 at T = C. So the relative mass loss from the weighing tank due to vapor would be 2.3%. Actual measurements indicate that the vapor temperature is at minimum -80 C which leads to a relative mass loss of 1%. This quantification sets the requirement of vapor gas meter accuracy to be better than 5% in order to contribute 0.05% on the accuracy of the calibration result Balance weight [kg] Readout point Mass Balance kg CMF kg/s USM l/s Op. Pressure Bar Offset Op.Press Bar Gas MF kg/s SubcoolMargin C Flow rates, Op pressure, Subcool margin [kg/s, l/s, kg/m3, barg, C] :11:31 15:12:58 15:14:24 15:15:50 15:17:17 15:18:43 15:20:10 15:21:36 15:23:02 Figure 3 Graph showing a number of relevant parameters before and during a calibration run Error! Reference source not found. Figure 4 gives more details about the system behavior during a calibration run: Return vapor gas: The oscillating behavior of the return vapor flow rate (light blue line) is caused by the on/off pattern of the pressure control.(light green line). During active gas venting and releasing, the vapor generation inside the weighing tank increases and vice versa;
7 The surface beneath the vapor mass flow line is representative for the total vented vapor released from the weighing tank (integral Q vapor(t) *dt). The calculations support the assumption of a relative mass loss of 1% due to vapor displacement and venting; Stability of density: The (fast response) density output of the CFM is a measure for the stability of the liquid condition. The dark blue line shows the delta between ρ average and ρ actual within the time window of the calibration; Stability of process pressure: The light green line depicts a control band within 30 mbar The derivative of the balance mass output cannot be used directly as reference in the mass flow rate domain. The balance output electronics cause a delay together with a non uniform readout. However for calibration purpose of the gas meter it can be used prevailing some statistic processing is used (See next sections) Flow rates, delta&deviation density, Oper. pressure [kg/s, %, barg] Op. Pressure Bar CMF kg/s dbalance kg/s delta dens % Dev. Dens % Gas MF kg/s :11:31 15:12:58 15:14:24 15:15:50 15:17:17 15:18:43 15:20:10 15:21:36 15:23: Gas mass flow rate [kg/s] Figure 4 Graph showing dynamics of some essential parameters. 4.2 Results of a mass flow meter calibration Two different types of 2 coriolis mass flow meters were calibrated with LIN (Liquid Nitrogen). The CMF was adjusted to read zero deviation when calibrated with ambient water. The results of the extensive calibration of a CMF are depicted as a function of mass flow rate in the following graph. Keep in mind that the indicated mass readout from the MuT is based upon a fixed pulse factor which was based upon the zero deviation water calibration for comparison. In 2009 this meter was also tested at NIST, Boulder Colorado. At NIST, the meter was tested in open air (no extra thermal insulation) whereas at VSL, the meter was mounted inside of a cold box filled with insulating perlite powder. According to the meter manufacturer these differences in set-up should not generate any difference. The VSL standard is not provided with an extra active cool-unit to maintain the liquid subcooled like in Boulder. Subcooling at VSL is achieved by increasing process pressure prior to calibration run. Most of the tests are carried out in flying start and stop mode (purple diamonds), were the MuT was stabilized in a pre-run at approximately the calibration flow rate setpoint. Data output collection is achieved with a fast counter and accumulated mass is related to the time window between the diverter switch trigger moments.
8 Some tests are done in Static start and stop mode (blue squares.) In this case, the MuT is ramped up from zero flow, kept a while at the setpoint flow and then ramped down to zero again. The timing issue does not play a role of importance, it is only the accumulated mass that counts and that will be directly compared to the balance mass increase. The main disadvantage is that the meter output has to cross potentially non linear regions over its range. 0.5 Overview all testresults Mass flow meter, sept oct 2012 Deviation MuT [%] Mass Flow Rate [kg/s] VSL 2012 static strt/stp 1.5 VSL 2012 flying strt/stp NIST Figure 5 Calibration results using the new VSL primary LNG flow standard and the NIST liquid nitrogen flow standard. The linepack volume correction can have a significant impact on the calculated deviation and it compensates for the instabilities caused by the (small) changes in density between the connections of MuT and weighing tank during the calibration time window. Linepack correction (C Linepack ) is calculated as follows: In which: ρ τ1 = density of the liquid in the linepack volume at time τ1 (end of calibration) ρ τ0 = density of the liquid in the linepack volume at time τ0 (start of calibration) V linepack =connected volume between flanges of CFM and weighing tank Mass Liquid =accumulated amount of liquid during the calibration. The sensitivity of density change due to variations in temperature is very high close to the saturation temperature: approximately 0.5%/ C. While the liquid density is slowly decreasing over time due to temperature rise, the corrections are in the order of 0 to +0.2%. It is however crucial to determine the representative temperature change in the connected volume. If linepack effects would not be considered, the results end up as depicted in Figure 6:
9 Overview all testresults Mass flow meter, sept oct 2012 Deviation MuT [%] Mass Flow Rate [kg/s] VSL 2012 no linepack correction VSL 2012 with linepack corrections Figure 6 Calibration results using the new VSL primary LNG flow standard with and without linepack corrections applied. In this case, the standard deviations appears to increase slightly if the corrections are applied. In other words, the results are not correlated with the linepack correction. The conclusion may be drawn that either: 1. The model for linepack correction is incorrect or not applicable or, 2. The measured density change at the position of the CMF is not representative for the actual density change in the linepack volume. From VSL experience scenario 2 is most likely and more efforts should be taken to determine the mixed cup temperature drift of the linepack volume. In any case, on average the density during the test run is slightly decreasing as expected. 4.3 Results of balance calibrations, effects of flexible hose connection The balance is calibrated numerous times with the deadweight disk semi automatic manifold prior and after calibration series. The flexible filling hose is constructed from two bellow-shaped tubes with vacuum insulation in between. The exterior of the hose is protected with steel wire braiding. The braiding is essential to prevent the hose from stretching during pressurization. This is however disadvantageous for the parasitic forces onto the weighing tank. Stable force offsets to the weighing tank are not of any concern because we are interested in the weight change after filling. The liquid flowing through the filling hose may however cause shocks, vibrations and mechanical hysteresis. The parasitic force changes caused by this are partly non-reversible and will attribute significantly to the normal hysteresis loop of the balance. The typical loop amplitude of a normally used weighing device of this quality level amounts to a few gram. In our configuration the maximum observed hysteresis was 280 grams. Figure 7 shows the balance deviation as a function of loading and for different process pressure.
