Guardian A Wearable Medical Alert System
|
|
|
- Brooke Simmons
- 10 years ago
- Views:
Transcription
1 Guardian A Wearable Medical Alert System A Design Report Presented to the Engineering Division of the Graduate School of Cornell University in Partial Fulfillment of the Requirements for the Degree of Master of Engineering, Electrical and Computer Engineering Submitted by Eileen Liu Advised by Dr. Bruce Land Date: December 2014
2 ebl43 Page 2 of 28 Abstract Master of Electrical Engineering Program Cornell University Design Project Report Project Title: Guardian: A Wearable Medical Alert System Author: Eileen Liu Abstract: Guardian is a medical alert system in the form of a wearable device that monitors rapid shifts in blood pressure and acceleration to automatically detect when a user has fallen. When a fall is detected, the device will connect to a mobile device to call for emergency help. The focus of the design project is to develop the blood pressure subsystem, which is what differentiates Guardian from existing alert systems for seniors. This subsystem is comprised of two pulse plethysmography units that connect with an ATMega1284P microcontroller to continuously measure user blood pressure based on pulse wave transit time. The microcontroller measures the pulse wave transit time between the two pulse plethysmographs and applies an experimentallyderived regression model to calculate blood pressure. Rapid shifts in pressure indicate shifts in body orientation, suggesting the user has fallen. The goal of this project is to develop this optical blood pressure monitoring system and to see if it is a plausible addition to existing wearable medical alert systems.
3 ebl43 Page 3 of 28 Executive Summary This design project was pursued to investigate more reliable means for fall detection among seniors. The goal of the project is to create an optical blood pressure monitoring system through a microcontroller based design. This system is also assessed to see if it could be beneficial to existing medical alert systems. The system should be noninvasive, robust, and reliable to be used commercially. The complete design uses two pulse plethysmographs placed at two different locations along the anterior side of a user s forearm, spaced five inches apart. Each pulse plethysmograph is comprised of a 940nm wavelength infrared LED and an OPT101 monolithic photodiode. When a pulse wave passes under a photodiode, a voltage peak is generated. The analog voltage outputs from each photodiode feed into analog-to-digital converter pins on an ATMega1284P microcontroller. The microcontroller implements peak detection by checking for values above a defined threshold. After defining where the peaks are, the microcontroller measures the amount of time it takes for a pulse wave to travel between the two points on the user s arm. This time measurement is then used in a regression model to convert the pulse wave travel time to a blood pressure reading. Overall, the cost of materials was less than $22 and the system showed potential results for optical blood pressure monitoring on a user s forearm.
4 ebl43 Page 4 of 28 Table of Contents Introduction... 6 Background... 6 Existing Products... 7 Design Problem... 7 System Requirements... 8 Range of Solutions... 8 System Design... 9 Theory of Operation Pulse Wave Transit Time Pulse Plethysmography Background Math Blood Pressure Subsystem High Level Design Blood Pressure Subsystem Hardware Design Sensing Subsystem Filtering and Amplification Subsystem Microcontroller Hardware Testing Blood Pressure Subsystem Software Design Analog-to-Digital Converter Peak Detection Timing Hardware and Software Tradeoffs Results Speed of Execution Accuracy Safety Interference with Other Designs Usability Conclusions... 20
5 ebl43 Page 5 of 28 Work Accomplished Future Work Standards Ethical Considerations References Data Sheets Appendices Cost and Budget Circuit Schematic Code Listing List of Figures Figure 1. Guardian high level system design... 9 Figure 2. Guardian system form factor Figure 3. Illustration of PWTT between ECG and pulse plethysmograph Figure 4. Principle of reflectance pulse plethysmography, illustrating the optical sensor and the different layers in the skin Figure 5. Breadboard prototype layout Figure 6. User arm over photodiodes on breadboard Figure 7. ATMega1284P microcontroller on custom PCB Figure 8. Regression model relating systolic blood pressure to pulse wave travel time List of Tables Table 1. Behavioral description template for "user wears device for fall detection" use case Table 2. PWTT and systolic blood pressure (SBP) measurements for sitting, supine, and standing orientations Table 3. Project schedule... 21
6 ebl43 Page 6 of 28 Introduction Background Older adults are at a higher risk for many types of injuries that can lead to death or disability. According to the Centers for Disease Control and Prevention, falls are the leading cause of injuries and death for Americans aged 65 years and older. 1 Each year, about 35% to 40% of adults 65 and older fall at least once. 1 With a population of roughly 43 million people in this age demographic, this equates to more than 17 million falls a year. 2 Medical alert systems were first introduced in the 1970s as basic push-button devices worn around the neck. 3 These devices called for help by transmitting a signal a base station that was connected to a home phone line that could then alert a call-center operator. However, these alert systems require the user to press a button when they have fallen. Only when the user has pressed this button will emergency services be contacted. 4,5 There are often situations where the user is incapacitated and unable to reach for this button when they have fallen. This has led to Fall Detection technology used in newer medical alert systems. 6 This technology uses accelerometers and gyroscopes to analyze changes in a body s position for fall detection. While this automated fall detection seems to be an ideal solution, this detection has limited functionality and issues with reliability. 7 Fall detection devices must be sensitive enough to separate everyday activities from falls. While newer medical alert systems have implemented rough algorithms for this detection, the technology must still be refined to catch edge cases, such as falls caused by dizziness, users sliding down from chairs to the floor, or when the user grips an object during the fall. False alarms could also be triggered by rapid movements by the user, inconveniencing both the user and the medical system dispatchers. The proposed system incorporates continuous blood pressure monitoring as a means to supplement accelerometer and gyroscope readings for more robust fall detection. According to the American Heart Association, a sudden decline in blood pressure can cause dizziness or fainting. 8 Changes in blood pressure can also indicate changes in body orientation. Using this information, a more robust medical alert system can be implemented. Papers have shown that systolic blood pressure can be measured using pulse wave transit time (PWTT), which is the time it takes for blood to reach an extremity from the heart. 9 The idea is that blood pressure affects the amount of time it takes for blood to travel through arteries. By creating a linear regression model to establish the relationship between PWTT and measured blood pressure, a system can be calibrated to measure blood pressure by applying pulse plethysmography at two different points on a user s body.
7 ebl43 Page 7 of 28 Existing Products The creators of current medical alert systems emphasize that their products will aid users who might have fallen and cannot stand back up. The detection portion of these existing system designs can be divided into two categories: 1. simple push button 2. push button with motion-sensitive fall detection technology Systems such as Life Alert and Bay Alarm Medical require the user to press a button to indicate that they have fallen. 4,5 However, the user is sometimes incapacitated when they have fallen and will not be able to reach for the button. Improved systems, such as Alert-1 include fall detection via readings from a gyroscope and accelerometer. 6 While this design is an improvement on the simple push-button design, it has limited functionality and has issues with reliability. To resolve the issues with present day solutions, the inclusion of a blood pressure monitoring subsystem is explored as a supplement to the fall detection technology. Design Problem The goal of this project is to design a proof of concept system for a continuous optical blood pressure monitoring system that can supplement existing fall detection methods. It is only one component of a complete medical alert system, named Guardian. Guardian is a wearable medical alert system that monitors rapid shifts in blood pressure and acceleration to automatically detect when a user has fallen. When a fall is detected, the device will connect to a mobile device to call for emergency help. The pulse wave travel times measured by the blood pressure subsystem should be compared to blood pressure as captured by a commercially available blood pressure cuff to create a regression model, establishing a relationship between pulse wave travel time and blood pressure. This deliverable will show how blood pressure changes can indicate changes in body orientation. Future work will be necessary to incorporate the accelerometer, gyroscope, and alert system to create a complete system. The complete system will be particularly useful for seniors who might be living alone and have a high risk of falling and becoming unconscious. By having an automated alert system, seniors will be able to live safely at home with a feeling of security that emergency help will come if needed.
