THE EFFECTS OF PITCH AND FLUTING ON INSERTION TORQUE AND PULLOUT STRENGTH OF MINISCREW IMPLANTS

Size: px
Start display at page:

Download "THE EFFECTS OF PITCH AND FLUTING ON INSERTION TORQUE AND PULLOUT STRENGTH OF MINISCREW IMPLANTS"

Transcription

1 THE EFFECTS OF PITCH AND FLUTING ON INSERTION TORQUE AND PULLOUT STRENGTH OF MINISCREW IMPLANTS Christine L. Brinley, B.S., D.M.D. An Abstract Presented to the Faculty of the Graduate School of Saint Louis University in Partial Fulfillment of the Requirements for the Degree of Master of Science in Dentistry 2008

2 ABSTRACT Purpose: This study was designed to isolate two design characteristics of miniscrew implants (MSIs), pitch and fluting, and to determine their effects on primary stability. Methods: Maximum placement torque and pullout strength of MSIs specifically fabricated for this experiment were compared to a control MSI using synthetic and cadaver bone models. The first experimental condition compared MSIs with 0.75 mm and a 1.25 mm pitch to a control MSI with a 1.0 mm pitch. The second experimental condition compared MSIs with 3 longitudinal flutes to a control without flutes. A total of 60 MSIs (15 of each design) were placed in the synthetic bone model; a split-mouth cadaver model was used to compare the three experimental designs (15 of each design) against the control MSIs (total of 45 MSIs). Maximum placement torque was measured as MSIs were inserted with intermittent rotation. To test pull-out strength, a vertical force was applied until failure. Results: The synthetic bone model showed that placement torque and pull-out strength were higher for the 0.75 pitch MSIs than the 1.0 mm and 1.25 mm pitch MSIs, but difference were significant (p<.05)only for pull-out strength. The cadaver model showed no significant differences in 1

3 placement torque or pullout strength associated with pitch. In both the synthetic and cadaver bone model, MSIs with flutes had significantly (p<.05) higher placement torque and pullout strength. Spearman correlations between placement torque and pullout strength were statistically significant both in the synthetic (r=.502) and cadaver (r=.504) bone models. Conclusion: Pitch and flutes affect the primary stability of MSIs. Within limits, decreasing MSI pitch increases pullout strength and fluting increases both placement torque and pullout strength. 2

4 THE EFFECTS OF PITCH AND FLUTING ON INSERTION TORQUE AND PULLOUT STRENGTH OF MINISCREW IMPLANTS Christine L. Brinley, B.S., D.M.D. A Thesis Presented to the Faculty of the Graduate School of Saint Louis University in Partial Fulfillment of the Requirements for the Degree of Master of Science in Dentistry 2008

5 COMMITTEE IN CHARGE OF CANDIDACY: Adjunct Professor Peter H. Buschang, Chairperson and Advisor Professor Rolf G. Behrents Associate Professor Sridhar Condoor Assistant Professor Ki Beom Kim i

6 DEDICATION To my best friend and husband, Michal, for his tremendous support and understanding during a very difficult time. It s all for you. ii

7 ACKNOWLEDGEMENTS I would like to acknowledge the following individuals: Dr. Peter Buschang for chairing my thesis committee. Thank you for your guidance, patience, insights and time. You have gone over and beyond what I could have ever expected and I will never forget your kindness. Dr. Rolf Behrents for serving on my committee. I truly appreciate your assistance with the writing of my thesis. Dr. Ki Boem Kim for serving on my committee. I appreciate all of your advice throughout my thesis and orthodontic education. Dr. Sridhar Condoor for serving on my committee. This project would not have been possible without your insight and suggestions. Drs. Cooper and Tolbert for assistance with the cadaver specimens. Karen Fete for help both mental and physical in the cadaver lab. I would have not been able to do it without you there. Joe Tricamo for help with the set up and manufacturing of all devices needed for testing. iii

8 TABLE OF CONTENTS List of Tables...vi List of Figures...vii Chapter 1: Introduction...1 Chapter 2: Review of the Literature Implant Stability...5 Primary Stability...5 Secondary Stability...7 Terminology...8 Design Terminology...10 Biomechanical Terminology...12 Determinants of Primary Stability...14 Placement Torque...16 Pullout Strength...18 Combination Testing...20 Factors Affecting Primary Stability...22 Bone...23 Bone Density...23 Cortical Thickness...25 MSI to Bone Interaction...26 Bone Material Properties...28 Screw-Type Implant Design Character..29 Length...30 Diameter...31 Thread Depth...32 Thread Design...33 Self-Drilling/Self-Tapping...35 Taper...38 Multiple Threads...39 Pitch...40 Flutes...42 Purpose...44 References Cited...45 Chapter 3: Journal Article Abstract...51 Introduction...53 Materials and Methods...56 Miniscrew Implants...56 Experimental Designs...56 Control Design...58 iv

9 Testing Methods...58 Synthetic Bone Model...59 Cadaver Bone Model...59 Mechanical Testing...60 Placement Torque...61 Pullout Strength...61 Statistical Analysis...62 Results...62 Synthetic Bone Model...62 Pitch...62 Fluted...65 Cadaver Bone Model...67 Pitch...67 Fluted...71 Discussion...74 Conclusions...77 References Cited...78 Vita Auctoris...82 v

10 LIST OF TABLES Table 2.1: Design Characteristics of a Sample of Commercially Available Miniscrew Implants...9 Table 2.2: Placement Torque Values of Various Screw-Type Implants...18 Table 3.1: Median, 25 th, and 75 th Quartile for Placement Torque and Pullout Testing of MSIs with Pitch Variation in a Synthetic Model...63 Table 3.2: Median, 25 th, and 75 th Quartile for Placement Torque and Pullout Testing of MSIs with and without Fluting in a Synthetic Model...66 Table 3.3: Torque and Pullout Comparisons of a 0.75 mm and a 1.0 mm Pitch MSI in a Cadaver Model...68 Table 3.4: Torque and Pullout Comparisons of a 1.0 mm and a 1.25 mm Pitch MSI in a Cadaver Model...70 Table 3.5: Mean and Standard Deviation for Placement Torque and Pullout Testing of MSIs with and without Fluting in a Cadaver Model...72 vi

11 LIST OF FIGURES Figure 1: MSI Length...10 Figure 2: Outer and Inner Diameters and Pitch...10 Figure 3: Symmetric Thread Design...11 Figure 4: Asymmetric Thread Design...11 Figure 5: Stress-Strain Curve...13 Figure 6: A mm Pitch Experimental MSI, B. 1.0 mm Pitch Control MSI, C mm Pitch Experimental MSI...57 Figure 7: Fluted MSI...58 Figure 8: Median Placement Torque of 0.75 mm, 1.0 mm (Control), and 1.25 mm Pitch MSIs in Synthetic Bone with Error Bars Representing the 25 th and 75 th Quartiles...64 Figure 9: Median Pullout Strength of 0.75 mm, 1.0 mm (Control), and 1.25 mm Pitch MSIs in Synthetic Bone with Error Bars Representing the 25 th and 75 th Quartiles...65 Figure 10: Median Placement Torque of Fluted and Control MSIs in Synthetic Bone with Error Bars Representing the 25 th and 75 th Quartiles...66 Figure 11: Median Pullout Strength of Fluted and Control MSIs in Synthetic Bone with Error Bars Representing the 25 th and 75 th Quartiles...67 Figure 12: Median Placement Torque of 0.75 mm and 1.0 mm (Control) Pitch MSIs in Cadaver Bone with Error Bars Representing the 25 th and 75 th Quartiles...68 vii

12 Figure 13: Median Pullout Strength of 0.75 mm and 1.0 mm (Control) Pitch MSIs in Cadaver Bone with Error Bars Representing the 25 th and 75 th Quartiles...69 Figure 14: Median Placement Torque of 1.0 mm (Control) and 1.25 mm Pitch MSIs in Cadaver Bone with Error Bars Representing the 25 th and 75 th Quartiles...70 Figure 15: Median Pullout Strength of 1.0 mm (Control) and 1.25 mm Pitch MSIs in Cadaver Bone with Error Bars Representing the 25 th and 75 th Quartiles...71 Figure 16: Median Placement Torque of Fluted and Control MSIs in Cadaver Bone with Error Bars Representing the 25 th and 75 th Quartiles...72 Figure 17: Median Pullout Strength of Fluted and Control MSIs in Cadaver Bone with Error Bars Representing the 25 th and 75 th Quartiles...73 viii

13 CHAPTER 1: INTRODUCTION Newton s 3 rd Law states that every action has an equal and opposite reaction. Orthodontic practice is subject to the effects of this law particularly whenever teeth are used as anchorage for moving other teeth. Unfortunately it is often the case in orthodontics that the movement of some teeth is desired while the reciprocal movement of other teeth is not (i.e., anchorage). A majority of the time this results in a compromise in treatment because absolute anchorage is, for the most part, not possible when using teeth as anchors. Extra oral devices have also been used as anchorage, but they are dependant upon patient compliance, and as a result, anchorage preservation varies and is outside the control of the practitioner. Miniscrew implants have made absolute anchorage a possibility. They allow the application of force to be controlled by the operator and provide an excellent alternative when traditional orthodontic anchorage is impossible, unattainable, or impractical. 1 According to contemporary procedures small titanium miniscrews can be placed at various locations within the oral cavity to serve as anchorage and facilitate controlled tooth movement. The 1

14 benefits that miniscrew implants offer orthodontics have led to a rapid rise in their popularity. Osseous screw-type implants have been studied with regards to both medical and dental applications. The use of bone screws for health-related purposes emerged during the mid-nineteenth century with the work of Cucel and Rigaud. 2,3 The orthodontic literature first mentioned the use of screw type implants in 1945 in the work of Gainsforth and Higley, 4 but clinical use did not gain popularity until recent years. Because of the rather new evolution of miniscrews and miniscrew mechanics, investigators have not yet had time to thoroughly investigate the engineering principles of miniscrew design. Although a fair amount of literature exists in orthopedics and prosthetics, the findings in those areas do not always apply to orthodontics due to the very different applications and force levels involved. Disciplines other than orthodontics use much larger screwtype implants which need to withstand the heavier force loads and fracture produced by chewing. The orthopedic and prosthetic literature has extensively examined bone properties, 5-7 stability testing, 8-10 screw design characteristics, 11,12 insertion torque, 8,12-15 and pullout strength Orthodontic studies have touched upon the effects of miniscrew implant(msi)diameter, length, 20 2

15 placement method, 19,20 insertion torque, 18,21 and pullout strength. 18 However, most of the orthodontic literature consists of case studies and force application research. There has been little attention paid to miniscrew implant design and its effect on stability. Existing miniscrew designs are relatively similar and have remained basically unchanged since their introduction. New designs may lead to more creative anchorage possibilities. For instance, present MSIs must maintain a minimum diameter due to the risk of MSI fracture during torsion. Maximizing stability while minimizing placement torque could allow for smaller MSIs, which could potentially have more clinical uses. Lower placement torque could also minimize osseous trauma and lead to an increased stability. Before such development occurs, however, some basic questions regarding the relationship between MSI design and stability have yet to be answered. For example, which design characteristics influence placement torque and/or stability? Are there design possibilities yet to be introduced that could improve on the basic designs presently available? Is there a relationship between MSI placement torque and primary stability? The purpose of the present study is to provide information about miniscrew implant design that might 3

16 provide a foundation for the development of improved MSIs. This investigation will attempt to isolate one design characteristic, while keeping all other parameters the same, to determine the qualities of any given design characteristic. To date, there have been no carefully crafted studies designed in this manner to evaluate miniscrew implants used for orthodontic purposes. The goal is to systematically evaluate the effects that MSI pitch and fluting have on insertion torque and pullout strength. The following hypotheses will be tested: 1) There are no differences in insertion torque and pullout strength between MSIs with 0.75 mm, 1.00 mm and 1.25 mm pitch. 2) There is no difference in insertion torque and pullout strength between MSIs with and without flutes. 3) There is no relationship between MSI insertion torque and pullout strength. 4

17 CHAPTER 2: REVIEW OF THE LITERATURE IMPLANT STABILITY Implant stability at the time of placement and throughout implant use is critical to success. Implants for various medical and dental uses are anchored to bone by basic mechanisms such as mechanical interlocking between the bone and the implant or soft tissue attachment. Failure of these mechanisms could be due to a number of factors including design, adverse host response, poor placement technique, or excessive loading. 22 Implant stability can be classified as either primary or secondary based upon timing and bone response. Primary stability is most critical during the early phases of miniscrew mechanics, while secondary stability becomes increasingly important as treatment progresses. PRIMARY STABILITY Primary stability refers to the lack of miniscrew movement at the time of initial placement. This type of 5

18 stability is considered critical in orthodontics because it allows for immediate loading. 23 A lack of primary stability frequently leads to miniscrew mobility and subsequent loss. 24 It is a mechanical phenomenon thought to be a function of the local bone quality and quantity, design of the implant, and placement technique. 9 These three factors significantly influence the two major determinants of primary implant stability: bone-to-implant contact and the compressive stresses at the bone to implant interface. 9 Bone-to-implant contact has been demonstrated to be greater in dense cortical bone than in trabecular bone and better stability has been reported for implants placed in dense bone areas than in less dense areas. 9 Also, there is a marked compression of local bone upon implant insertion resulting in circumferential hoop stresses. Hoop stresses are compressive stresses generated in the bone around the implant threads that help to enhance primary stability. However, when hoop stresses are too high they can produce local ischemia and bone necrosis. 9 Alternately, compressive stresses that are too low will offer very little primary stability. For optimal stability, implants should be placed in cortical bone with compressive stresses that are neither too high or too low. 21 6

19 SECONDARY STABILITY Bone regeneration and remodeling following implant placement results in an increase in stability referred to as secondary stability. Primary stability and secondary stability are related in that poor primary stability has a detrimental effect on subsequent secondary stability. Limited areas of contact between the miniscrew and bone at insertion may permit micromovement which could lead to bone resorption and the formation of a fibrous capsule resulting in clinical failure Osseous trauma and overload during miniscrew use or placement can necessitate healing and the formation of woven bone. Woven bone is part of the normal healing process, but it is poorly organized, which makes its strength too low to support loading. 28 To maximize implant retention, care must be taken to minimize trauma at the time of insertion to avoid extensive bone healing and turnover. 7

20 TERMINOLOGY A miniscrew implant can be defined as a device that is temporarily fixed to the bone for the purpose of enhancing orthodontic anchorage. 24 Miniscrew implants are, as the name indicates, miniature screws, however, a universal name has yet to be adopted. They have been referred to as TADs, microscrews, mini-implants, and a variety of other names. Many of the names in use today for orthodontic miniscrew implants do not provide accurate descriptions. The name TAD or temporary anchorage device is too general. Headgears, Nance buttons, and lingual arches, for example, are also temporary anchorage devices. The use of the term microscrew can be misleading because the word micro refers to something that must be viewed with the aid of a microscope. These screws are small, but not microscopic making miniature a more accurate descriptor. In addition, mini-implant is not ideal because it does not indicate design characteristics of the screw. In this discussion they will be referred to as miniscrew implants or MSIs because they are small, implanted screws. The miniscrews on the market today vary in terms of their length, diameter, thread depth, thread design, self- 8

