Abstract. NHL Journal of Medical Sciences/Jan 2014/Vol 3/Issue 1 32
|
|
|
- Jeffry Bennett
- 9 years ago
- Views:
Transcription
1 Abstract Research article Impedance platelet count in severe microcytosis-study of 161 patients N J Ninama*, Nirali K Shah** *Assistant Professor, Dept of Physiology; **Resident, Dept of Pathology Smt NHL Municipal Medical College, Ahmedabad method was developed, enabling platelets to be easily Introduction- Platelet count estimation is an important element of the diagnostic and treatment process in many disorders. The present study aims to determine the incidence of improper impedance platelet count in severe microcytosis and to state the sensitivity and specificity of various parameters like histogram, red cell distribution width (RDW), mean platelet volume (MPV), platelet flags Methods-Total 161 patients with mean corpuscular volume less than 60 femtoliter and impedance platelet count more than 150 x 10 9 /L were included. The samples were analyzed on CellDyn 3700 cell counter. Manual platelet count and stained peripheral smear review was done in each case. ROC (Receiver Operating Characteristics) analysis was used for MPV, RDW and histogram. Results %(29/161) patients did not show corelation between impedance and manual platelet count. For Do not corelate group mean platelet volume(>12.7 femtoliter)showed 79.55% specificity and 55.17% sensitivity, Abnormal Platelet/RBC histogram(>40%) showed 76.51% specificity and % sensitivity and RDW > 27 % showed % sensitivity and % specificity. Conclusion-Impedance platelet count is not always reliable in cases of severe microcytosis. Abnormal MPV and RBC/platelet histogram gives good hint for the same. Key wards- Automated cell counter, Microcytosis, Platelet count Introduction Platelet count estimation is an important element of the diagnostic and treatment process in many disorders. 1 In patients with thrombocytopenia, especially in the case of platelet transfusion, the reliability of the platelet estimation is highly desired and necessary to provide appropriate treatment. 2,3 The assessment of platelet count is essential in clinical haematology. Any numerical deficiency or defect in their function may lead to bleeding. The normal platelet count at all ages is widely quoted as x 10 9 /L of whole blood. 4 At present manual counting using phase contrast microscopy, impedance analysis, optical light scatter or fluorescence analysis and immunoplatelet by flowcytometry are various avalaible methods for platelet count.early methods to enumerate platelets in blood were usually inaccurate and irreproducible until the mid-20th century. In 1953, the manual phase contrast microscopy discriminated from lysed red cells within a counting chamber or haemocytometer. 5 The manual method is time-consuming, subjective and tedious and results in high levels of imprecision with typical interobserver coefficient of variations (CV) in the range of 10 25%. 6 At low platelet numbers, because fewer cells are counted, observed CVs increase proportionally. Although relatively imprecise, the manual method still offers a relatively inexpensive, simple and viable means to enumerate platelets in the nonspecialized laboratory. Although the development of the Coulter Principle 7 revolutionized blood counting, platelet counts were only added to the automated full blood count in the late 1970s.Majority of cell counter uses impedance method to enumerate platelet and RBC. The particles analyzed are suspended in an electrolyte solution and dilution is passed through an aperture that links two chambers one containing a positive and the other a negative electrode. As cells pass through the orifice they cause a momentary increase in electrical resistance, which is registered as a pulse. One pulse represents a cell and the size of the pulse is proportional to size of cell. Platelet and RBC analyzed in same chamber(s) are discriminated according to their volume. 8 The introduction of automated blood cell counters using impedance technology resulted in a dramatic improvement in precision, with typical coefficient of variation of <3% because much higher total numbers of platelets are counted. 9 A major disadvantage of the electrical impedance method for counting platelets is the difficulty in distinguishing large platelets from extremely microcytic or fragmented red cells, even with the use of hydrodynamic focusing methods. False increase in the platelet count will occur when red cell or white cell fragments, microcytic red cells, immune complexes, bacteria or cell debris are included in the [3] reported platelet count. False decreases in the impedance platelet count will occur in the presence of large platelets and if there is platelet clumping as seen with pseudo-thrombocytopenia by ethylene diamine tetracetic acid (EDTA) dependent agglutinins. 8 In optical light scattering method, platelets are counted and sized by a flow cytometry system in which the cells in a suitable diluent pass through a narrow beam of light (i.e.,helium-neon laser). The illumination and light scatter by each cell is measured at a single angle (2 to 3 ).This allows assessment of the number of electrical pulses generated in proportion to the number of cells and cell volume. To improve discrimination of platelets accurately from nonplatelet particles, two-dimensional laser light scatter was developed which measure two NHL Journal of Medical Sciences/Jan 2014/Vol 3/Issue 1 32
2 angles of laser light scatter at 2 3 and at In optical fluorescent platelet count polymethine dye is used to stain the RNA/DNA of reticulated cells and platelet membrane and granules. The fluorescence intensity of each cell is analyzed, which allows the separation of platelets from red cells and reticulocytes. The fluorescent staining of the platelets not only allows the exclusion of nonplatelet particles from the count, but also allows the inclusion of large or giant platelets. 10 The immunologic platelet count involves labeling EDTA-anticoagulated blood with a suitable antiplatelet monoclonal antibody, which has been fluorescently conjugated with, for example, fluorescein isothiocyanate.the method simply derives the platelet count from the ratio of fluorescent platelets to red cells within the sample. Immunological platelet counting is simple, rapid, reliable and easily transferable to any laboratory with a flow cytometer. There is good correlation with the existing recommended manual method at normal platelet counts. Superior precision is demonstrated compared with the manual method especially in thrombopenic samples. 10 More recently, the combination/convergence of flowcytometric principles (e.g., laser light and hydrodynamic focusing) with impedance technology has facilitated the simultaneous detection of multiple light scatter and fluorescent parameters of platelets and cells within a range of modern automated hematology analysers. Modern analysers can therefore offer a range of platelet counting methods including impedance counting, optical counting (light scatter or fluorescence), and an immunological platelet counting method. These advances have improved the ability of automated analysers to discriminate platelets and should theoretically increase the accuracy of platelet counting in many clinical conditions. 