8 ebl43 Page 8 of 28 System Requirements Since the product is designed for seniors, a main priority is to keep the device easy-to-use with minimal maintenance. The system requirements are as following: - the system shall be low cost (< $50) - the system shall be compact and worn on the forearm - the system shall be useable by multiple users A main constraint for this project is that it is to be executed by one person within three credits. Because of this, many tasks were re-scoped unlike the original plan, multiple design iterations were not possible and data collection to create the regression model was limited. Range of Solutions A range of solutions were explored before selecting the current design for blood pressure monitoring. These solutions included different methods to measure pulse wave travel time. The pulse wave travel time could have been calculated by the time difference between an electrocardiogram (ECG) and a pulse plethysmograph on the wrist. This method requires electrodes on both sides of the heart for the ECG. Another option was to calculate pulse wave travel time between the index fingers or wrists of both arms, using bluetooth modules to wirelessly communicate between the disjoint areas. While both these solutions could have rendered more accurate blood pressure readings, both options are bulky and would be inconvenient to use on a daily basis. The primary objective for this subsystem is to detect rapid changes in blood pressure to aid in fall detection. Thus, the precision of the blood pressure reading is not as important as detecting the changing blood pressure. For the purposes of this subsystem, it is a fair tradeoff to aim for a compact device with less accuracy. To achieve a more compact form factor, the current design implements a sleeve form factor with embedded pulse plethysomgraphy sensors spaced five inches apart. This sleeve will measure pulse wave transit time along a user s forearm. This design, which is confined to a single forearm, is less intrusive to a user s daily life.
9 ebl43 Page 9 of 28 System Design The full fall detection system is divided into multiple subsystems. While the full system is defined in Figure 1, the focus for this project is solely the blood pressure subsystem. Fall Subsystem Alert Subsystem Blood Pressure Subsystem Figure 1. Guardian high level system design A fall is defined when the accelerometer and gyroscope first detect a rapid change in acceleration or when the blood pressure subsystem detects a rapid change in blood pressure. When one subsystem is triggered, the microcontroller unit senses this change and checks the other subsystem to confirm if a fall was also detected. If a fall is confirmed by both the fall and blood pressure subsystems, the microcontroller provides an audio and visual indication that a fall has been detected. If the user does not disable the system within five seconds, the microcontroller will automatically connect to a mobile device via a bluetooth module to contact emergency services. The following table is a behavioral description template (BDT) for the device s primary use case: when the user wears the device for fall detection. The table demonstrates in detail how the device subsystems interact with one another and the user
10 ebl43 Page 10 of 28 Table 1. Behavioral description template for "user wears device for fall detection" use case Initial Condition: device is on and at rest User Gyroscope/Acc elerometer Subsystem Blood Pressure Subsystem wears Guardian on forearm falls fall detected rapid shift in blood pressure Microcontroller measures changing pulse wave travel time confirms fall detected 5 second time delay for user to turn off system if false positive sends bluetooth transmission to trigger alert system Alert Subsystem system contacts call center operator for emergency services The following diagram shows the form factor of the full Guardian system on a user s arm. = pulse plethysmograph unit = gyroscope/accelerometer = microcontroller unit = sleeve Theory of Operation Figure 2. Guardian system form factor Pulse Wave Transit Time Pulse wave transit time (PWTT) is the amount of time it takes for a pulse pressure wave to propagate through a length of the arterial tree. 10 When the heart pumps blood from the left ventricle into the aorta, it creates a pressure wave that travels along the arteries ahead of the pumped blood itself. This pressure wave travels at a speed of a few meters per second.
11 ebl43 Page 11 of 28 While PWTT is typically measured using readings from an electrocardiogram (ECG) and a peripheral pulse wave via a pulse plethysmograph, it can also be measured between two points on a user s body with two pulse plethysmographs. Figure 3. Illustration of PWTT between ECG and pulse plethysmograph Because the speed of this pressure wave depends on the tension of the arterial walls, PWTT correlates with a user s blood pressure. The higher the user s blood pressure, the faster the pressure wave travels, resulting in a shorter PWTT. The lower the user s blood pressure, the slower the pressure wave travels, resulting in a longer PWTT. Using this knowledge, a linear regression model can be created for a correlation between PWTT and blood pressure. Pulse Plethysmography Pulse plethysmography is a method used to measure a user s heart rate. It is a commonly used method because it is noninvasive, inexpensive to implement, and can be very compact. When the heart beats, a pressure wave moves away from the heart along the. This pressure wave creates an increase in blood volume in tissue wherever the wave is. Using an infrared LED and photodiode, this volumetric change in blood from the pressure wave can be detected. The infrared light is directed into a region of tissue with a paired photodiode pointed next to that region. Infrared light is used because it is relatively unaffected by changes in arterial oxygen saturation. Some of the light is absorbed by the tissue and deoxyhemoglobin while the rest of the light is reflected back. The photodiode detects the amount of infrared light that is reflected back from the tissue and outputs an analog voltage that linearly corresponds to the amount of infrared light detected. When the pulse wave reaches the region that the LED and photodiode are pointed at, the reflectivity of the tissue increases, causing an increase in voltage output from the photodiode. The composition of surrounding tissues remains the same as blood volume changes, which can add a DC component to the photodiode output.
12 ebl43 Page 12 of 28 Figure 4. Principle of reflectance pulse plethysmography, illustrating the optical sensor and the different layers in the skin 11 Peaks are detected in this reflectivity measurement when the pulse wave is below the photodetector, a peak will occur. Thus, using pulse plethysomgraphy at different locations on the user s body can help monitor the location and speed of the pulse wave. Background Math Moens-Korteweg equation is describes the relation between blood pressure and pulse wave velocity (PWV, which is inversely correlated to PWTT). 12 It assumes that the pulse wave velocity in a short elastic vessel is obtained from its geometric and elastic properties as defined in the following equation: (1) PWV = distance PWTT = Eh ρ2r where E = elasticity modulus of the vessel wall, h = thickness of vessel wall, ρ =density of blood, r = radius of the vessel. Blood pressure and PWV are then correlated using the Hughes equation: (2) E = E 0 e αp where α mmhg -1 and P = mean arterial pressure. Along, each of the parameters in both equations are difficult to measure. However, calibration functions can be derived by combining these equations to translate PTT to blood pressure, assuming constant vessel thickness and radius. After combining the two equations, the following relationship is derived: (3) Blood Pressure = A ln PTT + B
13 ebl43 Page 13 of 28 where A and B are constant factors that need to be experimentally derived. These constants can be derived by simultaneously measuring PWTT with the designed system and measuring blood pressure with a commercial blood pressure cuff. After plotting this data on the same set of axes, a regression model of the form of equation 3 can be fit to the data. This derived regression model can then be used to determine the relationship between blood pressure and PWTT. Blood Pressure Subsystem High Level Design The blood pressure subsystem uses two pulse plethysmographs placed at two different locations along the anterior side of a user s forearm, spaced five inches apart. One pulse plethysmograph is positioned on the anterior side of the user s forearm closer to the wrist while the other plethysmograph is positioned on the anterior side closer to the antecubital fossa (the area on the arm opposite the elbow). Each pulse plethysmograph is comprised of a 940nm wavelength infrared LED and an OPT101 monolithic photodiode. The infrared light is pointed toward the anterior side of the user s forearm next to the OPT101 which measures the amount of infrared light reflectance from the user. When a pulse wave reaches the tissue beneath the photodiode, the increased blood saturation increases the amount of light that is reflected to the photodiode, creating an analog voltage peak with the pulse wave. The analog voltage outputs from each photodiode feed into analog-to-digital converter (ADC) pins on an ATMega1284P microcontroller. The microcontroller implements peak detection by checking for values above a defined threshold. After defining where the peaks are, the microcontroller measures the amount of time it takes for a pulse wave to travel between the two points on the user s arm. This time measurement is then used in a regression model to convert the pulse wave travel time to a blood pressure reading. Blood Pressure Subsystem Hardware Design In the blood pressure subsystem, the main components are a sensing subsystem, a filtering and amplification subsystem, and a microcontroller unit. While the final form factor of the Guardian system will be a sleeve worn on the forearm, the current prototype resides on a breadboard where the user lays his or her arm above the sensors. For a full circuit schematic, refer to the appendix.
