Cumhuriyet Dental Journal
|
|
|
- Ross Caldwell
- 9 years ago
- Views:
Transcription
1 Cumhuriyet Dental Journal Volume 18 Issue 1 doi: /cdj available at RESEARCH ARTICLE Numerical analysis of the effect of implant geometry to stress distributions of the three different commercial dental implant system Tolga Topkaya, MSc a, Murat Yavuz Solmaz, MSc, PhD b, Serkan Dündar, DDS, PhD c, Abubekir Eltas,DDS, PhD d a Department of Mechanical Engineering, Faculty of Engineering, Batman University, Batman, Turkey b Department of Mechanical Engineering, Faculty of Engineering, Firat University, Elazığ, Turkey c Department of Periodontics, Faculty of Dentistry, Fırat University, Elazığ, Turkey d Department of Periodontics, Faculty of Dentistry, İnonu University, Malatya, Turkey ARTICLE INFO Article history: Received 25 June 2014 Accepted 19 October 2014 Keywords: Finite element analysis, Dental implant design, Stress distribution ABSTRACT Objectives: The success of dental implants is related to the quality, quantity of local bones, implant design and surgical technique. Implant diameter and length are accepted as key factors. Present work focuses to investigate the effect of titanium implant geometry to stress distributions in implant system. Materials and Methods: For this purpose three different implant models which are currently being used in clinical cases constructed by using ANSYS Workbench The stress distributions on components of implant system under static loadings were analyzed for all models. Results: The maximum stress values that occurred in all components happen in the case of loading in which the Nucleoss T-4 (Nucleoss, Turkiye) implant is used, but the occurred lowest stress values happen in the case of Fı loading in which Nobel Active (Nobel Biocare, Zurich, Switzerland) implant is used. In all models, the maximum tensions have occurred in the neck region of the implants. Conclusions: The crestal bone loss in the neck region of the implants reduced the long-term survival rate of implants.the length and the size of the implant are the two important factors in the stress distribution. INTRODUCTION Since the osseointegration of Brenemark 1 was defined as the directly structural and functional connection, without having a fibrous tissue between the living bone tissue and implant surface under loading in1960s, the dental implant-supported prosthesis have been scientifically accepted and have Correspondence to: Dr.Serkan Dundar, Fırat University, Faculty of Dentistry, Department of Periodontics, 23119, Elazıg, Campus, Turkey. [email protected], [email protected] Note: A part of the information presented in this study was reported (oral presentation) at the 11 th International Symposium, Computer Methods in Biomechanics and Biomedical Engineering April 3-7, 2013, Salt Lake City, Utah, USA 17
2 been a commen treatment choice in the case reconstructing of partial or total tooth loss. 2 While the basic reason of early term loss of implant in endosseous dental implantation is the infection of periimplant tissues, the basic reason of implant-loss after the loading and osseointegration is the loss of bone which occurs on the implant neck region. The occurance of loss of periimplant bone after loading derives from the excessive stress that comes along the long axis of implant and/or having a wrong direction. The type of stress, the features of materials of which the implants and prosthesis are made, the implant geometry and it s surface structure, the quality and the quantity of bones around implant, the structure between bone and implant are the factors that determine the stress which affect the bones around implants. 3 Numerous implant concepts and implant types which have different shapes, dimensions, materials and different surface features have been presented to be use. 3-7 Determining the reasons of the loss of implants, the analysis of the mechanic relations between bone and implant are important for planning of an effective, useful and dependable implant system. 4,5 FEM (Finite Element Method) has taken part in literature as being a useful method to determine the tensions that occur on bone-implant intersurface during mechanic loading. With the help of FEM analysis, it has s suggested that the highest stress rate in endosseous dental implantation occurs in the occlusal part of cortical bone around implant. 4 Many researchers have done lots of studies to increase the contact area on between the bone and implant intersurface and to lessen the crestal bone loss by diminishing the stress which affects the cortical alveolar bone. 4,6,7,20,21 The studies which aim to increase the connection fields on the bone and implant intersurface concentrate on the size of implants, the implant geometry and/or the length of the implants. This study focuses to investigate the effect of titanium implant geometry to stress distributions in implant system. MATERIALS & METHODS In this study, the implant and mandible, of which solid models were formed by using Solidworks Programme, were transferred to ANYSY Workbench Programme and finite element analysis were realized. The mandible has been modeled in Solidworks Programme with the help of computed tomography images. Figure 1 shows, the images obtained from computed tomography datums, Figure 2 shows, the three-dimensional model of the mandible. In implant applications, the three implants used clinically have been modeled with the help of manufacturer data. The used implant models have been given in Figure 3. Nucleoss T4 (Nucleoss, Izmir, Turkiye), Nobel Replace (Nobel Biocare, Zurich, Switzerland) and Nobel Active (Nobel Biocare, Zurich, Switzerland) dental implants has been choosen owing to being a few of the types of commercial implants frequently used in the clinics of our country. While the Nobel Replace (Nobel Biocare, Zurich, Switzerland) is a tissue level implant; Nucleoss T4 (Nucleoss, Izmir, Turkiye) and Nobel Active (Nobel Biocare, Zurich, Switzerland) are bone level implants. All three implants have grooves in the neck of them. The apical is right angled in all three implants. While Nobel Replace (Nobel Biocare, Zurich, Switzerland) and the Nobel Active (Nobel Biocare, Zurich, Switzerland) implants have v-shaped grooves, Nucleoss T4 (Nucleoss, Izmir, Turkiye) implant has double hellical groove. Nucleoss T4 (Nucleoss, Izmir, Turkiye) implant is smaller than Nobel Replace (Nobel Biocare, Zurich, Switzerland) and Nobel Active (Nobel Biocare, Zurich, Switzerland) implants in diameter and length. 18
