1 TM Volume 5, Number 2 March 2008 Garrett Krushefski Scientific & Technical Services Manager Air Removal from the Sterilization Chamber I recently received a question from a customer asking why we state in the EZTest instructions for use to Place the EZTest biological indicators in a horizontal position When sterilizing by steam, it is imperative to first remove ambient air not only from the chamber but also from the devices being sterilized. Failure to do so will impede steam contact with all surfaces which can result in non-sterile items being used from a seemingly acceptable sterilization cycle. There are two general means for removing ambient air from the sterilization chamber: dynamic air removal or gravity-displacement. Dynamic-air-removal sterilization cycles can be either a prevacuum cycle, in which case air is actively removed by a series of pressure and vacuum phases prior to admission of steam, or by steam-flush-pressure pulsing (SFPP) in which case a series of steam flushes and pulses of pressure are employed. Gravity-displacement is a passive means of air removal in which incoming steam displaces or pushes residual air out through the drain of the chamber. ANSI/AAMI ST79:2006 notes, The dynamic-air-removal cycle is generally preferred to a gravity-displacement cycle because of more efficient air removal, a shorter exposure time at higher temperatures, and a vacuum drying phase, resulting in an overall reduction in cycle time. While the above ANSI/AAMI note is accurate, do not think that you must start upgrading and replacing your gravity-displacement sterilizers with new machinery. Gravity displacement is still an effective means of sterilization as long as one employs a few precautions. Such precautions are not necessary in a dynamic-air-removal process because the air is actively sucked-out of the chamber and load. Without the aid of active air removal, one needs to facilitate removal of air from the load when using gravity-displacement systems. This is achieved by positioning items in such a manner that does not create any potential air traps. To illustrate this point, start
2 with a simple example such as a standard laboratory beaker. Of the three orientations shown in Figure 1, which are appropriate for a gravity-displacement steam sterilizer? Figure 1 A C B NOTE: In all of the photos non-packaged items are shown for visual clarity. In actual practice these items would be in sterilization pouches, allowing for post-exposure protection from contamination. The answer is (B & C). The ambient air that is contained within the beaker is heavier (i.e. more dense) than the steam that will be introduced into the chamber. In the upright orientation (A), the steam will not be able to displace or push the ambient air out of the beaker because the air has no means of escape (i.e. the beaker does not have a drain ). Note that the orientation shown for (B) does allow the air to drain from the beaker. The next example is an Erlenmeyer flask which is a bit trickier. Review the four orientations shown in Figure 2 and decide which are the most appropriate for a gravitydisplacement cycle.
3 Figure 2 D G E F It is easy to see that (F) or (G) is the best option and (D) must be avoided. But what about (E)? This orientation is also problematic because there is a zone within the Erlenmeyer flask where ambient air will become trapped and steam contact to the surfaces in that zone will be impeded (see outline of dashed red zone in Figure 3). A very effective way to demonstrate the effects of trapped air is to run a standard sterilization cycle with three flasks oriented as shown in D, E and F (see Figure 3 below). Place an EZTest biological indicator at each location marked in the photo and incubate the units after exposure. The EZTest from locations 1 and 2 will show surviving spores while the EZTest from location 3 will have no growth. This is a perfect example to show that the parametric data for the cycle (time, temperature, and pressure readings) will indicate acceptable cycle performance but the biological indicator indicates non-sterility.
4 Figure 3 D E F When conducting a validation several years ago, I performed the above recommended test in a 65-minute exposure. The penetration thermocouples (with attached EZTest BIs) recorded accumulated Fo values in excess of 80 minutes even from locations 1 and 2. The physical data indicated a lethal delivery that was well in excess of what was needed to kill the spore challenge, yet the EZTest from locations 1 and 2 were positive while location 3 was negative. How could the spores survive 80 Fo? The reason is the prevailing condition at locations 1 and 2 was NOT saturated steam; rather, it was C dry heat. Without an escape route, the less-dense steam molecules could not displace the heavier ambient air and air pockets (dashed red zones in Figure 3) become trapped in the flasks. The penetration thermocouples recorded temperatures of 121 to 123 C for most of the 65-minute dwell time, but unlike the spores in the biological indicator, the thermocouples can not differentiate between dry heat conditions and saturated steam. When reviewing the cycle data, time, temperature and pressure were all within acceptable levels, thus indicating acceptable cycle performance, yet the spores survived. The biological results indicated non-sterility and did not agree with the physical data which indicated sterilization was achieved. We refer to this condition as a catastrophic sterilization failure. SPORES DON T LIE! Air pockets within the materials being sterilized can not be detected by any instrumentation available today. The only system available to detect these pockets is the biological indicator. Air pockets occur in the worst case to sterilize locations in the
5 product or load. All standards direct the user to challenge the worst case locations with biological indicators. As a final example, consider the lengths of tubing shown in Figure 4. When oriented as shown in (G) multiple air pockets will form at various low spots along the length of the tube. One must carefully coil the tube in a helix shape (H) to facilitate air removal and steam penetration. Figure 4 G H Using colored water, one can illustrate the multiple low spots that exist in (G) where air pockets will form. For Figure 5, both tubes were filled with purple water and placed on the perforated stainless tray. All of the colored water drained from (H). Compare that to the amount and location of water that was unable to escape the (G) configuration. The colored water allows one to visually discern where the trapped air will form pockets and thus, the areas in the tube where steam will not penetrate.