21 drilling/tapping characteristics, pitch, taper, head design, and material. Some of the design characteristics of a few of the more commonly used miniscrews are listed in Table 2.1. Table 2.1: Design characteristics of a sample of commercially available miniscrew implants Manufacturer/ Name Dentos AbsoAnchor IMTEC Ortho Implant Dentaurum TOMAS pin HDC Company Spider Screw Mondeal LOMAS System Length Diameter (mm) (mm) 6,7,8, Major:1.2-10, Minor: NS 6,8,10 Major:1.8 Minor:1.6 8,10 Major:1.6 Minor:1.2 6,8,10 Major:1.5 Minor:NS 7,9,11 Major:2.0 Minor:NS 7,9,11 Major: , 2.3 Minor: NS Thread Depth Thread Design Self- Drilling? NS* Symmetric Yes 0.2 Buttress Yes 0.9 Symmetric No NS* Buttress No NS* NS* Yes * Many design characteristics are not specified (NS) by the manufacturer for reasons unknown. Because miniscrew implant terminology used to describe design characteristics can be confusing, it is best to clarify some terms used for MSI design. Biomechanical terms used to evaluate MSI stability also need to be defined in order to understand the results of the available studies. 9

22 DESIGN TERMINOLOGY The length of a miniscrew implant is defined as the length of the threaded body and not the length of the entire screw (Figure 1). MSI lengths range from 5.0 to 12.0 mm. 24 The major and minor diameters should always be Figure 1: MSI Length= the Threaded Body Figure 2: Outer & Inner Diameters and Pitch specified when describing MSIs (Figure 2). The major diameter is the maximum diameter determined by the outer diameter of the threads. Major diameters of MSIs range from 1.0 to 2.5 mm. 24 The minor diameter refers to the inner (or core) diameter and it ranges from 0.2 to 1.6 mm. The thread depth (Figure 2) is half the difference between the major and minor diameter. The MSI threads form angles with the core. The leading angle is closest to the miniscrew tip and comes into bone contact first upon insertion, while the trailing angle is closest to the MSI head and is the last to contact during placement. Thread design is often defined 10

23 as asymmetric or symmetric. A symmetric thread design is V- shaped, with leading and trailing angles that are congruent (Figure 3). An asymmetric, or buttress, thread is one with a leading angle (towards the tip) that is at 45 degrees to the long axis of the shaft and a trailing angle (towards the head) that is 90 degrees to the long axis of the shaft 24 (Figure 4). The pitch is the distance between the threads (Figure 2). When the threads are spaced far apart the miniscrew has a high pitch; conversely, when the threads are close together the miniscrew is low-pitched. Figure 3: Symmetric Thread Design Figure 4: Assymmetric Thread Design A miniscrew implant can be self-drilling, selftapping, both, or neither. A self-drilling MSI is one that does not require a pilot hole and has either a sharp, tapered apex to allow placement, 24 or a notch in the tip to drill through the cortex. All MSIs are self-tapping and able to create their own thread as they advance. There are 11

24 two different types of self-tapping designs: thread-forming and thread-cutting. The thread-forming design compresses the bone around the thread as the miniscrew advances. The thread-cutting design has either a notch at the tip parallel to the long axis or a sharpened thread that actually cuts threads into the bone as the miniscrew is inserted. 24 Biomechanical terminology used to describe stability testing can be as difficult to understand as design terminology. The literature is not consistent in the terms used and many terms are synonymous. BIOMECHANICAL TERMINOLOGY Engineering divides stresses into two categories; normal and shear. Normal stresses can be either tensile or compressive and are always directed perpendicular to a plane. Tensile stress is generated when a force acts away from the plane and compressive stress involves the direction of force pushing towards the plane. Shear stress is different because it is directed parallel to the plane of reference. 29 The effect that a stress has on a material is typically represented with a stress-strain curve (Figure 12

25 5). The linear portion of the curve can be described as the modulus of elasticity. It measures the strength of the material, its stiffness, and resistance to load. The modulus of elasticity of a material represents the amount of stress that can be withstood without permanent deformation. With increasing stress the yield point of a material is eventually reached. The yield point is the point on the stress-strain curve where strain continues to increase without an increase in the stress indicating material deformation. 29,30 Strain Figure 5: Stress-Strain Curve To understand screw mechanics, it is also important to define cutting resistance, stripping torque, and holding strength. Cutting resistance is the resistance of the miniscrew implant to placement which is influenced by bone 13

26 density, 10,31-33 compression, 9 and friction. 34 Stripping torque can be described as the point at which the MSI freely rotates in the bone with no further increase in torque. 35 The holding strength of a MSI is its resistance to displacement (or removal). DETERMINANTS OF PRIMARY STABILITY Various techniques have been used to test the primary stability of endosseous implants. In living animals, tests of primary stability must be performed immediately after placement to assure that no healing or bony adaptations have taken place; it can be tested at any time in nonviable tissue. 36 Implant stability depends on the stiffness of the implant, the implant-tissue interface, and the surrounding tissues. 9 Some methods used to test primary stability are more destructive than others. Non-invasive tests include radiographs, percussion, resonance frequency, placement torque, and reverse removal torque. 9 Some of these noninvasive methods can be unreliable due to operator sensitivity and error. 9 Radiographs are problematic because they are a two-dimensional and not easily standardized. Percussion is also not reliable because the test results 14

27 can differ depending upon the vertical position and angle of the testing instrument as it strikes the implant. In addition, resonance frequency tests are not very accurate because the ear is not sensitive enough to discriminate such frequencies. 9 Although reverse torque is typically classified as a non-invasive test, arguments can be made that such a test is in fact quite invasive because damage can occur at the bone-to-implant interface. 9 While noninvasive methods for testing can be somewhat unreliable, they are the most ethical and practical in clinical situations. Of the tests to assess primary stability, placement torque and resonance frequency seem to be the least contraindicated in clinical situations due to their innocuous nature. 34 Invasive tests of primary stability are typically used for research purposes. Histologic and histomorphometric techniques are available to assess the implant-tissue interface using light or electron microscopy. 9 Histologic techniques evaluate the microscopic nature of the bone surrounding the implant while histomorphometric techniques quantify the percentage of bone-to-implant contact. Both of the aforementioned techniques are problematic with regard to accuracy because of difficulties in specimen preparation due to the marked difference in the physical properties of 15

28 the implant and the bone. 9 Preparation techniques that allow for the visualization of the metal implant can be destructive to the surrounding bone. Removal torque is another invasive test which measures the amount of force necessary to remove an implant or disrupt the bone to implant interface. Removal torque is a measurement of interfacial strength in shear and is confounded by the geometry of the implant. 9 Pullout testing is an additional invasive test used to measure the pullout resistance of an implant. It involves complete removal of the implant by disrupting the bone to implant interface. It is a common test to measure the influence of implant design on purchase strength in porous materials. 36,37 PLACEMENT TORQUE Torque can be defined as an angular moment of force 18 or, for the purpose of this discussion, the rotational force required to insert a screw into bone. 14,38 A miniscrew implant is designed to convert torque into compressive force between it and the object it engages (such as the MSI and the bone). The MSI thread acts like an inclined plane and produces an axial force as it is placed. 3 The torque of 16

29 insertion upon miniscrew implant placement is dependant upon both the design of the MSI and bone properties. Many studies in various disciplines have found peak insertion torque to be a significant factor in predicting the holding strength of screw-type implants. 5,15,39,40 Motoyoshi et al. tested 124 orthodontic miniscrew implants and found the proper miniscrew placement torque to be a matter of balance between force levels that were neither too high nor too low. 21 High placement torque has been found to be related to better initial stability; 5 however, too high a placement torque can cause osseous damage and MSI loss. 41,42 Some have found low placement torque levels to be ideal; 19 although, torque levels too low are also undesirable because a certain level of torsion must be reached to achieve primary stability. 9,13 Different ranges of ideal placement torque levels have been reported in the literature, but they vary substantially possibly due to the wide variations in implant design, method of placement, and type of bone in which they were placed. Table 2.2 provides placement torque in N/cm reported in the literature from a variety of disciplines. 17

30 Table 2.2: Placement torque values of various screwtype implants Study Discipline Pilot Hole Range (N/cm) Friberg et al 13 Prosthodontics Yes 5-19 (maxilla) 5-25 (mandible) Motoyoshi et al 21 Phillips et al 42 Orthodontics Yes Maxillofacial Yes 40 (average) Surgery No 52 (average) O Sullivan Prosthodontics Yes (Type 2 bone)* et al (Type 3 bone)* 4-20 (Type 4 bone)* *Bone type is based upon bone quality and density. -Type 1- dense with limited vascularity -Type 2- better vascularity than Type 1 -Types 3 & 4- soft textures with Type 4 being the least successful PULLOUT STRENGTH Pullout is a popular test of holding power commonly used in orthopedics, orthodontics, neurosurgery, and maxillofacial surgery. 36 A pullout test directed vertically, with forces parallel to the long-axis of a screw, is one that tests the primary force a screw is designed to resist. Pullout tests are used to evaluate the design of a screwtype implant and the impact it has on the surrounding 18

31 bone. 37 In contrast, shear pullout loads, oriented directions other than parallel to the long axis of the MSI, test the diameter and material properties of a screw and have very little to do with design. 43 When pullout tests are performed either immediately after placement or in nonviable tissue no adaptive healing responses can occur; so it is a test of primary stability. Pullout tests provide an indication of anchorage values for immediately loaded miniscrew implants. 36 Most information about pullout testing in human bone comes from the orthopedic and neurosurgical literature and involves the vertebrae. 15,37,39,44,45 Although there is information about pullout testing in dog and synthetic bone models, pullout testing of MSIs in a human model has yet to be reported in orthodontics. Although providing useful clinical information, human models add variation to testing because of the differences in density and cortical thickness not only among samples, but within the same sample. 7,11,46 An argument can be made that pullout testing is more applicable to orthodontics because orthodontic forces are applied directly to the screw head, while in the other disciplines the implants are used to stabilize fractures and are not directly loaded. 36 Orthodontic research has 19

32 used pullout testing to evaluate the stability of miniscrew implants in dog and synthetic models. Huja et al. performed pullout tests on 56 MSIs placed in dog mandibles and found a positive correlation between cortical bone thickness and resistance to vertical displacement. 36 They reported the peak load at failure to be N in the anterior and N in the posterior regions. In addition, Carano et al. tested three different MSI systems and found a positive correlation between mechanical properties and pullout resistance. Peak failure loads ranged between 280 N and 370 N, depending upon the design. 18 COMBINATION TESTING At times, invasive and non-invasive methods of primary stability testing are combined to make an overall, two-fold assessment of stability. Combining measures of placement torque and pullout strength would be one such combination. The paring of these tests is important for analyzing the characteristics of the screw thread as a function of its insertion and stablilty. 18 Insertion torque has been found to correlate linearly with pull-out strength 39,47 and the strength of screw fixation is mainly determined by both of these parameters. 44,47 However, some research has 20

33 hypothesized that osseous damage accumulates around the MSI with high levels of torsion upon insertion, resulting in a decreased pullout strength. 45,48 This would suggest a parabolic effect in which desirable placement torque levels should be neither too low or too high. Low pullout strength has been attributed to placement torque that is above or below a certain range 21 and it has been shown that up to 76% of the variability in pullout force can be attributed to placement torque. 15 In other words, placement torque and pullout strength go hand in hand. The combination of placement torque and pullout strength testing of the same MSI have yet to be performed in orthodontics; however such combination testing is fairly common in other disciplines. High levels of placement torque do not always result in failure in the bone, but it can also lead to implant bending or fracture. 42 This is an increased risk when the insertion is impeded by something such as a tooth root, so care must be taken to assess the integrity of the implant when an object other then the bone is contacted. 38 If a root is contacted, one would expect a noticeable rise in placement resistance. 21

34 FACTORS AFFECTING PRIMARY STABILITY Primary stability is affected by the qualities of both the bone and miniscrew implant. The holding power of a miniscrew implant is not only determined by its design but also by the trauma-induced changes in the bone that result from insertion. 17 Failure can occur within the bone, at the bone-msi interface, or within the MSI itself; it is a function of the weakest link in the system. The most common mode of failure is shearing within the bone threads that are formed around the screw; 17,33,38 this causes the bone to strip and the screw to fail. 18 Because the influence of miniscrew implant design on primary stability is often obscured by research about bone quality and placement technique, 37 a method of assessing the influence of changes in geometry of the implant system is desirable. 34 However, miniscrew implant design and bone quality are not mutually exclusive because MSI design is most critical when bone quality is not optimal. 5 The following section examines the influence of bone and MSI design characteristics on primary stability. 22

35 BONE Bone plays an important role in miniscrew implant primary stability. Bone density, cortical thickness, MSI to bone interaction, and bone material properties have all been found to be correlated with placement torque and pullout resistance. Bone Density Bone density has been shown to have an influence on primary stability. The present knowledge that relates bone density properties to implant stability comes from the prosthodontic and orthopedic literature. Both disciplines have found bone density to significantly correlate with both torque and pullout strength. 15,44,45 The correlation has been found to be reliable enough for equations and classification systems to be developed based upon the relationship between torque and pullout strength to bone density. For example, Ryken et al. 15 developed an equation using bone mineral density and torque to predict pullout strength. In addition, classification systems to grade bone quality 49,50 have been based upon the significant correlation between implant placement torque and bone density. 10,

36 Johansson and Strid developed a classification system widely used in prosthetic dental situations that grades the bone by the resistance of implant placement. 50 Dense bone correlates with high placement torque values 10,13,31-33,50 and most screw-type implant failures, or loss, have been found to occur when torque values were high in medium to high density bone. 13 Bone mineral density also positively relates to pullout strength 15,33,44,45 and variation in mean pullout strength has been attributed to anatomical variation in density. 44 However, bone density has only been found to account for 28.3% of the variability in pullout force; in contrast, it accounts for 76% of the variablilty in placement torque. The literature is contradictory because it has been shown that an increase in bone density leads to an increase in both placement torque and pullout strength, while high levels of torsion at placement has been shown to decrease pullout strength. A plethora of variables, including the techniques used, bone and MSI characteristics, etc., could account for this discrepancy. Understanding bone density variability can be useful in the treatment planning of miniscrew implants. Differences in bone density have been noted among different orthodontic classifications 51 and greater density has been recorded in specimens from older patients. 10 Bone density 24