11 In India, particularly Gujarat state has high prevalence of iron deficiency and beta thalassemia trait (BTT) which are associated with severe microcytosis. 12 As severe microcytosis can lead to false increase of impedance platelet count, it is very important to identify actual platelet count. The person with BTT and severe iron deficiency anemia can have normal impedance platelet count even though actual platelet count is low or borderline low. In high volume laboratories and laboratories with less trained manpower, it is important to identify and correct impedance platelet count in severe microcytosis from cell counter data or by other available methods. The correct platelet count should be informed to the clinician as in common conditions like malaria and dengue fever, where actual platelet count is reduced but impedance platelet count is normal or near normal and clinician try to follow the platelet count report for response to the therapy or recovery. 13,14. Aims of study were (1)To determine the incidence of improper impedance platelet count in severe microcytosis.(2) To state sensitivity and specificity of various parameters like abnormal platelet/rbc histogram, abnormal MPV, platelet flags and RDW as indicators of improper impedance platelet count. Material and methods All blood samples were drawn in 3ml K 3 EDTAcollection tube.total 191 samples were included in the study.30 samples from normal individuals were analysed as control group. The 161 patients with mean corposuclar volume (MCV) value less than 60 femtoliter and impedence platelet count more than 150 x 10 9 /L were included. The patients who received any blood transfusion (whole blood or components) or haematinics within last 3 months were excluded. The samples were analyzed in auto sampler mode on Cell Dyn 3700 (Abbott diagnostics, Santa Clara, CA) cell counter within 6 hours of phlebotomy. The Cell Dyn 3700 uses impedance principal for red blood cell (RBC) and platelet count. The software version 1.3 is used in that cell counter. The cell counter is maintained as per manufacturer s instructions. Cell counter was calibrated according to manufacturers specifications by field service representatives. The manual platelet count was done by ammonium oxalate method. For each patient manual platelet count was performed by two trained persons. The manual platelet count is considered as gold standard. Because of high intraindiviual variation and intermethodology variation cut off 30% is considered as significant. If the difference between impedance platelet and manual platelet count is more than 30%, it is stamped as Do not correlate group. If the difference is less than 30% it is stamped as Correlate group.peripheral blood smears stained with Leishman stain were reviewed in all study samples for the presence of erythrocyte or leukocyte fragments, marked microcytosis, giant platelets, platelet clumps,fibrin strands and number of platelets.the persons who performed manual platelet count and smear review were unaware of cell counter data. The MPV more than >12.7 femtoliter and MPV (>>>,>>>> and no value ) on cell counter were considered as abnormal MPV. In Cell Dyn 3700 URI (Upper Region Interferance) and LURI (Lower and Upper Region Interferance) are platelet flags which indicate presence of microcytic RBC, giant platelet, schistocytes,sickle cell and platelet clumps. The histogram of RBC and platelet is consider significant when platelet peak merge with RBC peak ( Valley ) at >40% where height platelet peak is considered as 100 %.The height of platelet peak and valley is measured in centimetres from printout of histogram. NHL Journal of Medical Sciences/Jan 2014/Vol 3/Issue 1 33
3 Figure-1 Abnormal RBC/platelet histogram (>40%), where left side peak is of platelet and right side peak is of RBC. The platelet peak measure 1.5 centimetre and subsequent valley measures 0.97 centimetre. X- axis size femtoliter, Y-axis numbers of events PLT- Platelets, RBC- Red blood cells Figure-2 Correlation of the results of Platelet counts using the impedance method and the manual method in blood samples from patients with microcytosis. Mannual platelet Impedance Vs Mannual platelet y = x R 2 = Impedance platelet The data analysis was done on Microsoft office 2003 Excel. ROC analysis and other statistics were done on MedCalc software version The beta thalassemia trait status was confirmed on D-10 VARIANT (Bio-Rad Laboratories, Hercules, CA). Results In control group total 30 normal individuals data were included. The mean MCV was 83.6 femtoliter (Standard deviation- 4.16), mean platelet count was x 10 9 /L (Standard deviation ), mean MPV was 7.63 femtoliter (Standard deviation -1.02). None of them show abnormal MPV, platelet flags and abnormal RBC platelet histogram (>40%). In study group, correlate group showed average MCV,RDW and MPV 56.18, and 9.05 respectively with standard deviation 3.17, 3.52 and 1.36 respectively. Do not correlate group showed average MCV,RDW and MPV 55.60, and 9.36 respectively with standard deviation 3.29,3.98 and 3.10 respectively. Table-1 Distribution of various parameters in study group. Correlate (%) Do not correlate (%) Specificity (%) Sensitivity (%) Total Participant (n=161) 132 /161 (81.98) 29/161 (18.01) Abnormal MPV(n=42) 27/132 (20.45) 15/29 (51.72) Platelet flags (URI,LURI) 11/132 (8.34) 27/29 (84.61) Histogram(RBC-platelet >40%) (n=46) 31/132 (23.48) 15/29 (51.72) RDW>27(N=29) 06/132 (4.54) 10/29 (34.48) BTT (n=12) 3/132 (2.72) 9/29 (31.03) MPV + Histogram (RBC-platelet >40%) (n=35) 21/132 (15.90) 14/29 (48.29) BTT+ Abnormal MPV (n=06) 2/132 (1.51) 4/29 (13.79) BTT+ Histogram (RBC-platelet >40%) (n=05) 1/132 (0.75) 4/29 (13.79) BTT+ Histogram (RBC-platelet >40%)+ Abnormal MPV (n=05) 1/132 (0.75) 4/29 (13.79) MPV-Mean platelet volume, BTT- Beta thalassemia trait, RBC-Red blood cells, URI-upper region Interferance, LURI- lower and upper region Interferance NHL Journal of Medical Sciences/Jan 2014/Vol 3/Issue 1 34