6 Figure 5 G H The water visualization method is a valuable tool that can be used when training personnel to use steam sterilizers. One must position items in the chamber such that if they were filled with water, the water would completely drain from the item. If any water were retained, the positioning is not suitable (as shown for the Erlenmeyer flask in Figure 6).
7 Figure 6 The colored water is unable to drain from the flask and shows where trapped air will collect and impede steam contact. Now back to the original question, Why is it necessary to place the EZTest biological indicators in a horizontal position? The EZTest, by virtue of its design is an excellent detector of trapped air. The EZTest itself is not unlike the Erlenmeyer flask in that it too can harbor ambient air which would impede steam contact with the spores on the inoculated paper strip. Thus, just like the beaker shown in (B), this is the correct orientation for the EZTest BI. Is it necessary to use the horizontal orientation in a dynamic-air-removal sterilization cycle? In dynamic-air-removal cycles, one is relying on the prevacuum or SFPP to remove air from the chamber, the load and the EZTest (which may or may not be in a test pack). Regardless, it is always a good idea to use the horizontal orientation for your EZTest as it does help avoid inadvertent trapping of air, even in dynamic-air-removal systems.
8 Finally, consider Figure 7; can you see the flaw in (I) and why (J) is the preferred set-up? I J Figure 7 Recall the earlier statement, the EZTest itself is not unlike the Erlenmeyer flask in that it too can harbor ambient air which would impede steam contact with the spores on the inoculated paper strip. In both (I & J) above, the Erlenmeyer flask is inverted to the extent that if filled with water, all of the water would drain from the flask. HOWEVER, in (I) the EZTest is in a slight cap-up orientation and if the EZTest biological indicator were filled with water, it would be trapped inside the EZTest. In (J) a slight cap-down orientation prevails (by virtue of the inverted positioning of the Erlenmeyer flask) and thus we have a drain which allows the ambient air from inside the EZTest to escape the unit. SPORES DON T LIE Please us with topics you would like to see addressed in Spore News.
3039878000or8009926372
3039878000or8009926372 bi s a l e s @me s a l a bs. c om Regulatory officials and sterilization experts have voiced concerns regarding the appropriateness of using a Biological Indicator (BI) Ampoule interchangeably
Steam Sterilization Cycles for Lab Applications
Steam Sterilization Cycles for Lab Applications Presented by Gary Butler STERIS Life Sciences August 2009 Early Steam Sterilizers Koch Upright Sterilizer First Pressurized Sterilizer First built by Pasteur-Chamberlain
Biological Indicators and
& and BIs play an important role in the monitoring of sterilizers, along with physical and chemical indicators. However, only BIs contain live spores that directly measure the lethality of a sterilization
Steam Sterilization and the 2007 Revision of PDA Technical Report 1
Steam Sterilization and the 2007 Revision of PDA Technical Report 1 14 November 2007 Presented By: Mike Finger (Tunnell Consulting) Don Drew (Abbott Bioresearch Center) Agenda: Introduction Sterilization
3 Attest Biological Monitoring System
3 Attest Biological Monitoring System 1 Technical Product Profile A Technical Service Publication 2 Table of Contents Introduction...5 Product Description...5 Indicators... 5 Test Frequency... 8 Usage
AUTOCLAVE STEAM STERILIZERS
AUTOCLAVE STEAM STERILIZERS (BENCH TOP) Image: HS-1606VD Hanshin Bench Top Steam Sterilizer Small and easy to operate Three preset sterilization cycles Adjustable cycle presets Rapid heat up times Option
Comparison Study of the Growth Promotion Capabilities of Self Contained Biological Indicator Culture Medium in Extended Steam Sterilization Cycles
Comparison Study of the Growth Promotion Capabilities of Self Contained Biological Indicator Culture Medium in Extended Steam Sterilization Cycles Charles Hughes, Gary Socola, Donald Tumminelli, and Mike
Because You Can t See Sterile.SM
Because You Can t See Sterile.SM Sterilization EVERY LOAD. Sure-Check Sterilization Pouches 3 / 29 PMS 201C EVERY DAY. 50% BLACK STEAMPlus Class 5 Sterilization Integrators EVERY WEEK. ConFirm 10 In-Office
Steam Sterilization Principles
