BENDABLE VS. ANGULATED DENTAL IMPLANTS: CONSIDERATION OF ELASTIC AND PLASTIC MATERIAL PROPERTIES BASED ON EXPERIMENTAL IMPLANT MATERIAL DATA AND FEA
|
|
|
- Chrystal Greene
- 9 years ago
- Views:
Transcription
1 Biomed Pap Med Fac Univ Palacky Olomouc Czech Repub. 2008, 152(2): T. Goldmann, S. Ihde, J. Kuzelka, L. Himmlova 309 BENDABLE VS. ANGULATED DENTAL IMPLANTS: CONSIDERATION OF ELASTIC AND PLASTIC MATERIAL PROPERTIES BASED ON EXPERIMENTAL IMPLANT MATERIAL DATA AND FEA Tomas Goldmann a, Stefan Ihde b *, Jiri Kuzelka a, Lucie Himmlova c a Czech Technical University in Prague, Laboratory of Biomechanics, Department of Mechanics, Biomechanics and Mechatronics, Faculty of Mechanical Engineering, Technicka 4, Prague, Czech Republic b Gommiswald Dental Clinic, Dorfplatz 11, 8737 Gommiswald/SG, Switzerland c Institute of Dental Research 1 st Medical Faculty of Charles University and General Medical Hospital, Vinohradska 48, Prague 2, Czech Republic [email protected] Received: May 9, 2008; Accepted: September 29, 2008 Key words: FEM analysis/basal implants/bendable implants/angulated implants Aim: Bendable and angulated single-piece implants are used alternatively to screwable abutments in two-piece dental implant designs. Though used frequently, data on the stress distribution within such implants are not available and the question whether the bending contributes to fracture resistance has not been addressed. Methods: We used the method of finite element to identify von Mises stresses and maximum stresses in bent and non-bent but angulated implants. Implants with one (e.g. applicable to screw designs) or two (applicable to basal implants) bending areas were the variables under investigation. Results: For bends up to 13 degrees we discovered that if there is only one bend, the maximum stress is in the bent area. If two bends are made in two different bending areas, the maximum stresses are distributed between the two and, if either one of the bent areas is machined, there are no residual stresses within the implant body in this area. The maximum stresses are always located near the base-plates. The absolute value of the maximum stress is higher because no residual stresses are available to compensate stresses that stem from loading. Conclusion: Assuming that all other parameters are equal, bendable (basal) implants show a more even stress distribution along the vertical implant region than identically shaped implants with a machine-angulated area. Bendable basal implants therefore probably resist masticatory forces better than pre-angulated, machined implants, and unbent implants which provide a thin region in the vertical implant area. INTRODUCTION Bendable and angulated single-piece dental implants are used as an alternative to designs which include screwable abutments 1. The advantage of using one-piece implants is avoidance of gaps between implant body and abutments and to obviate otherwise frequent loosening of abutments or prosthetic screws 2. One consequence of jaw bone atrophy may be that no bone is present in those jaw areas where the teeth are to be placed. This creates problems for the implantologist and has led to the desire to perform bone augmentations. Basal implants 3 overcome this problem in a different manner than cylindrical ones. Their baseplates are anchored in small, native bone areas often far from the actual clinical tooth, i.e. distant in a vertical and/or sagittal and/or horizontal direction. Although clinical experience has led to the development of a variety of basal implant designs, the surgical situation may demand variable positioning of the interconnection implant part (screw hole or cementing post). For this reason, the surgeon has to bend the vertical part of the implants before or after the insertion, in order to allow easy prosthetic access of the connection area. Although bending of implant necks has been used in crestal implantology for a long time, no literature was found on the effects of bending on the development of residual stresses within the implant and alteration of the stress distribution during loading. The description of the effects takes into account that plastic bending affects the internal structure of the implant body because the described deformations exceed the yield limits of the material. We have developed a model that takes into account the following: the process of bending an inserted implant is assumed to be harmless to the bone if the resulting forces on the bone interface remain within the range of elastic deformation of the bone, thus creating no or only a limited number of cracks or micro-cracks the forces must be within a range that can be applied inside the oral cavity by the surgeon. In previous articles we created models to describe the interaction between bone and basal implants using the
2 310 T. Goldmann, S. Ihde, J. Kuzelka, L. Himmlova method of finite element analysis. Thereby we proposed a novel approach for modelling different interactions between cortical and cancellous bone 4,5. This approach has been used for describing different stages of osseointegration and remineralization. In daily clinical practice as well as during previous computations, the problem of stress distribution within dental implant bodies undergoing intra-operative bending vs. implant bodies which have been machined in angulated designs out of one piece of metal became of interest. MATERIAL AND METHODS The understanding of dental implants mechanical behavior during bending process comes from clinical practice fi ndings, where the problem of the durability prediction during implantation procedure arises. A set of finite element models describing possible bending cases formed at implantation was chosen as a suitable method. However material experiment takes also part in this study (Titanium Grade 2), where results are used to improve material model. As a fi rst step to obtaining reliable results, we performed a tensile testing experiment on a Titanium Grade 2 specimen. In this way the material model is extremely precise over a wide range of loading. We used rod specimens for this experiment with a total length of 190 mm and a diameter of 10 mm. The specimen was connected to the testing machine through M16 threads at both ends (Fig. 1a, b). We tested 5 specimens and the results were extremely similar. Average values and standard deviations (tensile yield strength σ k = 326 ± and ultimate tensile yield strength σ p = ± 2.41) were used for calculation. Our model is based on the stress-strain-curve shown in Fig. 2. This figure shows three lines (0 A; A B, B C), with the line 0 A corresponding to the elastic material behaviour, lines A B and B C approximate the properties of the material in the plastic zone. The line decreases on the right side of C since the cross sectional area of the specimen diminishes during the experiment. Since we found a considerable difference in repeated measurements and the available data in the literature 6 (Fig. 3), we decided to trust our own measurements and used the model obtained values from our material tests. In the above reference 6 the ultimate strength is reported to be higher than the maximum value of the yield strength, which we consider unusual. Also for this reason we decided to perform a test of our own. As a second step to approach the questions arising from angulated or bendable dental implants we chose to observe the stresses within the implants and the influence of the bending action, without considering the stresses at the interface to bone. Thus we modeled through the method of finite elements several bending situations for basal implants with one baseplate. The finite element mesh of the mandible was created in the system ABAQUS (Abaqus Inc., Providence, RI , USA) by the C3D4 element