Stress Distribution in Mandible Regulated by Bone and Dental Implant Parameters: Part I - Methodology
|
|
|
- Arron Horace Harvey
- 9 years ago
- Views:
Transcription
1 Stress Distribution in Mandible Regulated by Bone and Dental Implant Parameters: Part I - Methodology R. C. van Staden 1, H. Guan 1, Y. C. Loo 1, N. W. Johnson 2, N. Meredith 3 1 Griffith School of Engineering, Griffith University Gold Coast Campus, Australia 2 School of Dentistry and Oral Health, Griffith University Gold Coast Campus, Australia 3 Neoss Ltd, Harrogate, United Kingdom Abstract: The complicated inter-relationships between mandibular bone components and dental implants have attracted the attention of many a structural mechanics researcher as well as many a dental practitioner. This paper describes the methodology and analysis techniques employed to enable accurate evaluation of a vast range of the implant and bone parameters. The complex material and geometric properties of the bone and implant are modelled using two-dimensional (2D) triangular and quadrilateral plane strain elements. Assumptions made in the analysis include: (a) 50% osseointegration between bone and implant; (b) linear relationships exist between the stress value and the Youngs moduli of the cancellous and cortical bone at any specific point. In the companion paper (Part II) various bone, implant and loading parameters are evaluated for their influence on the stress distribution within the bone, in particular in the mandible. Keywords: bone, implant and loading parameters, finite element technique. 1 Introduction Dental implants are biocompatible screw-like titanium fixtures that are surgically placed into the jawbone to replace missing teeth. The mechanism by which an implant is bio-mechanically accepted by the jawbone is called osseointegration [1,2]. Stimulus of the bone through applied stresses has been well documented to influence the success or failure of an implant [3,4]. An optimum stress profile is required in order to maintain a strong and healthy jawbone: a stress that is too high may cause irreversible damage to the jawbone; one that is too low may fail to stimulate the jawbone sufficiently for satisfactory healing of the wound and thus, for osseointegration. Furthermore, the primary stability is especially critical because the bone is still in a state of repair and necessitates applied stresses that promote bone growth. Research by Himmlova et al. [5], Plikçiolu et al. [6] and Pierrisnard et al. [7] have all recognised the fact that the implant dimensions influence the magnitude and profile of stresses within the bone. It is commonly understood that increasing the implant length and/or diameter reduces the stresses within the bone. Furthermore, based on clinical experience practitioners appreciate the fact that if the bone is weak then a wider implant is required. However, these decisions are based primarily on clinical judgement but not supported by any theoretical data. It is critical for the practitioner to fully grasp the relationship between various combinations of bone, implant and loading parameters and the resulting stresses in the bone. 2 Modelling of Implant-Bone System Modelling and simulation are performed using the Strand7 Finite Element Analysis System [8]. The first step of the modelling is to define the bone and implant geometry. This is then followed by specifying the material behaviour in terms of the Youngs modulus, Poissons ratio and density for various mandibular bone components and the implant. After applying the appropriate loading and restraint conditions, the various parameters and their contributions to the stress profile within the bone, in particular in the mandible, can be evaluated. A 2D representation of the bone and implant is analysed because this is considered to be similarly accurate and more efficient in terms of computation time, as compared to a three dimensional equivalent. Data acquisition for the bone dimensions are carried out using Computed Tomography scanned images. The different types of bone, i.e. cancellous and cortical, are distinguished and the boundaries are identified in order to assign different material properties within the finite element model. Figure 1 shows a mandible section and implant with the loading and restraint conditions. Also shown in the figure are the detailed parameters considered in this study. The implant is cylindrical with 2 degrees of taperage and has a helical thread. For a particular finite element model with D = 4.5mm, L = 11mm, Tcor = 1.2mm, the total numbers of elements are respectively 2870 for the implant, 5956 for cancellous bone, and 1094 and for cortical bone. The total number of nodal points in the entire model is 9969.
2 Loading Conditions Horizontal Loading, FH = N Vertical Loading, FH = N Moment, M = Nmm F H M HH H 1 H 2 F V V 1 Tcor 2 Fixed Constraint Dental Implant Length, L = 7, 9, 11, 13, 15 mm Diameter, D = 3.5, 4.0, 4.5, 5.5 mm 0.2mm VV L Taper Cancellous Bone Young's Modulus, Ecan = 1, 2, 3, 4, 5, 6, 7, 8, 9, 10, 11, 12, 13, 14 GPa D V 2 Cortical Bone Thickness, Tcor = 0.3, 0.6, 0.9, 1.2, 1.5, 1.8, 2.1 mm Young's Modulus, Ecor = 7, 8, 9, 10, 11, 12, 13, 14, 15, 16, 17, 18, 19, 20 GPa Cross-Section in Distal Direction Figure 1: Finite element model of the mandible and dental implant As indicated in Figure 1, the von Mises stresses along the lines VV for cancellous bone and HH for cortical bone are recorded for all possible parameter combinations. The stress characteristics along these two lines are believed to be critical in evaluating the performance of implantation [9]. Note that both lines VV and HH are chosen on the right side of the implant because the highest stress magnitudes occur on the right. This is due to the loading characteristics. The start and finish points of VV (i.e. V 1 and V 2 ) and HH (i.e. H 1 and H 2 ) are also identified in Figure 1 for ease of discussion. A cross-sectional slice is taken from the mandible (refer to Figure 2). The arc length of the mandible is comparable to the width and depth of the slice. When the slice is subjected to in plane (x-y) masticatory forces (F H, F V and M), it is restrained from deforming out-of-plane (in the z-axis). Therefore, it can be assumed that all the strains are confined in the z-axis. To accurately represent the mechanical behaviour of the bone and implant, 3-node triangular (Tri3) and 4- node quadrilateral (Quad4) plane strain elements are therefore used for the construction of the finite element models. Note that for the plane strain elements each node has a complete set of s patial degrees of freedom including u and v. This means that each node can only translate in the x- (u) and y- (v) directions. Mandible 2D Slice M y x F H F v z Figure 2: Location of 2D slice in a mandible