Reprinted from PHARMACEUTICAL ENGINEERING THE OFFICIAL TECHNICAL MAGAZINE OF ISPE NOVEMBER/DECEMBER 2013, VOL 33, NO 6 Copyright ISPE 2013 www.pharmaceuticalengineering.org facilities and equipment by
Lab Exercise 3: Media, incubation, and aseptic technique
Lab Exercise 3: Media, incubation, and aseptic technique Objectives 1. Compare the different types of media. 2. Describe the different formats of media, plate, tube etc. 3. Explain how to sterilize it,
Sterilization of narrow channels: how to ensure effectiveness in daily routine? Dr.ir. Josephus Paulus Clemens Maria Van Doornmalen Gomez Hoyos
April 2015 Sterilization of narrow channels: how to ensure effectiveness in daily routine? Dr.ir. Josephus Paulus Clemens Maria Van Doornmalen Gomez Hoyos 3M Deutschand GmbH, IPD, Decontamination University
Laboratory Line. Vertical and Bench-Top Life Science Autoclaves
Laboratory Line Vertical and Bench-Top Life Science Autoclaves The Advanced Laboratory Line The Full Spectrum of Laboratory Sterilizers Sterilization in a laboratory environment has its unique requirements.
Medical Market Laboratory EMEA
Medical Market Laboratory EMEA Space for 3M Montage Penetration tests in according to the ISO 11140 Josephus P.C.M van Doornmalen Gomez Hoyos Croatia, March 27, 2009 Outline Purpose penetration tests Original
Infection Control in the Dental Practice through Proper Sterilization
Midmark Corporation 60 Vista Drive P.O. Box 286 Versailles, Ohio 45380-0286 937-526-3662 midmark.com Infection Control in the Dental Practice through Proper Sterilization In today's world, the need for
FACT SHEET : Using Autoclaves Safely
CSULA Environmental Health and Safety Biosafety Office FACT SHEET : Using Autoclaves Safely Most science research laboratories on campus require the use of autoclaves. The primary purpose of the autoclave
Application Note USTR 805. Steam Sterilization of Pall Filter Assemblies Utilizing Replaceable Filter Cartridges
Application Note USTR 805 Steam Sterilization of Pall Filter Assemblies Utilizing Replaceable Filter Cartridges 0 Table of Contents Table of Contents 1. Important Recommendations...1 1.1 Installation
THE BASICS OF STERILIZATION
THE BASICS OF STERILIZATION Objectives: State the importance of sterilization to patient care Review three key essentials of the sterilization process Describe sterilization methods used, application,
Vertical and Benchtop Autoclaves for Life Sciences. Laboratory Line
Vertical and Benchtop Autoclaves for Life Sciences Laboratory Line Designed for Your Laboratory Applications Advanced Laboratory Autoclaves Tuttnauer laboratory autoclaves have been designed to provide
STANDARD OPERATING PROCEDURES DIVISION OF COMPARATIVE MEDICINE UNIVERSITY OF SOUTH FLORIDA
STANDARD OPERATING PROCEDURES DIVISION OF COMPARATIVE MEDICINE UNIVERSITY OF SOUTH FLORIDA SOP#: 1002.9 Date Issued: 10/98 Date Revised: 5/16 Page 1 of 5 TITLE: SCOPE: RESPONSIBILITY: PURPOSE: Animal Care
Sterile Cleanroom Management
Sterile Cleanroom Management Lynn Stanard, Sr. Quality Manager, Berkshire Corporation When manufacturing in an aseptic environment, it is critical to ensure that the various cleanroom consumables, such
Sterilization A REVIEW OF THE BASICS. By Gayla Tilton and Myrna Kauffman 66 MANAGING INFECTION CONTROL
Sterilization A REVIEW OF THE BASICS By Gayla Tilton and Myrna Kauffman 66 MANAGING INFECTION CONTROL Although sterilization has been performed in healthcare facilities for many decades, the process is
Medical Market Laboratory EMEA
Medical Market Laboratory EMEA Space for 3M Montage Penetration tests according to the ISO 11140 Josephus P.C.M van Doornmalen Gomez Hoyos Croatia, March 27, 2009 Outline Purpose penetration tests Original
Fact Sheet on Steam Sterilizers at Stanford University
Fact Sheet on Steam Sterilizers at Stanford University By: Environmental Quality and Water Efficiency Group, Stanford University; December 2013 Jennifer Fitch, Marty Laporte, Julia Nussbaum, Adam Kern,
Federal Wage System Job Grading Standard For Laboratory Working, 3511. Table of Contents
Federal Wage System Job Grading Standard For Laboratory Working, 3511 Table of Contents WORK COVERED... 2 WORK NOT COVERED...2 TITLES... 2 GRADE LEVELS... 2 NOTE TO USERS... 3 LABORATORY WORKER, GRADE
Penetration tests according to the ISO 11140
Space for 3M Montage Penetration tests according to the ISO 11140 Josephus Paulus Clemens Maria van Doornmalen Gomez Hoyos Outline Purpose penetration tests ETS Alternatives for original Bowie and Dick
Calibration of Volumetric Glassware