type. The ABAQUS system was also used for all remaining pre-processing, computational and result evaluation works. The practically usable cross sectional diameter of the bending area for a dental implant (made from c.p. Titanium Grade 2) can only range from 1.7 mm to 2.0 mm. Smaller diameters should not be considered because they are prone to fracture, while implants with higher diameters do not allow bending inside the oral cavity. For our calculations, the shaft thickness was assumed to be 1.95 mm and the resulting upper and lower bending angles after plastic deformation and elastic relapse were each 13 degrees. The vertical implant part had no contact with bone and the baseplates were rigidly anchored. Four cases of bending or double bending of the vertical implant part were calculated (Fig. 4). Since we found in preliminary calculations 4 that only very little stress occurs within the baseplates in these implant designs, we were able to simplify the model into a plane symmetrical part of implant and leave out the desk (i.e. the fixating baseplate; Fig. 5). In this way our findings can also be applied to the crestal implant designs, i.e. screws. However in screw designs only the upper bending area is of clinical use. The analysis of each case (Fig. 4) considers a sequence of mechanical actions imposed on the implants: Step 1: before any bending, the implant was fixed below the particular bending area Step 2: the implant part above the notch was then bent, applying kinematical load. The initial bending angle was 15 degrees Step 3: immediately after releasing the necessary loading, an elastic relapse was assumed to occur, resulting in a total bending angle of 13 degrees Step 4: after the bending, stresses within the pre-bent implants under loading conditions were calculated. Note that von Mises stresses and main stresses were calculated and displayed. the actual loading was calculated in two ways: 1. Main stresses and von Mises stresses were calculated for all four cases, assuming static loading of 450 N, with the load being applied perpendicular to the crestal abutment surface. 2. Results were compared to implants which had not been bent but manufactured by CNC/machining into exactly the same angulation. These implants showed no residual stresses inside the area of angulation because they have not been bent. A reduced force of only 390 N was applied in all cases where the upper bending area was bent. We had to use a lower force because the computation would be inaccurate if the value of stress determined at point C (Fig. 2) was reached or exceeded. In the cases of two bendings (case 3 and 4, Fig. 4), we assumed that the results of the second bending procedure did not affect the first bending. The implant material, considered in this model, is c.p. Titanium grade 2, ISO :1993. Elastic properties are described by Young s modulus of elasticity E = 1.05e5 MPa and Poisson s ratio ν = The plastic prop-
3 Bendable vs. angulated dental implants: Consideration of elastic and plastic material properties based on experimental implant material data and FEA 311 Fig. 1a, b. Heckert VFP400 tensile testing device used for our testing, with the test specimen made from Titanium Grade 2. Fig. 2. Results of tensile testing of the Ti Grade 2 rod. A (yield point): x = 0,0123; y = 329,3, determining the end of the elastic zone; B (a second defined point for our material model): x = 0,0604; y = 483,7; C (endpoint of our model definition): x = 0,01209; y = 494,2. erties are taken from our material test described above where the stress-strain curve was approximated by three points (A, B, C). The above-mentioned material properties are shown in Figure 2 for illustration. The material in the plastic stage is modelled as incompressible (ν = 0.5). The material is assumed to change its material properties isotropically during plastic deformation but these changes have not been quantified to our knowledge. RESULTS 1. Bent implants Examples of Von Mises stress distribution [MPa] in range from the yield stress to the failure stress at the moment of maximal bend (Step 2 15 bending), the release of the kinematical loading (Step 3 13 bending) or final loading by the force (Step N or 390 N
4 312 T. Goldmann, S. Ihde, J. Kuzelka, L. Himmlova σ [MPa] ε [-] Fig. 3. The range of elastic deformation of Titanium Grade 2 reaches up to approx. 275 MPa (lower value before mechanical working while the almost horizontal, upper part of the line represents plastic material behaviour). Fig. 4. The four cases of bent or double-bent implants calculated in this publication are shown above. Two notches are integral part of the vertical part of the basal implants under examination. The upper notch is positioned underneath the prosthetic connection parts, the lower notch is found slightly above the baseplate. loading) for all four cases are shown in Figs The coloured range corresponds to the plastic zone. Residual von Mises stress distribution is distinguished from stresses under loading and from stresses during the actual bending phase. This value is slightly higher than failure stress in some cases. Moreover the tensile stress is positive and thus there is a serious risk for cracks. Figures 6, 7 and 8 show von Mises stress distribution for Case 1. Figure 6 displays the situation after kinematical 15 bending (Step 2); the maximum stress here is 495 MPa. This maximal stress surprisingly increases after the release of the bending load (Step 3) up to 530 MPa (Fig. 7) which is caused by the residual stress. The stress decreases again (515 MPa) after loading the implant by a force of 450 N (Fig. 8). This shows that loading of the implant after bending affects the residual stress and a more balanced stress distribution is the result. A more or less similar situation occurs again for remaining Cases 2, 3 and 4. The Figure 9 displays von Mises stress distribution at a maximum load of 390 N (Step 4) after one single bend had been applied to the upper bending area (Case 2). Note that there is also a deformation in the area of the lower bending zone stemming from the loading. (Maximum stress is 528 MPa but note that the loading force is different from Fig. 8). Figure 10 shows the residual stresses (Max stress is 530 MPa) in the vertical implant after two bends had been released (Case3, Step 3) and Figure 11 shows the stress increase (up to 591 MPa) after loading the implant by a force of 390 N (Case 3, Step 4). Case 3 is the only case where the loading in Step 4 causes no residual stresses and produces the most balanced stress distribution. Figures 12 and 13 describe Steps 3 and 4 for Case 4. The greater stress distribution (530 MPa) is again in the state of the bend release (Step 3) than in the state of final loading (Step 4 maximum stress is 519 MPa). 2. Unbent implants, machined with one angulation in the vertical implant shaft If the internal structure of the vertical implant part is homogenous, no residual stresses are present in the (machined) upper angulation area. In this case off-axis loading will lead to a peak stress near the baseplate. This is also true for cases without a circular notch (not shown here): in these cases the peak stresses may be displaced nearer to the baseplate (Fig. 14). The same applies to basal implants if two angulations are machined into the vertical part (Fig. 15). Our findings may be summarized as follows: if there is only one bending carried out, the maximum stresses are always found in the bent area if two bends are carried out, the maximum stresses are distributed between the two bent areas if only one of the bent area is machined, there are no residual stresses within the implant body in this area and maximum stresses are always located near the baseplates. The absolute value of the maximum stresses is higher because no residual stresses are available to compensate stresses that stem from loading. DISCUSSION The method of the finite element model is frequently used in evaluating the design of dental implants 7, 8. Most 9, 10 studies evaluate either the stresses imposed on the bone or on the prosthetic workpieces 11, 12. We found one study showing that staggered implant placement produces similar stresses to straight placement 13. No evaluations of bendable implant designs were found in the existing literature and hence the latter provides only limited grounds for comparison with the results of others. It is possible to apply bends to the vertical implant part more than 13 degrees. However, we chose the maxi-