14 ebl43 Page 14 of 28 Figure 5. Breadboard prototype layout The main sensing element for each pulse plethysmograph is the OPT101 sensor, which includes a linear light intensity to voltage conversion and a unity gain buffer. A single infrared LED at 940nm wavelength is used in each pulse plethysmograph to inject light into the region the OPT101 is monitoring. The signal from each sensor is buffered, filtered (both actively and passively), and amplified with two gain stages of 50 each before the signals are passed into ADC pins on the microcontroller unit. Sensing Subsystem Each sensing subsystem is comprised of a LTE4208 infrared LED at 940nm wavelength and an OPT101 monolithic photodiode. This particular photodiode was selected for this design because it includes a photodiode and a transimpedance amplifier to output voltage linearly with photocurrent. The configuration of the LED and photodiode have a significant impact on the results. In this setup, the LED is positioned next to the photodiode, bent at an angle of about 120 degrees to direct its light toward the sensor. Both the LED and photodiode point toward the user s forearm.
15 ebl43 Page 15 of 28 Figure 6. User arm over photodiodes on breadboard Filtering and Amplification Subsystem Three types of filters are used in this design. To remove the DC offset from amplification, the signal from the OPT101 is first passed through a passive low pass filter with a cutoff frequency of 10 Hz before the input to the first amplification stage. An active bandpass filter/amplification stage as well as several passive low pass filters are then used to remove 100 Hz noise from overhead lights. The bandpass filter only allows frequencies between 1 Hz and 8 Hz to pass through. Typical human heart rates range between 30 to 180 beats per minute the 1 Hz high pass removes low frequency noise from user movements while the 8 Hz low pass removes high frequency noise from ambient light and power lines. Two gain stages with 30 gain each are used to increase the input signal to the microcontroller for easier peak detection. The gain stage is split into 2 op amps because too much gain in a single stage saturated the signal. Microcontroller The microcontroller used for the Guardian system is the ATmega1284P mounted on a custom printed circuit board. The design uses an ATmega1284P for its wide range of peripherals and to accommodate future integration of the fall detection subsystem. The blood pressure monitoring system uses two ADC input pins and a timer on the microcontroller. Future design iterations could implement the low-power mode of the microcontroller or use a lower powered CPU, such as the ATTiny85. Figure 7. ATMega1284P microcontroller on custom PCB
16 ebl43 Page 16 of 28 Hardware Testing To verify proper functionality of the circuit, each pulse plethysmograph circuit was systematically tested. Testing was performed on each individual filter and gain stage within the circuit by using a function generator as an ideal source. After ensuring each stage behaved as expected individually, stages were incorporated one by one to ensure proper hardware functionality. Blood Pressure Subsystem Software Design The main tasks of the microcontroller software include acquiring data from the photodiodes via ADC pins A.0 and A.1, processing the acquired data to extract PWTT measurements in real time, and converting the PWTT measurements into blood pressure readings. While this software was written, it was not implemented in conjunction with the hardware due to reliability issues in the hardware. Analog-to-Digital Converter The ADC is initialized using VCC as the voltage reference. The ADC value is then left aligned in the data registers to for 8-bit measurements rather than 10-bit. After these initializations, the ADC is enabled with a prescaler of 1. The analog signals from each pulse plethysmograph are passed into ADC pins A.0 and A.1 after going through filtering and amplification. These signals are converted into digital values ranging from 0 to 255. Peak Detection Peaks for the pulse wave are currently detected in software using thresholding. If the current ADC value is above a predetermined value, the system recognizes the data to be a peak. This simplified method is susceptible to noise and drifting analog signals due to changes in body orientation or LED positioning. Thus, peak detection should be performed by saving a buffer of data from the last ten seconds. The data should first be checked to see if it is valid. To do this, the number of detected major peaks in this buffer should be checked. The standard deviation of time between each peak should be used to determine if the signal is periodic and the heart rate in beats per minute (bpm) should be checked to see if it falls within typical human bpm rates ( bpm). After a valid signal is established, the DC component of the signal should be removed by taking a fast fourier transform of the signal and removing the zero frequency component. An inverse fast fourier transform should then be performed. Peaks that occurred due to noise should then be removed by rejecting all peaks less than 25% of the amplitude of the maximum peak in the buffer. This creates a dynamic threshold for peak detection that constantly changes with the current data, which is more accurate than peak detection by simple thresholding.
17 ebl43 Page 17 of 28 Timing Timer 0 is set up as a 1 millisecond time base for the system. This is done by enabling Timer 0 compare match ISR and enabling the clear-on-match. In the Timer 0 ISR, the subroutine continuously increments the counter time1. When a peak is detected for the pulse plethysmograph closer to the elbow, the time variable is set to zero. This variable is incremented in the ISR for every passing millisecond. When the peak is detected for the pulse plethysmograph closer to the wrist, the milliseconds that have passed is recorded. This value is the PWTT that can then be used to calculate the user s blood pressure. Hardware and Software Tradeoffs There were a couple of signal processing components that could have been performed in either hardware or software. In general, anything that could reasonably be performed in software was calculated in software to minimize the size of the device. For example peak detection was performed in software to minimize the amount of hardware used. In hardware, this detection could implemented using a comparator. On the contrary, the signal filtering was performed in hardware rather than in software. To perform digital filtering, data would have to be stored in memory. The ATMega1284P has 16kB of RAM, which could limit the processing capabilities of the microcontroller which could slow down the software. With such a short distance between the two plethysmographs, it was more important to keep the software lean and quick. Thus, analog filtering was used. Results The system was tested on a single user with the following characteristics: female, age 22, height 5 7. Using the designed blood pressure subsystem, PWTT s were measured in various user orientations: standing, sitting, and supine positions. Five data points were taken for each user orientation. PWTT s for this plot were measured using a Tektronix oscilloscope and were compared to blood pressure readings measured by an Omron Healthcare s digital wrist blood pressure monitor model HEM-608.
18 Systolic Blood Pressure [mm Hg] ebl43 Page 18 of 28 Table 2. PWTT and systolic blood pressure (SBP) measurements for sitting, supine, and standing orientations Position PWTT [ms] SBP [mm Hg] sitting supine standing The data in Table 2 show a higher systolic blood pressure reading from the commercial blood pressure cuff when standing and a lower systolic blood pressure reading when laying down. This trend supports the idea that changes in blood pressure can indicate changes in body orientation which could be useful in detecting when a user falls. However, the data has a range of 22 mmhg which is a narrow window. More data should be collected to develop a stronger relationship between PWTT and blood pressure Regression Model Relating Systolic Blood Pressure to Pulse Wave Travel Time 0 y = -13.0ln(x) R² = Pulse Wave Travel Time [ms] Figure 8. Regression model relating systolic blood pressure to pulse wave travel time
19 ebl43 Page 19 of 28 The best fit regression model for the data was: blood pressure = ln(pwtt) with an R 2 value of 0.109, indicating poor correlation between the two variables. In Proenca et al 12, a best equation of blood pressure = -92 ln(pwtt) - 29 was derived. While these derived constants are very different from the values derived from Figure 8, Proenca et al 12 have high variability in their values with a +/- factors of up to 333. Speed of Execution If the software were implemented with the hardware, the speed of execution would be limited by ADC conversion rate, which for an ATMega1284P is about 300μs. This is fast enough for the system to measure PWTT readings since travel times are on the scale of milliseconds. Accuracy The data collected for the regression model was very limited. The model was based on fifteen measurements taken from one female user with a defined height and age. These measurements were only taken in three positions: sitting, supine, and standing. Because of this, the range for measured systolic blood pressure is very small a window of only 22 mm Hg. This is a small enough range and sample size of data that the accuracy of the collected data is uncertain. This regression model could be improved by increasing the range for systolic blood pressure by elevating blood pressure through exercise or lowering blood pressure through squeezing the brachial artery. The regression model could also be improved by collecting data from more users with greater diversity in demographics. The distance between the two pulse plethysmographs is a very short distance of five inches. PWTT is typically measured between a user s heart and finger. This shorter distance for the pulse wave to travel can amplify the effects of any confounding components that can decrease the system accuracy. Another component that has a significant effect on the accuracy of the blood pressure reading is the positioning of the sensors and infrared LED s on the user s arm. With flexible leads, the positions of these components can change each time the system is used. Thus, the collected PWTT readings might not have been based on the same positioning of LED s. On top of this, the reported testing results were performed in a stationary setting. Realistically, a user would move his or her arms while wearing the device. This arm movement could affect the system readings and the blood pressure itself. This could result in movement artifacts that could cause false peak detection and decreased accuracy. The amount of tension in the muscle under the photodiode can also affect the reflectance of infrared light. Overall, there are multiple causes for inaccurate readings. The best way to overcome these factors in the given system design is to increase the amount of data collection to decrease the impact of inaccurate readings.
20 ebl43 Page 20 of 28 Safety The safety concerns for this project are minimal due to the noninvasive nature of optical methods. The most dangerous components within the design are the low power infrared LED s. According to the International Commission on Non-Ionizing Radiation Protection (ICNIRP) standards on infrared radiation, direct eye exposure to small infrared sources for more than 0.25 seconds at a time should be limited to 1.9E07 W*m -2 *sr -1 to prevent thermal injury in the cornea. 13 Users will be warned against looking directly into the infrared LED s. The LED s will also be embedded in a sleeve, so users will not be looking at the LED s directly. To prevent thermal injury to the skin, the heat should be limited 2*t 3/4 J*cm -2. Assuming the cross sectional area of skin exposure to the infrared LED is about 1cm 2 and that 80mW of power is transmitted from the infared LED with ideal transmission over thirty seconds, heat should be limited to a maximum of 2.4 J/cm 2. Interference with Other Designs The blood pressure subsystem itself does not have any components that could interfere with other devices. However, the addition of the bluetooth module for connecting to mobile devices has potential to interfere with other devices. This would be unlikely as long as the integrated bluetooth module conforms to standards outlined by the Bluetooth Special Interests Group, it should not interfere with other devices. Usability Since the device is meant to be used by seniors, usability is a key component. The prototype in its current form factor on a breadboard is not portable for a user. However, the complete Guardian design will be portable and easy to use. With an elastic sleeve design that contains all subsystems on a single forearm, the device should be easily portable and should not interfere with a user s daily habits. If enough user testing is performed to create a robust platform of regression models that can be tailored to the user s age, height, weight, and gender, the device should be able to be used for all users. The concerns for the device usability are potential false positives that could be triggered by the user swinging his or her arms. Another concern is system sensitivity to the infrared LED orientation in comparison to the photodiode. On the breadboard, it was very difficult to consistently place the LED s in an orientation that could produce usable data. Thus, a different form factor that is worn on the torso should be considered. The torso is likely to have less movements in comparison to a user s arm. Conclusions This optical blood pressure monitoring system was built to supplement existing methods of fall detection using gyroscopes and accelerometers for fall detection. To enhance existing methods,
21 1-Sep 8-Sep 15-Sep 22-Sep 29-Sep 6-Oct 13-Oct 20-Oct 27-Oct 3-Nov 10-Nov 17-Nov 24-Nov 1-Dec 8-Dec 15-Dec ebl43 Page 21 of 28 the blood pressure monitoring system must first be reliable on its own. The project shows potential to be integrated in future systems for increased fall detection reliability. However, further testing needs to be conducted. Regression models need to developed for various users based on age, height, weight, and gender for the product to be applicable for a wide range of users. This will require abundant data collection. Although the current design was favored over alternate designs in the beginning of the project, different form factors should be explored for the design to be more robust. With the current design, the positioning of the infrared LED s and photodiodes can be inconsistent, which can have great impact on readings. Additionally, users tend to move their arms around even while stationary. This effect can cause false readings. A form factor on the torso area, which experiences less movement, should be explored. Work Accomplished The main tasks that were completed are summarized in Table 3. A prototype for continuous optical blood pressure monitoring was developed. An initial linear regression model was also developed to correlate blood pressure with pulse wave transit time as measured by the system. Table 3. Project schedule September October November December Task Brainstorm project ideas Decide on project and write abstract Perform research on similar products Develop high level system design Write project proposal Build blood pressure subsystem prototype Develop linear regression Determine effectiveness of method Write final report Future Work The reliability of the blood pressure monitoring system could be enhanced by pursuing different form factors. One option is to measure pulse wave transit time by applying electrocardiography
22 ebl43 Page 22 of 28 and pulse plethysmography on a user s deltoid muscle, which has shown potential for providing ECG data. This form factor would be a sleeve worn on the user s upper arm as opposed to the user s lower arm. Another option would be to measure PWTT by applying electrocardiography and pulse plethysmography across a user s chest in the form factor of a band that could be worn under a user s shirt. More importantly, to measure blood pressure more accurately, more data needs to be collected. This data needs to span a wider demographic of users in a greater variety of use cases. Because it is difficult to artificially decrease blood pressure, the use cases should aim to increase blood pressure to create a greater window of data for the linear regression model. Such use cases could include measuring PWTT after rigorous exercise or while squeezing the brachial artery. Machine learning could also be applied to help reduce the amount of data collection required to develop robust regression models. Standards The C code embedded in the microcontroller complies with the C language standards set by the American National Standards Institute and the IEEE Standard System Language. Ethical Considerations During the development process of this prototype design, the IEEE Code of Ethics was followed. Conscious design choices were made to ensure safety for the public and for myself while developing a quality product. While constructing the system, I constantly checked for short circuits, corrupt components, and made sure all electrical components were properly insulated. Another aspect of ethics that was considered was the trust that seniors would have in this device. By advertising this device to be supplementary to existing methods, users would expect this device to perform more reliably. Thus, the final system design will include the same basic functionality as existing devices on the market, including a push button and accelerometer/gyroscope fall technology.
23 ebl43 Page 23 of 28 References 1. Injuries Among Older Adults. Centers for Disease Control and Prevention, 9 Oct Web. 10 Nov < 2. Older Persons Health. Centers for Disease Control and Prevention, 14 July Web. 10 Nov < 3. What to look for in a medical alert system. ConsumerReports.org, July Web. 10 Nov < 4. Life Alert, Web. 10 Nov < 5. Bay Alarm Medical, Web. 10 Nov < 6. Fall Detection Technology. Alert-1, Web. 10 Nov < 1.com/content/fall-detection-technology/1390>. 7. Automatic Fall Detection Systems. Bay Alarm Medical, 10 June Web. 10 Nov < 8. Low Blood Pressure. American Heart Association, 3 Sept Web. 10 Nov < AboutHighBloodPressure/Low-Blood-Pressure_UCM_301785_Article.jsp>. 9. Meigas, K., Kattai, R., Lass J. Continuous Blood Pressure Monitoring Using Pulse Wave Delay. 25 Oct PWTT: How it Works. Nihon Kohden. Web. 10 Nov < 11. Mendelson, Y., Ochs, B. Noninvasive Pulse Oximetry Utilizing Skin Reflectance Photoplethysmography, IEEE Transactions on Biomedical Engineering, 35(10): pp , Oct Proenca, J., Muehlsteff, J., Aubert X., Carvalho, P. Is pulse transit time a good indicator of blood pressure changes during short physical exercise in a young population? 32nd Annual International Conference of the IEEE EMBS, pp , Sept ICNIRP guidelines on limits of exposure to incoherent visible and infrared radiation, Health Physics, 105(1), <
24 ebl43 Page 24 of 28 Data Sheets Infrared LED LTE Monolithic Photodiode OPT ATMega1284P Microcontroller Omron Healthcare digital wrist blood pressure monitor model HEM-608
25 ebl43 Page 25 of 28 Appendices Cost and Budget Part Vender Cost/Unit Quantity Total Cost Infrared LED LTE-4208 Digikey $ $1.00 Monolithic Photodiode OPT-101 Digikey $ $12.74 LM358 Op Amp Digikey $ $1.00 ATMega1284P already owned $ $5.00 Capacitors lab stock $ $2.00 Resistors lab stock $ $0.06 Total Cost $21.80 Circuit Schematic
26 ebl43 Page 26 of 28 Code Listing #include <util/delay.h> #include <stdio.h> #define F_CPU UL //frequency of CPU crystal #define PP1_thresh 100 //peak detection threshold #define PP2_thresh 150 //peak detection threshold #define k //scaling factor for regression model #define k //offset factor for regression model volatile unsigned char time1; //time counter for PWTT volatile unsigned char PWTT; //pulse wave transit time volatile unsigned float BP; //blood pressure // timer0 overflow ISR - used to increment a timer, keep a millisecond // time base ISR (TIMER0_COMPA_vect) { //Increment the time time1++; } // Reads adc value at pin ch // Code borrowed from ADC tutorial: // ch which pin (which transceiver) to read: 0=PP1, 1=PP2 adc-converted reading uint8_t adc_read(uint8_t ch) { // select the corresponding channel 0~7 // ANDing with '7' will always keep the value // of 'ch' between 0 and 7 ch &= 0b ; //AND operation with 7 ADMUX = (ADMUX & 0xF8) ch; //clears the bottom 3 bits before ORing // start single conversion - write '1' to ADSC ADCSRA = (1<<ADSC); // wait for conversion to complete // ADSC becomes '0' again // till then, run loop continuously while(adcsra & (1<<ADSC)); } return (ADCH); // initialize adc
27 ebl43 Page 27 of 28 void adc_init() { // AREF = AVcc ADMUX = (1<<REFS0); ADMUX = (1<<ADLAR); // left align ADC value in data registers // (turn measurement from 10 bits to 8 bits) } // ADC Enable and prescaler of 1 ADCSRA = (1<<ADEN); // Set up the MCU void initialize(void) { // set up timer 0 for 1 msec timebase TIMSK0 = (1<<OCIE0A); //turn on timer 0 cmp match ISR OCR0A = 249; //set the compare reg to 250 time ticks TCCR0A = (1<<WGM01); // turn on clear-on-match time1 = 0; // init the task timer DDRA &= ~(0x03); // set a.0 and a.1 to input sei(); // crank up the ISRs } // Check for peak in pulse pleth 1 (closer to elbow) int checkpp1(void) { if (adc_read(0) > PP1_thresh) return 1; } // Check for peak in pulse pleth 2 (closer to wrist) void checkpp2(void) { while (adc_read(1) < PP2_thresh); //wait until peak detected PWTT = time1; } int main(void) { initialize(); adc_init(); // main task scheduler loop while(1) {
28 ebl43 Page 28 of 28 } } if (checkpp1() == 1) { time1 = 0; checkpp2(); BP = A*log(PWTT)+B; }
Efficient Heart Rate Monitoring
Efficient Heart Rate Monitoring By Sanjeev Kumar, Applications Engineer, Cypress Semiconductor Corp. Heart rate is one of the most frequently measured parameters of the human body and plays an important
Measuring Laser Power and Energy Output
Measuring Laser Power and Energy Output Introduction The most fundamental method of checking the performance of a laser is to measure its power or energy output. Laser output directly affects a laser s
WAITER: A Wearable Personal Healthcare and Emergency Aid System
Sixth Annual IEEE International Conference on Pervasive Computing and Communications WAITER: A Wearable Personal Healthcare and Emergency Aid System Wanhong Wu 1, Jiannong Cao 1, Yuan Zheng 1, Yong-Ping
Pressure Sensor Based Estimation of Pulse Transit Time
International Journal of Information & Computation Technology. ISSN 0974-2239 Volume 4, Number 13 (2014), pp. 1321-1328 International Research Publications House http://www. irphouse.com Pressure Sensor
AMPLIFIED HIGH SPEED FIBER PHOTODETECTOR USER S GUIDE
AMPLIFIED HIGH SPEED FIBER PHOTODETECTOR USER S GUIDE Thank you for purchasing your Amplified High Speed Fiber Photodetector. This user s guide will help answer any questions you may have regarding the
Hands On ECG. Sean Hubber and Crystal Lu
Hands On ECG Sean Hubber and Crystal Lu The device. The black box contains the circuit and microcontroller, the mini tv is set on top, the bars on the sides are for holding it and reading hand voltage,
Electrocardiogram (ECG) Monitoring System using Bluetooth technology
Electrocardiogram (ECG) Monitoring System using Bluetooth technology Zarina Md Amin, Suryani Ilias, Zunuwanas Mohamad Department of Electrical Engineering, Polytechnic of Sultan Abdul Salahuddin Abdul
Using Arduino Microcontrollers to Sense DC Motor Speed and Position
ECE480 Design Team 3 Using Arduino Microcontrollers to Sense DC Motor Speed and Position Tom Manner April 4, 2011 page 1 of 7 Table of Contents 1. Introduction ----------------------------------------------------------
If an occupancy of room is zero, i.e. room is empty then light source will be switched off automatically
EE389 Electronic Design Lab Project Report, EE Dept, IIT Bombay, Nov 2009 Fully-automated control of lighting and security system of a Room Group No: D2 Bharat Bhushan (06d04026) Sravan
AVR126: ADC of megaavr in Single Ended Mode. Introduction. Features. AVR 8-bit Microcontrollers APPLICATION NOTE
AVR 8-bit Microcontrollers AVR126: ADC of megaavr in Single Ended Mode APPLICATION NOTE Introduction Atmel megaavr devices have a successive approximation Analog-to- Digital Converter (ADC) capable of
Example #1: Controller for Frequency Modulated Spectroscopy
Progress Report Examples The following examples are drawn from past student reports, and illustrate how the general guidelines can be applied to a variety of design projects. The technical details have
AND9035/D. BELASIGNA 250 and 300 for Low-Bandwidth Applications APPLICATION NOTE
BELASIGNA 250 and 300 for Low-Bandwidth Applications APPLICATION NOTE Introduction This application note describes the use of BELASIGNA 250 and BELASIGNA 300 in low bandwidth applications. The intended
Logging of RF Power Measurements
Logging of RF Power Measurements By Orwill Hawkins Logging of measurement data is critical for effective trend, drift and Exploring the use of RF event analysis of various processes. For RF power measurements,
Arbeitskreis Hardware. Prof. Dr. Michael Rohs, Dipl.-Inform. Sven Kratz [email protected] MHCI Lab, LMU München
Arbeitskreis Hardware Prof. Dr. Michael Rohs, Dipl.-Inform. Sven Kratz [email protected] MHCI Lab, LMU München Schedule (preliminary) Date Topic (preliminary) 2.5. Introduction to embedded interaction,
Design and Development of a Wireless Remote POC Patient Monitoring System Using Zigbee
Design and Development of a Wireless Remote POC Patient Monitoring System Using Zigbee A. B. Tagad 1, P. N. Matte 2 1 G.H.Raisoni College of Engineering & Management, Chas, Ahmednagar, India. 2 Departments
The Heartbeat behind Portable Medical Devices: Ultra-Low-Power Mixed-Signal Microcontrollers
The Heartbeat behind Portable Medical Devices: Ultra-Low-Power Mixed-Signal Microcontrollers The proliferation of sophisticated yet affordable personal medical devices is transforming the health care industry,
Miniaturizing Photoplethysmography for Use in a Multifunctional Health Monitoring Device with Applications for Asthma Analysis
Miniaturizing Photoplethysmography for Use in a Multifunctional Health Monitoring Device with Applications for Asthma Analysis Brinnae Bent North Carolina State University Faculty Mentor: Alper Bozkurt
Android based Alcohol detection system using Bluetooth technology
For more Project details visit: http://www.projectsof8051.com/android-based-alcohol-detection-system-usingbluetooth-technology/ Code 1435 Project Title Android based Alcohol detection system using Bluetooth
Heartbeat Monitoring Alert via SMS
Heartbeat Monitoring Alert via SMS Warsuzarina Mat Jubadi Dept. of Electronics Engineering University Tun Hussein Onn Malaysia Batu Pahat, Johor, Malaysia Email: [email protected] Siti Faridatul Aisyah
Application Note Noise Frequently Asked Questions
: What is? is a random signal inherent in all physical components. It directly limits the detection and processing of all information. The common form of noise is white Gaussian due to the many random
Encoders for Linear Motors in the Electronics Industry
Technical Information Encoders for Linear Motors in the Electronics Industry The semiconductor industry and automation technology increasingly require more precise and faster machines in order to satisfy
ends of transmission line are used for relay operation [70]. Another type of
69 CHAPTER-4 TRANSMISSION LINE DIFFERENTIAL PROTECTION USING WIRELESS TECHNOLOGY 4.1 INTRODUCTION The probability of fault occurrence on the overhead lines is much more due to their greater lengths and
How to measure blood pressure using digital monitors
How to measure blood pressure using digital monitors What is high blood pressure? Blood pressure is the pressure exerted by circulating blood upon the walls of blood vessels. Systolic pressure is the blood
Automatic Wrist Blood Pressure Monitor with
Instruction Manual Automatic Wrist Blood Pressure Monitor with Model HEM-609 ENGLISH ESPAÑOL CONTENTS Before Using the Monitor Introduction...................................... 3 Important Safety Notes.............................
Copyright is owned by the Author of the thesis. Permission is given for a copy to be downloaded by an individual for the purpose of research and
Copyright is owned by the Author of the thesis. Permission is given for a copy to be downloaded by an individual for the purpose of research and private study only. The thesis may not be reproduced elsewhere
Frequency Response of Filters
School of Engineering Department of Electrical and Computer Engineering 332:224 Principles of Electrical Engineering II Laboratory Experiment 2 Frequency Response of Filters 1 Introduction Objectives To
Evaluation copy. Blood Pressure. Project PROJECT DESIGN REQUIREMENTS
Blood Pressure Project 9 Blood pressure is a measure of the fluid pressure within the circulatory system. This pressure is required to ensure the delivery of oxygen and nutrients to, and the removal of
Adding Heart to Your Technology
RMCM-01 Heart Rate Receiver Component Product code #: 39025074 KEY FEATURES High Filtering Unit Designed to work well on constant noise fields SMD component: To be installed as a standard component to
MSAN-001 X-Band Microwave Motion Sensor Module Application Note
1. Introduction HB Series of microwave motion sensor module are X-Band Mono-static DRO Doppler transceiver front-end module. These modules are designed for movement detection, like intruder alarms, occupancy
Fiber Optics. Integrated Photo Detector Receiver for Plastic Fiber Plastic Connector Housing SFH551/1-1 SFH551/1-1V
Fiber Optics Integrated Photo Detector Receiver for Plastic Fiber Plastic Connector Housing SFH551/1-1 Features Bipolar IC with open-collector output Digital output, TTL compatible Sensitive in visible
Sensors and Cellphones
Sensors and Cellphones What is a sensor? A converter that measures a physical quantity and converts it into a signal which can be read by an observer or by an instrument What are some sensors we use every
Normally-Off Technologies for
ASP-DAC 2014 Normally-Off Technologies for Healthcare Appliance Shintaro Izumi 1, Hiroshi Kawaguchi 1, Yoshikazu Fujimori 2, and Masahiko Yoshimoto 1 1 Kobe University, Kobe, Japan, 2 Rohm, Kyoto, Japan
Software Manual RS232 Laser Merge Module. Document # SU-256521-09 Rev A
Laser Merge Module Document # SU-256521-09 Rev A The information presented in this document is proprietary to Spectral Applied Research Inc. and cannot be used for any purpose other than that for which
Data Transmission. Data Communications Model. CSE 3461 / 5461: Computer Networking & Internet Technologies. Presentation B
CSE 3461 / 5461: Computer Networking & Internet Technologies Data Transmission Presentation B Kannan Srinivasan 08/30/2012 Data Communications Model Figure 1.2 Studying Assignment: 3.1-3.4, 4.1 Presentation
Electrocardiography I Laboratory
Introduction The body relies on the heart to circulate blood throughout the body. The heart is responsible for pumping oxygenated blood from the lungs out to the body through the arteries and also circulating
AN4368 Application note
Application note Signal conditioning for pyroelectric passive infrared (PIR) sensors Sylvain Colliard-Piraud Introduction Pyroelectric passive infrared (PIR) sensors are widely used in daily life. They
Conversion Between Analog and Digital Signals
ELET 3156 DL - Laboratory #6 Conversion Between Analog and Digital Signals There is no pre-lab work required for this experiment. However, be sure to read through the assignment completely prior to starting
Impedance 50 (75 connectors via adapters)
VECTOR NETWORK ANALYZER PLANAR TR1300/1 DATA SHEET Frequency range: 300 khz to 1.3 GHz Measured parameters: S11, S21 Dynamic range of transmission measurement magnitude: 130 db Measurement time per point:
Calibration and Use of a Strain-Gage-Instrumented Beam: Density Determination and Weight-Flow-Rate Measurement
Chapter 2 Calibration and Use of a Strain-Gage-Instrumented Beam: Density Determination and Weight-Flow-Rate Measurement 2.1 Introduction and Objectives This laboratory exercise involves the static calibration
DDX 7000 & 8003. Digital Partial Discharge Detectors FEATURES APPLICATIONS
DDX 7000 & 8003 Digital Partial Discharge Detectors The HAEFELY HIPOTRONICS DDX Digital Partial Discharge Detector offers the high accuracy and flexibility of digital technology, plus the real-time display
Ultrasound Distance Measurement
Final Project Report E3390 Electronic Circuits Design Lab Ultrasound Distance Measurement Yiting Feng Izel Niyage Asif Quyyum Submitted in partial fulfillment of the requirements for the Bachelor of Science
1. Learn about the 555 timer integrated circuit and applications 2. Apply the 555 timer to build an infrared (IR) transmitter and receiver
Electronics Exercise 2: The 555 Timer and its Applications Mechatronics Instructional Laboratory Woodruff School of Mechanical Engineering Georgia Institute of Technology Lab Director: I. Charles Ume,
LM 358 Op Amp. If you have small signals and need a more useful reading we could amplify it using the op amp, this is commonly used in sensors.
LM 358 Op Amp S k i l l L e v e l : I n t e r m e d i a t e OVERVIEW The LM 358 is a duel single supply operational amplifier. As it is a single supply it eliminates the need for a duel power supply, thus
Display Board Pulse Width Modulation (PWM) Power/Speed Controller Module
Display Board Pulse Width Modulation (PWM) Power/Speed Controller Module RS0 Microcontroller LEDs Motor Control Pushbuttons Purpose: To demonstrate an easy way of using a Freescale RS0K2 microcontroller
Accurate Measurement of the Mains Electricity Frequency
Accurate Measurement of the Mains Electricity Frequency Dogan Ibrahim Near East University, Faculty of Engineering, Lefkosa, TRNC [email protected] Abstract The frequency of the mains electricity supply
Section 3. Sensor to ADC Design Example
Section 3 Sensor to ADC Design Example 3-1 This section describes the design of a sensor to ADC system. The sensor measures temperature, and the measurement is interfaced into an ADC selected by the systems
Making Accurate Voltage Noise and Current Noise Measurements on Operational Amplifiers Down to 0.1Hz
Author: Don LaFontaine Making Accurate Voltage Noise and Current Noise Measurements on Operational Amplifiers Down to 0.1Hz Abstract Making accurate voltage and current noise measurements on op amps in
OEM MAXNIBP Frequently Asked Questions
Frequently Asked Questions Why does the monitor sometimes inflate the BP cuff, then shortly thereafter reinflate the cuff? How will I know if the monitor is experiencing motion artifact during a measurement?
AVR125: ADC of tinyavr in Single Ended Mode. 8-bit Microcontrollers. Application Note. Features. 1 Introduction
AVR125: ADC of tinyavr in Single Ended Mode Features Up to 10bit resolution Up to 15kSPS Auto triggered and single conversion mode Optional left adjustment for ADC result readout Driver source code included
Current Loop Tuning Procedure. Servo Drive Current Loop Tuning Procedure (intended for Analog input PWM output servo drives) General Procedure AN-015
Servo Drive Current Loop Tuning Procedure (intended for Analog input PWM output servo drives) The standard tuning values used in ADVANCED Motion Controls drives are conservative and work well in over 90%
How To Use A Watt Saver On A Microcontroller (Watt Saver) On A Cell Phone Or Mp3 Player
Watt Saver for a Cell Phone AC Adapter Reference Design Document Number: DRM130 Rev 1, 10/2013 2 Freescale Semiconductor, Inc. Contents Section number Title Page Chapter 1 Introduction 1.1 Overview...5
AVR Timer/Counter. Prof Prabhat Ranjan DA-IICT, Gandhinagar
AVR Timer/Counter Prof Prabhat Ranjan DA-IICT, Gandhinagar 8-bit Timer/Counter0 with PWM Single Compare Unit Counter Clear Timer on Compare Match (Auto Reload) Glitch-free, Phase Correct Pulse Width Modulator
Ignition Alert Anti-Theft Security System for Motorbikes with Remote Control Amit Yadav, Anushri Jha, Neelesh Gupta amitrinku007@yahoo.
Ignition Alert Anti-Theft Security System for Motorbikes with Remote Control Amit Yadav, Anushri Jha, Neelesh Gupta [email protected] Abstract There are many vehicle security system are available
4. Optical position sensor, photo-current measurement
4. Optical position sensor, photo-current measurement Task 1. Measure the output signal of photo-electric sensors Measure the relative dependency of output signal versus light intensity. The dependence
How to Turn an AC Induction Motor Into a DC Motor (A Matter of Perspective) Steve Bowling Application Segments Engineer Microchip Technology, Inc.
1 How to Turn an AC Induction Motor Into a DC Motor (A Matter of Perspective) Steve Bowling Application Segments Engineer Microchip Technology, Inc. The territory of high-performance motor control has
INFRARED PARTS MANUAL
INFRARED PARTS MANUAL PIR325 FL65 GLOLAB CORPORATION Thank you for buying our Pyroelectric Infrared components. The goal of Glolab is to produce top quality electronic kits, products and components. All
Capacitive Sensing Solutions from Silicon Labs. In this section, we are going to cover the Silicon Labs Capacitive Sensing solutions.
Capacitive Sensing Solutions from Silicon Labs In this section, we are going to cover the Silicon Labs Capacitive Sensing solutions. 1 Agenda What are we measuring? Capacitance measurement techniques Relaxation
Implementation of Cloud-Computing Healthcare based on Xenon Uploading System and Hilbert Huang Transform
Implementation of Cloud-Computing Healthcare based on Xenon Uploading System and Hilbert Huang Transform Terry B J Kuo 1,2,3, Yu-Cheng Lin 1,2, I-Te Hsieh 1,2, Cheryl C H Yang 1,2,3, Kuan-Liang Kuo 4,
Wireless Medical Telemetry Laboratory
Wireless Medical Telemetry Laboratory 0 Introduction The development of wireless medical telemetry has become an increasingly popular application in recent years. As the elderly population continues to
FFT Frequency Detection on the dspic
FFT Frequency Detection on the dspic by Jac Kersing Nov 2012 Abstract In this article Jac explains how he solved a practical challenge using powerful dspic devices and Fast Fourier Transform algorithms.
Germanium Diode AM Radio
Germanium Diode AM Radio LAB 3 3.1 Introduction In this laboratory exercise you will build a germanium diode based AM (Medium Wave) radio. Earliest radios used simple diode detector circuits. The diodes
ACTIVE INFRARED MOTION DETECTOR FOR HOUSE SECURITY SYSTEM MIOR MOHAMMAD HAFIIZH BIN ABD. RANI UNIVERSITI MALAYSIA PAHANG
ACTIVE INFRARED MOTION DETECTOR FOR HOUSE SECURITY SYSTEM MIOR MOHAMMAD HAFIIZH BIN ABD. RANI UNIVERSITI MALAYSIA PAHANG ABSTRACT Nowadays, house security system becomes the best solution to overcome house
Application Note AN1
TAKING INVENTIVE STEPS IN INFRARED. MINIATURE INFRARED GAS SENSORS GOLD SERIES UK Patent App. No. 799A USA Patent App. No. 9/78,7 World Patents Pending SENSOR OVERVIEW Application Note AN The Dynament
Application Note 58 Crystal Considerations with Dallas Real Time Clocks
www.dalsemi.com Application Note 58 Crystal Considerations with Dallas Real Time Clocks Dallas Semiconductor offers a variety of real time clocks (RTCs). The majority of these are available either as integrated
isim ACTIVE FILTER DESIGNER NEW, VERY CAPABLE, MULTI-STAGE ACTIVE FILTER DESIGN TOOL
isim ACTIVE FILTER DESIGNER NEW, VERY CAPABLE, MULTI-STAGE ACTIVE FILTER DESIGN TOOL Michael Steffes Sr. Applications Manager 12/15/2010 SIMPLY SMARTER Introduction to the New Active Filter Designer Scope
Buffer Op Amp to ADC Circuit Collection
Application Report SLOA098 March 2002 Buffer Op Amp to ADC Circuit Collection Bruce Carter High Performance Linear Products ABSTRACT This document describes various techniques that interface buffer op
POCKET SCOPE 2. The idea 2. Design criteria 3
POCKET SCOPE 2 The idea 2 Design criteria 3 Microcontroller requirements 3 The microcontroller must have speed. 3 The microcontroller must have RAM. 3 The microcontroller must have secure Flash. 3 The
QUICK START GUIDE FOR DEMONSTRATION CIRCUIT 956 24-BIT DIFFERENTIAL ADC WITH I2C LTC2485 DESCRIPTION
LTC2485 DESCRIPTION Demonstration circuit 956 features the LTC2485, a 24-Bit high performance Σ analog-to-digital converter (ADC). The LTC2485 features 2ppm linearity, 0.5µV offset, and 600nV RMS noise.
Department of Electrical and Computer Engineering Ben-Gurion University of the Negev. LAB 1 - Introduction to USRP
Department of Electrical and Computer Engineering Ben-Gurion University of the Negev LAB 1 - Introduction to USRP - 1-1 Introduction In this lab you will use software reconfigurable RF hardware from National
Computer Aided Design of Home Medical Alert System
Computer Aided Design of Home Medical Alert System Submitted to The Engineering Honors Committee 119 Hitchcock Hall College of Engineering The Ohio State University Columbus, Ohio 43210 By Pei Chen Kan
The Energy Harvesting Tipping Point for Wireless Sensor Applications
The Energy Harvesting Tipping Point for Wireless Sensor Applications Ever since the first watermills and windmills were used to generate electricity, energy harvesting has been an attractive source of
The Periodic Moving Average Filter for Removing Motion Artifacts from PPG Signals
International Journal The of Periodic Control, Moving Automation, Average and Filter Systems, for Removing vol. 5, no. Motion 6, pp. Artifacts 71-76, from December PPG s 27 71 The Periodic Moving Average
LVDS Technology Solves Typical EMI Problems Associated with Cell Phone Cameras and Displays
AN-5059 Fairchild Semiconductor Application Note May 2005 Revised May 2005 LVDS Technology Solves Typical EMI Problems Associated with Cell Phone Cameras and Displays Differential technologies such as
How To Know If You Are Safe To Use An Antenna (Wired) Or Wireless (Wireless)
1 2 The range of RF spans 3 KHz (3000 Hz) to 300 GHz (300 million Hz) Frequencies of RF devices range from the low frequency AM broadcasts (80 MHz) to higher frequency mobile phones (1900 MHz) smart meters
Lecture N -1- PHYS 3330. Microcontrollers
Lecture N -1- PHYS 3330 Microcontrollers If you need more than a handful of logic gates to accomplish the task at hand, you likely should use a microcontroller instead of discrete logic gates 1. Microcontrollers
Clock Recovery in Serial-Data Systems Ransom Stephens, Ph.D.
Clock Recovery in Serial-Data Systems Ransom Stephens, Ph.D. Abstract: The definition of a bit period, or unit interval, is much more complicated than it looks. If it were just the reciprocal of the data
Designing a Printed Circuit Board
Designing a Printed Circuit Board Jamie Jacobs Design Team 4 4/03/2009 Abstract When the development stage of a circuit board is complete and working correctly, it is then necessary to take this breadboard
Doppler. Doppler. Doppler shift. Doppler Frequency. Doppler shift. Doppler shift. Chapter 19
Doppler Doppler Chapter 19 A moving train with a trumpet player holding the same tone for a very long time travels from your left to your right. The tone changes relative the motion of you (receiver) and
A Low-Cost, Single Coupling Capacitor Configuration for Stereo Headphone Amplifiers
Application Report SLOA043 - December 1999 A Low-Cost, Single Coupling Capacitor Configuration for Stereo Headphone Amplifiers Shawn Workman AAP Precision Analog ABSTRACT This application report compares
Blood Pressure Monitor Using Flexis QE128 Gabriel Sanchez RTAC Americas
Freescale Semiconductor Application Note Document Number: AN3500 Rev. 0, 08/2007 Blood Pressure Monitor Using Flexis QE128 by: Gabriel Sanchez RTAC Americas 1 Introduction Product designers and developers
Movement artifacts electrical noise caused by the movement of the sensor surface with respect to the skin surface
BIOHARNESS 42 Aero Camino, Goleta, CA 93117 Tel (805) 685-0066 Fax (805) 685-0067 [email protected] www.biopac.com 12.15.08 Field Comparison: Zephyr BioHarness / Polar HR Monitor Introduction The Zephyr
Hand Gestures Remote Controlled Robotic Arm
Advance in Electronic and Electric Engineering. ISSN 2231-1297, Volume 3, Number 5 (2013), pp. 601-606 Research India Publications http://www.ripublication.com/aeee.htm Hand Gestures Remote Controlled
Troubleshooting accelerometer installations
Troubleshooting accelerometer installations Accelerometer based monitoring systems can be tested to verify proper installation and operation. Testing ensures data integrity and can identify most problems.
Application Report. 1 Introduction. 2 Resolution of an A-D Converter. 2.1 Signal-to-Noise Ratio (SNR) Harman Grewal... ABSTRACT
Application Report SLAA323 JULY 2006 Oversampling the ADC12 for Higher Resolution Harman Grewal... ABSTRACT This application report describes the theory of oversampling to achieve resolutions greater than
Constructing a precision SWR meter and antenna analyzer. Mike Brink HNF, Design Technologist.
Constructing a precision SWR meter and antenna analyzer. Mike Brink HNF, Design Technologist. Abstract. I have been asked to put together a detailed article on a SWR meter. In this article I will deal
Patient Health Monitoring Using Wireless Body Area Network
Patient Health Monitoring Using Wireless Body Area Network Hsu Myat Thwe, Hla Myo Tun Abstract: wadays, remote patient health monitoring using wireless technology plays very vigorous role in a society.
Various Technics of Liquids and Solids Level Measurements. (Part 3)
(Part 3) In part one of this series of articles, level measurement using a floating system was discusses and the instruments were recommended for each application. In the second part of these articles,
MEASUREMENT UNCERTAINTY IN VECTOR NETWORK ANALYZER
MEASUREMENT UNCERTAINTY IN VECTOR NETWORK ANALYZER W. Li, J. Vandewege Department of Information Technology (INTEC) University of Gent, St.Pietersnieuwstaat 41, B-9000, Gent, Belgium Abstract: Precision
Harmonics and Noise in Photovoltaic (PV) Inverter and the Mitigation Strategies
Soonwook Hong, Ph. D. Michael Zuercher Martinson Harmonics and Noise in Photovoltaic (PV) Inverter and the Mitigation Strategies 1. Introduction PV inverters use semiconductor devices to transform the
Vibrations can have an adverse effect on the accuracy of the end effector of a
EGR 315 Design Project - 1 - Executive Summary Vibrations can have an adverse effect on the accuracy of the end effector of a multiple-link robot. The ability of the machine to move to precise points scattered
Use and Application of Output Limiting Amplifiers (HFA1115, HFA1130, HFA1135)
Use and Application of Output Limiting Amplifiers (HFA111, HFA110, HFA11) Application Note November 1996 AN96 Introduction Amplifiers with internal voltage clamps, also known as limiting amplifiers, have
FREQUENCY RESPONSE ANALYZERS
FREQUENCY RESPONSE ANALYZERS Dynamic Response Analyzers Servo analyzers When you need to stabilize feedback loops to measure hardware characteristics to measure system response BAFCO, INC. 717 Mearns Road
MATRIX TECHNICAL NOTES
200 WOOD AVENUE, MIDDLESEX, NJ 08846 PHONE (732) 469-9510 FAX (732) 469-0418 MATRIX TECHNICAL NOTES MTN-107 TEST SETUP FOR THE MEASUREMENT OF X-MOD, CTB, AND CSO USING A MEAN SQUARE CIRCUIT AS A DETECTOR
Signal Extraction Technology
Signal Extraction Technology Technical bulletin Introduction Masimo SET pulse oximetry is a new and fundamentally distinct method of acquiring, processing and reporting arterial oxygen saturation and pulse
mdocinabox with Arduino UNO and Parallax BOE, 2013 mdocinabox Rattanaphan Boonbutra of Computer Science, Hood College.
mdocinabox Rattanaphan Boonbutra of Computer Science, Hood College. Index Terms Arduino UNO, Parallax BOF, BlueSMiRF, Amarino 2.0, Amarino plugin bundle, pressure sensor, the MEAS vibra sensor, and the
Product Introduction MyoMuscle. Telemyo DTS-System. Telemetry system for EMG and Biomechanical Sensors. www.noraxon.com
Product Introduction MyoMuscle Telemyo DTS-System Telemetry system for EMG and Biomechanical Sensors DTS Device Family 3 major device categories can be used for the Direct Transmission System technology
Synchronization of sampling in distributed signal processing systems
Synchronization of sampling in distributed signal processing systems Károly Molnár, László Sujbert, Gábor Péceli Department of Measurement and Information Systems, Budapest University of Technology and
Wireless Device for Patient Monitoring
Wireless Device for Patient Monitoring Robert G. Lupu, Andrei Stan, Florina Ungureanu Abstract The increasing request of patients, suffering of chronic diseases, who wish to stay at home rather then in
Optimizing VCO PLL Evaluations & PLL Synthesizer Designs
Optimizing VCO PLL Evaluations & PLL Synthesizer Designs Today s mobile communications systems demand higher communication quality, higher data rates, higher operation, and more channels per unit bandwidth.
Design of Bidirectional Coupling Circuit for Broadband Power-Line Communications
Journal of Electromagnetic Analysis and Applications, 2012, 4, 162-166 http://dx.doi.org/10.4236/jemaa.2012.44021 Published Online April 2012 (http://www.scirp.org/journal/jemaa) Design of Bidirectional