37 variation has also been noted between the maxilla and mandible 13,21 and at different sites within the same jaw. 7,10,13,52,53 However, a difference in density has not been found when comparing dentate and edentulous mandibles or genders. 7 The idea of taking bone density into account when planning for implants is not a new concept because prosthetic implant selection protocol involves an assessment of the bone to determine the type of implant and placement technique. 10 In bone with low density an implant with a greater surface area is preferred. 33 This could be applied to orthodontic situations when considering the proper placement site and corresponding MSI. Cortical Thickness The relationship between cortical thickness and placement torque has not been established, but a significant correlation has been found between cortical thickness and pullout strength. 36,48 Cortical bone thickness, like density, varies depending upon location. Values for pullout strength could be influenced, at least partially, by cortical thickness. Pullout strength has been found to be significantly different between the maxilla and the mandible and even 25

38 when comparing anterior and posterior locations within the same jaw. 11,36 The bone located in the anterior region of the maxilla and mandible(excluding the mental protuberance) typically has the thinnest cortical bone and the lowest pullout strength; while the bone in the posterior has thicker cortical bone and higher pullout strengths. 36 Cortical bone thickness should always be taken into account when analyzing the cause of MSI failure. MSI to Bone Interaction at Placement Miniscrew implant primary stability has to do with its interaction with the bone at the time of placement. 32 As a miniscrew is inserted and meets resistance, a shear stress is developed along the length of the screw. 14 This shear force can cause microfractures within the bone which can effect the healing process 42,48,54 and lead to degeneration and necrosis at the interface. 9,17,54 The bone damaged during insertion will be targeted for repair and could result in screw loosening. 55 Cutting resistance of the bone is a function of the friction at the bone to screw interface 34 and is increased as bone material packs around the miniscrew implant. 13 This, in combination with the increased screw surface area 26

39 interacting with the bone, explains why placement torque increases as the MSI is inserted. Friction increases the torque required to place a MSI. During placement, applied torque is used to cut threads and to overcome friction between screw threads and the bone. 3 Tension can only be induced into the system when the torque needed to overcome friction is exceeded. 14 Only about 15% of the torque applied to a MSI is converted into axial force; the rest is lost due to factors such as friction between the bone and the MSI. 3 A reduction in friction could help reduce the torque at placement and thus limit osseous damage. To reduce friction, it has been recommended to wet the MSI and insert it with continuous rotation. Wetting with blood or saline prior to insertion produces an effect similar to oil in a crankshift. 35,56 In addition, a screw-type implant that is inserted with continuous rotation generates less torque than one that is inserted intermittently. A pause in rotation allows the bone to relax around the implant, resulting in more friction to be overcome as rotation resumes and the generation of more torque. 35,56 27

40 Bone Material Properties and Response Bone material properties play an important role in the primary stability of miniscrew implants. The holding power of a screw-type implant is related to the shearing strength of the material into which it is inserted. 14 Pullout is a test of shear strength. 43 The MSI s interaction with the bone upon insertion can also compromise its holding potential because the strength and integrity of the bone surrounding the screw are partially a function of the trauma caused during insertion. 17 As mentioned earlier, the interdigitation of miniscrew implants and bone offer two potential sources for failure: either in the threads of the bone or the MSI threads. 38 Failures tend to occur due to the bending of bone threads interdigitated between the MSI threads. Osseous failure occurs parallel to the long axis of the MSI core in a plane of natural cleavage(i.e., the plane that is parallel to the long axis at the extent of the major diameter). 17 This failure is related to the strength of the bone for a distance of about two millimeters around the screw. 17 The reasons for failure of miniscrew implants can be identified by examining the bone following pullout testing. Bone damage during placement will theoretically shear in a 28

41 line parallel to the outer edge of the MSI threads, causing a cylindrical defect upon pullout. 12 When the MSI failure actually happens at the time of pull-out testing, a conical piece of bone strips. 3 Examination of the defect after testing provides valuable information about the cause of failure and can aid in the determination of further research projects. SCREW-TYPE IMPLANT DESIGN CHARACTERISTICS There are a variety of screw-type implants available for medical use, but their designs have not been fully studied. 56 The limited amount of literature available on design mostly comes from disciplines other than orthodontics. Implant design has been shown to influence primary stability; different designs have produced different results. 34 Geometry and material are significant factors in miniscrew implant biomechanics and their alteration could change the MSI s mechanical strength. 18 Various studies have found that characteristics such as length, diameter, thread depth, thread design, and pitch to be important in determining holding power. 11,43,45,56 However, available MSIs differ in many ways, including geometry, 29

42 material, dimension, and insertion technique (such as the presence or absence of a pilot hole). 17 In the next sections, characteristics of miniscrew implants will be examined individually. Some of these characteristics have been more fully investigated than others. Determining the effects of each characteristic and its role in the overall design will aid in the planning of new designs. New designs could result in a decrease of trauma and an increase in treatment planning options. Length The orthodontic literature does not contain information on the effect of miniscrew implant length on primary stability. However, the orthopedic literature shows that implant length is one of the most important variables in mechanical strength. 18 The relationship between length and placement torque has not been investigated, but a correlation with pullout strength has been shown. Hitchon et al. tested 201 screw-type implants in frozen human cadaver specimens and found length to have a significant effect on pullout strength, with longer screws having a higher resistance to displacement. 45 This might be expected 30

43 because holding power is directly proportional to the amount of thread engagement. 54 Diameter As with length, the relationship between miniscrew implant diameter and primary stability has yet to be examined in orthodontics. Other disciplines have found diameter to be the one of the most important variables for mechanical strength. 18 It has influence on both the torque and pullout strength. Using fresh bovine bone, Nunamaker and Perren tested placement torque of screw-type implants of various diameters and found insertion torque to increase with an increase in minor diameter. 57 The increase in surface area that results from increasing the screw minor diameter creates more friction and increases torque. 3 The minor diameter of a screw-type implant can become too small and higher torque levels can cause breakage; this is referred to as breaking torque. The breaking torque of a screw is equal to the ratio of the cube of its respective minor diameter. 57 Carano et al. suggested that a minor diameter reduction of as little as 0.2 mm may reduce the resistance to breakage of the miniscrew implants by 50%. An overall 31

44 minor diameter of less than 1.5 mm is not recommended for orthodontic purposes because humans can apply enough torsional force to break smaller screws. 18 However, if placement torque could be reduced through the addition of other design features, it is theoretically possible to further reduce screw size. While placement torque is correlated with minor diameter, resistance to pullout correlates with the major diameter. Major diameter has been found to be one of the main factors in primary stability. 17,58 DeCoster et al. isolated screw-type implant characteristics and found major diameter to increase linearly with resistance to pullout in a synthetic bone model. 11 It is one of the most important dimensions for pullout strength because the tensile resistance of the screw-type implant corresponds to the cube of its major diameter. 58 Thread Depth While thread depth has yet to be considered in the orthodontic literature, it is fairly commonplace in orthopedics. It can be described as half the difference between the major and minor diameters. Shallow threads are typically used for cortical bone and deep threads are used 32

45 for cancellous bone. 3 Greater depth is thought to provide more holding power because of an increased bone volume between the threads and increased bone to screw contact. 12 The relationship between thread depth and placement torque is unknown, but information about the effect of thread depth on pullout is available. Chapman et al. tested the pullout strength of 12 different screw-type implant designs and found that an increased thread depth allows for greater purchase strength and pullout resistance in porous materials. 11,43 DeCoster et al. performed pullout on six different screw-type implants and came to the same conclusion. 11 Based upon these findings, a greater thread depth would be recommended for orthodontic applications to maximize resistance to displacement under force loads. Thread Design The thread is designed to facilitate placement, prevent loosening, offer strength, and withstand axial loads. 3 Threads are most commonly designed in two ways: symmetric or asymmetric. The literature provides different hypotheses as to why one design should be used over another. Perren et al. postulated that symmetric threads produce increased friction at placement, but greater 33

46 pullout strength. The increase in pullout strength is due to an increase in the amount of bone at the outer edge of the thread in a symmetric design since the height of the screw threads reaches zero at the outer edge. 3 They also stated that the symmetric thread offers mechanical advantages by contributing to the stiffness of the screw. 3 With regard to the asymmetric thread design, Perren et al. hypothesized that the asymmetric design offers the advantages of the symmetrical design without the disadvantages by having a contacting flank that is perpendicular to the long axis to resist the force of displacement. 3 In addition, Carano et al. found that the asymmetric threads have a geometry that facilitates insertion while obstructing removal. 18 Orthodontic miniscrew implants use both designs, but the literature supporting one design over another is lacking. However, it was hypothesized by Carano et al. that a combination of insertion and pullout tests are important for the analysis of thread design as a function of screw stability. 18 They also found that MSI resistance to torsional stress during placement is related to the design of the screw thread. 18 More testing is needed to determine which design is best for orthodontic purposes. 34

47 Self-Drilling/Self-Tapping Research pertaining to the effects of self-drilling and self-tapping characteristics on primary stability of screw-type implants comes largely from the orthopedic literature. Although self-drilling and self-tapping miniscrew implants are commonly used, research in orthodontics has yet to compare the differences between MSIs requiring a pilot hole and those that are selftapping. The relatively small size of orthodontic MSIs gives them the advantage of more than one option in tip design. In orthodontics there is confusion concerning the definitions of self-drilling or self-tapping designs, possibly due to the variety of designs available. Little has been scientifically proven regarding the effects that self-drilling characteristics have on primary stability, but educated hypotheses have been offered. Selfdrilling screw-type implants are thought to have greater bone contact than those requiring a pilot hole 36 Greater bone contact should positively effect stability. Selfdrilling screws are also thought to have a higher placement torque due to friction which could prove detrimental. In addition, Perren et al. found that a self-drilling tip must be able to cut a pilot hole at the same rate at which the 35

48 thread advances the screw into the bone. For example, an orthopedic screw with a pitch of 1.75 mm and a 4.5 mm diameter cannot be matched by the drilling effect of any type of tip because the pitch will not allow for efficient cutting and advancement with the given diameter. 3 A balance in design is required to prevent the development of undue stress in the bone during insertion. Self-tapping is the ability of a miniscrew to create its own thread as it advances. The literature pertaining to the effects of self-tapping screws on placement torque and initial stability is controversial. Phillips et al. and Al- Nawas et al. have shown that self-tapping implants have a higher placement torque and greater initial stability when compared with non-self-tapping. 8,42 The higher value of placement torque with the self-tapping screw was thought to be due to the excess torque needed to overcome the friction of cutting threads. 42 In contrast, Ansell and Scales have shown less torque during the insertion of self-tapping screws; this could be a result of the screw-type implants being inserted with intermittent rotation. 56 The method of placement (i.e. continuous versus intermittent rotation) should be considered when deciding whether or not to use a self-tapping design. Ansell et al. found high placement torque levels while placing non-self- 36

49 tapping screws with intermittent rotation. 56 Non-selftapping screws have to overcome more friction than screws which are self-tapping because the threads are not cutting. A pause in movement allows the bone to relax around the non-self-tapping thread, which increases friction and placement torque. 42 Some of the discrepancy in the literature with regard to the effects of self-tapping characteristics on primary stability could be as a result of inconsistent or flawed designs. Self-tapping screw-type implants can be either thread forming or thread cutting. The typical threadcutting design available in orthodontics has a cutting flute which extends only partially along the length of the tapered portion of the miniscrew implant. This does not allow for the cutting of threads beyond the point of flute extension and, as the major diameter continues to increase, sufficient space has not been tapped to allow for the insertion of the remaining threaded portion. Perren et al. demonstrated that thread cutting self-tapping implants should have flutes that continue until the cylindrical portion of the implant is reached. 3 They also suggested that high torque at insertion, reduced holding strength, and a rise in temperature at insertion can be fairly well controlled by proper design of the screw tip. 3 37

50 There are many conflicting reports regarding the placement torque and pullout strength of self-tapping and non-self-tapping screw implants. Some of the disparity may be due to the failure to compare the differences between thread-forming and thread-cutting self-tapping screws. More research should be done to better understand these characteristics of screw design. Taper The taper of screw-type implants ranges from completely cylindrical to entirely tapered. Most orthodontic miniscrew implants are at least partially tapered, with many having a combined design that starts tapered until the full diameter is reached and then progresses on cylindrically. Although Carano et al. were not specifically studying taper during testing, they noted that cylindrical screws have a constant insertion torque in any given material as long as the pitch is held constant. 18 This also holds true for the cylindrical portion of a combined design. In comparison, they also found that the placement torque of tapered screws increase throughout placement. 18 The effects of MSI taper have yet to be fully studied. 38

51 Multiple Threads A screw-type implant with a multiple thread design is one which has a series of threads wrapped around the core. A common example of such a design is a water bottle or plastic milk container. It was first described by Bechtol et al. 59 in 1959 and then later introduced into dentistry by Nobel Biocare as the Mark IV prosthetic implant. 5 Multiple-threaded designs are not common in either medicine or dentistry. In addition, there is no mention of this design characteristic in the orthodontic literature. The existing knowledge of the effects of a multiplethreaded design is limited, but there has been mention of its influence on both placement torque and pullout strength. O Sullivan et al. found the double-threaded Mark IV system to have a higher insertion torque than those with a single thread, but greater stability. They hypothesized that the double thread allowed faster placement without increasing the energy at the bone-to-implant interface. 5 In support, da Cunha et al. found a double-threaded design to reduce placement time by 50% without an increase in heat production. 34 Boyle et al., who included a multiple-threaded design in a study of surgical emergency screws, found those 39

52 with at least three threads have the largest pullout resistance, and thus greatest stability. 35 It has been suggested that the stability of the multiple-threaded design is less affected by the quality of the bone. 5 This is important because many studies have reported bone quality is correlated with placement torque and pullout strength. 10,14,15 A design that is less affected by parameters outside the control of the clinician, such as bone quality, is preferable. Pitch Pitch is another characteristic of miniscrew implant design that has yet to be studied in orthodontics. Like many of the characteristics previously mentioned, most of the information available comes from the orthopedic literature. The important role of thread pitch has been established. It has been suggested that pitch is clinically more important than major diameter because the major diameter is limited by the bone site and space available. 11 Pitch is not limited by available space and its effect on primary stability has been noted. The effect of pitch on placement torque has not yet been studied. The pitch of a screw-type implant determines 40

53 the amount a screw advances with each turn, 3 which should, in theory, have an effect on placement torque. A greater amount of advancement per turn creates higher axial force which may well increase torsion at placement. However, available studies indicate that pitch is negatively related to pullout strength. Chapman et al. tested the pullout strength of 12 different screw designs and found that decreases in pitch increase the resistance to pullout. 43 DeCoster et al., who performed pullout testing on three different designs, also found that a decrease in pitch significantly increases the pullout resistance. 11 A decrease in pitch is thought to increase the screw purchase strength in porous materials. 43 Since the significance of the bone-to-miniscrew implant contact has been established; the thickness of the bone in which a screw is to be inserted must be taken into consideration when choosing a pitch. In bone that is only 1-2 mm thick, a screw with a pitch angle of mm will only have 2-4 threads in contact with cortical bone. This limited area of contact may produce micromovements leading to bone resorption and loss of the implant

