VALIDATION OF WET AND DRY LASER DIFFRACTION PARTICLE CHARACTERISATION METHODS.

Size: px
Start display at page:

Download "VALIDATION OF WET AND DRY LASER DIFFRACTION PARTICLE CHARACTERISATION METHODS."

Transcription

1 VALIDATION OF WET AND DRY LASER DIFFRACTION PARTICLE CHARACTERISATION METHODS. Ward-Smith, R.S., Gummery, N. and Rawle, A.F. Malvern Instruments Ltd, Enigma Business Park, Grovewood Road, Malvern, Worcs, U.K., WR14 1XZ. Tel: Fax: INTRODUCTION With the introduction of both ISO (Particle size analysis Laser Diffraction methods) [1] and NIST (Practice Guide Particle size Characterisation) [2] the user of a laser diffraction instrument has been supplied with a wealth of useful information on theory and guidance on both dispersion and sampling. The amount of information available on how to validate the method of analysis once it has been developed is a lot more limited. The Pharmaceutical Analytical Sciences Group (PASG) [3] have attempted to lay down some guidelines, but these did not provide sufficient detail regarding the actual experiments that need to be performed. This paper sets out to define the important areas of method development and provide both guidelines and example data to suggest what experiments might be done in order to validate a method. However just because these experiments have been performed, the authors cannot guarantee that an auditing body will always approve and rubberstamp the method. Rather the author s are suggesting a minimal amount of work that we d expect to be performed. Any more work will only strengthen the case of the candidate method. Lerke and Adams [4] have recently published their ideas on the subject in a useful paper. Their paper covers both method development and method validation, whereas this paper is mainly concerned with the latter. The purpose of creating a validated method is to ascertain the robustness and integrity of a particle size method by testing all possible parameters that could cause variation in the reported size. Validation is defined by the FDA as Establishing documentary evidence which provides a high degree of assurance that a specific process will consistently produce a product meeting its predetermined specification and quality attributes. METHOD VALIDATION It goes without saying that development of such methods should be performed on an instrument, which has validated software and is regularly tested to confirm its performance (both in the form of an IQ/OQ (at least yearly) and a secondary standard). Validated software is software, which has both lifecycle documentation and the ability to have its numeracy tested with the use of a peer package such as Excel. It is essential in the current climate that the software is also CFR21 Rule 11 compliant. This paper will now consider each of the main variables in turn.

2 SAMPLING The first step any analyst will take is the extraction of a representative sample from the bulk material. If the sample is not representative the result can be taken to be atypical and the experiment is a pointless exercise. The question that every analyst should ask themselves before they sub-sample is Is the small bag of powder I have representative of the production process? Is the use of devices such as sample thieves (to extract sample from the bulk) a regular part of the production process. It can be seen from table 1 that the use of a spinning riffler produces sub-samples of powder, which vary less than other more commonly used sampling techniques. Riffling works best for free flowing particles, but can take a great deal of time if a large amount of powder is to be dealt with. If the sample is a suspension then this brings its own set of problems. Whereas with a powder, settling in transit leads to the large particles at the top of the heap and the fine at the bottom, a suspension will show the reverse. Large particles must be adequately re-suspended without bias. Use of certain stirrers (especially magnetic fleas) can lead to the large particles being thrown to the outside of the container where they are not sampled. Method Relative Standard Deviation (%) Cone & Quartering 6.81 Scoop Sampling 5.14 Table Sampling 2.09 Chute Riffling 1.01 Spin Riffling Table 1: Sources of sampling error (Reproduced from Allen [5] ). SAMPLE PREPARATION Bell et al [3] defined sample preparation as the pre-treatment and the presentation of the sample to the measuring technique in a meaningful manner. This will depend on the interests of the user, sometimes it is the primary particle size that is important, and the sample should be dispersed to this, sometimes it is the natural agglomerated state that is of interest. In either case, the dispersion medium, air (for dry dispersion) or a liquid, should not cause irreversible change to the particle (dissolution, milling or aggregation). The whole sample preparation procedure should be carried out as part of method development rather than method validation. RANGE The range of the instrument should ideally cover the size range of the sample. For modern laser diffraction instrumentation with size ranges between 20nm and 2000µm this normally presents no problem for most pharmaceutical samples. Older instrumentation often has many lenses to cover the same range. In this case the lens which covers the largest proportion of the particle size distribution should be used or the result from two lenses could be blended together (though this is not recommended as the result depends on the mathematical efficacy of the blending routines).

3 SPECIFICITY Is the technique appropriate? There is no strict requirement to perform this exercise in particle size analysis as it is assumed that the operator is using the most suitable particle size technique for the material being analysed. The appropriate technique should have already been established in the method development. Of course different methods of particle size analysis will lead to different answers for the same sample. Also, due to the nature of the laser diffraction technique, it is difficult to develop a particle size analysis method that will distinguish the scattering from different components within a product dosage form. In the case of a suspension, it may be necessary to prepare a placebo mix and analyse this separately. ROBUSTNESS The robustness of an analytical method is an indication of its ability to remain unaffected by small variations in the test parameters and so provide assurance of its reliability during routine usage. The method robustness should be considered before repeatability, reproducibility and intermediate precision exercises are undertaken. The variations that are normally considered are that of measurement duration and measurement stability. Variables such as air pressure (dry measurement) and pump / stir rates (wet measurement), normally considered as part of method development, are also important and will be briefly touched on here. a) Measurement Duration The purpose of this exercise is to determine a suitable measurement duration, which will then be used for the repeatability, reproducibility and intermediate precision exercises. A sample should be prepared in accordance with the method under investigation. A cycle of ten measurements should be performed for durations of 2, 5, 7, 10 and 15 seconds. The individual and mean readings for each duration should be recorded. The results can be over-plotted and any shift in particle size distribution can be noted. The appropriate duration period can be selected by looking at the relative standard deviation of the median particle size. For particles with a median size (D(v,0.5)) of >10 micron the Relative Standard Deviation (RSD) should be <3% and for particles with a median size <10 micron the RSD must be <6%. This reflects the fact that smaller particles are more difficult to disperse. Table 2 shows how the D(v,0.5) measured for a lactose sample varies with measurement duration.