10 Balance deviation as function of loading and for different process pressure Deviation: Balance indicated deadweight disks [g] Sum deadweight upon balance [kg] at 3 bar(g) 3 at 3 bar(g) 4 at 3 bar(g) 5 at 0 bar(g) 6 at 0 bar(g) 1750 Figure 7 Balance deviation as function of loading showing hysteresis In Table 1 the influence for each separate calibration hysteresis loop on the measurement uncertainty has been estimated. For the uncertainty model the hysteresis is conservatively described with a rectangular distributed probability function. batch U balance due to hysteresis in % (batch measurement) [%] 2s equivalent [kg] series 2 series 3 series 4 series 5 series Table 1 Uncertainty of balance reading due to hysteresis Series 2, 3 and 4 were subject to 3 bar(g), series 5 and 6 to atmospheric pressure. The preliminary conclusion could be drawn that different operating pressures do not affect hysteresis amplitude and/or deviation curve significantly. Furthermore the uncertainty evaluation of batch mass determination shows that batches less than 150 kg should be avoided. 4.4 Displaced vapor correction The displaced vapour is measured using a specially designed vapor return gas flow meter. The uncertainty requirement was set at 5% of reading under normal test conditions.the design is partly based on the ISO 5167 standard for orifice plates. Several designs were considered to find the best possible solution for gas metering under these demanding conditions: Temperature variations of more than 40 C during the test run;
11 Non stationary flow conditions; Fast changes of gas flow rate; Possible damage due to liquid overflow of the weighing tank; Suitable for cold Nitrogen as well as Methane gas; The gas temperature sensor is placed in axial direction in the centreline of the orifice throat and outlet piping. Since this is not in accordance to the ISO 5167 a laboratory or field calibration should be carried out. In the next figure, the configuration of this vapour gas meter is explained. Temperature sensor Orifice Vanes and stem support Anti radiation tube Figure 8 Schematic drawing of the return vapour flow meter To improve the representativeness of the measured gas temperature, the sensor is provided with an anti-radiation baffle and is in contact with the gas over 50 times stem diameter-length to prevent heat in-leak from the pipe wall to the sensor tip. The orifice plate armature is calibrated with air under atmospheric conditions and the correction factors are being fine tuned to match with the Reynolds number of the expected cold gas conditions for N 2 and CH 4. The model for the gas meter is derived from ISO 5167 and simplified to: In which ΔP = differential pressure over the orifice [bar] P upstream = Pressure just upstream of the orifice [bar] T throat =Gas temperature in the orifice throat [K] C 0,1 = constants depending on gas type N 2 or CH 4 The solution is solved by iteration after which the returned vapour mass is totalized according to: Where dt= reading interval (1 second) With a balance resolution of 1 g the vapour gas flow meter could be calibrated in the field by comparison with the derivate of the balance weight. Figure 9 shows the results when cold gas is vented from the weighing tank through the vapor gas meter:
12 Balance derivative and Vapor gas flow meter 50 Vapor mass flow rate [g/s] According to balance According to balance (crosscorrelated, moving averaged) Vapor Flow Meter output Elapsed time [s] Figure 9 Comparison of the balance weight derivative and the vapor gas flow meter readings The mass loss in the weighing tank is determined based on the balance readings at 1 second interval. Due to mechanical vibrations and time delay in the balance electronics, the calculated instantaneous mass loss cannot be directly used as a reference value for calibration purpose. Therefore, the data are processed by moving average filtering of the data and time lag compensation. The time lag is determined by a cross correlation technique. The processed data are used for the calibration of the vapor gas flow meter. Figure 10 shows that the vapor gas flow meter is over reading below 30 g/s. The origin of the offset of the gas meter at zero gas flow is probably caused by a ΔP transmitter zero shift. 50 Calibration of Vapor mass flow meter Vapor mass flow rate according to balance [g/s] based on rough balance data Based on processed balance data Vapor mass flow rate according to gas meter [g/s] Figure 10 Correlation Vapor gas flow meter related to Balance reading derivatives The next step is to operate the primary standard with LNG. Nitrogen gas will still be used to pressurize the system. Although this is expected to have a stabilizing effect on the operating pressure it will also cause additional uncertainty on the density of the displaced vapor being a mixture of nitrogen and natural gas. The vapor in the return line will need to be sampled to study the variations in composition.