4 As depicted in table % (132/161) patients showed correlation, while 18.01% (29/161) patients did not show correlation with manual platelet count. Out of 29 patients, 2 patients showed false decreased and 27 patients showed false increased impedance platelet count. In Correlate group histogram (Platelet-RBC >40%) showed highest incidence of 23.48% (31/132). In Do not correlate group MPV and histogram (Platelet-RBC >40%) showed highest incidence of 51.72%(15/129).The cut-off of MPV,RDW and histogram(platelet-rbc) are derived by ROC analysis. The correlation of coefficient between impedance and manual platelet count in severe microcytosis is 0.692, which show poor correlation. Discussion As impedance platelet count is based only on the estimation of volume, it proves imperfect in the presence of nonplatelet elements similar in size to platelets. Severe microcytosis interfere with impedance platelet counts to various degrees. Excessive results in the x 10 9 /L area of the scope or slight thrombocytosis do not constitute clinical problems. False-negative results in the case of thrombocytopenia may jeopardize life. In present study total 18.01% (29/161) patients showed improper impedance platelet count, while 16.66% (5/30) showed improper impedance platelet count in study done by R.Pinkowski 1. In R.Pinkowski study MCV range from femtoliter, while in present study all 161 patients had MCV less than 60 femtoliter. Total 27 patients out 29 showed falsely increased impedance platelet count. This is due to severe microcytic RBC which comes in platelet region as the Cell dyn 3700 measure platelet between 1-35 femtoliter. 2 cases showed false decrease impedance platelet count. Both patients showed increased MPV and giant platelets on smear, this can be explained by the fact that giant platelet goes outside of upper limit of detection for impedance platelet (35 femtoliter in Cell Dyn 3700).The coefficient of correlation for impedance and manual platelet was as compared to the coefficient of correlation of 0.98 from R Pinkowski study. The difference might be because of different number of samples and MCV range. R.Pinkowski studied 30 patients data with iron deficiency and MCV ranges from femtoliter, whereas present study 161 patients with MCV less than 60 femtoliter. The cell counter used in present study was Cell-Dyn 3700, R.Pinkowski study was done on Cell- Dyn 4000, both cell counter work on same impedance principal for platelet and RBC. Software version 1.3 is used in present study, the detail of software version was not provided in R.Pinkowski study. In laboratory scenario of high workload, high incidence of severe microcytosis, non availability of high end cell counter, less trained manpower and less turnaround time, it is important to do correction with manual platelet count and peripheral smear. If we can derive a factor for correction by comparing impedance platelet count with immunoplatelet count it will be very useful but it will vary from cell counter to cell counter even with same manufacturer company. There is significant variation in the platelet count obtained on different analyser with the same sample. It might be because of difference in the method of analysis, linearity over the entire measuring range and number of events actually counted. 15 The Beckman Coulter and Sysmex impedance platelet counts showed better correlation with the reference method than the optical platelet counts by the Advia and the Sysmex. 16 The differences between impedance platelet and optical platelet count had an inverse correlation with MCV patients with microcytosis. [1] Therefore all the patients less than MCV 60 femtoliter are included in present study to increase the chances of finding improper impedance platelet count. The abnormal RBC/Platelet histogram more than 40 % was considered only giving more specificity to this indicator for correction of impedance platelet count. Lower cut off will increase the sensitivity but decrease specificity while increased cutoff will increase specificity and reduce sensitivity. Abnormal high MPV give hint about giant platelet and abnormal platelet/rbc histogram from cell counter gives idea regarding severe microcytic RBC entering the impedance platelet detection area or giant platelets are entering in the impedance RBC detection area. An abmormal MPV and platelet/rbc histogram both are appear it means microcytic RBC and/or giant platelets are cause for improper impedance platelet. An abnormal MPV without abnormal platelet/rbc histogram appears it suggests giant platelets are cause of concern for improper impedance platelet count. An abnormal platelet/rbc histogram without abnormal MPV suggests microcytic RBC to be a cause for improper impedance platelet count. MPV, histogram analysis and RDW can be put in practice for knowing the reliability of impedance platelet count in severe microcytosis (MCV less than 60 femtoliter) particularly in near normal impedance platelet count. The authorised signatory of that report should make a remark on report that impedance platelet count is corrected so clinical or laboratory physician can take a note of the same. Following conclusions derived from present study- 1: Impedance platelets count is not always reliable in cases of severe microcytosis. 2: Abnormal MPV, RDW and platelet/rbc histogram can provide very important hint for correction of impedance platelet count. 3: Peripheral smear review in severe microcytosis is very helpful. NHL Journal of Medical Sciences/Jan 2014/Vol 3/Issue 1 35
5 References 1. I Nkowski RP. Difference between impedance and optical platelet count methods in patients with microcytosis of red blood cells. Lab Hematol 1999; 5: Ault KA. Platelet counting: Is there room for improvement? Lab Hematol 1996; 2: Kckler TS, Rothe M, Blosser L, Schisano T, Vanhove L: Improving platelet transfusion therapy using the Immuno PLT method on the CellDyn Lab Hematol 1998; 4:80 4. Briggs C., Harrison P., Machin SJ. Continuing developments with the automated platelet count. Int. Jnl. Lab. Hem 2007; 29: Brecher G., Schneiderman M. & CronkiteE.P. The reproducibility of the platelet count. American Journal of Clinical Pathology 1953; 23: Harrison P., Horton A., Grant D., Briggs C. & Machin, S. J. Immunoplatelet counting: a proposed new reference procedure. British Journal of Haematology 2000;108: Coulter W.H. Means for Counting Particles Suspended in a Fluid. US Patent : Zandecki M., Genevieve F., Gerard J., Godon A. Spurious counts and spurious results on haematology analysers: a review. Part I: platelets. Int. Jnl. Lab. Hem 2007;29: Bentley SA. Johnson A & Bishop CA. A parallel evaluation of four automated haematology analysers. American Journal of Clinical Pathology 1993;100: Kunicka J.E., Fischer G., Murphy J. & Zelmanovic D. Improving platelet counting using two-dimensional laser light scatter. American Journal of Clinical Pathology 2000;114: Harrison P., Segal H., Briggs C., Murphy M. et al. Impact of Immunological Platelet Counting (by the Platelet/RBC Ratio) on Hematological Practice. Cytometry Part B (Clinical Cytometry) 2005;67: Rathod DA., Kaur A., Patel V., Patel K. et al. Usefulness of Cell Counter Based Parameters and Formulas in Detection of β-thalassemia Trait in Areas of High Prevalence. Am J Clin Pathol 2007;128: Jadhav U. M.; Patkar V. S. ; Kadam N. N. Thrombocytopenia in malaria: Correlation with type and severity of malaria. Journal of Association of Physicians of India 2004;52: G Ahhluwalia., SK Shrama. Dengue: Current Trends and Challenges An Indian Perspective Journal of association of physicians of India 2004;52: Briggs C., Machin SJ. Discrepancy between impedance and immunofluorescence platelet counting has implications for clinical decision making in patients with idiopathic thrombocytopenia purpura. British Journal of Haematology 2006; 135: Sandhaus LM., Ebenezer SO., Agrawal NN., Dillman CA. et al. Platelet Counting by the Coulter LH 750, Sysmex XE 2100, and Advia 120 A Comparative Analysis Using the RBC/Platelet Ratio Reference Method. Am J Clin Pathol 2002;118: NHL Journal of Medical Sciences/Jan 2014/Vol 3/Issue 1 36
Keywords: RBC histogram, Peripheral blood smear, red cell disorders.