Chemistry 119: Experiment 2 Calibration of Volumetric Glassware For making accurate measurements in analytical procedures, next in importance to the balance is volumetric equipment. In this section volumetric
Top 10 considerations when validating an autoclave
Top 10 considerations when validating an autoclave Validating an autoclave is a daunting and time-consuming task. This white paper details the tricks, tips and traps to such a validation project from how
STEAM HEATING SYSTEM TROUBLESHOOTING GUIDE
Page 1 of 9 PURPOSE Steam is the most commonly used heating medium for maintaining process temperatures. Compared to other heating media, steam contains a significant amount of heat energy, and this heat
1/22/2012. When is it a good time to learn about sterile processing in your facility? Continuing Education Contact Hours. Learning Objectives
Continuing Education Contact Hours Participants must complete the entire presentation/seminar to achieve successful completion and receive contact hour credit. Partial credit will not be given. All of
STANDARD OPERATING PROCEDURES Safe Autoclave Operations
Department of Environmental Health & Safety Biosafety & Biosecurity STANDARD OPERATING PROCEDURES Safe Autoclave Operations The purpose of this document is to provide standard operating procedures for
GUIDE TO AUTOCLAVING PIONEERS IN STERILIZATION TECHNOLOGY
GUIDE TO AUTOCLAVING PIONEERS IN STERILIZATION TECHNOLOGY What is Sterilization? Sterilization may be defined as the statistically complete destruction of all microorganisms including the most resistant
Effective Use of Autoclaves
Effective Use of s Overview An autoclave is a commonly used piece of equipment in microbiology laboratories. s (steam sterilizers) are metal pressure vessels that are used for steam sterilization of media,
The Molar Mass of a Gas
The Molar Mass of a Gas Goals The purpose of this experiment is to determine the number of grams per mole of a gas by measuring the pressure, volume, temperature, and mass of a sample. Terms to Know Molar
KGI MEDICAL WASTE MANAGEMENT PLAN
1revised 2012 I. Facility Information KGI MEDICAL WASTE MANAGEMENT PLAN a. Contact Person: Barbara Erwin Director Research Operations/Chemical and Biological Laboratory Safety/Chair. 909-607-0160 i. Facility:
EH&S. Sheet. Fact. Safe and Effective Use of Autoclaves. What are autoclaves? Factors for effective sterilization. Dry heat cycle - when to use
Please post or circulate Fact heet nvironment, ealth and afety Information for the Berkeley Campus No. 33 Revised 04/04/11 afe and ffective Use of Autoclaves Autoclaves are easy to use but can pose a safety
Effective Heat Sterilization in CO 2 Incubators
Effective Heat Sterilization in CO 2 Incubators Volume 4, Number 3 Key Words Key Words Class 100 air Contamination control Contamination elimination HEPA filtration Incubator Heat Sterilization White Paper
Sterilizers with extra class
Sterilizers with extra class ClassB Ready for the future The European standard EN 13060 specifies the performance requirements and test methods for small steam sterilizers. These are used for hygiene and
Validation Guide. USTR 805j (5) Steam sterilization of Pall filter assemblies which utilize replaceable filter cartridges
Validation Guide USR 805j (5) Steam sterilization of Pall filter assemblies which utilize replaceable filter cartridges CONENS 1. Important Recommendations 4 1.1 Installation 4 1.2 Integrity esting 4 1.3
Decontamination and Waste Management www.biosecurity.sandia.gov
Decontamination and Waste Management www.biosecurity.sandia.gov SAND No. 2006-3684C Sandia is a multiprogram laboratory operated by Sandia Corporation, a Lockheed Martin Company, for the United States
The Supplier Perspective: Sterilization. Presented by: Thad Wroblewski STERIS Isomedix Services
The Supplier Perspective: Sterilization Presented by: Thad Wroblewski STERIS Isomedix Services Agenda Introduction Contractor selection criteria Pre-planning with sterilization in mind Key elements of
TECHNIQUES FOR STEAM STERILIZING LABORATORY WASTE
TECHNIQUES FOR STEAM STERILIZING LABORATORY WASTE by Timothy E. Cooney Research Scientist AMSCO Research and Development SUMMARY The traditional method for steam sterilizing petri dishes and other laboratory
ArtisanLink Staining System is an automated special stains slide
Technical Tips Tips on using the ArtisanLink Special Staining System Jamie Nowacek, BS, HT(ASCP) CM, QIHC, PMP Dako North America, Inc. Carpinteria, CA, USA ArtisanLink Staining System is an automated
PORTABLE LABORATORY MODEL AVAILABLE!