13 5. Conclusions This paper reports on the liquid nitrogen validation of the primary LNG mass flow standard. This is a fundamental and first step on the roadmap towards sound metrological support for the new as well as the traditional LNG markets. The validation results show reasonable reproducibility figures. At flow rates of 1, 2 and 3 kg/s the reproducibility is respectively: 0.14%, 0.07% and 0.28% (2 times standard deviation of average). The main contributions to the irreproducibility come from balance hysteresis (nonreversible parasitic forces from the flex-tubing connection), vapor gas flow measurement and timing. It is believed that the coriolis meter under test was not significantly contributing; however this has to be confirmed with e.g. Youden analyses. The potential for further improvements of the reproducibility is realistic. The deviations of the VSL results compared to the results at NIST Cryogenic Flow measurement Facility (CFMF) are subject to further investigations in the near future. Further research / LNG validation Further work is required to finalize the uncertainty analyses for liquid Nitrogen. In the first half of 2013 the calibration standard will be validated with LNG with a focus on: Displacement measurements of flex-tubing to gain better understanding of possible parasitic forces and torques behavior to balance; Optimization of vapor gas flow measurements; Sampling of return vapor to determine the maximum variation in composition (natural gas + nitrogen gas) and density. Feed-forward control of pump speed for flow stability. Modeling of characteristic behavior of the system on basis of experience gained from previous Nitrogen tests; Youden analyses based on two meters under test to find out attributors; Calibrating different brands flow meters for LNG (Coriolis- and Ultrasonic meters); Improving response time and representativiness of temperature metering especially for better i.e. more accurate linepack determination, check for correlation improvement; VSL has, as a national standards institute, not only a responsibility to metrology as a science but also to the origin of metrology, viz. the need of society for reliable and traceable measurements as a base for fair trade, quality control and process metering. Development, continuity and stability of Reference Values is an important element in the function and service of a National Metrology Institute to society. Realization of reference values for LNG flow metering is one of VSL s focal points for the past and coming years. Acknowledgements We are grateful to the following people and organizations: Erik Smits and Maurice Heemskerk from VSL for discussions and support, Krister Stolt and Anders Anderson from SP for discussions and collaboration in executing the validation; Asaad Kenbar and Calum Hardy from TüV NEL, Jan Geršl from CMI and Michael Lewis and Daniel Friend from NIST for discussions. This research was co-funded by the Dutch Ministry of Economic Affairs and through the European Metrology Research Program (EMRP) which is jointly funded by the EMRP participating countries within EURAMET and the European Union. References: 1. Progress Report of the NIST Nitrogen Flow Facility, J.L. Scott and M.A. Lewis, Natl. Inst. Stand. Technol. Note 1364 (Apr.1994) 2. EMRP project website:
191: Calibration and Standards in Flow Measurement. Richard Paton National Engineering Laboratory, Scotland, UK 1 GENERAL PRINCIPLES
191: Calibration and Standards in Measurement Richard Paton National Engineering Laboratory, Scotland, UK 1 General Principles 1 2 Gravimetric Calibration of Liquid meters 2 3 Volumetric Calibration of
Grant Agreement No. 228296 SFERA. Solar Facilities for the European Research Area SEVENTH FRAMEWORK PROGRAMME. Capacities Specific Programme
Grant Agreement No. 228296 SFERA Solar Facilities for the European Research Area SEVENTH FRAMEWORK PROGRAMME Capacities Specific Programme Research Infrastructures Integrating Activity - Combination of
Gas Custody Transfer Calibration
Gas Custody Transfer Calibration Using multi variable temperature / pressure calibrators for flowmeter calibration 2013 Introduction Gas custody transfer flow computers require special calibration to perform
Aeration Air & Digester Gas Flow Metering Using Thermal Mass Technology. HWEA 2011 Conference Craig S. Johnson
Aeration Air & Digester Gas Flow Metering Using Thermal Mass Technology HWEA 2011 Conference Craig S. Johnson Presentation Overview Introduction Aeration Air & Digester gas challenges Gas flow metering
LNG Fuel Tank System LNG Tank overview
Index 1. LNG Tank Overview 2. 3. Gas Flow Diagram 4. Tank Operation 5. 6. Maintenance Boil off line Slave tank Master tank manual gas line vacuum control manual liquid line Check nonreturn filling line
Testing methods applicable to refrigeration components and systems
Testing methods applicable to refrigeration components and systems Sylvain Quoilin (1)*, Cristian Cuevas (2), Vladut Teodorese (1), Vincent Lemort (1), Jules Hannay (1) and Jean Lebrun (1) (1) University
BB-18 Black Body High Vacuum System Technical Description
BB-18 Black Body High Vacuum System Technical Description The BB-18 Black Body is versatile and is programmed for use as a fixed cold target at 80 K or variable target, at 80 K- 350 K no extra cost. The
Fiscal Measurement Natural Gas
White Paper FSG-WP-0012 February 2014 Fiscal Measurement Natural Gas Fiscal Measurement How much? Fiscal Measurement must not be confused with Custody Transfer; in fact, fiscal measurement is a more general
Regulations for custody transfer measurements: meeting the needs? Focus on liquid flow measurement systems/applications
Regulations for custody transfer measurements: meeting the needs? Focus on liquid flow measurement systems/applications Contents - Requirements in OIML publications - Uncertainty effects on trade - Back
A basic introduction to steam
A basic introduction to steam FOR HOT, COLD, MOIST AND DRY, FOUR CHAMPIONS FIERCE. STRIVE HERE FOR MASTERY Milton 1666 Steam Wonderful Steam Very high heat content Recyclable Clean, non toxic Biodegradable
Gravimetric determination of pipette errors
Gravimetric determination of pipette errors In chemical measurements (for instance in titrimetric analysis) it is very important to precisely measure amount of liquid, the measurement is performed with
Precision Mass Flow Metering For CVD Applications.