4 Measurement Measurement Duration Number 2 sec 5 sec 7 sec 10 sec 15 sec Mean / Microns %RSD Table 2: Variation of lactose D(v,0.5) with measurement duration. The RSD can be seen to increase over time, so an interim point where the RSD was low (5 seconds) was chosen for measurement. From looking at the data overall it can be deduced that 2 seconds is inadequate as it appears that it is not giving the larger particles sufficient time to be sampled, so the D(v,0.5) is significantly lower. b) Repeatability / Sample Measurement Stability In order to determine whether samples are stable over the period of analysis and not subject to agglomeration, de-agglomeration or dissolution, it is necessary to monitor the particle size distribution at known time points. A sample should be prepared in accordance with the method under investigation. It is recommended that at least five measurements at the previously established duration are taken. Measurements could be taken after 1, 3, 5, 7 and 10 minutes. At least five repeat measurements should be obtained at each time point. The mean and relative standard deviation for the D(v,0.1), D(v,0.5) and D(v,0.9) values should then be determined. For material that has a D(v,0.5) greater than 10 microns an acceptable time point is defined when the following conditions are met: D(v,0.5) RSD <3%, D(v,0.1) and D(v,0.9) RSD <5%. For material that has a D(v,0.5) less than 10 microns an acceptable time point is defined when the following conditions are met: D(v,0.5) RSD <6%, D(v,0.1) and D(v,0.9) RSD <10%. This reflects the fact that smaller particles are more difficult to disperse. These are the limits of acceptability as laid down in ISO13320 [1] and for many samples much lower RSDs should be easily achieved. Typical measurement stability results obtained for a lactose sample are show in table 3 and figure 1.

5 Measurement Stabilisation Time / Minutes D(v,0.1) / Microns %RSD D(v,0.5) / Microns %RSD D(v,0.9) / Microns %RSD Table 3: Variation of lactose size over time (mean of 5 repeat measurements are shown). Figure 1: Sample stability measurement summary. From the above results it is clear that the lactose sample is stable whilst in suspension. It was decided from these results that measurements should be taken after 1 minute in order to give the sample time to equilibrate. c) Air Pressure (Dry measurement) / Ultrasound (wet measurement) A pressure titration should have previously been performed as part of method development (see ISO section [1] ). A suitable pressure is one at which dispersion of the particles is achieved but milling of the particles is not occurring. Most pharmaceuticals are friable and will be ground in any dry powder feeder if the pressure is too high. Often both dispersion and milling occur simultaneously (which leads to a broadening of the distribution) [6]. By measuring the same amount of sample (ideally sampled to single shot size by a spinning riffler) at different pressures, the pressure at which maximum dispersion is achieved without milling can be ascertained. The best way of proving that no attrition is occurring is to achieve near identical results for both wet and dry dispersion [1].

6 Figure 2 shows a pressure titration on a typical material. A pressure of 1 bar is not enough to disperse the material. As long as the pressure is greater than 2 bar good results are obtained. Figure 2: Pressure titration data obtained for a typical lactose sample. The use of ultrasound in wet measurements is very similar. ISO recommends that ultrasound can be used to assist dispersion [1]. Too much ultrasound can fracture friable pharmaceutical particles, although this is extremely unusual. As part of method development the affect of varying the duration and power of ultrasound on the particle size distribution should be examined. Ideally measurements should be taken before, during and after ultrasound to examine what effect sonication has on the robustness of the measurement. As well as separating particles, ultrasound can also increase the rate of particle particle collision and can actually cause agglomeration d) Pump and Stir Rates The pump and stir speeds used during a measurement should be examined as part of method development. The chosen conditions should be capable of suspending all the material without causing air entrainment (a particular problem if surfactants are being used). Figure 3 shows how the result obtained for a lactose sample varies according to the stirrer settings. As can be seen, the result reaches a plateau above 2000rpm. It is at this point that all of the material is correctly suspended and dispersed. Sample sedimentation causes the results to be smaller than expected at stirrer rates below 2000rpm.

7 Figure 3: Effect of varying the stirrer rate on the result obtained for a typical lactose sample. e) Confirmation of refractive index choice As part of the method development work, the choice of refractive index should be examined. Index matching fluids can be used to provide experimental evidence of the real refractive index. The Fraunhofer approximation should not be used if there are particles smaller than 40λ (25 microns at the He-Ne laser wavelength of µm) present in the distribution [1] as it may erroneously report the presence of fine material. LINEARITY / OBSCURATION Obscuration is a measure of the amount of light scattered by the sample and is defined here as (1-Transmission)*100. For most particle size distributions within a given concentration range the particle size should be independent of concentration. At extremely low concentration results with large RSDs may be obtained (owing to high signal to noise ratios) and at extremely high concentrations, the result may be smaller than expected due to multiple scattering. It is suggested that obscurations of 5, 10, 15, 20 and 25% are investigated in the same way that the measurement duration test was performed, with the acceptable RSDs being similarly specified. REPRODUCIBILITY Reproducibility is defined by Bell et al [3] as an indicator of precision between laboratories. It can indeed show this, but in the authors experience is far more likely to show how effective the sampling regime in use is i.e. how homogenous or otherwise the sample is. It can also be used to flag differences between different instruments (be it of the same or different models). A number of samples (at least five) should be taken from the same batch and tested in accordance with the method

8 under investigation. For each sample at least five repeat measurements should be taken and the individual and average results should be obtained. From the average result for each of the five repeat measurements, the RSD should be determined for the D(v,0.1), D(v,0.5) and D(v,0.9) for all the samples taken. For material with a D(v,0.5) greater then 10 micron, good reproducibility is shown if the D(v,0.5) RSD <3%,D(v,0.1) and D(v,0.9) RSD<5%. For material that has a D(v,0.5) less than 10 micron, good reproducibility is shown if the D(v,0.5) RSD<6%, D(v,0.1) and D(v,0.9) RSD<10%. Again these are the limits set out within ISO [1]. Sample Number D(v,0.1) / Microns D(v,0.5) / Microns D(v,0.9) / Microns Mean %RSD Table 4: Variation in the results obtained for seven separate scooped-sampled lactose samples. Scoop sampling is being used here, and the RSDs are in line with those reported in table 1. INTERMEDIATE PRECISION The checking of the precision of the technique should be done by the use of a second analyst or a second instrument (or both). This should in essence be a repeat of the reproducibility test. The RSD of the D(v,0.5) on the second test should again be <3%. Both sets of results should then be combined to give a pooled mean and a pooled RSD (which should be <3%). Sample Number D(v,0.1) / Microns D(v,0.5) / Microns D(v,0.9) / Microns Mean %RSD Table 5: Results obtained for a second analyst measuring the same lactose sample.