Research article Automated red blood cell analysis compared with routine red blood cell morphology by smear review Dr.Poonam Radadiya*, Dr.Nandita Mehta*, Dr.Hansa Goswami***, Dr.R.N.Gonsai**** *3 rd year
Cell Counting Using the BD Accuri C6 Flow Cytometer
Cell Counting Using the BD Accuri C6 Flow Cytometer Koen Verbrugghe, PhD Marketing Applications Specialist BD Biosciences 23-14763- The BD Accuri C6 Flow Cytometer System An affordable, full-featured,
XE-2100 Automated Hematology System. Fast, Accurate, Dependable
XE-2100 Automated Hematology System Fast, Accurate, Dependable Advanced Technology Solutions to Meet your Lab s Needs Even with these challenges, the need for hematology testing has remained steady or
Accuracy of platelet counting haematology analysers in severe thrombocytopenia and potential impact on platelet transfusion
research paper Accuracy of platelet counting haematology analysers in severe thrombocytopenia and potential impact on platelet transfusion H. C. Segal, 1 C. Briggs, 2 S. Kunka, 2 A. Casbard, 3 P. Harrison,
MANUAL PLATELET COUNT
MANUAL PLATELET COUNT LAB OBJECTIVE The student will be able to perform, within K20% accuracy as compared to the automated result, five manual platelet counts using the Unopette system. and the light microscope.
RDW-- Interpreting the Full Blood Count
RDW-- Interpreting the Full Blood Count The most important components of a Full Blood Count report are, of course, the Haemoglobin, the White Cell Count and Differential and the Platelet Count. However,
Collect and label sample according to standard protocols. Gently invert tube 8-10 times immediately after draw. DO NOT SHAKE. Do not centrifuge.
Complete Blood Count CPT Code: CBC with Differential: 85025 CBC without Differential: 85027 Order Code: CBC with Differential: C915 Includes: White blood cell, Red blood cell, Hematocrit, Hemoglobin, MCV,
MANUAL WBC LAB OBJECTIVE
MANUAL WBC LAB OBJECTIVE The student will be able to perform, within K20% accuracy as compared to the automated result, five manual white blood cell counts using the Unopette system. PRINCIPLE Free-flowing
Laboratory Studies in the Diagnosis of Iron Deficiency, Latent Iron Deficiency and Iron Deficient Erythropoiesis
Laboratory Studies in the Diagnosis of Iron Deficiency, Latent Iron Deficiency and Iron Deficient Erythropoiesis General Comments The laboratory studies listed below are helpful in the diagnosis and management
MANUAL PLATELET COUNT
MANUAL PLATELET COUNT Principle Whole blood is diluted with a 1% ammonium oxalate solution. The isotonic balance of the diluent is such that all erythrocytes are lysed while the leukocytes, platelets,
WHOLE BLOOD LYSING SOLUTION FOR FLOW CYTOMETRIC APPLICATIONS
FOR IN-VITRO DIAGNOSTIC USE INVITROGEN CAL-LYSE TM Lysing Solution WHOLE BLOOD LYSING SOLUTION FOR FLOW CYTOMETRIC APPLICATIONS CAL-LYSE TM CATALOG No. GAS-010 250 tests 25 ml CATALOG No. GAS-010S-100
STAINING OF PBF AND INTERPRETATION OF NORMAL AND ABNORMAL RED CELL MORPHOLOGY
9 STAINING OF PBF AND INTERPRETATION OF NORMAL AND ABNORMAL RED CELL MORPHOLOGY 9.1 INTRODUCTION A peripheral blood smear (peripheral blood film) is a glass microscope slide coated on one side with a thin
BC-5800. Auto Hematology Analyzer
BC-5800 Auto Hematology Analyzer BC-5800 Auto Hematology Analyzer 5-part differentiation, 29 parameters, 2 histograms + 2 scattergrams Up to 90 samples per hour Laser scatter + Chemical dye + Flow cytometry
Technical Bulletin. Threshold and Analysis of Small Particles on the BD Accuri C6 Flow Cytometer
Threshold and Analysis of Small Particles on the BD Accuri C6 Flow Cytometer Contents 2 Thresholds 2 Setting the Threshold When analyzing small particles, defined as particles smaller than 3.0 µm, on the
Hematology Morphology Critique
Survey Slide: History: 60-year-old female presenting with pneumonia Further Laboratory Data: Hgb : 90 g/l RBC : 2.92 10 12 /L Hct : 0.25 L/L MCV : 87 fl MCH : 30.8 pg MCHC : 355 g/l RDW : 17.7 % WBC :
Katharina Lückerath (AG Dr. Martin Zörnig) adapted from Dr. Jörg Hildmann BD Biosciences,Customer Service
Introduction into Flow Cytometry Katharina Lückerath (AG Dr. Martin Zörnig) adapted from Dr. Jörg Hildmann BD Biosciences,Customer Service How does a FACS look like? FACSCalibur FACScan What is Flow Cytometry?