54 Flutes A flute is a longitudinal groove in the threaded portion of a screw used for cutting and clearing bone chips. Flutes vary in length and number and are a common feature in self-tapping screw-type implants. The variation in the length and number of flutes has a significant effect on the insertion torque and pullout strength. The effect of flutes has yet to be studied by orthodontics, but literature is available in the disciplines of surgery and biomechanics. Cutting flutes have a significant influence on the placement torque of screw-type implants. Yerby et al. found that as the number and length of cutting flutes increases, there is a decrease in placement torque and cortical damage. 47 Higher torque levels have been found for screwtype implants with fewer or shorter cutting flutes because the design allows for less clearance of bone chips which tend to accumulate around the screw threads and provide resistance to insertion. 12,47,60 This would also explain why fluted designs have been found to be easier to insert in comparison to non-fluted designs. 47,59,61,62 The study by Yerby et al. recommends a minimum of three, full-length flutes to avoid stripping threads and facilitate advancement

55 The effect flutes have on pullout strength is not clear. It has been reported that the presence of flutes both decreases 47,59,61 and increases 35,62 pullout strength. Arguments can be made to support both notions depending upon the timing of pullout testing. A decreased resistance to pullout should be seen in primary stability testing because the fluted region of a screw does not have as great a holding power as the fully threaded region. 33,59 An increased resistance might be expected with secondary stability because of the possibility of bone forming in the fluted region. The temporary nature of orthodontic miniscrew implants could be problematic for the fluted design. This is because bone may grow into the flutes and cause problems with removal. 3,56 Problems may also arise as the bone chips fill the flutes during placement, which could increase the chances of splitting the bone. 35 This can be alleviated by placing grooves in the shank, or core, of the screw implant thereby creating a space for the bone chips to collect. 35 The aforementioned is not problematic for larger screw-type implants which are used in other disciplines, but grooves may compromise the strength of miniscrew implants because of their small dimensions. 43

56 PURPOSE The purpose of the present study is to isolate several characteristics of miniscrew implant design in order to determine their effect on primary stability. One design characteristic will be altered at a time and placement torque and pullout tests will be performed in comparison to a control to assess the characteristic s influence on MSI stability. Such an experiment has never been performed and could lead to better MSI designs. Orthodontic miniscrews are often loaded immediately 23 making this experiment clinically applicable. Even though pullout tests do not have direct clinical involvement because orthodontic forces are not sufficient to strip miniscrew implants from their sites, 18 they are useful in analyzing MSI geometry as a function of stability

57 REFERENCES CITED 1. Cheng S, Tseng I, Lee J, Kok S. A prospective study of the risk factors associated with failure of mini-implants used for orthodontic anchorage. Int J Oral Maxillofac Implants 2004;19: Kakria HL. Evolution in fracture management. MJAFI 2005;61: Perren SM, Cordey J, Baumgart F, Rahn BA, Schatzer J. Technical and biomechanical aspects of screws used for bone surgery. Int J Orthoped Trauma 1992;2: Gainsforth BL, Higley LB. A study of orthodontic anchorage possibilities in basal bone. Am J Orthod Oral Surg 1945;31: O'Sullivan D, Sennerby L, Meredith N. Measurements comparing the initial stability of five designs of dental implants: a human cadaver study. Clin Implant Dent Relat Res 2000;2: Fanuscu MI, Chang TL. Three-dimensional morphometric analysis of human cadaver bone: microstructural data from maxilla and mandible. Clin Oral Impl Res 2004;15: Schwartz-Dabney CL, Dechow PC. Edentulation alters material properties of cortical bone in the human mandible. J Dent Res 2002;81: Al-Nawas B, Wagner W, Grotz KA. Insertion torque and resonance frequency analysis of dental implant systems in an animal model with loaded implants. Int J Oral Maxillofac Implants 2006;21: Meredith N. Assesment of implant stability as a prognastic determinant. Int J Prost 1998;II: Turkyilmaz I, Tozum TF, Tumer C, Ozbek EN. Assessment of correlation between computerized tomography values of the bone, and maximum torque and resonance frequency values at dental implant placement. J Oral Rehab 2006;33:

58 11. DeCoster TA, Heetderks DB, Downey DJ, Ferries JS, Jones W. Optimizing bone screw pullout force. J Orthoped Trauma 1990;4: Johnson NL, Galuppo LD, Stover SM, Taylor KT. An in vitro biomechanical comparison of the insertion variables and pullout mechanical properties of AO 6.5-mm standard cancellous and 7.3-mm self-tapping, cannulated bone screws in foal femoral bone. Vet Surg 2004;33: Friberg B, Sennerby L, Grondahl K, Bergstrom C, Back T, Lekholm U. On cutting torque measurements during implant placement: a three-year clinical prospective study Clin Implant Dent Rel Res 1999;I: Hughes AN, Jordan BA. The mechanical properties of surgical bone screws and some aspects of insertion practice. Injury 1972;4: Ryken TC, Clausen JD, Traynelis VC, Goel VK. Biomechanical analysis of bone mineral density, insertion technique, screw torque, and holding strength of anterior cervical plate screws. J Neurosurg 1995;83: Pitzen TF, Barbier F, Steudel WI. Insertion torque and pullout force of rescue screws for anterior cervical plate fixation in a fatigued initial pilot hole. J Neurosurg: Spine 2004;1: Schatkzer J, Sanderson R, Murnaghan JP. The holding power of orthopedic screws in vivo. Clin Orthoped Rel Res 1975: Carano A, Lonardo P, Velo S, Incorvati C. Mechanical properties of three different commercially available miniscrews for skeletal anchorage. Prog Orthod 2005;6: Kim J, Ahn S, Chang Y. Histomorphometric and mechanical analysis of the drill-free screw as orthodontic anchorage. Am J Orthod Dentofacial Orthop 2005;128: Kravitz ND, Kusnoto B. Risks and complications of orthodontic miniscrews. Am J Orthod Dentofacial Orthop 2007;131:S43-S51. 46

59 21. Motoyoshi M, Hirabayashi M, Uemura M, Shimizu N. Recommended placement torque when tightening an orthodontic mini-implant. Clin Oral Impl Res 2006;17: Brunski JB. Biomechanical factors affecting the bonedental implant interface. Clin Mater 1992;10: Chen YJ, Chen YH, Lin LD, Yao CC. Removal torque of miniscrews used for orthodontic anchorage- a preliminary report. Int J Oral & Maxillofac Implants 2006;21: Cope JB. OrthoTADs: The Clinical Guide and Atlas. Dallas: Under Dog Media; Heller JG, Bradley T, Estes MS, Diop A. Biomechanical study of screws in the lateral masses: Variables affecting pull-out resistance J Bone Joint Surg [Am] 1996;78: Carter DR, Giori NJ. Effect of mechanical stress on tissue differentiation in the bony implant bed. In: Davies JE, editor. The bone-biomaterial interface. Toronto: University of Toronto Press; 1991: p Goodman S, Wang JS, Aspenberg P. Difference in bone ingrowth after one versus two daily episodes of micromotion: experiments with titianium chambers in rabbits J Biomed Mater Res 1993;27: Roberts WE, Smith RK, Zilberman Y, Mozsary PG, Smith RS. Osseous adaptation to continuous loading of rigid endosseous implants. Am J Orthod Dentofacial Orthop 1984;86: Byars EF, Snyder RD. Engineering Mechanics of Deformable Bodies. Scranton: International Text Co.; Morrow HW. Statistics and Strength of Materials. Columbus: Prentice Hall; Akca K, Chang TL, Tekdemir I, Fanuscu MI. Biomechanical aspects of initial intraosseous stability and implant design: a quantitative micro-morphometric analysis. Clin Oral Implants Res 2006;17:

60 32. Friberg B, Sennerby L, Roos J, Lekholm U. Identification of bone quality in conjunction with insertion of titianium implants. A pilot study in jaw autopsy specimens. Clin Oral Implants Res 1995;6: Ikumi N, Tsutsumi S. Assessment of correlation between computerized tomography values of the bone and cutting torque values at implant placement. Int J Oral Maxillofac Implants 2005;20: da Cunha HA, Francischone CE, Filho HN, de Oliveira RC. A comparison between cutting torque and resonance frequency in the assessment of primary stability and final torque capacity of standard and TiUnite single-tooth implants under immediate loading. Int J Oral Maxillofac Implants 2004;19: Boyle III JM, Frost DE, Foley WL, Grady JJ. Torque and pullout analysis of six currently available self-tapping and "emergency" screws. J Oral Maxillofac Surg 1993;51: Huja SS, Litsky AS, Beck FM, Johnson KA, Larsen PE. Pull-out strength of monocortical screws placed in the maxillae and mandibles of dogs. Am J Orthod Dentofacial Orthop 2005;127: Pfeiffer M, Gilbertson LG, Goel VK, Griss P, Keller JC, Ryken TC, Hoffman HE. Effect of Specimen Fixation Method on Pullout Tests of Pedical Screws. Spine 1996;21: Collinge CA, Stern S, Cordes S, Lautenschlager EP. Mechanical properties of small fragment screws. Clin Orthoped Rel Res 2000: Zdeblick TA, Kunz MS, Cooke ME, McCabe R. Pedicle Screw Pullout Strength: Correlation with Insertional Torque. Spine 1993;18: Ottoni JM, Oliveira ZF, Mansini R, Cabral AM. Correlation between placement torque and survival of single-tooth implants. Int J Oral & Maxillofac Implants 2005;20: Niimi A, Ozeki K, Ueda M, Nakayama B. A comparative study of removal torque of endosseous implants in the fibula, iliac crest, and scapula of cadavers: preliminary report Clin Oral Impl Res 1997;8:

61 42. Phillips JH, Rahn BA. Comparison of compression and torque measurements of self-tapping and pretapped screws. Plast Reconstr Surg 1989;83: Chapman JR, Harrington RM, Lee KM, Anderson PA, Tencer AF, Kowalski D. Factors affecting the pullout strength of cancellous bone screws. J Biomech Eng 1996;118: Pitzen T, Barbier D, Tintinger F, Steudel WI, Strowitzki M. Screw fixation to the posterior cortical shell does not influence peak torque and pullout in anterior cervical plating. Eur Spine J 2002;11: Hitchon PW, Brenton MD, Coppes JK, From AM, Torner JC. Factors affecting the pullout strength of self-drilling and self-tapping anterior cervical screws. Spine 2003;28: Wagenknecht M, Andrianne Y, Burny F, Donkerwolcke M. Study of the mechanical characteristics of external fixation pin anchorage. Preliminary results. Orthoped 1984;7: Yerby S, Scott CC, Evans NJ, Messing KL, Carter DR. Effect of cutting flute design on cortical bone screw insertion torque and pullout strength. J Orthoped Trauma 2001;15: Cleek TM, Reynolds KJ, Hearn TC. Effect of screw torque level on cortical bone pullout strength. J Orthop Trauma 2007;21: Lekholm U, Zarb GA. Tissue-Integrated Prostheses: Osseointegration in Clinical Dentistry. In: Zarb GA, Albrektsson T. editor. Patient selection and preparation. Chicago: Quintessence; 1985: p Johansson P, Strid KG. Assessment of bone quality from placement resistance during implant surgery. Int J Oral Maxillofac Implants 1994;9: Jensen O. Site classification for the osseointegrated implant. J Prost Dent 1989;1: Klinge B, Johansson CB, Albrektsson T, Hallstrom H, Engdahl T. A new method to obtain bone biopsys at implant sites peri-operatively: technique and bone structure. Clin Oral Impl Res 1995;6:

62 53. Ulm CW, Kneissel M, Hahn M, Solar P, Matejka M, Donath K. Characteristics of the cancellous bone of edentulous mandibles. Clin Oral Impl Res 1997;8: Lyon WF, Cochran JR, Smith L. Actual holding power of various screws in bone. Annals of Surgery 1941;114: Seebeck J, Goldhahn J, Morlock MM, et al. Mechanical behavior of screws in normal and osteoporotic bone. Osteoporos Int 2005;16:S Ansell RM, Scales JT. A study of some factors which effect the strength of screws and their insertion and holding power in bone. J Biomech 1968;1: Nunamaker DM, Perren SM. Force measurements in screw fixation. J Biomech 1976;9: Heam TC, Schatzker J, Wolfson N. Extraction strength of cannuated cancellous bone screws. J Orthop Trauma 1993;7: Bechtol CO, Ferguson AB, Lang PG. Metals and Engineering in Bone and Joint Surgery. In: Bechtol CO, editor. Internal fixation with plates and screws. Baltimore: Williams and Wilkins; Uhl RL. The biomechanics of screws. Orthoped Review 1989;18: Koranyi E, Bowman CE, Knecht CD, et al. Holding power of orthopedic screws in bone. Clin Orthop 1970;72: Evans M, Spencer M, Wang Q, et al. Design and testing of external fixator bone screws. J Biomed Eng 1990;12:

63 CHAPTER 3: JOURNAL ARTICLE ABSTRACT Purpose: This study was designed to isolate two design characteristics of miniscrew implants (MSIs), pitch and fluting, and to determine their effects on primary stability. Methods: Maximum placement torque and pullout strength of MSIs specifically fabricated for this experiment were compared to a control MSI using synthetic and cadaver bone models. The first experimental condition compared MSIs with 0.75 mm and a 1.25 mm pitch to a control MSI with a 1.0 mm pitch. The second experimental condition compared MSIs with three longitudinal flutes to a control without flutes. A total of 60 MSIs (15 of each design) were placed in the synthetic bone model; a split-mouth cadaver model was used to compare the three experimental designs against the control MSIs (total of 45 MSIs). Maximum placement torque was measured as MSIs were inserted with intermittent rotation. To test pull-out strength, a vertical force was applied until failure. Results: The synthetic bone model showed that placement torque and pullout strength were higher for the 0.75 pitch MSIs than the 1.0 mm and 1.25 mm pitch MSIs, but difference were significant (p<.05) only for pull-out strength. The cadaver 51

64 model showed no significant differences in placement torque or pullout strength associated with pitch. In both the synthetic and cadaver bone model, MSIs with flutes had significantly (p<.05) higher placement torque and pullout strength. Spearman correlations between placement torque and pullout strength were statistically significant both in the synthetic (r=.504) and cadaver (r=.502) bone models. Conclusion: Pitch and flutes affect the primary stability of MSIs. Within limits, decreasing MSI pitch increases pullout strength and fluting increases both placement torque and pullout strength. 52

65 INTRODUCTION Anchorage control is crucial for successful orthodontic practice. Using teeth for anchorage can result in unwanted tooth movement and a compromised treatment result. Extra oral devices are less than ideal because they are dependant upon patient compliance and outside the control of the practitioner. Miniscrew implants (MSIs) offer a solution to both of these problems. 1 Since the success of MSIs is dependant upon their stability, design characteristics that might enhance primary stability need to be fully understood. Insertion torque and pullout tests are typically used to analyze a screw s geometry as a function of its primary stability. 2,3-6 Both tests provide an indication of anchorage values of immediately loaded miniscrew implants. 7 Peak insertion torque has been shown to be a significant factor in determining the holding power of screw-type implants. 3-6 Pullout tests are commonly used to measure primary stability in orthopedics, orthodontics, neurosurgery, and maxillofacial surgery. 7 Because pullout tests directed vertically, with forces parallel to the long axis of the screw, evaluate the primary force a screw was designed to resist, they are often used to evaluate the design of a screw-type implants. 8 53