9 Parameter Pooled Mean Value Mean Standard Deviation 0.68 %RSD 2.98 Table 6: Pooled mean values. The data here has a RSD<3% and shows acceptable intermediate precision. Validation terms such as detection limit and quantification limit do not apply to laser diffraction methods and are therefore outside the scope of this paper. The detection limit or instrument resolution will depend on the instrument in question. CONCLUSION This paper has attempted to lay down guidelines on how to validate a particle sizing method. Guidelines on achievable standard deviations have also been supplied. REFERENCES [1] ISO Particle Size Analysis Laser Diffraction Methods Part 1: General Principles (1999) [2] Jillavenkatesa, A, Dapkunas, S. J. and Lum, L. S., (2001) Particle Size Characterization Practice Guide, N.I.S.T [3] Bell, R., Dennis, A., Hendriksen, B., North, N. and Sherwood, J., (1999) Position paper on Particle Sizing: Sample Preparation, Method Validation and Data Presentation, Pharmaceutical Technology Europe, November 1999 [4] Lerke, S.A. and Adams, S.A, (2002) Development and Validation of a Particle Size Distribution Method for Analysis of Drug Substance, American Pharmaceutical Review, Fall 2002 [5] Allen, T. Particle Size Measurement, (1999) 5 th edition, Volume 1,p 38, Chapman and Hall, London [6] Rawle, A. F., (2000) "Attrition, dispersion and sampling effects in dry and wet particle size analysis using laser diffraction" Paper th International Congress of Chemical and Process Engineering "CHISA'2000", August, Praha, Czech Republic

Method Development and Validation for Particle Size and Shape Measurements

Method Development and Validation for Particle Size and Shape Measurements Method Development and Validation for Particle Size and Shape Measurements Ulf Willén Divisional Product Manager Analytical Imaging Systems Malvern Instruments Ltd, Malvern, UK. FDA guidance: when should

More information

Validation and Calibration. Definitions and Terminology

Validation and Calibration. Definitions and Terminology Validation and Calibration Definitions and Terminology ACCEPTANCE CRITERIA: The specifications and acceptance/rejection criteria, such as acceptable quality level and unacceptable quality level, with an

More information

Molysulfide. Particle Size Analysis

Molysulfide. Particle Size Analysis Molysulfide Particle Size Analysis INTRODUCTION Molysulfide is the Climax Molybdenum Company trade name for lubricant grade molybdenum disulfide (MoS 2 ). The mineral designation for MoS 2 is Molybdenite

More information

ANALYTICAL METHODS INTERNATIONAL QUALITY SYSTEMS

ANALYTICAL METHODS INTERNATIONAL QUALITY SYSTEMS VALIDATION OF ANALYTICAL METHODS 1 GERT BEUVING INTERNATIONAL PHARMACEUTICAL OPERATIONS TASKS: - Internal auditing - Auditing of suppliers and contract manufacturers - Preparing for and guiding of external

More information

VALIDATION OF ANALYTICAL PROCEDURES: TEXT AND METHODOLOGY Q2(R1)

VALIDATION OF ANALYTICAL PROCEDURES: TEXT AND METHODOLOGY Q2(R1) INTERNATIONAL CONFERENCE ON HARMONISATION OF TECHNICAL REQUIREMENTS FOR REGISTRATION OF PHARMACEUTICALS FOR HUMAN USE ICH HARMONISED TRIPARTITE GUIDELINE VALIDATION OF ANALYTICAL PROCEDURES: TEXT AND METHODOLOGY

More information

Step-by-Step Analytical Methods Validation and Protocol in the Quality System Compliance Industry

Step-by-Step Analytical Methods Validation and Protocol in the Quality System Compliance Industry Step-by-Step Analytical Methods Validation and Protocol in the Quality System Compliance Industry BY GHULAM A. SHABIR Introduction Methods Validation: Establishing documented evidence that provides a high

More information

DYNAMIC LIGHT SCATTERING COMMON TERMS DEFINED

DYNAMIC LIGHT SCATTERING COMMON TERMS DEFINED DYNAMIC LIGHT SCATTERING COMMON TERMS DEFINED Abstract: There are a number of sources of information that give a mathematical description of the terms used in light scattering. However, these will not

More information

PARTICLE SIZE. By: Philo Morse, MS, and Andrew Loxley, PhD. INTRODUCTION are not as stringent as for scattering techniques.

PARTICLE SIZE. By: Philo Morse, MS, and Andrew Loxley, PhD. INTRODUCTION are not as stringent as for scattering techniques. D I S TR I BU T I O N Light Microscopic Determination of Particle Size Distribution in an Aqueous Gel By: Philo Morse, MS, and Andrew Loxley, PhD of diffracted light at the detectors, though dilution requirements

More information

Improving the consistency of particle size measurement. Improving the consistency of particle size measurement

Improving the consistency of particle size measurement. Improving the consistency of particle size measurement A NATIONAL MEASUREMENT GOOD PRACTICE GUIDE A NATIONAL MEASUREMENT GOOD PRACTICE GUIDE No. 111 No. 111 Improving the consistency of particle size measurement Improving the consistency of particle size measurement

More information

Guidance for Industry

Guidance for Industry Guidance for Industry Q2B Validation of Analytical Procedures: Methodology November 1996 ICH Guidance for Industry Q2B Validation of Analytical Procedures: Methodology Additional copies are available from:

More information

ICH Topic Q 2 (R1) Validation of Analytical Procedures: Text and Methodology. Step 5

ICH Topic Q 2 (R1) Validation of Analytical Procedures: Text and Methodology. Step 5 European Medicines Agency June 1995 CPMP/ICH/381/95 ICH Topic Q 2 (R1) Validation of Analytical Procedures: Text and Methodology Step 5 NOTE FOR GUIDANCE ON VALIDATION OF ANALYTICAL PROCEDURES: TEXT AND

More information

Particle size. Integrated systems for particle sizing. detailed specification sheets from www.malvern.co.uk/ms2000

Particle size. Integrated systems for particle sizing. detailed specification sheets from www.malvern.co.uk/ms2000 Particle size 2000 Integrated systems for particle sizing detailed specification sheets from /ms2000 Comprehensive particle sizing The Mastersizer 2000 is a practical, reliable solution to the everyday

More information

Setting Size Specifications

Setting Size Specifications SETTING PARTICLE SIZE SPECIFICATIONS The creation of a meaningful and product-appropriate particle size specification requires knowledge of its effect on product performance in addition to an understanding

More information

The Basic Principles of Sieve Analysis

The Basic Principles of Sieve Analysis The Basic Principles of Sieve Analysis Introduction Many natural and manufactured materials occur in a disperse form, which means that they consist of differently shaped and sized particles. The particle

More information

LIFE SCIENCE I TECHNICAL BULLETIN ISSUE N 11 /JULY 2008

LIFE SCIENCE I TECHNICAL BULLETIN ISSUE N 11 /JULY 2008 LIFE SCIENCE I TECHNICAL BULLETIN ISSUE N 11 /JULY 2008 PARTICLE CHARACTERISATION IN EXCIPIENTS, DRUG PRODUCTS AND DRUG SUBSTANCES AUTHOR: HILDEGARD BRÜMMER, PhD, CUSTOMER SERVICE MANAGER, SGS LIFE SCIENCE

More information

GUIDELINES FOR THE VALIDATION OF ANALYTICAL METHODS FOR ACTIVE CONSTITUENT, AGRICULTURAL AND VETERINARY CHEMICAL PRODUCTS.