Stepcount. Product Description: Closed transparent tubes with a metal screen, including a white matrix at the bottom. Cat. Reference: STP-25T
Product Description: Closed transparent tubes with a metal screen, including a white matrix at the bottom Cat. Reference: STP-25T Reagent provided:: 25 Stepcount tubes for 25 test INTENDED USE. Immunostep
Hematology Reagents and Controls Our products for your critical process
Part of a pure process Hematology Reagents and Controls Our products for your critical process 2 Reagents you can count on Mallinckrodt Baker s chemical expertise and 30 years of experience are the foundation
CELL CYCLE BASICS. G0/1 = 1X S Phase G2/M = 2X DYE FLUORESCENCE
CELL CYCLE BASICS Analysis of a population of cells replication state can be achieved by fluorescence labeling of the nuclei of cells in suspension and then analyzing the fluorescence properties of each
Introduction to flow cytometry
Introduction to flow cytometry Flow cytometry is a popular laser-based technology. Discover more with our introduction to flow cytometry. Flow cytometry is now a widely used method for analyzing the expression
Application Information Bulletin: Set-Up of the CytoFLEX Set-Up of the CytoFLEX* for Extracellular Vesicle Measurement
Application Information Bulletin: Set-Up of the CytoFLEX Set-Up of the CytoFLEX* for Extracellular Vesicle Measurement Andreas Spittler, MD, Associate Professor for Pathophysiology, Medical University
Introduction to Flow Cytometry
Introduction to Flow Cytometry presented by: Flow Cytometry y Core Facility Biomedical Instrumentation Center Uniformed Services University Topics Covered in this Lecture What is flow cytometry? Flow cytometer
Reagent kit for the flow cytometric quantitation of platelet associated immunoglobulin. Antibodies and reagents for 25 tests
page 1 of 8 Reagent kit for the flow cytometric quantitation of platelet associated immunoglobulin Antibodies and reagents for 25 tests Key to symbols used Glycotope Biotechnology GmbH Czernyring 22 69115
ABO-Rh Blood Typing Using Neo/BLOOD
ABO-Rh Blood Typing Using Neo/BLOOD Objectives Determine the ABO and Rh blood type of unknown simulated blood samples. Prepare a simulated blood smear. Examine a prepared blood smear under the microscope
Preparation of Blood Films
Preparation of Blood Films Principle: Blood film enables us to evaluate WBC, RBC, and PLT morphology, also, allows us to perform differential WBC count, furthermore estimation of WBC and platelets counts
APPLICATION INFORMATION
DRAFT: Rev. D A-2045A APPLICATION INFORMATION Flow Cytometry 3-COLOR COMPENSATION Raquel Cabana,* Mark Cheetham, Jay Enten, Yong Song, Michael Thomas,* and Brendan S. Yee Beckman Coulter, Inc., Miami FL
No-wash, no-lyse detection of leukocytes in human whole blood on the Attune NxT Flow Cytometer
APPLICATION NOTE Attune NxT Flow Cytometer No-wash, no-lyse detection of leukocytes in human whole blood on the Attune NxT Flow Cytometer Introduction Standard methods for isolating and detecting leukocytes
ST. VINCENT'S. MEDICAL CENTER St. Vincent's Healthcare
ST. VINCENT'S MEDICAL CENTER St. Vincent's Healthcare Medical Technology St. Vincent s Schools of Medical Science Throughout Northeast Florida and Southern Georgia, St. Vincent s HealthCare is well known
Standardization, Calibration and Quality Control
Standardization, Calibration and Quality Control Ian Storie Flow cytometry has become an essential tool in the research and clinical diagnostic laboratory. The range of available flow-based diagnostic
use. 3,5 Apoptosis Viability Titration Assays
Introduction Monitoring cell viability can provide critical insights into the effectiveness of biological protocols and the stability of biological systems including: the efficacy of cell harvesting procedures,
Automated Cerebrospinal Fluid Cytology
Automated Cerebrospinal Fluid Cytology Limitations and Reasonable Applications Herwig Strik, M.D., Hilmar Luthe, Ph.D., Ingelore Nagel, Bettina Ehrlich, and Mathias Bahr, M.D. OBJECTIVE: To evaluate the
Hematology Case Studies: Platelets
California Association for Medical Laboratory Technology Distance Learning Program Hematology Case Studies: Platelets by Helen M. Sowers, MA, CLS Dept of Biological Science (ret.) California State University,
CELL CYCLE BASICS. G0/1 = 1X S Phase G2/M = 2X DYE FLUORESCENCE
CELL CYCLE BASICS Analysis of a population of cells replication state can be achieved by fluorescence labeling of the nuclei of cells in suspension and then analyzing the fluorescence properties of each
EFFECT OF DAILY VERSUS WEEKLY IRON FOLIC ACID SUPPLEMENTATION ON THE HAEMOGLOBIN LEVELS OF CHILDREN 6 TO 36 MONTHS OF URBAN SLUMS OF VADODARA
ORIGINAL ARTICLE. EFFECT OF DAILY VERSUS WEEKLY IRON FOLIC ACID SUPPLEMENTATION ON THE HAEMOGLOBIN LEVELS OF CHILDREN 6 TO 36 MONTHS OF URBAN SLUMS OF VADODARA K Sharma 1, P Parikh 2, F Desai 3 1 PhD Advisor,
Automated Quadratic Characterization of Flow Cytometer Instrument Sensitivity (flowqb Package: Introductory Processing Using Data NIH))
Automated Quadratic Characterization of Flow Cytometer Instrument Sensitivity (flowqb Package: Introductory Processing Using Data NIH)) October 14, 2013 1 Licensing Under the Artistic License, you are
Pentra DX Nexus. Process efficiency in Hematology
Pentra DX Nexus Process efficiency in Hematology Pentra DX Nexus Be productive and flexible 120 samples per hour 28 parameters Automatic reflex testing Automatic validation of results Integrated cytology
Enumeration and Viability of Nucleated Cells from Bone Marrow, Cord Blood, and Mobilized Peripheral Blood
Application Note Primary Cell Analysis Enumeration and Viability of Nucleated Cells from Bone Marrow, Cord Blood, and Mobilized Peripheral Blood Cell Viability Introduction: Bone Marrow, Cord Blood, Stem
HEMATOCRIT/HCT AND CALCULATED HEMOGLOBIN/HB
HEMATOCRIT/HCT AND CALCULATED HEMOGLOBIN/HB Hematocrit is determined conductometrically. The measured conductivity, after correction for electrolyte concentration, is inversely related to the hematocrit.