Precision Mass Flow Metering For CVD Applications. Ir. H.J. Boer Research Department of Bronkhorst High-Tech B.V. Nijverheidsstraat 1A, 7261 AK Ruurlo The Netherlands. Tel: +31 (0)573 458800 Fax: +31 (0)573
The paper addresses the boil-off in the cryogenic industry and details the specifics of it when applied to vehicle LNG tanks.
What is Boil-off? Scope... 1 Boil-off in the cryogenic industry... 1... 1 Measures for boil-off... 2 LNG vehicle tanks... 2 Boil-off for the vehicle LNG Tank... 2 Heat management and types of fuel delivery
Michael Montgomery Marketing Product Manager Rosemount Inc. Russ Evans Manager of Engineering and Design Rosemount Inc.
ASGMT / Averaging Pitot Tube Flow Measurement Michael Montgomery Marketing Product Manager Rosemount Inc. Russ Evans Manager of Engineering and Design Rosemount Inc. Averaging Pitot Tube Meters Introduction
AN INSTRUMENT FOR GRAVIMETRIC CALIBRATION OF FLOW DEVICES WITH CORROSIVE GASES. J. O. Hylton C. J. Remenyik Oak Ridge National Laboratory *
AN INSTRUMENT FOR GRAVIMETRIC CALIBRATION OF FLOW DEVICES WITH CORROSIVE GASES J. O. Hylton C. J. Remenyik Oak Ridge National Laboratory * Paper prepared for the 4 th International Symposium on Fluid Flow
GOOD PRACTICE GUIDE THE CALIBRATION OF FLOW METERS
GOOD PRACTICE GUIDE THE CALIBRATION OF FLOW METERS www.tuvnel.com The calibration of flow meters This good practice guide covers the general principles of calibration when applied to the flow meters or
Refrigeration Basics 101. By: Eric Nelson
Refrigeration Basics 101 By: Eric Nelson Basics Refrigeration is the removal of heat from a material or space, so that it s temperature is lower than that of it s surroundings. When refrigerant absorbs
A PRIMARY CALIBRATION SYSTEM FOR THE SUPPORT OF HIGH PERFORMANCE GAS FLOW TRANSFER STANDARDS
Presented 2006 May 16-18 2006 ISFFM Queretaro, Mexico A PRIMARY CALIBRATION SYSTEM FOR THE SUPPORT OF HIGH PERFORMANCE GAS FLOW TRANSFER STANDARDS P. Delajoud, M. Bair, C. Rombouts, M. Girard DH Instruments,
EASIDEW PORTABLE HYGROMETER INSTALLATION, OPERATION AND MAINTENANCE MANUAL
EASIDEW PORTABLE HYGROMETER INSTALLATION, OPERATION AND MAINTENANCE MANUAL Issue February 2004 2 TABLE OF CONTENTS SECTION PAGE 1. INTRODUCTION 3 1.1 General 3 1.2 Ceramic Sensing Element 3 1.3 Calibration
Flow Metrology for Liquefied Natural Gas (LNG)
Flow Metrology for Liquefied Natural Gas (LNG) EMRP Project Introduction and overview October 2011 Lars Poder, FORCE Technology About LNG Liquefied Natural Gas is produced by cooling down natural gas below
Saeid Rahimi. Effect of Different Parameters on Depressuring Calculation Results. 01-Nov-2010. Introduction. Depressuring parameters
Effect of Different Parameters on Depressuring Calculation Results Introduction Saeid Rahimi 01-Nov-2010 Emergency depressuring facilities are utilized to accomplish at least one of the following objectives:
AN_6521_035 APRIL 2009
71M6521 Energy Meter IC A Maxim Integrated Products Brand APPLICATION NOTE AN_6521_035 APRIL 2009 This document describes how to use software to compensate the real time clock (RTC) in Teridian meter chips.
Fundamentals of Mass Flow Control
Fundamentals of Mass Flow Control Critical Terminology and Operation Principles for Gas and Liquid MFCs A mass flow controller (MFC) is a closed-loop device that sets, measures, and controls the flow of
Thermal Dispersion Mass Flow
Thermatel Thermal Dispersion Mass Flow Measurement Handbook Table of Contents Introduction...1 What is Mass Flow Measurement?...1 Types of Flow Transmitters...4 Differential Pressure...4 Orifice...4 Venturi...6
The On-Board Refueling Vapor Recovery (ORVR) Evaporative Emission (EVAP) system.
«1A: Description and Operation» Overview The Evaporative Emission (EVAP) system prevents fuel vapor build-up in the sealed fuel tank. Fuel vapors trapped in the sealed tank are vented through the vapor
OPTIMASS 6000 Technical Datasheet
OPTIMASS 6000 Technical Datasheet Sensor for mass flow The high performance meter for process industries Cryogenic, high temperature and high pressure options Supreme liquid and gas performance with CT
Sentinel LCT. Panametrics Ultrasonic Flowmeter for Liquid Custody Transfer Measurement. GE Measurement & Control Solutions. Benefits.