GUIDELINES FOR THE VALIDATION OF ANALYTICAL METHODS FOR ACTIVE CONSTITUENT, AGRICULTURAL AND VETERINARY CHEMICAL PRODUCTS. GUIDELINES FOR THE VALIDATION OF ANALYTICAL METHODS FOR ACTIVE CONSTITUENT, AGRICULTURAL AND VETERINARY CHEMICAL PRODUCTS October 2004 APVMA PO Box E240 KINGSTON 2604 AUSTRALIA http://www.apvma.gov.au

More information

Particle size measurement of lactose for dry powder inhalers

Particle size measurement of lactose for dry powder inhalers Particle size measurement of lactose for dry powder inhalers MCC Starch Lactose Inhalation Superdisintegrants 1 Introduction The particle size of lactose has been shown to be important for dry powder

More information

Top and Bottom Spray Fluid Bed Granulation Process

Top and Bottom Spray Fluid Bed Granulation Process Top and Bottom Spray Fluid Bed Granulation Process Use of Lasentec FBRM In-Process Particle Sizing PAT Technique to Study Top and Bottom Spray Fluid Bed Granulation Process Presented November 10, 2005

More information

Revision of The Dissolution Procedure: Development and Validation 1092

Revision of The Dissolution Procedure: Development and Validation 1092 Page 1 of 5 STIMULI TO THE REVISION PROCESS Stimuli articles do not necessarily reflect the policies of the USPC or the USP Council of Experts Revision of The Dissolution Procedure: Development and Validation

More information

IN-LINE PARTICLE SIZE MEASUREMENTS FOR CEMENT AND OTHER ABRASIVE PROCESS ENVIRONMENTS

IN-LINE PARTICLE SIZE MEASUREMENTS FOR CEMENT AND OTHER ABRASIVE PROCESS ENVIRONMENTS IN-LINE PARTICLE SIZE MEASUREMENTS FOR CEMENT AND OTHER ABRASIVE PROCESS ENVIRONMENTS 1998 A.P. Malcolmson Malvern, Inc. 10 Southville Road Southborough, MA D. J. Holve Malvern/Insitec, Inc. 2110 Omega

More information

Analytical Test Method Validation Report Template

Analytical Test Method Validation Report Template Analytical Test Method Validation Report Template 1. Purpose The purpose of this Validation Summary Report is to summarize the finding of the validation of test method Determination of, following Validation

More information

OMCL Network of the Council of Europe QUALITY MANAGEMENT DOCUMENT

OMCL Network of the Council of Europe QUALITY MANAGEMENT DOCUMENT OMCL Network of the Council of Europe QUALITY MANAGEMENT DOCUMENT PA/PH/OMCL (12) 77 7R QUALIFICATION OF EQUIPMENT ANNEX 8: QUALIFICATION OF BALANCES Full document title and reference Document type Qualification

More information

Guideline on Process Validation

Guideline on Process Validation 1 2 3 4 29 March 2012 EMA/CHMP/CVMP/QWP/70278/2012-Rev1 Committee for Medicinal Products for Human Use (CHMP) Committee for Medicinal Products for Veterinary Use (CVMP) 5 6 Draft Draft Agreed by CHMP /

More information

RESOLUTION-RDC Nr- 48, OF OCTOBER 6, 2009

RESOLUTION-RDC Nr- 48, OF OCTOBER 6, 2009 RESOLUTION-RDC Nr- 48, OF OCTOBER 6, 2009 This resolution provides for the alteration, inclusion, suspension, reactivation and cancellation after registration of medications and sets forth other provisions.

More information

GENERAL GUIDANCE FOR INSPECTORS ON HOLD-TIME STUDIES

GENERAL GUIDANCE FOR INSPECTORS ON HOLD-TIME STUDIES February 2013 RESTRICTED GENERAL GUIDANCE FOR INSPECTORS ON HOLD-TIME STUDIES DRAFT FOR COMMENT Should you have any comments on the attached text, please send these to Dr Sabine Kopp, Manager, Medicines

More information

International GMP Requirements for Quality Control Laboratories and Recomendations for Implementation

International GMP Requirements for Quality Control Laboratories and Recomendations for Implementation International GMP Requirements for Quality Control Laboratories and Recomendations for Implementation Ludwig Huber, Ph.D. ludwig_huber@labcompliance.com Overview GMP requirements for Quality Control laboratories

More information

Commercial Manufacturing - Qualification & Validation-related GMP Deficiencies and Other Lifecycle Considerations

Commercial Manufacturing - Qualification & Validation-related GMP Deficiencies and Other Lifecycle Considerations Commercial Manufacturing - Qualification & Validation-related GMP Deficiencies and Other Lifecycle Considerations Kevin O Donnell PhD Market Compliance Manager, IMB PDA / FDA Conference Pharmaceutical

More information

UV/VIS/IR SPECTROSCOPY ANALYSIS OF NANOPARTICLES

UV/VIS/IR SPECTROSCOPY ANALYSIS OF NANOPARTICLES UV/VIS/IR SPECTROSCOPY ANALYSIS OF NANOPARTICLES SEPTEMBER 2012, V 1.1 4878 RONSON CT STE K SAN DIEGO, CA 92111 858-565 - 4227 NANOCOMPOSIX.COM Note to the Reader: We at nanocomposix have published this

More information

SPECIFICATIONS AND CONTROL TESTS ON THE FINISHED PRODUCT

SPECIFICATIONS AND CONTROL TESTS ON THE FINISHED PRODUCT SPECIFICATIONS AND CONTROL TESTS ON THE FINISHED PRODUCT Guideline Title Specifications and Control Tests on the Finished Product Legislative basis Directive 75/318/EEC as amended Date of first adoption

More information

Journal of Chemical and Pharmaceutical Research

Journal of Chemical and Pharmaceutical Research Available on line www.jocpr.com Journal of Chemical and Pharmaceutical Research ISSN No: 0975-7384 CODEN(USA): JCPRC5 J. Chem. Pharm. Res., 2011, 3(2):892-898 World Health Organization s Guidelines for

More information

Reversed Phase High Presssure Liquid Chromatograhphic Technique for Determination of Sodium Alginate from Oral Suspension