66 Various implant design characteristics have been shown to influence primary stability 9 and mechanical strength. 2 Due to their recent development, our understanding of MSI design characteristics comes primarily from the orthopedic and prosthetic literature. Longer screws exhibit greater pullout strength than shorter screws, 10 while screws with greater diameter provide greater resistance to pullout than their counterparts with smaller diameters Screw-type implants with greater thread depth, or half the difference between major and minor diameter, have greater purchase strength in porous materials and thus higher primary stability. 11,14,15 It has also been shown that asymmetrical thread designs, with a 45 leading and 90 trailing angle, facilitate insertion while making removal more difficult. 2 Pitch and fluting are design characteristics of MSIs that have yet to be fully understood. Decreasing the surface area of a MSI by either increasing the pitch or adding flutes should, theoretically, decrease friction and placement torque. Pitch has been shown to be negatively related with pullout strength; 11,14 decreased pitch is thought to increase the screw purchase strength in porous materials. 14 However, MSIs advance more per turn as pitch increases, which might be expected to create higher torque levels. 16 It has also been shown that there is a decrease in 54

67 placement torque and cortical damage as the number and length of flutes increases. 17 This is thought to be due to an increased clearance of bone chips which tend to accumulate around the threads and provide the resistance. 15,17,18 However, bone chip clearance is also dependant upon an adequate flute dimensions. The presence of flutes has been reported to both decrease 17,19,20 and increase 21,22 pullout strength. Fluted screws should exhibit decreased resistance to pullout during primary stability testing because they are thought to have less holding power than fully threaded scews. 23,19 The purpose of this in vitro study was to evaluate the effects that pitch and fluting have on primary stability of miniscrew implants. This experiment is novel because it isolates one design characteristic while holding all other characteristics constant. 55

68 MATERIALS AND METHODS MINISCREW IMPLANTS The effects of pitch and fluting were tested by comparing experimental designs to a control. This was achieved by altering one design characteristic while maintaining all others. Two different miniscrew implant designs were tested. Pitch and fluting were chosen because 1) there are studies demonstrating their significance, 2) engineering theory supports their influence, and 3) both characteristics can be changed without increasing the size of the MSI. Experimental Designs The experimental MSIs were designed by altering or adding one characteristic of the control MSI. This makes it possible to isolate and test the effect a specific design characteristic. Thread pitch is the distance between the threads. To evaluate the effects of pitch, 1.0 mm pitch MSIs (Figure 6B) were compared to 0.75 mm (Figure 6A) and 1.25 mm (Figure 6C) pitch MSIs. Titanium has a relative strength much higher than that of bone. To match the strength of the thin, titanium MSI threads to that of bone the volume of bone between each thread must be increased. A

69 difference in pitch was chosen because increase in volume of bone between each thread may substantially increase its relative strength, to closer match the strength of the thin titanium threads. A mm B. 1.0 mm C mm Figure 6: A mm Pitch Experimental MSI, B. 1.0 mm Pitch Control MSI, C mm Pitch Experimental MSI. To evaluate the effects of fluting, three longitudinal flutes extending the full length of the threaded portion were added to the control MSI design (Figure 7). The depth of each flute extended through the threads to the core and was 0.225mm wide. The surfaces of the flutes were cutting to facilitate placement and removal. 57

70 Figure 7: Fluted MSI: The Same Dimensions as the Control MSI with the Addition of Three Longitudinal Flutes. Control Design The experimental miniscrew designs were all compared to the control MSI design (Figure 1B). The control was a surgical grade titanium implant 6 mm in length, with a major diameter of 1.8 mm, a minor diameter of 1.6 mm, and a corresponding thread depth of 0.1 mm. The thread was a 90 asymmetric buttress design with a 1.0 mm pitch. The apical 3 mm of the MSI was tapered; it was self-drilling and selftapping. TESTING METHODS Two models were used for this investigation, including a synthetic bone model, used to ensure uniformity of the sample and a cadaver model, used to mimic clinical 58

71 applicability. Testing MSIs in these two different materials serves to validate the results, assuming similarities in performance in both materials. Synthetic Bone Model A synthetic polyurethane cancellous bone model (Sawbones, Vashon, WA) was used to evaluate insertion torque and pullout strength. Synthetic bone has been commonly used to evaluate design characteristics of screws 2,11,14 due to its consistent material properties. The synthetic model had a density of 15 pounds per cubic foot, a compressive strength 5.2 MPa, a tensile strength of 3.7 MPa, shear strength of 3.2 MPa, and a moduli of 156 MPa, 173 MPa, and 42 MPa respectively. Square blocks (10mm³) of synthetic bone were cut from a larger block for testing. Fifteen control MSIs and 45 experimental MSIs (15 each)were evaluated in the synthetic bone. Cadaver Bone Model Partially dentate male cadavers between 68 and 91 years of age with formalin perfusion were used. This method of preservation has been shown to maintain mechanical properties similar to fresh bone. 24 All subjects were free of overt osseous pathology. The soft tissues were dissected 59

72 and the mandibles were sectioned at the symphysis to facilitate testing. A split-mouth design (15 MSIs per side) was used to evaluate the effects of pitch and fluting, with a control MSI randomly assigned to one side and an experimental MSI placed in the same location on the other side. The placement sites were restricted to the alveolus and posterior to the external oblique ridge. MECHANICAL TESTING Placement torque was chosen as one of the mechanical tests of primary stability because: 1) it is commonly used 2) it has been found to be a good predictor of bone-toimplant interface failure 5 3) it can be paired with other tests without compromising the integrity of the miniscrew implant. Pullout testing was performed because it is a test of implant design often used for evaluating the effects of MSI design on primary stability. 4,7,8 In addition, orthodontic MSIs are loaded in such a way that pullout testing is arguably more clinically valid than other methods of testing. 60

73 Placement Torque Measurements of placement torque were obtained using the Mecmesin Advanced Force and Torque Indicator (Mecmesin, Ltd, West Sussex, UK). All miniscrews were handplaced with intermittent rotation into specimens secured in a custom device using the Mecmesin Static Torque Screwdriver. The torque screwdriver was braced and maintained in the same position throughout insertion. All MSIs were fit into the torque screwdriver with an adapter manufactured by the Saint Louis University Engineering Department. Overall maximum insertion torque (MIT) was recorded in Ncm. Pullout Strength For pullout testing, the specimens were mounted on an Instron Machine Model 1011 (Instron Corp, Canton, MA) using custom holding devices (one for synthetic and one for cadaver bone) that allowed flexure of the bone by leaving an 8 mm 2 cylindrical area surrounding the unsupported MSI. Forces were applied to the miniscrews by threading two inch stainless steel ligatures through the MSI head securing in ligation to the Instron machine, and applying a vertical force of 10 mm/min oriented parallel to the long axis until failure. Load-displacement data was recorded and 61

74 the peak load at failure was obtained from the readout and reported in Newtons (N). STATISTICAL ANALYSIS The skewness and kurtosis statistics showed significant (p< 0.05) departures from normality for a number of distributions. On that basis, medians (50 th percentile) and interquartile ranges (25 th,75 th percentile) were used for descriptive purposes. Pitch was first evaluated using Kruskal-Wallis test followed by pairwise comparison using a Mann-Whitney test. The effect of fluting was evaluated using a Mann-Whitney test. The relationship between insertion torque and pullout strength was assessed using Spearman rank order correlation. RESULTS SYNTHETIC BONE MODEL Pitch Placement torque with 0.75 mm pitch was higher than placement torque with 1.0 mm and 1.25 mm pitch MSIs, but the differences were not statistically significant (Table 3.1; Figure 8). The maximum pullout force at failure was significantly different (p=.001) between the three groups of MSI. Post hoc test showed it was significantly higher 62

75 for the 0.75 mm pitch MSIs than the MSIs with a 1.0 mm pitch (p<.001). Differences in maximum pullout between the 1.0 mm and 1.25 mm pitch MSIs were not statistically significant (Table 3.1; Figure 9). Pullout of the 0.75 mm pitch MSIs was statistically greater (p=.015) than pullout with a 1.25 mm pitch. Table 3.1: Median, 25 th, and 75 th Quartile for Placement Torque and Pullout Testing of MSIs with Pitch Variation in a Synthetic Model mm 1.0 mm 1.25 mm 50 th 25 th, 50 th 25 th, 50 th 25 th, Diff Prob 75 th 75 th 75 th Placement Torque (Ncm) Pullout (N)

76 Ncm Pitch Figure 8: Median Placement Torque of 0.75 mm, 1.0 mm (Control), and 1.25 mm Pitch MSIs in Synthetic Bone with Error Bars Representing the 25 th and 75 th Quartiles 64

77 N Pitch Figure 9: Median Pullout Strength of 0.75 mm, 1.0 mm (Control), and 1.25 mm Pitch MSIs in Synthetic Bone with Error Bars Representing the 25 th and 75 th Quartiles Fluted Placement torque (Table 3.2; Figure 10) and pullout strength (Table 3.2; Figure 11) were significantly greater (p<.001) for the MSIs with flutes than the control MSIs in the synthetic bone model. 65

78 Table 3.2: Median, 25 th, and 75 th Quartile for Placement Torque and Pullout Testing of MSIs with and without Fluting in a Synthetic Model Fluted 50 th 25 th, Control 50 th 25 th, Diff Prob 75 th 75 th Placement Torque (Ncm) <.001 Pullout Strength (N) < Ncm Fluted Control MSI Type Figure 10: Median Placement Torque of Fluted and Control MSIs in Synthetic Bone with Error Bars Representing the 25 th and 75 th Quartiles 66

79 N Fluted MSI Type Control Figure 11: Median Pullout Strength of Fluted and Control MSIs in Synthetic Bone with Error Bars Representing the 25 th and 75 th Quartiles CADAVER BONE MODEL Pitch No significant differences in placement torque or pullout strength were found between the 0.75 mm and 1.0 mm pitch MSIs (Table 3.3; Figures 12 and 13). 67

80 Table 3.3: Torque and Pullout Comparisons of a 0.75 mm and a 1.0 mm Pitch MSI in a Cadaver Model 0.75 mm 1.0 mm 50 th 25 th, 50 th 25 th, Diff Prob 75 th 75 th Placement Torque (Ncm) Pullout Strength (N) Ncm Pitch Figure 12: Median Placement Torque of 0.75 mm and 1.0 mm (Control) Pitch MSIs in Cadaver Bone with Error Bars Representing the 25 th and 75 th Quartiles 68

81 N Pitch Figure 13: Median Pullout Strength of 0.75 mm and 1.0 mm (Control) Pitch MSIs in Cadaver Bone with Error Bars Representing the 25 th and 75 th Quartiles There were also no significant differences in placement torque or pullout strength between the 1.0 mm and 1.25 mm pitch MSIs (Table 3.4; Figures 14 and 15). 69

82 Table 3.4: Torque and Pullout Comparisons of a 1.0 mm and a 1.25 mm Pitch MSI in a Cadaver Model 1.0 mm 1.25 mm 50 th 25 th, 50 th 25 th, Diff Prob 75 th 75 th Placement Torque (Ncm) Pullout Strength (N) Ncm Pitch Figure 14: Median Placement Torque of 1.0 mm (Control) and 1.25 mm Pitch MSIs in Cadaver Bone with Error Bars Representing the 25 th and 75 th Quartiles 70

83 N Pitch Figure 15: Median Pullout Strength of 1.0 mm (Control) and 1.25 mm Pitch MSIs in Cadaver Bone with Error Bars Representing the 25 th and 75 th Quartiles Fluted Placement torque of the fluted MSIs was significantly greater (p<.001) than the control MSIs (Table 3.5; Figure 16). Pullout strength was also significantly greater (p=.027) for the fluted MSIs in comparison to the control MSIs (Table 3.5; Figure 17). 71

84 Table 3.5: Mean and Standard Deviation for Placement Torque and Pullout Testing of MSIs with and without Fluting in a Cadaver Model Fluted 50 th 25 th, Control 50 th 25 th, Diff Prob 75th 75th Placement Torque (Ncm) Pullout Strength (N) Ncm Fluted Control MSI Type Figure 16: Median Placement Torque of Fluted and Control MSIs in Cadaver Bone with Error Bars Representing the 25 th and 75 th Quartiles 72

85 N Fluted MSI Type Control Figure 17: Median Pullout Strength of Fluted and Control MSIs in Cadaver Bone with Error Bars Representing the 25 th and 75 th Quartiles Placement torque and pullout strength were significantly correlated in both the synthetic (r=.504;p<.001) and cadaver (r=.502;p<.001) bone models. 73

86 DISCUSSION MSI with 0.75 mm pitch provided greater primary stability than 1.0 mm pitch MSIs. The synthetic model showed that pullout resistance of MSIs with 0.75 mm pitch was significantly higher than those with a pitch of 1.0 mm. Although the differences were not statistically significant, the 0.75 mm MSIs also had higher placement torque than the 1.0 mm MSIs. A pitch of 0.75 mm provides a greater surface area than the 1.0 mm pitch, which increases torque due to increased friction at the bone-to-screw interface. 9 Although not statistically significant, the cadaver model displayed similar tendencies. These findings agree with studies showing significant increases in pullout resistance with decreasing pitch. 11,14 A decrease in MSI pitch is thought to increase resistance to pullout due to an increase in MSI purchase strength in porous materials. 14 Unexpectedly, no significant differences were noted between the 1.25 mm and 1.0 mm pitch MSIs for either placement torque or pullout strength. The expected decreases in torque and pullout strength with the 1.25 mm pitch MSIs might have been compensated for by higher compressive torque. MSIs are designed to convert torque into compressive force between the screw and the bone. The inclined plane of the MSI thread produces an axial force as 74

87 it is placed, 16 causing shear stress to develop during screw insertion. 25 MSI pitch determines how much the screw advances with each turn, 16 with greater amounts of advancement creating higher axial forces. A pitch of 1.25 mm advances the MSI more per turn than a 1.0 mm pitch; it also creates more axial stress (or compression) and torque. The thickness of the bone should be also considered because primary stability could be compromised when inserting a MSI with a pitch of 1.25 mm into cortical bone that is only 1-2 mm thick, due to the limited space available for thread engagement. 26 Cutting flutes had a significant effect on both placement torque and pullout strength of the miniscrew implants. The fluted MSIs had significantly higher placement torques than the control MSIs, which could be attributed to a couple of possibilities. First, bone chips may have accumulated around the threads of the fluted MSIs, providing greater friction and resistance to insertion. 15,17,18 Secondly, since the MSIs were inserted with intermittent rotation, the bone would have the opportunity to relax into the flutes during the rotational pauses, 27 making it necessary to exert more force to turn the MSI. 21,27 The significant increase in pullout resistance with fluted MSIs can be explained by the engineering principal of 75