GE Measurement & Control Solutions Sentinel LCT Panametrics Ultrasonic Flowmeter for Liquid Custody Transfer Measurement Benefits Extremely reliable and highly accurate flowmeter, designed specifically
TEMPERATURE MEASURING GALVANIC SEPARATORS CONVERTORS, PRESSURE MEASURING
DELIVERY PROGRAMME TEMPERATURE MEASURING - Resistance temperature sensors into a reservoir, - with a reservoir, with protection tube, with cable - Resistance temperature sensors Ex to zones 2, 1 - - and
DIRECT STEAM INJECTION HOT WATER SYSTEMS FOR JACKETED HEATING
By Philip Sutter Pick Heaters, Inc. DIRECT STEAM INJECTION HOT WATER SYSTEMS FOR JACKETED HEATING INTRODUCTION Many process plants currently use steam or hot water to heat jacketed devices such as tanks,
High Precision Measurement of the Target Mass of the Daya Bay Neutrino Detectors
High Precision Measurement of the Target Mass of the Daya Bay Neutrino Detectors Tom Wise on behalf of Daya Bay collaboration Day Bay is a reactor based experiment dedicated to measurement of the last
Select the Right Relief Valve - Part 1 Saeid Rahimi
Select the Right Relief Valve - Part 1 Saeid Rahimi 8-Apr-01 Introduction Selecting a proper type of relief valve is an essential part of an overpressure protection system design. The selection process
ORGANIC SAMPLE PREPARATION
ORGANIC SAMPLE PREPARATION W W W.LA BT E C H S R L.CO M WSPE MANUAL VACUUM MANIFOLD SPE Process control of the flow rate is critical to guarantee reproducible extractions. Differently then any other systems,
Calibration and Use of a Strain-Gage-Instrumented Beam: Density Determination and Weight-Flow-Rate Measurement
Chapter 2 Calibration and Use of a Strain-Gage-Instrumented Beam: Density Determination and Weight-Flow-Rate Measurement 2.1 Introduction and Objectives This laboratory exercise involves the static calibration
C. starting positive displacement pumps with the discharge valve closed.
KNOWLEDGE: K1.04 [3.4/3.6] P78 The possibility of water hammer in a liquid system is minimized by... A. maintaining temperature above the saturation temperature. B. starting centrifugal pumps with the
CFD SIMULATION OF SDHW STORAGE TANK WITH AND WITHOUT HEATER
International Journal of Advancements in Research & Technology, Volume 1, Issue2, July-2012 1 CFD SIMULATION OF SDHW STORAGE TANK WITH AND WITHOUT HEATER ABSTRACT (1) Mr. Mainak Bhaumik M.E. (Thermal Engg.)
Heating Water by Direct Steam Injection
Heating Water by Direct Steam Injection Producing hot water by direct steam injection provides a solution where large volumes of hot water at precise temperatures are required, and where energy and space
Energy and Flow Measurement for Hydronic Systems
Energy and Flow Measurement for Hydronic Systems Presented By: George Szyszko Applications Consultant MCR for ONICON Incorporated Why Measure Thermal Energy? To control something, you must first measure
A Practical Guide to Free Energy Devices
A Practical Guide to Free Energy Devices Part PatE17: Last updated: 24th November 2006 Author: Patrick J. Kelly This is a patent application from Stephen Meyer, brother of the late Stan Meyer. While this
COMBIMASS. Technical Data COMBIMASS eco-bio +
COMBIMASS Technical Data THE SYSTEM COMBIMASS The field transmitters of the COMBIMASS eco series are suitable for gas flow measurement and cover a wide range of different applications. The instruments
Subminiature Load Cell Model 8417
w Technical Product Information Subminiature Load Cell 1. Introduction... 2 2. Preparing for use... 2 2.1 Unpacking... 2 2.2 Using the instrument for the first time... 2 2.3 Grounding and potential connection...
Laminar and Turbulent flow. Flow Sensors. Reynolds Number. Thermal flow Sensor. Flow and Flow rate. R = Mass Flow controllers
Flow and Flow rate. Laminar and Turbulent flow Laminar flow: smooth, orderly and regular Mechanical sensors have inertia, which can integrate out small variations due to turbulence Turbulent flow: chaotic
The Secret of Hydraulic Schematics. BTPHydraulics www.iranfluidpower.com
The Secret of Hydraulic Schematics BTPHydraulics www.iranfluidpower.com www.iranfluidpower.com Table of Contents The Secret to Reading and Interpreting Hydraulic Schematics... 1 Hydraulic System Schematics...