Department of Pathology and Laboratory Medicine Capital District Health Authority Nova Scotia
Department of Pathology and Laboratory Medicine Capital District Health Authority Nova Scotia Doc #: 10902 Section: \\Management System\PLM\Point of Care Testing\CBC Diff\ Document Owner: ES Team Lead
Below is a list of things you should be aware of before you schedule your sort.
Sorting at the Flow Cytometry Facility At the present time, the assistance of a trained cell sorter operator is needed for all sorting applications. As such, if you are planning a first time sort you will
Selected Topics in Electrical Engineering: Flow Cytometry Data Analysis
Selected Topics in Electrical Engineering: Flow Cytometry Data Analysis Bilge Karaçalı, PhD Department of Electrical and Electronics Engineering Izmir Institute of Technology Outline Compensation and gating
QUALITY CONTROL (QC) INFORMATION AND TROUBLESHOOTING GUIDE HEMATOLOGY
QUALITY CONTROL (QC) INFORMATION AND TROUBLESHOOTING GUIDE HEMATOLOGY This document is not intended to replace the information in your Instructions for Use Manuals (IFU). Information in the Instructions
EDUCATIONAL COMMENTARY RED BLOOD CELLS AND WHITE BLOOD CELLS IN URINALYSIS
URINALYSIS Educational commentary is provided through our affiliation with the American Society for Clinical Pathology (ASCP). To obtain FREE CME/CMLE credits click on Earn CE Credits under Continuing
HAEMATOLOGY LABORATORY
HAEMATOLOGY LABORATORY Head of Unit : Dr Raudhawati Osman Phone : +603-26155281 Email : [email protected] 1. INTRODUCTION The Haematology Unit, Department of Pathology, HKL provides tertiary diagnostic
Cold Agglutination Titer detecting Cold Reacting Antibodies
Objectives: Cold Agglutination Titer detecting Cold Reacting Antibodies 1. Perform a serial dilution to determine the amount of cold reacting antibody present in a patient specimen with the results obtained
123count ebeads Catalog Number: 01-1234 Also known as: Absolute cell count beads GPR: General Purpose Reagents. For Laboratory Use.
Page 1 of 1 Catalog Number: 01-1234 Also known as: Absolute cell count beads GPR: General Purpose Reagents. For Laboratory Use. Normal human peripheral blood was stained with Anti- Human CD45 PE (cat.
Study of serum protein electrophoresis in suspected cases of Multiple Myeloma.
Journal homepage: http://www.journalijar.com INTERNATIONAL JOURNAL OF ADVANCED RESEARCH RESEARCH ARTICLE Study of serum protein electrophoresis in suspected cases of Multiple Myeloma. Dr. Dharmishtha N.
Evaluation of the Patient with Suspected Platelet Refractory State
Evaluation of the Patient with Suspected Platelet Refractory State NOTE: While evaluating the patient for suspected immune refractory state provide ABO matched platelets if available. 1. Determine if the
Immunophenotyping peripheral blood cells
IMMUNOPHENOTYPING Attune Accoustic Focusing Cytometer Immunophenotyping peripheral blood cells A no-lyse, no-wash, no cell loss method for immunophenotyping nucleated peripheral blood cells using the Attune
510(k) SUBSTANTIAL EQUIVALENCE DETERMINATION DECISION SUMMARY
510(k) SUBSTANTIAL EQUIVALENCE DETERMINATION DECISION SUMMARY A. 510(k) Number: k103555 B. Purpose for Submission: New submission C. Measurand: Platelet aggregation D. Type of Test: Platelet aggregometer
Measures of diagnostic accuracy: basic definitions
Measures of diagnostic accuracy: basic definitions Ana-Maria Šimundić Department of Molecular Diagnostics University Department of Chemistry, Sestre milosrdnice University Hospital, Zagreb, Croatia E-mail
Medical Laboratory Technology Program. Student Learning Outcomes & Course Descriptions with Learning Objectives
Medical Laboratory Technology Program Student Learning Outcomes & Course Descriptions with Learning Objectives Medical Laboratory Technology Student Learning Outcomes All Colorado Mesa University associate
Flow Cytometry A Basic Overview
Flow Cytometry A Basic Overview Overview Flow cytometry is a powerful technology for investigating many aspects of cell biology and for isolating cells of interest. Flow cytometry utilizes highly focused,
1) Write the following as an algebraic expression using x as the variable: Triple a number subtracted from the number
1) Write the following as an algebraic expression using x as the variable: Triple a number subtracted from the number A. 3(x - x) B. x 3 x C. 3x - x D. x - 3x 2) Write the following as an algebraic expression
Flow cytometry basics fluidics, optics, electronics...
Title Flow cytometry basics fluidics, optics, electronics... RNDr. Jan Svoboda, Ph.D. Cytometry and Microscopy Core Facility IMB, CAS, v.v.i Vídeňská 1083 Fluorescence Fluorescence occurs when a valence
Supplementary Material. Free-radical production after post-thaw incubation of ram spermatozoa is related to decreased in vivo fertility
10.1071/RD14043_AC CSIRO 2015 Supplementary Material: Reproduction, Fertility and Development, 2015, 27(8), 1187 1196. Supplementary Material Free-radical production after post-thaw incubation of ram spermatozoa
Platelet Transmission Electron Microscopy and Flow Cytometry 11/15/2015
Welcome to Mayo Medical Laboratories Hot Topics. These presentations provide short discussion of current topics and may be helpful to you in your practice. Today our topic looks at hereditary platelet
Introduction to Flow Cytometry
Outline Introduction to Flow Cytometry Basic Concept of Flow Cytometry Introduction to Instrument Subsystems Daisy Kuo Assistant Product Manager E-mail: [email protected] BDBiosciences Application Examples
Notes on titering antibodies
Home Events This Week Bulletin Boards Lab Personnel Lab Policies Protocols MolBio Tetramers Proteins Flow Cytometry Immunoassays Instruments Computers Lab Publications Scientiific Literature Science on
Whole Blood Flow Cytometry
Whole Blood Flow Cytometry y Nailin Li Department t of Medicine, i Clinical i l Pharmacology Unit Karolinska Institute/University Hospital, 171 76 Stockholm Department of Pathology & Pathophysiology Zhejiang
Spherotech, Inc. 27845 Irma Lee Circle, Unit 101, Lake Forest, Illinois 60045 1
78 Irma Lee Circle, Unit 0, Lake Forest, Illinois 00 SPHERO TM Technical Note STN- Rev B 008 DETERMINING PMT LINEARITY IN FLOW CYTOMETERS USING THE SPHERO TM PMT QUALITY CONTROL EXCEL TEMPLATE Introduction
1.5 Function of analyte For albumin, see separate entry. The immunoglobulins are components of the humoral arm of the immune system.
Total protein (serum, plasma) 1 Name and description of analyte 1.1 Name of analyte Total protein 1.2 Alternative names None 1.3 NMLC code 1.4 Description of analyte This is a quantitative measurement
Calcium. Table 1: Difference between method means in percent
Calcium Measurement of total calcium is widely used for both the diagnosis and the monitoring of a range of conditions related to the bones, heart, nerves, and kidneys. Total calcium measurements include
CyFlow SL. Healthcare Immunology. Portable FCM System for 3-colour Immunophenotyping
CyFlow SL Healthcare Immunology Portable FCM System for 3-colour Immunophenotyping 01 COMPANY Flow Cytometry made in Germany New sophisticated applications and increasing requirements for reliable results
METHODS OF VITAMIN ANALYSIS
METHODS OF VITAMIN ANALYSIS Specimen requirements; Fasting plasma or serum Lithium Heparin is the anticoagulant of choice for vitamins such as; thiamine, riboflavin, retinol, tocopherols & cholecalciferol
SEED Haematology. Sysmex Educational Enhancement and Development February 2013
SEED Haematology Sysmex Educational Enhancement and Development February 2013 The role of the peripheral blood smear in the modern haematology laboratory Automated haematology cell counting The laboratory
Method Validation/Verification. CAP/CLIA regulated methods at Texas Department of State Health Services Laboratory
Method Validation/Verification CAP/CLIA regulated methods at Texas Department of State Health Services Laboratory References Westgard J. O.: Basic Method Validation, Westgard Quality Corporation Sarewitz
Fundamentals of modern UV-visible spectroscopy. Presentation Materials
Fundamentals of modern UV-visible spectroscopy Presentation Materials The Electromagnetic Spectrum E = hν ν = c / λ 1 Electronic Transitions in Formaldehyde 2 Electronic Transitions and Spectra of Atoms
Urinalysis and Body Fluids CRg
Urinalysis and Body Fluids CRg Unit 2; Session 1 Urine Microscopic Examination The Complete Urinalysis Physical properties already covered Chemical analysis in the next unit Microscopic our current focus
NHS FORTH VALLEY B12 and Folate: A Practical Guide
NHS FORTH VALLEY B12 and Folate: A Practical Guide Date of First Issue 27/05/2011 Approved 28/06/2011 Current Issue Date 24/06/2013 Review Date 24/06/2015 Version 1.0 EQIA 27/05/2011 Author / Contact Group
Thibodeau: Anatomy and Physiology, 5/e. Chapter 17: Blood
Thibodeau: Anatomy and Physiology, 5/e Chapter 17: Blood This chapter begins a new unit. In this unit, the first four chapters deal with transportation one of the body's vital functions. It is important
EXERCISE 5: ERYTHROCYTES SEDIMENTATION RATE - ESR, SED RATE
EXERCISE 5: ERYTHROCYTES SEDIMENTATION RATE - ESR, SED RATE Textbook: Skills: None 10 points Objectives: 1. State the principle of the Erythrocytes Sedimentation Rate - ESR. 2. List two factors which may
Reticulocyte hemoglobin equivalent (Ret He) and assessment of iron-deficient states
Clin. Lab. Haem. 06, 28, 3 8 C. BRUGNARA*, B. SCHILLER, J. MORAN doi:.1111/j.1365-2257.06.00812.x Reticulocyte hemoglobin equivalent (Ret He) and assessment of iron-deficient states *Department of Laboratory
Variables Control Charts
MINITAB ASSISTANT WHITE PAPER This paper explains the research conducted by Minitab statisticians to develop the methods and data checks used in the Assistant in Minitab 17 Statistical Software. Variables
BLOOD BANK SPECIMEN COLLECTION PROCEDURE
BLOOD BANK SPECIMEN COLLECTION PROCEDURE INTRODUCTION Scientific and technical advances in blood group serology have made the transfusion of blood a relatively safe procedure, but serious adverse effects
LAB: Blood Exploration
Name: Period: Date: LAB: Blood Exploration Introduction A liquid called plasma makes up about half (55%) of the content of blood. Plasma contains proteins that help blood to clot, transport substances
D-Facts. Vitamin D Competitive Evaluation
D-Facts Vitamin D Competitive Evaluation Abstract 101 patient samples collected across sites in Germany were used to compare the performance of the Siemens ADVIA Centaur Vitamin D Total (VitD), Abbott
R&D Systems Hematology Products
R&D Systems Hematology Products Controls Calibrators Linearity Materials Quality Control Program TABLE OF CONTENTS R&D SYSTEMS HEMATOLOGY PRODUCTS R&D Systems, Inc., a specialty manufacturer of biological
Uses of Flow Cytometry
Uses of Flow Cytometry 1. Multicolour analysis... 2 2. Cell Cycle and Proliferation... 3 a. Analysis of Cellular DNA Content... 4 b. Cell Proliferation Assays... 5 3. Immunology... 6 4. Apoptosis... 7
The Blood Budget: how can we reduce costs and influence best practice? Sue Redfearn Blood Transfusion Manager Poole Hospital NHS Foundation Trust
The Blood Budget: how can we reduce costs and influence best practice? Sue Redfearn Blood Transfusion Manager Poole Hospital NHS Foundation Trust Poole Hospital vital statistics 680 beds 2011 (reduced
Chapter 6: Antigen-Antibody Interactions
Chapter 6: Antigen-Antibody Interactions I. Strength of Ag-Ab interactions A. Antibody Affinity - strength of total noncovalent interactions between single Ag-binding site on an Ab and a single epitope
Absolute counting using flow cytometry. Bright Count Microspheres
Absolute counting using flow cytometry Single-platform system for absolute counts I PRODUCT MANUAL [REF] IQP-391 [REF] IQP-392 s 50 tests s 100 tests [IVD] In Vitro Diagnostic medical device IQ Products
Flow cytometric analysis of stem cells derived from umbilical cord blood CD34+/CD45+ enumeration and viability
Flow cytometric analysis of stem cells derived from umbilical cord blood CD34+/CD45+ enumeration and viability Doretha Thompson Dissertation in fulfillment of the requirements for the degree Magister Scientiae
JOINT COMMISSION INTERNATIONAL ACCREDITATION STANDARDS FOR. 2nd Edition
JOINT COMMISSION INTERNATIONAL ACCREDITATION STANDARDS FOR CliniCAl laboratories 2nd Edition Effective 1 April 2010 International Patient Safety Goals (IPSG) Goals The following is a list of all goals.