3 Figure 1. Computed tomography images Figure 2. The three-dimensional model of the mandible The constructed mandible models were divided into finite elements and boundry conditions and they were applied by using the mounting of porselein prosthesis, implant models, abutment, metal support and adhesive layer. In the Table 1, the mechanic features of the used materials have been given. Dental implants are exposed to different loads. Three different loads were applied to models in order to examine the effect of different loads to tension distributions. In Figure 4, the application of the load to models has been given. In the figure, θ shows the angle degree of the load, F shows the amount of the load. If the applied loads are vertical, Fı is 100 N 4, if F at an agle of 15 degrees with the vertical axis, Fıı 100 N 6, if F at an angle of 30 degrees with the vertical axis, Fııı 150 N. 8 RESULTS When the result of the work were examined, the maximum stress values were obtained in Nucleoss T-4 (Nucleoss, Izmir, Turkiye) for all components the lowest stress values were obtained in Nobel Active (Nobel Biocare, Zurich, Switzerland). In Figure 5, abutment and screws which are the primary elements that exposed to stress are shown. When we look the figure, we see. Additionally, in the model in which only Nucleoss T-4(Nucleoss, Izmir, Turkiye) implant is used, it has been seen that the stress occuring in abutment during Fıı and Fııı loading is higher than the stress value occuring in implant. Within 3 different kinds of implant, the stress distributions of Fııı during the loading in which value of maximum stress occurs has been given in Figure 6. When we examine the stress distributions, we notice that the stresses concentrate on the neck region of the implant and the maximum stress values occur on the neck region of the implant in all three implants. DISCUSSION FEM is a numerical method that is used to analyze the stress and deformations occuring in the structure of a geometrical 19
4 Table 1. The mechanical features of the used materials have been given. a b Figure 3. The implant models used in this study: a- Nobel Replace (Nobel Biocare, Zurich, Switzerland). b- Nucleoss T-4 (Nucleoss,Izmir,Turkiye), c- Nobel Active (Nobel Biocare, Zurich, Switzerland). c method used in the evalution of various clinic and protetic choices. 3,4,6-8 This method aims to solve complex problems with mathematical methods by dividing them into simple and small structures related with each other. 3 Figure 4. The application of the load to models model. FEM has taken part in implant tecnology as not only being a method used in root-form implants and for the evalution and analysis of forces that affect bone/implant interface but also being a In this study, three different commercial osseintegrated implants were examined with FEM. All the studied implants on the same model have been planted on the mandible first molar region. Until now in most studies, the changes occurred in implant and surrounding bone have been examined by looking the stress values of Von Misses. 9, 10 So, we approved to examine Von Misses stress values based on these studies. The reason why biomechanical behaviours obtained from this study are 20
5 Figure 5. The maximum stress values occuring in implant, abutment and screws a b Figure 6. The stress distributions occuring in implants (a- Nobel Replace (Nobel Biocare, Zurich, Switzerland), b- Nucleoss T-4(Nucleoss,Izmir,Turkiye), c- Nobel Active (Nobel Biocare, Zurich, Switzerland)). c different is that because implant systems have different parameter shapes. The reason why the obtained results are different for all is that, in accordance with the other studies, the transmission of stress between bone and implant is thougt to be basically affected by the shape and size of 11, 12 implant. According to former studies done, the length and diameter of the implant are emphasized to be the two important parameters on the stress distribution to bone around implant. 13 In accordance with this, the basic reason of the stresses occurred in in Nucleoss T-4 (Nucleoss, Izmir, Turkiye) model to be higher than the other models is that their length and diameter values are smaller than the other models. The amount and distributions of the stresses coming to bone/implant interface are affected by the length of the implant. 14 In FEM analysis done in former studies, it was put forward that an increase in implant length causes a decrease in crestal bone stress values. 6 In accordance with this results, the clinic studies state that, contrary to short implants, the long ones can be kept in the mouth for a longer time. 15, 16 At the same time, there are studies which point that the diameter of 21