RITTER Multiple Sonic Nozzle Calibration System
RITTER Multiple Sonic Nozzle Calibration System »Wouldn t it be great to eliminate doubtful measurement results by using proven measurement technology? Providing the most precise results ensures and increases
Dynamic Process Modeling. Process Dynamics and Control
Dynamic Process Modeling Process Dynamics and Control 1 Description of process dynamics Classes of models What do we need for control? Modeling for control Mechanical Systems Modeling Electrical circuits
www.kem-kueppers.com [email protected]
www.kem-kueppers.com [email protected] Calibration of flow meters according to DIN EN ISO/IEC 17025:2005 Introduction... 2 Calibration laboratory... 3 Description of the calibration process... 3 Calibration
Effects of Atmospheric Pressure on Gas Measurement
Effects of Atmospheric Pressure on Gas Measurement Class # 1390.1 March 2012 / White paper by Denis Rutherford Regional Sales Manager Central US Schneider Electric Telemetry & Remote SCADA Solutions 6650
Measurement of Specific Heat Capacity Using Differential Scanning Calorimeter
INL/EXT-08-15056 Measurement of Specific Heat Capacity Using Differential Scanning Calorimeter J. E. Daw November 2008 The INL is a U.S. Department of Energy National Laboratory operated by Battelle Energy
How To Clean Up A Reactor Water Cleanup
General Electric Systems Technology Manual Chapter 2.8 Reactor Water Cleanup System TABLE OF CONTENTS 2.8 REACTOR CLEANUP SYSTEM... 1 2.8.1 Introduction... 2 2.8.2 System Description... 2 2.8.3 Component
Unsteady Pressure Measurements
Quite often the measurements of pressures has to be conducted in unsteady conditions. Typical cases are those of -the measurement of time-varying pressure (with periodic oscillations or step changes) -the
Experiment 3 Pipe Friction
EML 316L Experiment 3 Pipe Friction Laboratory Manual Mechanical and Materials Engineering Department College of Engineering FLORIDA INTERNATIONAL UNIVERSITY Nomenclature Symbol Description Unit A cross-sectional
F ox W hi t e Paper. Reduce Energy Costs and Enhance Emissions Monitoring Systems
F ox W hi t e Paper Reduce Energy Costs and Enhance Emissions Monitoring Systems A Technical White Paper from Fox Thermal Instruments Rich Cada, VP Sales & Marketing, Fox Thermal Instruments, Inc. 399
Flow System Calibration in Medical and Bio-medical Production Processes.
Flow System Calibration in Medical and Bio-medical Production Processes. Speaker / Author: Mark Finch Fluke Calibration Norfolk, Norwich, NR6 6JB, UK e-mail: [email protected] Phone: + 44 (0)1908 678
Flow Measurement Options for Pipeline and Open Channel Flow
Flow Measurement Options for Pipeline and Open Channel Flow October 2013 Presented by Molly Skorpik - 2013 Montana Association of Dam and Canal Systems Conference Irrigation Training and Research Center
Investigations of a Long-Distance 1000 MW Heat Transport System with APROS Simulation Software
th International Conference on Structural Mechanics in Reactor Technology (SMiRT ) Espoo, Finland, August 9-4, 9 SMiRT -Division 3, Paper 56 Investigations of a Long-Distance MW Heat Transport System with
FORMATION DAMAGE AND WELL TREATMENT SYSTEM
FORMATION DAMAGE AND WELL TREATMENT SYSTEM FDWTS-10K Includes ability to test Liquid Permeability & Gas Permeability Fines migration & critical velocity Static Filtration Drilling mud invasion evaluation
جامعة البلقاء التطبيقية
AlBalqa Applied University تا سست عام 997 The curriculum of associate degree in Air Conditioning, Refrigeration and Heating Systems consists of (7 credit hours) as follows: Serial No. Requirements First
STEAM HEATING SYSTEM TROUBLESHOOTING GUIDE
Page 1 of 9 PURPOSE Steam is the most commonly used heating medium for maintaining process temperatures. Compared to other heating media, steam contains a significant amount of heat energy, and this heat
Experimental Evaluation of the Discharge Coefficient of a Centre-Pivot Roof Window
Experimental Evaluation of the Discharge Coefficient of a Centre-Pivot Roof Window Ahsan Iqbal #1, Alireza Afshari #2, Per Heiselberg *3, Anders Høj **4 # Energy and Environment, Danish Building Research
Procon Engineering. Technical Document PELR 1002. TERMS and DEFINITIONS
Procon Engineering Technical Document PELR 1002 TERMS and DEFINITIONS The following terms are widely used in the weighing industry. Informal comment on terms is in italics and is not part of the formal
THE THERMAL FLOW METER, A GAS METER FOR ENERGY MEASUREMENT
THE THERMAL FLOW METER, A GAS METER FOR ENERGY MEASUREMENT Kazuto Otakane, Tokyo Gas Co., Ltd Katsuhito Sakai, Tokyo Gas Co., Ltd Minoru Seto, Tokyo Gas Co., Ltd 1. INTRODUCTION Tokyo Gas s new gas meter,
Precision Miniature Load Cell. Models 8431, 8432 with Overload Protection
w Technical Product Information Precision Miniature Load Cell with Overload Protection 1. Introduction The load cells in the model 8431 and 8432 series are primarily designed for the measurement of force
DETERMINATION OF THE HEAT STORAGE CAPACITY OF PCM AND PCM-OBJECTS AS A FUNCTION OF TEMPERATURE. E. Günther, S. Hiebler, H. Mehling
DETERMINATION OF THE HEAT STORAGE CAPACITY OF PCM AND PCM-OBJECTS AS A FUNCTION OF TEMPERATURE E. Günther, S. Hiebler, H. Mehling Bavarian Center for Applied Energy Research (ZAE Bayern) Walther-Meißner-Str.