3 3 Parameters 3.1 Jawbone An extensive literature review by van Staden et al. [10] indicated that the commonly assumed range of Youngs modulus values for cancellous bone is between 0.08 and 7.93GPa. For cortical bone it is normally assumed to vary from 5.57 to 22.8GPa. The present study adopts a range of Youngs modulus for cancellous bone (Ecan) from 1 to 14GPa and for cortical bone (Ecor) from 7 to 20GPa, as summarised in Figure 1. The selection of such a range is based on the understanding that in some exceptional cases the Youngs modulus of cancellous bone can go up to 14GPa when that of cortical bone is also at a high range. Assuming 1GPa intervals for these parameters results in fourteen different Youngs modulus values for cancellous and cortical bone respectively. With such a large range of material parameters to be considered, the corresponding densities of cancellous and cortical bone must be determined. Equations 1 and 2 show the mean relationships between the Youngs modulus and the density of cancellous ( can ) and cortical ( cor ) bone, respectively based on published works by Abendschein and Hyatt [11], Lotz et al. [12,13], Carter et al. [14] and Ciarelli et al. [15]. ρ can = ( Ecan) ) (1 ρ cor = ( Ecor ) (2 3 ) The cortical thickness (Tcor) is another parameter to be evaluated. Previous studies by Mellal et al. [16] assumed a thickness of 1mm while Natali et al. [17] assumed 0.8 and 1.9mm. In the present study a more extensive range of Tcor, viz 0.3, 0.6, 0.9, 1.2, 1.5, 1.8 and 2.1mm is selected, see Figure 1. The Poissons ratios, v, for the implant (grade 4 commercially pure titanium), cancellous and cortical bone are 0.3, 0.3 and 0.35 respectively. Note also that the cortical bone is constrained along the left and right faces of the cross-section in the distal direction (see Figure 1) thus representing a realistic function of the mandible with an implant. 3.2 Implant and loading The dimensions of the Neoss [18] implant system are employed for the purpose of this study. Various dimensions are considered including lengths of 7, 9, 11, 13 and 15mm, and diameters of 3.5, 4.0, 4.5 and 5.5mm. This study only focuses on the stresses within the cancellous and cortical bone therefore the entire implant system has been simplified to being solely the implant, excluding the abutment, abutment screw and crown. In reality, the masticatory force is acting on the top of the crown. Neglecting the implant components above the implant itself means that the masticatory force is transferred to the head of the implant. This requires the introduction of a moment about the z- axis along with the forces acting in the vertical (y-axis), Fv, and horizontal (xaxis), F H, directions. Force, F M (to produce M) Rigid Link F M (N) F M F M y-axis (v) z Plane Strain Element Node x-axis (u) z-axis (w) Moment (located at the centre of the implant top) Figure 3: Moment applied to plane strain element
4 The F H, F V and M of 2 5 to 250N, 50 to 500N and to 1625Nmm respectively are applied to the head of the implant, as shown in Figure 1, to simulate the masticatory forces. Note that M is applied in the negative z-axis and is based on the horizontal load laterally transferred at a distance (crown height) of 6.5mm. Based on the clinical study by Aparicio and Orozco [19], these loading conditions are considered comprehensive. In this study, F H /F V /M combinations are chosen from the minimum values 25N/50N/162.5Nmm to the maximum values 250N/500N/1625Nmm together with half maximum values 125N/250N/812.5Nmm. Such a load combination and increment are used for all the parametric studies. Due to the z-axis not being considered for the 2D plane strain elements, a moment must be applied through the use of rigid links to nodal points surrounding a specified node (N) at which the moment is applied (refer to Figure 3). At the end points of each rigid link, an orthogonal force F M is applied thus creating a moment around the connecting point (N) of the rigid links. 4 Assumptions In addition to assuming a 2D representation of the bone and implant, other assumptions are: (a) 50% osseointegration between the bone and implant;,(b) linear relationships exist between the stress value and Youngs moduli of the cancellous and cortical bone at any specific point. 4.1 Bone implant interface The interface surrounding the implant exhibits both blood and bone fragments. This interface is considered clinically ideal for osseointegration. However it is mechanically unfavourable. The stage after insertion and before full osseointegration occurs is critical for the surrounding bone because it exhibits the most distinct stress concentrations. However post osseointegration, when the lamina dura is formed, the implant stability drastically increases. In a finite element study by Natali et al. [17] two formation stages of the lamina dura were modelled by having Youngs moduli of 0.3 and 1.5GPa, respectively. The formation of lamina dura improved the stress distribution more than cancellous bone and is a clinically favourable course of healing. A study by Berglundh et al. [20] assumed that the blood interface extended 0.13mm into the thread chamber. Based on these findings the present study assumes a 50% osseointegration between the bone and implant. The voids between the bone and implant are modelled as blood, see Figure 4. The Youngs modulus and Poissons ratio of the blood are taken as 0.7GPa and 0.3 respectively. 4.2 Linear relationship between stress and Youngs modulus The possible number of parameter combinations for the bone, implant and loading exceeds 82,320. This number is based on five implant lengths, four implant diameters, fourteen Youngs moduli each for cancellous and cortical bone, seven cortical thicknesses and three loading conditions (see Figure 1). It would be unviable to complete such a huge number of simulations. As a result, further assumptions are made to reduce the number of simulations hence yielding a manageable amount of data for analysis Predicted and actual stresses for Ecan and Ecor Implant Hard Tissue (cancellous bone) Blood 0.13mm Figure 4: Proposed modelling of blood and hard tissue The feasibility of a linear relationship for the stress variation for all combinations of Ecan and Ecor is exploited. In Figure 5 a methodology for predicting the linear relationship between the stress and the Youngs modulus is shown. The actual stress values at any specific point along either VV or HH are recorded at the four corners representing the combinations of the minimum and maximum values of Ecan
5 and Ecor. Between any two actual stress values where either Ecan or Ecor is constant, a linear relationship between the stress and Ecan (or Ecor) is predicted. This prediction is then extended to any combination of Ecan and Ecor, provided that one of them remains constant. Ecan = 1GPa Ecor = 20GPa Actual stress values Ecan = 14GPa Ecor = 20GPa Predicted stress values 14 Combinations of Ecan and Ecor Ecan = 1GPa Ecor = 7GPa Actual stress values Ecan = 14GPa Ecor = 7GPa Figure 5: Methodology for predicting the linear relationship For the purpose of validation, a finite element analysis is carried out to obtain the actual von Mises stress values for L = 11 mm, D = 4.5mm, Tcor = 1.2mm, F V = 500N, F H = 250N and M = 1625Nmm. Varying Ecan from 1 to 14GPa for every possible value of Ecor shows little variation from linearity [9]. The actual stress at mid-point of line VV is recorded for all combinations of Ecan and Ecor and compared to the respective linear prediction. For Ecan = 7GPa and Ecor = 13GPa, the predicted stresses vary from the actual ones by an average of 5.7% and 4.3% respectively. The stresses recorded at mid-point of line HH produced similar percentage errors (4.9% and 6.5% for Ecan and Ecor respectively) to VV Parameter combinations With the assumption of linear relationships between the stress and Ecan or Ecor, only the minimum and maximum values of Youngs modulus are considered for each type of bone instead of the initial data set of fourteen. Parameters summarised in Figure 6 are thus evaluated. This means that the total number of analyses is reduced to D (mm) L (mm) Tcor (mm) Ecor (GPa) Ecan (GPa) F V, F H (N) and M (Nmm) 50, 25 and , 125 and , 250 and Discussion and Conclusion Figure 6: Parameter combinations This paper presents the methodology and analysis techniques that are used in the companion paper (Part II) to evaluate the stress distribution characteristics in the bone due to the effect of various parameters. Realistic geometries, material properties, loading and support conditions for the jawbone and implant are taken into consideration. Modelling assumptions in terms of (a) 50% osseointegration between bone and implant; (b) linear relationships between the stress value and the Youngs moduli of the cancellous and cortical bone at any specific point, are discussed in some detail in this paper.