Flow Data Analysis. Qianjun Zhang Application Scientist, Tree Star Inc. Oregon, USA FLOWJO CYTOMETRY DATA ANALYSIS SOFTWARE
Flow Data Analysis Qianjun Zhang Application Scientist, Tree Star Inc. Oregon, USA Flow Data Analysis From Cells to Data - data measurement and standard Data display and gating - Data Display options -
Young fetus: site of hematopoiesis together with the liver and bone marrow. Hgb WBC >----------< Plt Hct. Retic =
Learning Objectives At the completion of this program, the participants will be able to: 1. Identify the components of the CBC and Differential and their clinical implications. 2. Identify normal pediatric
Doppler. Doppler. Doppler shift. Doppler Frequency. Doppler shift. Doppler shift. Chapter 19
Doppler Doppler Chapter 19 A moving train with a trumpet player holding the same tone for a very long time travels from your left to your right. The tone changes relative the motion of you (receiver) and
Biochemistry Validation Form
Biochemistry Validation Form Method: Piccolo Xpress Chemistry 13 panel Manufacturer: Abaxis Cat no: 400-0029 CE marked: Yes: x No: Location of bench book: X:\Bio\Patricia\Piccollo Xpress evaluation (August
Blood Transfusion. There are three types of blood cells: Red blood cells. White blood cells. Platelets.
Blood Transfusion Introduction Blood transfusions can save lives. Every second, someone in the world needs a blood transfusion. Blood transfusions can replace the blood lost from a serious injury or surgery.
Exercise 9: Blood. Readings: Silverthorn 5 th ed, 547 558, 804 805; 6 th ed, 545 557, 825 826.
Exercise 9: Blood Readings: Silverthorn 5 th ed, 547 558, 804 805; 6 th ed, 545 557, 825 826. Blood Typing The membranes of human red blood cells (RBCs) contain a variety of cell surface proteins called
Spectrophotometry and the Beer-Lambert Law: An Important Analytical Technique in Chemistry
Spectrophotometry and the Beer-Lambert Law: An Important Analytical Technique in Chemistry Jon H. Hardesty, PhD and Bassam Attili, PhD Collin College Department of Chemistry Introduction: In the last lab
Preparation "Clinical Laboratory Technologist and Technician Overview"
Clinical Laboratory Technologist and Technician Overview The Field - Preparation - Day in the Life - Earnings - Employment - Career Path Forecast - Professional Organizations The Field Clinical laboratory
PROTOCOL. Immunostaining for Flow Cytometry. Background. Materials and equipment required.
PROTOCOL Immunostaining for Flow Cytometry 1850 Millrace Drive, Suite 3A Eugene, Oregon 97403 Rev.0 Background The combination of single cell analysis using flow cytometry and the specificity of antibody-based
Diagnosis, Treatment and Prevention of Typhoid Fever in the Children of Bangladesh: A Microbiologist s View.
Diagnosis, Treatment and Prevention of Typhoid Fever in the Children of Bangladesh: A Microbiologist s View. Samir K Saha, Ph.D. Department of Microbiology Dhaka Shishu Hospital Bangladesh Typhoid Fever:Salmonella
Cat S O SSM. Services. Reporting Particle Count by ISO Code. Management Guide. Two methods of reading particles
Cat S O SSM Services Management Guide Two methods of reading particles Measuring contamination ISO Codes ISO 11171:1999 Reporting Particle Count by ISO Code Introduction... 3 What ISO Means... 3 Two Methods
Yvette Marie Miller, M.D. Executive Medical Officer American Red Cross October 20, 2012 45 th Annual Great Lakes Cancer Nursing Conference Troy, MI
Yvette Marie Miller, M.D. Executive Medical Officer American Red Cross October 20, 2012 45 th Annual Great Lakes Cancer Nursing Conference Troy, MI Overview of Hematology, http://www.nu.edu.sa/userfiles/mhmorsy/h
UNIVERSITY OF PÉCS MEDICAL SCHOOL FLOW CYTOMETRY AND CELL SEPARATION BIOPHYSICS 2. 2015 4th March Dr. Beáta Bugyi Department of Biophysics Flow cytometry and cell separation FLOW = STREAM OF FLUID in a
Validation and Calibration. Definitions and Terminology
Validation and Calibration Definitions and Terminology ACCEPTANCE CRITERIA: The specifications and acceptance/rejection criteria, such as acceptable quality level and unacceptable quality level, with an