Using meters to measure steam flow
Page 1 of 6 April 1998 Plant Engineering SELECTING NEW TECHNOLOGY: PROBLEM SOLVING Using meters to measure steam flow Jesse L. Yoder, Ph.D., Senior Analyst, Automation Research Corp., Dedham, MA Key Concepts
LOCATION DEPENDENCY OF POSITIONING ERROR IN A 3-AXES CNC MILLING MACHINE
th International & 26 th All India Manufacturing Technology, Design and Research Conference (AIMTDR 214) December 12 th 14 th, 214, IIT Guwahati, Assam, India LOCATION DEPENDENCY OF POSITIONING ERROR IN
Triple Point Experiment
Equipment List Rotary vane vacuum pump 2-stage, 2 to 7 CFM. For example, Edwards 2M2 (2 CFM), Edwards RV5 (3.5 CFM), Edwards E2M8 (6.7 CFM) or equivalent. Bell jar Nalgene polycarbonate plastic, 5-5/8
APPLICATION OF DENSITOMETERS TO LIQUID MEASUREMENT Class # 2010
APPLICATION OF DENSITOMETERS TO LIQUID MEASUREMENT Class # 2010 Colin B. Blakemore Sr. Product Specialist Pipeline & Gas Measurement Products AMETEK Process Instruments 455 Corporate Boulevard Newark,
The Versatile Differential Pressure Transmitter. By David Gunn Honeywell Process Solutions
The Versatile Differential Pressure Transmitter By David Gunn Honeywell Process Solutions The Versatile Differential Pressure Transmitter 2 Table of Contents Abstract... 3 Pressure Fundamentals... 3 Applications...
DEVELOPMENT OF A TWIN SCREW EXPRESSOR AS A THROTTLE VALVE REPLACEMENT FOR WATER-COOLED CHILLERS
DEVELOPMENT OF A TWIN SCREW EXPRESSOR AS A THROTTLE VALVE REPLACEMENT FOR WATER-COOLED CHILLERS J J Brasz, Carrier Corporation, Syracuse, NY, 13221, USA [email protected] I K Smith and N Stosic
A New Technique Provides Faster Particle Size Analysis at a Lower Cost Compared to Conventional Methods
A New Technique Provides Faster Particle Size Analysis at a Lower Cost Compared to Conventional Methods Howard Sanders and Akshaya Jena Porous Material Inc. Ithaca, NY The technique described here calculates
QNET Experiment #06: HVAC Proportional- Integral (PI) Temperature Control Heating, Ventilation, and Air Conditioning Trainer (HVACT)
Quanser NI-ELVIS Trainer (QNET) Series: QNET Experiment #06: HVAC Proportional- Integral (PI) Temperature Control Heating, Ventilation, and Air Conditioning Trainer (HVACT) Student Manual Table of Contents
Rusty Walker, Corporate Trainer Hill PHOENIX
Refrigeration 101 Rusty Walker, Corporate Trainer Hill PHOENIX Compressor Basic Refrigeration Cycle Evaporator Condenser / Receiver Expansion Device Vapor Compression Cycle Cooling by the removal of heat
Refrigerant Charging Unit ICOGD. 020AH1000 Operating Manual. FR.8.2.4-09 İ-COLD 12.03.2014 Rev. 00
E Refrigerant Charging Unit ICOGD 020AH1000 Operating Manual FR.8.2.4-09 İ-COLD 12.03.2014 Rev. 00 Contents Technical Specifications... 20 Safety... 21 A/C System... 22 Components... 23 Control Panel...
User checks and maintenance of laboratory balances
Technical Note 13 March 2014 Issued: July 1995 Amended and reissued: July 2005, July 2007, October 2007, May 2010, March 2013, March 2014 User checks and maintenance of laboratory balances Copyright National
M&V of Hot Water Boiler Plant. Boban Ratkovich, P. Eng, CEM, BESA, LEED AP President CES Engineering Ltd
M&V of Hot Water Boiler Plant Boban Ratkovich, P. Eng, CEM, BESA, LEED AP President CES Engineering Ltd AIA Quality Assurance Learning Objectives 1. Case study examples of procedures and obstacles in measuring
Application of the Orifice Meter for Accurate Gas Flow Measurement page 1. Application of the Orifice Meter for Accurate Gas Flow Measurement.
Application of the Orifice Meter for Accurate Gas Flow Measurement page 1 DANIEL MEASUREMENT AND CONTROL WHITE PAPER Application of the Orifice Meter for Accurate Gas Flow Measurement www.daniel.com Summary
Correcting a Coriolis Meter for Two Phase Oil & Gas Flow
Correcting a Coriolis Meter for Two Phase Oil & Gas Flow Chris Mills, NEL 1 INTRODUCTION In the oil & gas industry, the requirement for flow measurement of two phase oil and gas flow is prevalent. This
saving energy in comfort Factsheet Recair Enthalpy
saving energy in comfort Factsheet Recair Enthalpy Introduction Enthalpy Exchanger The enthalpy exchanger is able to transfer both sensible heat and latent heat (moisture) from the cooled to the heated
The Fundamentals of Gas Flow Calibration
The Fundamentals of Gas Flow Calibration Application Note Introduction Understanding the fundamentals of gas flow calibration is essential for evaluating calibration systems, estimating the magnitude of
Head Loss in Pipe Flow ME 123: Mechanical Engineering Laboratory II: Fluids
Head Loss in Pipe Flow ME 123: Mechanical Engineering Laboratory II: Fluids Dr. J. M. Meyers Dr. D. G. Fletcher Dr. Y. Dubief 1. Introduction Last lab you investigated flow loss in a pipe due to the roughness
Physics 9e/Cutnell. correlated to the. College Board AP Physics 1 Course Objectives
Physics 9e/Cutnell correlated to the College Board AP Physics 1 Course Objectives Big Idea 1: Objects and systems have properties such as mass and charge. Systems may have internal structure. Enduring
In semiconductor applications, the required mass flows