S E L E C T I O N G U I D E LARGEST TABLETOP STERILIZER AVAILABLE! EASY TO PROGRAM TABLETOP STERILIZERS! QUICK PURE WATER FOR YOUR STEAM STERILIZER! FAST INDICATION AND EASY TO USE! HARVEY STERILIZERS
Revised 4-3-2012 EHS Biosafety. 1 Select appropriate containers/bags for autoclaving.
Requirements for Decontamination by Autoclaving Summary: Steam sterilization (autoclaving) is one of the most effective methods for decontaminating biohazardous material. Autoclaves use saturated steam
OPERATOR MANUAL Amsco Century TM Medium Steam Sterilizer 26" x 26" (660 x 660 mm) Scientific. (12/16/05) P129373-635 Rev. 0
OPERATOR MANUAL Amsco Century TM Medium Steam Sterilizer 26" x 26" (660 x 660 mm) Scientific (12/16/05) P129373-635 Rev. 0 A WORD FROM STERIS CORPORATION This manual contains important information on
TUTTNAUER OLD AUTOCLAVE OPERATING PROCEDURE
1 of 6 TUTTNAUER OLD AUTOCLAVE OPERATING PROCEDURE Aim The aim of this operating procedure is to equip the user with the knowledge to safely and successfully operate the old pressure chamber. This autoclave
Sterilization methods and equipment Lab 1-2
Sterilization methods and equipment Lab 1-2 PHT 434 Sterilization Sterilization a process that by which all viable M.O are removed or destroyed, based on a probability function. Sterilization concept Sterilization
BESTPRACTICE. HTM01-05 Q&A: You can now follow us on Twitter! We have launched our HTM01-05 questions and answers service on Twitter.
ISSUE 4 BESTPRACTICE For Essential Requirements, hand washing is still a viable method of cleaning your instruments, but how do you validate for compliance with HTM01-05? Ensuring your instruments are
Bench Autoclave. Standard Operating Procedure. For Installation, Use and Maintenance
Bench Autoclave Standard Operating Procedure For Installation, Use and Maintenance 1. Introduction This SOP is intended for use with the following model, in a laboratory context: Type: Nuve Bench Top Steam
Oil Sight Glass BENEFITS
8 L-4090 Schroeder es provide maintenance and lubrication management professionals a complete and immediate visual oil analysis. Constructed of durable cast acrylic, they withstand most petroleum products
Model Voltage (V/Hz) Power (W) Capacity (L) Chamber size Outside size(mm) Weight (kg) STE-TEN 220/50 1100 1 Ø90x160mm W200xH385xD450 30 (Net:28)
Table Top STE-TEN STE-TEN, 1 Liter Autoclave It adopts B-wclass (pred-13060), with 3-times pre-vacuum preceding vacuum drying. Residual humidity is 0.2, vacuum measurement reaches -0.7bar. It sterilizes
AUTOCLAVES. Table Top. STE-TEN, 1 Liter Autoclave
Table Top STE-TEN, 1 Liter utoclave STE-TEN It adopts B-wclass (pred-13060), with 3-times pre-vacuum preceding vacuum drying. Residual humidity is 0.2, vacuum measurement reaches -0.7bar. It sterilizes
DISINFECTION & STERILIZATION INFECTION CONTROL GUIDELINES SECTION 4 QUALITY MANAGEMENT IN STERILIZIATION
DISINFECTION & STERILIZATION INFECTION CONTROL GUIDELINES SECTION 4 QUALITY MANAGEMENT IN STERILIZIATION Section 4 - Quality Management in Sterilization Quality management in sterilizing services Quality
AUTOCLAVE PROGRAM. SOP Bio-006 FOR THE USE OF AUTOCLAVE FOR STERILIZATION OF MATERIALS AND BIOLOGICAL WASTE SOP
ENVIRONMENTAL AND EMERGENCY MANAGEMENT ENVIRONMENTAL HEALTH AND SAFETY 175 Cabot St. Wannalancit Suite 311 Lowell MA 01854 Ruth Medina, Ph.D. Tel: 978-934-2778 Senior Biosafety Specialist/Biosafety Officer
Partnership. Large Steam Autoclaves for the Life Sciences. Laboratory Line
Partnership Large Steam Autoclaves for the Life Sciences Laboratory Line Large Steam Autoclaves for the Life Sciences Sterilization in a laboratory environment has its unique requirements. Tuttnauer laboratory
Water Baths. with BioCote antimicrobial protection
Water Baths with BioCote antimicrobial protection Water Baths are one of the most common pieces of laboratory equipment used today, the Stuart range offers three sizes 6, 15 and 24L with digital control.