88 matched deformations. This principal states that components must be designed to deform similarly under a given load in order to avoid uneven distribution of load and high stress concentration. 28 The flutes created breaks in the titanium threads, making them more flexible and decreasing the stiffness of the MSI so as to more closely match the bone. Similar load deformations of two interacting systems even out the load distribution and increase the strength of the system. Moderate positive correlations were found between placement torque and pullout strength in both the synthetic and cadaver bone models, which supports studies in the orthopedic literature by Yerby et al. 17 and Zdeblick et al. 5 Various disciplines have shown that peak insertion torque is a significant factor for predicting the holding strength of screw-type implants. 3-6 Ryken et al. have suggested that up to 76% of the variability in pullout strength can be attributed to placement torque. 4 The relationship suggests that increased torque reflects tighter bone-to-implant contact and increased stability. Finally, this study showed a high coefficient of variability for insertion torque and pullout strength, especially for the cadaver models. This could be explained partially by variability in the design of the MSI heads. 76

89 Some of the MSIs were more difficult to remove from the driver than others, which could increase variability by damaging the integrity of the surrounding bone during MSI removal. Secondly, the use of human cadaver models adds variation to testing because of differences in density and cortical thickness not only among the samples, but also within each specimen. 11,29,30 Finally, even though all tests were performed in a standardized manner by one operator with calibrated instruments, measurement variability was obviously introduced based on the results from the synthetic bone model. CONCLUSIONS - Pullout strength significantly increases as pitch decreases from 1.0 mm to 0.75 mm. - There is no significant difference in placement torque or pullout strength of MSIs with a 1.0 mm and 1.25 mm pitch. - MSIs with flutings have significantly higher placement torque and pullout strength than MSI without flutes. - There is a positive correlation between placement torque and pullout strength. 77

90 REFERENCES CITED 1. Cheng SJ, Tseng IY, Lee JJ, Kok SH. A prospective study of the risk factors associated with failure of mini-implants used for orthodontic anchorage. The Int J Oral Maxillofac Implants 2004;19: Carano A, Lonardo P, Velo S, Incorvati C. Mechanical properties of three different commercially available miniscrews for skeletal anchorage. Prog Orthod 2005;6: O'Sullivan D, Sennerby L, Meredith N. Measurements comparing the initial stability of five designs of dental implants: a human cadaver study. Clin Implant Dent Relat Res 2000;2: Ryken TC, Clausen JD, Traynelis VC, Goel VK. Biomechanical analysis of bone mineral density, insertion technique, screw torque, and holding strength of anterior cervical plate screws. J Neurosurg 1995;83: Zdeblick TA, Kunz MS, Cooke ME, McCabe R. Pedicle Screw Pullout Strength: Correlation with Insertional Torque. Spine 1993;18: Ottoni JM, Oliveira ZF, Mansini R, Cabral AM. Correlation between placement torque and survival of single-tooth implants. Int J Oral Maxillofac Implants 2005;20: Huja SS, Litsky AS, Beck FM, Johnson KA, Larsen PE. Pull-out strength of monocortical screws placed in the maxillae and mandibles of dogs. Am J Orthod Dentofac Orthop 2005;127: Pfeiffer M, Gilbertson LG, Goel VK, Griss P, Keller JC, Ryken TC, Hoffman HE. Effect of Specimen Fixation Method on Pullout Tests of Pedical Screws. Spine 1996;21:

91 9. da Cunha HA, Francischone CE, Filho HN, de Oliveira RC. A comparison between cutting torque and resonance frequency in the assessment of primary stability and final torque capacity of standard and TiUnite singletooth implants under immediate loading. Int J Oral Maxillofac Implants 2004;19: Hitchon PW, Brenton MD, Coppes JK, From AM, Torner JC. Factors affecting the pullout strength of selfdrilling and self-tapping anterior cervical screws. Spine 2003;28: DeCoster TA, Heetderks DB, Downey DJ, Ferries JS, Jones W. Optimizing bone screw pullout force. J Orthoped Trauma 1990;4: Schatkzer J, Sanderson R, Murnaghan JP. The holding power of orthopedic screws in vivo. Clin Orthoped Rel Res 1975: Heam TC, Schatzker J, Wolfson N. Extraction strength of cannuated cancellous bone screws. J Orthop Trauma 1993;7: Chapman JR, Harrington RM, Lee KM, Anderson PA, Tencer AF, Kowalski D. Factors affecting the pullout strength of cancellous bone screws. J Biomech Eng 1996;118: Johnson NL, Galuppo LD, Stover SM, Taylor KT. An in vitro biomechanical comparison of the insertion variables and pullout mechanical properties of AO 6.5- mm standard cancellous and 7.3-mm self-tapping, cannulated bone screws in foal femoral bone. Vet Surg 2004;33: Perren SM, Cordey J, Baumgart F, Rahn BA, Schatzer J. Technical and biomechanical aspects of screws used for bone surgery. Int J of Orthoped Trauma 1992;2: Yerby S, Scott CC, Evans NJ, Messing KL, Carter DR. Effect of cutting flute design on cortical bone screw insertion torque and pullout strength. J Orthoped Trauma 2001;15: Uhl RL. The biomechanics of screws. Orthopaedic Review 1989;18:

92 19. Bechtol CO, Ferguson AB, Lang PG. Metals and Engineering in Bone and Joint Surgery. In: Bechtol CO, editor. Internal fixation with plates and screws. Baltimore: Williams and Wilkins; Koranyi E, Bowman CE, Knecht CD, et al. Holding power of orthopedic screws in bone. Clin Orthop 1970;72: Boyle III JM, Frost DE, Foley WL, Grady, JJ. Torque and pullout analysis of six currently available selftapping and "emergency" screws. J Oral Maxillofac Surg 1993;51: Evans M, Spencer M, Wang Q, et al. Design and testing of external fixator bone screws. J Biomed Eng 1990;12: Fanuscu MI, Chang, TL. Three-dimensional morphometric analysis of human cadaver bone: microstructural data from maxilla and mandible. Clin Oral Impl Res 2004;15: Wingerter S, Calvert G, Tucci M, Binghuzzi H, Russell G, Puckett A. Mechanical strength repercussions of various fixative storage methods on bone. Biomed Sci Instrum 2006;42: Hughes AN, Jordan BA. The mechanical properties of surgical bone screws and some aspects of insertion practice. Injury 1972;4: Phillips JH, Rahn BA. Comparison of compression and torque measurements of self-tapping and pretapped screws. Plast Reconstr Surg 1989;83: Ansell RM, Scales JT. A study of some factors which effect the strength of screws and their insertion and holding power in bone. J Biomech 1968;1: Pahl G, Beitz W. Engineering Design: A Systematic Approach. London: Springer-Verlag;

93 29. Schwartz-Dabney CL, Dechow PC. Edentulation alters material properties of cortical bone in the human mandible. J Dent Res 2002;81: Wagenknecht M, Andrianne Y, Burny F, Donkerwolcke M. Study of the mechanical characteristics of external fixation pin anchorage. Preliminary results. Orthopedics 1984;7:

94 VITA AUCTORIS Christine Lynn Brinley was born on December 9, 1978 in Belleville, Illinois to Alfred and Clara Brinley. She was raised with her parents and three younger sisters in Collinsville, Illinois. She graduated from Southern Illinois University in Edwardsville, Illinois in 2001 with a Bachelor of Science Degree in Biology. From 2001 to 2005, she attended Southern Illinois University School of Dental Medicine in Alton, Illinois. Graduating Cum Laude, she was awarded a Doctorate of Dental Medicine in It is anticipated that in May 2008 Christine will graduate from Saint Louis University with a Master of Science degree in Dentistry with an emphasis in Orthodontics and enter private practice in Maryville, Illinois. In August of 2006, Christine married Michal Matras of Pulawy, Poland. They now reside in Illinois. 82

Structural Integrity Analysis

Structural Integrity Analysis Structural Integrity Analysis 1. STRESS CONCENTRATION Igor Kokcharov 1.1 STRESSES AND CONCENTRATORS 1.1.1 Stress An applied external force F causes inner forces in the carrying structure. Inner forces

More information

Solid shape molding is not desired in injection molding due to following reasons.

Solid shape molding is not desired in injection molding due to following reasons. PLASTICS PART DESIGN and MOULDABILITY Injection molding is popular manufacturing method because of its high-speed production capability. Performance of plastics part is limited by its properties which

More information

MINI FRAGMENT SYSTEM. Instruments and implants for 1.5 mm, 2.0 mm, and 2.4 mm plate fixation PRODUCT OVERVIEW

MINI FRAGMENT SYSTEM. Instruments and implants for 1.5 mm, 2.0 mm, and 2.4 mm plate fixation PRODUCT OVERVIEW MINI FRAGMENT SYSTEM Instruments and implants for 1.5 mm, 2.0 mm, and 2.4 mm plate fixation PRODUCT OVERVIEW TABLE OF CONTENTS INTRODUCTION Mini Fragment System 2 PRODUCT INFORMATION Plates 4 Screws 6

More information

IMPLANT DENTISTRY EXAM BANK

IMPLANT DENTISTRY EXAM BANK IMPLANT DENTISTRY EXAM BANK 1. Define osseointegration. (4 points, 1/4 2. What are the critical components of an acceptable clinical trial? (10 points) 3. Compare the masticatory performance of individuals

More information

Dental Updates. Excerpted Article e-mail: [email protected]. Why Implant Screws Loosen Part 1. Richard Erickson, MS, DDS

Dental Updates. Excerpted Article e-mail: re777@comcast.net. Why Implant Screws Loosen Part 1. Richard Erickson, MS, DDS ¼ ½ ¾ µ mw/cm 2 Volume 17; 2007 Dental Updates "CUTTING EDGE INFORMATION FOR THE DENTAL PROFESSIONAL " 200 SEMINARS AND 30 JOURNALS REVIEWED YEARLY FOR THE LATEST, CUTTING EDGE INFORMATION Excerpted Article

More information

ORTHODONTIC MINI IMPLANTS Clinical procedure for positioning. Orthodontics and Implantology

ORTHODONTIC MINI IMPLANTS Clinical procedure for positioning. Orthodontics and Implantology ORTHODONTIC MINI IMPLANTS Clinical procedure for positioning Orthodontics and Implantology 2 All rights are reserved. Any reproduction of the present publication is prohibited in whole or in part and by

More information

Zimmer Small Fragment Universal Locking System. Surgical Technique

Zimmer Small Fragment Universal Locking System. Surgical Technique Zimmer Small Fragment Universal Locking System Surgical Technique Zimmer Small Fragment Universal Locking System 1 Zimmer Small Fragment Universal Locking System Surgical Technique Table of Contents Introduction

More information

CRITERIA FOR PRELOADED BOLTS

CRITERIA FOR PRELOADED BOLTS National Aeronautics and Space Administration Lyndon B. Johnson Space Center Houston, Texas 77058 REVISION A JULY 6, 1998 REPLACES BASELINE SPACE SHUTTLE CRITERIA FOR PRELOADED BOLTS CONTENTS 1.0 INTRODUCTION..............................................

More information

Don t Let Life Pass You By Because Of Oral Bone Loss

Don t Let Life Pass You By Because Of Oral Bone Loss Don t Let Life Pass You By Because Of Oral Bone Loss Ask For Dental Implant Solutions From BIOMET 3i Scan With Your Smartphone! In order to scan QR codes, your mobile device must have a QR code reader

More information

The SATURN implant by Cortex Dental Industries

The SATURN implant by Cortex Dental Industries The SATURN implant by Cortex Dental Industries By Dr. Zvi Laster DMD W e P r o v e I t E v e r y D a y A case report using a newly designed implant specifically designed for immediate post-extraction loading

More information

Fixing Dental Implant Component Failure and Fracture: Retrieving the Fractured Implant Abutment Screw. Charles A. Mastrovich, DDS

Fixing Dental Implant Component Failure and Fracture: Retrieving the Fractured Implant Abutment Screw. Charles A. Mastrovich, DDS Fixing Dental Implant Component Failure and Fracture: Retrieving the Fractured Implant Abutment Screw Charles A. Mastrovich, DDS [email protected] www.mastrovichdental.com Evaluating component failure

More information

Force measurement. Forces VECTORIAL ISSUES ACTION ET RÉACTION ISOSTATISM

Force measurement. Forces VECTORIAL ISSUES ACTION ET RÉACTION ISOSTATISM Force measurement Forces VECTORIAL ISSUES In classical mechanics, a force is defined as "an action capable of modifying the quantity of movement of a material point". Therefore, a force has the attributes

More information

Dental Implant Options in Atrophic Jaws

Dental Implant Options in Atrophic Jaws Dental Implant Options in Atrophic Jaws Orthopedic Application Jay B. Reznick, D.M.D., M.D. Diplomate, American Board of Oral and Maxillofacial Surgery Tarzana, CA Endopore Dental Implant System Screw-Type

More information

Aesculap Veterinary Orthopaedics. Targon VET Interlocking Nail

Aesculap Veterinary Orthopaedics. Targon VET Interlocking Nail Aesculap Veterinary Orthopaedics Targon VET Interlocking Nail Flexibility Stability Dynamics The Targon VET is based on new, technology which reverses the known principles of interlocking nails. The rotation-stabilising

More information

8.2 Elastic Strain Energy

8.2 Elastic Strain Energy Section 8. 8. Elastic Strain Energy The strain energy stored in an elastic material upon deformation is calculated below for a number of different geometries and loading conditions. These expressions for

More information

Torsion Tests. Subjects of interest

Torsion Tests. Subjects of interest Chapter 10 Torsion Tests Subjects of interest Introduction/Objectives Mechanical properties in torsion Torsional stresses for large plastic strains Type of torsion failures Torsion test vs.tension test

More information

Technique Guide. Large Fragment LCP Instrument and Implant Set. Part of the Synthes locking compression plate (LCP) system.

Technique Guide. Large Fragment LCP Instrument and Implant Set. Part of the Synthes locking compression plate (LCP) system. Technique Guide Large Fragment LCP Instrument and Implant Set. Part of the Synthes locking compression plate (LCP) system. Table of Contents Introduction Large Fragment LCP Instrument and Implant Set

More information

Contact: Steve Hurson VP, R & D Nobel Biocare 22715 Savi Ranch Pkwy Yorba Linda, CA 92887 (714) 282-5072 steve.hurson@nobelbiocare.

Contact: Steve Hurson VP, R & D Nobel Biocare 22715 Savi Ranch Pkwy Yorba Linda, CA 92887 (714) 282-5072 steve.hurson@nobelbiocare. Use of CP Titanium and Titanium alloys for Dental Implants Steve Hurson, Nobel Biocare U.S.A. Yorba Linda CA Abstract Professor Per Ingvar Branemark of Sweden published his results on 15 years of research

More information

Technique Guide. 3.5 mm and 4.5 mm Curved Locking Compression Plates (LCP). For minimally invasive osteosynthesis.

Technique Guide. 3.5 mm and 4.5 mm Curved Locking Compression Plates (LCP). For minimally invasive osteosynthesis. Technique Guide 3.5 mm and 4.5 mm Curved Locking Compression Plates (LCP). For minimally invasive osteosynthesis. Table of Contents Introduction 3.5 mm and 4.5 mm Curved Locking Compression 2 Plates (LCP)

More information

Microimplant Anchorage in Orthodontics

Microimplant Anchorage in Orthodontics Microimplant Anchorage in Orthodontics *Hee-Moon Kyung, **Bong-Gyu chang,*** Seong-Min Bae *Professor, ** Post-graduate student, *** Clinical Associate Professor Department of Orthodontics Dental School,

More information

Replacement of the upper left central incisor with a Straumann Bone Level Implant and a Straumann Customized Ceramic Abutment

Replacement of the upper left central incisor with a Straumann Bone Level Implant and a Straumann Customized Ceramic Abutment Replacement of the upper left central incisor with a Straumann Bone Level Implant and a Straumann Customized Ceramic Abutment by Dr. Ronald Jung and Master Dental Technician Xavier Zahno Initial situation

More information

Biomechanics of Joints, Ligaments and Tendons.

Biomechanics of Joints, Ligaments and Tendons. Hippocrates (460-377 B.C.) Biomechanics of Joints, s and Tendons. Course Text: Hamill & Knutzen (some in chapter 2 and 3, but ligament and tendon mechanics is not well covered in the text) Nordin & Frankel

More information

Resorptive Changes of Maxillary and Mandibular Bone Structures in Removable Denture Wearers

Resorptive Changes of Maxillary and Mandibular Bone Structures in Removable Denture Wearers Resorptive Changes of Maxillary and Mandibular Bone Structures in Removable Denture Wearers Dubravka KnezoviÊ-ZlatariÊ Asja»elebiÊ Biserka LaziÊ Department of Prosthodontics School of Dental Medicine University

More information

Molar Uprighting Dr. Margherita Santoro Division of Orthodontics School of Dental and Oral surgery. Consequences of tooth loss.

Molar Uprighting Dr. Margherita Santoro Division of Orthodontics School of Dental and Oral surgery. Consequences of tooth loss. Molar Uprighting Dr. Margherita Santoro Division of Orthodontics School of Dental and Oral surgery Molars The wide occlusal surface is designed for food grinding. The surface needs to be aligned with the

More information

ICUC One-Page Paper. Mixing a lag screw with a splinting protection plate Alberto Fernández Dell Oca, Pietro Regazzoni, Stephan Perren May 2016

ICUC One-Page Paper. Mixing a lag screw with a splinting protection plate Alberto Fernández Dell Oca, Pietro Regazzoni, Stephan Perren May 2016 Mixing a lag screw with a splinting protection plate Alberto Fernández Dell Oca, Pietro Regazzoni, Stephan Perren May 2016 The 20th century has seen three essential improvements of surgical fracture fixation:

More information

Zimmer Periarticular Proximal Tibial Locking Plate

Zimmer Periarticular Proximal Tibial Locking Plate Zimmer Periarticular Proximal Tibial Locking Plate Surgical Technique The Science of the Landscape Zimmer Periarticular Proximal Tibial Locking Plate 1 Table of Contents Introduction 2 Locking Screw Technology

More information

System. Stability. Security. Integrity. 150 Helical Anchor

System. Stability. Security. Integrity. 150 Helical Anchor Model 150 HELICAL ANCHOR System PN #MBHAT Stability. Security. Integrity. 150 Helical Anchor System About Foundation Supportworks is a network of the most experienced and knowledgeable foundation repair

More information

METU DEPARTMENT OF METALLURGICAL AND MATERIALS ENGINEERING

METU DEPARTMENT OF METALLURGICAL AND MATERIALS ENGINEERING METU DEPARTMENT OF METALLURGICAL AND MATERIALS ENGINEERING Met E 206 MATERIALS LABORATORY EXPERIMENT 1 Prof. Dr. Rıza GÜRBÜZ Res. Assist. Gül ÇEVİK (Room: B-306) INTRODUCTION TENSION TEST Mechanical testing

More information

Technical Notes 3B - Brick Masonry Section Properties May 1993

Technical Notes 3B - Brick Masonry Section Properties May 1993 Technical Notes 3B - Brick Masonry Section Properties May 1993 Abstract: This Technical Notes is a design aid for the Building Code Requirements for Masonry Structures (ACI 530/ASCE 5/TMS 402-92) and Specifications

More information

Ridge Reconstruction for Implant Placement

Ridge Reconstruction for Implant Placement Volume 1, No. 5 July/August 2009 The Journal of Implant & Advanced Clinical Dentistry Ridge Reconstruction for Implant Placement 2 Hours of CE Credit Oral Implications of Cancer Chemotherapy Immediate

More information

Influence of Biomechanical Factors on Restoration of Devitalized Teeth

Influence of Biomechanical Factors on Restoration of Devitalized Teeth Influence of Biomechanical Factors on Restoration of Devitalized Teeth Adnan atoviê 1 Davor Seifert 1 Renata Poljak-Guberina 1 Boris KvasniËka 2 1 Department of Fixed Prosthodontics School of Dental Medicine

More information

Objectives. Experimentally determine the yield strength, tensile strength, and modules of elasticity and ductility of given materials.

Objectives. Experimentally determine the yield strength, tensile strength, and modules of elasticity and ductility of given materials. Lab 3 Tension Test Objectives Concepts Background Experimental Procedure Report Requirements Discussion Objectives Experimentally determine the yield strength, tensile strength, and modules of elasticity

More information

Universal Screw Removal System (USR)

Universal Screw Removal System (USR) Craniomaxillofacial Surgery 3 Universal Screw Removal System (USR) Craniomaxillofacial rigid fixation systems are available from a variety of manufacturers. The USR system is a complete screwdriver array

More information

More than a fixed rehabilitation.

More than a fixed rehabilitation. More than a fixed rehabilitation. A reason to smile. In combination with: Patient expectations drive dental treatments for fixed edentulous immediate restorations. Patients today have increasingly high

More information

Implants in your Laboratory: Abutment Design

Implants in your Laboratory: Abutment Design 1/2 point CDT documented scientific credit. See Page 41. Implants in your Laboratory: Abutment Design By Leon Hermanides, CDT A patient s anatomical limitations have the greatest predictive value for successful

More information

In Class IV arch: Fulcrum line passes through two abutments adjacent to single edentulous space.

In Class IV arch: Fulcrum line passes through two abutments adjacent to single edentulous space. It is that part of removable partial denture which assists the direct retainers in preventing displacement of distal extension denture bases by resisting lever action from the opposite side of the fulcrum

More information

TriLock 1.5 Implants for the Phalanges

TriLock 1.5 Implants for the Phalanges P R O D U C T IN F O RM AT I O N TriLock 1.5 Implants for the Phalanges APTUS Hand 2 TriLock 1.5 Implants for the Phalanges TriLock 1.5 The Smallest Locking System Small, slender, strong Low profile and

More information

The mechanical behavior of the bone microstructure around dental implants

The mechanical behavior of the bone microstructure around dental implants BUDAPEST UNIVERSITY OF TECHNOLOGY AND ECONOMICS DEPARTMENT OF STRUCTURAL MECHANICS The mechanical behavior of the bone microstructure around dental implants Summary and Theses of PhD dissertation ÉVA LAKATOS

More information

How To Fix A Radial Head Plate

How To Fix A Radial Head Plate Mayo Clinic CoNGRUENT RADIAL HEAD PLATE Since 1988 Acumed has been designing solutions to the demanding situations facing orthopedic surgeons, hospitals and their patients. Our strategy has been to know

More information

Chapter Outline. Mechanical Properties of Metals How do metals respond to external loads?

Chapter Outline. Mechanical Properties of Metals How do metals respond to external loads? Mechanical Properties of Metals How do metals respond to external loads? Stress and Strain Tension Compression Shear Torsion Elastic deformation Plastic Deformation Yield Strength Tensile Strength Ductility

More information

NCB Distal Femur System. Surgical Technique

NCB Distal Femur System. Surgical Technique NCB Distal Femur System Surgical Technique NCB Distal Femur System Surgical Technique 3 Surgical Technique NCB Distal Femur System Table of Contents Introduction 4 Indications 8 Preoperative Planning

More information

CDT 2015 Code Change Summary New codes effective 1/1/2015

CDT 2015 Code Change Summary New codes effective 1/1/2015 CDT 2015 Code Change Summary New codes effective 1/1/2015 Code Nomenclature Delta Dental Policy D0171 Re-Evaluation Post Operative Office Visit Not a Covered Benefit D0351 3D Photographic Image Not a Covered

More information

Standard Internal Hex

Standard Internal Hex Standard Internal Hex Touareg TM -OS Touareg TM -S Swell TM Touareg -S Touareg -OS Swell About ADIN Adin Dental Implant Systems Ltd., designs, manufactures and markets state of the art, technologically

More information

Bending, Forming and Flexing Printed Circuits

Bending, Forming and Flexing Printed Circuits Bending, Forming and Flexing Printed Circuits John Coonrod Rogers Corporation Introduction: In the printed circuit board industry there are generally two main types of circuit boards; there are rigid printed

More information

A Comprehensive Explanation

A Comprehensive Explanation Dental Implants A Comprehensive Explanation Overview Since the 1980s, dental implants have become more popular among dentists and patients. 1 In some clinical situations, implants may be the best treatment

More information

Removing chips is a method for producing plastic threads of small diameters and high batches, which cause frequent failures of thread punches.

Removing chips is a method for producing plastic threads of small diameters and high batches, which cause frequent failures of thread punches. Plastic Threads Technical University of Gabrovo Yordanka Atanasova Threads in plastic products can be produced in three ways: a) by direct moulding with thread punch or die; b) by placing a threaded metal

More information

LOMAS / MONDEFIT ORTHODONTICS. The most innovative anchorage method for tooth correction

LOMAS / MONDEFIT ORTHODONTICS. The most innovative anchorage method for tooth correction LOMAS / MONDEFIT The most innovative anchorage method for tooth correction No extraction of healthy teeth Completion of tooth correction without gaps Symmetrical, beautiful smile For young people and young

More information

What Dental Implants Can Do For You!

What Dental Implants Can Do For You! What Dental Implants Can Do For You! Putting Smiles into Motion About Implants 01. What if a Tooth is Lost and the Area is Left Untreated? 02. Do You Want to Restore Confidence in Your Appearance? 03.

More information

Orthopedic Foot Instruments. Dedicated instruments for reconstructive foot surgery.

Orthopedic Foot Instruments. Dedicated instruments for reconstructive foot surgery. Orthopedic Foot Instruments. Dedicated instruments for reconstructive foot surgery. Surgical Technique This publication is not intended for distribution in the USA. Instruments and implants approved by

More information

Screw Thread Design. Rev. 3-4-09

Screw Thread Design. Rev. 3-4-09 Screw Thread Design Screw Thread Fundamentals A screw thread is defined as a ridge of uniform section in the form of a helix on either the external or internal surface of a cylinder. Internal threads refer

More information

Precision Miniature Load Cell. Models 8431, 8432 with Overload Protection

Precision Miniature Load Cell. Models 8431, 8432 with Overload Protection w Technical Product Information Precision Miniature Load Cell with Overload Protection 1. Introduction The load cells in the model 8431 and 8432 series are primarily designed for the measurement of force

More information

due to uncertainty. This, in turn, has a direct impact on equipment availability and maintenance costs. Unfortunately, due to misconceptions and

due to uncertainty. This, in turn, has a direct impact on equipment availability and maintenance costs. Unfortunately, due to misconceptions and due to uncertainty. This, in turn, has a direct impact on equipment availability and maintenance costs. Unfortunately, due to misconceptions and pressure from plant operators to get "back on line", it

More information

CHAPTER 10 RESTS AND PREPARATIONS. 4. Serve as a reference point for evaluating the fit of the framework to the teeth.

CHAPTER 10 RESTS AND PREPARATIONS. 4. Serve as a reference point for evaluating the fit of the framework to the teeth. CHAPTER 10 RESTS AND DEFINITIONS A REST is any rigid part of an RPD framework which contacts a properly prepared surface of a tooth. A REST PREPARATION or REST SEAT is any portion of a tooth or restoration

More information

There When You Need Them: 10 Principles of Successful RPD Treatment

There When You Need Them: 10 Principles of Successful RPD Treatment There When You Need Them: 10 Principles of Successful RPD Treatment Jeff Scott, DMD [email protected] 239 2 nd Ave South Suite 100 St. Petersburg, FL 33701 The West Coast District Dental

More information

SCREW THREADS C H A P T E R 17

SCREW THREADS C H A P T E R 17 C H A P T E R 17 SCREW THREADS Screw threads are of prime importance in machine drawing. It is a functional element used as temporary fasteners such as bolt, stud, nut and screw etc. These are constructed

More information

SCD Case Study. Treatment Considerations for Implant Rehabilitation

SCD Case Study. Treatment Considerations for Implant Rehabilitation SCD Case Study Treatment Considerations for Implant Rehabilitation Multiple surgical and restorative factors play a role in the treatment planning of implant restorations for the edentulous patient (Ali

More information

CHAPTER 12 SURVEY LINES. portion of the tooth is undercut to the path of placement of the denture. DEFINITIONS

CHAPTER 12 SURVEY LINES. portion of the tooth is undercut to the path of placement of the denture. DEFINITIONS CHAPTER 12 portion of the tooth is undercut to the path of placement of the denture. SURVEY LINES DEFINITIONS A SURVEY LINE is a line produced on a cast by a surveyor or scribe marking the greatest prominence

More information

NARROW DIAMETER implant

NARROW DIAMETER implant ND NARROW DIAMETER implant TABLE OF CONTENTS ND - NARROW DIAMETER implant Implant characteristics page 04 Dental implant page 05 Open Tray Impression Transfer page 06 Titanium Abutments page 07 O-Ball

More information

The new wide Ponto implant design clinical and surgical aspects

The new wide Ponto implant design clinical and surgical aspects The new wide Ponto implant design clinical and surgical aspects Partik Westerkull, MSc. Eng. Ph., Senior Research Consultant to Oticon Medical Lars Jinton, Director of Engineering, Oticon Medical With

More information

Ando A., Nakamura Y., Kanbara R., Kumano H., Miyata T., Masuda T., Ohno Y. and Tanaka Y.

Ando A., Nakamura Y., Kanbara R., Kumano H., Miyata T., Masuda T., Ohno Y. and Tanaka Y. 11. The Effect of Abutment Tooth Connection with Extracoronal Attachment using the Three Dimensional Finite Element Method - Part 2. The Construction of Finite Element Model from CT Data - Ando A., Nakamura

More information

CEEN 162 - Geotechnical Engineering Laboratory Session 7 - Direct Shear and Unconfined Compression Tests

CEEN 162 - Geotechnical Engineering Laboratory Session 7 - Direct Shear and Unconfined Compression Tests PURPOSE: The parameters of the shear strength relationship provide a means of evaluating the load carrying capacity of soils, stability of slopes, and pile capacity. The direct shear test is one of the

More information

Straumann Dental Implant System. Implant Selection Guide.

Straumann Dental Implant System. Implant Selection Guide. Straumann Dental Implant System. Implant Selection Guide. STRAUMANN's IMPLANT PORTFOLIO The Straumann Dental Implant System offers two implant lines with diverse body and neck designs ranging from the

More information

NOTCHES AND THEIR EFFECTS. Ali Fatemi - University of Toledo All Rights Reserved Chapter 7 Notches and Their Effects 1

NOTCHES AND THEIR EFFECTS. Ali Fatemi - University of Toledo All Rights Reserved Chapter 7 Notches and Their Effects 1 NOTCHES AND THEIR EFFECTS Ali Fatemi - University of Toledo All Rights Reserved Chapter 7 Notches and Their Effects 1 CHAPTER OUTLINE Background Stress/Strain Concentrations S-N Approach for Notched Members

More information

REHAB 442: Advanced Kinesiology and Biomechanics INTRODUCTION - TERMS & CONCEPTS

REHAB 442: Advanced Kinesiology and Biomechanics INTRODUCTION - TERMS & CONCEPTS Rehab 442: Introduction - Page 1 REHAB 442: Advanced Kinesiology and Biomechanics INTRODUCTION - TERMS & CONCEPTS Readings: Norkin & Levangie, Chapters 1 & 2 or Oatis, Ch. 1 & 2 (don't get too bogged down

More information

Implants in Orthodontics - A Litterature Rewiev

Implants in Orthodontics - A Litterature Rewiev Implants in Orthodontics - A Litterature Rewiev Thorir Schiöth University of Berne, Switzerland School of Dental Medicine Department of Orthodontics Implants in Orthodontics Published reports with specific

More information

Stress Strain Relationships

Stress Strain Relationships Stress Strain Relationships Tensile Testing One basic ingredient in the study of the mechanics of deformable bodies is the resistive properties of materials. These properties relate the stresses to the

More information

DESS. Screws. Tijuana Ventas: (664) 685 6294/95 [email protected]. For all major implant systems!! www.dabocorp.com

DESS. Screws. Tijuana Ventas: (664) 685 6294/95 hirambogarin@dabocorp.com. For all major implant systems!! www.dabocorp.com Screws Screws for definitive use. Made from medical grade 5 ELI Titanium Integrity and soundness guaranteed by stress tests of up to 150% of their nominal torques. Their designs ensure a perfect fit with

More information

Computer Aided Engineering (CAE) Techniques Applied To Hip Implant

Computer Aided Engineering (CAE) Techniques Applied To Hip Implant International Journal Of Computational Engineering Research (ijceronline.com) Vol. 3 Issue. 3 Computer Aided Engineering (CAE) Techniques Applied To Hip Implant 1, M. S. Abo_Elkhair, 2, M. E. Abo-Elnor,

More information

Healing Abutment Selection. Perio Implant Part I. Implant Surface Characteristics. Single Tooth Restorations. Credit and Thanks for Lecture Material

Healing Abutment Selection. Perio Implant Part I. Implant Surface Characteristics. Single Tooth Restorations. Credit and Thanks for Lecture Material Healing Abutment Selection Perio Implant Part I Credit and Thanks for Lecture Material Implant Surface Characteristics!CAPT Robert Taft!CAPT Greg Waskewicz!Periodontal Residents NPDS and UMN!Machined Titanium!Tiunite!Osseotite

More information

Mechanical Properties of Metals Mechanical Properties refers to the behavior of material when external forces are applied

Mechanical Properties of Metals Mechanical Properties refers to the behavior of material when external forces are applied Mechanical Properties of Metals Mechanical Properties refers to the behavior of material when external forces are applied Stress and strain fracture or engineering point of view: allows to predict the

More information

Lentur Cable System. Surgical Technique

Lentur Cable System. Surgical Technique Lentur Cable System Surgical Technique Contents Introduction... Page 1 System Design Features And Benefits... Page 2 Implants... Page 3 Instrumentation... Page 4 Surgical Technique... Page 5 Single Cable...

More information

Anterior Lumbar Interbody Fusion (ALIF). Instrument set supports placement of ALIF spacers using anterior or anterolateral approach.

Anterior Lumbar Interbody Fusion (ALIF). Instrument set supports placement of ALIF spacers using anterior or anterolateral approach. Anterior Lumbar Interbody Fusion (ALIF). Instrument set supports placement of ALIF spacers using anterior or anterolateral approach. Technique Guide Instruments and implants approved by the AO Foundation

More information

International Journal of Engineering Research-Online A Peer Reviewed International Journal Articles available online http://www.ijoer.

International Journal of Engineering Research-Online A Peer Reviewed International Journal Articles available online http://www.ijoer. RESEARCH ARTICLE ISSN: 2321-7758 DESIGN AND DEVELOPMENT OF A DYNAMOMETER FOR MEASURING THRUST AND TORQUE IN DRILLING APPLICATION SREEJITH C 1,MANU RAJ K R 2 1 PG Scholar, M.Tech Machine Design, Nehru College

More information

HUS-V Screw anchor. HUS-V Screw anchor. Basic loading data (for a single anchor) Mean ultimate resistance

HUS-V Screw anchor. HUS-V Screw anchor. Basic loading data (for a single anchor) Mean ultimate resistance HUS-V Screw anchor Anchor version HUS-V 8 / 10 Carbon steel concrete screw with hexagonal head Benefits - High productivity less drilling and fewer operations than with conventional anchors - Technical

More information

bi directional loading). Prototype ten story

bi directional loading). Prototype ten story NEESR SG: Behavior, Analysis and Design of Complex Wall Systems The laboratory testing presented here was conducted as part of a larger effort that employed laboratory testing and numerical simulation

More information

METHOD OF STATEMENT FOR STATIC LOADING TEST

METHOD OF STATEMENT FOR STATIC LOADING TEST Compression Test, METHOD OF STATEMENT FOR STATIC LOADING TEST Tension Test and Lateral Test According to the American Standards ASTM D1143 07, ASTM D3689 07, ASTM D3966 07 and Euro Codes EC7 Table of Contents

More information

B.TECH. (AEROSPACE ENGINEERING) PROGRAMME (BTAE) Term-End Examination December, 2011 BAS-010 : MACHINE DESIGN

B.TECH. (AEROSPACE ENGINEERING) PROGRAMME (BTAE) Term-End Examination December, 2011 BAS-010 : MACHINE DESIGN No. of Printed Pages : 7 BAS-01.0 B.TECH. (AEROSPACE ENGINEERING) PROGRAMME (BTAE) CV CA CV C:) O Term-End Examination December, 2011 BAS-010 : MACHINE DESIGN Time : 3 hours Maximum Marks : 70 Note : (1)

More information

Universal Crown and Bridge Preparation

Universal Crown and Bridge Preparation Universal Crown and Bridge Preparation The All-Ceramic Crown Preparation Technique for Predictable Success According to Dr. Ronald E. Goldstein Expect the Best. Buy Direct. The Universal * Crown and Bridge

More information

Solid Mechanics. Stress. What you ll learn: Motivation

Solid Mechanics. Stress. What you ll learn: Motivation Solid Mechanics Stress What you ll learn: What is stress? Why stress is important? What are normal and shear stresses? What is strain? Hooke s law (relationship between stress and strain) Stress strain

More information

Workshops & Courses. For Further Information and Registeration. Tel.:+966 12 640 2000 Ext. 22264 / 73061 / 21206. By Art House : 0503684163

Workshops & Courses. For Further Information and Registeration. Tel.:+966 12 640 2000 Ext. 22264 / 73061 / 21206. By Art House : 0503684163 Workshops & Courses By Art House : 0503684163 For Further Information and Registeration http://fdc.kau.edu.sa e-mail: [email protected] Tel.:+966 12 640 2000 Ext. 22264 / 73061 / 21206 Scan to Register

More information

The Mandibular Two-Implant Overdenture First-Choice. Standard of Care for the Edentulous Denture Patient

The Mandibular Two-Implant Overdenture First-Choice. Standard of Care for the Edentulous Denture Patient The Mandibular Two-Implant Overdenture First-Choice Standard of Care for the Edentulous Denture Patient Joseph R. Carpentieri, DDS Dennis P. Tarnow, DDS ii Preface Preface The prosthetic management of

More information

DUPONT PERFORMANCE POLYMERS

DUPONT PERFORMANCE POLYMERS DUPONT PERFORMANCE POLYMERS SELF TAPPING SCREWS: HOW TO CHOOSE THE RIGHT ONE Self-tapping screws provide an economical means of assembling components, especially where dissimilar materials must be joined

More information

LOAD-CARRYING CAPACITY OF AXIALLY LOADED RODS GLUED-IN PERPENDICULAR TO THE GRAIN

LOAD-CARRYING CAPACITY OF AXIALLY LOADED RODS GLUED-IN PERPENDICULAR TO THE GRAIN LOAD-CARRYING CAPACITY OF AXIALLY LOADED RODS GLUED-IN PERPENDICULAR TO TE GRAIN Prof. Dr.-Ing..J. Blaß, Dipl.-Ing. B. Laskewitz Universität Karlsruhe (T), Germany Abstract Glued-in rods have been used

More information

SPECIFICATIONS, LOADS, AND METHODS OF DESIGN

SPECIFICATIONS, LOADS, AND METHODS OF DESIGN CHAPTER Structural Steel Design LRFD Method Third Edition SPECIFICATIONS, LOADS, AND METHODS OF DESIGN A. J. Clark School of Engineering Department of Civil and Environmental Engineering Part II Structural

More information

DRAFTING MANUAL. Gears (Bevel and Hypoid) Drafting Practice

DRAFTING MANUAL. Gears (Bevel and Hypoid) Drafting Practice Page 1 1.0 General This section provides the basis for uniformity in engineering gears drawings and their technical data for gears with intersecting axes (bevel gears), and nonparallel, nonintersecting

More information

Saudi Fellowship In Dental Implant (SF-DI)

Saudi Fellowship In Dental Implant (SF-DI) Saudi Fellowship In Dental Implant (SF-DI) Prepared and Updated by Dr. Arwa AL-Sayed Consultant Periodontics and Dental Implants M E M B E R S Dr. Arwa AL-Sayed Dr. Abdulhadi Abanmy Dr. Ali AL-Ghamdi Dr.

More information

GIVE YOUR PATIENTS THE FREEDOM TO EAT, SPEAK AND LAUGH AGAIN.

GIVE YOUR PATIENTS THE FREEDOM TO EAT, SPEAK AND LAUGH AGAIN. Now Distributed By GIVE YOUR PATIENTS THE FREEDOM TO EAT, SPEAK AND LAUGH AGAIN. OPENING MINDS AND EXPANDING PRACTICE REVENUE OPPORTUNITIES Since the McGill Consensus in 2002, the dental industry has recognized

More information

RESTORING STRAUMANN IMPLANTS WITH LOCATOR ABUTMENTS

RESTORING STRAUMANN IMPLANTS WITH LOCATOR ABUTMENTS RESTORING STRAUMANN IMPLANTS WITH LOCATOR ABUTMENTS Straumann is the industrial partner of the ITI (International Team for Implantology) in the areas of research, development, and education. CONTENTS Product

More information

Technique Guide. Screw Removal Set. Instruments for removing Synthes screws.

Technique Guide. Screw Removal Set. Instruments for removing Synthes screws. Technique Guide Screw Removal Set. Instruments for removing Synthes screws. Table of Contents Introduction Screw Removal Set 2 Surgical Technique Preoperative Planning and Preparation 6 Removal of Intact

More information

The elements used in commercial codes can be classified in two basic categories:

The elements used in commercial codes can be classified in two basic categories: CHAPTER 3 Truss Element 3.1 Introduction The single most important concept in understanding FEA, is the basic understanding of various finite elements that we employ in an analysis. Elements are used for

More information

PERIPROSTHETIC IMPLANTS

PERIPROSTHETIC IMPLANTS PERIPROSTHETIC IMPLANTS PRODUCT OVERVIEW CLINICAL SOLUTIONS Periprosthetic fractures present unique challenges, such as how to gain fixation when the medullary canal is occupied. Special techniques and

More information

Supervisors: Dr. Farhan Raza Khan

Supervisors: Dr. Farhan Raza Khan 1 Presenter: Dr. Sana Ehsen Supervisors: Dr. Farhan Raza Khan 2 A dental implant (also known as an endosseous implant or fixture) is a surgical component that interfaces with the bone of the jaw to support

More information

MATERIALS AND METHODS

MATERIALS AND METHODS REDISTRIBUTION OF LOAD OF INJURED LOWER CERVICAL SPINE UNDER AXIAL COMPRESSION USING FEM E. C. Teo 1 and H. W. Ng 1 1 School of Mechanical and Production Engineering, Nanyang Technological University,

More information

Titanium versus Zirconium Implants

Titanium versus Zirconium Implants Titanium versus Zirconium Implants A dental materials comparison Dr. Gerald Regni, DMD!!" Christian St. Phila. PA 1213" T: (""") %&'-!"#" E: [email protected] Titanium versus Zirconium Implants

More information

BASIC ORTHODONTICS. And why we believe in THE TIP EDGE TECHNIQUE (Differential Straight Arch)

BASIC ORTHODONTICS. And why we believe in THE TIP EDGE TECHNIQUE (Differential Straight Arch) BASIC ORTHODONTICS And why we believe in THE TIP EDGE TECHNIQUE (Differential Straight Arch) MOVING TEETH Light force + Time = Tooth Movement Any technique of moving teeth involves light force and time.

More information

4-1-2005. Dental Clinical Criteria and Documentation Requirements

4-1-2005. Dental Clinical Criteria and Documentation Requirements 4-1-2005 Dental Clinical Criteria and Documentation Requirements Table of Contents Dental Clinical Criteria Cast Restorations and Veneer Procedures... Pages 1-3 Crown Repair... Page 3 Endodontic Procedures...

More information

LATERAL BONE EXPANSION FOR IMMEDIATE PLACEMENT OF ENDOSSEOUS DENTAL IMPLANTS

LATERAL BONE EXPANSION FOR IMMEDIATE PLACEMENT OF ENDOSSEOUS DENTAL IMPLANTS LATERAL BONE EXPANSION FOR IMMEDIATE PLACEMENT OF ENDOSSEOUS DENTAL IMPLANTS Department of Oral Maxillofacial Surgery, Chisinau Abstract: The study included 10 using the split control expansion technique

More information

Removable appliances II. Functional jaw orthopedics

Removable appliances II. Functional jaw orthopedics Removable appliances II. Functional jaw orthopedics Melinda Madléna DMD, PhD Associate professor Department of Pedodontics and Orthodontics Faculty of Dentistry Semmelweis University Budapest Classification

More information