6 The parametric analysis covering the influence of varying each parameter on the stress characteristics in the mandible is presented in the companion paper (Part II). 6 References [1] Branemark PI, Adell R, Breine U, Hansson BO, Lindstrom J, Ohlsson A. Intra-osseous anchorage of dental prostheses. I. Experimental studies. Scand J Plast Reconstr Surg. 1969;3(2): [2] Branemark PI, Hansson BO, Adell R, Breine U, Lindstrom J, Hallen O, Ohman A. Osseointegrated implants in the treatment of the edentulous jaw. Experience from a 10-year period. Scand J Plast Reconstr Surg. 1977;16: [3] Mellal A, Wiskott HW, Botsis J, Scherrer SS, Belser UC. Stimulating effect of implant loading on surrounding bone. Comparison of three numerical models and validation by in vivo data. Clin Oral Implants Res. 2004;15(2): [4] Carter DR, Orr TE, Fyhrie DP, Schurman DJ. Influences of mechanical stress on prenatal and postnatal skeletal development. Clin Orthop Relat Res. 1987;(219): [5] Himmlova L, Dostalova T, Kacovsky A, Konvickova S. Influence of implant length and diameter on stress distribution: a finite element analysis. J Prosthet Dent. 2004;91(1): [6] Plikçiolu H, Akça K. Comparative evaluation of the effect of diameter, length and number of implants supporting three-unit fixed partial prostheses on stress distribution in the bone. J Dent. 2002;30(1): [7] Pierrisnard L, Renouard F, Renault P, Barquins M. Influence of implant length and bicortical anchorage on implant stress distribution. Clin Implant Dent Relat Res. 2003;5(4): [8] Strand7 Pty Ltd (2004), Strand7 Theoretical Manual, Sydney, Australia. [9] van Staden RC, Guan H, Loo YC. Johnson NW, Ivanovski S, Meredith N. Influence of Bone and Dental Implant Parameters on Stress Distribution in Mandible A Finite Element Study. Int J Oral Maxillofac Implants. (Submitted). [10] van Staden RC, Guan H, Loo YC. Application of the finite element method in dental implant research. Comput Methods Biomech Biomed Engin. 2006;9(4): [11] Abendschein W, Hyatt GW. Ultrasonics and selected physical properties of bone. Clin Orthop Relat Res. 1970;69: [12] Lotz JC, Gerhart TN, Hayes WC. Mechanical properties of trabecular bone from the proximal femur: a quantitative CT study. J Comput Assist Tomogr. 1990;14(1): [13] Lotz JC, Gerhart TN, Hayes WC. Mechanical properties of metaphyseal bone in the proximal femur. J Biomech. 1991;24(5): [14] Carter DR, Schwab GH, Spengler DM. Tensile fracture of cancellous bone. Acta Orthop Scand. 1980;51(5): [15] Ciarelli MJ, Goldstein SA, Kuhn JL, Cody DD, Brown MB. Evaluation of orthogonal mechanical properties and density of human trabecular bone from the major metaphyseal regions with materials testing and computed tomography. J Orthop Res. 1991;9(5): [16] Mellal A, Wiskott HW, Botsis J, Scherrer SS, Belser UC. Stimulating effect of implant loading on surrounding bone. Comparison of three numerical models and validation by in vivo data. Clin Oral Implants Res. 2004;15(2): [17] Natali AN, Pavan PG, Ruggero AL. Analysis of bone-implant interaction phenomena by using a numerical approach. Clin Oral Implants Res. 2006;17(1): [18] Neoss Limited (2006), Neoss Implant System Surgical Guidelines, UK. [19] Aparicio C, Orozco P. Use of 5-mm-diameter implants: Periotest values related to a clinical and radiographic evaluation. Clin Oral Implants Res. 1998;9(6): [20] Berglundh T, Abrahamsson I, Lang NP, Lindhe J. De novo alveolar bone formation adjacent to endosseous implants. Clin Oral Implants Res. 2003;14(3): Stress Distribution in Mandible Regulated by Bone and Dental Implant Parameters: Part II Parametric Analysis R. C. van Staden 1, H. Guan 1, Y. C. Loo 1, N. W. Johnson 2, N. Meredith 3 1 Griffith School of Engineering, Griffith University Gold Coast Campus, Australia 2 School of Dentistry and Oral Health, Griffith University Gold Coast Campus, Australia 3 Neoss Ltd, Harrogate, United Kingdom
7 Abstract: In this paper the finite element procedure is used to evaluate various bone, implant and loading parameters for their influence on the stress distribution within the bone, in particular in the mandible. The methodology and analysis techniques employed to enable accurate modelling of the implant and bone systems are discussed in the companion paper (Part I). The analysis results show that an increase in the Youngs moduli for the cancellous and cortical bone yields elevated stress magnitudes in the bone. Thinner cortical bone results in elevated stress levels within itself as well as in the cancellous bone. In both cancellous and cortical bone, the implant length demonstrates to be more influential, in terms of differences between the minimum and maximum stress values, as compared to the diameter. The masticatory forces exhibit a more significant influence on the stress than all the other parameters. Therefore loading should be considered an imperative factor when planning an i mplant placement. Keywords: bone, implant and loading parameters, finite element technique. 1 Introduction The Finite Element Method (FEM) is becoming widely used to advance dental technologies [1,2]. The use of the FEM allows for an improved understanding of stresses within the implant and bone systems. Areas of application in dental implantology include studies on jawbone and implant properties as well as the bone implant interface [3-7]. Using the finite element technique the stress profile can be evaluated within both cancellous and cortical bone. This stress profile can be validated with the stress-strain behaviour of both cancellous and cortical as defined by Burstein et al. [8] and Currey [9]. To date no published research appears to have investigated the stress profile in the bone when various combinations of bone, implant and loading parameters are considered. The outcome of this study may help dental practitioners to identify the stimulus state of the bone - hence, they would be able to predict success or failure for all combinations of implant length and diameter with given cortical bone thickness, and material properties of cancellous and cortical bone, under a range of masticatory forces. 2 Parametric Analysis As discussed in the companion paper (Part I), modelling and simulation of a total of 1680 parametric combinations are performed using the Strand7 Finite Element Analysis System [10]. The first subsection includes a summary of the von Mises stress profiles, recorded along the lines VV and HH for cancellous and cortical bone respectively, where average values are chosen for the remaining parameters except for the masticatory forces. The results due to three combinations of the masticatory forces are presented, viz the minimum values of 25N/50N/162.5Nmm, the maximum values of 250N/500N/1625Nmm and the median values of 125N/250N/812.5 Nmm. For example when presenting the stress for different values of L, the Ecan, Ecor, D and Tcor are all set to their mean values. Note that for this subsection, when F V, F H and M are varied all other parameters are set to their mean values. The sec ond subsection summarises the most and least influential parameters where all the results presented correspond to the maximum masticatory forces. 2.1 Influence of varying each parameter Figures 1 (a) and (b) present the von Mises stress recorded along the lines VV and HH for all the possible values of D. Other parameters are set to their average values: L = 11mm, Tcor = 1.2mm, Ecan = 7GPa and Ecor = 13GPa. When D increases the stress along the line VV oscillates more significantly following the thread profile (Figure 1 (a)). Such an increased oscillation is a result of reduced distance between the implant thread and cortical bone and the subsequent reduced volume of cancellous bone. This reduces the capacity of cancellous bone in distributing the stress more evenly. The stresses along the line HH exhibit a reduction when D is increased (Figure 1 (b)). This reduction is caused by the larger diameter i mplant supporting an increased portion of the load thereby reducing the stress in the cortical bone. Generally higher stresses are found at the implant neck for all values of D. Note that the stress levels along the lines VV and HH increase as F H, F V and M increase, which is evident in Figures 1 (a) and (b).
8 Varying diameter, D Varying length, L (a) (b) (c) Varying cortical thickness, Tcor (d) (e) (f)
9 Varying Youngs Modulus of cancellous bone, Ecan (g) Varying Youngs Modulus of cortical bone, Ecor (h) (i) Figure 1: Stress profiles with varying parameters The von Mises stress recorded along the lines VV and HH, for all possible implant lengths, are shown in Figures 1 (c) and (d). The stresses decrease within both cancellous and cortical bone (Figures 1 (c) and (d) respectively) as L increases because the surface area contact between the bone and implant also increases leading to the implant absorbing more load. Figures 1 (e) and (f) present the von Mises stress recorded along the lines VV and HH for all possible values of Tcor. As Tcor decreases, the ability of the cortical bone in supporting the load also decreases leading to a slight increase in the magnitude of stresses along the line VV (Figure 1 (e)). No discernable trend is found for the stress profiles along the line HH, as shown in Figure 1 (f). With reference to Figure 1 (g), as Ecan increases the stress magnitude increases. This is due to the cancellous being able to support a greater portion of the load. The stress within the cortical bone, Figure 1 (h), increases as Ecan decreases. This is because the cancellous bone is supporting less load and therefore an increased portion of the load has to be supported by the cortical bone. As evident in Figure 1 (i), as Ecor decreases the stress magnitude increases slightly in the cancellous bone, which is due to the cancellous having to support an increased portion of the load. However the stress within the cortical bone, Figure 1 (j), increases as Ecor increases. This is due to the cortical bone having increased resistance to the load. (j)
10 2.2 Parameter ranking To identify the most and least influential parameters, all the six parameters given in Figure 6 of the companion paper (Part I) are ranked in this subsection. Based on the data shown in Figure 1, the average differences (AD) are analysed and discussed, where AD is defined as the difference between the stresses when a single parameter is set to its minimum and maximum values while all the other parameters are set to their mean values. The differences in stress at distances 1, 4.5, 9mm from V 1 and 0.2, 0.6, 1mm from H 1 along the lines VV and HH respectively (see Figure 1 of companion paper), are given in Table 1. However, for D = 3.5mm the differences in stress along the line HH are recorded at distances 0.2, 0.4 and 0.8mm from H 1. Note that F-M represents F H, F V and M by presenting the average stress difference that result from varying each load from minimum to maximum. Table 1. von Mises stress (MPa) in cancellous and cortical bone (stress along the line HH shown in brackets) VV/HH (mm) D L Tcor Ecan Ecor F-M 1 / , (16.86) 16.82, (21.92) 4.25, (53.29) 26.12, (66.80) 1.85, (52.98) 41.70, (87.80) 4.5 / , (25.70) 18.93, (18.19) 4.26, (16.24) 6.44, (81.33) 1.87, (55.80) 27.95, (83.27) 9 / , (22.27) 20.33, (15.17) 8.70, (3.06) 6.14, (52.02) 2.81, (41.58) 30.15, (69.66) AD 3.00, (21.61) 18.69, (18.43) 5.74, (24.20) 12.90, (66.72) 2.18, (50.12) 33.27, (80.24) When considering the AD in stresses along the line VV, the ranking order is F-M>L>Ecan>Tcor>D>Ecor. Where > indicates greater difference in stress than the next parameter. The applied loading (F-M) has a more significant influence on the stress differences than all other parameters. However, Ecan demonstrated differences in stress at the distance 1mm along the line VV that are closely matched by that of L. The Youngs modulus of cortical bone (Ecor) exhibits the lowest AD, which is especially evident at a distance of 1mm along the line VV. The ranking order for the AD in stress along the line HH is F-M>Ecan>Ecor>Tcor>D>L. As found for cancellous bone varying F-M has a major influence on the stress differences within the cortical bone. Varying Tcor, especially at 0.6mm along the line HH, has the least influence on the differences. 3 Stress Validation Very similar to Curreys experimental results [9], the uniaxial compressive stress values for the human femora and tibiae measured by Burstein et al. [8] (Figure 2) are used as a reference to determine which parameters lead to either cancellous or cortical bone fracturing. Note also in this study that the compressive stresses on the right side of the implant are obtained under biaxial stress condition. Therefore a discrepancy is 200 Cortical Bone expected when the present results are compared to the uniaxial compressive stresses obtained by Burstein et al. [8]. 100 Fracture Point Derived from the stress-strain curve at the fracturing point are the uniaxial compressive stresses of 40MPa and 190MPa, for the cancellous and cortical bone respectively. Shown in Table 2 are the absolute maximum biaxial compressive stresses recorded along the lines VV and HH that are found close to the implant neck. Figure 3 presents the compressive stress profiles within the cancellous and cortical bone. Note that, for the data shown in Table 2 and Figure 3, all other parameters are set to their average values, with maximum loading combinations (F H /F V /M = 250N/500N/1625Nmm). Table 2. Absolute maximum compressive stresses (MPa) in cancellous and cortical bone Parameter Cancellous Bone D (mm) L (mm) Tcor (GPa) Ecan (GPa) Ecor (GPa) Line VV HH Stress (MPa) Strain (%) Figure 2: Stress-strain behaviour (Burstein et al. [10])
11 (a) Cancellous bone (b) Cortical bone Figure 3: Absolute maximum compressive stress profiles in cancellous and cortical bone For the majority of parameters the fracture stress of cancellous bone (40MPa) is exceeded (refer to Figures 3 (a) and (b)) close to the implant head along the line VV, with D exhibiting the highest stress out of all the parameters considered (refer to Table 2). However, the fracture stress for cortical bone is not exceeded for any of the parameters. O'Mahony et al. [6] found that the greatest compressive stresses in the cancellous and cortical bone were 20 and 115MPa respectively. Discrepancies between the results from the present study and those of OMahony et al. could be due to the different assumptions adopted for loading, modelling and restraint conditions: they assumed a 6 mm off-axial vertical load of 490N. On the other hand, the current study assumes F H, F V and M of 250N, 500N and 1625Nmm, respectively. Also OMahony et al. used an implant with no thread and they applied fixed constraints to the base of a two dimensional mandible section. 4 Discussion and Conclusion A good agreement is found with previous research [11-17] which stated that an increase in implant length and diameter leads to a reduction in stress magnitudes within the cancellous bone. Note in this study however, when D increases the stress along the line VV oscillates more significantly following the thread profile (Figure 1 (a)). A statement by Himmlova et al. [18] and Tawil et al. [16] indicates that the implant diameter (D) is more important for improved stress distribution than the implant length (L). However, it is shown herein that an increase in length reduces the stress, within both the cancellous and cortical bone, for a wider range of parameters. This is more obvious than the influence of varying diameter. In contrast, Pierrisnard et al. [12] found that the stress within the bone was virtually constant, independent of implant length and bicortical anchorage. No research to date has considered the influence that the cortical thickness (Tcor) has on the stress in the cancellous and cortical bone. In this study it is concluded that decreased Tcor leads to cortica l bone supporting less load leading to a slight increase in the magnitude of stresses in the cancellous bone. Generally, as the strength of either cancellous (Ecan) or cortical (Ecor) bone increases, their ability to carry the load increases. This is evident when considering the stress along the line VV, as when Ecan increases the stress magnitude increases, which is due to the cancellous being able to support more load. In addition and as expected, when decreasing Ecor the stress magnitude increases within the cancellous bone, which is due to the cancellous having to support a greater portion of the load. The stress within the cortical bone increases as Ecor increases, which is due to the cortical bone offering more resistance to the load. It is found common during this study that e levated stress levels occur at the implant neck for all the parameter combinations. This phenomenon is supported by the study of Meijer et al. [5] which concluded that the regions of bone exposed to maximum stress are located around the implant neck. Note that the present study also found
12 that the stress level at the implant neck, within the cancellous bone, is influenced by the fact t hat elevated stress magnitudes are transferred through its contact with the cortical bone. In general, the implant length has a noteworthy characteristic relationship with the stress i n the bone. The length is more influential, in terms of the stress difference in the cancellous bone, than the diameter. However the diameters is more influential in the stress difference in the cortical bone. Overall, in both cancellous and cortical bone the applied load and moment are most influential and therefore it is considered an imperative factor when planning an implant placement. When comparing the absolute maximum compressive stress values within both cancellous and cortical bone with published stress-strain data it is found that the cancellous bone experiences fracturing for all parameter combinations when maximum loading is applied. However, this is not the case when the loading is set to a magnitude that is half of the maximum. For the cortical bone the fracture point is never reached for all parameter combinations even when subjected to maximum loading. It is anticipated that these new findings will assist the clinician to perform a patient specific implant treatment in a more quality-controlled manner. Evaluating the influence of other parameters which affect the stress characteristics that govern osseointegration and bone growth, may constitute the main aim of a future study. These parameters include implant taperage, pitch and design of implant thread, implant neck offset, different percentages of osseointegration and implant orientation within the bone. Acknowledgements A special thank you goes to John Divitini and Fredrik Engman from Neoss Limited for their technical advice. 5 References [21] DeTolla DH, Andreana S, Patra A, Buhite, R, Comella B. Role of the finite element model in dental implants. J Oral Implantol. 2000;26(2): [22] Geng JP, Ma QS, Xu W, T an KB, Liu GR. Finite element analysis of four thread-form configurations in a s tepped screw implant. J Oral Rehabil. 2004;31(3): [23] Menicucci G, Mossolov A, Mozzati M, Lorenzetti M, Preti G. Tooth-implant connection: some biomechanical aspects based on finite element analyses. Clin Oral Implants Res. 2002;13(3): [24] Canay S, Hersek N, Akpinar I, Asik Z. Comparison of s tress distribution around vertical and angled implants with finiteelement analysis. Quintessence Int. 1996;27(9): [25] Meijer HJ, Starmans FJ, Steen WH, Bosman F. Loading conditions of endosseous implants in an edentulous human mandible: a three -dimensional, finite -element study. J Oral Re habil. 1996;23: [26] O'Mahony A, Bowles Q, Woolsey G, Robinson SJ, Spencer P. Stress distribution in the single-unit osseointegrated dental implant: finite element analyses of axial and off -axial loading. Implant Dent. 2000;9 (3): [27] Patra AK, DePaolo JM, D'Souza KS, DeTolla D, Meenaghan MA. Guidelines for analysis and redesign of dental implants. Implant Dent. 1998;7(4): [28] Burstein AH, Reilly DT, Martens M. Aging of bone tissue: mechanical properties. J Bone Joint Surg Am. 1976;58(1): [29] Currey JD. The Mechanical Adaptation of Bones. New Jersey: Princeton University Press; 1984:3-36. [30] Strand7 Pty Ltd (2004), Strand7 Theoretical Manual, Sydney, Australia. [31] Plikçiolu H, Akça K. Comparative evaluation of the effect of d iameter, length and number of implants supporting threeunit fixed partial prostheses on stress distribution in the bone. J Dent. 2002;30(1): [32] Pierrisnard L, Renouard F, Renault P, Barquins M. Influence of implant length and bicortical anchorage on implant stress distribution. Clin Implant Dent Relat Res. 2003;5(4): [33] Aparicio C, Orozco P. Use of 5 -mm-diameter implants: Periotest values related to a clinical and radiographic evaluation. Clin Oral Implants Res. 1998;9(6): [34] Iplikcioglu H, Akca K. Comparative evaluation of the effect of di ameter, length and number of implants supporting three-unit fixed partial prostheses on stress distribution in the bone. J Dent. 2002;30(1): [35] Misch CE. Implant design considerations for the posterior regions of the mouth. Implant Dent. 1999;8(4): [36] Tawil G, Aboujaoude N, Younan R. Influence of prosthetic parameters on the survival and complication rates of short implants. Int J Oral Maxillofac Implants. 2006;21(2): [37] Zhang L, Zhou Y, Meng W. A three-dimensional finite element analysis of the correlation between lengths and diameters of the implants of fixed bridges with proper stress distribution. Hua Xi Kou Qiang Yi Xue Za Zhi. 2000;18(4): [38] Himmlova L, Dostalova T, Kacovsky A, Konvickova S. Influence of implant length and diameter on stress distribution: a finite element analysis. J Prosthet Dent. 2004;91(1):20-25.
Influence of Bone and Dental Implant Parameters on Stress Distribution in the Mandible: A Finite Element Study
Influence of Bone and Dental Parameters on Stress Distribution in the Mandible: A Finite Element Study Hong Guan, PhD 1 /Rudi van Staden, BEng 2 /Yew-Chaye Loo, PhD 3 / Newell Johnson, PhD 4 /Saso Ivanovski,
Stress distribution characteristics in dental implant influenced by its wall thickness
Stress distribution characteristics in dental implant influenced b its wall thickness R.C. van Staden, H. Guan & Y.C. Loo Griffith School of Engineering, Griffith Universit, Gold Coast Campus, Queensland
A REVIEW ON APPLICATION OF CAD AND FEM TECHNOLOGY IN DESIGN OF TAPER DENTAL IMPLANT
A REVIEW ON APPLICATION OF CAD AND FEM TECHNOLOGY IN DESIGN OF TAPER DENTAL IMPLANT Dhiraj Agnihotri 1, Amol More 2, Dr. P.G.Patil 3 M.Tech CAD/CAM, Department of Mechanical Engineering, G.H.R.C.E,, Nagpur,
Stress Evaluation of Dental Implant Wall Thickness using Numerical Techniques
Stress Evaluation of Dental Implant Wall Thickness using Numerical Techniques Rudi C. van Staden 1, Hong Guan 1, Yew-Chae Loo 1, Newell W. Johnson 2, Neil Meredith 3, 1 Grif th School of Engineering, Grif
A THREE-DIMENSIONAL FINITE ELEMENT ANALYSIS OF MANDIBULAR OVERDENTURE SUPPORTED BY CYLINDRICAL AND CONICAL IMPLANTS
A THREE-DIMENSIONAL FINITE ELEMENT ANALYSIS OF MANDIBULAR OVERDENTURE SUPPORTED BY CYLINDRICAL AND CONICAL IMPLANTS Evandro Afonso Sartori, [email protected] Josué Ricardo Broilo Luiz Oscar
Ridge Reconstruction for Implant Placement
Volume 1, No. 5 July/August 2009 The Journal of Implant & Advanced Clinical Dentistry Ridge Reconstruction for Implant Placement 2 Hours of CE Credit Oral Implications of Cancer Chemotherapy Immediate
Cumhuriyet Dental Journal
Cumhuriyet Dental Journal Volume 18 Issue 1 doi: 10.7126/cdj.58140.5000037693 available at http://dergipark.ulakbim.gov.tr/cumudj/ RESEARCH ARTICLE Numerical analysis of the effect of implant geometry
Ando A., Nakamura Y., Kanbara R., Kumano H., Miyata T., Masuda T., Ohno Y. and Tanaka Y.
11. The Effect of Abutment Tooth Connection with Extracoronal Attachment using the Three Dimensional Finite Element Method - Part 2. The Construction of Finite Element Model from CT Data - Ando A., Nakamura
Application of finite element method in dental implant research
Application of finite element method in dental implant research R. C. VAN STADEN, H. GUAN, Y. C. LOO School of Engineering, Griffith University Gold Coast Campus, Australia Address correspondence to Dr
Lap Fillet Weld Calculations and FEA Techniques
Lap Fillet Weld Calculations and FEA Techniques By: MS.ME Ahmad A. Abbas Sr. Analysis Engineer [email protected] www.advancedcae.com Sunday, July 11, 2010 Advanced CAE All contents Copyright
The mechanical behavior of the bone microstructure around dental implants
BUDAPEST UNIVERSITY OF TECHNOLOGY AND ECONOMICS DEPARTMENT OF STRUCTURAL MECHANICS The mechanical behavior of the bone microstructure around dental implants Summary and Theses of PhD dissertation ÉVA LAKATOS
LATERAL BONE EXPANSION FOR IMMEDIATE PLACEMENT OF ENDOSSEOUS DENTAL IMPLANTS
LATERAL BONE EXPANSION FOR IMMEDIATE PLACEMENT OF ENDOSSEOUS DENTAL IMPLANTS Department of Oral Maxillofacial Surgery, Chisinau Abstract: The study included 10 using the split control expansion technique
Life Table Analysis for Evaluating Curative-effect of One-stage Non-submerged Dental Implant in Taiwan
Journal of Data Science 6(2008), 591-599 Life Table Analysis for Evaluating Curative-effect of One-stage Non-submerged Dental Implant in Taiwan Miin-Jye Wen 1, Chuen-Chyi Tseng 2 and Cheng K. Lee 3 1 National
Oftentimes, as implant surgeons, we are
CLINICAL AVOIDING INJURY TO THE INFERIOR ALVEOLAR NERVE BY ROUTINE USE OF INTRAOPERATIVE RADIOGRAPHS DURING IMPLANT PLACEMENT Jeffrey Burstein, DDS, MD; Chris Mastin, DMD; Bach Le, DDS, MD Injury to the
Computer Aided Engineering (CAE) Techniques Applied To Hip Implant
International Journal Of Computational Engineering Research (ijceronline.com) Vol. 3 Issue. 3 Computer Aided Engineering (CAE) Techniques Applied To Hip Implant 1, M. S. Abo_Elkhair, 2, M. E. Abo-Elnor,
BIOMECHANICAL ANALYSIS OF NORMAL AND IMPLANTED TOOTH USING BITING FORCE MEASUREMENT
BIOMECHANICAL ANALYSIS OF NORMAL AND IMPLANTED TOOTH USING BITING FORCE MEASUREMENT B K Biswas 1 S Bag 2 & S Pal 3 1. Dept. of Dental Surgery, Associate Professor, KPC Medical College & Hospital, Kolkata,
Ideal treatment of the impaired
RESEARCH IMPLANTS AS ANCHORAGE IN ORTHODONTICS: ACLINICAL CASE REPORT Dale B. Herrero, DDS KEY WORDS External anchorage Pneumatized Often, in dental reconstruction, orthodontics is required for either
Numerical Analysis of Independent Wire Strand Core (IWSC) Wire Rope
Numerical Analysis of Independent Wire Strand Core (IWSC) Wire Rope Rakesh Sidharthan 1 Gnanavel B K 2 Assistant professor Mechanical, Department Professor, Mechanical Department, Gojan engineering college,
Stress and deformation of offshore piles under structural and wave loading
Stress and deformation of offshore piles under structural and wave loading J. A. Eicher, H. Guan, and D. S. Jeng # School of Engineering, Griffith University, Gold Coast Campus, PMB 50 Gold Coast Mail
BENDABLE VS. ANGULATED DENTAL IMPLANTS: CONSIDERATION OF ELASTIC AND PLASTIC MATERIAL PROPERTIES BASED ON EXPERIMENTAL IMPLANT MATERIAL DATA AND FEA
Biomed Pap Med Fac Univ Palacky Olomouc Czech Repub. 2008, 152(2):309 316. T. Goldmann, S. Ihde, J. Kuzelka, L. Himmlova 309 BENDABLE VS. ANGULATED DENTAL IMPLANTS: CONSIDERATION OF ELASTIC AND PLASTIC
Outcomes of Placing Short Dental Implants in the Posterior Mandible: A Retrospective Study of 124 Cases
DENTAL IMPLANTS J Oral Maxillofac Surg 67:713-717, 2009 Outcomes of Placing Short Dental in the Posterior Mandible: A Retrospective Study of 124 Cases Bao-Thy N. Grant, DDS,* Franklin X. Pancko, DDS, and
Optimum Selection of the Dental Implants According to Length and Diameter Parameters by FE Method in the Anterior Position
International Journal of Bioscience, Biochemistry and Bioinformatics, Vol. 4, No. 4, July 2014 Optimum Selection of the Dental Implants According to Length and Diameter Parameters by FE Method in the Anterior
METU DEPARTMENT OF METALLURGICAL AND MATERIALS ENGINEERING
METU DEPARTMENT OF METALLURGICAL AND MATERIALS ENGINEERING Met E 206 MATERIALS LABORATORY EXPERIMENT 1 Prof. Dr. Rıza GÜRBÜZ Res. Assist. Gül ÇEVİK (Room: B-306) INTRODUCTION TENSION TEST Mechanical testing
Three-dimensional finite element stress analysis of SKY implant system
Journal of the Serbian Society for Computational Mechanics / Vol. 4 / No. 2, 2010 / pp. 87-96 (UDC: 616.314-77) Three-dimensional finite element stress analysis of SKY implant system M. Kalanović 1*, N.
Placement of dental implants is a highly successful
Shape Optimization of Dental Implants Li Shi, BEng, MSc 1 /Haiyan Li, BEng, MSc, PhD 2 /Alex S. L. Fok, BEng, MSc, PhD, CEng 3 /Cemal Ucer, BDS, MSc 4 / Hugh Devlin, BDS, BSc, MSc, PhD 5 /Keith Horner,
The Effect of Dental Implant Materials and Thread Profiles A Finite Element and Statistical Study
RESEARCH ARTICLE Copyright 2013 American Scientific Publishers All rights reserved Printed in the United States of America Journal of Medical Imaging and Health Informatics Vol. 3, 1 5, 2013 The Effect
Straumann Dental Implant System. Implant Selection Guide.
Straumann Dental Implant System. Implant Selection Guide. STRAUMANN's IMPLANT PORTFOLIO The Straumann Dental Implant System offers two implant lines with diverse body and neck designs ranging from the
Dentistry Fatigue test for endosseous dental implants. Report No.: 1 ASTRA System Ti Design
Dentistry Fatigue test for endosseous dental implants Report No.: 1 ASTRA System Ti Design Client: NT - Trading Time of execution: Mai 2011 Subject of work: - - S - 24285 Abutment Ti DesignTM - RESULTS
Supervisors: Dr. Farhan Raza Khan
1 Presenter: Dr. Sana Ehsen Supervisors: Dr. Farhan Raza Khan 2 A dental implant (also known as an endosseous implant or fixture) is a surgical component that interfaces with the bone of the jaw to support
Resorptive Changes of Maxillary and Mandibular Bone Structures in Removable Denture Wearers
Resorptive Changes of Maxillary and Mandibular Bone Structures in Removable Denture Wearers Dubravka KnezoviÊ-ZlatariÊ Asja»elebiÊ Biserka LaziÊ Department of Prosthodontics School of Dental Medicine University
Histologic comparison of biologic width around teeth versus implants: The effect on bone preservation
Clinical Histologic comparison of biologic width around teeth versus implants: The effect on bone preservation Kazuto Makigusa 1 Abstract Histological analysis of the biological width surrounding primate
Restoration of the Edentulous Maxilla: The Case for the Zygomatic Implants
CLINICAL CONTROVERSIES IN ORAL AND MAXILLOFACIAL SURGERY: PART ONE J Oral Maxillofac Surg 62:1418-1422, 2004 Restoration of the Edentulous Maxilla: The Case for the Zygomatic Implants Eric D. Ferrara,
Settlement of Precast Culverts Under High Fills; The Influence of Construction Sequence and Structural Effects of Longitudinal Strains
Settlement of Precast Culverts Under High Fills; The Influence of Construction Sequence and Structural Effects of Longitudinal Strains Doug Jenkins 1, Chris Lawson 2 1 Interactive Design Services, 2 Reinforced
Healing Abutment Selection. Perio Implant Part I. Implant Surface Characteristics. Single Tooth Restorations. Credit and Thanks for Lecture Material
Healing Abutment Selection Perio Implant Part I Credit and Thanks for Lecture Material Implant Surface Characteristics!CAPT Robert Taft!CAPT Greg Waskewicz!Periodontal Residents NPDS and UMN!Machined Titanium!Tiunite!Osseotite
Chapter Outline. Mechanical Properties of Metals How do metals respond to external loads?
Mechanical Properties of Metals How do metals respond to external loads? Stress and Strain Tension Compression Shear Torsion Elastic deformation Plastic Deformation Yield Strength Tensile Strength Ductility
Clinical and Laboratory Procedures for Fixed Margin Implant Abutments
Clinical and Laboratory Procedures for Fixed Margin Implant Abutments Dr. Carl Drago DDS, MS, American Board of Prosthodontics Director, Dental Research BIOMET 3i, Adjunct Faculty Department of Prosthodontics,
Quality from Switzerland 01.01.2014
Quality from Switzerland 01.01.2014 TRI Dental Implants: An example of the highest standards of quality We deliver quality We supply quality. From patent to patient Modern production Compliance with the
Custom Design of Knee Joint Prosthesis By Using Computerized Tomography (CT) Images and 3D Modelling
Custom Design of Knee Joint Prosthesis By Using Computerized Tomography (CT) Images and 3D Modelling Qasem Alyazji Near East University, Department of Biomedical Engineering Nicosia, North Cyprus, Turkey
Optimum proportions for the design of suspension bridge
Journal of Civil Engineering (IEB), 34 (1) (26) 1-14 Optimum proportions for the design of suspension bridge Tanvir Manzur and Alamgir Habib Department of Civil Engineering Bangladesh University of Engineering
Technology of EHIS (stamping) applied to the automotive parts production
Laboratory of Applied Mathematics and Mechanics Technology of EHIS (stamping) applied to the automotive parts production Churilova Maria, Saint-Petersburg State Polytechnical University Department of Applied
Bearing strength of stainless steel bolted plates in tension
NSCC29 Bearing strength of stainless steel bolted plates in tension G. KIYMAZ 1 1 Department of Civil Engineering, Istanbul Kultur University, Ataköy Campus, Bakırköy, Istanbul, Turkey ABSTRACT: A study
Technical Notes 3B - Brick Masonry Section Properties May 1993
Technical Notes 3B - Brick Masonry Section Properties May 1993 Abstract: This Technical Notes is a design aid for the Building Code Requirements for Masonry Structures (ACI 530/ASCE 5/TMS 402-92) and Specifications
Contact: Steve Hurson VP, R & D Nobel Biocare 22715 Savi Ranch Pkwy Yorba Linda, CA 92887 (714) 282-5072 steve.hurson@nobelbiocare.
Use of CP Titanium and Titanium alloys for Dental Implants Steve Hurson, Nobel Biocare U.S.A. Yorba Linda CA Abstract Professor Per Ingvar Branemark of Sweden published his results on 15 years of research
The Impact of Oral Implants Past and Future, 1966 2042. Tomas Albrektsson, MD, PhD, ODhc Ann Wennerberg, DDS, PhD
P R O F E S S I O N A L I S S U E S The Impact of Oral Implants Past and Future, 1966 202 Tomas Albrektsson, MD, PhD, ODhc Ann Wennerberg, DDS, PhD A b s t r a c t This paper traces the history of oral
Dental Updates. Excerpted Article e-mail: [email protected]. Why Implant Screws Loosen Part 1. Richard Erickson, MS, DDS
¼ ½ ¾ µ mw/cm 2 Volume 17; 2007 Dental Updates "CUTTING EDGE INFORMATION FOR THE DENTAL PROFESSIONAL " 200 SEMINARS AND 30 JOURNALS REVIEWED YEARLY FOR THE LATEST, CUTTING EDGE INFORMATION Excerpted Article
Objectives. Experimentally determine the yield strength, tensile strength, and modules of elasticity and ductility of given materials.
Lab 3 Tension Test Objectives Concepts Background Experimental Procedure Report Requirements Discussion Objectives Experimentally determine the yield strength, tensile strength, and modules of elasticity
Prosthetic treatment planning on the basis of scientific evidence.
Prosthetic treatment planning on the basis of scientific evidence. Pjetursson BE, Lang NP. J Oral Rehabil. 2008 Jan;35 Suppl 1:72-9. Faculty of Odontology, University of Iceland, Reykjavik, Iceland, and
MASTER DEGREE PROJECT
MASTER DEGREE PROJECT Finite Element Analysis of a Washing Machine Cylinder Thesis in Applied Mechanics one year Master Degree Program Performed : Spring term, 2010 Level Author Supervisor s Examiner :
The Basics of FEA Procedure
CHAPTER 2 The Basics of FEA Procedure 2.1 Introduction This chapter discusses the spring element, especially for the purpose of introducing various concepts involved in use of the FEA technique. A spring
Dental Implant Options in Atrophic Jaws
Dental Implant Options in Atrophic Jaws Orthopedic Application Jay B. Reznick, D.M.D., M.D. Diplomate, American Board of Oral and Maxillofacial Surgery Tarzana, CA Endopore Dental Implant System Screw-Type
IMPLANT CONSENT FORM WHAT ARE DENTAL IMPLANTS?
IMPLANT CONSENT FORM WHAT ARE DENTAL IMPLANTS? Dental implants are a very successful and accepted treatment option to replace lost or missing teeth. A dental implant is essentially an artificial tooth
MATERIALS AND METHODS
REDISTRIBUTION OF LOAD OF INJURED LOWER CERVICAL SPINE UNDER AXIAL COMPRESSION USING FEM E. C. Teo 1 and H. W. Ng 1 1 School of Mechanical and Production Engineering, Nanyang Technological University,
Introduction to Mechanical Behavior of Biological Materials
Introduction to Mechanical Behavior of Biological Materials Ozkaya and Nordin Chapter 7, pages 127-151 Chapter 8, pages 173-194 Outline Modes of loading Internal forces and moments Stiffness of a structure
Luigi Vito Stefanelli,D.Eng.,DDS
LYON, 27-28/09/2013 Luigi Vito Stefanelli,D.Eng.,DDS Assistant Researcher, Department of Oral Sciences, Sapienza University of Rome, Italy, School of Dentistry. Private practitioner dental office Stasi
The SATURN implant by Cortex Dental Industries
The SATURN implant by Cortex Dental Industries By Dr. Zvi Laster DMD W e P r o v e I t E v e r y D a y A case report using a newly designed implant specifically designed for immediate post-extraction loading
Design of Steel Structures Prof. S.R.Satish Kumar and Prof. A.R.Santha Kumar. Fig. 7.21 some of the trusses that are used in steel bridges
7.7 Truss bridges Fig. 7.21 some of the trusses that are used in steel bridges Truss Girders, lattice girders or open web girders are efficient and economical structural systems, since the members experience
Tired of THe Gooey mess?
Tired of THe Gooey mess? There is a better option. if you Are WeArinG conventional full or partial dentures you may relate To THese daily complications. Constant denture movement frequent adjustments you
New approaches in Eurocode 3 efficient global structural design
New approaches in Eurocode 3 efficient global structural design Part 1: 3D model based analysis using general beam-column FEM Ferenc Papp* and József Szalai ** * Associate Professor, Department of Structural
Rehabilitation of Endondontically Failed Anterior teeth by Immediate Replacement and Loading of an Implant supported Crown: A Case Report.
RESEARCH AND REVIEWS: JOURNAL OF DENTAL SCIENCES Rehabilitation of Endondontically Failed Anterior teeth by Immediate Replacement and Loading of an Implant supported Crown: A Case Report. Lalit Kumar 1
Solid shape molding is not desired in injection molding due to following reasons.
PLASTICS PART DESIGN and MOULDABILITY Injection molding is popular manufacturing method because of its high-speed production capability. Performance of plastics part is limited by its properties which
Flapless Implant Surgery for Replacement of Posterior Teeth
Course Number: 108.2 Flapless Implant Surgery for Replacement of Posterior Teeth Authored by J. Steven Cloyd, DDS Upon successful completion of this CE activity 1 CE credit hour may be awarded A Peer-Reviewed
CRITERIA FOR PRELOADED BOLTS
National Aeronautics and Space Administration Lyndon B. Johnson Space Center Houston, Texas 77058 REVISION A JULY 6, 1998 REPLACES BASELINE SPACE SHUTTLE CRITERIA FOR PRELOADED BOLTS CONTENTS 1.0 INTRODUCTION..............................................
Leslie Laing Gibbard, BSc, BEd, MSc, PhD, DDS George Zarb, BChD, DDS, MS, MS, FRCD(C)
A P P L I E D R E S E A R C H A 5-Year Prospective Study of Implant-Supported Single-Tooth Replacements Leslie Laing Gibbard, BSc, BEd, MSc, PhD, DDS George Zarb, BChD, DDS, MS, MS, FRCD(C) A b s t r a
TRIGEN INTERTAN vs Stryker Gamma3 summary
TRIGEN INTERTAN vs Stryker Gamma3 summary Initial stiffness of INTERTAN construct almost 4% greater than Gamma3 INTERTAN femoral head rotation and varus collapse initially up to 84% lower than Gamma3 After
Tensile fracture analysis of blunt notched PMMA specimens by means of the Strain Energy Density
Engineering Solid Mechanics 3 (2015) 35-42 Contents lists available at GrowingScience Engineering Solid Mechanics homepage: www.growingscience.com/esm Tensile fracture analysis of blunt notched PMMA specimens
Zimmer Trabecular Metal Dental Implant
Zimmer Trabecular Metal Dental Implant TRABECULAR METAL MATERIAL: Designed to Enhance Secondary Stability Through Bone Ingrowth. Artistic Rendering osseoincorporation TRABECULAR METAL MATERIAL Designed
Finite Elements for 2 D Problems
Finite Elements for 2 D Problems General Formula for the Stiffness Matrix Displacements (u, v) in a plane element are interpolated from nodal displacements (ui, vi) using shape functions Ni as follows,
A Comprehensive Explanation
Dental Implants A Comprehensive Explanation Overview Since the 1980s, dental implants have become more popular among dentists and patients. 1 In some clinical situations, implants may be the best treatment
IMPLANT DENTISTRY EXAM BANK
IMPLANT DENTISTRY EXAM BANK 1. Define osseointegration. (4 points, 1/4 2. What are the critical components of an acceptable clinical trial? (10 points) 3. Compare the masticatory performance of individuals
While the prosthetic rehabilitation of
Restoring Mandibular Single Teeth with the Inclusive Tooth Replacement Solution Go online for in-depth content by Bradley C. Bockhorst, DMD While the prosthetic rehabilitation of full-arch cases provides
Straumann Dental Implants Confident smiles
Straumann Dental Implants Confident smiles STRAUMANN DENTAL IMPLANTS - DESIGNED TO LAST A LIFETIME Dental Implants The decision to replace missing teeth with dental implants is an excellent investment
What is a dental implant?
What is a dental implant? Today, the preferred method of tooth replacement is a dental implant. They replace missing tooth roots and form a stable foundation for replacement teeth that look, feel and function
Factors Influencing the Resonance Frequency of Dental Implants
SCIENTIFIC ARTICLES J Oral Maxillofac Surg 61:1184-1188, 2003 Factors Influencing the Resonance Frequency of Dental Implants Haw-Ming Huang, MS,* Ching-Lai Chiu, DDS, MS, Ching-Ying Yeh, PhD, and Sheng-Yang
Solved with COMSOL Multiphysics 4.3
Vibrating String Introduction In the following example you compute the natural frequencies of a pre-tensioned string using the 2D Truss interface. This is an example of stress stiffening ; in fact the
Three-Dimensional Analysis Using Finite Element Method of Anterior Teeth Inclination and Center of Resistance Location
Three-Dimensional Analysis Using Finite Element Method of Anterior Teeth Inclination and Center of Resistance Location Allahyar GERAMY 1, Ahmad SODAGAR 1, Mehdi HASSANPOUR 1 Objective: To locate the centre
Restoring quality to life. Dental implants. A naturally better solution. Patient Education
Restoring quality to life. Dental implants. A naturally better solution. Patient Education Dental implants: A better treatment option. What are dental implants? Dental implants are a safe, medically proven,
3M ESPE MDI. Mini Dental Implants. Literature Review. Espertise. Scientific Facts
3M ESPE MDI Mini Dental Implants Literature Review Espertise Scientific Facts 3M ESPE MDI Mini Dental Implant Denture Stabilization Literature Review June 2009 Objective: To review the published literature
Pragmatic multi-scale and multi-physics analysis of Charles Bridge in Prague
Pragmatic multi-scale and multi-physics analysis of Charles Bridge in Prague Jiří Šejnoha 1,2 Michal Šejnoha 1,2 Jan Zeman 1 Jan Novák 1,2 with Zdeněk Janda, Jiří Maděra, Jan Vorel and Jan Sýkora 1 Department
Molar Uprighting Dr. Margherita Santoro Division of Orthodontics School of Dental and Oral surgery. Consequences of tooth loss.
Molar Uprighting Dr. Margherita Santoro Division of Orthodontics School of Dental and Oral surgery Molars The wide occlusal surface is designed for food grinding. The surface needs to be aligned with the
Bending, Forming and Flexing Printed Circuits
Bending, Forming and Flexing Printed Circuits John Coonrod Rogers Corporation Introduction: In the printed circuit board industry there are generally two main types of circuit boards; there are rigid printed
What Dental Implants Can Do For You!
What Dental Implants Can Do For You! Putting Smiles into Motion About Implants 01. What if a Tooth is Lost and the Area is Left Untreated? 02. Do You Want to Restore Confidence in Your Appearance? 03.
Mechanical Properties of Metals Mechanical Properties refers to the behavior of material when external forces are applied
Mechanical Properties of Metals Mechanical Properties refers to the behavior of material when external forces are applied Stress and strain fracture or engineering point of view: allows to predict the
Finite Element Formulation for Plates - Handout 3 -
Finite Element Formulation for Plates - Handout 3 - Dr Fehmi Cirak (fc286@) Completed Version Definitions A plate is a three dimensional solid body with one of the plate dimensions much smaller than the
EXPERIMENTAL AND NUMERICAL ANALYSIS OF THE COLLAR PRODUCTION ON THE PIERCED FLAT SHEET METAL USING LASER FORMING PROCESS
JOURNAL OF CURRENT RESEARCH IN SCIENCE (ISSN 2322-5009) CODEN (USA): JCRSDJ 2014, Vol. 2, No. 2, pp:277-284 Available at www.jcrs010.com ORIGINAL ARTICLE EXPERIMENTAL AND NUMERICAL ANALYSIS OF THE COLLAR
Validation of Cable Bolt Support Design in Weak Rock Using SMART Instruments and Phase 2
Validation of Cable Bolt Support Design in Weak Rock Using SMART Instruments and Phase 2 W.F. Bawden, Chair Lassonde Mineral Engineering Program, U. of Toronto, Canada J.D. Tod, Senior Engineer, Mine Design
P. Lu, Sh. Huang and K. Jiang
416 Rev. Adv. Mater. Sci. 33 (2013) 416-422 P. Lu, Sh. Huang and K. Jiang NUMERICAL ANALYSIS FOR THREE-DIMENSIONAL BULK METAL FORMING PROCESSES WITH ARBITRARILY SHAPED DIES USING THE RIGID/VISCO-PLASTIC
Standard Internal Hex
Standard Internal Hex Touareg TM -OS Touareg TM -S Swell TM Touareg -S Touareg -OS Swell About ADIN Adin Dental Implant Systems Ltd., designs, manufactures and markets state of the art, technologically
DESS. Screws. Tijuana Ventas: (664) 685 6294/95 [email protected]. For all major implant systems!! www.dabocorp.com
Screws Screws for definitive use. Made from medical grade 5 ELI Titanium Integrity and soundness guaranteed by stress tests of up to 150% of their nominal torques. Their designs ensure a perfect fit with