STERILIZATION AND DISINFECTION
STERILIZATION AND DISINFECTION Importance of hand washing shown by Semmelweis STERILIZATION A physical or chemical process that destroys or eliminates all forms of microbial life, including spores. A satisfactory
Laboratory & Scientific Line. Bulk Steam Sterilizers. Since
Laboratory & Scientific Line Bulk Steam Sterilizers Since 1925 Bulk Steam Sterilizers for Laboratory & Scientific Applications The Tuttnauer line of bulk autoclaves are designed to handle the most critical
Assembly and Installation Procedures
Assembly and Installation Procedures for Pall Pharmaceutical Grade Capsule Assemblies 1. Introduction The following procedures must be followed for the installation of Pall pharmaceutical grade capsule
Horizontal Sterilizer MU 012.B / I (09/07) Page 2/46. 1 Index... 2
0932 Horizontal Sterilizer MU 012.B / I (09/07) Page 2/46 1 Index 1 Index... 2 2 Description of the Inputs / Outputs... 4 2.1 Inputs / Module OMRON ID201 / ID211...4 2.2 Outputs / Module OMRON OC211...5
Heat Transfer. Condensation. Revaporization
Objectives Wrapping a Towel Around the Problem of Wet Packs Identify causes of wet packs in steam sterilization Review suggested documentation that can assist in solving wet pack problems Why are Wet Packs
Managing Steam Sterilization Process By Patricia H. Nothum, RN, BSN, CRCST, CHL
Managing Steam Sterilization Process By Patricia H. Nothum, RN, BSN, CRCST, CHL A research paper submitted to the International Association of Central Service Health Care Material Management as Partial
Sterile E Online NetworkTutorials
3M Sterilization Assurance Products EO Blue bar must enter area Sterile E Online NetworkTutorials EO Blue bar must enter area Understanding the Differences Between Class 5 s and Class 6 Emulating Indicators
LABOKLAV. Autoclaves Made in Germany. www.shp-steriltechnik.de
LABOKLAV Autoclaves Made in Germany www.shp-steriltechnik.de 2 LABOKLAV STEAM STERILIZERS Steam sterilization is the sterilization method with the widest range of applications efficient, safe, reproducible
Chemical Indicators & Accessories from Albert Browne International Ltd. Assurance through Technology
Chemical & Accessories from Albert Browne International Ltd Assurance Assurance What are Browne Chemical? Chemical The Browne range of chemical indicators has been categorised into six areas, broadly defined
Model Voltage (V/Hz) Power (W) Capacity (L) Chamber size Outside size(mm) Weight (kg) STE-TEN 220/50 1100 1 Ø90x160mm W200xH385xD450 30 (Net:28)
Table Top STE-TEN STE-TEN, 1 Liter Autoclave It adopts B-class (pred-13060), with 3-times pre-vacuum preceding vacuum drying. Residual humidity is 0.2, vacuum measurement reaches -0.7bar. It sterilizes
The Power of Innovation. For better Sterilization. Pass-through autoclaves: Systec D-Series 2D Horizontal front loading bench-top autoclaves: Systec
The Power of Innovation. For better Sterilization. Pass-through autoclaves: Systec D-Series 2D Horizontal front loading bench-top autoclaves: Systec D-Series t h e a u t o c l a v e c o m p a n y Systec
Steripro cgmp. Steam Sterilizers
Steripro cgmp Steam Sterilizers Steripro cgmp Steam Sterilizers BMT worldwide QUALITY, RELIABILITY, INNOVATIO N BMT USA is a member of the MMM Group, one of the world s leading suppliers of sterilization
Experiment 3: Extraction: Separation of an Acidic, a Basic and a Neutral Substance