Reversed Phase High Presssure Liquid Chromatograhphic Technique for Determination of Sodium Alginate from Oral Suspension International Journal of PharmTech Research CODEN (USA): IJPRIF ISSN : 0974-4304 Vol.2, No.2, pp 1634-1638, April-June 2010 Reversed Phase High Presssure Liquid Chromatograhphic Technique for Determination

More information

Design of Experiments for Analytical Method Development and Validation

Design of Experiments for Analytical Method Development and Validation Design of Experiments for Analytical Method Development and Validation Thomas A. Little Ph.D. 2/12/2014 President Thomas A. Little Consulting 12401 N Wildflower Lane Highland, UT 84003 1-925-285-1847 drlittle@dr-tom.com

More information

Reference Materials for Environmental Performance Testing Dr Steve Wood Head of Regulatory and Legislative Services. ISPRA 25 June 2009

Reference Materials for Environmental Performance Testing Dr Steve Wood Head of Regulatory and Legislative Services. ISPRA 25 June 2009 Reference Materials for Environmental Performance Testing Dr Steve Wood Head of Regulatory and Legislative Services ISPRA 25 June 2009 Outline Background to LGC UK MCERTS scheme Reference materials production

More information

105 Adopted: 27.07.95

105 Adopted: 27.07.95 105 Adopted: 27.07.95 OECD GUIDELINE FOR THE TESTING OF CHEMICALS Adopted by the Council on 27 th July 1995 Water Solubility INTRODUCTION 1. This guideline is a revised version of the original Guideline

More information

ISO 21501 A Standard Methodology to Optical Particle Counter Calibration and What It Means to Cleanroom Owners

ISO 21501 A Standard Methodology to Optical Particle Counter Calibration and What It Means to Cleanroom Owners ISO 21501 A Standard Methodology to Optical Particle Counter Calibration and What It Means to Cleanroom Owners Tony Harrison and Bob Latimer ISO 21501- A Standard Methodology to Optical Particle Counter

More information

ZETA POTENTIAL ANALYSIS OF NANOPARTICLES

ZETA POTENTIAL ANALYSIS OF NANOPARTICLES ZETA POTENTIAL ANALYSIS OF NANOPARTICLES SEPTEMBER 2012, V 1.1 4878 RONSON CT STE K SAN DIEGO, CA 92111 858-565 - 4227 NANOCOMPOSIX.COM Note to the Reader: We at nanocomposix have published this document

More information

Analysis of the Vitamin B Complex in Infant Formula Samples by LC-MS/MS

Analysis of the Vitamin B Complex in Infant Formula Samples by LC-MS/MS Analysis of the Vitamin B Complex in Infant Formula Samples by LC-MS/MS Stephen Lock 1 and Matthew Noestheden 2 1 AB SCIEX Warrington, Cheshire (UK), 2 AB SCIEX Concord, Ontario (Canada) Overview A rapid,

More information

Mitos P-Pump. product datasheet

Mitos P-Pump. product datasheet product datasheet page Product description 2 Main benefits 3 Why choose the Mitos P-Pump? 3 Closed-loop flow control 4 Product specifications 5 Accessories - overview 6 Mitos P-Pump Basic 7 Mitos P-Pump

More information

Analytical Methods: A Statistical Perspective on the ICH Q2A and Q2B Guidelines for Validation of Analytical Methods

Analytical Methods: A Statistical Perspective on the ICH Q2A and Q2B Guidelines for Validation of Analytical Methods Page 1 of 6 Analytical Methods: A Statistical Perspective on the ICH Q2A and Q2B Guidelines for Validation of Analytical Methods Dec 1, 2006 By: Steven Walfish BioPharm International ABSTRACT Vagueness

More information

Particle Size Analysis

Particle Size Analysis Particle Size Analysis Why measure particle size of pharmaceuticals??? Particle size can affect Processability of powder (API or excipient) Flow, mixing, compaction Final formulation: performance, appearance,

More information

Assay Development and Method Validation Essentials

Assay Development and Method Validation Essentials Assay Development and Method Validation Essentials Thomas A. Little Ph.D. 10/13/2012 President Thomas A. Little Consulting 12401 N Wildflower Lane Highland, UT 84003 1-925-285-1847 drlittle@dr-tom.com

More information

Appendix 5 Overview of requirements in English

Appendix 5 Overview of requirements in English Appendix 5 Overview of requirements in English This document is a translation of Appendix 4 (Bilag 4) section 2. This translation is meant as a service for the bidder and in case of any differences between

More information

Getting Started. MAN 0101 Issue1.3 August 1997

Getting Started. MAN 0101 Issue1.3 August 1997 Getting Started MAN 0101 Issue1.3 August 1997 Malvern Instruments Ltd. 1995, 1997. Malvern Instruments makes every effort to ensure that this document is correct. However, due to Malvern Instruments policy

More information

IVHHN GUIDELINES FOR GRAIN-SIZE DISTRIBUTION ANALYSIS

IVHHN GUIDELINES FOR GRAIN-SIZE DISTRIBUTION ANALYSIS IVHHN GUIDELINES FOR GRAIN-SIZE DISTRIBUTION ANALYSIS Introduction Grain-size distribution (GSD) analysis is a vital step in the assessment of the health hazard of volcanic ash, giving the quantity of

More information

IUCLID 5 COMPOSITION AND ANALYSIS GUIDANCE DOCUMENT: IRON ORES, AGGLOMERATES [EINECS NUMBER 265 996 3, CAS NUMBER 65996 65 8] IRON ORE PELLETS

IUCLID 5 COMPOSITION AND ANALYSIS GUIDANCE DOCUMENT: IRON ORES, AGGLOMERATES [EINECS NUMBER 265 996 3, CAS NUMBER 65996 65 8] IRON ORE PELLETS IUCLID 5 COMPOSITION AND ANALYSIS GUIDANCE DOCUMENT: IRON ORES, AGGLOMERATES [EINECS NUMBER 265 996 3, CAS NUMBER 65996 65 8] IRON ORE PELLETS INTRODUCTION Each REACH registrant is required to file its

More information

Fundamentals of modern UV-visible spectroscopy. Presentation Materials

Fundamentals of modern UV-visible spectroscopy. Presentation Materials Fundamentals of modern UV-visible spectroscopy Presentation Materials The Electromagnetic Spectrum E = hν ν = c / λ 1 Electronic Transitions in Formaldehyde 2 Electronic Transitions and Spectra of Atoms

More information

Overview of Pre-Approval Inspections

Overview of Pre-Approval Inspections Overview of Pre-Approval Inspections Presented by: Kelli F. Dobilas NWJ-DO Pre-Approval Manager Pre-Approval Drug Inspections What are Pre-Approval Inspections? One of the last reviews of the drug approval

More information

MILESTONE H E L P I N G C H E M I S T S

MILESTONE H E L P I N G C H E M I S T S MILESTONE H E L P I N G C H E M I S T S 8 good reasons to choose the Milestone DMA-80 1. No sample preparation The DMA-80 does not require any sample preparation or other wet chemistry prior the analysis.

More information

METHOD 9075 TEST METHOD FOR TOTAL CHLORINE IN NEW AND USED PETROLEUM PRODUCTS BY X-RAY FLUORESCENCE SPECTROMETRY (XRF)

METHOD 9075 TEST METHOD FOR TOTAL CHLORINE IN NEW AND USED PETROLEUM PRODUCTS BY X-RAY FLUORESCENCE SPECTROMETRY (XRF) METHOD 9075 TEST METHOD FOR TOTAL CHLORINE IN NEW AND USED PETROLEUM PRODUCTS BY X-RAY FLUORESCENCE SPECTROMETRY (XRF) 1.0 SCOPE AND APPLICATION 1.1 This test method covers the determination of total chlorine

More information

OMCL Network of the Council of Europe QUALITY MANAGEMENT DOCUMENT

OMCL Network of the Council of Europe QUALITY MANAGEMENT DOCUMENT OMCL Network of the Council of Europe QUALITY MANAGEMENT DOCUMENT PA/PH/OMCL (11) 204 3R HANDLING AND USE OF REFERENCE STANDARDS IN THE OMCL NETWORK Full document title and reference Document type Legislative

More information

Why System Suitability Tests Are Not a Substitute for Analytical Instrument Qualification

Why System Suitability Tests Are Not a Substitute for Analytical Instrument Qualification White paper System Suitability Tests and AIQ R.D.McDowall PhD Principal, McDowall Consulting Paul Smith European Validation Program Manager, PerkinElmer Nicola Vosloo PhD European Market Development Leader,

More information

Multivariate Tools for Modern Pharmaceutical Control FDA Perspective

Multivariate Tools for Modern Pharmaceutical Control FDA Perspective Multivariate Tools for Modern Pharmaceutical Control FDA Perspective IFPAC Annual Meeting 22 January 2013 Christine M. V. Moore, Ph.D. Acting Director ONDQA/CDER/FDA Outline Introduction to Multivariate

More information

C 5. chemical development contract research custom synthesis cgmp API manufacturing commercial production. Welcome to

C 5. chemical development contract research custom synthesis cgmp API manufacturing commercial production. Welcome to C 5 chemical development contract research custom synthesis cgmp API manufacturing commercial production Welcome to ChemCon Company profile Company profile ChemCon offers outstanding chemical services

More information

Validating Methods using Waters Empower TM 2 Method. Validation. Manager

Validating Methods using Waters Empower TM 2 Method. Validation. Manager Validating Methods using Waters Empower TM 2 Method Validation Manager (MVM) Anders Janesten Nordic Informatics Sales Specialist 2008 Waters Corporation Decision-centric centric method development information

More information

ASSURING THE QUALITY OF TEST RESULTS

ASSURING THE QUALITY OF TEST RESULTS Page 1 of 12 Sections Included in this Document and Change History 1. Purpose 2. Scope 3. Responsibilities 4. Background 5. References 6. Procedure/(6. B changed Division of Field Science and DFS to Office

More information

USE OF REFERENCE MATERIALS IN THE LABORATORY

USE OF REFERENCE MATERIALS IN THE LABORATORY USE OF REFERENCE MATERIALS IN THE LABORATORY What is a reference material? A general definition of a reference material is a material or substance one or more of whose property values are sufficiently

More information

GMP ANNEX 1 REVISION 2008, INTERPRETATION OF MOST IMPORTANT CHANGES FOR THE MANUFACTURE OF STERILE MEDICINAL PRODUCTS

GMP ANNEX 1 REVISION 2008, INTERPRETATION OF MOST IMPORTANT CHANGES FOR THE MANUFACTURE OF STERILE MEDICINAL PRODUCTS PHARMACEUTICAL INSPECTION CONVENTION PHARMACEUTICAL INSPECTION CO-OPERATION SCHEME PI 032-2 8 January 2010 RECOMMENDATION GMP ANNEX 1 REVISION 2008, INTERPRETATION OF MOST IMPORTANT CHANGES FOR THE MANUFACTURE

More information

Measuring Line Edge Roughness: Fluctuations in Uncertainty

Measuring Line Edge Roughness: Fluctuations in Uncertainty Tutor6.doc: Version 5/6/08 T h e L i t h o g r a p h y E x p e r t (August 008) Measuring Line Edge Roughness: Fluctuations in Uncertainty Line edge roughness () is the deviation of a feature edge (as

More information

UNIVERSITY OF PARMA INTERDEPARTMENTAL CENTRE FOR INNOVATION IN HEALTH PRODUCTS, BIOPHARMANET-TEC

UNIVERSITY OF PARMA INTERDEPARTMENTAL CENTRE FOR INNOVATION IN HEALTH PRODUCTS, BIOPHARMANET-TEC UNIVERSITY OF PARMA INTERDEPARTMENTAL CENTRE FOR INNOVATION IN HEALTH PRODUCTS, BIOPHARMANET-TEC RULES FOR THE USE OF THE CENTER EQUIPMENTS ACQUIRED WITHIN THE AGREEMENT BETWEEN THE UNIVERSITY OF PARMA

More information

White paper: FDA Guidance for Industry Update Process Validation

White paper: FDA Guidance for Industry Update Process Validation White paper: FDA Guidance for Industry Update Process Validation In January 2011, the FDA released the final version of its long-awaited update to its Process Validation Guidance for Industry. Since then,

More information

BASIC PRINCIPLES OF PARTICLE

BASIC PRINCIPLES OF PARTICLE T E C H N I C A L P A P E R BASIC PRINCIPLES OF PARTICLE SIZE ANALYSIS Written by Dr. Alan Rawle,Malvern Instruments Limited, Enigma Business Park, Grovewood Road, Malvern, Worcestershire, WR14 1XZ, UK.

More information

AMBERLITE IRP64 Pharmaceutical Grade Cation Exchange Resin (Polacrilex Resin)

AMBERLITE IRP64 Pharmaceutical Grade Cation Exchange Resin (Polacrilex Resin) AMBERLITE IRP64 Pharmaceutical Grade Cation Exchange Resin (Polacrilex Resin) Description AMBERLITE IRP64 [1] resin is an insoluble, weakly acidic, hydrogen form, cation exchange resin supplied as a dry,

More information

QbD Approach to Assay Development and Method Validation

QbD Approach to Assay Development and Method Validation QbD Approach to Assay Development and Method Validation Thomas A. Little Ph.D. 11/105/2014 President Thomas A. Little Consulting 12401 N Wildflower Lane Highland, UT 84003 1-925-285-1847 drlittle@dr-tom.com

More information

Gold Nanoparticles & Custom Conjugation

Gold Nanoparticles & Custom Conjugation Gold Nanoparticles & Custom Conjugation p 02 Our unique gold has been used for over 20 years ensuring you get superior quality and performance. Gold nanoparticles for lateral flow development and maufacture

More information

Guidance for Industry

Guidance for Industry Guidance for Industry Investigating Out-of-Specification (OOS) Test Results for Pharmaceutical Production U.S. Department of Health and Human Services Food and Drug Administration Center for Drug Evaluation

More information

INSTRUCTIONS 56-1190-98. Edition AC

INSTRUCTIONS 56-1190-98. Edition AC Sephacryl S-100 High Resolution Sephacryl S-200 High Resolution Sephacryl S-300 High Resolution Sephacryl S-400 High Resolution Sephacryl S-500 High Resolution INSTRUCTIONS Sephacryl High Resolution chromatography

More information

OLIVÉR BÁNHIDI 1. Introduction

OLIVÉR BÁNHIDI 1. Introduction Materials Science and Engineering, Volume 39, No. 1 (2014), pp. 5 13. DETERMINATION OF THE ANTIMONY- AND STRONTIUM- CONTENT OF ALUMINIUM ALLOYS BY INDUCTIVELY COUPLED PLASMA ATOM EMISSION SPECTROMETRY

More information

POWDER PROPERTIES LABORATORY

POWDER PROPERTIES LABORATORY Ground Rules POWDER PROPERTIES LABORATORY You will work as a team of no more than 6 students. At the end of this laboratory session each team will turn in a single report. The report will be reviewed,

More information

About Coffee and Refractometers 2008-2010 Voice Systems Technology, Inc. (VST)

About Coffee and Refractometers 2008-2010 Voice Systems Technology, Inc. (VST) About Coffee and Refractometers 2008-200 Voice Systems Technology, Inc. (VST) www.mojotogo.us.0 Coffee and Refractive Index Refractive Index measurements have been used for process control in the food

More information

Module 13 : Measurements on Fiber Optic Systems

Module 13 : Measurements on Fiber Optic Systems Module 13 : Measurements on Fiber Optic Systems Lecture : Measurements on Fiber Optic Systems Objectives In this lecture you will learn the following Measurements on Fiber Optic Systems Attenuation (Loss)

More information

Dispersing Powders in Liquid

Dispersing Powders in Liquid Dispersing Powders in Liquid Mark Bumiller mark.bumiller@horiba.com Definitions Typical powder: individual particles and clumps stuck together by weak or strong forces Agglomerates: assemblage of particles

More information

ICH Topic Q 1 A Stability Testing Guidelines: Stability Testing of New Drug Substances and Products

ICH Topic Q 1 A Stability Testing Guidelines: Stability Testing of New Drug Substances and Products The European Agency for the Evaluation of Medicinal Products Human Medicines Evaluation Unit CPMP/ICH/380/95 ICH Topic Q 1 A Stability Testing Guidelines: Stability Testing of New Drug Substances and Products

More information

Assay Qualification Template for Host Cell Protein ELISA

Assay Qualification Template for Host Cell Protein ELISA Assay Qualification Template for Host Cell Protein ELISA Introduction: With ever increasing importance being placed on host cell protein (HCP) removal during the purification process, it is critical to

More information

VISCOSE FIBRES WITH NEW FUNCTIONAL QUALITIES

VISCOSE FIBRES WITH NEW FUNCTIONAL QUALITIES VISCOSE FIBRES WITH NEW FUNCTIONAL QUALITIES Walter Roggenstein Kelheim Fibres GmbH, Regensburger Str. 109, 93309 Kelheim, Germany Phone: (+49) 9441 99-489; Fax: (+49) 9441 99-1489; Email: walter.roggenstein@kelheim-fibres.com

More information

The importance of normalisation when comparing tablet properties

The importance of normalisation when comparing tablet properties The importance of normalisation when comparing tablet properties Tablet quality definition The properties of a tablet, both during manufacturing and in vivo, are determined by the properties of the materials

More information

DMA-1 Direct Mercury Analyzer

DMA-1 Direct Mercury Analyzer DMA-1 Direct Mercury Analyzer www.milestonesrl.com MILESTONE H E L P I N G C H E M I S T S 8 good reasons to choose the milestone DmA-1 1. no sample preparation 2. ease of use 3. lowest cost of analysis

More information

Workshop B Control Strategy

Workshop B Control Strategy ICH-GCG ASEAN Workshop B Control Strategy Jean-Louis ROBERT, Ph.D. National Health Laboratory, Luxembourg Chair-person ICH IWG Q8, Q9, Q10 Kuala Lumpur, 26-28 July 2010 International Conference on Harmonisation

More information

Qualitative NIR Analysis for Ingredients in the Baking Industry

Qualitative NIR Analysis for Ingredients in the Baking Industry Overview The challenge to all baking companies in today s economy is to operate plants as efficiently as possible, with a focus on quality and keeping costs in check. With regulatory issues becoming more

More information

Thermo Scientific SOLAµ SPE Plates Technical Guide. Consistent excellence. for bioanalysis

Thermo Scientific SOLAµ SPE Plates Technical Guide. Consistent excellence. for bioanalysis Thermo Scientific SOLAµ SPE Plates Technical Guide Consistent excellence for bioanalysis SOLAµ - delivering reproducible low volume extractions. Everytime! Thermo Scientific SOLAµ plates are designed for

More information

INTERNATIONAL JOURNAL OF PHARMACEUTICAL RESEARCH AND BIO-SCIENCE

INTERNATIONAL JOURNAL OF PHARMACEUTICAL RESEARCH AND BIO-SCIENCE Rajgor VM,, 2014; Volume 3(3): 188-197 INTERNATIONAL JOURNAL OF PHARMACEUTICAL RESEARCH AND BIO-SCIENCE ANALYTICAL METHOD DEVELOPEMENT AND VALIDATION FOR THE SIMULTANEOUS ESTIMATION OF MEMANTINE HCL AND

More information

Calibration & Preventative Maintenance. Sally Wolfgang Manager, Quality Operations Merck & Co., Inc.

Calibration & Preventative Maintenance. Sally Wolfgang Manager, Quality Operations Merck & Co., Inc. Calibration & Preventative Maintenance Sally Wolfgang Manager, Quality Operations Merck & Co., Inc. Calibration: A comparison of two instruments or measuring devices one of which is a standard of known

More information

Determination of Insoluble Solids in Pretreated Biomass March 2008 Material

Determination of Insoluble Solids in Pretreated Biomass March 2008 Material National Renewable Energy Laboratory Innovation for Our Energy Future A national laboratory of the U.S. Department of Energy Office of Energy Efficiency & Renewable Energy Determination of Insoluble Solids

More information

Refractive Index Measurement Principle

Refractive Index Measurement Principle Refractive Index Measurement Principle Refractive index measurement principle Introduction Detection of liquid concentrations by optical means was already known in antiquity. The law of refraction was

More information

A Guide to Acousto-Optic Modulators

A Guide to Acousto-Optic Modulators A Guide to Acousto-Optic Modulators D. J. McCarron December 7, 2007 1 Introduction Acousto-optic modulators (AOMs) are useful devices which allow the frequency, intensity and direction of a laser beam

More information

Applying Statistics Recommended by Regulatory Documents

Applying Statistics Recommended by Regulatory Documents Applying Statistics Recommended by Regulatory Documents Steven Walfish President, Statistical Outsourcing Services steven@statisticaloutsourcingservices.com 301-325 325-31293129 About the Speaker Mr. Steven

More information

Relative Humidity Sensors, Transmitters and Calibration Products Relative Humidity Measurement and Calibration

Relative Humidity Sensors, Transmitters and Calibration Products Relative Humidity Measurement and Calibration Relative Humidity Measurement and Calibration The widest range of humidity products from a pioneer of RH sensor development The best in Relative Humidity technology to optimize your process Whether you

More information

Statistical estimation using confidence intervals

Statistical estimation using confidence intervals 0894PP_ch06 15/3/02 11:02 am Page 135 6 Statistical estimation using confidence intervals In Chapter 2, the concept of the central nature and variability of data and the methods by which these two phenomena

More information

Particle Monitoring Requirements in Pharmaceutical Cleanrooms

Particle Monitoring Requirements in Pharmaceutical Cleanrooms Particle Monitoring Requirements in Pharmaceutical Cleanrooms All drugs must be manufactured in accordance with the current Good Manufacturing Practice (cgmp) regulations. Pharmaceutical manufacturers

More information

Coating Technology: Evaporation Vs Sputtering

Coating Technology: Evaporation Vs Sputtering Satisloh Italy S.r.l. Coating Technology: Evaporation Vs Sputtering Gianni Monaco, PhD R&D project manager, Satisloh Italy 04.04.2016 V1 The aim of this document is to provide basic technical information

More information

Quality by Design Approaches to Analytical Methods -- FDA Perspective. Yubing Tang, Ph.D. FDA/CDER/ONDQA AAPS, Washington DC October 25, 2011

Quality by Design Approaches to Analytical Methods -- FDA Perspective. Yubing Tang, Ph.D. FDA/CDER/ONDQA AAPS, Washington DC October 25, 2011 Quality by Design Approaches to Analytical Methods -- FDA Perspective Yubing Tang, Ph.D. FDA/CDER/ONDQA AAPS, Washington DC October 25, 2011 1 Outline What is Quality by Design (QbD) Role of Analytical

More information

Lifecycle Management of Analytical Procedures; What is it all about? Jane Weitzel Independent Consultant

Lifecycle Management of Analytical Procedures; What is it all about? Jane Weitzel Independent Consultant Lifecycle Management of Analytical Procedures; What is it all about? Jane Weitzel Independent Consultant 2 USP Stimuli Article Lifecycle Management of Analytical Procedures: Method Development, Procedure

More information

INTRODUCING THE NEW. Xcelodose S. New Powder Micro-dosing System. Even Faster Time to First in Man

INTRODUCING THE NEW. Xcelodose S. New Powder Micro-dosing System. Even Faster Time to First in Man INTRODUCING THE NEW Xcelodose S New Powder Micro-dosing System Even Faster Time to First in Man Xcelodose S Benefits Even Faster Time to First in Man with precision, speed and accuracy Fills with precision

More information

1 Quality Assurance and Quality Control Project Plan

1 Quality Assurance and Quality Control Project Plan 1 Quality Assurance and Quality Control Project Plan The purpose of this section is to describe the quality assurance/quality control program that will be used during the system specific field testing

More information

Warrants, Certificates and other products

Warrants, Certificates and other products Interconnection Trading System Warrants, Certificates and other products MARKET MODEL DESCRIPTION January 2015 TABLE OF CONTENTS 1. INTRODUCTION 4 1.1. Background 4 1.2. Institutional market configuration

More information

Environmental Chamber. Stability/ GMP 2000 Series. Stability Testing Chambers

Environmental Chamber. Stability/ GMP 2000 Series. Stability Testing Chambers GMP 2000 Series ISO 9001:2000 21 CFR PART 11 COMPLIANCE SOFTWARE Stability/ Environmental Chamber Stability Testing Chambers Walk-in Test Chambers for Stability Tests on Pharmaceutical Products according

More information

Annex 2 Stability testing of active pharmaceutical ingredients and finished pharmaceutical products

Annex 2 Stability testing of active pharmaceutical ingredients and finished pharmaceutical products World Health Organization WHO Technical Report Series, No. 953, 2009 Annex 2 Stability testing of active pharmaceutical ingredients and finished pharmaceutical products 1. Introduction 1.1 Objectives of

More information

Evaluating System Suitability CE, GC, LC and A/D ChemStation Revisions: A.03.0x- A.08.0x

Evaluating System Suitability CE, GC, LC and A/D ChemStation Revisions: A.03.0x- A.08.0x CE, GC, LC and A/D ChemStation Revisions: A.03.0x- A.08.0x This document is believed to be accurate and up-to-date. However, Agilent Technologies, Inc. cannot assume responsibility for the use of this

More information

FDA Guidance for Industry Update - Process Validation

FDA Guidance for Industry Update - Process Validation FDA Guidance Update: Process Validation: General Principles and Practices White Paper FDA Guidance for Industry Update - Process Validation The changing face of Validation; are IQ, OQ and PQ really dead

More information

TABLET ROUGHNESS INSPECTION WITH 3D PROFILOMETRY

TABLET ROUGHNESS INSPECTION WITH 3D PROFILOMETRY TABLET ROUGHNESS INSPECTION WITH 3D PROFILOMETRY Prepared by Benjamin Mell 6 Morgan, Ste156, Irvine CA 92618 P: 949.461.9292 F: 949.461.9232 nanovea.com Today's standard for tomorrow's materials. 2011

More information

Using the Surface Zeta Potential Cell Kit

Using the Surface Zeta Potential Cell Kit Using the Surface Zeta Potential Cell Kit Note: Malvern has a video that goes through all of these steps in detail. There is a link on the NBTC Zetasizer information page and it is bookmarked on the Zetasizer

More information