5 Bendable vs. angulated dental implants: Consideration of elastic and plastic material properties based on experimental implant material data and FEA 313 Fig. 5. Plane symmetrical model of the implant, showing the areas of fixation. The term rotation of the ABAQUS terminology ( describes the bent(-over) part of the vertical implant section. Fig. 6. Case 1, Step 2 von Mises stress distribution in MPa (Maximum stress is 495 MPa). Fig. 7. Case 1, Step 3 von Mises stress distribution in MPa (Maximum stress is 530 MPa).
6 314 T. Goldmann, S. Ihde, J. Kuzelka, L. Himmlova Fig. 8. Case 1, Step 4 von Mises stress distribution in MPa (Maximum stress is 515 MPa). Fig. 9. Case 2, Step 4 von Mises stress distribution in MPa (Maximum stress is 528 MPa). Fig. 10. Case 3, Step 3 von Mises stress distribution in MPa (Maximum stress is 530 MPa). Fig. 11. Case 3, Step 4 von Mises stress distribution in MPa (Maximum stress is 591 MPa).
7 Bendable vs. angulated dental implants: Consideration of elastic and plastic material properties based on experimental implant material data and FEA 315 Fig. 12. Case 4, Step 3 von Mises stress distribution in MPa (Maximum stress is 530 MPa). Fig. 13. Case 4, Step 4 von Mises stress distribution in MPa (Maximum stress is 519 MPa). Fig. 14. Von Mises stress distribution within a basal implant with a machined upper angulation area (in MPa). The same applies to basal implants if two angulations have been machined into the vertical part of Fig. 15. Fig. 15. Von Mises stresses within a basal implant with two machined angulations in the vertical implant portion.
8 316 T. Goldmann, S. Ihde, J. Kuzelka, L. Himmlova mum degree of bending (after Step 2) since when we tried to calculate with larger bending angles, the computation failed. In our model the situation changed so dramatically with bends of more than 13 degrees, that any calculation was unfeasible. Larger bends are clinically possible and also performed in daily clinical practice but our model did not allow us to compute them. We did not calculate machined pre-angulated implants with a greater angulation because the bendable counterpart was not available for comparison in our model. The highest stresses were found when exerting a maximum bending of 15 degrees. This stress consists of elastic and plastic components. After the elastic relapse to a final bending of 13 degrees, the plastic deformation and the residual stresses remain, while the elastic stresses vanish. The strong internal forces from plastic deformation are not visible. However, they affect the stress distribution during subsequent loading. Our findings are applicable to a wide range of materials. However a large number of materials will not allow bending under the conditions applicable inside the oral cavity. Interestingly the different types of bending (Cases 1-4) do not alter the maximum stresses dramatically. One of the weaknesses of this study is the fact, that the interface to bone was not considered. Hence the results are valid only for basal implants immediately after the implant placement, in cases where the implant is placed in highly mineralized, cortical bone providing a rigid fixation of the baseplate with no contact between the vertical implant part and the bone. The situation during the dual process of osseointegration of a basal implant has been described for non-angulated, unbent implants and oblique loading 5. However, calculations for angulated or bent osseointegrated implants and the determination of forces acting on the bone while implants are bent, remain to be performed in future. Also the effect of cyclic loading on the bending areas was not calculated, since fatigue computations are not routinely performed in finite element analysis. When transferring the results from basal to screw-type implants and the clinical reality, we must keep in mind that the insertion of bendable implants into more mineralized bone (Types D1, D2) may lead to extremely high torque forces, which exceed the forces during bending by far. These forces may lead to fracture of the bending area, although the implant may have been well able to withstand the bending and the following cyclic masticatory loading without damage, in the case that the implant had only been inserted without damage. CONCLUSIONS Assuming that all other parameters are equal, bendable basal dental implants show a more even stress distribution along the vertical implant part compared to identical angulated implants. Bent implants of this type probably therefore resist masticatory forces better than angulated implants. Our results can be transferred to screw implants with thin, bendable necks: those implants should undergo bending whenever they are used, in order to improve material properties through residual stresses and thereby increase mechanical resistance in the bending area. ACKNOWLEDGEMENTS This study was funded by the Czech Ministry of Education project No. MSM and by the Grant Agency of the Czech Republic under project No. 106/06/0849. REFERENCES 1. Bauer E. Dentalimplantat EP B1; Binon PP. Evaluation of the effectiveness of a technique to prevent screw loosening. J Prosthet Dent 1998; 42: Scortecci G, Misch CE, Benner K. Implants and Restorative Dentistry. 2000, Martin Dunitz,: Martin Dunitz, London; Ihde S, Goldman T, Himmlova L, Aleksic Z. The Use of Finite Element Analysis to Model Bone-Basal Implant Contact. Oral Surg Oral Med Oral Pathol Oral Radiol Endod. 2008; 106(1): Ihde S, Goldman T, Himmlova L, Aleksic Z, Kuzelka J. Implementation of contact definitions calculated by FEA to describe the healing process of basal implants. Biomed Pap Med Fac Univ Palacky Olomouc Czech Repub 2008; 152(1): Titanium grade 2 composition, available from: com/search/specificmaterial.asp?bassnum=mtu DeTolla DH, Andreana S, Patra A, Buhite R, Comella B. Role of the finite element model in dental implants. J Oral Implantol 2000; 26(2): Geng JP, Tan KB, Liu GR. Application of finite element analysis in implant dentistry: a review of the literature. J Prosthet Dent 2001 Jun; 85(6): Geng JP, Xu W, Tan KBC, Liu GR. Finite Element Analysis of an Osseointegrated Stepped Screw Dental Implant. J Oral Implantol 2004; 30: Holmgren EP, Seckinger RJ, Kilgren LM, Mante F. Evaluating Parameters of Osseointegrated Dental Implants using finite element Analysis a two-dimensional Comparative study examining the effects of implant diameter, implant shape, and load direction. J Oral Implantol 1998; 22(2): Ciftci Y, Canay S. Stress distribution on the metal framework of the implant supported fixed prosthesis using different veneering materials. Int J Prosthod 2001; 14(5): Geramy A, Morgano SM. Finite element analysis of three designs of an implant-supported molar crown. J Prosthet Dent 2004; 92(5): Kivanc A, Haldun I. Finite element analysis of the influence of staggered versus straight placement of dental implants. Int J Oral Maxillofac Impl 2001; 16(5):
Contact: Steve Hurson VP, R & D Nobel Biocare 22715 Savi Ranch Pkwy Yorba Linda, CA 92887 (714) 282-5072 steve.hurson@nobelbiocare.
Use of CP Titanium and Titanium alloys for Dental Implants Steve Hurson, Nobel Biocare U.S.A. Yorba Linda CA Abstract Professor Per Ingvar Branemark of Sweden published his results on 15 years of research
CH 6: Fatigue Failure Resulting from Variable Loading
CH 6: Fatigue Failure Resulting from Variable Loading Some machine elements are subjected to static loads and for such elements static failure theories are used to predict failure (yielding or fracture).
Computer Aided Engineering (CAE) Techniques Applied To Hip Implant
International Journal Of Computational Engineering Research (ijceronline.com) Vol. 3 Issue. 3 Computer Aided Engineering (CAE) Techniques Applied To Hip Implant 1, M. S. Abo_Elkhair, 2, M. E. Abo-Elnor,
A THREE-DIMENSIONAL FINITE ELEMENT ANALYSIS OF MANDIBULAR OVERDENTURE SUPPORTED BY CYLINDRICAL AND CONICAL IMPLANTS
A THREE-DIMENSIONAL FINITE ELEMENT ANALYSIS OF MANDIBULAR OVERDENTURE SUPPORTED BY CYLINDRICAL AND CONICAL IMPLANTS Evandro Afonso Sartori, [email protected] Josué Ricardo Broilo Luiz Oscar
A REVIEW ON APPLICATION OF CAD AND FEM TECHNOLOGY IN DESIGN OF TAPER DENTAL IMPLANT
A REVIEW ON APPLICATION OF CAD AND FEM TECHNOLOGY IN DESIGN OF TAPER DENTAL IMPLANT Dhiraj Agnihotri 1, Amol More 2, Dr. P.G.Patil 3 M.Tech CAD/CAM, Department of Mechanical Engineering, G.H.R.C.E,, Nagpur,
Numerical Analysis of Independent Wire Strand Core (IWSC) Wire Rope
Numerical Analysis of Independent Wire Strand Core (IWSC) Wire Rope Rakesh Sidharthan 1 Gnanavel B K 2 Assistant professor Mechanical, Department Professor, Mechanical Department, Gojan engineering college,
Objectives. Experimentally determine the yield strength, tensile strength, and modules of elasticity and ductility of given materials.
Lab 3 Tension Test Objectives Concepts Background Experimental Procedure Report Requirements Discussion Objectives Experimentally determine the yield strength, tensile strength, and modules of elasticity
Dentistry Fatigue test for endosseous dental implants. Report No.: 1 ASTRA System Ti Design
Dentistry Fatigue test for endosseous dental implants Report No.: 1 ASTRA System Ti Design Client: NT - Trading Time of execution: Mai 2011 Subject of work: - - S - 24285 Abutment Ti DesignTM - RESULTS
A Study of Durability Analysis Methodology for Engine Valve Considering Head Thermal Deformation and Dynamic Behavior
A Study of Durability Analysis Methodology for Engine Valve Considering Head Thermal Deformation and Dynamic Behavior Kum-Chul, Oh 1, Sang-Woo Cha 1 and Ji-Ho Kim 1 1 R&D Center, Hyundai Motor Company
METU DEPARTMENT OF METALLURGICAL AND MATERIALS ENGINEERING
METU DEPARTMENT OF METALLURGICAL AND MATERIALS ENGINEERING Met E 206 MATERIALS LABORATORY EXPERIMENT 1 Prof. Dr. Rıza GÜRBÜZ Res. Assist. Gül ÇEVİK (Room: B-306) INTRODUCTION TENSION TEST Mechanical testing
DESS. Screws. Tijuana Ventas: (664) 685 6294/95 [email protected]. For all major implant systems!! www.dabocorp.com
Screws Screws for definitive use. Made from medical grade 5 ELI Titanium Integrity and soundness guaranteed by stress tests of up to 150% of their nominal torques. Their designs ensure a perfect fit with
Sheet metal operations - Bending and related processes
Sheet metal operations - Bending and related processes R. Chandramouli Associate Dean-Research SASTRA University, Thanjavur-613 401 Table of Contents 1.Quiz-Key... Error! Bookmark not defined. 1.Bending
EDEXCEL NATIONAL CERTIFICATE/DIPLOMA MECHANICAL PRINCIPLES OUTCOME 2 ENGINEERING COMPONENTS TUTORIAL 1 STRUCTURAL MEMBERS
ENGINEERING COMPONENTS EDEXCEL NATIONAL CERTIFICATE/DIPLOMA MECHANICAL PRINCIPLES OUTCOME ENGINEERING COMPONENTS TUTORIAL 1 STRUCTURAL MEMBERS Structural members: struts and ties; direct stress and strain,
ALL-CERAMIC DENTAL IMPLANT SOLUTIONS
ALL-CERAMIC DENTAL IMPLANT SOLUTIONS Scientific Evidence Bone-to-implant contact of 78% at 3 months. One piece implant = no prosthetic connections References 1. One-year follow-up of first consecutive
Fatigue Performance Evaluation of Forged Steel versus Ductile Cast Iron Crankshaft: A Comparative Study (EXECUTIVE SUMMARY)
Fatigue Performance Evaluation of Forged Steel versus Ductile Cast Iron Crankshaft: A Comparative Study (EXECUTIVE SUMMARY) Ali Fatemi, Jonathan Williams and Farzin Montazersadgh Professor and Graduate
EDEXCEL NATIONAL CERTIFICATE/DIPLOMA MECHANICAL PRINCIPLES AND APPLICATIONS NQF LEVEL 3 OUTCOME 1 - LOADING SYSTEMS TUTORIAL 3 LOADED COMPONENTS
EDEXCEL NATIONAL CERTIICATE/DIPLOMA MECHANICAL PRINCIPLES AND APPLICATIONS NQ LEVEL 3 OUTCOME 1 - LOADING SYSTEMS TUTORIAL 3 LOADED COMPONENTS 1. Be able to determine the effects of loading in static engineering
Mechanical Properties of Metals Mechanical Properties refers to the behavior of material when external forces are applied
Mechanical Properties of Metals Mechanical Properties refers to the behavior of material when external forces are applied Stress and strain fracture or engineering point of view: allows to predict the
Solid Mechanics. Stress. What you ll learn: Motivation
Solid Mechanics Stress What you ll learn: What is stress? Why stress is important? What are normal and shear stresses? What is strain? Hooke s law (relationship between stress and strain) Stress strain
Ando A., Nakamura Y., Kanbara R., Kumano H., Miyata T., Masuda T., Ohno Y. and Tanaka Y.
11. The Effect of Abutment Tooth Connection with Extracoronal Attachment using the Three Dimensional Finite Element Method - Part 2. The Construction of Finite Element Model from CT Data - Ando A., Nakamura
Stress Strain Relationships
Stress Strain Relationships Tensile Testing One basic ingredient in the study of the mechanics of deformable bodies is the resistive properties of materials. These properties relate the stresses to the
Implants of high-performance ceramic. vk-lange 12.03.2007 14:44 Uhr Seite 1. One-piece precision implant. Made in Germany
OMNIS Implants of high-performance ceramic vk-lange 12.03.2007 14:44 Uhr Seite 1 One-piece precision implant C M Y CM MY CY CMY K Dieter Lange Verkaufsleiter Labor Made in Germany creamed nge 12.03.2007
Optimum Selection of the Dental Implants According to Length and Diameter Parameters by FE Method in the Anterior Position
International Journal of Bioscience, Biochemistry and Bioinformatics, Vol. 4, No. 4, July 2014 Optimum Selection of the Dental Implants According to Length and Diameter Parameters by FE Method in the Anterior
Tooth Replacement Options
Dr. Jordan Johnson Johnson Dental Associates http://www.beta.mydentalhub.com/ada/test/ (800) 947-4746 Tooth Replacement Options If you re missing one or more teeth, you may be all too aware of their importance
Technology of EHIS (stamping) applied to the automotive parts production
Laboratory of Applied Mathematics and Mechanics Technology of EHIS (stamping) applied to the automotive parts production Churilova Maria, Saint-Petersburg State Polytechnical University Department of Applied
Ideal treatment of the impaired
RESEARCH IMPLANTS AS ANCHORAGE IN ORTHODONTICS: ACLINICAL CASE REPORT Dale B. Herrero, DDS KEY WORDS External anchorage Pneumatized Often, in dental reconstruction, orthodontics is required for either
Dental Updates. Excerpted Article e-mail: [email protected]. Why Implant Screws Loosen Part 1. Richard Erickson, MS, DDS
¼ ½ ¾ µ mw/cm 2 Volume 17; 2007 Dental Updates "CUTTING EDGE INFORMATION FOR THE DENTAL PROFESSIONAL " 200 SEMINARS AND 30 JOURNALS REVIEWED YEARLY FOR THE LATEST, CUTTING EDGE INFORMATION Excerpted Article
MECHANICAL PRINCIPLES HNC/D PRELIMINARY LEVEL TUTORIAL 1 BASIC STUDIES OF STRESS AND STRAIN
MECHANICAL PRINCIPLES HNC/D PRELIMINARY LEVEL TUTORIAL 1 BASIC STUDIES O STRESS AND STRAIN This tutorial is essential for anyone studying the group of tutorials on beams. Essential pre-requisite knowledge
FATIGUE CONSIDERATION IN DESIGN
FATIGUE CONSIDERATION IN DESIGN OBJECTIVES AND SCOPE In this module we will be discussing on design aspects related to fatigue failure, an important mode of failure in engineering components. Fatigue failure
Straumann Dental Implant System. Implant Selection Guide.
Straumann Dental Implant System. Implant Selection Guide. STRAUMANN's IMPLANT PORTFOLIO The Straumann Dental Implant System offers two implant lines with diverse body and neck designs ranging from the
ENGINEERING COUNCIL CERTIFICATE LEVEL
ENGINEERING COUNCIL CERTIICATE LEVEL ENGINEERING SCIENCE C103 TUTORIAL - BASIC STUDIES O STRESS AND STRAIN You should judge your progress by completing the self assessment exercises. These may be sent
NARROW DIAMETER implant
ND NARROW DIAMETER implant TABLE OF CONTENTS ND - NARROW DIAMETER implant Implant characteristics page 04 Dental implant page 05 Open Tray Impression Transfer page 06 Titanium Abutments page 07 O-Ball
Modern Tooth Replacement Strategies & Digital Workflow
Modern Tooth Replacement Strategies & Digital Workflow Case Studies by Dr Maurice Salama, DMD AS PUBLISHED BY Dentistry Today, June 2014 Complete Implant Restoration System FACTS: Implant Dentistry Has
While the prosthetic rehabilitation of
Restoring Mandibular Single Teeth with the Inclusive Tooth Replacement Solution Go online for in-depth content by Bradley C. Bockhorst, DMD While the prosthetic rehabilitation of full-arch cases provides
Dental Implants - the tooth replacement solution
Dental Implants - the tooth replacement solution Are missing teeth causing you to miss out on life? Missing teeth and loose dentures make too many people sit on the sidelines and let life pass them by.
EXPERIMENTAL AND NUMERICAL ANALYSIS OF THE COLLAR PRODUCTION ON THE PIERCED FLAT SHEET METAL USING LASER FORMING PROCESS
JOURNAL OF CURRENT RESEARCH IN SCIENCE (ISSN 2322-5009) CODEN (USA): JCRSDJ 2014, Vol. 2, No. 2, pp:277-284 Available at www.jcrs010.com ORIGINAL ARTICLE EXPERIMENTAL AND NUMERICAL ANALYSIS OF THE COLLAR
SCD Case Study. Treatment Considerations for Implant Rehabilitation
SCD Case Study Treatment Considerations for Implant Rehabilitation Multiple surgical and restorative factors play a role in the treatment planning of implant restorations for the edentulous patient (Ali
Force measurement. Forces VECTORIAL ISSUES ACTION ET RÉACTION ISOSTATISM
Force measurement Forces VECTORIAL ISSUES In classical mechanics, a force is defined as "an action capable of modifying the quantity of movement of a material point". Therefore, a force has the attributes
Implant Parts. A Radford Heath Guide http://www.radfordheath.com 1
Implant Parts A Radford Heath Guide http://www.radfordheath.com 1 Disclaimer The information given in this document has been provided in good faith for basic information purposes only and the information
Torsion Tests. Subjects of interest
Chapter 10 Torsion Tests Subjects of interest Introduction/Objectives Mechanical properties in torsion Torsional stresses for large plastic strains Type of torsion failures Torsion test vs.tension test
FRETTING FATIGUE OF STEELS WITH IFFERENT STRENGTH
FRETTING FATIGUE OF STEELS WITH IFFERENT STRENGTH Václav LINHART, Martin ČIPERA, Dagmar MIKULOVÁ SVÚM, a.s., Podnikatelská 565, 190 11 Praha 9- Běchovice,Czech Republic Abstract The investigation of fretting
Chapter Outline. Mechanical Properties of Metals How do metals respond to external loads?
Mechanical Properties of Metals How do metals respond to external loads? Stress and Strain Tension Compression Shear Torsion Elastic deformation Plastic Deformation Yield Strength Tensile Strength Ductility
Implant Systems MAX LOGIC MONO. Today s name in tomorrow s implant technology
Implant Systems MAX LOGIC MONO Today s name in tomorrow s implant technology Implant Systems Etgar Medical Implant Systems brings to the market an entire spectrum of dental implant products based on its
Lap Fillet Weld Calculations and FEA Techniques
Lap Fillet Weld Calculations and FEA Techniques By: MS.ME Ahmad A. Abbas Sr. Analysis Engineer [email protected] www.advancedcae.com Sunday, July 11, 2010 Advanced CAE All contents Copyright
Introduction to Solid Modeling Using SolidWorks 2012 SolidWorks Simulation Tutorial Page 1
Introduction to Solid Modeling Using SolidWorks 2012 SolidWorks Simulation Tutorial Page 1 In this tutorial, we will use the SolidWorks Simulation finite element analysis (FEA) program to analyze the response
AN EXPLANATION OF JOINT DIAGRAMS
AN EXPLANATION OF JOINT DIAGRAMS When bolted joints are subjected to external tensile loads, what forces and elastic deformation really exist? The majority of engineers in both the fastener manufacturing
Tensile Testing of Steel
C 265 Lab No. 2: Tensile Testing of Steel See web for typical report format including: TITL PAG, ABSTRACT, TABL OF CONTNTS, LIST OF TABL, LIST OF FIGURS 1.0 - INTRODUCTION See General Lab Report Format
Validation of Cable Bolt Support Design in Weak Rock Using SMART Instruments and Phase 2
Validation of Cable Bolt Support Design in Weak Rock Using SMART Instruments and Phase 2 W.F. Bawden, Chair Lassonde Mineral Engineering Program, U. of Toronto, Canada J.D. Tod, Senior Engineer, Mine Design
NOTCHES AND THEIR EFFECTS. Ali Fatemi - University of Toledo All Rights Reserved Chapter 7 Notches and Their Effects 1
NOTCHES AND THEIR EFFECTS Ali Fatemi - University of Toledo All Rights Reserved Chapter 7 Notches and Their Effects 1 CHAPTER OUTLINE Background Stress/Strain Concentrations S-N Approach for Notched Members
Stress distribution characteristics in dental implant influenced by its wall thickness
Stress distribution characteristics in dental implant influenced b its wall thickness R.C. van Staden, H. Guan & Y.C. Loo Griffith School of Engineering, Griffith Universit, Gold Coast Campus, Queensland
TEPZZ 69 _ZA T EP 2 692 310 A2 (19) (11) EP 2 692 310 A2. (12) EUROPEAN PATENT APPLICATION published in accordance with Art.
(19) TEPZZ 69 _ZA T (11) EP 2 692 3 A2 (12) EUROPEAN PATENT APPLICATION published in accordance with Art. 13(4) EPC (43) Date of publication: 0.02.14 Bulletin 14/06 (21) Application number: 1276632.0 (22)
TENSILE TESTING PRACTICAL
TENSILE TESTING PRACTICAL MTK 2B- Science Of Materials Ts epo Mputsoe 215024596 Summary Material have different properties all varying form mechanical to chemical properties. Taking special interest in
IMPLANT DENTISTRY EXAM BANK
IMPLANT DENTISTRY EXAM BANK 1. Define osseointegration. (4 points, 1/4 2. What are the critical components of an acceptable clinical trial? (10 points) 3. Compare the masticatory performance of individuals
Oftentimes, as implant surgeons, we are
CLINICAL AVOIDING INJURY TO THE INFERIOR ALVEOLAR NERVE BY ROUTINE USE OF INTRAOPERATIVE RADIOGRAPHS DURING IMPLANT PLACEMENT Jeffrey Burstein, DDS, MD; Chris Mastin, DMD; Bach Le, DDS, MD Injury to the
Strip Flatness Prediction in a 4 High Tandem Mill Using a Dynamic Model.
Strip Flatness Prediction in a 4 High Tandem Mill Using a Dynamic Model. M. A. Bello-Gomez 1, M. P. Guerrero-Mata 1, L. A. Leduc Lezama 1, T. P. Berber- Solano 1, L. Nieves 2, F. Gonzalez 2, H. R. Siller
The Basics of FEA Procedure
CHAPTER 2 The Basics of FEA Procedure 2.1 Introduction This chapter discusses the spring element, especially for the purpose of introducing various concepts involved in use of the FEA technique. A spring
Technical Report Example (1) Chartered (CEng) Membership
Technical Report Example (1) Chartered (CEng) Membership A TECHNICAL REPORT IN SUPPORT OF APPLICATION FOR CHARTERED MEMBERSHIP OF IGEM DESIGN OF 600 (103 BAR) 820MM SELF SEALING REPAIR CLAMP AND VERIFICATION
Fatigue. 3. Final fracture (rough zone) 1. Fatigue origin. 2. Beach marks (velvety zone)
Fatigue Term fatigue introduced by Poncelet (France) 1839 progressive fracture is more descriptive 1. Minute crack at critical area of high local stress (geometric stress raiser, flaws, preexisting cracks)
Measurement of Residual Stress in Plastics
Measurement of Residual Stress in Plastics An evaluation has been made of the effectiveness of the chemical probe and hole drilling techniques to measure the residual stresses present in thermoplastic
Clinical and Laboratory Procedures for Fixed Margin Implant Abutments
Clinical and Laboratory Procedures for Fixed Margin Implant Abutments Dr. Carl Drago DDS, MS, American Board of Prosthodontics Director, Dental Research BIOMET 3i, Adjunct Faculty Department of Prosthodontics,
Considerations on Dental Implants Fixation
MECAHITECH 1 Bucharest, 23-24 September 21 Considerations on Dental Implants Fixation Conf. Sorin Kostrakievici *, As. Mircea Iulian Nistor *, S.l. Doina Bucur *, Medic Cristian Dobreci **, *Politehnica
The SATURN implant by Cortex Dental Industries
The SATURN implant by Cortex Dental Industries By Dr. Zvi Laster DMD W e P r o v e I t E v e r y D a y A case report using a newly designed implant specifically designed for immediate post-extraction loading
Luigi Vito Stefanelli,D.Eng.,DDS
LYON, 27-28/09/2013 Luigi Vito Stefanelli,D.Eng.,DDS Assistant Researcher, Department of Oral Sciences, Sapienza University of Rome, Italy, School of Dentistry. Private practitioner dental office Stasi
PERIPROSTHETIC IMPLANTS
PERIPROSTHETIC IMPLANTS PRODUCT OVERVIEW CLINICAL SOLUTIONS Periprosthetic fractures present unique challenges, such as how to gain fixation when the medullary canal is occupied. Special techniques and
The LOCATOR concept. Simplicity and versatility for prosthesis fixation
The concept Simplicity and versatility for prosthesis fixation The concept Experience the freedom in prosthesis fixation Simple and secure fixation of implant-supported prostheses is essential for successful
The elements used in commercial codes can be classified in two basic categories:
CHAPTER 3 Truss Element 3.1 Introduction The single most important concept in understanding FEA, is the basic understanding of various finite elements that we employ in an analysis. Elements are used for
Information on the product
Information on the product Information on the product -Implantat.doc Page 1 of 15 This document composes of the following items 1. Name of the product 2. Raw materials 3. Structure and dimensions 3.1.
Contents. Cement retained restoration. Screw retained restoration. Overdenture retained restoration. TS Implant System. 70 ComOcta Gold Abutment
Contents TS Implant System Cement retained restoration Screw retained restoration 06 Cement-retained bridges with the Solid abutment system (non- 72 Screw retained crown with the ComOcta Gold abutment
Zimmer Small Fragment Universal Locking System. Surgical Technique
Zimmer Small Fragment Universal Locking System Surgical Technique Zimmer Small Fragment Universal Locking System 1 Zimmer Small Fragment Universal Locking System Surgical Technique Table of Contents Introduction
SEISMIC RETROFITTING TECHNIQUE USING CARBON FIBERS FOR REINFORCED CONCRETE BUILDINGS
Fracture Mechanics of Concrete Structures Proceedings FRAMCOS-3 AEDIFICA TIO Publishers, D-79104 Freiburg, Germany SEISMIC RETROFITTING TECHNIQUE USING CARBON FIBERS FOR REINFORCED CONCRETE BUILDINGS H.
3. Test Methods for Evaluation of ESCR of Plastics
3. Test Methods for Evaluation of ESCR of Plastics A common laboratory request for ESC-prone polymers is to check ESCR performance for quality control, competitive product evaluations, and research and
FEM analysis of the forming process of automotive suspension springs
FEM analysis of the forming process of automotive suspension springs Berti G. and Monti M. University of Padua, DTG, Stradella San Nicola 3, I-36100 Vicenza (Italy) [email protected], [email protected].
Metal Injection Molding (MIM) of components made of Titanium and its alloys
Metal Injection Molding (MIM) of components made of Titanium and its alloys 1 Presentation content Introduction to Metal Injection Molding (MIM) Technology - explaination Products - examples Company brief
Properties of Materials
CHAPTER 1 Properties of Materials INTRODUCTION Materials are the driving force behind the technological revolutions and are the key ingredients for manufacturing. Materials are everywhere around us, and
9. TIME DEPENDENT BEHAVIOUR: CYCLIC FATIGUE
9. TIME DEPENDENT BEHAVIOUR: CYCLIC FATIGUE A machine part or structure will, if improperly designed and subjected to a repeated reversal or removal of an applied load, fail at a stress much lower than
Restoration of a screw retained single tooth restoration in the upper jaw with Thommen Titanium base abutment.
Restoration of a screw retained single tooth restoration in the upper jaw with Thommen Titanium base abutment. Dr. med. dent. David McFadden, Dallas County, USA Initial situation (single X-ray) Tooth 16
The mechanical behavior of the bone microstructure around dental implants
BUDAPEST UNIVERSITY OF TECHNOLOGY AND ECONOMICS DEPARTMENT OF STRUCTURAL MECHANICS The mechanical behavior of the bone microstructure around dental implants Summary and Theses of PhD dissertation ÉVA LAKATOS
Regular C/X Prosthetics. Prosthetics
Regular C/X Prosthetics /X C/ Prosthetics ANKYLOS C/X Prosthetics For more than 20 years, the ANKYLOS system developed by Prof. Dr. G.-H. Nentwig and Dr. Dipl.-Ing. Walter Moser with its TissueCare Connection
2.75 6.525 Problem Set 1 Solutions to ME problems Fall 2013
2.75 6.525 Problem Set 1 Solutions to ME problems Fall 2013 2. Pinned Joint problem Jacob Bayless a) Draw a free-body diagram for the pin. How is it loaded? Does the loading depend on whether the pin is
DENTAL IMPLANT THERAPY
DENTAL IMPLANT THERAPY PATIENT WELCOME PACK Dr. Syed Abdullah BDS, MSc (Dental Implants) What are dental implants? In the early 1950s, a Swedish Scientist, Per-Ingvar Branemark observed that titanium metal
Tensile fracture analysis of blunt notched PMMA specimens by means of the Strain Energy Density
Engineering Solid Mechanics 3 (2015) 35-42 Contents lists available at GrowingScience Engineering Solid Mechanics homepage: www.growingscience.com/esm Tensile fracture analysis of blunt notched PMMA specimens
Figure 1: Typical S-N Curves
Stress-Life Diagram (S-N Diagram) The basis of the Stress-Life method is the Wohler S-N diagram, shown schematically for two materials in Figure 1. The S-N diagram plots nominal stress amplitude S versus
Structural Integrity Analysis
Structural Integrity Analysis 1. STRESS CONCENTRATION Igor Kokcharov 1.1 STRESSES AND CONCENTRATORS 1.1.1 Stress An applied external force F causes inner forces in the carrying structure. Inner forces
Hardened Concrete. Lecture No. 14
Hardened Concrete Lecture No. 14 Strength of Concrete Strength of concrete is commonly considered its most valuable property, although in many practical cases, other characteristics, such as durability
Ridge Reconstruction for Implant Placement
Volume 1, No. 5 July/August 2009 The Journal of Implant & Advanced Clinical Dentistry Ridge Reconstruction for Implant Placement 2 Hours of CE Credit Oral Implications of Cancer Chemotherapy Immediate
Description of mechanical properties
ArcelorMittal Europe Flat Products Description of mechanical properties Introduction Mechanical properties are governed by the basic concepts of elasticity, plasticity and toughness. Elasticity is the
What Dental Implants Can Do For You!
What Dental Implants Can Do For You! Putting Smiles into Motion About Implants 01. What if a Tooth is Lost and the Area is Left Untreated? 02. Do You Want to Restore Confidence in Your Appearance? 03.
Prosthetic treatment planning on the basis of scientific evidence.
Prosthetic treatment planning on the basis of scientific evidence. Pjetursson BE, Lang NP. J Oral Rehabil. 2008 Jan;35 Suppl 1:72-9. Faculty of Odontology, University of Iceland, Reykjavik, Iceland, and
All-on-4 treatment concept with NobelSpeedy Groovy
All-on-4 treatment concept with NobelSpeedy Groovy Product overview Immediate Function for high patient satisfaction Immediately loaded fixed provisional prosthesis on the day of surgery. Immediate improvement
THE COMPOSITE DISC - A NEW JOINT FOR HIGH POWER DRIVESHAFTS
THE COMPOSITE DISC - A NEW JOINT FOR HIGH POWER DRIVESHAFTS Dr Andrew Pollard Principal Engineer GKN Technology UK INTRODUCTION There is a wide choice of flexible couplings for power transmission applications,
BIOMECHANICAL ANALYSIS OF NORMAL AND IMPLANTED TOOTH USING BITING FORCE MEASUREMENT
BIOMECHANICAL ANALYSIS OF NORMAL AND IMPLANTED TOOTH USING BITING FORCE MEASUREMENT B K Biswas 1 S Bag 2 & S Pal 3 1. Dept. of Dental Surgery, Associate Professor, KPC Medical College & Hospital, Kolkata,
Stress Evaluation of Dental Implant Wall Thickness using Numerical Techniques
Stress Evaluation of Dental Implant Wall Thickness using Numerical Techniques Rudi C. van Staden 1, Hong Guan 1, Yew-Chae Loo 1, Newell W. Johnson 2, Neil Meredith 3, 1 Grif th School of Engineering, Grif
ORTHODONTIC MINI IMPLANTS Clinical procedure for positioning. Orthodontics and Implantology
ORTHODONTIC MINI IMPLANTS Clinical procedure for positioning Orthodontics and Implantology 2 All rights are reserved. Any reproduction of the present publication is prohibited in whole or in part and by
Lentur Cable System. Surgical Technique
Lentur Cable System Surgical Technique Contents Introduction... Page 1 System Design Features And Benefits... Page 2 Implants... Page 3 Instrumentation... Page 4 Surgical Technique... Page 5 Single Cable...
Tensile Testing Laboratory
Tensile Testing Laboratory By Stephan Favilla 0723668 ME 354 AC Date of Lab Report Submission: February 11 th 2010 Date of Lab Exercise: January 28 th 2010 1 Executive Summary Tensile tests are fundamental