1 Experiment 3: Extraction: Separation of an Acidic, a Basic and a Neutral Substance Read pp 142-155, 161-162, Chapter 10 and pp 163-173, Chapter 11, in LTOC. View the videos: 4.2 Extraction (Macroscale);
Thermo Scientific VARIOKLAV Modular Laboratory Autoclaves Versatile, space-saving performance for steam sterilization applications
Thermo Scientific VARIOKLAV Modular Laboratory Autoclaves Versatile, space-saving performance for steam sterilization applications Part of Thermo Fisher Scientific Thermo Scientific Varioklav Laboratory
Vacuum Tube Solar Water heater Non-pressure. User manual
Vacuum Tube Solar Water heater Non-pressure User manual Please read this manual carefully before beginning installation Specification Tubes 12 --- 30 58X1800 Nominal volume tank capacity 100L --- 300L
2. Laboratory equipment
2. Laboratory equipment The choice of laboratory equipment must take into account national regulations and the technical premises that determine the requirements for its appropriate use. The more com plex
OPERATOR MANUAL Lab 250 and Lab 110 Life Sciences Small Sterilizers. (11/13/06) P387352-364 Rev. 0
OPERATOR MANUAL Lab 250 and Lab 110 Life Sciences Small Sterilizers (11/13/06) P387352-364 Rev. 0 A WORD FROM STERIS This manual contains important information on proper use of this Lab 250 and 110 Series
Autoclave Safety. Autoclaves are sterilizers using high pressure and high temperature steam. The potential safety risks for the operators are:
Autoclave Safety Purpose: Sterilization refers to the complete killing of all living organisms, including spores. Common sterilization techniques include the application of wet heat, dry heat, chemicals,
Total Heat Versus Sensible Heat Evaporator Selection Methods & Application
Total Heat Versus Sensible Heat Evaporator Selection Methods & Application Scope The purpose of this paper is to provide specifying engineers, purchasers and users of evaporators in industrial refrigeration
Microbiological Evaluation of the STI Series 2000 Medical Waste Treatment Process
WNWN International,Inc. WNWN International Phone: 860-675-1217 Fax 860-675-1311 PO Box 1164 Burlington, CT. 06013 USA Microbiological Evaluation of the STI Series 2000 Medical Waste Treatment Process January
FVA THIRD-GENERATION VERTICAL SATURATED-STEAM STERILIZERS
The key difference between a small vertical or bench-top sterilizer and industrial sterilizer is performance. Small saturated-steam sterilizers with vertical or horizontal chambers were devised many years
A GUIDELINE FOR THE SAFE USE OF AUTOCLAVES UNIVERSITY OF OTTAWA. Environmental Health and Safety Service
A GUIDELINE FOR THE SAFE USE OF AUTOCLAVES UNIVERSITY OF OTTAWA Environmental Health and Safety Service Last date of revision July 9, 2003 TABLE OF CONTENT 1.0 PURPOSE...1 2.0 INTRODUCTION...1 3.0 PERSONAL
Laboratory- and sterilisation equipment HMC EUROPE
Laboratory- and sterilisation equipment HMC PROFILE HMC Laboratory- and sterilisation equipment Consistent marketing of technologies and the conversion of new ideas create product ranges, that are tailor-made
Laboratory- and sterilisation equipment HMC EUROPE
Laboratory- and sterilisation equipment HMC EUROPE PROFILE HMC EUROPE Laboratory- and sterilisation equipment Consistent marketing of technologies and the conversion of new ideas create product ranges,
Autoclave Testing Service, Inc.
Autoclave Testing Service, Inc. Sterilizer Monitoring Since 1986 Toll Free 877/287-1199 Celebrating 28 Years 3 Serving the healthcare industry since 1986 One of the first mail-to-lab spore testing companies
OPERATION OF THE AUTOCLAVES
OPERATION OF THE AUTOCLAVES Part 1. General autoclave wisdom from Howard Part 2. Specific instructions for the "new" and "old" autoclaves An introduction to happy autoclaving The sterilization of materials
the stainless steel water-heater for your woodstove USER S MANUAL
the stainless steel water-heater for your woodstove USER S MANUAL Get Into Hot Water... Heating water while heating the home with renewable wood has been done for centuries and is now reemmerging as a
Laboratory & Scientific Line. Horizontal Steam Sterilizers
Laboratory & Scientific Line Horizontal Steam Sterilizers Horizontal Steam Sterilizers for Laboratory & Scientific Applications Sterilization in a laboratory environment has its unique requirements. The
Autoclave Quality Control Policy
Northwestern University Autoclave Quality Control Policy Office of Research Safety Office of the Vice President for Research Oct 2010, Aug 2005, Aug 2004, May 2002 1 Table of Contents 1.0 INTRODUCTION
TMI-Orion CERAMICS TILES BRICKS VALUE PROPOSITION HIGH TECHNOLOGY DATA LOGGING SOLUTIONS. www.tmi-orion.com
TMI-Orion CERAMICS TILES BRICKS VALUE PROPOSITION HIGH TECHNOLOGY DATA LOGGING SOLUTIONS www.tmi-orion.com TMI-Orion Ceramics Tiles Bricks TMI-Orion presents a full range of high technology data loggers
Pure Solid Compounds: Molecules held together in rigid formations by intermolecular forces.
Recrystallization: Purification of Solid Compounds Pure Solid Compounds: Molecules held together in rigid formations by intermolecular forces. Types of Intermolecular Forces? 1. Van der Waal s forces London
#002. Flash Sterilization. S t u d y G u i d e. The Hot Issues of. A Continuing Education Study Guide presented by
#002 S t u d y G u i d e The Hot Issues of Flash Sterilization A Continuing Education Study Guide presented by For more information, visit the education page of our website www.steris.com The Hot Issue
Fundamental Autoclave Techniques
Fundamental Autoclave Techniques Version 3, September 2010 Veronika Tatarinoff GSBmE University of NSW Brandy Nelson University of Kentucky The following information is directed at aseptic/sterile surgical,
IDENTIFICATION BROCHURE FOR Central RESIDENTIAL WET SPRINKLERS
IDENTIFICATION BROCHURE FOR Central RESIDENTIAL WET SPRINKLERS 1 INTRODUCTION TO THE VOLUNTARY REPLACEMENT PROGRAM This brochure provides information that will assist you in identifying whether the sprinklers
Standard laryngoscope blade assemblies Directions for use
1 2014 Welch Allyn, Inc. MM 721358 Ver. C Standard laryngoscope blade assemblies Directions for use Intended use About this document The rigid laryngoscope is a device used to examine and visualize a patient
Subpart 1. Installation. All plumbing systems must be. installed and tested according to this chapter and chapter 4715,
4658.4500 PLUMBING SYSTEMS; NEW CONSTRUCTION. Subpart 1. Installation. All plumbing systems must be installed and tested according to this chapter and chapter 4715, the Minnesota Plumbing Code. Subp. 2.
STANDARD OPERATING PROCEDURE
STANDARD OPERATING PROCEDURE Title: Evaluation using Western Blot SOP#: M-103 Version #: 1 Author: R. Saul Date Approved: Feb. 5, 2009 Date Modified: 1. PURPOSE The purpose of this document is to describe
STEAM STERILIZERS. DLOV & DLOM series. more than fifty years of experience in sterilization. DE LAMA s Quality Management System is certified:
STEAM STERILIZERS DLOV & DLOM series DE LAMA s Quality Management System is certified: more than fifty years of experience in sterilization UNI EN ISO 9001:2000 - Cert. n 0351 UNI EN ISO 13485 - Cert.
Guidance on safe use of Autoclaves
Safety Office Reviewed: Dec 2013 Reviewed: DD Month Year Guidance on safe use of Autoclaves 1. Hazards & Uses There are several different types of hazard associated with the use of autoclaves. The main
BSL3 and BSL4 Autoclaves. Laboratory Line
BSL3 and BSL4 Autoclaves Laboratory Line BSL3 and BSL4 Autoclaves In today s society the need for safe bio-containment systems is increasingly important. This is especially true in the sterilization process
Medical Waste Autoclaves. Medical Waste Autoclaves
Medical Waste Autoclaves Medical Waste Autoclaves Safe, Cost-Effective, Medical Waste Sterilization Treat Medical Waste with high performance Tuttnauer Medical Waste autoclave systems designed for small,
Refrigeration Installation 101
Refrigeration Installation 101 Hundreds of Installations Refrigerated Cases 2 Compressor Racks 3 Condensing Units 4 Walk-in Boxes 5 Some were done very well 6 Some had some issues... 7 8 Some seemingly
PAPER CHROMATOGRAPHY
PAPER CHROMATOGRAPHY INTRODUCTION Chromatography is a technique that is used to separate and to identify components of a mixture. This analytical technique has a wide range of applications in the real