6 implant is more important than the length of the implant, too. 8, 13, 17 According to FEM studies, the possible reason of this result is that because the stress distribution in bone socket is uneven and the bone on implant neck exposed to maximum stress. 17 In FEM studies on titanium implants, it is informed that stress intensity occurs in implant neck region. 18 In this study, stresses coming out in implants concentrate on implant neck region and the maximum Von-Misses stress values emerge on these points too. The result obtained from clinic and histomorphometric studies is that bone loss occurred on implant neck region is an important parameter in implant loss during the period after loading In Nucleoss T-4 (Nucleoss, Izmir, Turkiye) model which is smaller in diameter than the other implants, the stress of neck region is higher. The implants which have a larger neck region can resist more to masticatory forces, can be kept in mouth longer-terms and can cause less crestal bone loss. 4,17 In literature, when three-dimensional finite element analysis on dental implants are examined, it is seen that the non-axial forces are more destructive than the axial forces, and non-axial forces couse more stress accumulation in periimplant bone than axial forces do. 20, 21 From this study, in all three implant types, it is clearly seen that a rise in loading force angle with vertical axis increases the stress distribution in implant system. These results are consistent with previous studies. In addidion to excessive stress in all forces in Nucleos T-4 (Nucleoss, Izmir, Turkiye) model, the reason of the stresses coming out in abutments in especially Fıı and Fııı forces is, as mentioned previously, Nucleos T-4 (Nucleoss, Izmir, Turkiye) implant are smaller in length and diameter than the others. 12,13 In most single molar implant-supported prostheses, the bone loss occurred as a result of excessive oclusal loading have been stated to come ahead of implant structures. 22 As the increased oclusal forces can couse implant losses by leading to crestal bone loss, it can also couse abutment and implant losses by leading to abutment screw loosening. 23 In implant losses after loading, implant neck fractures should be taken in consideration. The writers have determined an implant fracture rate of % 14 in a study on standard Nobel Biocare (Zurich, Switzerland) implant-supported single molars. 22 This study has been done on mandible posterior region. When obtained results have been examined, maximum stress values occurred in all components have been determined to come out in Nucleoss T-4 (Nucleoss, Izmir, Turkiye) implant-used model in the case of Fııı loading. The reoccurance of most stresses in abutment in both Fıı and Fııı forces that have angle with vertical axis, the increasing of stress distributions in direct proportion to the increase in the angle with vertical load are remarkable in terms of axial forces. Choosing largediameter and long implants by taking the length of bone and anatomic limitations into account because of excessive forces occuring during the selection of implant on mandible posterior region can increase the survival rate of the implant in the mouth. Large-diameter and long implants used on mandible posterior region can perform better biomechanical properties with the increasing surface, increasing fracture resistance and increasing abutment stability. 23 CONCLUSION Based on the limited results of this study, we can say that implant geometry is an important parameter in the distribution of biomechanic forces on implant and periimplant tissues. High-emergence of stress distribution, in implants which are smaller in length and diameter put 22
7 forward that length and diameter are the two important factors in the distribution of forces. In all models, maximum stresses occur on implant neck region. The crestal bone loss on implant neck region diminishes the survival rate of implants in the mouth in long terms. In the implants which are exposed to more masticatory force on lower jaw posterior regions, especially the use of large-diameter and long implants can be evaluated as parameters which increase the survival rate in the mouth. There is need of more laboratory and clinical studies in order to obtain appropriate shapegeometry that provide longer term clinical use of implants. Disclosure: We have not a financial relationship with the organization that sponsored the research. REFERENCES 1. Branemark PI, Adell R, Breine U, Hansson BO, Lindstrom J, Ohlsson A. Intra-osseous anchorage of dental prostheses. I. Experimental studies. Scand J Plast Reconstr Surg 1969;3: Buser D, Mericske-Stern R, Bernard JP, Behneke A, Behneke N, Hirt HP, et al. Long-term evaluation of non-submerged ITI implants. Part 1: 8-year life table analysis of a prospective multi-center study with 2359 implants. Clin Oral Implants Res 1997;8: Bozkaya D, Muftu S, Muftu A. Evaluation of load transfer characteristics of five different implants in compact bone at different load levels by finite elements analysis. J Prosthet Dent 2004;92: Petrie CS, Williams JL. Comparative evaluation of implant designs: influence of diameter, length, and taper on strains in the alveolar crest. A three-dimensional finite-element analysis. Clin Oral Implants Res 2005;16: Acar A, İnan Ö. İmplant destekli protezlerde okluzyon. Cumhuriyet Üniversitesi Dişhekimliği Fakültesi Dergisi 2001;4,1: Chun HJ, Cheong SY, Han JH, Heo SJ, Chung JP, Rhyu IC, et al. Evaluation of design parameters of osseointegrated dental implants using finite element analysis. J Oral Rehabil 2002;29: Tada S, Stegaroiu R, Kitamura E, Miyakawa O, Kusakari H. Influence of implant design and bone quality on stress/strain distribution in bone around implants: a 3-dimensional finite element analysis. Int J