Model-Based Multi-GasMulti-Range Mass Flow Controllers With Single Gas Calibration and Tuning JOOST LÖTTERS A model has been developed that accurately predicts the behavior of the flow sensor combined
A Demonstration Plant of a Liquid Desiccant Air Conditioning Unit for Drying Applications
A Demonstration Plant of a Liquid Desiccant Air Conditioning Unit for Drying Applications Mustafa Jaradat, Klaus Vajen, Ulrike Jordan Institut für Thermische Energietechnik, Universität Kassel, 34125 Kassel,
STANDARD CLEANING AND CALIBRATION PROCEDURE FOR TGA-50(H) AND TGA-51(H)
STANDARD CLEANING AND CALIBRATION PROCEDURE FOR TGA-50(H) AND TGA-51(H) by Joseph F. Greenbeck, Jr. And Courtney W. Taylor INTRODUCTION The following procedure is to be used for the maintenance and calibration
Principles of Direct-Operated Regulators
Principles of Direct-Operated Regulators Introduction Pressure regulators have become very familiar items over the years, and nearly everyone has grown accustomed to seeing them in factories, public buildings,
Kompressoren. Adsorption Dryer AD
Kompressoren Adsorption Dryer AD Adsorption Drying: why? Modern industries require compressed air that is increasingly filtered with low dew point and condensate. Today, the equipment is more sophisticated
OPTIMASS 1400 Technical Datasheet
OPTIMASS 1400 Technical Datasheet Sensor for mass flow First choice for universal applications Best price-performance ratio A wide range of options available with no restrictions The documentation is only
Series DFMT Digital Paddlewheel Flowmeter
Series DFMT Digital Paddlewheel Flowmeter Bulletin F-DFMT Specifications - Installation Operating Instructions -/8 [00] -/ [8] -/8 [00] A The Series DFMT Digital Paddlewheel Flow Transmitter provides instantaneous
Free Fall: Observing and Analyzing the Free Fall Motion of a Bouncing Ping-Pong Ball and Calculating the Free Fall Acceleration (Teacher s Guide)
Free Fall: Observing and Analyzing the Free Fall Motion of a Bouncing Ping-Pong Ball and Calculating the Free Fall Acceleration (Teacher s Guide) 2012 WARD S Science v.11/12 OVERVIEW Students will measure
Pressure calibration equipment
Pressure calibration equipment Pressure calibration evolution 1925 Westertoren in Amsterdam 30 meter mercury column (±40 bar) 1950 comparison test pump 21 st century Stiko gas driven differential deadweight
PPB Dissolved Oxygen Measurement - Calibration and Sampling Techniques
PPB Dissolved Oxygen Measurement - Calibration and Sampling Techniques Introduction The amount of dissolved oxygen in process water is continually gaining importance in many industries as a critical parameter
Errors Related to Cable Resistance Imbalance in Three Wire RTDs
Errors Related to Cable Resistance Imbalance in Three Wire RTDs 1.0 Introduction There are multiple sources of error that can impact the accuracy of an RTD measurement. The cable incorporated into the
Eurovacuum EV Series. Single Stage Oil Sealed Rotary Vane Pump
Eurovacuum EV Series Single Stage Oil Sealed Rotary Vane Pump Eurovacuum EV Series Single Stage Oil Sealed Rotary Vane Pumps Eurovacuum Company is offering its various products to meet industrial vacuum
Tadahiro Yasuda. Introduction. Overview of Criterion D200. Feature Article
F e a t u r e A r t i c l e Feature Article Development of a High Accuracy, Fast Response Mass Flow Module Utilizing Pressure Measurement with a Laminar Flow Element (Resistive Element) Criterion D200
PSAC/SAIT Well Testing Training Program Learning Outcomes/Objectives Level B
s/objectives Level B CHEM 6014: Properties and Characteristics of Natural Gases Discuss the pertinent properties and uses of natural gases as they occur in gas processing. 1. Describe the physical properties
Air Eliminators and Combination Air Eliminators Strainers
Description Air Eliminators and Combination Air Eliminator Strainers are designed to provide separation, elimination and prevention of air in piping systems for a variety of installations and conditions.
RI-215A Operator s Manual. Part Number: 71-0045RK Revision 0 Released: 10/3/05
RI-215A Operator s Manual Part Number: 71-0045RK Revision 0 Released: 10/3/05 Warranty RKI Instruments, Inc., warrants gas alarm equipment sold by us to be free from defects in materials and workmanship,
Industrial Steam System Process Control Schemes
Industrial Steam System Process Control Schemes This paper was developed to provide a basic understanding of the different process control schemes used in a typical steam system. This is however a fundamental
AVL FUEL MASS FLOW METER & FUEL TEMPERATURE CONTROL
FUEL CONSUMPTION MEASUREMENT AVL FUEL MASS FLOW METER & FUEL TEMPERATURE CONTROL Description The combination of AVL Fuel Mass Flow Meter and AVL Fuel Temperature Control is a high precise and continuous
FULL ELECTRICAL LNG PLANTS: HIGHEST AVAILABILITY AND ENERGY EFFICIENCY THROUGH OVERALL SYSTEM DESIGN
FULL ELECTRICAL LN PLANTS: HIHEST AVAILABILITY AND ENERY EFFICIENCY THROUH OVERALL SYSTEM DESIN Dr. Edwin Lerch Siemens A Infrastructure and Cities Sector, IC S SE PTI, ermany Phone: 49-9131-7-34052 Fax: