Packaging for Sterilization
|
|
|
- Beatrice Ophelia Morton
- 9 years ago
- Views:
Transcription
1 Packaging for Sterilization IT S NOT JUST A BOX, A BAG OR GIFT WRAPPING! by Rose Seavey, RN, BS, MBA, CNOR, ACSP Objectives After completion of this self-study activity, the learner will be able to: 1. Describe available options for sterilization packaging and containment devices. 2. Discuss appropriate use for each type of sterilization packaging. 3. Develop a Quality Control Program for evaluation, selection and use of packaging systems for items to be sterilized. 4, Order the Association for the Advancement of Medical Instrumentation (AAMI) and Association of perioperative Registered Nurses (AORN) documents for your reference library. Test Questions True of False. Circle the correct answer. 1. If major changes are made to packaging such as switching the type of wrappers, peel pouches or containment devices used, or changing the package dimensions or weight of the set, facilities should conduct product testing with biological indicators (BIs) and chemical indicators (CIs) as well as moisture content assessment. 2. Both AAMI and AORN recommendations agree on a maximum weight limit of 25 pounds for a containment device, the instruments, and any accessories or wrapper due to ergonomic and drying issues. 3. Both solid and perforated bottom rigid containers can be sterilized by gravity or dynamic-air-removal steam sterilization cycles. Many thanks to the team at 3M Health Care for working with Managing Infection Control to provide the following accredited course. IAHCSMM has awarded one and one-half (1.5) contact points for completion of this continuing education lesson toward IAHCSMM recertification. The CBSPD has preapproved this inservice for one and one-half (1.5) contact hours for a period of five (5) years from the date of publication, and to be used only once in a recertification period. This inservice is 3M Health Care Provider approved by the California Board of Registered Nurses, CEP 5770 for one (1) contact hour. This form is valid up to five (5) years from the date of publication. Instructions for submitting results are on page 97. Managing Infection Control and 3M Health Care will be working collaboratively to provide continuing education courses in monthly editions of Managing Infection Control. 4. Rigid sterilization container systems should be cleaned after each use following the manufacturers written recommendations. 5. Room temperature and relative humidity has no effect on packaging materials and the ability to achieve sterilization. 6. Using the proper wrap size and grade is important and wrapping procedures should be standardized. 7. If packages are wrapped too loose, it could create low spots that can collect condensate. 8. If items are to be put in a double peel pouch, the inner pouch should fit inside the outer pouch without folding and the plastic side of the inner pouch should face the plastic side of the outer pouch. 9. Paper-plastic pouches should not be placed inside wrapped sets or containment devices. 10. Instrument count sheets should be placed on the inside of sterile packages. Introduction There are several types of sterilization packaging systems on the market today. These devices help to maintain sterility, as well as protect the contents of the package and support aseptic presentation. Most healthcare facilities use a variety of sterilization packaging systems. The available systems include woven and nonwoven wrappers, peel pouches of plastic, 82 MANAGING INFECTION CONTROL
2 paper or Tyvek, and containment devices such as reusable rigid sterilization containers, organizing trays and instrument case/cassettes available in both plastic and metal. There are also a wide variety of devices used to protect instruments or components. These include insert cases or trays, reusable accessories, paper-paper bags, foam and towels. Sterilization packaging is classified by the Food and Drug Administration (FDA) as a Class II medical device and is defined as: A sterilization wrap (pack, sterilization wrapper, bag or accessories) is a device intended to be used to enclose another medical device that is to be sterilized by a health care provider. It is intended to allow sterilization of the enclosed medical device and also to maintain sterility of the enclosed device until used. 1 Sterilization packaging is more complicated than you might think. It is not just a box, bag or wrap! As a Class II medical device, there are specific validated guidelines that must be followed with each type of system used. To ensure patient safety, manufacturers written instructions must be followed. Not all packaging systems are compatible with all types of sterilization processes. This article will cover the various types of sterility packaging and container systems, their intended use, and limitations. This educational offering will also discuss labeling, closures, tamper evidence, product evaluations and testing, weights and density of sets, compability issues, instrument inventory list placement, options for internal containment devices, and maintenance of reusable containers. Purpose of packaging The intention of any sterilization packaging system is to allow sterilization of the package contents, maintain sterility of contents until the package is opened, and permit delivery of contents without contamination. PACKAGING OPTIONS The choice of packaging systems includes: woven fabrics, usually 100 percent cotton, cottonpolyester blends and synthetic blends, either treated or untreated; nonwoven materials, made of plastic polymers, cellulose fibers or washed paper pulp bonded under pressure into sheets not woven on a loom. These are usually designed for single use; peel pouches of plastic and/or paper, made of a variety of materials, including paper, cellophane, polyethylene, Tyvek and various paper-plastic combinations; and rigid container systems which are specially designed metal or plastic containers. WHICH PACKING TO USE? Deciding on the appropriate packaging system for sterilizing devices has become a challenging quest. The development of high tech, complicated surgical devices especially for minimally invasive surgical procedures and an increasing variety of sterilization processes require greater diligence and critical thinking skills when choosing the type of packaging system to use for each device, or instrument set or sterilization process. Issues of cost, rapid turnover, ease of use, material compatibility and storage space need to be balanced with patient safety and instrument protection concerns. Healthcare professionals should work closely with the instrument, sterilizer and packaging manufacturers to determine the appropriate devices to wrap, package and contain sterilized items. EVALUATING PACKAGING SYSTEMS Before deciding on which packaging system(s) to purchase and use, an evaluation should take place to investigate suitability for items being sterilized, types and methods of sterilization used. According to the Association of perioperative Registered Nurses (AORN), sterilization packaging systems should: have a sufficient barrier to microbes, particulates and liquid; maintain sterility of contents until used; permit sterilant saturation and removal while allowing direct contact with all surfaces of each item; have no toxic ingredients or nonfast dyes; allow aseptic presentation to the sterile field; close completely and securely; protect items from physical damage; provide a sufficient, reliable tamper-evident seal; be resistant to tears or punctures; have very little lint; allow for content identification; be of sufficient size to uniformly distribute the items; be easy to use when preparing and opening the package or container; be cost effective and; have manufacturer s written instructions for use. 2 Important items to consider when evaluating a rigid container system include: Ergonomics of container design for ease of carrying; Empty container weight; Ease of locking and closing the container; Ability to stack containers for storage and transportation; Design of filter retention and ease of use; Types of accessories for organization; Ability to stack baskets internally; Various sizes to accommodate a wide variety of instrument sets; 84 MANAGING INFECTION CONTROL
3 Ability to clearly label sets; Ability to put container through an automatic washer; and Preventative maintenance and service. Work closely with the instrument, sterilizer and packaging manufacturer to determine the appropriate packaging to use for each device and instrument set. Evaluate packaging before purchasing. Product testing Whenever there are major changes made to packaging, e.g., switching the type of packaging material or containers used, and/or changes in the package dimensions or weight, product testing should always be performed. This test should include biological indicators (BIs) and chemical indicators (CIs) as well as moisture content (wet pack) assessment after the cycle is complete. 3 The BIs and CIs should be placed in the most challenging area of the packages being tested and in multiple locations. More than likely, that will be the center of a wrapped pack or the corners of containers, inside containment devices and on each level of multilevel packs and next to the largest heat sink. The placement of the indicators should be identified for record keeping purposes. The packages should be labeled as product tests and run in a full load using the medical device manufacturer s instructions. The packages are sacrificed and not used so the CIs can be read and the BIs incubated. Test results should be documented and maintained with the sterilization log and the quality assurance records. The products with the changes should not be placed into routine use until the CIs reach their endpoint and the BIs are negative. 2-4 For specific instructions on facility product testing, healthcare providers should refer to the Association for the Advancement of Medical Instrumentation s (AAMI) Comprehensive guide to steam sterilization and sterility assurance in health care facilities, ANSI/AAMI ST79:2006 section 10 Quality Control, 10.9 Periodic product quality assurance testing of routinely processed items. 3 Perform product testing whenever there are major changes made to packaging, e.g., switching the type of packaging material or containers used, and/or changes in the package dimensions or weight on loaner instruments. Weight and density of sets An ongoing issue is the instrument set weights. Instrument sets sometimes seem to take on a life of their own, growing at alarming rates. Some of the complete sets from the original manufactures are almost too heavy to lift. Often additional items get repeatedly added to sets creating a set with a metal mass that is not equally distributed. From time to time, due to lack of storage space or for staff convenience, we merge multiple sets into one. With all of these issues there is a growing concern with the weight of instrument sets. Due to ergonomic, sterilization and drying issues, both AAMI and AORN recommendations agree on a maximum weight limit of 25 pounds. AORN states: The total weight of instrument containment devices should not exceed 25 pounds, including the contents and containment method (eg., wrappers, rigid container systems, cassettes, organizing trays). Excessively heavy instrument sets may compromise sterilization and drying. The focus should be on the set configuration (i.e., how the instruments are distributed in the set) and the overall weight of the set. Lifting and moving heavy instrument sets may cause health care worker injury. 2 The AAMI Containment devices for reusable medical device sterilization, ANSI/AAMI ST77 and ANSI/AAMI ST79:2006 section 4 Requirements, recommends a maximum combined weight limit of 25 pound. This weight limit includes the containment device, the instruments, and any accessories or wrappers. 5 If you are using a containment device, the manufacturer should be consulted on the recommendations for weights and density of instrument sets. However, it is the facility s responsibility to verify that the instrument set can be effectively sterilized and dried in their sterilizers by doing product testing and following the medical device manufacture s instructions for use. 3 Four years ago, The Children s Hospital of Denver (TCH) chose to institute a maximum weight limit of 20 pounds for any sterilized package. This was instituted to comply with the container manufacturers recommendations of a pound limit as well as ergonomic issues for staff. The implementation of this weight limit was established not only for the rigid containers but for any package including wrapped trays and loaner instrumentation trays. This program has proven very successful in many ways. The complaints of back, neck, shoulder and wrist pain from employees have been significantly reduced as well as almost virtually eliminating any wet loads while decreasing some of the dry times. Here are steps for an instrument set weight reduction strategy: 86 MANAGING INFECTION CONTROL
4 Establish the maximum acceptable weight limits for your healthcare facility instrument sets with input from surgical services. Add the established weight limit to your policy and be sure to get the approval of surgical services, infection control and risk management; also inform the facility and loaner instrument vendors. Obtain written reprocessing and sterilization guidelines from all medical device manufacturers including rigid containers. Redesign the instrument sets to even out the metal mass. Perform product testing using biological and chemical indicators inside representative sets in the areas determined to be the most challenging and following the medical device manufacturers sterilization parameters. Place the sets into use. Routinely monitor sets to ensure the weight is not inching up. 4 Figure 1. Wrapped pan being weighed STEAM Steam sterilization is affected by such things as the amount of available humidity, the altitude, the type of packaging material, its contents, the size and configuration of the load, the size, weight and density of the pack or container, and the parameters of the sterilization cycle. It is important to follow the manufacturers written instructions for each type of packaging system used in the steam sterilizer. 2 ETHYLENE OXIDE (EO) When using EO sterilization, the packaging systems must be EO, moisture and air permeable, allow aeration of the gas, be constructed of a material recommended by the sterilizer and sterilant manufacturer, and be compatible with the process of EO sterilization. Peel-packs, wrappers, both woven and nonwoven and some rigid container systems are permeable to EO. 2 HYDROGEN PEROXIDE GAS PLASMA Sterilization packaging systems used for packages undergoing hydrogen peroxide gas plasma sterilization must allow vapor phase hydrogen peroxide to penetrate packaging materials, be compatible with the process, and be recommended by the manufacturer. Both the packaging and sterilizer manufacturer s written instructions should be followed. Because cellulose products are not compatible with hydrogen peroxide gas plasma, peel packs used in these sterilizers must be made of Tyvek /plastic. Only compatible container systems which have been validated for this application should be used. 2 OZONE Packages used for sterilization with Ozone must allow Ozone to penetrate the package material, be compatible with the process, and be recommended by the manufacturer. Both the packaging and sterilizer manufacturer s written instructions for use should be followed. Ozone compatible packaging includes uncoated nonwoven materials, Tyvek /plastic pouches and commercially available anodized aluminum containers using noncellulose disposable filters. 2 Establish a maximum weight limit of 25 pounds for a containment device, the instruments, and any accessories or wrapper to protect your employees health and to decrease drying times. Packaging systems compatibility Not all packaging systems are compatible with all types of sterilization processes. As a Class II medical device, packaging systems must be cleared by the FDA for their intended purposes as well as the method of sterilization used. Choose a packaging system that is cleared by the FDA for the purpose and sterilization method you are using. Types of packaging systems available RIGID STERILIZATION CONTAINER SYSTEMS Reusable rigid container systems are used for the packaging, transportation and storage of surgical instruments prior to, during and after sterilization. These self-contained closed sterilization containers use filters or valves as a barrier system to permit air removal and the sterilant to enter and exit the container. The reusable box-like structures act as a microbial barrier and must have a tamper proof seal. Rigid container systems have internal baskets, come in many sizes, and may have solid or perforated bottoms. The container manufacturer s MANAGING INFECTION CONTROL 87
5 written recommendations for sterilization cycle parameters should be followed. In general, the solid bottom containers are to be used in steam sterilizers using dynamic-air-removal cycles only. Perforated bottoms can be sterilized by gravity or dynamic-air-removal steam sterilization. Some containers have been validated for use in Ethylene Oxide, Hydrogen Peroxide Gas Plasma, Ozone and flash sterilization. Solid bottoms can also be used to transport contaminated moist instruments. Figure 2. Solid bottom rigid container Pre-assembly Before placing items in the tray, both the box and the lid of the container system should be inspected to ensure the entire container, top and bottom is free from dents or cracks. The gaskets should be pliable and securely fastened without breaks or cuts. The locking mechanism should function properly. All of the container s components (top, bottom, valve or filter mechanisms, securing or latching mechanisms) must function effectively as a unit. It is vital to the maintenance of sterility, that these components work together to allow air removal, to facilitate sterilant penetration and removal, and to inhibit contamination. Filter retention mechanisms, valves and fasteners such as screws and rivets should be secure and not loose. If the filter or valve is found loose, the sterility of the contents should be questioned. If any part of the container is not satisfactory or free of functional defects, the container should be repaired or replaced. Assembly To assure all instruments are exposed to the sterilization agents, the basket(s) placed in the container should be large enough to allow the metal mass of instruments to be distributed evenly. Only filter materials that have been tested and documented to be efficacious in the specific container system should be used. If disposable filters are used, a new filter should be used each time the container is sterilized. If mechanical filters are used, their functionality may need to be checked daily or before each use. Follow the manufacturers instructions. Internal stacking Some container manufacturers have validated internal stacking of two or more layers within a containment device. According to ANSI/AAMI ST77: , Internal stacking: If organizing trays are to be stacked within a containment device, the following factors shall be addressed for each sterilization method for which stacking is recommended: a) ease of removal of stacked items within a containment device; b) the maximum density of contents within the containment device to allow for proper sterilization and drying; c) adequate perforations to allow for sterilant penetration, sterilant evacuating, drying, and, if applicable, aeration; d) the stability of the stacked items during transport and handling; e) a dry outcome (see ); f) achievement of sterilization throughout the layers of stacked organizing trays and the recommended accessories (e.g., mats, holders) within the containment device. If any of the accessories are incompatible with a specific sterilization method, the incompatibility should be indicated in the instructions for use. 5 If trays are stacked internally, a chemical indicator/integrator (CI) should be placed in each layer in the area of the tray or container system considered to be the least accessible to sterilant penetration. 3 AORN Recommended Practices for Selection and Use of Packaging Systems recommends placing two chemical indicators/integrators in opposite corners of each level of trays. 2 Tamper-evident closure Self-contained sterilization containers require a tamper-evident closure (usually a plastic lock) to provide the user with visual confirmation that the container has not been opened after sterilization until the time of use. The tamper-evident closure system shall be effortlessly and safely removed. 5 Figure 3. Tamper-evidence closure or lock 88 MANAGING INFECTION CONTROL
6 Immediately before use Before it is opened, the container should be inspected for the appropriate appearance of the filters or valves and external chemical indicators and the physical integrity of the outer containers and tamper-evident devices. The 2007 AORN Recommended Practices for Maintaining a Sterile Field states: Rigid container systems should be opened on a separate surface. The external indicator should be verified for appropriate color change. Locks should be inspected for security to verify there has not been a breach of the container seal prior to use. The lid should be lifted toward the person opening the container and away form the container. The filter should be checked and changed according to the manufacturer s written instructions. 6 Care should be taken to ensure that there is no contact between the lid and the inner rim of the sterile contents, or any part of the inside of the container. Before removing the instrument basket, the surgically attired scrub person should check the internal chemical indicator(s) for the appropriate change. The scrub person should avoid all contact with the table, external surfaces of the container and the upper rim of the container. Before the instrument basket is placed on the sterile field, the container bottom should be inspected visually for the proper appearance of the filter or valve assembly; and the instruments, basket and container bottom should be inspected for moisture. Cleaning Reduction of bioburden is essential for sterilization. Rigid container systems should be cleaned after each use following the manufacturers written recommendations. The cleaning and decontamination of rigid sterilization containers systems is as important as the cleaning and decontamination of the soiled/contaminated contents. During routine cleaning, the container should be inspected for mechanical wear and stress. If the container system has removable filters, the filter holders (retention plates) should be removed to disengage the filter. If disposable, the MANAGING INFECTION CONTROL 89
7 Figure 4. Sequential double-wrapping: envelope fold entire filter should be disposed of. Reusable filters should be disassembled, cleaned and replaced according to the manufacturer s instructions. Most container systems can be cleaned and decontaminated in mechanical equipment. The cleaning method selected depends on the container manufacturer s instructions. If the container has valve-type closures, the manufacturer s instructions for frequency and method of removal, disassembly and cleaning should be followed. Follow the rigid container manufacturer s instructions for pre-assembly, assembly, internal stacking, tamper-evident closures, how to open and cleaning. Reprinted from ANSI/AAMI ST79:2006 with permission of the Association for the Advancement of Medical Instrumentation, Inc AAMI All rights reserved. Figure 6. Simultaneous double-wrapping: envelope fold Reprinted from ANSI/AAMI ST79:2006 with permission of the Association for the Advancement of Medical Instrumentation, Inc AAMI All rights reserved. PACKAGING MATERIALS (WRAPPERS) Sterilization wrappers used today for sterilization packaging are either disposable nonwoven fabrics made of natural or manmade fibers which are pressure-bonded together to form sheets of fabric; or reusable woven textiles made of natural and/or synthetic materials. Some are treated for water repellence. 1 Reusable materials must be free of detergents, bleaches or other chemicals that could react with the sterilant and cause discoloration and/or adversely affect the contents of the pack. 1 Wrappers made of reusable woven textile need to be rehydrated by laundering between every use. Superheating, which could limit the ability to achieve sterilization, may result if textiles are not re-laundered. 3 All reusable textiles need to be de-linted before using. Textile fabrics need to be inspected over a light table for defects before each use. Rips should not be sewn, which would increase the number of holes in the material. Defects should be repaired on both sides with a vulcanized patch applied using heat. 2 All packaging materials should be at room temperature (68-73 F/20-23 C) and at a relative humidity of 30 percent to 60 percent for at least two hours before use. Sustaining room temperature and moisture content of woven material allows sterilant penetration and prevents superheating during the sterilization process. 2 Single-use packaging material should be used for one sterilization cycle. Disposable packaging material should be discarded after opening. 2 Pre-assembly Packaging materials should be routinely examined for any holes, defects and inappropriate extraneous material. Wrap comes in various sizes and grade (sturdiness). Using the proper wrap size is important. The wrap must be of sufficient size and grade to adequately cover and support the weight and configuration of the medical device being wrapped. If the wrap is too large it will be hard to wrap and open aseptically. If the wrap is not sturdy enough for the weight or design of the items, it may have a tendency to tear during handling. Using manufacturers written recommendations, facilities should develop policies and procedures for all packaging techniques. 3 Wrapping procedures should be standardized. To decrease confusion for the end user, like packages should be wrapped in the same size wrapper every time. 90 MANAGING INFECTION CONTROL
8 Assembly When using wrappers, reusable or disposable, the package should be wrapped snug, but not too tightly. If loose it could cause low spots which could collect condensate and if wrapped too tightly it could result in strike-through. 3 Packages can be wrapped by the envelope fold method or the square fold method that is either sequential (see Figure 4 on page 90) or simultaneous (Figure 6 on page 90). When sequential wrapping is done the items are wrapped twice, creating a pack within a pack. When a pack is wrapped simultaneous the package is wrapped once but is done with a special double layered synthetic nonwoven material welded on the sides with a special double layered synthetic nonwoven material welded on the sides. Decreasing the chance for tears Some larger devices or items with sharp corners may benefit by placing a towel or foam under the device being wrapped (Figure 5 on page 92). This provides a type of cushion to help prevent rips or tears in the wrap. The user MANAGING INFECTION CONTROL 91
9 must make sure the cushion is compatible with the type of sterilization being used. Figure 5. Towel as cushion to decrease chance for tears designed specifically for sterile packaging are acceptable as outside closures only if the wrapper manufacturer explicitly recommends their use and only if care is taken to choose the proper size (relative to the length and width of the package) so that the elastomer band fits snugly yet does not constrict the package (e.g., create an hour-glass effect) or cause excessive wrinkles or folds in the package. 3 Follow recommended practices and manufacturer s instructions on pre-assembly, assembly, decreasing the chance for tears, and closures for wrapped packages. Transfer trays are another device to help protect the integrity of the wrap. These are specifically designed plastic trays that can help alleviate the possibility of tears due to manipulation onto and off of carts. The packages can be placed on the trays before or after sterilization. Figure 7. Transfer trays Closures for wrapped packages Once assembled, the sterilization package needs to be secured. The most common closure for packages is indicator tape. The devices must ensure the package remains together, allow for sterilization penetration, avoid constriction of the pack, and maintain integrity of the package. Tape (other than sterilization indicator tape) should not be used to secure packages, nor should safety pins, paper clips, staples, or other sharp objects. Elastomer bands PEEL PACKS Paper-plastic or Tyvek -plastic pouches are frequently used to package small lightweight, low-profile items (e.g., one or two needle holders or small retractors). One side of the peel pack is clear plastic which is important when it is necessary to view the items before they are opened. Pouches are not recommended for heavy or bulky items, because the seals may break open. 3 There are various size peel pouches and peel pack rolls (which can be cut to size) on the market today. Choosing the correct size and application of the pouch is important. If the pouch is too tight, the seal may not hold and if it is too loose, the item will move which could stress the seals to the point of rupturing. Proper sizing also allows for adequate air removal, steam penetration and drying. 3 Some peel pouches are self sealing, meaning they come with a sticky strip that will fold over to form a seal after the item(s) is placed inside. Some need to be heat sealed with a special machine that applies pressure and heat, bonding the two together to form a seal. The user needs to make sure there are no creases or gaps in the seal which could allow microorganisms to enter the package and contaminate it. A single peel pouch is adequate unless the package contains more than one item or would be complicated to open aseptically with only one peel package (e.g., very small items). If the item is to be double-packaged, two sequentially sized pouches should be used (i.e., the sealed inner pouch should fit inside the other pouch without folding). When assembled, the pouches should be positioned so that plastic of the inner one faces the plastic of the out one. 3 See Figure 8. Figure 8. Example of single- and double-packaging with paper-plastic pouches NOTE: Instruments should be oriented within paper-plastic pouches according to the healthcare facility s policies and procedures. Reprinted from ANSI/ AAMI ST79:2006 with permission of the Association for the Advancement of Medical Instrumentation, Inc AAMI All rights reserved. 92 MANAGING INFECTION CONTROL
10 Follow recommended practices and manufacturer s instructions on how to use peel pouches. When using double peel pouches, the inside peel pouch should fit without folding and be placed so the plastic of the inner pouch faces the plastic of the outer pouch. ORGANIZING DEVICES Figure 9. Containment devices Organizing devices are sometimes used inside of containers or trays to help keep items contained together (Figure 9). The devices are also used as protection for sharp and/or small items. Small basket-type accessory, containers with covers or lids should only be used if they have been designed and tested for this purpose. The user has responsibility to test and evaluate the effectiveness of sterilization and drying of protective organizing devices. Specific recommendations of the device manufacturer should be consulted. See Product testing above for more information. According to both AORN and AAMI: Paper-plastic pouches should not be used within wrapped sets or containment devices because the pouches cannot be positioned to ensure adequate air removal, sterilant contact, and drying. The practice of confining instruments in paper-plastic pouches and then including them in wrapped or containerized sets has not been validated as appropriate and efficacious by packing and container manufacturers. 2 Other options for containing devices include all paper bags (Figure 10), foam or woven material that are designed and tested for this purpose. Figure 10. Paper sterilization pouches In order to verify adequate sterilant contact, upon assembly, a CI should be placed inside each organizing device within the instrument set. Follow recommended practices and manufacturer s instructions on how to use peel pouches. Never place paper-plastic peel pouches inside a containment device because you can not ensure adequate air removal, sterilant contact, and drying. This would also be true of Tyvek plastic peel pouches. MANAGING INFECTION CONTROL 93
11 CHEMICAL INDICATOR/INTEGRATOR Chemical indicators/integrators, both internal and external, should be specific to the sterilization process being used. The end user should be knowledgeable and demonstrate competency regarding the performance characteristics of the monitoring products. 3 Internal CIs At least one chemical indicator/integrator should be used within each wrapped package. The CI may be a Class 3 (single parameter indicator), a Class 4 (multi-variable indicator), or Class 5 (integrating indicator). 2 AAMI ST79 states the Class 4 and Class 5 CIs provide more information than the Class 3 CIs that only measure temperature. The CI should be placed in the area of the package considered to be the least accessible to sterilant penetration and air removal. This may or may not be the geometric center of the package. For rigid containers, the corners and the area away from the filters are the likeliest location of air pockets, therefore, resulting in the biggest challenge for the CI. So place two CIs, one in each of two opposite corners of the inside basket. For multi-levels, either in rigid containers or wrapped sets, a CI should be placed on each level of the set in opposite corners. 3 End users should know how to interpret the CI and not use the package if the endpoint is not achieved. Other packages in the load should be quarantined until the BI results are available or other indicators are checked if the BI was not run. 3 External CIs An external chemical indicator, classified as Class 1, should be affixed to each package unless the internal CI is visible. The purpose of the external CI is to distinguish between processed and unprocessed devices. The external chemical indicator maybe located on the package, an identification card, tamper-proof locks, the filter, or elsewhere according to the policy of each healthcare facility. The indicator should be examined after sterilization and before use to make sure that the package and its contents were exposed to a sterilization process. 3 Place internal CIs inside each package in the areas that create the greatest challenge to air removal and sterilant penetration. Place an external CI on the outside of each package unless the internal CI is visible. If the external or internal CI have not reached their acceptable endpoint, do not use the package. Count sheets Instrument count sheets should not go inside wrapped or containerized instruments sets. Various types of copy paper (not medical grade paper) and toners are usually used to print count sheets. Currently, there is no available research regarding the effects of the copy paper or toners to the sterilization process. Paper and toner ink may pose a risk of reaction to some sensitized patients due to the chemicals used in the manufacturing of these products. 2 One option is to attach the instrument inventory count sheets to the outside of the container or wrapped package. Folding and then attaching the count sheet to the container handle or taping it to the wrap is one acceptable method to ensure the count sheet stays with the package (Figure 11 below). Figure 11. Count sheet placement on outside of package Do not place count sheets inside of wrapped or containerized instruments sets to avoid the patient contacting paper or toner ink which may cause a reaction in sensitized patients. Labeling the packages Packaging systems should be labeled with the description of the package contents, identification of the person assembling the package and a lot control number. To decrease confusion, packages should be labeled before being put into the sterilizer. 2 Package labels (i.e., process indicators, product identification, expiration statement labels) should be fixed firmly to the package all the way through processing, storage, and distribution to the point of use. If a marking pen is used to mark surgical packages, the information should be written on the indicator tape or attached label. Marking pens should only be used on the plastic side of peel-pouches. 2 Writing on the paper side of the pouch or on a wrapper (woven or nonwoven) could cause damage to the package (which might not be noticeable) and thereby compromise the barrier protection MANAGING INFECTION CONTROL
12 To avoid toxins being deposited on packs or instruments, the ink in marking pens should be nontoxic. When asking sterile processing professionals what they use for package labeling, they tell me they use a black felt tip marker. To my knowledge, the only felt tip marker tested to withstand extreme heat and steam as well as processing chemicals and cleaners is the Industrial Sharpie. It comes in fine point (item no ) or extra fine point (item no ). The Industrial Sharpie package states one of the specific uses is for sterilization such as autoclaving. Only write on the indicator tape, attached label or plastic side of a peel pouch so that toxins do not contact the medical devices. Extended cycles Today s sterile processing professionals are faced with the issue of extended steam sterilization cycles for surgical instruments (a cycle time longer than four minutes at F/ C in a dynamic-air-removal sterilizer). AORN s Sterile Processing/Materials Management Specialty Assembly (SP/MM SA) Coordinating Council released a general statement on extended cycles: Each medical facility needs to make sure that all their products used for sterilization (peel pouches, wrap, etc.) can withstand these longer steam sterilization cycles. Each medical facility needs to make sure that the products used in extended cycles are validated for these types of cycles. Many manufactures have tested their products for these longer cycle times. The manufacture of the products should supply information to the user for their records. 7 At the time of writing this article there are two manufacturers of sterilization packaging products that have documentation posted on AORN s SP/MM SA Web portal at communities.aorn.org/cop/spmatmngmt/filesharing/index. fusion that state their products can withstand sterilization cycles with extended exposure times up to 30 minutes. 7 Check with the manufacturer of the packaging material to make sure it can be used in extended cycles or cycles longer than four minutes at F/ C in a dynamic-air-removal sterilizer. Competency Employees responsible for selecting and using packaging systems should be knowledgeable about sterilization principles and each manufacturer s written directions and sterilization compatibility issues of each type of packaging. The competency of each staff member responsible for packaging systems should periodically be evaluated and documented. 2 All the information in this inservice should be part of the competency testing of all employees that select and use packaging systems. Summary Sterilization packaging is not a simple topic. It must adequately allow sterilization of the enclosed medical device and also maintain sterility of the contents until used. Classified by the FDA as a Class II medical device, it is much more than a box, bag or gift wrap! To ensure patient safety, specific manufacturers written validated guidelines for each type of system must be followed and staff must demonstrate competency in each packaging system used. Meanwhile, save the gift wrapping skills for the holidays! Glossary of Terms Aseptic presentation: Maintaining the sterility of the contents as a sterilized package is opened and the contents are removed. 5 Containment device: Reusable rigid sterilization container, instrument case, cassette or organizing tray intended for use in healthcare facilities for the purpose of containing reusable medical devices for sterilization. 5 Reusable rigid sterilization container: Sterilization containment device designed to hold medical devices for sterilization, storage, transportation, and aseptic presentation of contents. 5 Organizing tray: Reusable metal or plastic containment device that organizes and protects instruments and components in specified locations within the device, and that is usually wrapped with an approved wrapping material. 5 Insert case or tray: Reusable case or tray that is placed inside a reusable rigid sterilization container, that is fixed or not fixed in place, and that serves to group or protect instruments and components. 5 Reusable accessory: Optional component not essential in itself but that aids in the organization and protection MANAGING INFECTION CONTROL 95
13 of instruments within a sterilization container system. Such items include, but are not limited to, insert cases, organizing trays, cassettes, brackets, posts, partitions, instrument mats, racks and stringers. 5 Instrument case/cassette: Sterilization containment device that consists of a lid and a base with means to allow sterilant penetration and removal, and that is enclosed in a sterile barrier system if sterility is to be maintained. 5 Product testing: Routinely testing processed items with biological and chemical indicators whenever major changes are made in packaging, wraps, or load configuration, such as dimensional changes, weight changes, or changes in the type or material of packaging or wrapper. 3 Biological indicator(bi): Test system containing viable microorganisms providing a defined resistance to a specified sterilization process. 3 Chemical indicator(ci): Device used to monitor the presence or attainment of one or more of the parameters required for a satisfactory sterilization process, or used in specific tests of sterilization equipment. 3 Drying time: Time required to dry steam sterilized items inside the sterilizer. 5 Dynamic-air-removal sterilizers: Type of sterilization cycle in which air is removed from the chamber and the load by means of a series of pressure and vacuum excursions (prevacuum cycle) or by means of a series of steam flushes and pressure pulses above atmospheric pressure (steam-flush pressure-pulse [SFPP] cycle). 1 Gravity-displacement steam sterilizers: Type of sterilization cycle in which incoming steam displaces residual air through a port or drain in or near the bottom (usually) of the sterilizer chamber. 3 Ethylene oxide: Type of gaseous chemical sterilization cycle in which the four process variables are ethylene oxide (EO) gas concentration, exposure time, temperature and relative humidity. 5 Hydrogen peroxide gas plasma: Type of gaseous chemical sterilization cycle in which the four process variables are hydrogen peroxide concentration, time, temperature and gas plasma power level. 5 Ozone sterilization: Type of gaseous chemical sterilization cycle in which the four process variables are ozone dose injected, humidity, exposure time and temperature. 5 Flash sterilization: Process designed for the steam sterilization of patient care items for immediate use. 5 Ordering Information AAMI ANSI/AAMI ST79:2006, Comprehensive guide to steam sterilization and sterility assurance in health care facilities Order code: ST79 or ST79-PDF Available in an attractive binder featuring sturdy metal rings, ledger-weight pages, and a laminated tab for each section for easy navigation. AAMI will issue revised pages that can be substituted into the binder when changes are made. Also available in PDF format and as part of AAMI s electronic CD and subscription products. Price/Member discount price: $200/$100. AAMI documents can be purchased through AAMI by credit card using the following four options: 1. Internet: 2. Call: or Fax: Mail: AAMI, Customer Service Center, 1100 N. Glebe Road, Suite 220, Arlington, VA AORN AORN Standards can be purchased through AORN using the following options: 1. Internet: 2. Call: ext. 1 or ext. 1 (Monday- Friday, 8AM to 4:30PM mountain standard time) 3. Fax: By mail: AORN, Inc., Customer Service/Book Orders, 2170 South Parker Road, Suite 300, Denver, CO , USA 5. Payment can be made by: VISA, MasterCard, American Express, or Discover, either online or by mail/fax/phone. By authorized PO, and AORN will bill you ($100 minimum purchase). Sorry, we cannot accept POs for education conferences or services. (Not available for online orders). 6. A CDRom of the standards is available for the first time this year. References 1. Central Service Technical Manual. 6th ed. Chicago, International Association of Healthcare Central Service materiel Management, 2003, p Recommended Practices for Selection and Use of Packaging Systems. Association of perioperative Registered Nurses. AORN Standards, Recommended Practices, and Guidelines The Association for the Advancement of Medical Instrumentation. Comprehensive guide to steam sterilization and sterility assurance in health care facilities. ANSI/AAMI ST79: Seavey R. Just Say No! Don t Get Weighed Down by Instrument Sets That Are Too Heavy. Managing Infection Control. 2006, April volume Six Issue 4, p The Association for the Advancement of Medical Instrumentation. Containment devices for reusable medical device sterilization. ANSI/AAMI ST77: Recommended Practices for Maintaining a Sterile Field. Association of perioperative Registered Nurses. AORN Standards, Recommended Practices, and Guidelines AORN. Sterile Processing/Materials Management Specialty Assembly Portal. Available at: FileSharing/index.fusion. 96 MANAGING INFECTION CONTROL
14 Sterile Process and Distribution CEU Information CEU Applicant Name Address City State Zip Code Rose Seavey, RN, BS, MBA, CNOR, ACSP, is the director of the sterile processing department at The Children s Hospital of Denver. Ms Seavey is an active member of the Association of perioperative Registered Nurses (AORN) and was elected to the National AORN Nominating Committee for She was honored with AORN s award for Outstanding Achievement in Clinical Nurse Education in She served as the President of the American Society of Healthcare Central Service Professionals (ASHCSP) in 2003 and is the 2002 recipient of the ASHCSP National Educator of the Year award. She is currently the chair for the AORN speciality assembly for Sterile Processing Materials Management. Ms. Seavey is a member of several AAMI working group committees that are developing recommended practices and is currently a co-chair for the AAMI ST8, Hospital steam sterilizers. In addition she has lectured and authored many articles on various topics relating to perioperative services and sterile processing, locally, nationally and internationally. The CBSPD (Certification Board for Sterile Processing and Distribution) has pre-approved this inservice for one and one-half (1.5) contact hours for a period of five (5) years from the date of publication. Successful completion of the lesson and post test must be documented by facility management and those records maintained by the individuals until recertification is required. DO NOT SEND LESSON OR TEST TO CBSPD. For additional information regarding Certification contact: CBSPD, 121 State Hwy 31N, Suite 500, Flemington, NJ or call or visit the Web site at IAHCSMM has awarded one and one-half (1.5) Contact Points for completion of this continuing education lesson toward IAHCSMM recertification. Nursing CE Application Form This inservice is approved by the California Board of Registered Nurses, CEP 5770 for one (1) contact hour. This form is valid up to one (1) year from the date of publication. 1. Make a photocopy of this form. 2. Print your name, address and daytime phone number and position/title. 3. Add the last 4 digits of your social security number or your nursing license number. 4. Date the application and sign. 5. Answer the true/false CE questions. 6. Submit this form and the answer sheet to: Workhorse Publishing Managing Infection Control PO Box 25310, Scottsdale, AZ For questions or follow-up, contact [email protected]. 8. Participants who score at least 70% will receive a certificate of completion within 30 days of Managing Infection Control s receipt of the application. Application Please print or type. Name Mailing Address City, State, Country, Zip Daytime phone ( ) Position/Title Social Security or Nursing License Number Date application submitted ANSWERS 1. T 2. T 3. F 4. T 5. F 6. T 7. T 8. T 9. T 10. F Signature Offer expires January 2013 On a scale of 1-5, 5 being Excellent and 1 being Poor, please rate this program for the following: 1) Overall content 2) Met written objectives 3) Usability of content Reprint with permission from Workhorse Publishing L.L.C. Copyright 2008/Workhorse Publishing L.L.C./All Rights Reseved. <01/08> MANAGING INFECTION CONTROL 97
THE BASICS OF PACKAGING
THE BASICS OF PACKAGING Objectives: Review the regulatory requirements for packaging Present the purpose, function and essential characteristics of packaging Describe packaging options, their use, and
Sterilization Packaging Systems, Preparation and Loading for Steam Sterilization
Sterilization Packaging Systems, Preparation and Loading for Steam Sterilization by Rose Seavey, RN, MBA, CNOR, ACSP Many thanks to the team at 3M Health Care for working with Managing Infection Contro
AST Standards of Practice for Packaging Material and Preparing Items for Sterilization
AST Standards of Practice for Packaging Material and Preparing Items for Sterilization Introduction The following Standards of Practice were researched and authored by the AST Education and Professional
Sterilization of Surgical Implants: Did You Know
Sterilization of Surgical Implants: Did You Know by Rose Seavey, RN, MBA, CNOR, ACSP Many thanks to the team at 3M Health Care for working with Managing Infection Control to provide the following accre
Heat Transfer. Condensation. Revaporization
Objectives Wrapping a Towel Around the Problem of Wet Packs Identify causes of wet packs in steam sterilization Review suggested documentation that can assist in solving wet pack problems Why are Wet Packs
1/22/2012. When is it a good time to learn about sterile processing in your facility? Continuing Education Contact Hours. Learning Objectives
Continuing Education Contact Hours Participants must complete the entire presentation/seminar to achieve successful completion and receive contact hour credit. Partial credit will not be given. All of
Best Practices for Loaner Surgical Instruments
Best Practices for Loaner Surgical Instruments Copyright 2015, Sterile Processing University, LLC, All Rights Reserved. This material may not be copied or used without permission of the author. Nancy Chobin,
THE BASICS OF STERILIZATION
THE BASICS OF STERILIZATION Objectives: State the importance of sterilization to patient care Review three key essentials of the sterilization process Describe sterilization methods used, application,
Biological Indicators and
& and BIs play an important role in the monitoring of sterilizers, along with physical and chemical indicators. However, only BIs contain live spores that directly measure the lethality of a sterilization
Getting a Handle on Loaner Instrumentation
Getting a Handle on Loaner Instrumentation Essentials for a successful loaner program by Rose Seavey, RN, MBA, CNOR, CRCST, CSPDT Objectives After completion of this self-study activity, the learner will
Infection Control in the Dental Practice through Proper Sterilization
Midmark Corporation 60 Vista Drive P.O. Box 286 Versailles, Ohio 45380-0286 937-526-3662 midmark.com Infection Control in the Dental Practice through Proper Sterilization In today's world, the need for
Body Art Facility Infection Prevention And Control Plan Guideline
Body Art Facility Infection Prevention And Control Plan Guideline In accordance with the California Health and Safety Code, Section 119313, a body art facility shall maintain and follow a written Infection
Steam Sterilization Cycles for Lab Applications
Steam Sterilization Cycles for Lab Applications Presented by Gary Butler STERIS Life Sciences August 2009 Early Steam Sterilizers Koch Upright Sterilizer First Pressurized Sterilizer First built by Pasteur-Chamberlain
Sterilization A REVIEW OF THE BASICS. By Gayla Tilton and Myrna Kauffman 66 MANAGING INFECTION CONTROL
Sterilization A REVIEW OF THE BASICS By Gayla Tilton and Myrna Kauffman 66 MANAGING INFECTION CONTROL Although sterilization has been performed in healthcare facilities for many decades, the process is
LOW-TEMPERATURE STERILIZATION IS AN EFFECTIVE MEANS TO
CRCST Self-Study Lesson Plan Lesson No. CRCST 138 (Technical Continuing Education - TCE) Sponsored by: by Susan Klacik, ACE, CHL, CRCST, CIS, FCS, CSS Manager, St. Elizabeth Health Center, Youngstown,
FACT SHEET : Using Autoclaves Safely
CSULA Environmental Health and Safety Biosafety Office FACT SHEET : Using Autoclaves Safely Most science research laboratories on campus require the use of autoclaves. The primary purpose of the autoclave
Revised 4-3-2012 EHS Biosafety. 1 Select appropriate containers/bags for autoclaving.
Requirements for Decontamination by Autoclaving Summary: Steam sterilization (autoclaving) is one of the most effective methods for decontaminating biohazardous material. Autoclaves use saturated steam
STERILIZATION WRAP. STERISHEET user guide for the validation of the packaging THE BACTERIAL BARRIER FOR HUMAN PROTECTION
STERILIZATION WRAP STERISHEET user guide for the validation of the packaging ARJOWIGGINS in few words A large range of sterilization wrap from standard creped paper to non-woven A large range of standard
Standard laryngoscope blade assemblies Directions for use
1 2014 Welch Allyn, Inc. MM 721358 Ver. C Standard laryngoscope blade assemblies Directions for use Intended use About this document The rigid laryngoscope is a device used to examine and visualize a patient
STANDARD OPERATING PROCEDURES DIVISION OF COMPARATIVE MEDICINE UNIVERSITY OF SOUTH FLORIDA
STANDARD OPERATING PROCEDURES DIVISION OF COMPARATIVE MEDICINE UNIVERSITY OF SOUTH FLORIDA SOP#: 1002.9 Date Issued: 10/98 Date Revised: 5/16 Page 1 of 5 TITLE: SCOPE: RESPONSIBILITY: PURPOSE: Animal Care
Improving the Quality of Surgical Trays. By: Stephen M. Kovach
Improving the Quality of Surgical Trays By: Stephen M. Kovach Performance measurement in healthcare represents what is done and how well it is done. The goal is to accurately understand the basis for current
Top Ten Sterilization Issues in an Ambulatory Surgery Center
10 Top Ten Sterilization Issues in an Ambulatory Surgery Center by Peggy Prinz Luebbert, MS, MT(ASCP), CIC, CHSP Objectives After completion of this self-study activity, the learner will be able to: 1.
3 Attest Biological Monitoring System
3 Attest Biological Monitoring System 1 Technical Product Profile A Technical Service Publication 2 Table of Contents Introduction...5 Product Description...5 Indicators... 5 Test Frequency... 8 Usage
CENTRAL SERVICE REGULATIONS FOR AMBULATORY SURGERY CENTERS
CENTRAL SERVICE REGULATIONS FOR AMBULATORY SURGERY CENTERS Last updated July 2014 Legal disclaimer: The International Association of Healthcare Central Service Materiel Management (IAHCSMM) provides the
EH&S. Sheet. Fact. Safe and Effective Use of Autoclaves. What are autoclaves? Factors for effective sterilization. Dry heat cycle - when to use
Please post or circulate Fact heet nvironment, ealth and afety Information for the Berkeley Campus No. 33 Revised 04/04/11 afe and ffective Use of Autoclaves Autoclaves are easy to use but can pose a safety
DISINFECTION & STERILIZATION INFECTION CONTROL GUIDELINES SECTION 2 CLEANING & PACKAGING
DISINFECTION & STERILIZATION INFECTION CONTROL GUIDELINES SECTION 2 CLEANING & PACKAGING Section 2- Cleaning and Packaging Cleaning of instruments and equipment Cleaning is defined as the removal of all
Fundamental Autoclave Techniques
Fundamental Autoclave Techniques Version 3, September 2010 Veronika Tatarinoff GSBmE University of NSW Brandy Nelson University of Kentucky The following information is directed at aseptic/sterile surgical,
STANDARD OPERATING PROCEDURES Safe Autoclave Operations
Department of Environmental Health & Safety Biosafety & Biosecurity STANDARD OPERATING PROCEDURES Safe Autoclave Operations The purpose of this document is to provide standard operating procedures for
UX-30 PERFORMANCE TESTING SUMMARY REPORT Revision 0
UX-30 PERFORMANCE TESTING SUMMARY REPORT Revision 0 Prepared By: Patrick L. Paquin, EnergySolutions Date Reviewed By: Mirza I. Baig, EnergySolutions Date Approved By: Thomas F. Dougherty, Columbiana Hi
"ADOPTED STANDARDS FOR THE REGULATION OF MEDICAL WASTE" IN HEALTH CARE FACILITIES LICENSED BY THE MISSISSIPPI STATE DEPARTMENT OF HEALTH
"ADOPTED STANDARDS FOR THE REGULATION OF MEDICAL WASTE" IN HEALTH CARE FACILITIES LICENSED BY THE MISSISSIPPI STATE DEPARTMENT OF HEALTH REGULATED MEDICAL WASTE "Infectious medical wastes" includes solid
Australian/New Zealand Standard
AS/NZS 4187:2003 AS/NZS 4187 Australian/New Zealand Standard Cleaning, disinfecting and sterilizing reusable medical and surgical instruments and equipment, and maintenance of associated environments in
Appendix J IBC Biohazard Spill Management Plan
OVERVIEW Prevention is the most important part of any spill management plan. Be sure to read and understand standard operating procedures (SOP s) and protocols for safe manipulation of biohazards before
Federal Wage System Job Grading Standard For Laboratory Working, 3511. Table of Contents
Federal Wage System Job Grading Standard For Laboratory Working, 3511 Table of Contents WORK COVERED... 2 WORK NOT COVERED...2 TITLES... 2 GRADE LEVELS... 2 NOTE TO USERS... 3 LABORATORY WORKER, GRADE
Air Removal from the Sterilization Chamber
TM Volume 5, Number 2 March 2008 Garrett Krushefski Scientific & Technical Services Manager Air Removal from the Sterilization Chamber I recently received a question from a customer asking why we state
#002. Flash Sterilization. S t u d y G u i d e. The Hot Issues of. A Continuing Education Study Guide presented by
#002 S t u d y G u i d e The Hot Issues of Flash Sterilization A Continuing Education Study Guide presented by For more information, visit the education page of our website www.steris.com The Hot Issue
Packaging Perishable Shipments
Packaging Perishable Shipments To help prevent spoilage and maintain product integrity, follow these instructions for shipping your perishable items via FedEx Express services. Preparing Perishable Items
Cleaning. By the end of this chapter, you will be able to: Introduction. Definitions. Chapter 9
Chapter 9 By the end of this chapter, you will be able to: l Define the terms cleaning, decontamination, disinfection l List the factors that affect disinfection l Describe the aims of a cleaning schedule
Sterile Dressing Change with Tegaderm CHG for Central Venous Catheter (CVC)
Sterile Dressing Change with Tegaderm CHG for Central Venous Catheter (CVC) The dressing protects your catheter site. It also helps prevent infection at the site. Keep your dressing clean and dry at all
Packaging Perishable Shipments
Packaging Perishable Shipments To help prevent spoilage and maintain product integrity, follow these instructions for shipping your perishable items via FedEx Express services. Preparing Perishable Items
3039878000or8009926372
3039878000or8009926372 bi s a l e s @me s a l a bs. c om Regulatory officials and sterilization experts have voiced concerns regarding the appropriateness of using a Biological Indicator (BI) Ampoule interchangeably
Sterilization methods and equipment Lab 1-2
Sterilization methods and equipment Lab 1-2 PHT 434 Sterilization Sterilization a process that by which all viable M.O are removed or destroyed, based on a probability function. Sterilization concept Sterilization
3. It is recommended that the water supply into the facility can be obtained from two (2) separate water lines if possible.
PART VIII SANITATION AND MEDICAL WASTE 132 SANITATION 132.01 Water Supply. 1. If at all possible, all water shall be obtained from a public water supply. If not possible to obtain water from a public water
RULE 1102. DRY CLEANERS USING SOLVENT OTHER THAN PERCHLOROETHYLENE
(Adopted January 6, 1978)(Amended August 3, 1979)(Amended July 11, 1980) (Amended August 3, 1990)(Amended December 7, 1990) (Amended November 17, 2000) RULE 1102. DRY CLEANERS USING SOLVENT OTHER THAN
Gear PEPSI CAN STOVE INSTRUCTIONS
Gear PEPSI CAN STOVE INSTRUCTIONS [NOTE: Updated Instructions are now available. The new stove is less likely to develop flame leaks and the fuel/air mixture is improved. Instructions for a simmer ring
Laboratory Line. Vertical and Bench-Top Life Science Autoclaves
Laboratory Line Vertical and Bench-Top Life Science Autoclaves The Advanced Laboratory Line The Full Spectrum of Laboratory Sterilizers Sterilization in a laboratory environment has its unique requirements.
HUMULIN 70/30 KwikPen
1 Instructions for Use HUMULIN 70/30 KwikPen (70% human insulin isophane suspension 30% human insulin injection [rdna origin]) Read the Instructions for Use before you start taking HUMULIN 70/30 and each
FLEXIBLE CIRCUITS MANUFACTURING
IPC-DVD-37 FLEXIBLE CIRCUITS MANUFACTURING Below is a copy of the narration for DVD-37. The contents of this script were developed by a review group of industry experts and were based on the best available
Autoclave Safety. Autoclaves are sterilizers using high pressure and high temperature steam. The potential safety risks for the operators are:
Autoclave Safety Purpose: Sterilization refers to the complete killing of all living organisms, including spores. Common sterilization techniques include the application of wet heat, dry heat, chemicals,
1.4 Metallic Impurities Iron must not exceed 150 ppm and copper shall not exceed 6 ppm when tested according to TAPPI T 266.
Library of Congress Preservation Directorate Specification Number 100 132 09 Specifications for 3 x 10ʺ Paper Enclosures For the Storage of Strip Photographic Negatives This specification is provided as
Set-Up Instructions. (Liner-Based System) UL LISTED. Important Read set-up instructions before assembly. Report any shortages within 72 hours
INDUSTRIAL POWER Set-Up Instructions (Liner-Based System) UL LISTED Important Read set-up instructions before assembly. Report any shortages within 72 hours Specializing in Environmental and Fire-Life
Potentially Infectious Medical Waste
Potentially Infectious Medical Waste A Summary of Regulatory Requirements General Requirements Title XV of the Illinois Environmental Protection Act (Act) establishes statutory requirements to ensure that
Laboratory Biosafty In Molecular Biology and its levels
Laboratory Biosafty In Molecular Biology and its levels Workshop 16-17 Oct..2012 Guidelines Does not mean optional Laboratory Biosafety The Laboratory Biosafety Manual is an important WHO publication
Oxivir Tb and Accelerated Hydrogen Peroxide (AHP) Frequently Asked Questions
Oxivir Tb and Accelerated Hydrogen Peroxide (AHP) Frequently Asked Questions What is Accelerated Hydrogen Peroxide (AHP)? AHP is a patented synergistic blend of commonly used, safe ingredients that when
-1- SPECIFICATIONS 002085 CONE SETTER PLATFORM ATTACHMENT INDEX
-1-002085 CONE SETTER PLATFORM ATTACHMENT INDEX I. GENERAL EQUIPMENT : A. Intent Statement B. Cone Setter Platform II. III. 1. Understructure Frame 2. Mounting Components 3. Paint 4. Basket 5. Labels 6.
BC Children s Hospital Mistake-Proofing Equipment Reprocessing. Laurence Bayzand Janice Penner Sue Fuller Blamey
BC Children s Hospital Mistake-Proofing Equipment Reprocessing Laurence Bayzand Janice Penner Sue Fuller Blamey Case for Collaboration Patient Safety Learning System (PSLS) trending identified an increased
DIY Advanced Dry Bags
DIY Advanced Dry Bags You can make dry bags to fit specific items, such as sleep pads, chairs, etc., by taking a few measurements and making allowances for seams, the bottom pleat and the roll-down closure.
Medical Market Laboratory EMEA
Medical Market Laboratory EMEA Space for 3M Montage Penetration tests according to the ISO 11140 Josephus P.C.M van Doornmalen Gomez Hoyos Croatia, March 27, 2009 Outline Purpose penetration tests Original
3M Dual Lock Reclosable Fastener
3 Dual Lock Reclosable Fasteners SJ3550CF (Type 250) SJ3551CF(Type 400) SJ3552CF(Type 170) Dual Lock Reclosable Fastener Black with a clear acrylic adhesive Product Information February, 2007 Product Description
Scotchgard Multi-Layer Protective Film 1004
Scotchgard Multi-Layer Protective Film 1004 3 Technical Bulletin September, 2008 Scotchgard Multi-Layer Protective Film 1004 Application Kit (3M Part No. 70-0710-4666-1) includes: (clockwise) Heavy Duty
INFECTION CONTROL SURVEYOR WORKSHEET
Exhibit 351 Ambulatory Surgical Center (ASC) INFECTION CONTROL SURVEYOR WORKSHEET (Rev: 142, Issued: 07-17-15, Effective: 07-17-15, Implementation: 07-17-15) Name of State Agency or AO (please specify)
Paper Chromatography: Separation and Identification of Five Metal Cations
Paper Chromatography: Separation and Identification of Five Metal Cations Objectives Known and unknown solutions of the metal ions Ag +, Fe 3+, Co 2+, Cu 2+ and Hg 2+ will be analyzed using paper chromatography.
2. The mold is closed up and held under hydraulic pressure while the rubber material or compound cures.
Designing with Rubber Molding Processes Compression Molding Compression molding is the process of placing a pre-load of a rubber material or compound directly in the mold cavity and compressed to the shape
COMPRESSED GASES. 1.2 The contents of each cylinder and container must be clearly identified (by tag or stamp) on the cylinder.
Page 1 of 5 COMPRESSED GASES A compressed gas is defined as any mixture of gases in a container with a pressure exceeding 40 psi. at 70 o F, or 104 psi. at 130 o F; or any flammable liquid with an absolute
KGI MEDICAL WASTE MANAGEMENT PLAN
1revised 2012 I. Facility Information KGI MEDICAL WASTE MANAGEMENT PLAN a. Contact Person: Barbara Erwin Director Research Operations/Chemical and Biological Laboratory Safety/Chair. 909-607-0160 i. Facility:
How to Build a Printed Circuit Board. Advanced Circuits Inc 2004
How to Build a Printed Circuit Board 1 This presentation is a work in progress. As methods and processes change it will be updated accordingly. It is intended only as an introduction to the production
Bed Cleaning Procedure
This is an official Northern Trust policy and should not be edited in any way Bed Cleaning Procedure Reference Number: NHSCT/10/308 Target audience: Nursing and Midwifery Staff Sources of advice in relation
Sun Windows General Information Sun New Construction Window Installation Instructions
Sun Windows General Information Section 1 G F H H H A. Apply Weather Resistive Barrier B. Apply Sill Flashing Tape C. Place Waterproof Sill Shims D. Caulk Back of Nail Fin Sides & Top, Set Window, Fasten
MODEL #12006. Introduction
THE ORIGINAL PECO POWER SPRAYER PECO OWNER S MANUAL MODEL #12006 (Q0080) Revised: 1/8/2014 PECO POWER SPRAYER MODEL #12006 TABLE OF CONTENTS SECTION PAGE SECTION PAGE INTRODUCTION - - - - - - - - - - -
STERILIZATION د STERILIZATION
STERILIZATION. د STERILIZATION Is a term referring to any process that eliminates (removes) or kills all forms of microbial life, including transmissible agents (such as fungi, bacteria, viruses, spores
Comparison Study of the Growth Promotion Capabilities of Self Contained Biological Indicator Culture Medium in Extended Steam Sterilization Cycles
Comparison Study of the Growth Promotion Capabilities of Self Contained Biological Indicator Culture Medium in Extended Steam Sterilization Cycles Charles Hughes, Gary Socola, Donald Tumminelli, and Mike
CHARISMA SHOWER/TUB DOOR INSTALLATION INSTRUCTIONS
CHARISMA SHOWER/TUB DOOR INSTALLATION INSTRUCTIONS IMPORTANT DreamLine reserves the right to alter, modify or redesign products at any time without prior notice. For the latest up-to-date technical drawings,
Navico-Northstar 2kW JRC Radar Package, Scanner Cable Removal and Replacement
Navico-Northstar 2kW JRC Radar Package, Scanner Cable Removal and Replacement This work instruction describes the methods and means for which to remove and reinstall optional scanner cable configurations
CEILING CASSETTE FAN COIL UNIT
CEILING CASSETTE FAN COIL UNIT Operation and Installation Manual MHCFC4W-04, 08, 12, 16 Four way ceiling cassette fan coil unit Please read this manual before using the fan coil unit. Installation and
WHITE PAPER April 18, 2016
WHITE PAPER April 18, 2016 Nitrogen Dioxide Biodecontamination: A New, Effective and Cost-saving Option for Biodecontaminating Syringe Tubs Prior to the Filling Line Prepared by: David Opie, PhD Maura
Sterile Cleanroom Management
Sterile Cleanroom Management Lynn Stanard, Sr. Quality Manager, Berkshire Corporation When manufacturing in an aseptic environment, it is critical to ensure that the various cleanroom consumables, such
SPECIFIC INSTRUCTIONS ON HOW TO PERIODICALLY CHECK FOR STRESS CRACKING CAUSED BY CHEMICALS, HANDLING ABUSE & ENVIRONMENTAL DEGRADATION
SPECIFIC INSTRUCTIONS ON HOW TO PERIODICALLY CHECK FOR STRESS CRACKING CAUSED BY CHEMICALS, HANDLING ABUSE & ENVIRONMENTAL DEGRADATION Provides the same capability for checking your liquid bulk storage
Assembly and Installation Procedures
Assembly and Installation Procedures for Pall Pharmaceutical Grade Capsule Assemblies 1. Introduction The following procedures must be followed for the installation of Pall pharmaceutical grade capsule
1.1.2 Polypropylene The polypropylene must be a white opaque film, 1 2 mil thick. Biaxial orientation is preferred.
Library of Congress Preservation Directorate Specification Number 700 704 09 Specifications for Pressure Sensitive Adhesive Labels For Application to Single Paper Sheets and Text Pages of Bound Books For
PATIENT GUIDE. Care and Maintenance Drainage Frequency: Max. Drainage Volume: Dressing Option: Clinician s Signature: Every drainage Weekly
PATIENT GUIDE Care and Maintenance Drainage Frequency: Max. Drainage Volume: Dressing Option: Every drainage Weekly Clinician s Signature: ACCESS SYSTEMS Pleural Space Insertion Site Cuff Exit Site Catheter
Reflectex TM Reflective Concrete Interlayer Geotextile INSTALLATION GUIDELINES
Reflectex TM Reflective Concrete Interlayer Geotextile INSTALLATION GUIDELINES Page 1 Table of Contents WHAT IS REFLECTEX TM?... 2 STORAGE AND HANDLING... 3 SITE PREPARATION... 3 PREPARATION OF EXISTING
INSTALLATION INSTRUCTIONS for Bifold Doors (JII103)
Thank you for selecting JELD-WEN products. Attached are JELD-WEN s recommended installation instructions for premium composite, hollow and solid core molded Bifold Doors. Bifolds are designed for fast
SCREAMIN' EAGLE "HEAVY BREATHER" PERFORMANCE AIR CLEANER KIT
REV. 05-04- SCREAMIN' EAGLE "HEAVY BREATHER" PERFORMANCE AIR CLEANER KIT GENERAL Kit Numbers 95-08B, 9006-09B, 876-0A Models For model fitment information, see the P&A Retail Catalog or the Parts and Accessories
Regulations & Standards
Regulations & Standards Impact Safe Instrument Processing by Rose Seavey, MBA, BS, RN, CNOR, CRCST, CSPDT Learning Objectives 1. Discuss the role of federal agencies in the regulation of processing procedures
Infectious Waste Management Plan
Infectious Waste Management Plan Infectious Waste Management Plan USC Health & Safety Programs Unit 777-5269 POLICY: A. In keeping with the University of South Carolina's policy of providing protection
Control Box Wiring For PRSstandard Tool
888-680-4466 ShopBotTools.com Control Box Wiring For PRSstandard Tool Copyright 2016 ShopBot Tools, Inc. page 1 Copyright 2016 ShopBot Tools, Inc. page 2 Table of Contents Introduction:...5 Installation:...5
Water should be removed as soon as possible once the safety of the structure has been verified.
FACT SHEET Posted October 18 2005 Check List for Infection Control Concerns when Reopening Healthcare Facilities Closed due to Extensive Water and Wind Damage Prior to opening a healthcare facility that
Horizontal Steam Sterilizers. Healthcare Line. Since
Horizontal Steam Sterilizers Healthcare Line Since 1925 Horizontal Steam Sterilizers for Healthcare Applications The Tuttnauer line of healthcare sterilizers are designed for hospital operating rooms,
NEW ELECTRICAL SUBSTATION Building 300 SECTION 07 53 23 ETHYLENE-PROPYLENE-DIENE-MONOMER ROOFING WORK
ETHYLENE-PROPYLENE-DIENE-MONOMER ROOFING WORK PART 1 GENERAL 1.1 REFERENCES The publications listed below form a part of this specification to the extent referenced. The publications are referred to within
RAIL ADAPTER SYSTEM DRAFT TC-23-AIMSS CHAPTER 1 1-1 DESCRIPTION
DRAFT TC-23-AIMSS CHAPTER 1 RAIL ADAPTER SYSTEM 1-1 DESCRIPTION The Rail Adapter System (RAS) consists of a set of lightweight specially modified sections that replace the standard front handguards on
CLEAN UP FOR VOMITING & DIARRHEAL EVENT IN RETAIL FOOD FACILITIES
CLEAN UP FOR VOMITING & DIARRHEAL EVENT IN RETAIL FOOD FACILITIES GENERAL INFORMATION Noroviruses are a group of viruses that cause gastroenteritis [gas-trō-en-ter-ī-tis] in people. Gastroenteritis is
Instructions for Use HUMALOG KwikPen insulin lispro injection (rdna origin)
1 Instructions for Use HUMALOG KwikPen insulin lispro injection (rdna origin) Read the Instructions for Use before you start taking HUMALOG and each time you get a refill. There may be new information.
Window Installation Instructions
Caution The correct installation methods of windows in full frame or insert applications is critical towards achieving the tested performance of the window and longterm enjoyment and energy savings for
A discussion of condensate removal systems for clarifier and thickener drives for water and wastewater facilities.
A discussion of condensate removal systems for clarifier and thickener drives for water and wastewater facilities. Richard L. Dornfeld, BSME, P.E. Staff Engineer September 25, 2014 Moisture can be a problem
Sampling Solutions. Fuel and Lube Oil sampling Equipment. Convenient, Representative, Approved.
Sampling Solutions Fuel and Lube Oil sampling Equipment. Convenient, Representative, Approved. Oil Samples are used to obtain a clear indication of the operating status of your machinery. One of the most
Fact Sheet on Steam Sterilizers at Stanford University
Fact Sheet on Steam Sterilizers at Stanford University By: Environmental Quality and Water Efficiency Group, Stanford University; December 2013 Jennifer Fitch, Marty Laporte, Julia Nussbaum, Adam Kern,
Installation Instructions
Installation Instructions READ BEFORE INSTALLING UNIT For Low Profile Window Air Conditioner INSTALLATION WARNINGS AND CAUTION Carefully read the installation manual before beginning. Follow each step
A. Work on this project shall consist of, but is not limited to, the following:
PART 1 - GENERAL CLEANING AND RESTORATION OF SECTION 05900-1 1.1 SUMMARY A. Work on this project shall consist of, but is not limited to, the following: 1. Clean and treat all exposed anodized architectural
SCENIC VIEW SV 10 SECTION 08 87 13 SOLAR CONTROL FILMS
PART 1 GENERAL SCENIC VIEW SV 10 SECTION 08 87 13 SOLAR CONTROL FILMS 1.1 SUMMARY A. Optically clear dual reflective, neutral sputtered and aluminum metallized polyester film for solar control. 1.2 RELATED