Oral Maxillofac Implants 2003;18: Anitua E, Tapia R, Luzuriaga F, Orive G. Influence of implant length, diameter, and geometry on stress distribution: a finite element analysis. Int J Periodontics Restorative Dent 2010;30: Teixeira ER, Sato Y, Akagawa Y, Shindoi N. A comparative evaluation of mandibular finite element models with different lengths and elements for implant biomechanics. J Oral Rehabil 1998;25: Geng JP, Ma QS, Xu W, Tan KB, Liu GR. Finite element analysis of four thread-form configurations in a stepped screw implant. J Oral Rehabil 2004;31: Akpinar I, Demirel F, Parnas L, Sahin S. A comparison of stress and strain distribution characteristics of two different rigid implant designs for distal-extension fixed prostheses. Quintessence Int 1996;27: Holmgren EP, Seckinger RJ, Kilgren LM, Mante F. Evaluating parameters of osseointegrated dental implants using finite element analysis- -a two-dimensional comparative study examining the effects of 23
8 implant diameter, implant shape, and load direction. J Oral Implantol 1998;24: Van Staden RC, Guan H, Loo YC. Application of the finite element method in dental implant research. Comput Methods Biomech Biomed Engin 2006;9: Baggi L, Cappelloni I, Di Girolamo M, Maceri F, Vairo G. The influence of implant diameter and length on stress distribution of osseointegrated implants related to crestal bone geometry: a three-dimensional finite element analysis. J Prosthet Dent 2008;100: Winkler S, Morris HF, Ochi S. Implant survival to 36 months as related to length and diameter. Ann Periodontol 2000;5: Ivanoff CJ, Grondahl K, Sennerby L, Bergstrom C, Lekholm U. Influence of variations in implant diameters: a 3- to 5-year retrospective clinical report. Int J Oral Maxillofac Implants 1999;14: Himmlova L, Dostalova T, Kacovsky A, Konvickova S. Influence of implant length and diameter on stress distribution: a finite element analysis. J Prosthet Dent 2004;91: Geng JP, Tan KB, Liu GR. Application of finite element analysis in implant dentistry: a review of the literature. J Prosthet Dent 2001;85: Wiskott HW, Belser UC. Lack of integration of smooth titanium surfaces: a working hypothesis based on strains generated in the surrounding bone. Clin Oral Implants Res 1999;10: Papavasiliou G, Kamposiora P, Bayne SC, Felton DA. Three-dimensional finite element analysis of stressdistribution around single tooth implants as a function of bony support, prosthesis type, and loading during function. J Prosthet Dent 1996;76: Kitamura E, Stegaroiu R, Nomura S, Miyakawa O. Influence of marginal bone resorption on stress around an implant-a three-dimensional finite element analysis. J Oral Rehabil 2005;32: Rangert B, Krogh PH, Langer B, Van Roekel N. Bending overload and implant fracture: a retrospective clinical analysis. Int J Oral Maxillofac Implants 1995;10: Misch CE. Contemporaray Implant Dentisrty. Mosby Elsevier 2008;Third edition:337 How to cite this article: Tolga Topkaya, Murat Yavuz Solmaz, Serkan Dündar, Abubekir Eltas. Numerical analysis of the effect of implant geometry to stress distributions of the three different commercial dental implant system. Cumhuriyet Dent J 2015;18(1):
A THREE-DIMENSIONAL FINITE ELEMENT ANALYSIS OF MANDIBULAR OVERDENTURE SUPPORTED BY CYLINDRICAL AND CONICAL IMPLANTS
A THREE-DIMENSIONAL FINITE ELEMENT ANALYSIS OF MANDIBULAR OVERDENTURE SUPPORTED BY CYLINDRICAL AND CONICAL IMPLANTS Evandro Afonso Sartori, [email protected] Josué Ricardo Broilo Luiz Oscar
A REVIEW ON APPLICATION OF CAD AND FEM TECHNOLOGY IN DESIGN OF TAPER DENTAL IMPLANT
A REVIEW ON APPLICATION OF CAD AND FEM TECHNOLOGY IN DESIGN OF TAPER DENTAL IMPLANT Dhiraj Agnihotri 1, Amol More 2, Dr. P.G.Patil 3 M.Tech CAD/CAM, Department of Mechanical Engineering, G.H.R.C.E,, Nagpur,
Life Table Analysis for Evaluating Curative-effect of One-stage Non-submerged Dental Implant in Taiwan
Journal of Data Science 6(2008), 591-599 Life Table Analysis for Evaluating Curative-effect of One-stage Non-submerged Dental Implant in Taiwan Miin-Jye Wen 1, Chuen-Chyi Tseng 2 and Cheng K. Lee 3 1 National
Outcomes of Placing Short Dental Implants in the Posterior Mandible: A Retrospective Study of 124 Cases
DENTAL IMPLANTS J Oral Maxillofac Surg 67:713-717, 2009 Outcomes of Placing Short Dental in the Posterior Mandible: A Retrospective Study of 124 Cases Bao-Thy N. Grant, DDS,* Franklin X. Pancko, DDS, and
Influence of Bone and Dental Implant Parameters on Stress Distribution in the Mandible: A Finite Element Study
Influence of Bone and Dental Parameters on Stress Distribution in the Mandible: A Finite Element Study Hong Guan, PhD 1 /Rudi van Staden, BEng 2 /Yew-Chaye Loo, PhD 3 / Newell Johnson, PhD 4 /Saso Ivanovski,
Clinical and Laboratory Procedures for Fixed Margin Implant Abutments
Clinical and Laboratory Procedures for Fixed Margin Implant Abutments Dr. Carl Drago DDS, MS, American Board of Prosthodontics Director, Dental Research BIOMET 3i, Adjunct Faculty Department of Prosthodontics,
COMPARISON OF SHORT DENTAL IMPLANTS UNDER DIFFERENT LOADING CONDITIONS AND LOCATIONS
Süleyman Demirel Üniversitesi Mühendislik Bilimleri ve Tasarım Dergisi 2(3),ÖS:BiyoMekanik2014, 261-265, 2014 ISSN: 1308-6693 Suleyman Demirel University Journal of Engineering Sciences and Design 2(3),SI:BioMechanics2014,
Prosthetic treatment planning on the basis of scientific evidence.
Prosthetic treatment planning on the basis of scientific evidence. Pjetursson BE, Lang NP. J Oral Rehabil. 2008 Jan;35 Suppl 1:72-9. Faculty of Odontology, University of Iceland, Reykjavik, Iceland, and
Rehabilitation of Endondontically Failed Anterior teeth by Immediate Replacement and Loading of an Implant supported Crown: A Case Report.
RESEARCH AND REVIEWS: JOURNAL OF DENTAL SCIENCES Rehabilitation of Endondontically Failed Anterior teeth by Immediate Replacement and Loading of an Implant supported Crown: A Case Report. Lalit Kumar 1
Placement of dental implants is a highly successful
Shape Optimization of Dental Implants Li Shi, BEng, MSc 1 /Haiyan Li, BEng, MSc, PhD 2 /Alex S. L. Fok, BEng, MSc, PhD, CEng 3 /Cemal Ucer, BDS, MSc 4 / Hugh Devlin, BDS, BSc, MSc, PhD 5 /Keith Horner,
BENDABLE VS. ANGULATED DENTAL IMPLANTS: CONSIDERATION OF ELASTIC AND PLASTIC MATERIAL PROPERTIES BASED ON EXPERIMENTAL IMPLANT MATERIAL DATA AND FEA
Biomed Pap Med Fac Univ Palacky Olomouc Czech Repub. 2008, 152(2):309 316. T. Goldmann, S. Ihde, J. Kuzelka, L. Himmlova 309 BENDABLE VS. ANGULATED DENTAL IMPLANTS: CONSIDERATION OF ELASTIC AND PLASTIC
Stress Distribution in Mandible Regulated by Bone and Dental Implant Parameters: Part I - Methodology
Stress Distribution in Mandible Regulated by Bone and Dental Implant Parameters: Part I - Methodology R. C. van Staden 1, H. Guan 1, Y. C. Loo 1, N. W. Johnson 2, N. Meredith 3 1 Griffith School of Engineering,
Influence of Diameter and Length of Implant on Early Dental Implant Failure
J Oral Maxillofac Surg 68:414-419, 2010 Influence of Diameter and Length of Implant on Early Dental Implant Failure Sergio Olate, DDS, MSc,* Mariana Camilo Negreiros Lyrio, DDS, MSc, Márcio de Moraes,
The Effect of Dental Implant Materials and Thread Profiles A Finite Element and Statistical Study
RESEARCH ARTICLE Copyright 2013 American Scientific Publishers All rights reserved Printed in the United States of America Journal of Medical Imaging and Health Informatics Vol. 3, 1 5, 2013 The Effect
Restoration of the Edentulous Maxilla: The Case for the Zygomatic Implants
CLINICAL CONTROVERSIES IN ORAL AND MAXILLOFACIAL SURGERY: PART ONE J Oral Maxillofac Surg 62:1418-1422, 2004 Restoration of the Edentulous Maxilla: The Case for the Zygomatic Implants Eric D. Ferrara,
IMPLANT DENTISTRY EXAM BANK
IMPLANT DENTISTRY EXAM BANK 1. Define osseointegration. (4 points, 1/4 2. What are the critical components of an acceptable clinical trial? (10 points) 3. Compare the masticatory performance of individuals
BIOMECHANICAL ANALYSIS OF NORMAL AND IMPLANTED TOOTH USING BITING FORCE MEASUREMENT
BIOMECHANICAL ANALYSIS OF NORMAL AND IMPLANTED TOOTH USING BITING FORCE MEASUREMENT B K Biswas 1 S Bag 2 & S Pal 3 1. Dept. of Dental Surgery, Associate Professor, KPC Medical College & Hospital, Kolkata,
Retrospective study on the survival rate of IBS implant
Retrospective study on the survival rate of IBS implant Date : 30. 05. 2013 Written by : Dr. Je Won Wang, Director of research Approved by : Prof. Min Seung Ki - Contents - 1. Purpose Of Study 2. Materials
Ridge Reconstruction for Implant Placement
Volume 1, No. 5 July/August 2009 The Journal of Implant & Advanced Clinical Dentistry Ridge Reconstruction for Implant Placement 2 Hours of CE Credit Oral Implications of Cancer Chemotherapy Immediate
RESEARCH. Key Words: crown-implant ratios, proximal bone-to-implant contact. Journal of Oral Implantology 425
RESEARCH Crown-to-Implant Ratios of Short-Length Implants Hardeep Birdi, DMD, MS 1 * John Schulte, DDS, MSD 1 Alejandro Kovacs, DDS, MS 1 Meghan Weed, RDH 2 Sung-Kiang Chuang, DMD, MD 3 Excessive crown-implant
Radiographic Vertical Bone Loss Evaluation around Dental Implants Following One Year of Functional Loading
Original Article Radiographic Vertical Bone Loss Evaluation around Dental Implants Following One Year of Functional Loading AAR. Rasouli Ghahroudi 1, AR Talaeepour 2, A. Mesgarzadeh 3, AR. Rokn 4, 5, A.
Leslie Laing Gibbard, BSc, BEd, MSc, PhD, DDS George Zarb, BChD, DDS, MS, MS, FRCD(C)
A P P L I E D R E S E A R C H A 5-Year Prospective Study of Implant-Supported Single-Tooth Replacements Leslie Laing Gibbard, BSc, BEd, MSc, PhD, DDS George Zarb, BChD, DDS, MS, MS, FRCD(C) A b s t r a
The Impact of Oral Implants Past and Future, 1966 2042. Tomas Albrektsson, MD, PhD, ODhc Ann Wennerberg, DDS, PhD
P R O F E S S I O N A L I S S U E S The Impact of Oral Implants Past and Future, 1966 202 Tomas Albrektsson, MD, PhD, ODhc Ann Wennerberg, DDS, PhD A b s t r a c t This paper traces the history of oral
Stress distribution characteristics in dental implant influenced by its wall thickness
Stress distribution characteristics in dental implant influenced b its wall thickness R.C. van Staden, H. Guan & Y.C. Loo Griffith School of Engineering, Griffith Universit, Gold Coast Campus, Queensland
Dental Implant Treatment after Improvement of Oral Environment by Orthodontic Therapy
Dental implant treatment after impr Title environment by orthodontic therapy. Sekine, H; Miyazaki, H; Takanashi, Author(s) Matsuzaki, F; Taguchi, T; Katada, H Journal Bulletin of Tokyo Dental College,
While the prosthetic rehabilitation of
Restoring Mandibular Single Teeth with the Inclusive Tooth Replacement Solution Go online for in-depth content by Bradley C. Bockhorst, DMD While the prosthetic rehabilitation of full-arch cases provides
Straumann Dental Implant System. Implant Selection Guide.
Straumann Dental Implant System. Implant Selection Guide. STRAUMANN's IMPLANT PORTFOLIO The Straumann Dental Implant System offers two implant lines with diverse body and neck designs ranging from the
Supervisors: Dr. Farhan Raza Khan
1 Presenter: Dr. Sana Ehsen Supervisors: Dr. Farhan Raza Khan 2 A dental implant (also known as an endosseous implant or fixture) is a surgical component that interfaces with the bone of the jaw to support
Straumann. Time-tested
Straumann Soft TiSSue Level implant System Time-tested Why a Soft Tissue Level implant? Simplicity and efficiency by integrated soft tissue management Straumann Soft Tissue Level implants have a built-in
Histologic comparison of biologic width around teeth versus implants: The effect on bone preservation
Clinical Histologic comparison of biologic width around teeth versus implants: The effect on bone preservation Kazuto Makigusa 1 Abstract Histological analysis of the biological width surrounding primate
Implant therapy using osseointegrated implants has
Natural Tooth Intrusion Phenomenon With Implants: A Survey Lily T. Garcia, DDS, MS*/Larry J. Oesterle, DDS, MS** A common assumption when planning for treatment for a fixed partial denture potentially
Zimmer Trabecular Metal Dental Implant
Zimmer Trabecular Metal Dental Implant TRABECULAR METAL MATERIAL: Designed to Enhance Secondary Stability Through Bone Ingrowth. Artistic Rendering osseoincorporation TRABECULAR METAL MATERIAL Designed
Optimum Selection of the Dental Implants According to Length and Diameter Parameters by FE Method in the Anterior Position
International Journal of Bioscience, Biochemistry and Bioinformatics, Vol. 4, No. 4, July 2014 Optimum Selection of the Dental Implants According to Length and Diameter Parameters by FE Method in the Anterior
IMPLANT DENTISTRY CLINICAL SYLLABUS
IMPLANT DENTISTRY CLINICAL SYLLABUS RESD 535 MATS H. KRONSTROM, DDS, PhD Course Director DEPARTMENT OF RESTORATIVE DENTISTRY SCHOOL OF DENTISTRY UNIVERSITY OF WASHINGTON 2008 2009 RD 535 - SCHEDULE FALL
Stress Evaluation of Dental Implant Wall Thickness using Numerical Techniques
Stress Evaluation of Dental Implant Wall Thickness using Numerical Techniques Rudi C. van Staden 1, Hong Guan 1, Yew-Chae Loo 1, Newell W. Johnson 2, Neil Meredith 3, 1 Grif th School of Engineering, Grif
Stress analysis of different prosthesis materials in implant-supported fixed dental prosthesis using 3D finite element method
Original Research Stress analysis of different prosthesis materials in implant-supported fixed dental prosthesis using 3D finite element method Pedram Iranmanesh 1, Alireza Abedian 2, Naeimeh Nasri 1,
Flapless Implant Surgery for Replacement of Posterior Teeth
Course Number: 108.2 Flapless Implant Surgery for Replacement of Posterior Teeth Authored by J. Steven Cloyd, DDS Upon successful completion of this CE activity 1 CE credit hour may be awarded A Peer-Reviewed
TRI Product NewsFlash. December 2015
TRI Product NewsFlash December 2015 Study Overview 2015 Dear Partners Year in, year out, we are screening all major scientific journals to ensure that our TRI Performance Concept still reflects the latest
Importance of Crown to Root and Crown to Implant Ratios
Course Number: 135 Importance of Crown to Root and Crown to Implant Ratios Authored by Gary Greenstein, DDS, MS, and John S. Cavallaro Jr, DDS Upon successful completion of this CE activity 2 CE credit
The maintenance of crestal bone around dental implants
I clinical technique_ crestal bone loss The maintenance of crestal bone around dental Author_Mohammed A. Alshehri, BDS Fig.1_The heat generated during drilling and excessive pressure at the crestal region
Load Fatigue Performance of Four Implant- Abutment Interface Designs: Effect of Torque Level and Implant System
Load Fatigue Performance of Four Implant- Abutment Interface Designs: Effect of Torque Level and Implant System H. C. Quek, BDS, MDS, MRD, RCSEd 1 /Keson B. Tan, BDS, MSD 2 /Jack I. Nicholls, PhD 3 Purpose:
Guided bone regeneration (GBR) has made the
Case Report 684 Guided Bone Regeneration for Fenestration Defects in Dental Implants Hwey-Chin Yeh, DDS, MS; Kuang-Wei Hsu 1, DDS has been applied in implant dentistry for increasing the width and height
Implant rehabilitation in the edentulous jaw: the All-on-4 immediate function concept
Implant rehabilitation in the edentulous jaw: the All-on-4 immediate function concept By Christopher CK Ho, BDS (Hons), Grad.Dip.Clin.Dent (Implants), M.Clin.Dent (Pros) The All-on-4 technique involves
Dental Implant Options in Atrophic Jaws
Dental Implant Options in Atrophic Jaws Orthopedic Application Jay B. Reznick, D.M.D., M.D. Diplomate, American Board of Oral and Maxillofacial Surgery Tarzana, CA Endopore Dental Implant System Screw-Type
Ando A., Nakamura Y., Kanbara R., Kumano H., Miyata T., Masuda T., Ohno Y. and Tanaka Y.
11. The Effect of Abutment Tooth Connection with Extracoronal Attachment using the Three Dimensional Finite Element Method - Part 2. The Construction of Finite Element Model from CT Data - Ando A., Nakamura
Implant occlusion: biomechanical considerations for implant-supported prostheses
Review Article Implant occlusion: biomechanical considerations for implant-supported prostheses YU-YING CHEN 1,2 CHUNG-LING KUAN 1 YI-BING WANG 1,2 1 Department of Dentistry, Tri-Service General Hospital,
Telescopic Denture A Treatment Modality for Minimizing the Conventional Removable Complete Denture Problems: A Case Report
Dentistry Section Case Report ID: JCDR/2012/3886:2351 Telescopic Denture A Treatment Modality for Minimizing the Conventional Removable Complete Denture Problems: A Case Report Kunwarjeet Singh, Nidhi
LATERAL BONE EXPANSION FOR IMMEDIATE PLACEMENT OF ENDOSSEOUS DENTAL IMPLANTS
LATERAL BONE EXPANSION FOR IMMEDIATE PLACEMENT OF ENDOSSEOUS DENTAL IMPLANTS Department of Oral Maxillofacial Surgery, Chisinau Abstract: The study included 10 using the split control expansion technique
Healing Abutment Selection. Perio Implant Part I. Implant Surface Characteristics. Single Tooth Restorations. Credit and Thanks for Lecture Material
Healing Abutment Selection Perio Implant Part I Credit and Thanks for Lecture Material Implant Surface Characteristics!CAPT Robert Taft!CAPT Greg Waskewicz!Periodontal Residents NPDS and UMN!Machined Titanium!Tiunite!Osseotite
Quality from Switzerland 01.01.2014
Quality from Switzerland 01.01.2014 TRI Dental Implants: An example of the highest standards of quality We deliver quality We supply quality. From patent to patient Modern production Compliance with the
Reconstruction of the anterior maxilla with implants using customized zirconia abutments and all-ceramic crowns: a clinical case report
Vol. 34 No. 2, September 2015 Reconstruction of the anterior maxilla with implants using customized zirconia abutments and all-ceramic crowns: a clinical case report Sang-ki Byun, Yung-bin Lee, Woohyun
Implants for life? A critical review of implant-supported restorations
journal of dentistry 35 (2007) 768 772 available at www.sciencedirect.com journal homepage: www.intl.elsevierhealth.com/journals/jden Review Implants for life? A critical review of implant-supported restorations
Straumann Dental implant system ONE SYSTEM ONE INSTRUMENT KIT ALL INDICATIONS*
Straumann Dental implant system ONE SYSTEM ONE INSTRUMENT KIT ALL INDICATIONS* For all indications*! The Straumann Dental Implant System offers you and your patients efficient and esthetic implant-borne
Ideal treatment of the impaired
RESEARCH IMPLANTS AS ANCHORAGE IN ORTHODONTICS: ACLINICAL CASE REPORT Dale B. Herrero, DDS KEY WORDS External anchorage Pneumatized Often, in dental reconstruction, orthodontics is required for either
A Study of 25 Zygomatic Dental Implants with 11 to 49 Months Follow-up After Loading
A Study of 25 Zygomatic Dental Implants with 11 to 49 Months Follow-up After Loading Fredrik Ahlgren, DDS, MSc 1 / Kjell Størksen, DDS 2 /Knut Tornes, DDS, PhD 3 Purpose: The purpose of this study was
Photoelastic Comparison of Single Tooth Implant-Abutment-Bone of Platform Switching vs Conventional Implant Designs
10.5005/jp-journals-10024-1021 Fabiana Rossi et al ORIGINAL RESEARCH Photoelastic Comparison of Single Tooth Implant-Abutment-Bone of Platform Switching vs Conventional Implant Designs Fabiana Rossi, Adriana
Three-Dimensional Analysis Using Finite Element Method of Anterior Teeth Inclination and Center of Resistance Location
Three-Dimensional Analysis Using Finite Element Method of Anterior Teeth Inclination and Center of Resistance Location Allahyar GERAMY 1, Ahmad SODAGAR 1, Mehdi HASSANPOUR 1 Objective: To locate the centre
Implant Protected Occlusion
IOSR Journal of Dental and Medical Sciences (IOSR-JDMS) e-issn: 2279-0853, p-issn: 2279-0861.Volume 11, Issue 3 (Nov.- Dec. 2013), PP 20-25 Implant Protected Occlusion Yogeshwari Swaminathan¹, Gururaj
The SATURN implant by Cortex Dental Industries
The SATURN implant by Cortex Dental Industries By Dr. Zvi Laster DMD W e P r o v e I t E v e r y D a y A case report using a newly designed implant specifically designed for immediate post-extraction loading
Vertical augmentation using the mandibular body bone for repair of failed implants: a case report
Vol. 34 No. 2, September 2015 Vertical augmentation using the mandibular body bone for repair of failed implants: a case report Da-Nee Jeon, Han-Kyul Park, Yong-Deok Kim* Department of Oral and Maxillofacial
Teeth in a Day for the Maxilla and Mandible: Case Report
Teeth in a Day for the Maxilla and Mandible: Case Report Thomas J. Balshi, DDS, FACP* Glenn J. Wolfinger, DMD, FACP** Abstract Background: A growing body of evidence indicates that successful osseointegration
Electronic Thesis and Dissertations UCLA
Electronic Thesis and Dissertations UCLA Peer Reviewed Title: The Predictive Value of Resonance Frequency Analysis Measurements in the Surgical Placement and Loading of Endosseous Implants Author: Baltayan,
Post-extraction immediate implant placement in the maxillary anterior region: A case report
Merit Research Journal of Medicine and Medical Sciences (ISSN: 2354-3238) Vol. 3(1) pp. 001-009, January, 2015 Available online http://www.meritresearchjournals.org/mms/index.htm Copyright 2015 Merit Research
In vitro evaluation of the influence of the cortical bone on the primary stability of two implant systems
Journal section: Oral Surgery Publication Types: Research In vitro evaluation of the influence of the cortical bone on the primary stability of two implant systems Rodrigo Andrés-García 1, Nuria García
All-on-4 treatment concept with NobelSpeedy Groovy
All-on-4 treatment concept with NobelSpeedy Groovy Product overview Immediate Function for high patient satisfaction Immediately loaded fixed provisional prosthesis on the day of surgery. Immediate improvement
THE SINGLE-TOOTH IMPLANT TREATMENT FOR MAXILLARY CENTRAL INCISORS LOSS AFTER TRAUMA: CASE REPORTS
CLINICAL DENTISTRY AND RESEARCH 2011; 35(1): 47-52 THE SINGLE-TOOTH IMPLANT TREATMENT FOR MAXILLARY CENTRAL INCISORS LOSS AFTER TRAUMA: CASE REPORTS Hakan Tuna, DDS, PhD Assistant Professor, Department
More than an implant. A sense of trust. Straumann Dental Implant System
More than an implant. A sense of trust. Straumann Dental Implant System More than an implant. A sense of trust. Patients today are more aware of their options when it comes to tooth replacement. Partner
Clinical Effectiveness of Implant-Supported Removable Partial Dentures A Review of the Literature and Retrospective Case Evaluation
J Oral Maxillofac Surg 67:1941-1946, 2009 Clinical Effectiveness of Implant-Supported Removable Partial Dentures A Review of the Literature and Retrospective Case Evaluation Yoav Grossmann, DMD, MsHA,*
Immediate Molar Implant Placement: A Private Practice Clinical Investigation. Abstract
Immediate Molar Implant Placement: A Private Practice Clinical Investigation Gargiulo et al Alphonse Gargiulo, DDS, MS 1 Thomas Manos, DDS, MS 2 Mark Kolozenski, DDS, MS 3 Alex Tzanos, DDS, MSD 3 Michael
Analysis of the 619 Brånemark System TiUnite Implants: A Retrospective Study
Kobe J. Med. Sci., Vol. 58, No. 1, pp. E19-E28, 2012 Analysis of the 619 Brånemark System TiUnite Implants: A Retrospective Study YASUYUKI SHIBUYA, DDS, PhD* 1, NAOKI TAKATA, DDS, PhD 1, JUNICHIRO TAKEUCHI,
Augmentation in Proximity to the Incisive Foramen to Allow Placement of Endosseous Implants: A Case Series
DENTAL IMPLANTS J Oral Maxillofac Surg 68:2267-2271, 2010 Augmentation in Proximity to the Incisive Foramen to Allow Placement of Endosseous Implants: A Case Series Gerry M. Raghoebar, DDS, MD, PhD,* Laurens
Dental Updates. Excerpted Article e-mail: [email protected]. Why Implant Screws Loosen Part 1. Richard Erickson, MS, DDS
¼ ½ ¾ µ mw/cm 2 Volume 17; 2007 Dental Updates "CUTTING EDGE INFORMATION FOR THE DENTAL PROFESSIONAL " 200 SEMINARS AND 30 JOURNALS REVIEWED YEARLY FOR THE LATEST, CUTTING EDGE INFORMATION Excerpted Article
Short implants with single-unit restorations in posterior regions with reduced height a retrospective study
Original Article Braz J Oral Sci. October December 2010 - Volume 9, Number 4 Short implants with single-unit restorations in posterior regions with reduced height a retrospective study Jamal Hassan Assaf
Choosing Between Screw-Retained and Cement-Retained Implant Crowns
Choosing Between Screw-Retained and Cement-Retained Implant Crowns Go online for in-depth content by William F. Campbell, DDS, FAGD and Marc W. Herman, DDS, FAGD The choice of a screw-retained versus a
Replacing Hopeless Retained Deciduous Teeth in Adults Utilizing Dental Implants: Concepts and Case Presentation
Replacing Hopeless Retained Deciduous Teeth in Adults Utilizing Dental Implants: Concepts and Case Presentation by Michael Tischler, DDS Published: Dentistry Today November 2005 Photos at end of article
1 The Single Tooth Implant. The Ultimate Aesthetic Challenge
1 The Single Tooth Implant The Ultimate Aesthetic Challenge by Daniel G. Pompa, D.D.S. 2 Before starting any Maxillary Anterior Single Implant, or any case in the esthetic zone: TAKE A PHOTO OF YOUR PATIENT
Titanium endosseous implants have
Smoking and Complications of Endosseous Dental Implants* Devorah Schwartz-Arad, Naama Samet, Nachum Samet, and Avi Mamlider Background: The purpose of this study was to compare the incidence of the complications
Luigi Vito Stefanelli,D.Eng.,DDS
LYON, 27-28/09/2013 Luigi Vito Stefanelli,D.Eng.,DDS Assistant Researcher, Department of Oral Sciences, Sapienza University of Rome, Italy, School of Dentistry. Private practitioner dental office Stasi
RESTORING STRAUMANN IMPLANTS WITH LOCATOR ABUTMENTS
RESTORING STRAUMANN IMPLANTS WITH LOCATOR ABUTMENTS Straumann is the industrial partner of the ITI (International Team for Implantology) in the areas of research, development, and education. CONTENTS Product
ALL-CERAMIC DENTAL IMPLANT SOLUTIONS
ALL-CERAMIC DENTAL IMPLANT SOLUTIONS Scientific Evidence Bone-to-implant contact of 78% at 3 months. One piece implant = no prosthetic connections References 1. One-year follow-up of first consecutive
Oftentimes, as implant surgeons, we are
CLINICAL AVOIDING INJURY TO THE INFERIOR ALVEOLAR NERVE BY ROUTINE USE OF INTRAOPERATIVE RADIOGRAPHS DURING IMPLANT PLACEMENT Jeffrey Burstein, DDS, MD; Chris Mastin, DMD; Bach Le, DDS, MD Injury to the